|
Method no.: |
ID-189 |
|
|
Matrix: |
Air |
|
|
OSHA Permissible Exposure
Limits: |
5 µg/m3 (TWA) 2.5 µg/m3 (Action Level TWA) |
|
|
Collection Procedure: |
A known volume of air is drawn
through a 37-mm diameter filter cassette
containing a 0.8-µm
mixed cellulose ester membrane filter (MCEF). |
|
|
Recommended Air Volume: |
960 L |
|
|
Recommended Sampling Rate: |
2.0 L/min |
|
|
Analytical
Procedure: |
Air filter samples are digested
with nitric acid. After digestion, a small amount of
hydrochloric acid is added. The samples are then diluted to
volume with deionized water and analyzed by either flame
atomic absorption spectroscopy (AAS) or flameless atomic
absorption spectroscopy using a heated graphite furnace
atomizer(AAS-HGA). |
|
|
Detection
Limits: Qualitative: |
0.2 µg/m3 for a 200 L sample by
Flame AAS 0.007 µg/m3
for a 60 L sample by AAS-HGA |
|
|
Quantitative: |
0.70 µg/m3 for a 200 L sample by
Flame AAS 0.025 µg/m3
for a 60 L sample by AAS-HGA |
|
|
Precision and Accuracy: |
Flame AAS Analysis
AAS-HGA
Analysis |
|
|
Validation Level: |
2.5 to 10 µg/m3 for a 400 L air
vol |
1.25 to 5.0 µg/m3 for a 60 L air
vol | |
|
|
CV1(pooled): |
|
|
|
Analytical Bias: |
|
|
|
Overall Analytical Error: |
|
|
|
Method Classification: |
Validated |
|
|
Date: |
June, 1992 |
|
|
Inorganic Service Branch II OSHA Salt Lake
Technical Center Salt Lake City, Utah
Commercial manufacturers and products mentioned in this
method are for descriptive use only and do not constitute
endorsements by USDOL-OSHA. Similar products from other
sources can be substituted.
|
1.Introduction
1.1. Scope
This
method describes the collection of airborne elemental
cadmium and cadmium compounds on 0.8-µm mixed cellulose ester membrane
filters and their subsequent analysis by either flame atomic
absorption spectroscopy (AAS) or flameless atomic absorption
spectroscopy using a heated graphite furnace atomizer
(AAS-HGA). It is applicable for both TWA and
Action Level TWA Permissible Exposure Level (PEL)
measurements. The two atomic absorption analytical
techniques included in the method do not differentiate
between cadmium fume and cadmium dust samples. They also do
not differentiate between elemental cadmium and its
compounds.
1.2. Principle
Airborne elemental
cadmium and cadmium compounds are collected on a 0.8-µm mixed cellulose ester membrane filter
(MCEF). The air filter samples are digested with
concentrated nitric acid to destroy the organic matrix and
dissolve the cadmium analytes. After digestion, a small
amount of concentrated hydrochloric acid is added to help
dissolve other metals which may be present. The samples are
diluted to volume with deionized water and then aspirated
into the oxidizing air/acetylene flame of an atomic
absorption spectrophotometer for analysis of elemental
cadmium.
If the concentration of cadmium in a sample
solution is too low for quantitation by this flame AAS
analytical technique, and the sample is to be averaged with
other samples for TWA calculations, aliquots of the sample
and a matrix modifier are later injected onto a L'vov
platform in a pyrolytically-coated graphite tube of a Zeeman
atomic absorption spectrophotometer/graphite furnace
assembly for analysis of elemental cadmium. The matrix
modifier is added to stabilize the cadmium metal and
minimize sodium chloride as an interference during the high
temperature charring step of the analysis (5.1.,
5.2.)
.
1.3. History
Previously, two OSHA sampling
and analytical methods for cadmium were used concurrently
(5.3.,
5.4.).
Both of these methods also required 0.8-µm mixed cellulose ester membrane
filters for the collection of air samples. These cadmium air
filter samples were analyzed by either flame atomic
absorption spectroscopy (5.3.)
or inductively coupled plasma/atomic emission spectroscopy
(ICP-AES) (5.4.).
Neither of these two analytical methods have adequate
sensitivity for measuring workplace exposure to airborne
cadmium at the new lower TWA and Action Level TWA PEL levels
when consecutive samples are taken on one employee and the
sample results need to be averaged with other samples to
determine a single TWA.
The inclusion of two atomic
absorption analytical techniques in the new sampling and
analysis method for airborne cadmium permits quantitation of
sample results over a broad range of exposure levels and
sampling periods. The flame AAS analytical technique
included in this method is similar to the previous procedure
given in the General Metals Method ID-121
(5.3.)
with some modifications. The sensitivity of the
AAS-HGA analytical technique included in this
method is adequate to measure exposure levels at 1/10 the
Action Level TWA, or lower, when less than full-shift
samples need to be averaged together.
1.4.
Properties (5.5.)
Elemental cadmium is a silver-white, blue-tinged,
lustrous metal which is easily cut with a knife. It is
slowly oxidized by moist air to form cadmium oxide. It is
insoluble in water, but reacts readily with dilute nitric
acid. Some of the physical properties and other descriptive
information of elemental cadmium are given below:
CAS No. |
7440-43-9 |
Atomic Number |
48 |
Atomic Symbol |
Cd |
Atomic Weight |
112.41 |
Melting Point |
321°C |
Boiling Point |
765°C |
Density |
8.65 g/mL
(25°C) |
The properties of specific cadmium compounds
are described in reference 5.5.
1.5. Method Performance
A synopsis of method
performance is presented below. Further information can be
found in Section
4.
1.5.1. The qualitative and quantitative
detection limits for the flame AAS analytical technique
are 0.04 µg (0.004 µg/mL) and 0.14 µg (0.014 µg/mL) cadmium, respectively, for a 10
mL solution volume. These correspond, respectively, to 0.2
µg/m3 and 0.70 µg/m3 for a 200 L air
volume.
1.5.2. The qualitative and quantitative
detection limits for the AAS-HGA analytical
technique are 0.44 ng (0.044 ng/mL) and 1.5 ng (0.15
ng/mL) cadmium, respectively, for a 10 mL solution volume.
These correspond, respectively, to 0.007 µg/m3 and 0.025 µg/m3 for a 60 L air
volume.
1.5.3. The average recovery by the flame
AAS analytical technique of 17 spiked MCEF samples
containing cadmium in the range of 0.5 to 2.0 times the
TWA target concentration of 5 µg/m3 (assuming a 400 L air
volume) was 104.0% with a pooled coefficient of variation
(CV1) of 0.010. The flame analytical technique
exhibited a positive bias of +4.0% for the validated
concentration range. The overall analytical error (OAE)
for the flame AAS analytical technique was ±6.0%.
1.5.4. The average recovery by the
AAS-HGA analytical technique of 18 spiked
MCEF samples containing cadmium in the range of 0.5 to 2.0
times the Action Level TWA target concentration of 2.5
µg/m3 (assuming a 60
L air volume) was 94.2% with a pooled coefficient of
variation (CV1) of 0.043. The
AAS-HGA analytical technique exhibited a
negative bias of -5.8% for the validated concentration
range. The overall analytical error (OAE) for the
AAS-HGA analytical technique was ±14.2%.
1.5.5. Sensitivity in flame
atomic absorption is defined as the characteristic
concentration of an element required to produce a signal
of 1% absorbance (0.0044 absorbance units). Sensitivity
values are listed for each element by the atomic
absorption spectrophotometer manufacturer and have proved
to be a very valuable diagnostic tool to determine if
instrumental parameters are optimized and if the
instrument is performing up to specification. The
sensitivity of the spectrophotometer used in the
validation of the flame AAS analytical technique agreed
with the manufacturer specifications (5.6.);
the 2 µg/mL cadmium standard
gave an absorbance reading of 0.350 abs. units.
1.5.6. Sensitivity in graphite furnace
atomic absorption is defined in terms of the
characteristic mass, the number of picograms required to
give an integrated absorbance value of 0.0044
absorbance-second (5.7.).
Data suggests that under Stabilized Temperature Platform
Furnace (STPF) conditions (see Section
1.6.2.), characteristic mass values are transferable
between properly functioning instruments to an accuracy of
about 20% (5.2.).
The characteristic mass for STPF analysis of cadmium with
Zeeman background correction listed by the manufacturer of
the instrument used in the validation of the
AAS-HGA analytical technique was 0.35 pg. The
experimental characteristic mass value observed during the
determination of the working range and detection limits of
the AAS-HGA analytical technique was 0.41
pg. 1.6. Interferences
1.6.1. High concentrations of silicate
interfere in determining cadmium by flame AAS (5.6.).
However, silicates are not significantly soluble in the
acid matrix used to prepare the samples.
1.6.2. Interferences, such as background
absorption, are reduced to a minimum in the
AAS-HGA analytical technique by taking full
advantage of the Stabilized Temperature Platform Furnace
(STPF) concept. STPF includes all of the following
parameters (5.2.):
- Integrated Absorbance,
- Fast Instrument Electronics and Sampling Frequency,
- Background Correction,
- Maximum Power Heating,
- Atomization off the L'vov platform in a
pyrolytically coated graphite tube,
- Gas Stop during Atomization,
- Use of Matrix Modifiers.
1.7. Toxicology (5.14.)
Information listed within this section
is synopsis of current knowledge of the physiological
effects of cadmium and is not intended to be used as the
basis for OSHA policy.
IARC classifies cadmium and certain of
its compounds as Group 2A carcinogens (probably carcinogenic
to humans). Cadmium fume is intensely irritating to the
respiratory tract. Workplace exposure to cadmium can cause
both chronic and acute effects. Acute effects include
tracheobronchitis, pneumonitis, and pulmonary edema. Chronic
effects include anemia, rhinitis/anosmia, pulmonary
emphysema, proteinuria and lung cancer. The primary target
organs for chronic disease are the kidneys
(non-carcinogenic) and the lungs
(carcinogenic). 2. Sampling
2.1. Apparatus
2.1.1. Filter cassette unit for air sampling:
A 37-mm diameter mixed cellulose ester membrane filter
with a pore size of 0.8-µm
contained in a 37-mm polystyrene two- or three-piece
cassette filter holder (part no. MAWP 037 A0, Millipore
Corp., Bedford, MA). The filter is supported with a
cellulose backup pad. The cassette is sealed prior to use
with a shrinkable gel band.
2.1.2. A calibrated
personal sampling pump whose flow is determined to an
accuracy of ±5% at the recommended flow rate with the
filter cassette unit in line. 2.2. Procedure
2.2.1. Attach the prepared cassette to the
calibrated sampling pump (the backup pad should face the
pump) using flexible tubing. Place the sampling device on
the employee such that air is sampled from the breathing
zone.
2.2.2. Collect air samples at a flow rate of
2.0 L/min. If the filter does not become overloaded, a
full-shift (at least seven hours) sample is strongly
recommended for TWA and Action Level TWA measurements with
a maximum air volume of 960 L. If overloading occurs,
collect consecutive air samples for shorter sampling
periods to cover the full workshift.
2.2.3.
Replace the end plugs into the filter cassettes
immediately after sampling. Record the sampling
conditions.
2.2.4. Securely wrap each sample
filter cassette end-to-end with an OSHA Form 21 sample
seal.
2.2.5. Submit at least one blank sample with
each set of air samples. The blank sample should be
handled the same as the other samples except that no air
is drawn through it.
2.2.6. Ship the samples to
the laboratory for analysis as soon as possible in a
suitable container designed to prevent damage in transit.
3.
Analysis
3.1. Safety Precautions
3.1.1. Wear safety glasses, protective
clothing and gloves at all times.
3.1.2. Handle
acid solutions with care. Handle all cadmium samples and
solutions with extra care (see Section
1.7.). Avoid their direct contact with work area
surfaces, eyes, skin and clothes. Flush acid solutions
which contact the skin or eyes with copious amounts of
water.
3.1.3. Perform all acid digestions and acid
dilutions in an exhaust hood while wearing a face shield.
To avoid exposure to acid vapors, do not remove beakers
containing concentrated acid solutions from the exhaust
hood until they have returned to room temperature and have
been diluted or emptied.
3.1.4. Exercise care when
using laboratory glassware. Do not use chipped pipets,
volumetric flasks, beakers or any glassware with sharp
edges exposed in order to avoid the possibility of cuts or
abrasions.
3.1.5. Never pipet by mouth.
3.1.6. Refer to the instrument instruction manuals
and SOPs (5.8.,
5.9.)
for proper and safe operation of the atomic absorption
spectrophotometer, graphite furnace atomizer and
associated equipment.
3.1.7. Because metallic
elements and other toxic substances are vaporized during
AAS flame or graphite furnace atomizer operation, it is
imperative that an exhaust vent be used. Always ensure
that the exhaust system is operating properly during
instrument use. 3.2. Apparatus for Sample and
Standard Preparation
3.2.1. Hot plate, capable of reaching 150°C,
installed in an exhaust hood.
3.2.2. Phillips
beakers, 125 mL.
3.2.3. Bottles, narrow-mouth,
polyethylene or glass with leakproof caps: used for
storage of standards and matrix modifier.
3.2.4.
Volumetric flasks, volumetric pipets, beakers and other
associated general laboratory glassware.
3.2.5.
Forceps and other associated general laboratory equipment.
3.3. Apparatus for Flame AAS Analysis
3.3.1. Atomic absorption spectrophotometer
consisting of a(an):
Nebulizer and burner head.
Pressure regulating devices capable of maintaining
constant oxidant and fuel pressures.
Optical
system capable of isolating the desired wavelength of
radiation (228.8 nm).
Adjustable slit.
Light measuring and amplifying device.
Display, strip chart, or computer interface for
indicating the amount of absorbed radiation.
Cadmium hollow cathode lamp or electrodeless
discharge lamp (EDL) and power supply.
3.3.2.
Oxidant: compressed air, filtered to remove water, oil and
other foreign substances.
3.3.3. Fuel: standard
commercially available tanks of acetylene dissolved in
acetone; tanks should be equipped with flash arresters.
CAUTION: Do not use grades of acetylene
containing solvents other than acetone because they may
damage the PVC tubing used in some instruments.
3.3.4. Pressure-reducing valves: two gauge,
two-stage pressure regulators to maintain fuel and oxidant
pressures somewhat higher than the controlled operating
pressures of the instrument.
3.3.5. Exhaust vent
installed directly above the spectrophotometer burner
head. 3.4. Apparatus for AAS-HGA
Analysis
3.4.1. Atomic absorption spectrophotometer
consisting of a(an):
Heated graphite furnace
atomizer (HGA) with argon purge system.
Pressure-regulating devices capable of maintaining
constant argon purge pressure.
Optical system
capable of isolating the desired wavelength of radiation
(228.8 nm).
Adjustable slit.
Light
measuring and amplifying device.
Display, strip
chart, or computer interface for indicating the amount of
absorbed radiation (as integrated absorbance, peak area).
Background corrector: Zeeman or deuterium arc. The
Zeeman background corrector is recommended.
Cadmium hollow cathode lamp or electrodeless
discharge lamp (EDL) and power supply.
Autosampler
capable of accurately injecting 5 to 20 µL sample aliquots onto the L'vov
Platform in a graphite tube.
3.4.2.
Pyrolytically-coated graphite tubes containing solid,
pyrolytic L'vov platforms.
3.4.3. Polyethylene
sample cups, 2.0-2.5 mL, for use with the autosampler.
3.4.4. Inert purge gas for graphite furnace
atomizer: compressed gas cylinder of purified argon.
3.4.5. Two gauge, two-stage pressure regulator for
the argon gas cylinder.
3.4.6. Cooling water
supply for graphite furnace atomizer.
3.4.7.
Exhaust vent installed directly above the graphite furnace
atomizer. 3.5. Reagents
All reagents
should be ACS analytical reagent grade or better.
3.5.1. Deionized water with a specific
conductance of less than 10 µS.
3.5.2. Concentrated nitric acid, HNO3.
3.5.3. Concentrated hydrochloric acid, HCl.
3.5.4. Ammonium phosphate, monobasic,
NH4H2PO4.
3.5.5.
Magnesium nitrate, Hexahydrate,
Mg(NO3)2 · 6H2O.
3.5.6. Diluting solution (4%
HNO3, 0.4% HCl): Add 40 mL HNO3 and
4 mL HCl carefully to approximately 500 mL deionized water
and dilute to 1 L with deionized water.
3.5.7.
Cadmium standard stock solution, 1,000 µg/mL: Use a commercially available
certified 1,000 µg/mL cadmium
standard or, alternatively, dissolve 1.0000 g of cadmium
metal in a minimum volume of 1:1 HCl and dilute to 1 L
with 4% HNO3. Observe expiration dates of
commercial standards. Properly dispose of commercial
standards with no expiration dates or prepared standards
one year after their receipt or preparation date.
3.5.8. Matrix modifier for AAS-HGA
analysis: Dissolve 10.0 g NH4H2
PO4 and 2.6 g
Mg(NO3)2 · 6H2O in
approximately 200 mL deionized water. Add 1 mL
HNO3 and dilute to 500 mL with deionized water.
3.5.9 Nitric Acid, 1:1 HNO3/DI
H2O mixture: Carefully add a measured volume of
concentrated HNO3 to an equal volume of DI
H2O.
3.5.10. Nitric acid, 10% v/v:
Carefully add 100 mL of concentrated HNO3 to
500 mL of DI H 2O and dilute to 1 L.
3.6. Glassware Preparation
3.6.1. Clean Phillips beakers by refluxing
with 1:1 nitric acid on a hot plate in a fume hood.
Thoroughly rinse with deionized water and invert the
beakers to allow them to drain dry.
3.6.2. Rinse
volumetric flasks and all other glassware with 10% nitric
acid and deionized water prior to use. 3.7.
Standard Preparation for Flame AAS Analysis
3.7.1. Dilute stock solutions:
Prepare
1, 5, 10 and 100 µg/mL cadmium
standard stock solutions by making appropriate serial
dilutions of 1,000 µg/mL cadmium
standard stock solution with the diluting solution
described in Section
3.5.6.
3.7.2. Working
standards:
Prepare cadmium working standards
in the range of 0.02 to 2.0 µg/mL by making appropriate serial
dilutions of the dilute stock solutions with the same
diluting solution. A suggested method of preparation of
the working standards is given below.
|
Working
Standard (µg/mL)
|
Std
Solution (µg/mL)
|
Aliquot (mL)
|
Final
Vol (mL)
|
0.02 |
1 |
10 |
500 |
0.05 |
5 |
5 |
500 |
0.1 |
10 |
5 |
500 |
0.2 |
10 |
10 |
500 |
0.5 |
10 |
25 |
500 |
1 |
100 |
5 |
500 |
2 |
100 |
10 |
500 |
| Store the
working standards in 500-mL, narrow-mouth polyethylene or
glass bottles with leak proof caps. Prepare every twelve
months. 3.8.
Standard Preparation for AAS-HGA Analysis
3.8.1. Dilute stock solutions:
Prepare
10, 100 and 1,000 ng/mL cadmium standard stock solutions
by making appropriate ten-fold serial dilutions of the
1,000 µg/mL cadmium standard
stock solution with the diluting solution described in Section
3.5.6.
3.8.2. Working
standards:
Prepare cadmium working standards
in the range of 0.2 to 20 ng/mL by making appropriate
serial dilutions of the dilute stock solutions with the
same diluting solution. A suggested method of preparation
of the working standards is given below.
|
Working
Standard (ng/mL)
|
Std
Solution (ng/mL)
|
Aliquot (mL)
|
Final
Vol (mL)
|
0.2 |
10 |
2 |
100 |
0.5 |
10 |
5 |
100 |
1 |
10 |
10 |
100 |
2 |
100 |
2 |
100 |
5 |
100 |
5 |
100 |
10 |
100 |
10 |
100 |
20 |
1,000 |
2 |
100 |
| Store the
working standards in narrow-mouth polyethylene or glass
bottles with leakproof caps. Prepare monthly.
3.9. Sample
Preparation
3.9.1. Carefully transfer each sample filter
with forceps from its filter cassette unit to a clean,
separate 125-mL Phillips beaker along with any loose dust
found in the cassette. Label each Phillips beaker with the
appropriate sample number.
3.9.2. Digest the
sample by adding 5 mL of concentrated nitric acid
(HNO3) to each Phillips beaker containing an
air filter sample. Place the Phillips beakers on a hot
plate in an exhaust hood and heat the samples until
approximately 0.5 mL remains. The sample solution in each
Phillips beaker should become clear. If it is not clear,
digest the sample with another portion of concentrated
nitric acid.
3.9.3. After completing the
HNO3 digestion and cooling the samples, add 40
µL (2 drops) of concentrated HCl
to each air sample solution and then swirl the contents.
Carefully add about 5 mL of deionized water by pouring it
down the inside of each beaker.
3.9.4.
Quantitatively transfer each cooled air sample solution
from each Phillips beaker to a clean 10-mL volumetric
flask. Dilute each flask to volume with deionized water
and mix well. 3.10. Flame
AAS Analysis
Analyze all of the air samples for
their cadmium content by flame atomic absorption
spectroscopy (AAS) according to the instructions given
below.
3.10.1. Set up the atomic absorption
spectrophotometer for the air/acetylene flame analysis of
cadmium according to the SOP (5.8.)
or the manufacturer's operational instructions. For the
source lamp, use the cadmium hollow cathode or
electrodeless discharge lamp operated at the
manufacturer's recommended rating for continuous
operation. Allow the lamp to warm up 10 to 20 min or until
the energy output stabilizes. Optimize conditions such as
lamp position, burner head alignment, fuel and oxidant
flow rates, etc. See the SOP or specific instrument
manuals for details. Instrumental parameters for the
Perkin-Elmer Model 603 used in the validation of this
method are given in Attachment
1.
3.10.2. Aspirate and measure the absorbance
of a standard solution of cadmium. The standard
concentration should be within the linear range. For the
instrumentation used in the validation of this method a 2
µg/mL cadmium standard gives a
net absorbance reading of about 0.350 abs. units (see Section
1.5.5.) when the instrument and the source lamp are
performing to manufacturer specifications.
3.10.3.
To increase instrument response, scale expand the
absorbance reading of the aspirated 2 µg/mL working standard approximately
four times. Increase the integration time to at least 3
seconds to reduce signal noise.
3.10.4. Autozero
the instrument while aspirating a deionized water blank.
Monitor the variation in the baseline absorbance reading
(baseline noise) for a few minutes to insure that the
instrument, source lamp and associated equipment are in
good operating condition.
3.10.5. Aspirate the
working standards and samples directly into the flame and
record their absorbance readings. Aspirate the deionized
water blank immediately after every standard or sample to
correct for and monitor any baseline drift and noise.
Record the baseline absorbance reading of each deionized
water blank. Label each standard and sample reading and
its accompanying baseline reading.
3.10.6. It is
recommended that the entire series of working standards be
analyzed at the beginning and end of the analysis of a set
of samples to establish a
concentration-response curve, ensure that the
standard readings agree with each other and are
reproducible. Also, analyze a working standard after every
five or six samples to monitor the performance of the
spectrophotometer. Standard readings should agree within
±10 to 15% of the readings obtained at the beginning of
the analysis.
3.10.7. Bracket the sample readings
with standards during the analysis. If the absorbance
reading of a sample is above the absorbance reading of the
highest working standard, dilute the sample with diluting
solution and reanalyze. Use the appropriate dilution
factor in the calculations.
3.10.8. Repeat the
analysis of approximately 10% of the samples for a check
of precision.
3.10.9. If possible, analyze quality
control samples from an independent source as a check on
analytical recovery and precision.
3.10.10. Record
the final instrument settings at the end of the analysis.
Date and label the output. 3.11. AAS-HGA Analysis
Initially analyze all of the air samples for their
cadmium content by flame atomic absorption spectroscopy
(AAS) according to the instructions given in Section
3.10. If the concentration of cadmium in a sample
solution is less than three times the quantitative detection
limit [0.04 µg/mL (40 ng/mL) for
the instrumentation used in the validation] and the sample results are to be averaged with
other samples for TWA calculations, proceed with the
AAS-HGA analysis of the sample as described
below.
3.11.1. Set up the atomic absorption
spectrophotometer and HGA for flameless atomic absorption
analysis of cadmium according to the SOP (5.9.)
or the manufacturer's operational instructions and allow
the instrument to stabilize. The graphite furnace atomizer
is equipped with a pyrolytically coated graphite tube
containing a pyrolytic platform. For the source lamp, use
a cadmium hollow cathode or electrodeless discharge lamp
operated at the manufacturer's recommended setting for
graphite furnace operation. The Zeeman background
corrector and EDL are recommended for use with the L'vov
platform. Instrumental parameters for the Perkin-Elmer
Model 5100 spectrophotometer and Zeeman HGA-600 graphite
furnace used in the validation of this method are given in
Attachment
2.
3.11.2. Optimize the energy reading of the
spectrophotometer at 228.8 nm by adjusting the lamp
position and the wavelength according to the
manufacturer's instructions.
3.11.3. Set up the
autosampler to inject a 5-µL
aliquot of the working standard, sample or reagent blank
solution onto the L'vov platform along with a 10-µL overlay of the matrix modifier.
3.11.4. Analyze the reagent blank (diluting
solution, Section
3.5.6.) and then autozero the instrument before
starting the analysis of a set of samples. It is
recommended that the reagent blank be analyzed several
times during the analysis to assure the integrated
absorbance (peak area) reading remains at or near zero.
3.11.5. Analyze a working standard approximately
midway in the linear portion of the working standard range
two or three times to check for reproducibility and
sensitivity (see Sections 1.5.5.
and 1.5.6.)
before starting the analysis of samples. Calculate the
experimental characteristic mass value from the average
integrated absorbance reading and injection volume of the
analyzed working standard. Compare this value to the
manufacturer's suggested value as a check of proper
instrument operation.
3.11.6. Analyze the reagent
blank, working standard, and sample solutions. Record and
label the peak area (abs-sec) readings and the peak and
background peak profiles on the printer/plotter.
3.11.7. It is recommended the entire series of
working standards be analyzed at the beginning and end of
the analysis of a set of samples. Establish a
concentration-response curve and ensure
standard readings agree with each other and are
reproducible. Also, analyze a working standard after every
five or six samples to monitor the performance of the
system. Standard readings should agree within ±15% of the
readings obtained at the beginning of the analysis.
3.11.8. Bracket the sample readings with standards
during the analysis. If the peak area reading of a sample
is above the peak area reading of the highest working
standard, dilute the sample with the diluting solution and
reanalyze. Use the appropriate dilution factor in the
calculations.
3.11.9. Repeat the analysis of
approximately 10% of the samples for a check of precision.
3.11.10. If possible, analyze quality control
samples from an independent source as a check of
analytical recovery and precision.
3.11.11. Record
the final instrument settings at the end of the analysis.
Date and label the output. 3.12. Calculations
Note: Standards used for HGA analysis are in ng/mL.
Total amounts of cadmium from calculations will be in ng
(not µg) unless a prior conversion
is made.
3.12.1. Correct for baseline drift and noise
in flame AAS analysis by subtracting each baseline
absorbance reading from its corresponding working standard
or sample absorbance reading to obtain the net absorbance
reading for each standard and sample.
3.12.2. Use
a least squares regression program to plot a
concentration-response curve of net
absorbance reading (or peak area for HGA analysis) versus
concentration (µg/mL or ng/mL)
of cadmium in each working standard.
3.12.3.
Determine the concentration (µg/mL or ng/mL) of cadmium in each
sample from the resulting
concentration-response curve. If the
concentration of cadmium in a sample solution is less than
three times the quantitative detection limit [0.04 µg/mL (40 ng/mL) for the
instrumentation used in the validation of the method]
and if consecutive samples were taken on
one employee and the sample results are to be averaged
with other samples to determine a single TWA,
reanalyze the sample by AAS-HGA as described
in Section
3.11. and report the AAS-HGA analytical
results.
3.12.4. Calculate the total amount (µg or ng) of cadmium in each sample
from the sample solution volume (mL) :
W =
(C)(sample vol, mL)(DF)
Where: |
W |
= |
Total cadmium in sample |
|
C |
= |
Calculated concentration of
cadmium |
|
DF |
= |
Dilution Factor (if
applicable) |
3.12.5. Make a blank correction for each air
sample by subtracting the total amount of cadmium in the
corresponding blank sample from the total amount of
cadmium in the sample.
3.12.6. Calculate the
concentration of cadmium in an air sample
(mg/m3 or µg/m3) by using one of the
following equations:
mg/m3 =
Wbc/(Air vol sampled, L)
or
µg/m3 =
(Wbc)(1,000 ng/ µg)/(Air vol sampled, L)
Where: Wbc = blank corrected total
µg cadmium in the sample. (1
µg = 1,000 ng)
4. Backup Data
4.1. Introduction
4.1.1. The purpose of this
evaluation is to determine the analytical method recovery,
working standard range, and qualitative and quantitative
detection limits of the two atomic absorption analytical
techniques included in this method. The evaluation
consisted of the following experiments:
- An analysis of 24 samples (six samples each at 0.1,
0.5, 1 and 2 times the TWA-PEL) for the analytical
method recovery study of the flame AAS analytical
technique.
- An analysis of 18 samples (six samples each at 0.5,
1 and 2 times the Action Level TWA-PEL) for the
analytical method recovery study of the
AAS-HGA analytical technique.
- Multiple analyses of the reagent blank and a series
of standard solutions to determine the working standard
range and the qualitative and quantitative detection
limits for both atomic absorption analytical techniques.
4.1.2. The analytical
method recovery results at all test levels were calculated
from concentration-response curves and
statistically examined for outliers at the 99% confidence
level. Possible outliers were determined using the
Treatment of Outliers test (5.10.).
In addition, the sample results of the two analytical
techniques, at 0.5, 1.0 and 2.0 times their target
concentrations, were tested for homogeneity of variances
also at the 99% confidence level. Homogeneity of the
coefficients of variation was determined using the
Bartlett's test (5.11.).
The overall analytical error (OAE) at the 95% confidence
level was calculated using the equation (5.12.):
OAE = ±[|Bias| +
(1.96)(CV1(pooled))(100%)]
4.1.3. A derivation of the International
Union of Pure and Applied Chemistry (IUPAC) detection
limit equation (5.13.)
was used to determine the qualitative and quantitative
detection limits for both atomic absorption analytical
techniques:
Cld =
k(sd)/m |
(Equation
1) |
Where: |
|
Cld |
= |
the smallest reliable detectable
concentration an analytical instrument can determine
at a given confidence level. |
k |
= |
3 for the Qualitative Detection
Limit at the 99.86% Confidence Level |
|
= |
10 for the Quantitative Detection
Limit at the 99.99% Confidence Level. |
sd |
= |
standard deviation of the reagent
blank (Rbl) readings. |
m |
= |
analytical sensitivity or slope as
calculated by linear
regression. |
4.1.4. Collection efficiencies of metallic
fume and dust atmospheres on 0.8-µm mixed cellulose ester membrane
filters are well documented and have been shown to be
excellent (5.11.).
Since elemental cadmium and the cadmium component of
cadmium compounds are nonvolatile, stability studies of
cadmium spiked MCEF samples were not performed.
4.2. Equipment
4.2.1. A Perkin-Elmer (PE) Model 603
spectrophotometer equipped with a manual gas control
system, a stainless steel nebulizer, a burner mixing
chamber, a flow spoiler and a 10 cm. (one-slot) burner
head was used in the experimental validation of the flame
AAS analytical technique. A PE cadmium hollow cathode
lamp, operated at the manufacturer's recommended current
setting for continuous operation (4 mA), was used as the
source lamp. Instrument parameters are listed in Attachment
1.
4.2.2. A PE Model 5100 spectrophotometer,
Zeeman HGA-600 graphite furnace atomizer and AS-60 HGA
autosampler were used in the experimental validation of
the AAS-HGA analytical technique. The
spectrophotometer was equipped with a PE Series 7700
professional computer and Model PR-310 printer. A PE
System 2 cadmium electrodeless discharge lamp, operated at
the manufacturer's recommended current setting for
modulated operation (170 mA), was used as the source lamp.
Instrument parameters are listed in Attachment
2. 4.3. Reagents
4.3.1 J.T. Baker Chem. Co. (Analyzed grade)
concentrated nitric acid, 69.0-71.0%, and concentrated
hydrochloric acid, 36.5-38.0%, were used to prepare the
samples and standards.
4.3.2. Ammonium phosphate,
monobasic, NH4H2PO4 and
magnesium nitrate, Mg(NO3)2
· 6H2O, both
manufactured by the Mallinckrodt Chem. Co., were used to
prepare the matrix modifier for AAS-HGA
analysis. 4.4. Standard Preparation for Flame
AAS Analysis
4.4.1. Dilute stock solutions:
Prepared 0.01, 0.1, 1, 10 and 100 µg/mL cadmium standard stock solutions
by making appropriate serial dilutions of a commercially
available 1,000 µg/mL cadmium
standard stock solution (RICCA Chemical Co., Lot# A102)
with the diluting solution (4% HNO3, 0.4% HCl).
4.4.2. Analyzed Standards:
Prepared
cadmium standards in the range of 0.001 to 2.0 µg/mL by pipetting 2 to 10 mL of the
appropriate dilute cadmium stock solution into a 100-mL
volumetric flask and diluting to volume with the diluting
solution. (See Section
3.7.2.) 4.5. Standard Preparation for
AAS-HGA Analysis
4.5.1. Dilute stock
solutions:
Prepared 1, 10, 100 and 1,000 ng/mL
cadmium standard stock solutions by making appropriate
serial dilutions of a commercially available 1,000 µg/mL cadmium standard stock solution
(J.T. Baker Chemical Co., Instra-analyzed, Lot# D22642)
with the diluting solution (4% HNO3, 0.4% HCl).
4.5.2. Analyzed Standards:
Prepared
cadmium standards in the range of 0.1 to 40 ng/mL by
pipetting 2 to 10 mL of the appropriate dilute cadmium
stock solution into a 100-mL volumetric flask and diluting
to volume with the diluting solution. (See Section
3.8.2.) 4.6.
Detection Limits and Standard Working Range for Flame AAS
Analysis
4.6.1. Analyzed the
reagent blank solution and the entire series of cadmium
standards in the range of 0.001 to 2.0 µg/mL three to six times according to
the instructions given in Section
3.10. The diluting solution (4% HNO3, 0.4%
HCl) was used as the reagent blank. The integration time
on the PE 603 spectrophotometer was set to 3.0 seconds and
a four-fold expansion of the absorbance reading of the 2.0
µg/mL cadmium standard was made
prior to analysis. The 2.0 µg/mL
standard gave a net absorbance reading of 0.350 abs. units
prior to expansion in agreement with the manufacturer's
specifications (5.6.).
4.6.2. The net absorbance readings of the reagent
blank and the low concentration Cd standards from 0.001 to
0.1 µg/mL and the statistical
analysis of the results are shown in Table
I. The standard deviation, sd, of the six net
absorbance readings of the reagent blank is 1.05 abs.
units. The slope, m, as calculated by a linear regression
plot of the net absorbance readings (shown in Table
II) of the 0.02 to 1.0 µg/mL
cadmium standards versus their concentration is 772.7 abs.
units/(µg/mL).
4.6.3. If
these values for sd and the slope, m, are used in Eqn.
1 ( Section
4.1.3.), the qualitative and quantitative detection
limits as determined by the IUPAC Method are:
Cld |
= = |
(3)(1.05 abs. units)/(772.7 abs.
units/(µg/mL)) 0.0041
µg/mL for the qualitative
detection limit. |
|
Cld |
= = |
(10)(1.05 abs. units)/(772.7 abs.
units/µg/mL)) 0.014
µg/mL for the quantitative
detection limit. |
The qualitative and quantitative detection
limits for the flame AAS analytical technique are 0.041
µg and 0.14 µg cadmium, respectively, for a 10 mL
solution volume. These correspond, respectively, to 0.2
µg/m3 and 0.70 µg/m3 for a 200 L air
volume.
4.6.4. The recommended Cd standard working
range for flame AAS analysis is 0.02 to 2.0 µg/mL. The net absorbance readings of
the reagent blank and the recommended working range
standards and the statistical analysis of the results are
shown in Table
II. The standard of lowest concentration in the
working range, 0.02 µg/mL, is
slightly greater than the calculated quantitative
detection limit, 0.014 µg/mL.
The standard of highest concentration in the working
range, 2.0 µg/mL, is at the
upper end of the linear working range suggested by the
manufacturer (5.6..
Although the standard net absorbance readings are not
strictly linear at concentrations above 0.5 µg/mL, the deviation from linearity is
only about 10% at the upper end of the recommended
standard working range. The deviation from linearity is
probably caused by the four-fold expansion of the signal
suggested in the method. As shown in Table
II, the precision of the standard net absorbance
readings are excellent throughout the recommended working
range; the relative standard deviations of the readings
range from 0.009 to 0.064. 4.7. Detection Limits and Standard Working
Range for AAS-HGA Analysis
4.7.1. Analyzed the
reagent blank solution and the entire series of cadmium
standards in the range of 0.1 to 40 ng/mL according to the
instructions given in Section
3.11. The diluting solution (4% HNO3, 0.4%
HCl) was used as the reagent blank. A fresh aliquot of the
reagent blank and of each standard was used for every
analysis. The experimental characteristic mass value was
0.41 pg, calculated from the average peak area (abs-sec)
reading of the 5 ng/mL standard which is approximately
midway in the linear portion of the working standard
range. This agreed within 20% with the characteristic mass
value, 0.35 pg, listed by the manufacturer of the
instrument (5.2.).
4.7.2. The peak area (abs-sec) readings of the
reagent blank and the low concentration Cd standards from
0.1 to 2.0 ng/mL and statistical analysis of the results
are shown in Table
III. Five of the reagent blank peak area readings were
zero and the sixth reading was 1 and was an outlier. The
near lack of a blank signal does not satisfy a strict
interpretation of the IUPAC method for determining the
detection limits. Therefore, the standard deviation of the
six peak area readings of the 0.2 ng/mL cadmium standard,
0.75 abs-sec, was used to calculate the detection limits
by the IUPAC method. The slope, m, as calculated by a
linear regression plot of the peak area (abs-sec) readings
(shown in Table
IV) of the 0.2 to 10 ng/mL cadmium standards versus
their concentration is 51.5 abs-sec/(ng/mL).
4.7.3. If 0.75 abs-sec (sd) and 51.5
abs-sec/(ng/mL) (m) are used in Eqn.
1 (Section
4.1.3.), the qualitative and quantitative detection
limits as determined by the IUPAC method are:
Cld |
= = |
(3)(0.75 abs-sec)/(51.5
abs-sec/(ng/mL) 0.044 ng/mL for the qualitative
detection limit. |
|
Cld |
= = |
(10)(0.75 abs-sec)/(51.5
abs-sec/(ng/mL) 0.15 ng/mL for the quantitative
detection limit. |
The qualitative and quantitative detection
limits for the AAS-HGA analytical technique
are 0.44 ng and 1.5 ng cadmium, respectively, for a 10 mL
solution volume. These correspond, respectively, to 0.007
µg/m3 and 0.025 µg/m3 for a 60 L air
volume.
4.7.4. The peak area (abs-sec) readings of
the Cd standards from 0.2 to 40 ng/mL and the statistical
analysis of the results are given in Table
IV. The recommended standard working range for
AAS-HGA analysis is 0.2 to 20 ng/mL. The
standard of lowest concentration in the recommended
working range is slightly greater than the calculated
quantitative detection limit, 0.15 ng/mL. The deviation
from linearity of the peak area readings of the 20 ng/mL
standard, the highest concentration standard in the
recommended working range, is approximately 10%. The
deviations from linearity of the peak area readings of the
30 and 40 ng/mL standards are significantly greater than
10%. As shown in Table
IV, the precision of the peak area readings are
satisfactory throughout the recommended working range; the
relative standard deviations of the readings range from
0.025 to 0.083. 4.8.
Analytical Method Recovery for Flame AAS Analysis
4.8.1. Four sets of spiked MCEF samples were
prepared by injecting 20 µL of
10, 50, 100 and 200 µg/mL dilute
cadmium stock solutions on 37 mm diameter filters (part
no. AAWP 037 00, Millipore Corp., Bedford, MA) with a
calibrated micropipet. The dilute stock solutions were
prepared by making appropriate serial dilutions of a
commercially available 1,000 µg/mL cadmium standard stock solution
(RICCA Chemical Co., Lot# A102) with the diluting solution
(4% HNO3, 0.4% HCl). Each set contained six
samples and a sample blank. The amount of cadmium in the
prepared sets were equivalent to 0.1, 0.5, 1.0 and 2.0
times the TWA PEL target concentration of 5 µg/m3 for a 400 L air
volume.
4.8.2. The air-dried spiked filters were
digested and analyzed for their cadmium content by flame
atomic absorption spectroscopy (AAS) following the
procedure described in Section 3.
The 0.02 to 2.0 µg/mL cadmium
standards (the suggested working range) were used in the
analysis of the spiked filters.
4.8.3. The results
of the analysis are given in Table
V. One result at 0.5 times the TWA PEL target
concentration was an outlier and was excluded from
statistical analysis. Experimental justification for
rejecting it is that the outlier value was probably due to
a spiking error. The coefficients of variation for the
three test levels at 0.5 to 2.0 times the TWA PEL target
concentration passed the Bartlett's test and were pooled.
4.8.4. The average recovery of the six spiked
filter samples at 0.1 times the TWA PEL target
concentration was 118.2% with a coefficient of variation
(CV1) of 0.128. The average recovery of the
spiked filter samples in the range of 0.5 to 2.0 times the
TWA target concentration was 104.0% with a pooled
coefficient of variation (CV1) of 0.010.
Consequently, the analytical bias found in these spiked
sample results over the tested concentration range was
+4.0% and the OAE was ±6.0%. 4.9. Analytical
Method Recovery for AAS-HGA Analysis
4.9.1. Three sets of spiked MCEF samples were
prepared by injecting 15 µL of
5, 10 and 20 µg/mL dilute
cadmium stock solutions on 37 mm diameter filters (part
no. AAWP 037 00, Millipore Corp., Bedford, MA) with a
calibrated micropipet. The dilute stock solutions were
prepared by making appropriate serial dilutions of a
commercially available certified 1,000 µg/mL cadmium standard stock solution
(Fisher Chemical Co., Lot# 913438-24) with the diluting
solution (4% HNO3, 0.4% HCl). Each set
contained six samples and a sample blank. The amount of
cadmium in the prepared sets were equivalent to 0.5, 1 and
2 times the Action Level TWA target concentration of 2.5
µg/m3 for a 60 L air
volume.
4.9.2. The air-dried spiked filters were
digested and analyzed for their cadmium content by
flameless atomic absorption spectroscopy using a heated
graphite furnace atomizer following the procedure
described in Section 3.
A five-fold dilution of the spiked filter samples at 2
times the Action Level TWA was made prior to their
analysis. The 0.05 to 20 ng/mL cadmium standards were used
in the analysis of the spiked filters.
4.9.3. The
results of the analysis are given in Table
VI. There were no outliers. The coefficients of
variation for the three test levels at 0.5 to 2.0 times
the Action Level TWA PEL passed the Bartlett's test and
were pooled. The average recovery of the spiked filter
samples was 94.2% with a pooled coefficient of variation
(CV1) of 0.043. Consequently, the analytical
bias was -5.8% and the OAE was ±14.2%. 4.10.
Conclusions
The experiments performed in this
evaluation show the two atomic absorption analytical
techniques included in this method to be precise and
accurate and have sufficient sensitivity to measure airborne
cadmium over a broad range of exposure levels and sampling
periods.
5. References:
5.1. Slavin, W. Graphite
Furnace AAS - A Source Book; Perkin-Elmer Corp.,
Spectroscopy Div.: Ridgefield, CT, 1984; p. 18 and pp.
83-90.
5.2. Grosser, Z., Ed.; Techniques in Graphite Furnace Atomic
Absorption Spectrophotometry; Perkin-Elmer Corp.,
Spectroscopy Div.: Ridgefield, CT, 1985.
5.3. Occupational Safety and
Health Administration Salt Lake Technical Center:
Metal and Metalloid Particulate in Workplace Atmospheres
(Atomic Absorption) (USDOL/OSHA Method No. ID-121).
In OSHA Analytical Methods Manual 2nd ed. Cincinnati, OH:
American Conference of Governmental Industrial Hygienists,
1991.
5.4. Occupational Safety and Health Administration
Salt Lake Technical Center: Metal and Metalloid
Particulate in Workplace Atmospheres (ICP) (USDOL/OSHA
Method No. ID-125G). In OSHA Analytical
Methods Manual 2nd ed. Cincinnati, OH: American
Conference of Governmental Industrial Hygienists, 1991.
5.5. Windholz,
M., Ed.; The Merck Index,
10th ed.; Merck & Co.: Rahway, NJ, 1983.
5.6. Analytical Methods
for Atomic Absorption Spectrophotometry, The
Perkin-Elmer Corporation: Norwalk, CT, 1982.
5.7. Slavin, W.,
D.C. Manning, G. Carnrick, and E. Pruszkowska:
Properties of the Cadmium Determination with the Platform
Furnace and Zeeman Background Correction. Spectrochim. Acta 38B:1157-1170 (1983).
5.8. Occupational
Safety and Health Administration Salt Lake Technical
Center: Standard Operating
Procedure for Atomic Absorption. Salt Lake City, UT:
USDOL/OSHA-SLTC, In progress.
5.9. Occupational Safety and
Health Administration Salt Lake Technical Center:
AAS-HGA Standard Operating
Procedure. Salt Lake City, UT:
USDOL/OSHA-SLTC, In progress.
5.10. Mandel, J.:
Accuracy and Precision, Evaluation and Interpretation of
Analytical Results, The Treatment of Outliers. In Treatise On Analytical Chemistry, 2nd ed.,
Vol.1, edited by I. M. Kolthoff and P. J. Elving. New
York: John Wiley and Sons, 1978. pp. 282-285.
5.11. National Institute
for Occupational Safety and Health: Documentation of the NIOSH Validation
Tests by D. Taylor, R. Kupel, and J. Bryant
(DHEW/NIOSH Pub. No. 77-185). Cincinnati, OH: National
Institute for Occupational Safety and Health, 1977.
5.12. Occupational Safety and Health Administration
Analytical Laboratory: Precision and Accuracy Data
Protocol for Laboratory Validations. In OSHA Analytical Methods Manual 1st ed.
Cincinnati, OH: American Conference of Governmental
Industrial Hygienists (Pub. No. ISBN: 0-936712-66-X), 1985.
5.13. Long, G.L.
and J.D. Winefordner: Limit of Detection -- A Closer
Look at the IUPAC Definition. Anal.
Chem. 55:712A-724A (1983).
5.14. American Conference
of Governmental Industrial Hygienists: Documentation of Threshold Limit Values and
Biological Exposure Indices. 5th ed. Cincinnati, OH:
American Conference of Governmental Industrial Hygienists,
1986.
Table I
Cd Detection Limit
Study (Flame AAS Analysis)
|
STD (µg/mL) |
Absorbance
Reading at 228.8 nm |
Statistical Analysis |
|
Reagent
Blank |
5 4 4 |
2 3 3 |
n mean std dev CV |
= = =
= |
6 3.50 1.05 0.30 |
0.001 |
6 2 6 |
6 4 6 |
n mean std
dev CV |
= = = = |
6 5.00 1.67 0.335 |
0.002 |
5 7 7 |
7 3 4 |
n mean std
dev CV |
= = = = |
6 5.50 1.76 0.320 |
0.005 |
7 8 8 |
7 8 6 |
n mean std
dev CV |
= = = = |
6 7.33 0.817 0.111 |
0.010 |
10 10 10 |
9 13 10 |
n mean std
dev CV |
= = = = |
6 10.3 1.37 0.133 |
0.020 |
20 20 20 |
23 22 20 |
n mean std
dev CV |
= = = = |
6 20.8 1.33 0.064 |
0.050 |
42 42 42 |
42 42 45 |
n mean std
dev CV |
= = = = |
6 42.5 1.22 0.029 |
0.10 |
84 80 83 |
n mean std
dev CV |
= = = = |
3 82.3 2.08 0.025 |
Table II
Cd Standard
Working Range Study (Flame AAS Analysis)
|
STD (µg/mL) |
Absorbance
Reading at 228.8 nm |
Statistical Analysis |
|
Reagent
Blank |
5 4 4 |
2 3 3 |
n mean std dev CV |
= = =
= |
6 3.50 1.05 0.30 |
0.020 |
20 20 20 |
23 22 20 |
n mean std
dev CV |
= = = = |
6 20.8 1.33 0.064 |
0.050 |
42 42 42 |
42 42 45 |
n mean std
dev CV |
= = = = |
6 42.5 1.22 0.092 |
0.10 |
84 80 83 |
n mean std
dev CV |
= = = = |
3 82.3 2.08 0.025 |
0.20 |
161 161 158 |
n mean std
dev CV |
= = = = |
3 160.0 1.73 0.011 |
0.05 |
391 389 393 |
n mean std
dev CV |
= = = = |
3 391.0 2.00 0.005 |
1.00 |
760 748 752 |
n mean std
dev CV |
= = = = |
3 753.3 6.11 0.008 |
2.00 |
1416 1426 1401 |
n mean std
dev CV |
= = = = |
3 1414.3 12.6 0.009 |
Table III
Cd
Detection Limit Study (AAS-HGA
Analysis)
|
STD (ng/mL) |
Peak Area
Readings × 103 at 228.8 nm |
Statistical Analysis |
|
Reagent
Blank |
0 0 0 |
0 1 0 |
n mean std dev CV |
= = =
= |
6 0.167 0.41 2.45 |
0.1 |
8 5 13 |
6 7 7 |
n mean std
dev CV |
= = = = |
6 7.7 2.8 0.366 |
0.2 |
11 11 12 |
13 12 12 |
n mean std
dev CV |
= = = = |
6 11.8 0.75 0.064 |
0.5 |
28 26 28 |
33 28 30 |
n mean std
dev CV |
= = = = |
6 28.8 2.4 0.083 |
1.0 |
52 56 54 |
55 58 54 |
n mean std
dev CV |
= = = = |
6 54.8 2.0 0.037 |
2.0 |
101 110 110 |
112 110 110 |
n mean std
dev CV |
= = = = |
6 108.8 3.9 0.036 |
Table IV
Cd Standard
Working Range Study (AAS-HGA Analysis)
|
STD (ng/mL) |
Peak Area
Readings × 103 at 228.8 nm |
Statistical Analysis |
|
0.2 |
11 11 12 |
13 12 12 |
n mean std dev CV |
= = =
= |
6 11.8 0.75 0.064 |
0.5 |
28 26 28 |
33 28 30 |
n mean std
dev CV |
= = = = |
6 28.8 2.4 0.083 |
1.0 |
52 56 54 |
55 58 54 |
n mean std
dev CV |
= = = = |
6 54.8 2.0 0.037 |
2.0 |
101 110 110 |
112 110 110 |
n mean std
dev CV |
= = = = |
6 108.8 3.9 0.036 |
5.0 |
247 268 259 |
265 275 279 |
n mean std
dev CV |
= = = = |
6 265.5 11.5 0.044 |
10.0 |
495 523 516 |
520 513 533 |
n mean std
dev CV |
= = = = |
6 516.7 12.7 0.025 |
20.0 |
950 951 949 |
953 958 890 |
n mean std
dev CV |
= = = = |
6 941.8 25.6 0.027 |
30.0 |
1269 1303 1295 |
1291 1307 1290 |
n mean std
dev CV |
= = = = |
6 1293 13.3 0.010 |
40.0 |
1505 1535 1566 |
1567 1567 1572 |
n mean std
dev CV |
= = = = |
6 1552 26.6 0.017 |
Table V
Analytical
Method Recovery (Flame AAS Analysis)
|
Test Level |
|
0.5× |
|
1.0× |
|
2.0× |
µg Taken
|
µg Found
|
% Rec.
|
|
µg Taken
|
µg Found
|
% Rec.
|
|
µg Taken
|
µg Found
|
% Rec.
|
1.00 |
1.0715 |
107.2 |
|
2.00 |
2.0688 |
103.4 |
|
4.00 |
4.1504 |
103.8 |
1.00 |
1.0842 |
108.4 |
|
2.00 |
2.0174 |
100.9 |
|
4.00 |
4.1108 |
102.8 |
1.00 |
1.0842 |
108.4 |
|
2.00 |
2.0431 |
102.2 |
|
4.00 |
4.0581 |
101.5 |
1.00 |
1.0081* |
100.8* |
|
2.00 |
2.0431 |
102.2 |
|
4.00 |
4.0844 |
102.1 |
1.00 |
1.0715 |
107.2 |
|
2.00 |
2.0174 |
100.9 |
|
4.00 |
4.1504 |
103.8 |
1.00 |
1.0842 |
108.4 |
|
2.00 |
2.0045 |
100.2 |
|
4.00 |
4.1899 |
104.7 |
|
|
5 107.9 0.657 0.006 |
|
|
6 101.6 1.174 0.011 |
|
|
6 103.1 1.199 0.012 |
|
CV1(pooled) = 0.010 |
|
* - Rejected as
an outlier - this value did not pass the outlier T-test
at the 99% confidence level. |
|
Test Level |
|
|
0.1× |
|
|
µg Taken
|
µg Found
|
% Rec.
|
|
|
0.200 |
0.2509 |
125.5 |
|
|
0.200 |
0.2509 |
125.5 |
|
|
0.200 |
0.2761 |
138.1 |
|
|
0.200 |
0.2258 |
112.9 |
|
|
0.200 |
0.2258 |
112.9 |
|
|
0.200 |
0.1881 |
94.1 |
|
|
|
|
6 118.2 15.1 0.128 |
|
Table VI
Analytical
Method Recovery (AAS-HGA Analysis)
|
Test Level |
|
0.5× |
|
1.0× |
|
2.0× |
ng Taken
|
ng Found
|
% Rec.
|
|
ng Taken
|
ng Found
|
% Rec.
|
|
ng Taken
|
ng Found
|
% Rec.
|
75 |
71.23 |
95.0 |
|
150 |
138.00 |
92.0 |
|
300 |
258.43 |
86.1 |
75 |
71.47 |
95.3 |
|
150 |
138.29 |
92.2 |
|
300 |
258.46 |
86.2 |
75 |
70.02 |
93.4 |
|
150 |
136.30 |
90.9 |
|
300 |
280.55 |
93.5 |
75 |
77.34 |
103.1 |
|
150 |
146.62 |
97.7 |
|
300 |
288.34 |
96.1 |
75 |
78.32 |
104.4 |
|
150 |
145.17 |
96.8 |
|
300 |
261.74 |
87.2 |
75 |
71.96 |
95.9 |
|
150 |
144.88 |
96.6 |
|
300 |
277.22 |
92.4 |
|
|
6 97.9 4.66 0.048 |
|
|
6 94.4 2.98 0.032 |
|
|
6 90.3 4.30 0.048 |
|
CV1(pooled) =
0.043 |
Attachment 1
Instrumental Parameters for Flame AAS
Analysis
Atomic Absorption
Spectrophotometer (Perkin-Elmer Model 603)
Flame: |
Air/Acetylene -- lean,
blue |
Oxidant Flow: |
55 |
Fuel Flow: |
32 |
Wavelength: |
228.8 nm |
Slit: |
4 (0.7 nm) |
Range: |
UV |
Signal: |
Concentration (4 exp) |
Integration Time: |
3 sec |
Attachment 2
Instrumental Parameters for HGA Analysis
Atomic Absorption Spectrophotometer (Perkin-Elmer
Model 5100)
Siganl Type: |
Zeeman AA |
Slitwidth: |
0.7 nm |
Wavelength: |
228.8 nm |
Measurement: |
Peak Area |
Integration Time: |
6.0 sec |
BOC Time: |
5 sec |
|
BOC = Background Offset
Correction |
Zeeman Graphite Furnace (Perkin-Elmer Model
HGA-600)
Step |
Ramp Time (sec)
|
Hold Time (sec)
|
Temp. (°C)
|
Argon
Flow (mL/min)
|
Read (sec)
|
|
1)Predry |
5 |
10 |
90 |
300 |
-- |
2)Dry |
30 |
10 |
140 |
300 |
-- |
3)Char |
10 |
20 |
900 |
300 |
-- |
4)Cool Down |
1 |
8 |
30 |
300 |
-- |
5)Atomize |
0 |
5 |
600 |
0 |
-1 |
6)Burnout |
1 |
8 |
2500 |
300 |
-- | |
|
| |