CARBON MONOXIDE IN WORKPLACE ATMOSPHERES (DIRECT-READING
MONITOR)
Method Number: |
ID-209 |
|
Matrix: |
Air |
|
OSHA Permissible Exposure Limits
(PEL) Final Rule*: |
35 ppm Time Weighted Average (TWA) 200 ppm
Ceiling (5-min sample) 1,500 ppm Instantaneous* |
|
Transitional*: |
50 ppm TWA |
|
Monitoring Device: |
Workplace areas are monitored using a
direct-reading passive sampling instrument capable of
recording data at given time intervals. Data can be retrieved after
sampling. |
|
Recommended Sampling Times: TWA
Determination: Ceiling Determination: |
8 h 5 min |
|
Analytical Procedure: |
Direct-reading instrument - Datalogger. Data are
transferred to a computer, exposures are calculated, and results are
stored and/or sent to a printer for a hard copy. |
|
Detection Limits (Instrument TWA read-out -
also see Section 1.5.2.) |
Qualitative: Quantitative: |
1.2 ppm 4.1 ppm |
|
Precision and Accuracy |
|
CV2
(Pooled): |
|
Bias: |
|
Overall Error: |
|
Validation Ranges: |
16.1 to 70.2 ppm |
|
197.8
ppm | |
|
Special Requirements: |
IBM-compatible computer |
|
Method Classification: |
Validated method (Direct Reading Instrument) |
|
Chemist: |
Robert G. Adler |
|
Date: |
March, 1993 |
* This method was evaluated using the Final Rule
Limits stated above; the Transitional Limit of 50 ppm is within tested
ranges and the method should perform well when assessing compliance at
this concentration. Compliance with the 1,500 ppm Instantaneous Exposure
Limit (same as the Immediately Dangerous to Life and Health (IDLH)] cannot
be assessed using the specific equipment described in this method. Please
consult The OSHA Chemical Information File and Section 2.
of this method regarding instantaneous carbon monoxide exposure
determinations.
Commercial manufacturers and products mentioned in this method are
for descriptive use only and do not constitute endorsements by USDOL-OSHA.
Although the following Instrument Description, Backup Data, and Auxiliary
Sampling (Appendix
B) Sections of this report describe use of a specific carbon monoxide
monitor, similar monitors can be substituted provided they meet validation
requirements.
Branch of Inorganic Methods Development OSHA Salt Lake
Technical Center Salt Lake City, Utah
1. Introduction
1.1. History - Monitoring for Carbon Monoxide
The recent change in the OSHA Time Weighted Average (TWA)
Permissible Exposure Limit (PEL) for CO from 50 to 35 ppm (5.1.),
the inclusion of a Ceiling PEL of 200 ppm (5-min sampling period) (5.2.),
and the addition of a maximum Instantaneous limit of 1,500 ppm (5.3.)
stimulated a review of the methods used for the analysis of CO in
workplace atmospheres. This review included both
direct-reading monitoring procedures and classical
collection procedures.
The previous OSHA sampling and analytical method for CO required
the use of direct-reading procedures for monitoring (5.4.).
One direct-reading procedure involved the use of CO
short-term detector tubes (5.5.).
An evaluation at the OSHA Salt Lake Technical Center (OSHA-SLTC) was
conducted for several brands of these tubes (5.6.).
Short-term detector tubes offer only spot checks of the environment,
and sampling procedures capable of determining long-term CO
concentrations are preferred. A long-term detector tube for CO was
evaluated by OSHA and was determined to have significant lot-to-lot
variability, and results were affected by sampling temperature (5.7.).
For TWA compliance measurements, a long-term
direct-reading method was performed by OSHA compliance
officers using an electrochemical detector (Ecolyzer, Energetics
Science, Inc., Elmsford, NY). However, this instrument required
constant calibration, readings were subject to drift, and TWA
determinations were difficult to calculate. Due to the size of the
instrument, only area samples were normally taken. Personal monitoring
could be conducted by taking air samples in gas sampling bags and then
attaching them to the Ecolyzer at the end of the sampling period. Air
samples collected in gas bags or canisters have also been analyzed for
CO using infrared absorption spectrophotometry (5.8.),
electrochemical means (5.9.),
gas chromatography (GC) with a flame ionization detector (FID) (5.10.),
or GC with a helium glow discharge ionization detector (DID) (5.11.).
A recently developed OSHA Method ID-210
(5.12.)
involves the use of gas bags followed by analysis by GC-DID. This is a
reliable and sensitive method; however, it has the disadvantage of
requiring sample shipment for laboratory analysis, and storage
stability of the collected samples has to be addressed.
Recent developments in direct-reading instrumentation
provide increased ease in personal monitoring of workplace
atmospheres. These instruments are advertised as accurate and precise,
and capable of storing data points for specified sampling periods
(each period may be as small as 1 min). One such monitor is the
Draeger Model 190 CO Datalogger (National Draeger, Inc., Pittsburgh,
PA), which was the instrument evaluated for the current method. Other
instruments may be substituted for monitoring CO provided they meet
validation requirements.
1.2. Principle
Samples for monitoring exposures in regards to the TWA and
5-min Ceiling for CO are taken using a portable
direct-reading instrument. A datalogger system can be
used to reduce data to a TWA or 5-min Ceiling value.
Other labor intensive methods such as hand-calculated
averaging of readings, strip-chart-recordings, etc. can
also be employed to reduce data if necessary.
1.3. Instrument Description and Special Features
or Considerations
(Note: |
The Draeger Model 190 Datalogger is discussed
below. Other commercially available monitors for CO may have
similar characteristics.) |
1.3.1. The Draeger Model 190 CO Datalogger consists of a
patented 3-electrode diffusional electrochemical sensor with a
removable filter cap, a liquid crystal display (LCD) for providing
direct-reading information, and a memory for storing
the data. Logging of data may be started and ended by use of
function keys (red and blue) provided. Sampled data are stored in
1-min intervals for a specified sampling time. The data
can later be transferred to a computer or printer (5.13.).
1.3.2. A CO-specific filter designed to reduce or eliminate any
air contaminants which might interfere with the CO determination is
available. This removable filter cap also serves as a dust filter. A
plain dust filter can be attached to the sensor if the monitor is
used where interfering contaminants are not present.
1.3.3. The operating temperature range is 0 to 40°C. The optimum
range in which minimal temperature effect occurs is 10 to 30°C.
1.3.4. Recording of data:
- When the sensor is in operation, measured values are sent to
the microprocessor twice each second. The values (120 sequential
readings) are then computed into a 1-min average and
this value is stored in memory. When data are transferred to a
printer, the 1-min averages can be listed and/or
graphically plotted. The sensor has a response time of
approximately 1 min for large changes in concentration.
- The 1-min averages are truncated to integer ppm
CO values (i.e., for 6.9 ppm CO the Datalogger would give a value
of 6 ppm). The TWA value reported by the instrument is the average
of all truncated 1-min averages, and is also
truncated to an integer ppm value. The truncated values will
always be < the hand-calculated values.
Note: |
Hand-calculated averages were
also obtained during this study; in these cases, all
truncated 1-min averages were manually averaged
and rounded to the nearest 0.1 ppm
value. |
- The instrument also notes the highest 1-min
average over the sampling period, and reports this value as the
peak concentration. The specific time that this peak occurs is
indicated either graphically or in a sampling summary. Where
duplicate highest 1-min averages occur, only the
first occurrence is listed. Monitoring for the OSHA Ceiling PEL of
200 ppm CO requires the user to perform a
hand-calculated average of 1-min
readings over a 5-min sampling period. The peak
concentration function of the Datalogger cannot be used in this
instance because it reports out only the highest
1-min reading.
1.3.5. The instrument is not capable of measuring the
Instantaneous Exposure Limit of 1,500 ppm due to the instrument
range limitation of 0 to 999 ppm CO.
1.3.6. Calibration for routine monitoring should be performed
near the manufacturer-recommended flow rate of 0.20
L/min. Higher flow rates over the monitor sensor during calibration
result in higher monitor readings, possibly because of a slight
pressure increase at the face of the sensor under these conditions.
Under normal workplace monitoring conditions, the recommended
calibration procedure should give satisfactory results.
1.4. Advantages and Disadvantages
1.4.1. Advantages
- According to the manufacturer, the instrument is specific for
CO when the Datalogger is equipped with a CO-specific filter (P/N
4510184). Interferences besides those mentioned by the
manufacturer were discovered during the evaluation. See Section 1.6.
for further details.
- The monitor is equipped with an audible alarm for immediate
warning of dangerous CO levels.
- The monitor can be used to measure compliance with Indoor Air
Quality (IAQ) Standards for CO [9 ppm (8 h), 35 ppm (1 h)] (5.14.).
- The monitor is easy to set up and use and has a rapid response
time (~1 min).
- Data can be retrieved at the convenience of the industrial
hygienist.
- The monitor normally will not interfere with the worker's
activities and is intrinsically safe if not altered.
- Changes in humidity do not affect sample data collection.
- The monitors permit measurement of the fluctuations in the CO
concentration throughout the sampling period. Classical sample
collection techniques will give only a TWA concentration unless
multiple samples are taken.
- The TWA concentration for the sampling period is calculated by
the monitor. Peak concentrations are also indicated; however, a
5-min hand-calculated average is
necessary when comparing output to the OSHA Ceiling PEL for CO.
- Sample storage stability problems are not encountered.
1.4.2. Disadvantages regarding equipment required or used for
this procedure are:
- Without a CO-specific filter, significant interferences can be
encountered from ethylene, acetylene, hydrogen, hydrogen sulfide,
nitrogen dioxide, nitric oxide, and sulfur dioxide. See Section 1.6.
for other interferences.
- For the initial purchase, the monitors are relatively
expensive.
- Additional equipment is needed for calibration. For accurate
calibration, the calibration gas flow must be maintained at the
manufacturer's recommended rate (0.20 L/min). For OSHA compliance,
monitors are calibrated each time they are used.
- Recovery of the data requires transferring to a computer or
printer; sample results can be easily lost if the transfer is
performed incorrectly. Data can also be lost if battery voltage
drops significantly before the data can be transferred. A 9-volt
battery is used to operate the instrument.
- Samples taken in areas of high air velocities directed toward
the sensor may result in a positive bias to the CO concentrations
read.
- Because the instrument automatically performs integer math for
calculations, truncation error is possible for TWA values reported
by the instrument, especially at single digit levels (i.e., <10
ppm). Optimal performance can be achieved if a
hand-calculated average is obtained for all of the
1-min values reported for a sampling period; however,
this is a very tedious process and is not recommended for routine
applications. The comparability of the precision and accuracy
results in this study for the two calculation procedures indicates
that, for CO levels near the TWA PEL, the extra calculation work
is not necessary.
1.5. Method Performance (See Section 4
for more detailed information)
Note: |
During the evaluation, six instruments were used
for each experiment. Imprecision was likely greater than if only
one instrument were used and the experiment repeated six times;
however, the statistical values obtained will be more
representative of OSHA field office use. Many OSHA field offices
are supplied with five or more monitors and an industrial
hygienist may randomly take one or more instruments into the
field for monitoring. As previously discussed in Section 1.3.4.,
both truncated and hand-calculated averages were
used to determine instrument performance. The
hand-calculated average only appears beneficial if
detection limit is a significant factor as shown
below. |
1.5.1. Range
The upper analytical range used during the TWA PEL studies was
about 70 ppm. The Ceiling study was conducted at about 200 ppm. The
overall range according to the manufacturer is 0-999
ppm CO (5.13.).
1.5.2. Precision, Accuracy, Detection Limits
Values determined during the validation are listed below. For
further information see Section 4.
Statistical Data
|
Truncated
|
Hand-Calculated
|
|
Range* |
|
|
% Mean Recovery |
96.1 - 112.4 |
96.6 - 114.0 |
CV (Individual experiments) |
0.024 - 0.058 |
0.021 - 0.056 |
Bias |
-0.016 - 0.062 |
-0.007 - 0.075 |
CV2
(Pooled) |
0.040 - 0.048 |
0.032 - 0.045 |
* Includes different RHs and
concns - 30 to 80% RH, 16.1 to 70.2 ppm
|
Detection Limits |
|
|
Qualitative: |
1.2 ppm |
0.5 ppm |
Quantitative: |
4.1 ppm |
1.8 ppm |
|
Precision and Accuracy** |
|
|
CV2
(Pooled): |
0.040 |
0.041 |
Bias: |
-0.016 |
-0.007 |
Overall Error: |
±9.6% |
±8.8% |
** Only data for 0.5 to 2 times
PEL, 50% RH is listed |
1.6. Interferences
Note: |
The following paragraphs list specific
interferences associated with the Draeger Datalogger. If other
types of instruments are used, always consult with the
manufacturer or manufacturer-provided instructional
manuals concerning potential interferences before general use.
Additional information regarding interferences concerning the
Draeger Datalogger can be found in reference 5.13. |
The following interferences occur with the CO-specific filter
attached to the monitor. Acetylene and hydrogen have small positive
interferences (97 ppm acetylene will indicate 10 ppm CO, 1,000 ppm
hydrogen will indicate 40 ppm CO ) (5.13.).
During the evaluation of the Draeger Datalogger, it was noted high
concentrations of the lower alcohols produced large positive biases
when the monitor was exposed to headspace concentrations. Dataloggers
were also placed near the open tops of glass bottles containing the
chemicals listed below and monitor response was noted:
Chemical
|
Response
|
|
Chemical
|
Response
|
|
|
|
acetonitrile |
no response |
ethanol |
positive |
|
acetone |
no response* |
isobutanol |
positive |
|
butylamine |
no response |
isopropanol |
positive |
|
dimethylamine |
no response |
methanol |
positive |
|
ethyl acetate |
no response |
n-propanol |
positive |
|
formaldehyde |
no response** |
|
|
|
isooctane |
no response |
|
|
|
1-octanol |
no response |
|
|
|
toluene |
no response |
|
* |
Acetone gave a positive response when the
CO-specific filter was removed or when the filter had
diminished capacity (been in use over 1 year) |
** |
A 37% solution of formaldehyde gave a
response which was primarily due to the 10 to 15% methanol
used as a stabilizer. When formaldehyde was produced by
heating paraformaldehyde (no methanol present), a response
was not noted when the CO-specific filter was new. A
positive response was noted to the formaldehyde vapors
when the filter was removed or had diminished
capacity. | |
The responses indicated lower alcohols gave positive responses and
others structurally similar to CO (i.e. acetone) could give a response
if the CO-specific filter is removed or the filter has diminished
capacity. Placement of the monitors into atmospheres saturated with
each substance giving a positive response gave very large false
positive readings for CO and also tended to fatigue the sensor. In
addition, fine mists or droplets of the positive-interfering
substances will produce similar false positive results. These
interferences were produced using near-saturated or
saturated atmospheres and should not be significant at concentrations
near the PELs for these compounds. Concentration estimates for those
substances giving positive responses at saturated vapor conditions are
shown in Appendix
A (Note: These are only rough approximations). The industrial
hygienist should note whenever there is a significant amount of these
compounds present (i.e. some confined space monitoring may have
near-saturated or saturated atmospheres), or
operations where droplets or fine mists of these substances could
contact the surface of the filter/sensor of the Datalogger. If
large amounts or aerosols of these substances are expected,
monitoring for the interfering substances should take precedence, and
an alternate method such as OSHA Method No. ID-210
can be used for CO monitoring if necessary. It is highly
recommended that the CO-specific filters be changed
periodically.
Early reports from other Datalogger users indicated a positive bias
was observed when the monitor was placed near liquid aerosol streams.
To assess if small liquid droplets could trigger a positive reading,
aerosols of water or ethyl acetate were sprayed directly above the
monitor sensor and produced no response.
1.7. Physical Properties (5.15.,
5.16.)
and CAS No.
CAS No. (Carbon Monoxide)
|
630-08-0 |
|
|
Molecular weight |
28.01 |
Molecular formula |
CO |
Appearance |
Colorless, odorless gas |
Explosive limits in air |
12.5 to 74.2% (v/v) |
Autoignition temperature |
651°C |
Melting point |
-207°C |
Boiling point |
-191.3°C |
Specific gravity (air = 1) |
0.968 |
Density, gas at 0°C, 101.3 kPa
(760 mmHg) |
1.25 g/L |
Solubility at
0°C at 25°C |
3.54 mL/100 mL water 2.14 mL/100 mL
water |
1.8. Prevalence and Use of CO
With the single exception of CO2, the
total yearly emissions of CO exceed all other atmospheric pollutants
combined (5.16.).
Potential sources for CO emission and exposure are listed (5.16.,
5.17.):
Foundries Petroleum
refineries Fluid catalytic
crackers Fluid coking
operations Moving-bed catalytic crackers Kraft
pulp mills Carbon black manufacturers Steel
mills Coke ovens Basic oxygen
furnaces Sintering operations |
Formaldehyde
manufacturers Coal combustion
facilities Utility and large industrial
boilers Commercial and domestic furnaces Fuel
oil combustion operations Power
plants Industrial, commercial, and domestic
uses Charcoal manufacturers Meat smokehouses Sugarcane
processing operations Motor
vehicles |
1.9. Toxicology
Note: |
Information contained within this section is a
synopsis of the present knowledge of the physiological effects
of CO and is not necessarily intended to be used as the basis
for OSHA policy. |
Carbon monoxide has over a 200-fold greater affinity for hemoglobin
than has oxygen (5.18.,
5.19.).
Thus, it can make hemoglobin incapable of carrying oxygen to the
tissues. The presence of CO-hemoglobin (COHb) interferes
with the dissociation of the remaining oxyhemoglobin, further
depriving the tissues of oxygen (5.15.,
5.16.).
The signs and symptoms of CO poisoning include headache, nausea,
weakness, dizziness, mental confusion, hallucinations, cyanosis, and
depression of the S-T segment of an electrocardiogram. Although most
injuries in survivors of CO poisoning occur to the central nervous
system, it is likely that myocardial ischemia is the cause for many
CO-induced deaths (5.18.).
The uptake rate of CO by blood when air containing CO is breathed
increases from 3 to 6 times between rest and heavy work. The uptake
rate is also influenced by oxygen partial pressure and altitude (5.20.).
Carbon monoxide can be removed through the lungs when CO-free air
is breathed, with generally half of the CO being removed in 1 hour.
Breathing of 100% oxygen removes CO quickly.
Acute poisoning from brief exposure to high concentrations rarely
leads to permanent disability if recovery occurs. Chronic effects from
repeated exposure to lower concentrations have been reported. These
include visual and auditory disturbances and heart irregularities.
Where poisoning has been long and severe, long-lasting mental and/or
nerve damage has resulted (5.15.).
The following table gives the levels of COHb in the blood which
tend to form at equilibrium with various concentrations of CO in the
air and the clinical effects observed (5.21.):
Atmospheric CO (ppm) |
COHb in Blood (%) |
Symptoms |
|
70 |
10 |
Shortness of breath upon vigorous exertion;
possible tightness across the forehead. |
120 |
20 |
Shortness of breath with moderate exertion;
occasional headache with throbbing in the temples. |
220 |
30 |
Decided headache; irritability; easy
fatiguability; disturbed judgment; possible dizziness; dimness
of vision. |
350-520 |
40-50 |
Headache; confusion; collapse; fainting upon
exertion. |
800-1220 |
60-70 |
Unconsciousness; intermittent convulsions;
respiratory failure; death if exposure is prolonged. |
1950 |
80 |
Rapidly fatal. |
|
Adults (non-smokers) normally have about 1% COHb in the body.
Cigarette smokers generally have blood levels of 2 to 10% COHb (5.20.).
In examining the CO levels in an occupational environment, CO
generated from tobacco smoking may need to be considered. These
amounts may ordinarily be small, but when added to the amounts
generated by occupational activities, may aggravate conditions from an
already existing high concentration of CO (5.22.,
5.23.).
1.10. Other Hazardous Properties
Carbon monoxide is flammable and can be a dangerous fire and
explosion risk. The flammable limits in air range from 12 to 75% by
volume (5.19.).
2. Sampling
| Note: |
A training videotape (Draeger part no. 4505202) is
available from Draeger regarding Datalogger 190 use and data
transfer. |
2.1. Precautions
2.1.1. Be certain that the radio frequency shielding of the
monitor is intact.
2.1.2. Attach the monitor to the worker in such a manner that it
will not interfere with work performance or safety.
2.1.3. Follow all safety practices that apply to the work area
being sampled.
2.1.4. If the employee being monitored is smoking a
tobacco product during sampling, a positive contribution of CO from
the tobacco combustion may occur for personal samples. Ask the
employee to refrain from smoking during sampling so that only the
occupational exposure is considered.
2.1.5. Describe the audible alarms to the employee, and instruct
the employee on what to do if they occur. The alarm is normally set
to a level slightly above the OSHA Ceiling Exposure Limit of 200 ppm
CO.
Note: |
The Immediately Dangerous to Life and Health
(IDLH) and the OSHA Instantaneous Limit are the same
concentration, 1,500 ppm CO. The Datalogger discussed in this
method is unable to measure above 999 ppm. If the Datalogger
registers readings above 999 ppm, appropriate protective
measures should be taken to safeguard those exposed. If
necessary to document excursions above the IDLH or OSHA
Instantaneous Limit, detector tubes listed in the OSHA
Chemical Information File can be used provided measurements
can be safely taken. |
2.2. Equipment
2.2.1. CO Datalogger or other validated
direct-reading device (For information on specific
equipment needed with the Draeger Model 190 CO Datalogger, see Appendix
B. For interferences, please see Section 1.6.)
2.2.2. Gases for calibration:
- CO calibration gas
- CO-free nitrogen or CO-free air
2.3. Sampling Procedure
2.3.1. See Appendix
B regarding specific information on preparation, alarm
adjustment, calibration, and other procedures for the Draeger Model
190 Datalogger.
2.3.2. Calibration - Use only certified calibration gases. For
OSHA purposes, the recommended concentration of the calibration gas
(span gas) is 30 to 70 ppm CO (listed to the nearest 0.1 ppm). Gas
cylinders are available from the OCL (see Appendix
C)(5.24.).
The gas cylinders can be mailed or expressed to the site from OCL
using appropriate shipping procedures.
Notes: |
- Do not perform calibrations in the presence of an
operating radio transmitter. Calibrate the monitors in a
clean, well-ventilated environment; if possible, in a
well-ventilated exhaust hood. Pre- and post-calibrate the
monitors each time measurements are taken.
- In order to eliminate possible effects on calibration
due to variations in altitude (pressure difference),
calibrate the monitors at the same altitude at which they
are to be used. Altitude corrections (calibrating at one
altitude and taking measurements at another) were not
investigated in this evaluation.
- Calibrate the monitors using the
manufacturer-recommended flow rate of 0.20
L/min. Samples taken in areas of high air velocities
(directed toward the monitor sensor) may indicate high CO
readings. If necessary, gas bag samples can also be taken
and submitted for laboratory analysis [use OSHA Method No.
ID-210
(5.12.)].
Air velocity does not have a significant effect on gas bag
sample results.
- If possible, calibrate the monitors in the field. The
effects of vibration during shipping have not been
investigated.
|
2.3.3. General instructions - Sampling with a
direct-reading instrument:
- Start the monitor sampling. Record the time at the start of
sampling.
- Place the monitor in the breathing zone of the user,
preferably secured or attached in a breast pocket. Be sure the
sensor is not obstructed.
- Sample for the time indicated:
- TWA Determination: 8 h (if possible)
- Ceiling Determination: 5 min - If datalogging is available
in 5-min increments or less, a full-shift (8 h)
sample can be taken and any overexposure to Ceilings can be
determined as discussed below (Section 3.2.).
- Terminate sampling.
3. Data Recovery
3.1. Refer to instrument manuals and operating procedures for
proper operation of all instruments. See Appendix
B for information on the data recovery procedures for the Draeger
Model 190 Datalogger.
3.2. Calculate Ceiling exposures if they are not
calculated in 5-min increments by the instrument. Most
commercially available instruments will measure in 1-min
increments. Hand-calculate an average of 1-min segments
for each 5-min Ceiling exposure. The 1-min
increments must be consecutive.
3.3. Generate hard copies of the results. Show a display of the
results and a graph.
3.4. Report the results as ppm CO.
4. Backup Report
Experimental Protocol
The evaluation of the instrument consisted of the following
experimental protocol:
- Analysis of the sample data from six Dataloggers exposed to
dynamically generated CO atmospheres having concentration ranges of
approximately 0.5, 1, and 2 × TWA PEL.
- Evaluation at levels commonly encountered during IAQ
investigations (approximately 5 and 10 ppm CO).
- Evaluation at levels near the Ceiling PEL (approximately 200 ppm
CO).
- Determination of any variation in results when sampling at low and
high humidity levels.
- Assessment of any significant effect from varying the air flow
rate over the monitor sensor during the evaluation tests.
- Determination of the qualitative and quantitative detection limits
for the analysis of CO.
- Comparison of Datalogger results with a GC method (5.12.)
used for CO determinations in which the CO was analyzed using a
GC-DID.
- Assessment of the performance of this method and conclusions.
All results were statistically examined for outliers and, when
necessary to pool results, homogeneous variance. Possible outliers were
determined using the American Society for Testing and Materials (ASTM)
test for outliers (5.25.).
Homogeneity of the coefficients of variation was determined using the
Bartlett's test (5.26.).
Overall Error was calculated as:
OEi = ±(|mean
biasi| + 2CVi) ×
100%
where i is the respective instrument pool being examined (5.27.).
Six Dataloggers were used, and the statistical values for each set of
determinations are based on the analysis of data consisting of one
result from each of the six instruments.
This evaluation was split into two blocks of time [early work ("A"
experiments) and later work ("B" experiments)].
4.1. Analysis - Because of the nature of the monitor and the data
collection procedure, the sample spiking procedure normally performed
during the evaluation of a sampling and analytical method was not
possible; therefore, this experiment (analytical recovery,
CV1) is not relevant to this
direct-reading monitor.
4.2. Sampling and Analysis
Procedure: Three sets of samples (at 0.5, 1, and 2 × TWA
PEL) were collected and evaluated. Samples taken at 5 and 10 ppm CO
(50% RH) were also prepared and analyzed. Six Dataloggers were
evaluated at each of these sampling conditions. Samples were collected
according to the procedure listed below.
Block Diagram of the Laboratory Generation
System
Figure 1a
4.2.1. A dynamic gas generation system was
assembled as shown in Figure 1a.
Moisture and other contaminants were removed from the diluent air by
using a charcoal/Drierite/silica gel filtering system. A hopcalite
filtering system was also used in part of the later work ("B"
experiments) in an attempt to reduce any background CO
concentration. A humidity, temperature, and flow control system
(Model HCS-301, Miller-Nelson Research Inc., Monterey, CA) was used
to treat the diluent air to produce the stated RH at 25°C. Diluent
air flow was measured before and after each experiment using a dry
test meter (DTM-115, American Meter Co., Philadelphia, PA). The flow
control system had earlier been calibrated in-house for temperature
and humidity. The Dataloggers were connected to the sampling
manifold of the generation system using connectors (Figure 1b)
similar to the one used for calibration. The free end of the
Tygon® tubing was attached to a sampling
manifold exit port (Teflon® tubing) for
each Datalogger.
Figure 1b
For connectors, short pieces of rubber tubing slipped over rubber
stoppers were drilled in the sides to allow test atmospheres moving
down the Tygon® tubing to freely exit. The
connectors were then used to attach the Datalogger sensor and Tygon
tubing together. These connectors had slightly larger exit holes
than the calibration units shown in Appendix
B (Figure 8)
to allow the generated atmosphere to freely pass over the sensor
during testing. This type of sampling connection allowed the linear
face velocity to be altered during testing. Passive monitor test
chambers were used in additional experiments (Sections 4.4.
and 4.6.)
without the use of connectors.
4.2.2. A gas cylinder containing approximately 0.5% CO in
nitrogen was used as the CO source. For the "A" experiments, the gas
was from Linde Div., Union Carbide Corp., Denver, CO (concentration
certified at 0.50%). For the "B" experiments, the gas was from
Airco, Riverton, NJ (certified at 0.496%). The CO was introduced
into the flow system via a glass mixing chamber. Gas flow rates were
controlled using a mass flow controller (Model FC-261, Tylan Corp.,
Torrance, CA). Gas flows were measured immediately before and after
each experiment using a soap bubble flowmeter (Model 823-1, Mast
Development Co., Davenport, IA).
4.2.3. To assure a continuous generation of controlled
concentrations of CO and provide an additional verification of
concentrations, the flow system was continuously monitored during
each test by noting the LCD readings on the Dataloggers and, during
the "A" experiments, with a direct-reading instrument
[Model 7140 (for CO), Interscan Corp., Chatsworth, CA] connected to
the flow system. Calibration of the Interscan
direct-reading instrument was performed with a
40-ppm calibrating gas (CO in air, certified, Alphagaz,
Cambridge, MD).
4.2.4. Calibration of the Dataloggers during
the "A" experiments was performed with the 40-ppm
calibrating gas mentioned above. Calibration during the "B"
experiments was done using 104 ppm CO in nitrogen (certified, Airco,
Inc., Murray Hill, NJ). The CO content of this cylinder had been
confirmed in a previous study (5.12.).
For the "B" experiment calibrations, the gas flow over the face of
each Datalogger was regulated with a regulator-rotameter system and
the flow rates were measured using a soap bubble flowmeter (Model
M-5, A. P. Buck, Inc., Orlando, FL).
Note: |
Preliminary experiments and a personal
communication with the manufacturer indicated that the
concentration of the calibration gas did not appear to have a
significant effect on precision and accuracy as long as the
calibration concentration was in the vicinity of the PEL. The
manufacturer did state the higher calibration concentration
(104 ppm CO) would probably give slightly more accurate
results because of the greater linear
range. |
4.2.5. During the course of each "A" experiment, the monitors
were calibrated and then attached to the generation system. The
tempered CO/air mixture from the system was permitted to flow over
the sensors and datalogging was initiated. At the conclusion of the
experiment the logging was stopped and the monitors were removed
from the system. The collected data were then downloaded to a
computer.
For the "B" experiments a calibration was performed, then
readings from the monitor LCDs were taken with only air flowing
through the system. The CO flow was then started and, after
stabilizing, datalogging was begun. At the conclusion of the
experiment, datalogging was stopped, CO gas flow was stopped,
monitor readings of the generation system air were again obtained,
and then the monitors were removed from the system. (Note: A few
problems occurred during the generations and are further discussed
following the Results Section.)
Results: The results for generated samples at 0.5, 1, and
2 × PEL (50% RH) based on the TWA output values are listed in Table
1a.
These represent Datalogger outputs [i.e., results are rounded down
(truncated) to the next lower integer ppm value by the Datalogger].
The results for the generated samples based on the actual rounded
hand-calculated average of all 1-min ppm
values collected are listed in Table 1b.
In principle, these hand-calculated results should
represent a greater degree of accuracy because the truncation error
is not present; however, the hand-calculated average is
very tedious to determine. For CO levels in the region of the TWA
PEL, the precision and accuracy results for the two calculation
procedures were comparable.
The "A" experiments were conducted using calibration flow rates
slightly higher than the recommended rate of 0.2 L/min. This would
result in slightly lower Datalogger concentration readings during
the "A" experiments. However, uncompensated background CO in the
diluent air would result in a slightly larger generated CO
concentration than the theoretical calculations. Thus, one source of
error could have canceled the other.
The sampling and analysis data showed very good precision and
accuracy. All data passed the outlier and Bartlett's tests;
therefore, the data were pooled. The CV2
(Pooled) value based on the TWA output values is 0.040 for samples
taken in the range of 16 to 70 ppm CO (50% RH).
Note: |
Two problems were noted during the generations.
These were sensor face velocity dependence, and background
contamination during the "B" experiments.
Using the tubing devices similar in design to the Draeger
calibration adapter to connect the Dataloggers to the sampling
manifold of the generation system, it was found that
increasing the flow rate of CO-containing air over the
Datalogger sensor tended to result in higher CO concentration
readings, even though the concentration had not changed. This
was possibly due to a slight increase in pressure at the face
of the sensor caused by the high gas flow rates through the
connecting tubing. The pressure at each sampling port was
estimated by a water manometer to be 2 to 6 mm water above the
surrounding atmosphere. Flow rates used during the studies
tended to be higher than the 0.20 L/min recommended for normal
calibrating. Slight increases in CO recoveries possibly
occurred due to increased pressure and not from humidity. This
calibration flow rate dependence is further discussed in
Section 4.6.
For two of the "B" experiments (1 × PEL, 30% RH and 0.5 × PEL,
80% RH) an attempt was made to examine this effect by having
significantly different gas flows when calibrating and taking
the samples. The other "B" experiments were conducted by
closely matching the calibration gas flow with the generation
system flow.
Also during the "B" experiments, it was noted that
significant background levels of CO existed in the laboratory
atmosphere, in some cases up to 6 ppm. This was readily
apparent when observing the Dataloggers' operation after
zeroing the monitors with nitrogen. At the time of the "B"
experiments, a climatic temperature inversion occurred, and
higher than normal concentrations of CO were retained in the
lower atmosphere. Because the diluent air is prepared from
atmospheric air, this contamination could result in higher
than normally expected CO concentrations in the
generating system. Normal expected background CO
concentrations in the generation system were <1 ppm.
Prescrubbing the diluent air for the generation system through
hopcalite removed only about one-half the amount of CO. With
the hopcalite prescrubbing, the CO concentration was still
occasionally observed to drift slightly as the experiment
proceeded.
To compensate for the elevated CO background noted during
the "B" experiments, "blank addition" was performed. The
background CO concentration was determined by averaging a few
monitor readings after the monitors were connected to the
generation system with the dilution air passing over them
(prior to 0.5% CO addition) and also at the conclusion of the
experiment (after the 0.5% CO had been turned off). The
average CO concentration determined from these readings was
added to the theoretical concentration derived from the CO gas
and diluent air flows to obtain a corrected theoretical
concentration. These corrections ranged from 0.5 to 4 ppm
CO. |
4.3. Sampling and Analysis at Low CO Levels
Procedure: In order to evaluate the Dataloggers at CO levels
commonly encountered during IAQ investigations, test atmospheres at
about 5 and 10 ppm CO (50% RH) were prepared and Datalogger samples
were taken, using the same equipment and the sensor conditions
described in Section 4.2.
The effects of the gas flow rate over the sensor, as well as the CO
background concentration in the generator diluent air, were accounted
for in these studies. Six Dataloggers were evaluated at each
concentration.
Results: Recoveries for the samples generated at
approximately 5 and 10 ppm CO are given in Table 2a
(Datalogger truncated TWA output values) and Table 2b
(hand-calculated averages of the 1-min ppm values
collected by each Datalogger).
Results at 5 ppm gave a large OE value; however, this concentration
level is very low in relation to the TWA PEL for CO. The
hand-calculated results in Table 2b
are an improvement over TWA output values given in Table 2a.
This is to be expected because, at these low levels, recovery values
based on hand calculating are taken to the nearest 0.1 ppm (Table 2b).
The TWA output values (Table 2a)
are truncated to the next lowest whole ppm.
4.4. Sampling and Analysis at Ceiling CO Levels
Procedure: In order to evaluate the Dataloggers at CO levels
in the vicinity of the Ceiling PEL (for a 5-min exposure
period), generated samples near 200 ppm CO (50% RH) were also prepared
and analyzed. For this experiment, six dataloggers were calibrated
using the 40-ppm CO gas previously mentioned, and then
mounted in a Du Pont gas flow chamber [E. I. Du Pont de Nemours and
Co., Inc., Wilmington, DE. An excellent description of a similar
chamber and the mixing chamber (Figure 1a)
can be found in reference 5.28.]
designed for passive monitor evaluation. The chamber was connected to
the generation system using Teflon tubing. The cross-sectional area in
the chamber was 83.87 cm2 (5.29.).
The gas flow rate in the chamber during the experiment was estimated
by calculation as 1.1 m/min (3.6 ft/min). The experimental procedure
consisted of placing the Dataloggers along a rigid Teflon tube in the
chamber before assembling the system. The Dataloggers were mounted
with the sensors facing upward, toward the direction of air flow
through the chamber. Datalogging was initiated immediately before the
monitors were mounted in the chamber. After placing the monitors in
the chamber, air flow was started, then the 0.496% CO was turned on.
After the CO concentration in the chamber had stabilized
(determined by observing the monitor LCDs through the glass chamber
wall), two gas bag samples of the CO/air mixture were taken for GC
analysis from a sampling manifold connected to the Du Pont chamber. At
about 50 min after the CO concentration had stabilized, the gas flows
were stopped and the Dataloggers were removed and datalogging was
halted.
Results: Recoveries for the samples generated at
approximately 200 ppm CO are given in Table 3.
The hand-calculated averages of the 1-min
ppm values collected by each Datalogger for the time period of
interest were used. Results are given for:
- the first 5-min period after the CO concentration
readings began to stabilize, and
- the entire period from when the CO concentration began to
stabilize to the stopping of the CO flow. Calculated recoveries were
96.1% for the initial 5-min time period and 93.9% for
the entire period. As previously stated in Sections 1.3.,
1.4.,
and 3.2.,
the Datalogger system only gives 1-min averages and,
therefore, 5-min truncated values are not displayed by
the Dataloggers.
4.5. Humidity Study
Procedure: Samples were also generated at 30% and 80% RH
using the same equipment and conditions described in Section 4.2.
An experiment was not conducted for 2 × TWA PEL at 30% RH.
Results: The results for the generated samples collected at
the three RH levels are listed in the following Tables:
Table |
% Humidity |
Determination |
1a 1b 4a 4b 5a 5b |
50 50 30 30 80 80 |
Truncated values Hand-calculated
average values Truncated values Hand-calculated average
values Truncated values Hand-calculated average
values |
Two "B" experiment tests (1 × PEL at 30% RH and 0.5 × PEL at 80%
RH) showed a positive bias in the mean recoveries obtained. These were
initial "B" experiments, and were conducted with calibration flow
rates lower than generation system flow rates. The increased flow
across the sensor face as compared to the calibration flow was
suspected of causing the higher bias.
As shown in Table 6,
an analysis of variance (F test) (5.30.)
was performed on the data to determine any significant difference
among or within the various RH groups. Variance at each concentration
(0.5, 1, and 2 × TWA PEL) was compared across the three RH levels
(30%, 50%, and 80% RH). The variance among and within the different
concentration groups gave high calculated F values at 0.5 and 1 × PEL.
The large calculated F values appear to be mainly due to variability
in mean recoveries, and are most likely related to variance from
sensor face velocity differences within the experiments.
Recalculation of F values after making estimated corrections for
face velocity differences between calibrations and generation system
measurements reduced the F values considerably.
4.6. Assessment of the Rate of Gas Flow on the
Datalogger Readings
Procedure: As mentioned previously in Section 4.2.1.,
tubing devices similar to the Draeger calibration adapter were used to
connect the Dataloggers to the generation system. Under these
conditions, generation system flow rates over the Datalogger sensor
faces (linear face velocities) tended to be high, and therefore,
higher CO sample concentration readings were obtained.
An investigation of face velocity dependence of gas flowing through
the calibration adapter or similar tubing was conducted with six
Dataloggers which were calibrated with the 104-ppm CO certified
standard mentioned in Section 4.2.4.
Four different sets of tests were conducted:
- Two tests were performed to determine linear face velocity
effects on Datalogger readings; in one experiment the Dataloggers
were calibrated with the gas flow set at 0.60 L/min, and in the
other the calibration was conducted with the gas flow set at 0.20
L/min. In each experiment, readings were then taken from the
Dataloggers as the rate of flow of the 104-ppm CO through the
calibration adapter over the sensors was varied.
- In a test using three monitors, the tubing similar to the
calibration device was replaced with a short piece of glass tubing
which fit loosely over the monitor sensor cap. Gas flows ranging
from 0.20 to 2.50 L/min were permitted to flow over the sensors.
- A test was conducted to determine any effect of varying the
nitrogen flow over the sensor after zeroing the monitor. The
monitors were zeroed and then the flow rate of pure nitrogen was
increased while attached to a Datalogger. Any change from a zero
reading in the Datalogger output was noted.
- A test was conducted to evaluate the effect of fast (turbulent)
air flow past the Datalogger sensors when the monitors were
surrounded by the moving air, rather than being exposed only through
the adapter tubing over the sensor (linear flow). Four Dataloggers
were exposed in a chamber to air containing 102 ppm CO at 50% RH
derived from the generation system. In the chamber, the air was
stirred with a fan so that turbulent air at speeds of 15.2 to 91.4
m/min (50 to 300 ft/min) could be swept by the monitors and their
sensors.
Results: For test (1), the Datalogger readings as a function
of gas flow across the face of the sensor (face velocity) are given in
Figure 2.
It is evident that the readings tend to rise as the linear flow
increases, and a "steady state" is achieved where the readings stop
rising at a certain point.
Datalogger Readings vs. Flow Rate
Figure 2
Gas flow rates over the sensor in metric units are given below.
Gas Flow Rate Over
Sensor
|
L/min
|
m/min*
|
ft/min
|
0.0532 0.103 0.201 0.303 0.405 0.506 0.599 0.700 0.805 0.903 1.006 1.206 1.402 1.610 1.804 2.002 |
0.29 0.56 1.09 1.65 2.20 2.75 3.26 3.81 4.38 4.91 5.47 6.56 7.62 8.75 9.81 10.89 |
0.9 1.8 3.6 5.4 7.2 9.0 10.7 12.5 14.4 16.1 17.9 21.5 25.0 28.7 32.2 35.7 |
* Based on a measured sensor
exposure area of 1.839
cm2 |
For test (2) the rise in observed readings was less pronounced when
tubing similar to the calibration tubing was replaced with the
loose-fitting glass tube. It should be stressed in either test, the
flow of the CO/air mixture is being applied directly at the sensors.
The Datalogger dependence on face velocity appears important during
calibrations and appears partially due to increased pressure on the
sensor face. Although the curvature shown in Figure 2
is important during calibration, it follows a similar pattern noted
with other passive monitors. Face velocity dependence has been noted
previously (5.29.),
and it is not unusual for a monitor to have:
- a very low sampling rate at low face velocities (<5-10
ft/min),
- a relatively constant ("steady state") sampling rate over a
specific sampling range,
- a higher sampling rate at very large face velocities (>500
ft/min).
This effect of air flow rate on the results for passive monitors
has been discussed further in a report from the National Institute for
Occupational Safety and Health (5.31.).
Typical face velocities in general industry, according to the NIOSH
report (5.31.)
are 10 to 150 cm/s (approximately 6 to 90 m/min or 20 to 300 ft/min).
In test (3) where the nitrogen flow after zeroing was altered,
there was no significant change from zero as the flow rate of nitrogen
increased. This indicates that pressure alone is not responsible for
the elevated readings noted in tests (1) and (2).
In test (4) where the monitors were exposed to the rapidly moving
turbulent air, the monitor readings leveled off in the range of 103 to
114 ppm CO. The readings were similar even when the fan was turned
off. It appears the monitors are only slightly affected by turbulent
air flows of 15.2 to 91.4 m/min (50 to 300 ft/min) where a positive
bias of approximately 5 to 10% was noted. The monitor appears more
affected by linear flows. It is expected in many industrial settings
the air flow will move in a turbulent fashion.
The manufacturer has reported the following calibration flow
deviations when CO-containing gas flow rates different from the
recommended 0.20 L/min are used:
Ft/min
|
|
L/min
|
|
% Error
|
2.7 |
|
0.17 |
|
-8 |
2.9 |
|
0.18 |
|
-7 |
3.3 |
|
0.21 |
|
<3 |
3.7 |
|
0.23 |
|
<3 |
These manufacturer-determined values appear similar to
those shown in Figure 2.
The flows are clustered close to the recommended calibration flow rate
of 0.2 L/min. According to the manufacturer, experiments in which the
air velocity was varied from 0 to 6 m/s (1,200 ft/min), the
sensitivity was <±5% of the measured value (5.32.).
It is not known whether the test was conducted using linear or
turbulent flows.
Both series of internal and manufacturer-performed
experiments discussed above demonstrate a significant relationship of
Datalogger accuracy and calibration flow rate. Significantly different
results may be obtained during sampling if the Datalogger is
calibrated at flow rates different than recommended (0.2 L/min).
4.7. Detection Limits
Procedure: Both qualitative and quantitative limits for the
analysis of CO by the Dataloggers were calculated using the
International Union of Pure and Applied Chemistry (IUPAC) method for
detection limit determinations (5.33.).
The data collected at 50% RH and 5.6, 11.1, and 16.1 ppm (Tables 1a,
1b,
2a,
and 2b)
were used. Also included were data from a blank generation in which no
CO was added. The blank generation was performed when the background
CO level in the air was at a very low level (0 to 1 ppm).
Results: The detection limit results are listed in Table 8a
(for the truncated TWA output values) and Table 8b
(for the hand-calculated averages of all of the
1-min ppm values). A derivation of the IUPAC method was
used for the data in Table 8a
because the blank readings were zero. The standard deviation of the
lowest concentration standard was used in place of the blank for data
in Table 8a.
The results are summarized as follows:
Detection Limit
|
|
Truncated TWA Values
|
|
Hand-Calc. Ave of 1-min
Values
|
|
Qualitative: |
|
1.2 ppm |
|
0.5 ppm |
Quantitative: |
|
4.1 ppm |
|
1.8 ppm |
The lower detection limits for the "Hand-Calc. Ave." of all the
1-min ppm values is to be expected, since fractional
values for the ppm determination are calculated, averaged, and
rounded. This type of data (rounded rather than truncated) is also
more attuned to the gaussian statistical model used. Also,
hand-calculated averages are indicative of the ultimate
CO measuring capabilities of the instrument at low concentrations.
4.8. Comparison of Methods
Procedure: In order to confirm the accuracy of the CO
concentration of the 0.50% and 0.496% CO sources used in the
generation system, as well as the accuracy of generation, one or two
gas bag samples were taken from the generated gas stream during each
Datalogger study. Bags employed were 5-L five-layer aluminized gas
bags (Calibrated Instruments, Inc., Ardsley, NY). Analysis of the CO
content of the gas bags was performed with a Tracor Model No. 540 GC
equipped with a Model No. 706 DID according to OSHA Method ID-210
(5.12.).
Standards were prepared from the cylinder of 104-ppm CO in nitrogen
previously discussed in Section 4.2.4.
The CO concentration of this gas cylinder had been confirmed during
previous work by reaction with iodine pentoxide and titration of the
subsequent iodine produced (5.12.).
Results: The GC results were compared with theoretical
values (e.g., "GC Results" = GC Recovery/Theoretical Value).
Theoretical values were calculated from generation system flow
settings. As shown in Figures 3
- 5,
good agreement is noted between theoretical and GC results. The
Datalogger results ("Datalogger" = TWA Output Values, Calc Average =
Hand-Calculated Average Values) are also illustrated for comparison
purposes. It should be noted in Figure 4
the TWA output value (Datalogger result) for the Ceiling determination
is not present. In this case the results were
hand-calculated as previously mentioned in Section 4.4.
Figure 3
Figure 4
Figure 5
4.9. Conclusions
The experimental results indicate this direct-reading
monitor is an excellent means for sampling and analyzing CO in the
workplace if properly calibrated and if air flow is not directly at
the face of the monitor sensor (linear flow). Obtaining accurate
analytical results for other-than-normal conditions existing in the
working environment may be contingent on matching the flow of the
calibration gas. Although the precision and accuracy results were
still acceptable when the calibration/generation system flow ratios
were different by a factor of two (Tables 4a
and 5a),
the monitors should normally be calibrated using the
manufacturer-stated flow settings. If the face velocity
in the sampling environment simulates a linear flow (i.e. measurements
inside of air ducts), calibration flow can be matched to sampling
flows to provide for more accurate results. It should be noted that
turbulent flow at the face of the sensors did not produce the
magnitude of positive bias noted to occur during the linear face
velocity tests. The positive bias is likely partially due to increased
pressure on the CO sample directly above the sensor face. The
Dataloggers should be calibrated at the same altitude as the site
where sampling will be conducted.
Because the instrument automatically performs integer math for
calculations (truncating to the next lower integer ppm value),
rounding-off error for the resulting TWA output values is possible.
This does not appear problematic for CO concentrations normally
encountered in industry, and is aptly shown by comparable precision
and accuracy data obtained when testing in the vicinity of the PEL
(Tables 1a
and 1b).
The error does become significant at single digit levels (i.e., <10
ppm); however, these levels are well under the present TWA PEL. It
should also be pointed out that in workplace environments the CO level
will not be static and may quite often rise to high concentrations and
fall to single digit values in one monitoring period.
Determinations near the TWA PEL were well within NIOSH and OSHA
accuracy and precision guidelines (5.26.,
5.27.).
For the truncated TWA output values at 50% RH, the pooled coefficient
of variation (CV2) was 0.040 and the overall
recovery was 98.4%. An effect from humidity is not apparent. In two
cases where the OE values appear somewhat high (1 × PEL at 30% RH and
0.5 × PEL at 80% RH), face velocity and increased pressure at the
sensor face appear as contributory factors. The manufacturer of the
Datalogger may need to further investigate interferences caused by low
molecular weight alcohols and aldehydes which appear structurally
similar to the CO molecule. These interferences are not expected to be
significant in normal workplace monitoring but may cause problems
during confined space evaluations.
The OE values at approximately 5 and 10 ppm CO were somewhat high,
but within NIOSH and OSHA guidelines if non-truncated values are used.
The Dataloggers are very convenient to use. There is minimum
interference with work procedures by employees. The exposure data are
stored and easily recovered to produce a minute-by-minute record of
the working environment.
This sampling and analytical method is capable of accurate and
precise measurements to determine compliance with the 35-ppm TWA or
200 ppm Ceiling PEL for CO exposures.
5. References
5.1. "Air Contaminants; Final Rule" Federal
Register 29 CFR Part 1910 (19 Jan. 1989). pp.
2332-2983.
5.2. United States Department of Labor,
OSHA: "Memorandum, Updated Changes to 29 CFR 1910.1000, Air
Contaminants Standard." by Patricia Clark, Director Designate,
Directorate of Compliance Programs. United States Department of Labor,
OSHA, Washington, DC, June 1, 1990. [Memo].
5.3. "Air Contaminants; Corrections" Federal
Register 29 CFR Part 1910 (1 July 1992). pp.
29204-29206.
5.4. Occupational Safety and Health
Administration Salt Lake Technical Center: Chemical Information
File, Online Database-OSHA Information System. Salt Lake City, UT:
Occupational Safety and Health Administration Salt Lake Technical
Center, 1989.
5.5. Katz, M., ed.: Methods of Air
Sampling and Analysis. 2nd ed., APHA Intersociety Committee.
Washington, D.C.: Publication Office, American Public Health
Association, 1977. No. 132, pp. 368-369.
5.6. Occupational Safety and Health
Administration Salt Lake Technical Center: Carbon Monoxide
Detector Tubes (Short-Term) (USDOL/OSHA-SLTC PE-11). Salt Lake
City, UT: Occupational Safety and Health Administration Salt Lake
Technical Center, 1990.
5.7. Occupational Safety and Health
Administration Salt Lake Technical Center: A Laboratory
Evaluation of Draeger Long Duration Carbon Monoxide Detector Tubes
(USDOL/OSHA-SLTC PE-3). Salt Lake City, UT: Occupational Safety and
Health Administration Salt Lake Technical Center, 1981.
5.8. National Institute for Occupational
Safety and Health: NIOSH Manual of Analytical Methods. 2nd.
ed., Vol. 1 (Method No. P&CAM 112) (DHEW/NIOSH Pub. No. 77-157-A).
Cincinnati, OH: National Institute for Occupational Safety and Health,
1977.
5.9. National Institute for Occupational
Safety and Health: NIOSH Manual of Analytical Methods. 2nd.
ed., Vol. 4 (Method No. S340) (DHEW/NIOSH Pub. No. 78-175).
Cincinnati, OH: National Institute for Occupational Safety and Health,
1978.
5.10. Mine Safety and Health
Administration: Regular Mine Gas Analysis (MSHA Standard
Method No. 1). Denver, CO: MSHA, 1979.
5.11. Williams, D.M.: "Unique
Applications for New Helium Glow Discharge Ionization Detector for Gas
Chromatography." Paper presented at the 1988 Pittsburgh Conference,
New Orleans, LA, February 1988.
5.12. Occupational Safety and Health
Administration Salt Lake Technical Center: Carbon Monoxide in
Workplace Atmospheres (USDOL/OSHA-SLTC Method No. ID-210).
Salt Lake City, UT: Occupational Safety and Health Administration
Technical Center, 1991.
5.13. National Draeger, Inc.: Draeger
Model 190 Datalogger, CO, Operating Manual. Pittsburgh, PA:
National Draeger, Inc., No date specified.
5.14. American Society of Heating,
Refrigerating and Air-Conditioning Engineers Inc.: Ventilation
for Acceptable Indoor Air Quality. ASHRAE 62-1989. Atlanta, GA:
American Society of Heating, Refrigerating and Air-Conditioning
Engineers, 1989. pp. 1-26.
5.15. Sax, N.I.: Dangerous Properties
of Industrial Materials. 4th ed. New York: Van Nostrand Reinhold
Company, 1975. pp. 520-521.
5.16. National Institute for Occupational
Safety and Health: Criteria for a Recommended
Standard-Occupational Exposure to Carbon Monoxide (DHEW/NIOSH Pub.
HSM 73-11000). Cincinnati, OH: National Institute for Occupational
Safety and Health, 1972.
5.17. Coburn, R.F., Chairman: Carbon
Monoxide. Washington, DC: National Academy of Sciences, 1977.
5.18. Proctor, N.H. and J.P. Hughes:
Chemical Hazards of the Workplace. Philadelphia, PA: J.B.
Lippincott Co., 1978. pp. 151-153.
5.19. Sax, N.I. and R.J. Lewis, Sr.:
Hawley's Condensed Chemical Dictionary. 11th ed. New York: Van
Nostrand Reinhold Co., 1987. pp. 221-222.
5.20. American Conference of Governmental
Industrial Hygienists: Documentation of the Threshold Limit
Values for Substances in Workroom Air. 3rd ed. Cincinnati, OH:
American Conference of Governmental Industrial Hygienists, 1971. pp.
41-43.
5.21. Ellenhorn, M.J., and D.G.
Barceloux: Medical Toxicology. New York, NY: Elsevier
Science Publishing Co., Inc., 1988. p. 821.
5.22. Lee, H.K., T.A. McKenna, L.N. Renton,
and J. Kirkbride: Impact of a New Smoking Policy on Office Air
Quality. In Indoor Air Quality in Cold Climates: Hazards and
Abatement Measures, edited by D.S. Walkinshaw. Pittsburgh, PA: Air
Pollution Control Association, 1986. pp. 307-322. (NIOSH-00172085).
5.23. Leaderer, B.P., W.S. Cain, R. Isseroff,
and L.G. Berglund: Ventilation Requirements in Buildings. II.
Particulate Matter and Carbon Monoxide from Cigarette Smoking.
Atmospheric Environment 18, No. 1: 99-106 (1984).
(NIOSH-00137853).
5.24. U.S. Department of Labor - OSHA:
"Additions to Supplies Furnished by OCL, Memorandum for Regional
Administrators, Area Directors, ARA's for Technical Support, Health
Response Team." by Robert T. Williams. U.S. Department of Labor -
OSHA, Cincinnati Laboratory, Cincinnati, OH, May 8, 1992. [Memo].
5.25. Mandel, J.: Accuracy and Precision,
Evaluation and Interpretation of Analytical Results, The Treatment of
Outliers. In Treatise On Analytical Chemistry. 2nd ed., Vol. 1,
edited by I.M. Kolthoff and P.J. Elving. New York, NY: John Wiley and
Sons, 1978. pp. 282-285.
5.26. National Institute for Occupational
Safety and Health: Documentation of the NIOSH Validation
Tests by D. Taylor, R. Kupel and J. Bryant (DHEW/NIOSH Pub. No.
77-185). Cincinnati, OH: National Institute for Occupational Safety
and Health, 1977. pp. 1-12.
5.27. Occupational Safety and Health
Administration Salt Lake Technical Center: OSHA Analytical
Methods Manual. Vol. III (USDOL/OSHA-SLTC Method Validation
Guidelines). Cincinnati, OH: American Conference of Governmental
Industrial Hygienists (Pub. No. ISBN: 0-936712-66-X), 1985.
5.28. Freeland, L.T.: An industrial
hygiene calibration manifold. AIHAJ 38: 712-720 (1977).
5.29. Occupational Safety and Health
Administration Salt Lake Technical Center: Evaluation of a
Solid Sorbent Passive Dosimeter for Collecting Mercury Vapor
(USDOL/OSHA-SLTC Backup Report No. ID-140). Salt Lake City, UT:
Occupational Safety and Health Administration Technical Center,
revised 1989.
5.30. Dowdy, S. and S. Wearden:
Statistics for Research. New York: John Wiley and Sons, 1983.
Chapter 8.
5.31. Cassinelli, M.E., R.D. Hull, J.V.
Crable, and A.W. Teass (National Institute for Occupational Safety
And Health): Protocol for the Evaluation of Passive Monitors. In
Diffusive Sampling - An Alternative Approach to Workplace Air
Monitoring, The Proceedings of an International Symposium held in
Luxembourg, 22-26 September 1986, edited by A. Berlin, R.H. Brown, and
K.J. Saunders. London: The Royal Society of Chemistry, Burlington
House, 1987. pp. 190-202.
5.32. Nancy Hoblack: "Effect of Air
Velocity on Draeger CO Datalogger Readings." July 9, 1992. [Private
Communication]. Nancy Hoblack, National Draeger, Inc., 101 Technology
Drive, Pittsburgh, PA 15230.
5.33. Long, G.L. and J.D. Winefordner:
Limit of Detection - A Closer Look at the IUPAC Definition. Anal.
Chem. 55: 712A-724A (1983).
5.34. National Draeger, Inc.: Enhanced
Graphics Software (Version 1.0), Operating Manual. Pittsburgh, PA:
National Draeger, Inc.
5.35. U.S. Department of Labor - OSHA:
"National Draeger CO Dosimeters, Memorandum for Rick Cee." by Robert
T. Williams. U.S. Department of Labor - OSHA, Cincinnati Laboratory,
Cincinnati, OH, February 7, 1991. [Memo].
Appendix A
Estimates of Saturated Vapor Concentrations of Selected
Interferences
The saturated vapor concentrations for interferences discussed in
Section 1.6.
of this method were estimated from commonly found vapor pressure data. Six
values for the temperature versus vapor pressure were given for each
compound (the vapor pressure values were taken from the following
reference: Lide, David R., ed.: CRC Handbook of Chemistry and Physics.
73rd ed. Boca Raton, FL: CRC Press, 1992-1993. pp. 6-68 to
6-82). These values were plotted for each compound of
interest. The temperature in EC versus the logarithm of the vapor pressure
was plotted as shown below:
Temperature vs log(Vapor Pressure)
The vapor pressures listed below at 25°C were determined by
interpolation. The saturated vapor concentration was then calculated using
the equation:
Concentration = vp × 106/P
Where: |
vp P |
= = |
vapor pressure atmospheric
pressure |
The tests were performed at an atmospheric pressure of approximately
650 mmHg. These are estimates of the concentration near the surface of the
liquid; in practice, the Datalogger was held a short distance away from
the liquid surface. The actual concentration of each substance during
interference measurements was less than reported here.
|
Vapor |
----------Concentration
(ppm)---------- |
Name |
Pressure (mmHg) |
P@760 mmHg |
P@650 mmHg |
|
Acetone |
223.9 |
295,000 |
344,000 |
Ethanol |
57.5 |
75,700 |
88,500 |
Isobutanol |
12.6 |
16,600 |
19,400 |
Isopropanol |
43.7 |
57,500 |
67,200 |
Methanol |
123.0 |
162,000 |
189,000 |
Acetonitrile |
91.2 |
120,000 |
140,000 |
n-Butylamine* |
85.1 |
112,000 |
131,000 |
Ethyl acetate |
93.3 |
123,000 |
144,000 |
Isooctane** |
50.1 |
65,900 |
77,100 |
1-Octanol |
0.1*** |
130 |
150 |
Toluene |
28.2 |
37,100 |
43,400 |
n-propanol |
20.4 |
26,800 |
31,400 |
|
* |
Vapor pressures estimated from
boiling temp. of 77.8°C and vapor properties of related
amines. |
** |
2,2,4-trimethylpentane |
*** |
Value estimated by
extrapolation. | |
Two of the compounds tested for interferences with the Draeger
Dataloggers were available only in diluted solutions. These were:
Formaldehyde, 37% wt/wt in water, with 10-15% methanol
(preservative).
Dimethylamine, 26% wt/vol in water.
Values for the partial pressures of formaldehyde over its aqueous
solutions as a function of temperature and solution concentration were
obtained from the following reference:
Walker, J. Frederick: Formaldehyde. American Chemical
Society Monograph Series #120. New York: Reinhold Publishing Corp.,
1953. p. 92.
For 36.2 g formaldehyde/100 g water at 20°C, the partial pressure of
formaldehyde was given as 1.025 mmHg.
|
Partial |
----------Concentration
(ppm)--------- |
Name |
Pressure (mmHg) |
P@760 mmHg |
P@650 mmHg |
|
Formaldehyde |
1.025 |
1,350 |
1,580 |
This value is for HCHO in pure water, whereas the solution available in
the laboratory also contained methanol. No attempt was made to estimate
the partial pressure of the dimethylamine over its solution.
Appendix B
Equipment, Initial Preparation, Calibration, Sampling, and Data
Recovery Procedures
Draeger Model 190 CO Datalogger (5.13.,
5.34.)
1a. Equipment
1.1.a. Draeger Model 190 CO Datalogger (Figure 6)
Figure 6
1.2.a. CO-specific filter (filter housing in Figure 6)
1.3.a. Red and blue function keys for starting and ending of
datalogging (Figure 7)
Figure 7
1.4.a. Alkaline battery (9 volt)
1.5.a. A small screwdriver for adjusting the zero, span, and alarm
trimpots
1.6.a. Calibration equipment and supplies
1.6.1.a. Calibration cylinder gas flow regulator (set to 0.20
L/min) (available from Draeger)
1.6.2.a. Calibration adapter (unit consisting
of tubing attached to a perforated cap that fits over the monitor
detection cell filter, see Figure 8)
Figure 8
1.6.3.a. Gas cylinder with certified CO concentration for
calibration
1.7.a. Data reduction equipment
1.7.1.a. Draeger RS-232 Interface, consisting of a Converter Box
(25 pin) and Converter Box Power Supply for transferring the data
from the Datalogger to a computer (Figure 9)
CONVERTER BOX
Figure 9
1.7.2.a. Computer program [National Draeger Inc. Enhanced Graphic
Software (EGS) 2.0 Part Number 4510259] on a 32-in. (720K byte)
micro diskette or a 5¼-in. (360K byte) disk
Note: |
These versions will not transfer data
compatible to standard laser printers. Dot matrix printers
shall be used unless the laser printer is modified for output.
Contact Draeger for additional
information. |
1.7.3.a. MS-DOS operating system (Draeger did not stipulate in
their documentation what version of DOS is required; however, DOS
3.3 and above appear compatible)
1.7.4.a. Computer such as an IBM or compatible AT or XT computer
with an RS-232 (serial) port (Toshiba T1100PLUS Lap Top Computer)
1.7.5.a. Printer such as an IBM/Dotmatrix printer (Epson FX-86e).
2.a. Initial Preparation
2.1.a. Battery installation - Remove the battery cover by
loosening the two screws and lifting off the cover. Insert a fresh
9-volt alkaline battery into the battery compartment and
connect the battery terminal socket. The liquid crystal reading
initially is a very high value and the alarm may activate. As the
reading drops, eventually the alarm should become silent. Replace the
battery housing; ensure that the battery lead wires will not be caught
in between the covers. Seat the cover properly before the screws are
tightened. Usually, a battery will last for about 1 month. If the
Datalogger is not to be used for an extended period of time, the
battery should be removed.
Caution: Do not replace the battery in potentially hazardous
environments where a spark might cause an explosion.
2.2.a. Stabilization - When the battery is installed or a new unit
is placed in service, a 24-h stabilization period is necessary. If the
battery is only being replaced and the old battery indicates a low
battery condition (by a short audible tone emitted about every 10 s,
or an unstable LCD reading), allow a 2-h period for stabilization.
Carefully install the blue key to reset the memory.
Precautions
- When installing the red or blue function keys, turn only the
knurled ring at the base of the key. Do not attempt to twist the
entire key, as the electrical prongs can bend excessively.
- Replace the filter unit if it has become contaminated or has a
large amount of trapped dust present.
- After prolonged use, the electrochemical sensor may also need to
be replaced, as evidenced by erratic behavior of the values observed
on the LCD.
Note: |
The sensor contains sulfuric acid. Normally, use
of the sensor does not result in exposure; however, if the
container leaks or breaks, contact with the solution can cause
severe burns or eye damage. If contact with the solution occurs,
flush exposed body parts with cold water for about 10 min and
get immediate medical attention. |
2.3.a. Alarm adjustment procedure
Precautions
- If part of the circuit board remains under the edge of the cover
after alarm adjustment, the cover will not be completely closed and
the Datalogger will not be adequately shielded from radio frequency
(RF) radiation. Loss of RF shielding can result in false readings
and alarms if subjected to strong RF fields (5.35.).
- The alarm set point adjustment must be done in a clean air
environment to avoid false zero adjustment levels.
Remove the filter unit from the monitor sensor. Loosen the three
screws (located on the back side from the front panel) holding the
housing cover in place and remove the cover. Locate the zero and alarm
potentiometer adjustment screws (the alarm adjustment screw is between
the zero and span adjustment screws, marked "Z" and "S" respectively).
Adjust the zero potentiometer until the display indicates the ppm
value of the desired alarm set point. [For compliance data collection,
the alarm should be set to above 200 ppm CO (the Ceiling concentration
value)]. If the alarm is sounding at concentrations lower than these,
adjust the alarm set potentiometer clockwise until the alarm is
silenced, then set counterclockwise until the alarm sounds. If the
alarm is quiet at this concentration, adjust the alarm set
potentiometer counterclockwise until the alarm sounds. Reset the zero
potentiometer ("Z") to 0. Replace the housing. Be sure the battery
lead wires are routed around the screw post. Tighten the three screws.
Be sure the instrument circuit board in the upper chamber is
properly seated and the housing is properly in place before the screws
are tightened. Replace the dust filter.
3.a. Calibration
3.1.a. According to the manufacturer, "calibration must be
performed at least on a monthly basis to keep the Model 190 Datalogger
within published specifications" (5.13.);
however, for OSHA enforcement purposes, calibrate each time the
monitors are used.
3.2.a. The recommended concentration of the calibration gas (span
gas) should be 30 to 40 ppm CO. If necessary, higher concentrations of
CO, 100 ppm or more, may also be used for calibration of the Draeger
Dataloggers.
Note: |
The combination of the truncation feature of the
Draeger Datalogger and the recommendation that the calibration
gas should be 30 to 70 ppm CO increases the potential systematic
error up to 3%. Using calibration gases whose concentrations are
close to integer values (i.e. 39 ppm versus 39.8 ppm) will
reduce this potential error. |
3.3.a. Zero adjustment - Remove the blue function key, install the
red function key and push the button on the key to stop the logging
mode. Zero the instrument using the calibration adapter (see 1.6.2.a.
above) and a source of CO-free nitrogen or CO-free air. Allow the
nitrogen or air to flow at about 0.20 L/min until the reading
stabilizes. Locate the zero adjustment screw ("Z"). Adjust the zero
potentiometer until the display reads zero ppm CO. (Be certain that
the potentiometer is capable of being adjusted to within ±5 ppm.)
3.4.a. Calibration procedure (Span adjustment) - IF POSSIBLE,
PERFORM THIS PROCEDURE IN A WELL VENTILATED AREA SUCH AS AN EXHAUST
HOOD. Attach the monitor calibration adapter to the regulator on the
calibration gas cylinder. Turn on the regulator (set to a flow rate of
0.20 L/min) and purge the calibration adapter. Place the calibration
adapter over the filter housing. Wait until the LCD gives a constant
reading. (New sensors may take more than 2 min to stabilize.) Locate
the span adjustment screw ("S") on the side of the instrument. Adjust
the span potentiometer until the display indicates the ppm value of
the known calibration (span) gas (rounded to the nearest whole ppm).
Note: |
Do not perform calibrations in the presence of an
operating radio transmitter. |
4.a. Sampling Procedure
4.1.a. The unit is ready for use when the battery is installed.
Sufficient battery power is indicated by a periodic flash of the LED
visual alarm (about every 10 s). When the alarm is activated, more
power is needed. Insufficient battery power is indicated by a short
audible tone emitted about every 10 s; the battery should be replaced
at the end of any sampling period if the low battery warning is
activated. The industrial hygienist should be notified if this occurs.
4.2.a. Functional use - ALWAYS OBTAIN A PRINTOUT OF ANY USABLE
STORED DATA PRIOR TO INSTALLING THE BLUE KEY.
Installation of the blue key puts the Datalogger into "Sleep-mode,"
turns off the display, and reduces power consumption.
To start the Logger-mode and reset (clear) the memory, align the
blue key, insert it into the Datalogger, turn the knurled ring until
it is tight, and then remove the key. This starts the datalogging.
When doing this, data stored previously in the memory will be lost.
During data collection, the display will show consecutively: (1)
the immediate concentration (displayed for 5 s), (2) a "running" TWA
(displayed for 2.5 s), (3) the peak concentration recorded during the
time period sampled (displayed for 2.5 s). The Datalogger stops
sampling after 12 h and the display shows OFF. To terminate the
logging function at any time, attach the red key and depress the
button once. The display will then show a continuous monitoring of the
immediate concentration of the environment [as discussed in (1)
above]. Do not depress the red key again before transferring the data
or the data may be lost.
5.a. Data Recovery
5.1.a. Precautions
DO NOT INSTALL THE BLUE KEY PRIOR TO DATA TRANSFER,
or the stored data will be lost.
DO NOT TRANSFER DATA IN POTENTIALLY HAZARDOUS
ENVIRONMENTS where a spark might cause an explosion.
1. |
Converter Box |
2. |
9V AC/DC Adaptor |
3. |
Floppy disc - Enhanced Graphic
Software |
4. |
RS232 Interface Connector
(female) |
5. |
ON/OFF Switch (SW1) for starting
transmission |
6. |
AC/DC Supply Connector |
7. |
"Power On" LED (green) |
8. |
"Transmit" LED (red) |
9. |
Interface cable to Model 190
Datalogger |
10. |
Switch 2 (SW2) Selection Data Receive
Line |
11. |
Switch 3 (SW3) Connection DTR to
ground |
Figure 10
5.2.a. Computer equipment setup
5.2.1.a. Connect the RS232 connector of the Converter Box to the
RS232 port of the computer (Figure 10).
Use a suitable extension cable or adapter if needed.
5.2.2.a. Connect the AC/DC Adapter to the Converter Box and the
power outlet, making sure the switch (SW1) on the side of the
Converter Box is in the OFF position.
5.3.a. Programming of the computer
5.3.1.a. Load the MS-DOS system into the computer memory.
5.3.2.a. Load the Draeger EGS software into the computer (type
EGS, press Enter).
5.3.3.a. When the Draeger logo appears on the screen, press any
key. The following menu will appear:
DATALOGGER ANALYZING SYSTEM
Commands: |
|
0 = List files |
1 = Quit |
2 = Retrieve file |
3 = Load from Logger |
4 = Save file |
5 = Display graph |
6 = Print Report |
7 = Configuration |
|
Type Command -> |
|
5.4.a. Data transfer
5.4.1.a. Connect the Converter Box cable to the Datalogger.
5.4.2.a. Press 3 (after Type Command ->) on the
computer. This will set it up to receive data from the Datalogger.
5.4.3.a. Switch SW1 on the Converter Box to ON. The green LED
will be lit. Data transmission will start automatically and the red
LED will be lit. After a few moments, data transmission will be
complete and the red LED will extinguish. At this point, switch SW1
to OFF. The data stored in the Datalogger will remain intact.
5.4.4.a. Default text information (from any previous data entry)
will now appear on the screen. New text information can be added at
this point:
Name |
of user or operator |
Location |
where measured or headquarters |
Date |
when logged or downloaded |
Start Time |
when logging was started (military time
format) |
Comment |
regarding this record |
Gas |
(select number from list) |
Serial Number |
of instrument used |
5.4.5.a. Disconnect the Datalogger from the Converter Box. The
data are now stored in the computer. The Datalogger should still
have the data retained.
5.4.6.a. Screw the cap on the Datalogger key connection terminal.
If data retention is no longer necessary and the monitor is to be
stored, insert the blue key. This will reduce battery power
consumption.
5.5.a. Other DATALOGGER ANALYZING SYSTEM commands
Note: |
Check the configuration of the system before data
transfer (press 7) to assure data will be deposited to the
appropriate disk or area in the computer. |
5.5.1.a. Save file: Press 4 and then enter a file
name (up to eight characters). Do not supply the ".DAT"
extension; this is added by the software. The data will be stored in
the computer.
5.5.2.a. Commands used to retrieve data:
List files: Press 0 to obtain a list of other
Datalogger files in storage.
Retrieve file: Press 2 to retrieve another set of
data from storage and store it in the computer memory.
Display graph: Press 5 to display the graph of CO
concentration versus time. When the computer-printer system has been
set up, follow the instructions to print a copy of the graph. To
obtain a vertical printout, press F7. To obtain a horizontal
printout, press F8. To return to the menu, press
Enter.
Print report: When a printer system is installed, press
6 to print a copy (spread sheet) of the complete list of the
data from the Datalogger.
5.5.3.a. Quit: When data handling operations are complete,
press 1 to exit from the EGS program.
Appendix C
Calibration Gases from the OSHA Cincinnati Laboratory
The following gas blends containing CO are available for the OSHA
offices. Orders may be placed through the regional offices, as for other
supplies (5.24.):
BLEND #2 - |
Mixture of 35 ppm Carbon Monoxide, 1.6% Methane,
19.5% Oxygen and balance Nitrogen. The actual concentration of the
Carbon Monoxide may vary between 35-40 ppm, the Methane between 1.5
-1.7%, and the Oxygen between 19.5% to 19.0%. |
|
BLEND #3 - |
70 ppm ±5 ppm Carbon Monoxide in Air. |
|
BLEND #5 - |
Mixture of 50 ppm Carbon Monoxide, 0.55% ±.05%
n-Hexane (50% of Lower Flammable Limit), 19.5% -.2% +0% Oxygen and
the balance Nitrogen. [Note: Due to oxygen meter alarm
characteristics, the oxygen range for this blend may not exceed
19.5% (+0%) and may not be lower than 19.3% (-0.2%).]
|
Examples of the labels which may appear on the BLEND #2 gas cylinders
are shown below:
THIS CYLINDER CONTAINS THE NECESSARY CONCENTRATION OF THE
LISTED GASES TO SPAN OR CHECK THE ALARM OPERATION OF THE FOLLOWING
INSTRUMENTS:
DRAEGER MODEL 190 CO DOSIMETER MSA MICROGARD PORTABLE
ALARM SCOTT-ALERT MODEL S105 ISC CMX270, 271 & MX241
MONITORS
THE INSTRUMENT MUST SEE THE CORRECT FLOW OR ERRORS IN READING THE
GAS CONCENTRATIONS CAN OCCUR. USE THE APPROPRIATE REGULATOR THAT
SUPPLIES EACH INSTRUMENT THE PROPER FLOW AS LISTED BELOW:
DRAEGER MODEL 190 |
200 sccm |
MSA MICROGARD |
200--250 sccm |
SCOTT-ALERT MODEL S105 |
750--1000 sccm |
ISC CMX & MX MONITORS |
500--750 sccm |
To re-order this cylinder thru the OSHA Cincinnati Laboratory,
contact the ARA for Technical Support in your region. Order: BLEND #2.
(Note: sccm = standard cubic centimeters per minute)
Approximate pressure |
1000 psig |
Approximate volume |
100 liters @ 70°F |
ANALYSIS |
COMPRESSED GAS N.O.S. |
UN1956 |
CARBON MONOXIDE |
35.3 PPM |
METHANE |
1.59% |
OXYGEN |
19.5% |
NITROGEN |
BALANCE |
Date shipped |
10/16/91 |
Tables
Table 1a
Sampling and Analysis - TWA Output Values
(Truncated) [50% RH and 20-25°C]
(OSHA-TWA
PEL) |
Monitor
|
ppm Taken
|
ppm Found
|
F/T
|
N
|
Mean
|
Std Dev
|
CV
|
OE
|
(0.5 × TWA
PEL) ("A" experiment) |
|
#1 #2 #3 #4 #5 #6 |
16.1 16.1 16.1 16.1 16.1 16.1 |
15 16 16 16 17 16 |
0.932 0.994 0.994 0.994 1.056 0.994 |
|
|
6 |
0.994 |
0.039 |
0.040 |
8.5 |
(1 × TWA
PEL) ("A" experiment) |
|
#1 #2 #3 #4 #5 #6 |
39.2 39.2 39.2 39.2 39.2 39.2 |
36 37 37 38 40 38 |
0.918 0.944 0.944 0.969 1.020 0.969 |
|
|
6 |
0.961 |
0.035 |
0.036 |
11.2 |
(2 × TWA
PEL) ("B" experiment) |
|
#1 #2 #3 #4 #5 #6 |
70.2 70.2 70.2 70.2 70.2 70.2 |
73 69 68 70 74 66 |
1.040 0.983 0.969 0.997 1.054 0.940 |
|
|
6 |
0.997 |
0.043 |
0.043 |
9.0 |
F/T |
= |
Found/Taken |
OE |
= |
Overall Error (±
%) |
Bias |
= |
-0.016 |
CV2 (Pooled) |
= |
0.040 |
OE (Total) |
= |
±9.6% | | |
Table 1b
Sampling and Analysis - Hand-Calculated Average
Values* [50% RH and 20-25°C]
(OSHA-TWA
PEL) |
Monitor
|
ppm Taken
|
ppm Found
|
F/T
|
N
|
Mean
|
Std Dev
|
CV
|
OE
|
(0.5 × TWA
PEL) ("A" experiment) |
|
#1 #2 #3 #4 #5 #6 |
16.1 16.1 16.1 16.1 16.1 16.1 |
15.3 17.0 16.0 16.0 17.2 16.1 |
0.950 1.056 0.994 0.994 1.068 1.000 |
|
|
6 |
1.010 |
0.044 |
0.044 |
9.8 |
(1 × TWA
PEL) ("A" experiment) |
|
#1 #2 #3 #4 #5 #6 |
39.2 39.2 39.2 39.2 39.2 39.2 |
36.0 37.8 37.0 38.1 40.1 38.1 |
0.918 0.964 0.944 0.972 1.023 0.972 |
|
|
6 |
0.966 |
0.035 |
0.036 |
10.7 |
(2 × TWA
PEL) ("B" experiment) |
|
#1 #2 #3 #4 #5 #6 |
70.2 70.2 70.2 70.2 70.2 70.2 |
73.2 69.5 68.5 70.6 74.6 66.7 |
1.043 0.990 0.976 1.006 1.063 0.950 |
|
|
6 |
1.005 |
0.042 |
0.042 |
8.8 |
F/T |
= |
Found/Taken |
OE |
= |
Overall Error (±
%) |
Bias |
= |
-0.007 |
CV2 (Pooled) |
= |
0.041 |
OE (Total) |
= |
±8.8% | |
* Based on averages of
all ppm values collected by each
Datalogger | |
Table 2a
Sampling and Analysis - TWA Output Values
(Truncated) [50% RH and 20-25°C, Low Concn]
Monitor
|
ppm Taken*
|
ppm Found
|
F/T
|
N
|
Mean
|
Std Dev
|
CV
|
OE
|
(10 ppm)
("B" experiment) |
|
#1 #2 #3 #4 #5 #6 |
11.1 11.1 11.1 11.1 11.1 11.1 |
10 10 11 11 11 10 |
0.901 0.901 0.991 0.991 0.991 0.901 |
|
|
6 |
0.946 |
0.049 |
0.052 |
15.8 |
(5 ppm) ("B"
experiment) |
|
#1 #2 #3 #4 #5 #6 |
5.6 5.6 5.6 5.6 5.6 5.6 |
5 5 5 5 5 4 |
0.893 0.893 0.893 0.893 0.893 0.714 |
|
|
6 |
0.863 |
0.073 |
0.084 |
30.6 |
F/T |
= |
Found/Taken |
OE |
= |
Overall Error (±
%) | |
* Background CO is
included in ppm Taken value. Background CO concentrations in
the diluent air: |
10 ppm experiment ~ 1.6 ppm
CO |
|
5 ppm experiment ~ 0.5 ppm
CO | | |
Table 2b
Sampling and Analysis - Hand-Calculated Average
Values [50% RH and 20-25°C, Low Concn]
Monitor
|
ppm Taken*
|
ppm Found
|
F/T
|
N
|
Mean
|
Std Dev
|
CV
|
OE
|
(10 ppm)
("B" experiment) |
|
#1 #2 #3 #4 #5 #6 |
11.1 11.1 11.1 11.1 11.1 11.1 |
11.0 10.8 11.2 11.4 12.0 10.6 |
0.991 0.973 1.009 1.027 1.081 0.955 |
|
|
6 |
1.006 |
0.045 |
0.044 |
9.5 |
(5 ppm) ("B"
experiment) |
|
#1 #2 #3 #4 #5 #6 |
5.6 5.6 5.6 5.6 5.6 5.6 |
5.0 5.1 5.2 5.2 6.0 4.9 |
0.893 0.911 0.929 0.929 1.071 0.875 |
|
|
6 |
0.935 |
0.070 |
0.075 |
21.6 |
F/T |
= |
Found/Taken |
OE |
= |
Overall Error (±
%) | |
* Background levels of
CO in the generation stream are included in the Taken ppm
value. | |
Table 3
Sampling and Analysis - Hand-Calculated Average
Values [50% RH and 20-25°C, Ceiling Concn]
Monitor
|
ppm Taken
|
ppm Found
|
F/T
|
N
|
Mean
|
Std Dev
|
CV
|
OE
|
(Initial
5-min Period) |
#1 #2 #3 #4 #5 #6 |
197.8 197.8 197.8 197.8 197.8 197.8 |
195.0 191.4 181.0 198.6 185.8 189.0 |
0.986 0.968 0.915 1.004 0.939 0.956 |
|
|
6 |
0.961 |
0.032 |
0.033 |
10.5 |
(Start to
End of CO Flow, approx. 50 min) |
#1 #2 #3 #4 #5 #6 |
197.8 197.8 197.8 197.8 197.8 197.8 |
192.1 187.4 172.4 199.2 177.6 185.9 |
0.971 0.947 0.872 1.007 0.898 0.940 |
|
|
6 |
0.939 |
0.049 |
0.052 |
16.5 |
F/T |
= |
Found/Taken |
OE |
= |
Overall Error (±
%) | | |
Table 4a
Humidity Study - TWA Output Values (Truncated) [30% RH
and 20-25°C]
(OSHA TWA
PEL) |
Monitor
|
ppm Taken
|
ppm Found
|
F/T
|
N
|
Mean
|
Std Dev
|
CV
|
OE
|
(0.5 × PEL)
("B" experiment) |
#1 #2 #3 #4 #5 #6 |
19.3 19.3 19.3 19.3 19.3 19.3 |
19 19 18 19 20 17 |
0.984 0.984 0.933 0.984 1.036 0.881 |
|
|
6 |
0.967 |
0.054 |
0.055 |
14.3 |
(1 × PEL)
("B" experiment)* |
#1 #2 #3 #4 #5 #6 |
36.6 36.6 36.6 36.6 36.6 36.6 |
40 40 39 40 42 40 |
1.093 1.093 1.066 1.093 1.148 1.093 |
|
|
6 |
1.097 |
0.027 |
0.024 |
14.6 |
F/T |
= |
Found/Taken |
OE |
= |
Overall Error (±
%) |
Bias |
= |
0.032 |
CV2 (Pooled) |
= |
0.042 |
OE (Total) |
= |
±11.6% | |
* Gas flow rate (1.310
L/min) over sensor was larger than calibration gas flow rate
(0.60 L/min) (Ratio = 2.2). No compensation was made in the
results for the different
flows. | |
Table 4b
Humidity Study - Hand-Calculated Average Values [30% RH
and 20-25°C]
(OSHA TWA
PEL) |
Monitor
|
ppm Taken
|
ppm Found
|
F/T
|
N
|
Mean
|
Std Dev
|
CV
|
OE
|
(0.5 × PEL)
("B" experiment) |
#1 #2 #3 #4 #5 #6 |
19.3 19.3 19.3 19.3 19.3 19.3 |
19.2 19.1 19.0 19.1 20.1 17.7 |
0.995 0.990 0.984 0.990 1.041 0.917 |
|
|
6 |
0.986 |
0.040 |
0.040 |
9.5 |
(1 × PEL)
("B" experiment)* |
#1 #2 #3 #4 #5 #6 |
36.6 36.6 36.6 36.6 36.6 36.6 |
40.5 40.8 39.7 40.6 42.3 40.5 |
1.107 1.115 1.085 1.109 1.156 1.107 |
|
|
6 |
1.113 |
0.023 |
0.021 |
15.5 |
F/T |
= |
Found/Taken |
OE |
= |
Overall Error (±
%) |
Bias |
= |
0.050 |
CV2 (Pooled) |
= |
0.032 |
OE (Total) |
= |
±11.4% | |
* Gas flow rate (1.310
L/min) over sensor was larger than calibration gas flow rate
(0.60 L/min) (Ratio = 2.2). No compensation was made in the
results for the different
flows. | |
Table 5a
Humidity Study - TWA Output Values (Truncated) [80% RH
and 20-25°C]
(OSHA TWA
PEL) |
Monitor
|
ppm Taken
|
ppm Found
|
F/T
|
N
|
Mean
|
Std Dev
|
CV
|
OE
|
(0.5 × PEL)
("B" experiment)* |
#1 #2 #3 #4 #5 #6 |
21.2 21.2 21.2 21.2 21.2 21.2 |
24 22 24 25 25 23 |
1.132 1.038 1.132 1.179 1.179 1.085 |
|
|
6 |
1.124 |
0.055 |
0.049 |
22.2 |
(1 × PEL)
("A" experiment) |
#1 #2 #3 #4 #5 #6 |
39.9 39.9 39.9 39.9 39.9 39.9 |
43 40 41 41 42 44 |
1.078 1.003 1.028 1.028 1.053 1.103 |
|
|
6 |
1.048 |
0.037 |
0.035 |
11.9 |
(2 × PEL)
("A" experiment) |
#1 #2 #3 #4 #5 #6 |
74.4 74.4 74.4 74.4 74.4 74.4 |
82 74 74 78 75 69 |
1.102 0.995 0.995 1.048 1.008 0.927 |
|
|
6 |
1.013 |
0.059 |
0.058 |
12.8 |
F/T |
= |
Found/Taken |
OE |
= |
Overall Error (±
%) |
Bias |
= |
0.062 |
CV2 (Pooled) |
= |
0.048 |
OE (Total) |
= |
±15.8% | |
* Gas flow rate (1.340
L/min) over sensor was larger than calibration gas flow rate
(0.580 L/min) (Ratio = 2.3). No compensation was made in the
calculations for
this. | |
Table 5b
Humidity Study - Hand-Calculated Average Values [80% RH
and 20-25°C]
(OSHA TWA
PEL) |
Monitor
|
ppm Taken
|
ppm Found
|
F/T
|
N
|
Mean
|
Std Dev
|
CV
|
OE
|
(0.5 × PEL)
("B" experiment)* |
#1 #2 #3 #4 #5 #6 |
21.2 21.2 21.2 21.2 21.2 21.2 |
24.0 22.6 24.4 25.2 25.4 23.4 |
1.132 1.066 1.151 1.189 1.198 1.104 |
|
|
6 |
1.140 |
0.050 |
0.044 |
22.8 |
(1 × PEL)
("A" experiment) |
#1 #2 #3 #4 #5 #6 |
39.9 39.9 39.9 39.9 39.9 39.9 |
43.2 40.8 42.0 41.5 43.0 44.6 |
1.083 1.023 1.053 1.040 1.078 1.118 |
|
|
6 |
1.066 |
0.034 |
0.032 |
13.0 |
(2 × PEL)
("A" experiment) |
#1 #2 #3 #4 #5 #6 |
74.4 74.4 74.4 74.4 74.4 74.4 |
82.4 75.0 74.5 78.6 75.4 69.8 |
1.108 1.008 1.001 1.056 1.013 0.938 |
|
|
6 |
1.021 |
0.057 |
0.056 |
13.2 |
F/T |
= |
Found/Taken |
OE |
= |
Overall Error (±
%) |
Bias |
= |
0.075 |
CV2 (Pooled) |
= |
0.045 |
OE (Total) |
= |
±16.6% | |
* Gas flow rate (1.340
L/min) over sensor was larger than calibration gas flow rate
(0.580 L/min) (Ratio = 2.3). No compensation was made in the
calculations for
this. | |
Table 6
Humidity Study - TWA Output Values (Truncated)
|
F test
|
|
Recoveries
%
|
Level
|
F(calc)
|
F(crit)
|
|
30% RH
|
50% RH
|
80% RH
|
0.5 × PEL 1 ×
PEL 2 × PEL Average |
17.20* 26.41* 0.27 ---- |
6.36 6.36 10.04 ---- |
|
96.7 109.7** *** 103.2 |
99.4 96.1 99.7 98.4 |
112.4** 104.8 101.3 106.2 |
* |
Large values
appear to be due to variability in mean recoveries. This
variability appears related to the degree of matching
the calibration gas flow to the generation system gas
flow over the monitor sensors. |
|
If (Flow rate
generation)/(Flow rate calibration) >1, then CO concn
results will be high. |
** |
Results were
high as explained above. |
*** |
Data were not
collected for this
experiment. | | |
Table 8a
Qualitative and Quantitative Detection Limits (Truncated
TWA Output Values)
ppm
|
TWA
Values
|
Std
Dev
|
Blank 5.6 11.1 16.1 |
0 5 10 15 |
0* 5 10 16 |
0 5 11 16 |
0* 5 11 16 |
0* 5 11 17 |
0 4 10 16 |
0.000 0.408 0.548 0.632 |
* Datalogger
reported zero although a number of values of -1 were reported
for the 1-min readings. |
IUPAC Method
using the
equation: C =
k(sd)/m |
Where: |
C |
= |
the smallest
detectabel concentration an analytical instrument can
determine at a given confidence level. |
k |
= |
3 (Qualitative
detection limit, 99.86% confidence). |
|
= |
10 (Quantitative
detection limit, 99.99% confidence). |
sd |
= |
standard
deviation of blank readings. |
m |
= |
analytical
sensitivity or slope as calculated by linear
regression. | |
Note: |
For the
Dataloggers, the numerical responses (TWA values) are
nearly the same as the input values (ppm); therefore,
the slope is nearly equal to
unity. | |
Minimum
detectable signal (Qualitative detection
limit): |
C =
3(0.408)/1.0008 = 1.2 ppm |
For k = 10
(Quantitative detection limit): |
C = 4.1 ppm as a
reliable detectable
signal | | |
Table 8b
Qualitative and Quantitative Detection
Limits (Hand-Calculated Average Values)
ppm
|
TWA
Values
|
Std
Dev
|
Blank 5.6 11.1 16.1 |
0.00 5.0 11.0 15.3 |
-0.26 5.1 10.8 17.0 |
0.00 5.2 11.2 16.0 |
-0.42 5.2 11.4 16.0 |
-0.12 6.0 12.0 17.2 |
0.00 4.9 10.6 16.1 |
0.174 0.393 0.497 0.709 |
Minimum
detectable signal (Qualitative detection
limit): |
C =
3(0.174)/0.984848 = 0.5 ppm |
For k = 10
(Quantitative detection limit): |
C = 1.8 ppm as a
reliable detectable
signal | | |
|