|
Method no.: |
ID-121 |
|
|
Control no.: |
t-id121-fv-02-02021 |
|
|
Matrix: |
Air, Wipes, or Bulks |
|
|
OSHA Permissible Exposure
Limits: |
See Table 1 |
|
|
Collection
Procedure: |
Personal air samples are
collected on mixed-cellulose ester filters using
a calibrated sampling pump. Wipe or bulk samples are collected
using grab sampling techniques. |
|
|
Recommended Sampling Rate: |
2 L/min |
|
|
Recommended Air Volumes: |
|
|
|
Time Weighted Average
Samples: |
480 to 960 L |
|
|
Short-Term Exposure Limit
Samples: |
30 L |
|
|
Ceiling Samples: |
10 L* |
|
|
Analytical
Procedure: |
Samples are desorbed or digested
using water extractions or mineral acid digestions. Elemental
analysis of the prepared sample solutions is performed by
atomic absorption or emission spectroscopy. |
|
|
Detection Limits: |
See Table 2 |
|
|
Precision and Accuracy: |
See Table 3 |
|
|
Method Classification: |
Validated Analytical
Method |
|
|
Date (Date Revised): |
1985 (February 2002) |
|
|
* Alternate air
volumes may be necessary to achieve good analytical
sensitivity. |
|
|
Commercial manufacturers and products mentioned in this method
are for descriptive use only and do not constitute
endorsements by USDOL-OSHA. Similar products from other
sources can be substituted.
|
|
|
Division of Physical Measurements and Inorganic
Analyses OSHA Salt Lake Technical Center Sandy, Utah
|
1. Introduction
This method can determine the amount of specific
metal and metalloid particulates in the workplace atmosphere.
The airborne particulates are collected on filters using
calibrated sampling pumps. These samples are then analyzed
using flame atomic absorption or emission spectrometry. This
method can also determine specific metals and metalloids
contained in wipe and bulk samples. The identification and
quantification of the particulate is directly determined as
the element. The elements are:
Aluminum (Al) |
Gold (Au) |
Potassium
(K) |
Antimony (Sb) |
Hafnium (Hf) |
Selenium
(Se) |
Barium (Ba) |
Indium (In) |
Silver
(Ag) |
Bismuth (Bi) |
Iron (Fe) |
Sodium
(Na) |
Cadmium (Cd) |
Lead (Pb) |
Tellurium
(Te) |
Calcium (Ca) |
Lithium (Li) |
Thallium
(Tl) |
Cesium (Cs) |
Magnesium
(Mg) |
Tin (Sn) |
Chromium (Cr) |
Manganese
(Mn) |
Titanium
(Ti) |
Cobalt (Co) |
Molybdenum
(Mo) |
Yttrium
(Y) |
Copper (Cu) |
Nickel (Ni) |
Zinc (Zn) |
|
Platinum (Pt) |
Zirconium
(Zr) | For some
analytes, there are alternate methods or procedures which may
be more sensitive, accurate, or specific. When a separate OSHA
method or procedure exists, that method shall take precedence
over this method unless special circumstances render it
inapplicable. Elements or compounds having alternate methods
or stopgap procedures are:
Element or
Compound |
OSHA Method
No. |
Aluminum oxide |
ID-198SG or ID-109SG |
Barium sulfate |
ID-204 |
Cadmium |
ID-189 |
Chromic acid/chromates |
ID-103 |
Ferrovanadium |
ID-125G |
Vapors (i.e.
Ni(CO)4, H2Se, TeF6,
C5H4Mn(CO)3] |
In-House Methods |
Organic tin compounds |
ID-102SG |
Platinum (soluble) |
ID-130SG |
Selenium |
ID-133SG |
Solders |
ID-206 |
Stibine |
NIOSH 6008, In House |
Tetraethyl lead and
tetramethyl lead |
In-House Method |
Titanium dioxide |
ID-204 |
Welding fumes |
ID-125G |
Zinc oxide |
ID-143 | Depending on advances in technology or changes
in exposure limits, substances may be added or deleted from
the above lists.
1.1 History
Air and
wipe samples containing metal and metalloid particulate have
always been analyzed at the OSHA Salt Lake City Analytical
Laboratory using atomic absorption or emission spectrometry
(8.1). Constituents in bulk samples have been determined
semi-quantitatively using this technique.
1.2
Principle
Air samples of the workplace are taken
using calibrated sampling pumps with cassettes containing
either mixed cellulose ester (MCE) or polyvinyl chloride
(PVC) filters. These samples are prepared in the laboratory
using concentrated (concd) acids or extracted with deionized
water if a soluble fraction is required. The sample solution
is diluted to a known volume after any necessary matrix
modifiers are added. The sample is then aspirated into the
flame of an atomic absorption or emission spectrophotometer
(AAS or AES) and the molecules in the sample solution are
subjected to the following processes:
1) nebulization 2)
desolvation 3) liquefaction 4) vaporization 5)
atomization 6) excitation (atoms converted from
"ground" to excited state) 7)
ionization The
absorption or emission of light occurring during processes 5
and 6 is then measured at the characteristic wavelength for
the element of interest.
For absorption, a hollow
cathode lamp or an electrodeless discharge lamp (EDL) is
used as the light source. A double beam spectrophotometer is
normally used where the lamp radiation alternately passes
through and around a flame into which the sample is being
aspirated. The sample is atomized and the metal or metalloid
atoms absorb light from the source at their characteristic
wavelengths. This absorption is proportional to the
concentration of the element present in the sample solution.
A monochromator isolates the characteristic radiation of the
element being analyzed. A photosensitive device then
measures the intensity of the transmitted radiation from the
two light paths to determine the amount of absorbance
occurring in the flame.
For emission, a light source
is not used. The sample is introduced into the flame,
atomized and excited, and then the light emission from
excitation is isolated and measured. The intensity of the
light emitted is proportional to the concentration of the
element present.
The following flames are used in
this method for absorption or emission:
a) Air/Acetylene mixture
(Air/C2H2) b) Nitrous
oxide/Acetylene mixture
(N2O/C2H2) c)
Air/Hydrogen mixture (Air/H2)
The use of a specific flame
is dependent on the respective element's analytical
stability, sensitivity, and interferences.
1.3
Advantages and Disadvantages
1.3.1 This analytical
method is specific for the element to be determined and
does not distinguish different compounds. When an analysis
for a compound is requested, an elemental analysis is
performed on the sample. A gravimetric factor is then
applied to calculate the compound value (Note: For some
compounds, additional analytical procedures (i.e. ion
chromatography or X-ray diffraction) can be
used to confirm the presence of the particular compound.
1.3.2 The analysis will also not differentiate
between different particle size ranges, such as dusts and
fumes.
1.3.3 Metallic analytes having Permissible
Exposure Limits (PELs) designated as the soluble form
(i.e., iron soluble salts, nickel, etc.) can be analyzed
using this method. Samples for soluble analytes are
extracted with deionized water and an elemental analysis
is performed on the extract.
1.3.4 Some compounds
may not dissolve using the digestion procedures described
herein. In these cases, an alternative digestion method
should be used.
1.3.5. Several elements can be
determined from the same filter sample using this method;
however, digestion procedures may solubilize only certain
metals. If a combination of metals is requested on the
same filter, all of the metals must be soluble in the
digestion procedure used.
1.3.6 The equipment used
is inexpensive and does not require specialized
training. 1.4 Use of
Metal and Metalloid Compounds in Industry
Metals,
their alloys, and compounds are used in a wide variety of
industries. In certain operations (e.g., welding, smelting,
grinding, etc.), particulate matter containing metals and
their compounds may be released into the workplace
atmosphere. These substances pose a potential health hazard
to workers exposed to them (8.2-8.4). Further documentation
regarding industrial use, toxicity, and physical properties
may be found in NIOSH criteria documents for the particular
substance. 2. Analytical Range and
Sensitivity
This method uses detection limit,
linearity, and sensitivity terms which are characteristic of
atomic absorption. These terms are further defined in Appendix
A. Any detection limits, linear ranges, and
sensitivities mentioned in this method are for analyses using
the primary analytical wavelength, a flow spoiler, an
Air/C2H2 flame, and a hollow cathode
lamp unless otherwise noted.
2.1 The qualitative
detection limits listed in Table 2 were taken from reference
8.5. The analytical detection limits (8.1.) listed were
determined from routine laboratory analyses using the
definition listed in Appendix A. These limits are
approximate since they are dependent on instrument
performance and optimization, sample characteristics, and
the range of standards analyzed.
2.2 The upper
linear range for each element is also given in Table 2.
These ranges were taken from reference 8.6. Instrument
response is linear to greater concentrations if an alternate
wavelength is used; however, the detection limit may also
increase. Samples can be diluted to bring the concentration
of the element(s) within the linear range. The upper linear
range for most elements is usually found near 0.25 to 0.30
absorbance units (ABS).
2.3 The sensitivity for each
element is also listed in Table 2. These values are for a
nebulizer which has been optimized to give an ABS of 0.25
for an aqueous solution containing 5 µg/mL Cu (8.6.). The actual sensitivity
obtained will depend on the particular instrument and flame
used, the sample matrix, and instrument operating
parameters. 3. Method Performance - Precision and
Recoveries
Listed in Table 3 are data compiled
from quality control (QC) samples which were spiked with
aqueous solutions of various analytes and then analyzed in
single blind tests. Each analyte was spiked onto an individual
MCE filter, allowed to dry, and then prepared and analyzed
along with survey samples previously taken by industrial
hygienists. These samples were analyzed from 1986 to 1989. Due
to the limited number of survey samples received for a few
substances, QC samples were not prepared and analyzed for all
analytes included in this method.
4.
Interferences
Interferences occur at the
analytical level and can be characterized as chemical, matrix,
ionization, spectral, or as background absorption.
4.1 Chemical or condensed
phase interferences occur when the element of interest
combines with another species in the flame, thus altering
the number of atoms available for emission or absorption.
This can result in either a positive or negative bias
(usually negative) in the results obtained. Chemical
interferences can be controlled by using a hotter flame, or
by the addition of a releasing agent which inhibits the
reaction between the metal and the interfering species.
4.2 Matrix interferences occur when the physical
characteristics (viscosity, surface tension, etc.) of the
sample and standard solutions differ considerably. This may
occur when samples contain large amounts of dissolved salts
or acid, when different solvents are used for samples and
standards, or when the temperatures of samples and standards
are appreciably different. To control this, samples and
standards must be matrix matched, or the sample must be
diluted until any matrix effect becomes insignificant.
4.3 Ionization interferences occur when the flame
temperature is sufficiently high to ionize the atoms of
interest. This changes the absorption spectrum of the
analyte and effectively removes atoms from the flame,
causing a loss of sensitivity. Ionization interferences are
controlled by adding large amounts (usually >0.1%) of an
easily ionized metal such as Na, K, Cs, or rubidium (Rb).
The excess electrons released in the flame greatly reduces
the degree of ionization of the metal being determined.
4.4 Spectral interferences occur when an element
other than the one analyzed absorbs at the same wavelength.
This causes a positive bias in the results obtained when the
interfering element is present in the samples. In this case,
an alternate line should be used. Spectral interferences
also occur when a multielement hollow cathode lamp is used
which contains elements with absorbing wavelengths close to
one another and the analytical slit width used is wide
enough to allow the wavelengths of more than one element to
pass. If the sample contains two or more of these elements,
a positive bias will occur. To resolve this, a single
element lamp, an alternate wavelength, or in certain cases,
a narrower slit width can be used.
4.5 Background
absorption interferences include flame absorption, molecular
absorption, and light scattering:
a) Flame absorption is
most severe below 250 nm. This absorption can be
controlled by careful optimization of fuel and oxidant
flow rates. Other mechanisms of control are: Use of flames
which are more transparent at these wavelengths (i.e.,
Air/H2 or argon/hydrogen flames), or deuterium
arc background correction (DABC).
b) Molecular
absorption is controlled by using hotter flames to break
down molecular species or by DABC.
c) Light
scattering occurs at shorter wavelengths when samples have
a large salt content; this is controlled using
DABC. 4.6 Large
amounts of silicates or other particulates may interfere and
may also cause aspiration problems (8.7). If present, they
should be removed by filtration. The particulate should then
be re-digested and analyzed to ensure the analyte(s) of
interest have been completely extracted.
4.7 This
analytical method is normally not
compound-specific. Compounds are only
determined as the element, and a significant positive bias
can occur when any sample has additional analytes containing
the same element. Other analytical procedures may be
necessary to identify a specific compound. An assessment of
the industrial operation sampled may also provide
information regarding the potential existence of other
analytes that could cause a positive bias.
4.8
Potential interferences for several of the elements
determined by this method are listed in Appendix
B. 5.
Sampling
5.1 Equipment - Air Filter
Samples
5.1.1 Mixed cellulose
ester (MCE) filters (0.8 µm pore
size), cellulose backup pads, and cassettes,
37-mm diameter (part no. MAWP 037 A0,
Millipore Corp., Bedford, MA). Filters and cassettes
having a 25-mm diameter can also be used.
5.1.2 Gel bands (Omega Specialty Instrument Co.,
Chelmsford, MA) for sealing cassettes.
5.1.3
Sampling pumps capable of sampling at 2 liters per minute
(L/min).
5.1.4 Assorted flexible tubing.
5.1.5 Stopwatch and bubble tube or meter for pump
calibration. 5.2
Equipment - Wipe Samples
5.2.1 Smear tabs (part no.
225-24, SKC Inc., Eighty Four, PA), or wipe
filters (Whatman no. 41 or no. 42 filters, Whatman
Labsales Inc., Hillsboro, OR).
5.2.2 Deionized
water.
5.2.3 Scintillation vials,
20-mL (part no. 74515 or 58515, Kimble, Div.
of Owens-Illinois Inc., Toledo, OH) with
polypropylene or Teflon cap liners. Metal cap liners
should not be used. 5.3 Equipment - Bulk Samples
5.3.1 High-volume sampling
pump with appropriate sized MCE collection filters.
5.3.2 Scintillation vials, 20-mL
(same as Section 5.2.3). 5.4 Sampling Procedure - Air Filter
Samples
5.4.1 Place a MCE filter
and a cellulose backup pad in each two- or
three-piece cassette. Seal each cassette with
a gel band.
5.4.2 Calibrate each personal sampling
pump with a prepared cassette in-line to
approximately 2 L/min.
5.4.3 Attach prepared
cassettes to calibrated sampling pumps (the backup pad
should face the pump) and place in appropriate positions
on the employee or workplace area.
5.4.4 Collect
the samples at approximately 2 L/min for the recommended
sampling times (unless otherwise noted):
Time Weighted Average
Samples |
240 to 480 min |
Short-Term Exposure
Limit Samples |
15 min |
Ceiling Samples |
5
min* | The
analytical sensitivity of a specific analyte may dictate
the use of a different sampling time.
* When determining compliance with the Ceiling PEL for
sodium hydroxide, take 15-min
samples.
5.4.5 Place plastic end caps on each cassette after
sampling. Attach an OSHA-21 seal around each
cassette in such a way as to secure the end
caps. 5.5 Sampling
Procedure - Wipe Samples
Certain analytes may have a
skin designation (See Table 1).
5.5.1 Wear clean,
impervious, disposable gloves when taking each wipe
sample.
5.5.2 Moisten the wipe filters with
deionized water prior to use.
5.5.3 If possible,
wipe a surface area covering 100 cm².
5.5.4 Fold
the wipe sample with the exposed side in.
5.5.5
Transfer the wipe sample into a 20-mL
scintillation vial and seal with vinyl or electrical tape.
Securely wrap an OSHA-21 seal
length-wise from vial top to
bottom. 5.6 Sampling
Procedure - Bulk Samples
5.6.1 In order of
laboratory preference, bulk samples may be one of the
following:
1) a
high-volume (>1,000 L) filter sample of
the workplace area, 2) a representative settled dust
(rafter) sample, 3) a sample of the bulk material in
the workplace. 5.6.2 If possible, transfer the bulk
material or filter into a 20-mL scintillation
vial and seal with vinyl or electrical tape. Securely wrap
an OSHA-21 seal length-wise from
vial top to bottom. 5.7 Shipment
5.7.1 Submit at least one
blank sample with each set of air or wipe samples. Blank
filter samples should be handled in the same manner as
other samples, except that an air or wipe sample is not
taken.
5.7.2 The type of bulk sample should be
stated on the OSHA 91A and cross-referenced
to the appropriate air sample(s). Bulk samples should be
shipped with Material Safety Data Sheets (if available)
and should be sent separately from air samples. Check
current mailing restrictions and ship bulks to the
laboratory by an appropriate method.
5.7.3 Send
all samples to the laboratory with the OSHA 91A paperwork
requesting the specific analyte(s) of interest. If
analysis of a mixture of different elements or compounds
is necessary, contact the lab to ascertain which analytes
can be analyzed together.
6. Analysis
6.1 Safety
Precautions
6.1.1 Care should be
exercised when handling any acidic solutions. Acid
solution contact with work surfaces should be avoided. If
any acid contacts the eyes, skin, or clothes, flush the
area immediately with copious amounts of water. Medical
treatment may be necessary.
6.1.2 All work with
concd acids is potentially hazardous. Always wear safety
glasses and protective clothing. Prepare all mixtures,
samples, or dilutions in an exhaust hood. To avoid
exposure to acid vapors, do not
remove any beakers from the hoods until they have returned
to room temperature.
6.1.3 Extra care
should be used when handling perchloric acid
(HClO4). Perchloric acid should only be used in
a hood that has been approved for HClO4 use. In
this hood:
a) Organic reagents
should not be used or stored near HClO4.
b) A water washdown system for the ducts and
work surface must be installed and periodically used.
c) Precautions should be taken to ensure that
explosions or spontaneous ignition of sample material
from HClO4 is
prevented. Working
with HClO4 is very hazardous. Be sure to wear
safety glasses, a labcoat, and gloves. Always add nitric
acid (HNO3) with HClO4. When
digesting backup pads or other samples with
HClO4, watch them carefully since there is a
chance they could ignite. Always keep HNO3
nearby when using HClO4. In the event of sample
media ignition, quickly douse the sample with a small
portion of HNO3.
6.1.4 Care should be
exercised when using laboratory glassware. Chipped
pipettes, volumetric flasks, beakers, or any glassware
with sharp edges exposed should not be used.
6.1.5
Pipetting is always performed using an automatic pipet or
pipette bulb, never by mouth.
6.1.6 Before using
any instrument, the operator should consult the Standard
Operating Procedure (SOP) (8.8) and any instrument
manuals.
6.1.7 Since metallic elements and other
toxic substances are vaporized during flame operation, it
is imperative that an exhaust hood is installed and used
directly above the burner chamber of the spectrometer.
Always ensure the exhaust system is operating before
proceeding with the analysis. 6.2 Equipment
6.2.1 Atomic absorption
spectrophotometer consisting of a(an):
Nebulizer
and burner head.
Pressure-regulating devices
capable of maintaining constant oxidant and fuel
pressures.
Optical system capable of isolating the
desired wavelength of radiation.
Adjustable slit.
Light measuring and amplifying device.
Display, strip chart, or computer interface for
indicating the amount of absorbed or emitted radiation.
Deuterium Arc Background Corrector. This is
usually required for determinations at short (<250 nm)
wavelengths.
Light source for absorption:
a) Hollow cathode lamp
for the specific element or multielement (Note: Please
see specific limitations of multielement lamps in
Appendix B)
b) Electrodeless Discharge Lamp
(EDL) for the specific element. This type of lamp may
provide better sensitivity and detection limits for some
elements, especially Se, Sn, and Sb. If used, a separate
EDL power supply is usually
necessary. 6.2.2.Oxidant: Compressed, filtered air free
from water, oils and other contaminants.
6.2.3
Nitrous oxide (N2O).
6.2.4 Fuel (Use
flash arrestors when using flammable gases. Consult with
the manufacturer for appropriate use.):
a) Acetylene,
commercially available acetylene dissolved in acetone.
CAUTION: Do not use grades of acetylene that contain
solvents other than acetone. These solvents may damage
PVC tubing in some instruments. Do not use acetylene
when the tank pressure drops below 520 kPa (75 psi).
b) Hydrogen is used as the fuel in the
determination of certain elements.
6.2.5 Pressure
regulators, Two-stage.
6.2.6 Flash arrestors
(model 6103, Matheson Gas Products, East Rutherford, NJ).
6.2.7 Glassware
a) Conical beakers,
125- and 250-mL
b) Volumetric
flasks, Class A: 10-, 25-, 50-
and 100-mL
c) Pipettes, Class A: Assorted
sizes 6.2.8
Forceps.
6.2.9 Exhaust hood and hotplate, or
microwave digestion system (model no. MDS-81, CEM Corp.,
Matthews, NC).
6.2.10 Filtering apparatus
consisting of MCE filters, 0.45-µm pore size, 47-mm
diameter (cat. no. HAWP 047 00, Millipore Corp., Bedford,
MA) and filtering apparatus (cat. no. XX15 047 00,
Millipore).
6.2.11 Analytical balance (0.01
mg). 6.3 Reagents
(All chemicals should be reagent grade or better. Many of
the chemicals listed below are only used in specific
instances. Specific reagents are listed within the
additional procedures in Table 4 and also in Table
5.)
6.3.1 Deionized water (DI
H2O) with a specific conductance of less than
10 µS.
6.3.2 Ammonium
fluoride (NH4F) solutions (used for specific
insoluble compounds, see AP 6, Table 4).
a) Ammonium fluoride, 1
M: Dissolve 37.04 g NH4F and dilute to 1 L in
DI H2O. Store in a polyethylene bottle.
b) Ammonium fluoride, 0.1 M in 4%
HNO3: Carefully add 40 mL concd
HNO3 and 100 mL of the 1 M NH4F
solution to 500 mL DI H2O and dilute to 1 L
in a polyethylene volumetric flask. Store in a
polyethylene bottle since acidic solutions of
NH4F may form small amounts of HF and etch
glass containers. 6.3.3 Hydrogen peroxide
(H2O2), 30% (used for digestions of
Cr, see AP 5, Table 4).
6.3.4 Mineral acids (used
for digestions)
CAUTION: Refer to Section 6.1.2.
before using acids
a) Hydrochloric acid
(HCl), concd (36.5 to 38%).
b) Hydrofluoric acid
(HF), concd (49%).
c) Nitric acid
(HNO3), concd (69 to 71%).
d)
Perchloric acid (HClO4), concd (69 to 72%).
Please see Section 6.1.3 before using HClO4.
e) Sulfuric acid (H2SO4),
concd (95 to 98%).
f) Acid mixture for platinum
digestions: Prepare a mixture of HCl/HNO3 by
slowly and carefully adding 82 mL concd HCl to 18
mL concd HNO3 (CAUTION: Do not store this
solution; dispose of properly after
use). 6.3.5 Mineral
acids (used for dilutions or cleaning glassware)
CAUTION: Refer to Section 6.1.2. before using
acids.
a) Nitric acid, 1:1
HNO3/DI H2O mixture: Carefully add
a measured volume of concd HNO3 to an equal
volume of DI H2O.
b) Nitric acid, 4%
v/v: Carefully add 40 mL concd HNO3 to 500 mL
DI H2O and dilute to 1 L.
c) Nitric
acid 10% v/v: Carefully add 100 mL of concd
HNO3 to 500 mL of DI H2O and then
dilute to 1 L.
d) Nitric and hydrochloric acid
v/v mixture (4% HNO3 / X% HCl, where X% is
listed below): Carefully add the appropriate amount of
concd HCl to 500 mL of DI H2O:
4% HCl |
40 mL |
16% HCl |
160 mL |
32% HCl |
320
mL | Then
carefully add 40 mL concd HNO3 and dilute to
1 L with DI H2O. 6.3.6 Chemical or ionization interference
suppressants
a) Aluminum ion, 5,000
µg/mL: Dissolve 69.52 g
aluminum nitrate (Al(NO3)3·9H2O) and
dilute to 1 L in DI H2O.
b) Potassium
ion, 5,000 µg/mL: Dissolve
9.54 g potassium chloride (KCl) in DI H2O and
dilute to 1 L.
c) Sodium ion, 5,000 µg/mL: Dissolve 12.71 g sodium
chloride (NaCl) in DI H2O and dilute to 1
L. 6.3.7 Stock
standard solutions
Commercially available aqueous
standards are used. Expiration dates for standards should
be followed. If there is no expiration date, dispose of
after 1 year. As an alternative, standards can be prepared
using the procedures described in the SOP (8.8) or
instrument manufacturer manuals (i.e., 8.6, 8.9,
8.10). 6.4 Glassware
Preparation
6.4.1 Place the conical
beakers in an exhaust hood and add approximately 10 mL of
a 1:1 HNO3/DI H2O mixture in each
125- or 250-mL conical beaker.
Apply moderate heat until refluxing occurs. Decant the
acid mixture into a waste container and allow the beakers
to cool before removing from the hood. Rinse the beakers
thoroughly with DI H2O.
6.4.2 Rinse all
volumetric flasks with 10% v/v HNO3 and then
rinse thoroughly with DI
H2O. 6.5
Working Standards
6.5.1 Dilute stock
standard solutions to the appropriate ranges using a
diluent that will match the sample matrix. Use information
in Tables 1 and 2 as guides for the ranges; use Table 5
for matrices. The standard concentrations should bracket
the expected sample concentrations and the standard/sample
matrices should match.
6.5.2 Store standards in
appropriate containers. Protect Ag standards from light by
storing them in actinic or brown plastic bottles. Store
standards containing NH4F in polyethylene
containers. 6.6
Sample Preparation
Note: Always prepare blank samples with every sample
set. Prepare an additional blank media sample any time an
extra procedure is used (i.e. wiping out the particulate
contained inside a cassette with an MCE filter or preparing
a contaminated backup pad). This blank media should be from
the same manufactured lot as the prepared filter or backup
pad.
6.6.1 Preparation of air and wipe samples
Open the filter cassette or scintillation vial,
carefully remove the sample filter with forceps, and place
in a labeled conical beaker. Use 125-mL
beakers for air samples and smear tabs; use
250-mL beakers for large wipe samples. If the
cassette or vial contains loose dust, carefully pour the
dust into the beaker. Always rinse the cassette with DI
H2O and pour the water into the beaker and wipe
out the cassette with a moistened, clean filter and place
this filter in the sample beaker.
If the backup
pad appears to be discolored, it may be due to leakage of
air around the filter during sampling. In these cases, the
pad should also be prepared and analyzed.
6.6.2 Preparation of bulk
samples
Review any available material
safety data sheets to determine safe bulk handling. The
safety data may also offer a clue as to the aliquot amount
needed for adequate detection of the element(s) of
interest.
Measure by volume or weight an
appropriate aliquot of any liquid bulk sample.
Weigh the appropriate amount of any solid bulk
sample.
Note: Aliquot amounts of bulks are dependent on the
analytical sensitivity, detection limit, and solubility of
the material used. If uncertain, a 20- to
50-mg aliquot of a solid material can be
taken as a starting point. Make sure the aliquot taken is
representative of the entire bulk sample. If necessary,
use a mortar and pestle to grind any nonhomogenous
particulate bulk samples in an exhaust hood.
After measuring, transfer the aliquot to a
250-mL conical beaker.
6.6.3 Extraction or digestion - all samples
Consult Tables 4 and 5 to determine the reagents
used during extraction or digestion for each element to be
analyzed. Some elements (Ba, Sn, etc.) or compounds are
not digested with concd HNO3, but are prepared
using alternate procedures (APs) listed in Table 4. These
elements or compounds and their AP numbers are:
|
Ag |
AP 1 |
LiH |
AP 7 |
Te |
AP 1 |
Al (soluble) |
AP 2 |
MgO |
AP 3 |
TiO2 |
AP 8 |
Al (pyro powders) |
AP 3 |
Na cmpds |
AP 7 |
Tl (soluble) |
AP 2 |
Au |
AP 4 |
Ni (soluble) |
AP 2 |
Y |
AP 3 |
Ba (soluble) |
AP 2 |
Mo (soluble) |
AP 2 |
Zr |
AP 6 |
Ca cmpds |
AP 3 |
Mo (insoluble) |
AP 3 |
|
Cr (II or III) |
AP 2 |
Pb |
AP 1 |
|
Cr (metal) |
AP 5 |
Pt (metal) |
AP 4 |
|
CsOH |
AP 7 |
Sb |
AP 1 |
|
Fe (soluble) |
AP 2 |
Se |
AP 1 |
|
Hf |
AP 6 |
Sn (inorganic) |
AP 4 |
|
KOH |
AP 7 |
SnO |
AP 4 |
|
| For the
element or compounds listed above, follow the APs
recommended and then proceed with Section 6.6.2. For other
elements or compounds, follow the procedures a, b, or c
listed below:
a) All MCE air filters and smear tabs
requiring HNO3 digestion
Place
the beakers in an exhaust hood and add 3 to 5 mL concd
HNO3 to cover the filter. Place the beakers
on a hot plate and heat the samples until about 1 mL
remains. Add a second portion of approximately 1 to 2 mL
of concd HNO3. Apply heat until the
appropriate amount of HNO3 remains in the
beaker (1 mL of HNO3 will give a 4%
HNO3 matrix when diluted to 25 mL final
volume).
b) Large wipe, PVC
filters, or backup pads
Place the beakers
in an exhaust hood and add the following amount of concd
HNO3 to the beakers:
Large wipes and
backup pads 10 to 15 mL PVC filters 3 to 5 mL
Place the beakers on a hot plate and heat the
samples until about 1 mL remains. Add 2 mL of concd
HClO4 along with a second portion of 2 mL
HNO3, heat the sample, and then remove when
about 1 mL remains. (Note: Please see Section 6.1.3.
before using HClO4.) As an alternative, an
extraction of the backup pad or wipe sample using only
HNO3 may be used. Add HNO3 to the
media, digest on a hotplate, and continue to add
HNO3 until the solution becomes clear. Remove
the beaker from the hotplate when the appropriate amount
of HNO3 remains.
c) Bulk samples
Add 10 to 30 mL
HNO3, place the beaker on a hot plate, and
digest the bulk sample until the material dissolves and
the appropriate amount of solution remains (about 1 mL
if diluting samples to 25 mL, 2 mL if 50 mL final
volume, etc. After dilution this will give a final
volume of 4% HNO3). If necessary, use other
acids, or use a microwave digestion system to facilitate
digestion [For further information regarding microwave
digestion, see the Standard Operating Procedure
(8.11.)]. 6.6.4
Filtration - all samples
1) Samples Previously
Extracted:
Samples extracted with DI
H2O should normally be filtered. If
particulate is present, filter the extract through a
0.45-µm MCE filter. Save the
extract as the soluble portion. If necessary, digest the
particulate on both filters using procedure (a) above or
the applicable AP to prepare the remaining insoluble
material for additional analyses. To control for
potential contamination, prepare blank samples in the
same fashion as the filtered samples.
2) Samples
Previously Digested:
If particulate matter is
present after digesting, cool the sample, add
approximately 10 mL DI H2O, then filter
the solution through a 0.45-µm
MCE filter. Save the filtrate. Repeat digestion
procedure (a) above for the filter containing the
particulate. 6.6.5
Dilution - all samples
Allow all digested samples to cool to room
temperature in an exhaust hood before proceeding.
Additional sample or filtrate treatment may be required
for certain elements. Perform any special sample
treatments recommended in Table 5, and then quantitatively
transfer each sample and each filtrate solution to
individual volumetric flasks. Add any reagents necessary
to achieve the final solution concentrations listed in
Table 5 for specific analytes. Dilute to volume with DI
H2O and then mix well. Solution volumes are
dependent on the following factors:
a) The amount of sample
the industrial hygienist has collected (air volume
and/or filter loading).
b) The detection limit
of the analytical method.
c) The PEL of the
analyte.
d) The number of analytes
requested. Air
samples are normally diluted to 25 mL unless one or more
of the above factors suggests an alternate volume should
be used. For routine analysis, at least 1/10 of the OSHA
PEL should be detectable. Final solution volumes can be
estimated using the following equation:
FV
Factor = |
0.1 ×
PEL × air volume
QnDL × GF | Where:
PEL |
= |
Permissible Exposure
Limit (in mg/m3) |
Air Vol |
= |
Air Volume taken (in
L) |
QnDL |
= |
Quantitative Detection
Limit (in µg/mL) |
GF |
= |
Gravimetric Factor (if
required - some factors are listed in Table
6) | Quantitative detection limits are listed in
Table 2.
The FV factor assists in determining the
final volume. Sample solution volumes normally used are:
5-, 10-, 25-, 50-, or
100-mL. Final volumes of 50- and
100-mL are normally reserved for wipe or bulk samples. If
possible, FV should always be larger than the final
solution volume. For example, if a sample has a
200-L air volume, a PEL of 0.05
mg/m3, a GF of 1, and a QnDL of 0.09 µg/mL, then:
FV Factor = 11.1
and a final volume should be 10-mL.
Due to the limited amount of solution available for
analysis and the potential for sample loss during
transfer, 5-mL solution volumes are only used
when absolutely necessary. 6.7 Instrument Setup and Analysis
6.7.1 Set up the AAS or
AES according to the SOP (8.8) or the manufacturer's
instructions. Use the flame and wavelength recommended in
Table 7. If alternate conditions are necessary, consult
the instrument manufacturer's manual for other settings
and operating procedures. Install an EDL or hollow cathode
lamp for the element of interest and allow it to warm up
for 10-20 min or until the energy output
stabilizes. Optimize conditions such as lamp position,
burner head alignment, fuel and oxidant flow rates, etc.
See the SOP (8.8) or specific instrument manuals for
details.
6.7.2 Aspirate and measure the ABS of a
standard solution for the element of interest. The
standard concentration should be within the linear range
for the element. Compare the ABS to an expected
sensitivity value (Note: Some values are listed in
Table 7; these were adapted from reference 8.6 or
obtained at the OSHA laboratory). Then aspirate the
smallest standard to be used and assure the ABS reading is
above the background level of the instrument.
6.7.3 Make any adjustments necessary for the
particular analysis, such as: scale expansion, burner head
rotation, background correction, or alternate wavelength.
6.7.4 Aspirate and measure the ABS of a prepared
standard solution, then determine the baseline by
aspirating DI H2O and measuring the ABS.
6.7.5 Analyze standards, samples, and blanks.
Repeat the baseline determination after each solution is
analyzed. The baseline readings will assist in correcting
any instrument drift. If more than one solution has been
prepared for a sample (i.e. filtrate and sample, or
soluble and insoluble portions), analyze each for all
requested elements. Standards must bracket the sample
concentrations. Analyze a standard after every four or
five samples. Standard readings should be within 10 to 15%
of the readings obtained at the beginning of the analysis.
6.7.6 If any samples exceed the linear range, they
should be diluted. When diluting a sample, be sure that
the diluted sample has the same matrix as the original
sample and standards. If a number of samples must be
diluted, it may be more advantageous to use a less
sensitive wavelength. 6.8 Analytical Recommendations
6.8.1 When a fresh
standard is prepared, analyze the old and new standards
and compare results to verify the new standard is correct.
If two or more stock solutions are available for working
standard preparations, rotate the preparation from one
stock solution to the next to verify the quality.
6.8.2 Keep a permanent record of all standard
preparation and comparison data. Assign and follow
expiration dates for all standards.
6.8.3 Always
analyze blank samples along with the other samples. Treat
blanks in the same fashion as samples, including any
filtration steps.
6.8.4 When analyzing for Ag,
carry-over from a large concentration sample
or standard to the next sample can occur, causing
erroneous readings. To remedy this, aspirate 4%
HNO3 instead of water between samples.
6.8.5 In this method, many different matrices are
used to digest and keep analytes in solution.
Occasionally, during multiple element analysis of the same
sample, matrix effects can occur if standards are not
matrix-matched with samples. Also, it is
sometimes necessary to prepare samples in a matrix
substantially different from recommendations. If these
conditions occur, one or two standards should be prepared
in the same matrix to determine any matrix effects. A
reagent blank should also be prepared and analyzed to
determine any effect on the background signal. If a
significant difference is noted in the analytical signals
for the two different matrices, a full set of standards
should always be prepared in the sample matrix and
analyzed with the
samples. 7. Calculations
7.1 Subtract each baseline
ABS from the corresponding standard ABS, and plot the net
ABS versus the standard concentrations. Using a least
squares method, determine the equation for the best curve
fit.
7.2 Subtract each baseline ABS from the
corresponding sample or blank ABS, and use the standard
curve to calculate the concentration of each analyte in
µg/mL.
7.3 Calculate the
concentration for each air sample as:
C = |
[(A × SA
× D × GF) - (B × SB × GF)]
air volume | Where:
C |
= |
analyte
(mg/m3) |
A |
= |
concn of analyte in the
sample solution (µg/mL) |
B |
= |
concn of analyte in the
blank solution (µg/mL) |
SA |
= |
sample solution volume
(mL) |
SB |
= |
blank solution volume
(mL) |
D |
= |
dilution factor (if
any) |
GF |
= |
gravimetric factor (if
any; see Table 6) |
Air Vol |
= |
air volume sampled
(L) |
7.4
For wipe or bulk samples, calculate the total amount (in µg)
of analyte in each sample using the equation above. An air
volume is not used. Convert bulk sample analytes to %
composition using:
analyte %
(w/w) = |
(C)
(100%)
(sample wt) (1,000 µg/mg) |
(Bulk
Samples) | Where:
C |
= |
analyte amount (µg) |
Sample wt |
= |
aliquot (in mg) of bulk
taken in Section 6.6. |
7.5 Reporting Results to the Industrial
Hygienist
For those samples only extracted with DI
H2O, report the sample results as the soluble
fraction of the sample.
If more than one solution
exists for a sample, and it is not necessary to report
results separately, then combine these results. An example
is a sample that was filtered due to insoluble particulate.
The results from the filtrate plus results from the second
particulate digestion are added together.
7.5.1 Report air sample
results as mg/m3 analyte.
7.5.2 Report
wipe sample concentrations as total micrograms or
milligrams analyte.
7.5.3 Report bulk sample
results as approximate percent by weight analyte (note:
Sample results for bulk liquids may be reported as
approximate percent by volume if volumetric aliquots were
taken during sample preparation.) Due to differences in
sample matrices between bulks and standards, bulk results
are approximate. 8. References
8.1
Occupational Safety and Health Administration Analytical
Laboratory: OSHA Manual of
Analytical Methods edited by R.G. Adler (Method No.
I-1). Salt Lake City, UT. 1977.
8.2 Clayton, G.D. and F.E. Clayton,
ed.: Patty's Industrial Hygiene
and Toxicology. 3rd ed. New York: John Wiley and
Sons, 1978.
8.3 American Conference
of Governmental Industrial Hygienists: Documentation of the Threshold Limit Values and
Biological Exposure Indices. 5th Ed. Cincinnati, OH:
American Conference of Governmental Industrial Hygienists,
1986.
8.4 National Institute for
Occupational Safety and Health: The Industrial Environment--Its Evaluation and
Control. Washington, DC: Government Printing Office,
1973.
8.5 Slavin, S., W.B. Barnett,
and H.L. Kahn: The Determination of Atomic Absorption
Detection Limits by Direct Measurement. Atomic Absorption Newsletter 11:
37-41 (1972).
8.6 Perkin-Elmer Corp.: Analytical Methods for Atomic Absorption
Spectrophotometry. Norwalk, CT:
Perkin-Elmer Corp., 1973 and revised edition,
1982.
8.7 National Institute for
Occupational Safety and Health: NIOSH Manual of Analytical Methods. 2nd
ed. (Method no. 173) Cincinnati, OH: National Institute for
Occupational Safety and Health, 1977.
8.8 Occupational Safety and Health Administration
Technical Center: Standard
Operating Procedure for Atomic Absorption. Salt Lake
City, UT. In progress (unpublished).
8.9 Fisher Scientific Company: Atomic Absorption Methods Manual.
Waltham, MA: Fisher Scientific Co., 1977.
8.10 Instrumentation Laboratory
Inc.: Atomic Absorption Methods
Manual. Wilmington, MA: Instrumentation Laboratory
Inc., 1975.
8.11 Occupational
Safety and Health Administration Analytical
Laboratory: Standard Operating
Procedure for Microwave Digestions by D. Cook. Salt
Lake City, UT. 1989 (unpublished).
8.12 "Air
Contaminants; Final Rule": Federal
Register 54:12 (19 Jan. 1989). pp.
2923-2960 and also 54:127 (5 July 1989). pp.
28054-28061.
8.13 Occupational
Safety and Health Administration Analytical Laboratory:
OSHA Laboratory Quality Control
Division Data by B. Babcock, Salt Lake City, UT, 1989
(unpublished).
8.14 Slavin,
Walter: Atomic Absorption
Spectroscopy. New York: Interscience Publishers,
1968.
8.15 Ediger, R.D.:
Atomic Absorption Analysis with the Graphite Furnace using
Matrix Modification. Atomic Absorption
Newsletter. 14(5):
127-130 (1975).
Table 1 Air Contaminants - OSHA Permissible
Exposure Limits* |
|
|
Transitional
PEL |
---Final Rule
PEL--- |
Element |
Substance Exposed to |
----(mg/m3)---- |
----(mg/m3)---- |
|
TWA |
CEILING |
TWA |
STEL |
CEILING |
|
Ag |
Metal and soluble
cmpds (as Ag) |
0.01 |
|
0.01 |
|
|
Al |
Soluble salts (as Al) |
--- |
|
2 |
|
|
Pyro powders |
--- |
|
5 |
|
|
Ba |
Soluble compounds (as
Ba) |
0.5 |
|
0.5 |
|
|
Bi |
Bismuth telluride (Se
doped)** |
--- |
|
5 |
|
|
Ca |
Calcium oxide |
5 |
|
5 |
|
|
Calcium cyanamide |
--- |
|
0.5 |
|
|
Cd |
Fume |
0.1 |
0.3 |
0.1 |
|
0.3 |
|
Dust |
0.2 |
0.6 |
0.2 |
|
0.6 |
|
Co |
Metal dust and fume (as
Co) |
0.1 |
|
0.05 |
|
|
Cobalt Carbonyl
or hydrocarbonyl (as Co) |
--- |
|
0.1 |
|
|
Cr |
Cr (II or III)
compounds (as Cr) |
0.5 |
|
0.5 |
|
|
Cr metal (as Cr) |
1 |
|
1 |
|
|
Cs |
Cesium hydroxide |
--- |
|
2 |
|
|
Cu |
Fumes (as Cu) |
0.1 |
|
0.1 |
|
|
Dusts and mists (as Cu) |
1 |
|
1 |
|
|
Fe |
Dicyclopentadienyl iron |
|
|
Total dust |
15 |
|
10 |
|
|
Iron oxide fume (as
Fe2O3) |
10 |
|
10 |
|
|
Iron salts (soluble) (as
Fe) |
1 |
|
|
Hf |
Hafnium |
0.5 |
|
0.5 |
|
|
In |
Indium and compounds (as
In) |
--- |
|
0.1 |
|
|
K |
Potassium hydroxide |
--- |
|
2 |
|
|
Li |
Lithium hydride |
0.025 |
|
0.025 |
|
|
Mg |
Magnesium oxide fume |
|
|
Total
particulate |
15 |
|
10 |
|
|
Mn |
Mn compounds (as Mn) |
5 |
|
5 |
|
Mn fume (as Mn) |
|
5 |
1 |
3 |
|
Manganese tetroxide (as
Mn) |
--- |
|
1 |
|
|
Mo |
Soluble compounds (as
Mo) |
5 |
|
5 |
|
|
Insoluble compounds (as
Mo) |
|
|
Total dust |
15 |
|
10 |
|
|
Na |
Sodium bisulfite |
--- |
|
5 |
|
|
Sodium fluoroacetate |
--- |
|
0.05 |
0.15 |
|
|
Sodium hydroxide |
2 |
--- |
|
2 |
|
Sodium metabisulfite |
--- |
|
5 |
|
|
Tetrasodium
pyrophosphate*** |
--- |
|
5 |
|
|
Ni |
Metal and insoluble |
|
|
compounds (as Ni) |
1 |
|
1 |
|
|
Soluble compounds (as
Ni) |
1 |
|
0.1 |
|
|
Pb |
Inorganic (see
Code of Federal Regulations 1910.1025) |
|
Pt |
Pt metal |
--- |
|
1 |
|
|
Sb |
Sb and compounds (as Sb) |
0.5 |
|
0.5 |
|
|
Se |
Se and compounds (as Se) |
0.2 |
|
0.2 |
|
|
Sn |
Inorganic compounds |
|
|
except oxides
(as Sn) |
2 |
|
2 |
|
|
Tin oxide (as Sn) |
|
2 |
|
|
Te |
Te and compounds (as Te) |
0.1 |
|
0.1 |
|
|
Ti |
Titanium dioxide |
|
|
Total dust |
15 |
|
10 |
|
|
Tl |
+ Soluble compounds (as
Tl) |
0.1 |
|
0.1 |
|
|
Y |
Yttrium |
1 |
|
1 |
|
|
Zn |
Zinc chloride fume |
1 |
|
1 |
2 |
|
|
Zinc oxide fume |
5 |
|
5 |
10 |
|
|
Zinc oxide |
|
|
Total dust |
15 |
|
10 |
|
|
|
Zinc stearate |
|
|
Total dust |
15 |
|
10 |
|
|
Zr |
Zr compounds (as Zr) |
5 |
|
5 |
10 |
|
|
* From reference
8.12 - Final Rule PELs were voided by a court ruling and
are not applicable *** Also can be analyzed for total
phosphate content by ion chromatography. + Skin
Designation |
| Note:
Compounds having total and respirable dust PELs of 15 and 5
mg/m3, respectively, are normally analyzed
gravimetrically. Elements contained in these dust samples can
be identified by this or other methods, if necessary.
Table
2 Detection Limits, Sensitivities, and Ranges |
|
Element |
Qualitative |
Analytical |
Sensitivity* |
Upper Linear Range* |
|
DL* (µg/mL) |
DL* (µg/mL) |
(µg/mL) |
(µg/mL) |
|
Ag |
0.002 |
0.005 |
0.06 |
4 |
Al+ |
0.02 |
0.3 |
1 |
50 |
Au |
0.01 |
0.05 |
0.25 |
20 |
Ba+ |
0.008 |
0.5 |
0.4 |
25 |
Bi |
0.025 |
0.2 |
0.5 |
30 |
Ca |
<0.0005 |
0.03+ |
0.08 (0.029)+ |
7 |
Cd |
0.0002 |
0.004 |
0.025 |
2 |
Co |
0.01 |
0.04 |
0.15 |
5 |
Cr |
0.003 |
0.04 (0.04) + 0.1 |
(0.31)+ |
5(10)+ |
Cs |
0.005++ |
|
0.2 |
15 |
Cu |
0.001 |
0.005 |
0.09 |
5 |
Fe |
0.005 |
0.03 |
0.12 |
5 |
Hf+ |
2.0 |
|
15 |
500 |
In |
0.02 |
0.1 |
0.7 |
50 |
K |
<0.002 |
0.02 |
0.04 |
2 |
Li |
0.0003 |
0.004 |
0.035 |
2 |
Mg |
<0.0001 |
0.01 |
0.007 |
0.5 |
Mn |
0.002 |
0.01 |
0.055 |
3 |
Mo+ |
0.02 |
0.04 |
0.5 |
60 |
Na |
<0.0002 |
0.009 |
0.015 |
1 |
Ni |
0.002 |
0.1 |
0.15 |
5 |
Pb |
0.01 |
0.05 |
0.5 |
20 |
Pt |
|
2.0 |
13 |
|
Sb** |
(0.08) |
0.1 |
1.0 |
50 |
Se** |
(0.05) |
0.3 |
0.25 |
25 |
Sn** |
(0.01) |
0.1 |
0.6 |
40 |
Te |
0.05 |
0.2 |
1.0 |
25 |
Ti+ |
0.04 |
|
1.8 |
|
Tl |
0.03 |
0.05 |
0.5 |
20 |
Y |
0.05 |
0.7 |
1.8 |
200 |
Zn |
<0.01 |
0.01 |
0.018 |
1 |
Zr+ |
1.0 |
8 |
10.0 |
800 | * DL = Detection Limit. See Appendix A for
more information regarding definitions or calculations.
Analytical DLs are approximate.
** Alternate line of
231.2 nm was used with one exception: The qualitative
detection limit value is for the primary line (217.6 nm).
*** Air/H2 flame used with the exception of
the qualitative detection limit determination. This value is
for Air/C2H2 flame.
+
N2O/C2H2 flame used.
++ Flame emission used to determine qualitative
detection limit.
Table 3 Precision and Accuracy* |
|
Element |
CV |
% Ave Recovery |
Range** |
N |
Ag |
0.083 |
97.8 |
1-4 |
270 |
Al |
0.076 |
94.5 |
100-1500 |
27 |
Au |
--- |
--- |
--- |
|
Ba |
0.10 |
104.7 |
50-75 |
45 |
Bi |
--- |
--- |
--- |
|
Ca |
0.162 |
98.3 |
100-150 |
51 |
Cd |
0.087 |
99.5 |
10-15 |
93 |
Co |
0.052 |
99.3 |
10-15 |
39 |
Cr (Soluble) |
--- |
--- |
--- |
|
Cr (Insoluble) |
0.052 |
95.7 |
45-75 |
72 |
Cs |
--- |
--- |
--- |
|
Cu |
0.043 |
96.8 |
100-150 |
45 |
Fe |
0.084 |
98.2 |
300-400 |
69 |
Hf |
--- |
--- |
--- |
|
In |
--- |
--- |
--- |
|
K |
0.063 |
93.3 |
125-200 |
30 |
Li |
--- |
--- |
--- |
|
Mg |
0.073 |
112.1 |
100-300 |
24 |
Mn |
0.044 |
100.2 |
100-150 |
60 |
Mo (Soluble) |
--- |
--- |
--- |
|
Mo (Insoluble) |
0.075 |
91.2 |
100-250 |
27 |
Na |
0.058 |
97.5 |
100-250 |
68 |
Ni |
0.065 |
99.1 |
100-150 |
18 |
Pb |
0.047 |
99.3 |
20-40 |
300 |
Pt |
0.055 |
98.1 |
80-1800 |
24+ |
Sb |
0.081 |
98.4 |
50-75 |
36 |
Se |
0.122 |
104.9 |
20-100 |
30 |
Sn |
0.079 |
97.4 |
100-150 |
63 |
Te |
--- |
--- |
--- |
|
Tl |
--- |
--- |
--- |
|
Y |
--- |
--- |
--- |
|
Zn |
0.039 |
101.2 |
100-150 |
69 |
Zr |
--- |
--- |
--- |
|
| CV Coefficient
of Variation * Table updated January, 1990 (8.13) **
Range (in µg) of analyte spiked onto
MCE filters. Samples were spiked with aqueous solutions of
dissolved metals or their salts. All samples were prepared and
analyzed using conditions stated in the method. + These
samples were prepared by weighing the metal on filters. A
single blind study was not performed.
Table 4 Alternate Procedures
|
1) Digest
samples with HNO3. Heat until the liquid is nearly
gone. Allow the samples to cool to room temperature.
2) For 25 mL final sample solution volumes, add the
following amount of concd HCl (Adjust accordingly for
alternate solution volumes):
Analyte
Suspected to be Present |
Amount of
HCl |
|
Sb |
|
8 mL |
|
Pb or Ag |
|
4 mL |
|
Se or Te |
|
1
mL | 3) Warm
gently and swirl to dissolve the analyte. Allow samples to
cool and dilute to a 25-mL volume with DI
H2O.
AP 2: Soluble Compounds of Al, Ba, Cr (II or III),
Fe, Ni, Mo, Tl, Zn
| 1) Place the
sample in a beaker and add an aliquot of
room-temperature DI H2O into the
beaker (15 mL is typically used for a full-shift
sample).
2) Place the beaker in an ultrasonic bath for
approximately 10 min.
3) Filter the sample through a
0.45 µm MCE filter and transfer the
filtrate to a 25-mL volumetric flask. If an
insoluble fraction is also requested, digest both sample
filters according to the appropriate procedure.
4) Add
reagents to achieve the final solution concentrations
listed:
Analyte
Presence Suspected |
Final
Concentration |
Cr (II or III), Fe, Ni, Tl,
Zn (as ZnCl2) |
4% HNO3 |
|
Al, Ba |
4% HNO3/1,000
µg/mL Potassium ion |
|
Mo |
4% HNO3/1,000
µg/mL Aluminum
ion |
AP 3: Al (pyro powders), Ca, Mg, Mo (insoluble),
Y
| (1) Digest
the sample using the procedure described in Section 6.6.3.a.
(2) Transfer the sample to a volumetric flask.
(3) Dilute the samples and add ionization suppressants
to achieve the final solution concentrations listed:
Analyte
Suspected to be Present |
Final
Concentration |
Al (pyro powders), Ca, Mg,
Y |
4% HNO3/1,000
µg/mL Potassium ion |
Mo (insoluble) |
4% HNO3/1,000
µg/mL Aluminum
ion |
AP 4: Au, Pt (metal), Sn, or Tin Oxide
(SnO)
| (1) For Au,
Sn, or SnO, add 9 mL HCl to each beaker, swirl, and then add 2
mL HNO3. CAUTION: Make sure the entire filter or
sample is wetted with HCl and allow the filter/HCl solution to
sit for a period of at least 2 to 3 min before adding the
HNO3.
(2) Digest the sample on a hot plate
until nearly dry.
(3) Allow the samples to cool and
then quantitatively transfer the sample, using a small amount
of DI H2O to rinse the beaker, to a clean
volumetric flask. Dilute to volume, making the final solution
10% HCl. For example, add 2.5 mL concd HCl to a sample if the
total solution volume is 25 mL.
(4) Results for either
Sn or SnO are reported as total Sn.
AP 5: Cr [Samples which potentially contain
Cr(VI)]
| For samples
requiring analysis of total Cr, the following procedure should
be used. This procedure avoids the loss of any Cr(VI) as
chromyl chloride (CrO2Cl2). For chromate
or chromic acid analysis, see OSHA method no. ID-103.
(1) Digest the samples collected on MCE filters with
HNO3 and then allow to cool to room temperature. If
PVC filters were used, digest with HNO3 plus 2 mL
of HClO4 and then allow to cool.
(2) Add 1
or 2 mL of 30% H2O2 to the cooled
solution to reduce any Cr(VI) that may be present. Let the
sample sit for several minutes.
(3) Heat approximately
5 min to boil off the H2O2 and then
allow to cool. At this stage HCl may be added if needed to
dissolve other metals.
(4) Dilute to volume with DI
H2O and analyze.
NOTE: Do not add
HClO4 to the sample solution if a large amount of
HCl is already present [any Cr(VI) in the sample would be lost
as CrO2Cl2]. Add concd HNO3,
boil off the HCl, and then add the HClO4.
AP 6: Elements or Compounds* which are Insoluble in
Nitric Acid Digestions
| (1) For
compounds such as zirconium dioxide or hafnium dioxide, place
the sample filter in a platinum crucible, char at 300 °C, then
heat the residue at 800 °C in a muffle furnace. [As an
alternative, the digestion can be performed using a microwave
digestion system (8.11.).]
(2) Add 1 to 2 mL of HF,
swirl the solution, and then heat on a hot plate to dissolve
the residue.
(3) Evaporate the solution to
approximately 0.4 mL and then transfer to a 10-mL
polyethylene volumetric flask. Dilute to volume with a
solution of 0.1 M ammonium fluoride in 4% HNO3.
Another procedure can be used for elements which do
not need to be converted to their fluoride salts:
(1)
Heat the HF solution on the hot plate until the liquid is
nearly gone.
(2) Add 2 to 3 mL HCl, and warm the
solution until about 1 mL remains.
(3) Quantitatively
transfer the solution to a 10 mL volumetric flask and dilute
to volume with the appropriate diluents mentioned in Table 5.
(4) It is recommended to prepare quality control
samples of the substance of concern. Digest the samples and
analyze by the same procedure to check the recovery
efficiency.
For platinum:
(1) Place the sample
filter in a Teflon microwave digestion vessel and add 5
mL of the "acid mixture (HCl/HNO3) for platinum
digestions" prepared in Section 6.3.4, part f).
(2)
Digest the sample according to Microwave Digestion Standard
Operating Procedure (8.11) or manufacturer guidelines.
(3) Allow the sample to cool and then transfer to a
25-mL volumetric flask. Dilute to volume with DI
H2O.
* Some Zr compounds, such as the oxide
and sulfate, may be insoluble when using the HNO3
digestion (8.6, 8.7). Hafnium dioxide may also be
insoluble.
AP 7: CsOH, KOH, LiH, and Na Compounds
| (1) Place
the sample filter in a beaker and desorb with 15 mL of DI
H2O for approximately 5 min.
(2) Decant the
sample solution into a 25-mL volumetric flask and
add any reagents to achieve the final solution
concentrations:
Analyte Suspected to be Present |
Final Concentration |
CsOH |
DI
H2O/1,000 µg/mL
potassium ion |
LiH, Na
cmpds |
DI
H2O |
KOH |
DI
H2O/1,000 µg/mL
sodium ion | For example, add 5 mL of 5,000 µg/mL potassium ion for Cs analysis and
dilute to volume with DI H2O. Add 5 mL of 5,000
µg/mL sodium ion for KOH analysis.
(3) Analyze by flame emission or atomic
absorption.
(1) Digest
the filter with 1 mL HNO3 and 2 mL
H2SO4 in a conical beaker and heat until
about 1 mL remains.
(2) Quantitatively transfer the
solution to a 25-mL volumetric flask, add 5 mL of 5,000 µg/mL potassium ion, then dilute to volume
with DI H2O.
Table 5 Digestion or Extraction
Reagents |
|
Substance |
Reagents Used |
Final Volume Concentration+ |
|
Ag |
HNO3/HCl |
4% HNO3/16%
HCl |
Al (soluble cmpds) |
DI H2O |
4% HNO3/1,000
µg/mL K+ |
Al (pyro powders) |
HNO3 |
4% HNO3/1,000
µg/mL K+ |
Au |
HCl/HNO3 |
10% |
Ba (soluble cmpds) |
DI H2O |
4% HNO3/1,000
µg/mL K+ |
Bi2Te3 (Se doped) |
HNO3 |
4% HNO3 |
Ca & cmpds |
HNO3/HCl* |
4% HNO3/1,000
µg/mL K+ |
Cd |
HNO3 |
4% HNO3 |
Co & cmpds |
HNO3/HCl* |
4% HNO3 |
Cr (II or III) soluble
cmpds |
DI H2O |
4% HNO3 |
Cr metal |
HNO3/H2O2/HCl* |
4% HNO3 |
CsOH |
DI H2O |
DI H2O/1,000
µg/mL K+ |
Cu |
HNO3 |
4% HNO3 |
Fe & cmpds |
HNO3/HCl* |
4% HNO3 |
Fe (soluble salts) |
DI H2O |
4% HNO3 |
Hf |
HF |
4% HF/4%
HNO3/0.1 M NH4F |
In & cmpds |
HNO3 |
4% HNO3 |
KOH |
DI H2O |
DI H2O/1,000
µg/mL Na+ |
LiH |
DI H2O |
DI H2O |
MgO |
HNO3/HCl* |
4% HNO3/1,000
µg/mL K+ |
Mn & cmpds |
HNO3/HCl* |
4% HNO3 |
Mo (soluble cmpds) |
DI H2O |
4% HNO3/1,000
µg/mL Al |
Mo (insoluble cmpds) |
HNO3 |
4% HNO3/1,000
µg/mL Al |
Na & cmpds |
DI H2O |
DI H2O |
Ni metal & insoluble
cmpds |
HNO3/HCl* |
4% HNO3 |
Ni (soluble cmpds) |
DI H2O |
4% HNO3 |
Pb |
HNO3/HCl |
4% HNO3/16%
HCl |
Pt metal |
HCl/HNO3 |
4% HNO3/16%
HCl |
Sb & cmpds |
HNO3/HCl |
4% HNO3/32%
HCl |
Se & cmpds |
HNO3/HCl |
4% HNO3/4%
HCl |
Sn (and SnO) |
HCl/HNO3 |
10% HCl |
Te & cmpds |
HNO3/HCl |
4% HNO3/4%
HCl |
TiO2 |
HNO3/H2SO4 |
4%
H2SO4/1,000 µg/mL K+ |
Tl (soluble cmpds) |
DI H2O |
4% HNO3 |
Y |
HNO3 |
4% HNO3/1,000
µg/mL K+ |
ZnCl2 |
DI H2O |
4% HNO3 |
Zn & cmpds |
HNO3 |
4% HNO3 |
Zr & cmpds |
HF |
4% HF/4%
HNO3/0.1 M NH4F |
| + Standards
should be prepared in this matrix.
* After completing
the digestion with HNO3, add 1 or 2 drops of concd
HCl to facilitate particulate dissolution.
Table 6 Gravimetric Factors |
|
Element |
Compound |
Gravimetric Factor |
|
Bi |
Bismuth telluride
(Bi2Te3) |
1.916 |
Ca |
Calcium cyanamide
(CaCN2) |
1.998 |
Ca |
Calcium hydroxide
[Ca(OH)2] |
1.849 |
Ca |
Calcium oxide (CaO) |
1.399 |
Cs |
Cesium hydroxide (CsOH) |
1.128 |
Cr |
Chromic acid
(CrO3) |
1.923 |
Fe |
Dicyclopentadienyl iron
[(C5H5)2Fe] |
3.331 |
Fe |
Iron oxide
(Fe2O3) |
1.430 |
Li |
Lithium hydride (LiH) |
1.145 |
Mg |
Magnesium oxide (MgO) |
1.658 |
Na |
Sodium bisulfite
(NaHSO3) |
4.525 |
Na |
Sodium fluoroacetate
(FCH2COONa) |
4.351 |
Na |
Sodium hydroxide (NaOH) |
1.740 |
Na |
Sodium metabisulfite
(Na2S2O5) |
4.134 |
Na |
Tetrasodium pyrophosphate
(Na4P2O7) |
2.891 |
Ti |
Titanium oxide
(TiO2) |
1.668 |
Zn |
Zinc chloride
(ZnCl2) |
2.085 |
Zn |
Zinc oxide (ZnO) |
1.245 |
Zn |
Zinc stearate
[Zn(C18H35O2)2] |
9.671 |
|
Table 7 Analytical Parameters |
|
Element |
l(nm) |
Slit (nm) |
Optimization* |
Flame Used |
Comments |
Ag |
328.1 |
0.7 |
4 µg/mL=0.3 ABS |
1 |
For multielement
lamps containing Cu, use 0.2 nm slit. |
Al |
309.3 |
0.7 |
50 µg/mL=0.22 ABS |
3 |
|
Au |
242.8 |
0.7 |
15 µg/mL=0.26 ABS |
1 |
|
Ba |
553.6 |
0.4 |
15 µg/mL=0.16 ABS |
3 |
|
Bi |
223.1 |
0.2 |
20 µg/mL=0.18 ABS |
1 |
|
Ca** |
422.7 |
0.7 |
4 µg/mL=0.22 ABS |
4 |
|
Cd** |
228.8 |
0.7 |
2 µg/mL=0.35 ABS |
1 |
|
Co |
240.7 |
0.2 |
5 µg/mL=.015 ABS |
2 |
|
Cr** |
357.9 |
0.7 |
2 µg/mL=0.05 ABS |
3 |
|
Cs |
852.1 |
1.4 |
10 µg/mL=0.22 ABS |
1 |
|
Cu** |
324.7 |
0.7 |
5 µg/mL=0.25 ABS |
1 |
For multielement
lamps containing Ni or Fe, use 0.2 nm slit. |
Fe** |
248.3 |
0.2 |
5 µg/mL=0.18 ABS |
2 |
In the presence
of Co, |
|
248.8 |
0.2 |
5 µg/mL=0.11 ABS |
2 |
do not use a
multielement lamp containing Co at 248.3 nm. Use 248.8
or 372.0 nm. |
Hf |
286.6 |
0.2 |
300 µg/mL=0.2 ABS |
3 |
|
In |
303.9 |
0.7 |
25 µg/mL=0.15 ABS |
1 |
|
K |
766.5 |
1.4 |
2 µg/mL=0.3 ABS |
1 |
|
Li |
670.8 |
1.4 |
1 µg/mL=0.13 ABS |
1 |
|
Mg** |
285.2 |
0.7 |
0.3 µg/mL=0.19 ABS |
3 |
|
Mn** |
279.5 |
0.2 |
2 µg/mL=0.16 ABS |
1 |
|
Mo** |
313.5 |
0.7 |
2 µg/mL=0.20 ABS |
3 |
|
Na |
589.6 |
0.4 |
0.8 µg/mL=0.2 ABS |
1 |
|
Ni** |
232.0 |
0.2 |
5 µg/mL=0.15 ABS |
1 |
For multielement
lamps containing Fe, use the secondary Ni line, 352.4
nm. |
Pb |
283.3 |
0.7 |
20 µg/mL=0.18 ABS |
1 |
|
Pt |
265.9 |
0.7 |
100 µg/mL=0.033 ABS |
3 |
|
Sb |
217.6 |
0.2 |
20 µg/mL=0.18 ABS |
1 |
For determination
in 231.2 0.7 20 µg/mL=0.07 ABS
1 the presence of Pb, use the 231.2 nm line. |
Se |
196.0 |
2.0 |
20 µg/mL=0.18 ABS |
5 |
Use an EDL |
Sn |
224.6 |
0.7 |
50 µg/mL=0.28 ABS |
5 |
|
Te |
214.3 |
0.2 |
25 µg/mL=0.11 ABS |
1 |
|
Ti |
365.3 |
0.2 |
120 µg/mL=0.3 ABS |
3 |
|
Tl |
276.8 |
0.7 |
20 µg/mL=0.18 ABS |
1 |
|
Y |
410.2 |
0.2 |
100 µg/mL=0.24 ABS |
3 |
|
Zn** |
213.9 |
0.7 |
0.5 µg/mL=0.12 ABS |
1 |
In the presence
of Cu, do not use a multielement lamp containing
Cu. |
Zr |
360.1 |
0.2 |
400 µg/mL=0.17 ABS |
3 |
|
| * Adapted from
reference 8.6 or from laboratory determinations
** Due
to the limited upper linear range, samples may have to be
diluted, the burner head rotated, or an alternate wavelength
used. The burner head is routinely rotated for Fe and Mg
before analysis.
Flame Types:
- Air/Acetylene mixture, lean, blue flame
- Air/Acetylene mixture, very lean, blue flame
- Nitrous oxide/Acetylene mixture, rich, red flame
- Nitrous oxide/Acetylene mixture, slightly rich, red
flame
- Air/Hydrogen mixture
For the purposes of this method, the following
definitions are used:
Qualitative
detection limit
The concentration (µg/mL) of an element which would yield an
absorbance (ABS) equal to twice the standard deviation of a
series of measurements of an aqueous solution containing the
element. The signal obtained from the aqueous solution must be
distinctly greater than the baseline (8.10). These detection
limits were taken from reference 8.5.
Analytical detection limit
The
lowest concentration (µg/mL) of an
element that can be reliably quantitated. This limit is the
largest value obtained from any of the three calculations: a)
Three times the smallest possible non-zero
instrument reading,
b) Two times the average baseline
variation, or
c) The lowest standard used to construct
a concentration-response curve.
One-tenth the concentration of this standard is
considered to be the detection limit if:
The average
reading for this standard is within 20% of it's linear
response. The linearity is determined by the other standards
used to construct the least-squares curve fit.
If the lowest standard ABS reading is more than 20% in
error, then an algorithm is used and the concentration value
is increased in 10% increments until a concentration is
achieved that would display less than 20% error or until the
lowest standard concentration is reached.
Sensitivity
The concentration
(µg/mL) of an element in aqueous
solution which will produce an ABS of 0.0044 (8.6).
Linear Range
The
working range of a specific analyte. The range is considered
linear if doubling the concentration of a standard results in
at least a 75% increase in ABS.
Appendix B Potential
Interferences
|
Ag |
If a multielement lamp
containing Cu is used, a spectral interference may occur
when determining Ag in a sample containing Cu. A narrow
slit should be used in this instance (8.6). |
|
|
Thorium (Th) is a reported
chemical interference (8.14); however, this element is
extremely rare in workplace environments. Analyze the
sample for Th first if both are suspected to be
present. |
|
Al |
Acetic acid, fluoroborate,
Fe, and Ti enhance the Al signal. Ionization should be
controlled by adding an alkali salt (potassium or
lanthanum) to samples and standards. |
|
Au |
Spectral interferences from
Fe have been observed. Palladium, platinum, and cyanide
complexes are reported interferences (8.6). |
|
Ba |
This element is partially
ionized in the N2O/C2H2
flame. To control this interference, the samples and
standards should contain 1,000 µg/mL potassium ion (8.6). |
|
|
When analyzing using the
primary Ba line (553.6 nm), background correction should
be used if a large amount of Ca is present. The Ca can
cause molecular absorption at this line. |
|
Ca |
Sulfate, aluminate,
phosphate, and silicate decrease sensitivity (8.14).
Silicon (Si), Ti, Al, and Zr have also been reported as
interferences (8.6). Using a
N2O/C2H2 flame will
control these interferences; however, samples and
standards should contain 1,000 µg/mL potassium ion to control any
ionization. |
|
|
Acetone from acetylene
tanks has been reported to decrease sensitivity. Tanks
should be changed when the pressure drops below 75 to 85
psig to prevent acetone from entering the flame
(8.9). |
|
Cd |
A possible interference is
Si; however, Si is not significantly soluble using the
mentioned digestion procedures. |
|
Co |
A reported interference is
Ni in concentrations greater than 1,500 µg/mL (8.10). Such levels of Ni are
unusual in industrial environments. If a large amount is
expected, samples should be analyzed for Ni first and
then analyzed using an alternate Co line if Ni
concentrations exceed 1,500 µg/mL. |
|
Cr |
Co, Fe, Ni, Cu, Ba, Al, Mg,
Ca, Na, and other metals have been reported as chemical
interferences (8.6, 8.9, 8.10). Determining Cr in a lean
flame will control these interferences, but with a
decrease in sensitivity (8.9, 8.10). The instrument
should be optimized using a mixed standard containing Fe
and Ni in addition to the Cr when using the
Air/C2H2 flame. The above
interferences are not noticed when a
N2O/C2H2 flame is
used. |
|
Cs |
Solutions should contain
1,000 µg/mL potassium ion to
control ionization. |
|
|
Strong acids may suppress
the signal; therefore, samples and standards should be
matrix-matched. |
|
Cu |
Spectral interferences may
occur when Ni or Fe is contained in the multielement
lamp and in the sample solution. Use a single element Cu
lamp or a narrow slit to circumvent this problem. |
|
|
A large amount of Zn in the
sample may interfere but can be controlled by using a
lean flame (8.10.). |
|
Fe |
A spectral interference may
be observed if the multielement lamp and the sample
solution contain Co. An alternate line for Fe should be
used (8.6). |
|
|
Citric acid, Ni, and
HNO3 may interfere but can be controlled by
using a lean flame and by carefully optimizing burner
height (8.6, 8.10). Silica may also interfere (8.14),
but is not appreciably soluble in the acid digestion
procedures mentioned. |
|
Hf |
The presence of fluoride
greatly enhances the sensitivity in the determination of
Hf. Samples and standards should contain 0.1 M
NH4F to control this effect and to obtain the
best sensitivity (8.6). |
|
In |
A 100-fold or
greater excess of Al, Mg, Cu, Zn, or phosphate will
suppress the signal. |
|
Mg |
Al,
H2SO4, HNO3, Si, Ti,
and HF are reported to interfere. Addition of a
suppressant (lanthanum or potassium) will control these
interferences (8.6, 8.14). Interferences can also be
controlled using a
N2O/C2H2 flame. |
|
Mn |
Phosphate, perchlorate, Fe,
Ni, and Co may interfere but can be controlled by using
a lean flame (8.10). Tungsten (W), Mo, and Si have been
reported to interfere when the pressure in the acetylene
tanks is low (8.14). |
|
Mo |
Many interferences have
been reported for Mo including Fe, Mn, Ni, Cr, Si, and
strontium (Sr). Addition of Al controls these
interferences (8.9, 8.10, 8.14). |
|
Na |
Ionization in the flame can
occur; an ionization suppressant should be added to the
standards and samples (8.6). |
|
Ni |
A spectral interference
from Fe will result when determining Ni in a sample
containing Fe with a multielement lamp containing Fe. An
alternate line should be used. |
|
|
Cr, Co, and Fe (8.9), or
HCl and HClO4 in the presence of these metals
(8.10) have been reported as interferences. They are
controlled by using a lean
Air/C2H2 flame (8.10, 8.14). |
|
Pb |
Al, Be, Th, and Zr in a
1,000-fold molar excess over the Pb
concentration decrease sensitivity (8.14). The digestion
procedure used for Pb does not solubilize a significant
amount of Al, Be, or Zr for them to be a problem in the
analysis. Workplace environments rarely contain
significant amounts of Th along with Pb; however, if
suspected to be present, the sample should also be
analyzed for Th since it is very toxic. |
|
|
Phosphate, carbonate,
iodide, fluoride, and acetate at a 10-fold
excess may also interfere (8.10). Sulfate and Ca in
excess have also been reported as interferences
(8.7). |
|
Pt |
A number of elements
interfere with the determination when using an
Air/C2H2 flame (8.6). These
interferences are minimized when using a
N2O/C2H2 flame. |
|
Sb |
A spectral interference
occurs when Sb is determined at 217.6 nm in the presence
of large amounts of Pb, which has an adjacent line at
217.0 nm. It has been reported that large concentrations
of Cu also absorb at 217.6 nm. In either situation, the
alternate 231.2 nm line for Sb should be used (8.6,
8.7) |
|
|
Cu and Ni have been
reported to suppress Sb sensitivity, but can be
controlled by using a lean flame (8.9, 8.10). |
|
Se |
Background absorption is
severe at the wavelengths used to determine Se.
Background correction should be used (8.6). |
|
|
Large amounts of Ni, Co,
Fe, Cu, Mn, Pb, and other metals, if present in the
sample may form selenides in the flame, decreasing
sensitivity (8.9). |
|
|
Increased sensitivity is
noted when using an Air/H2 flame as compared
to an Air/C2H2 flame. For greatly
enhanced sensitivity, analyze Se by graphite furnace
atomic absorption using a modified matrix containing Ni
(8.15). |
|
Sn |
Alkali metals and alkaline
earths, Cu, Co, Zn, Al, Ti, phosphoric acid
(H3PO4), and
H2SO4 have been reported as
interferences when Air/H2 flames are used.
Interferences are reduced or eliminated in hotter
flames, but sensitivity is greatly reduced (8.6,
8.10). |
|
Te |
A spectral interference may
occur when Cu is contained in the multielement lamp and
in the sample (8.14). |
|
|
Enhanced sensitivity can be
obtained for this element using graphite furnace atomic
absorption analysis of sample solutions modified to
contain a Ni matrix (8.15). |
|
Ti |
Samples and standards
should contain 1,000 µg/mL
potassium ion to control ionization. |
|
|
The Ti signal is enhanced
by many other metals (8.6). |
|
Y |
Samples and standards
should contain 1,000 µg/mL
potassium ion to control ionization. |
|
|
Strong acids may suppress
the signal; therefore, samples and standards should be
matrix-matched. |
|
Zn |
A spectral interference may
occur if the multielement lamp and the sample contain Cu
(8.14). |
|
Zr |
Fluoride, chloride, and
ammonium enhance Zr sensitivity. Sulfate, nitrate, and
nickel bromide decrease sensitivity. Addition of
NH4F will control these interferences
(8.6). |
|
|
| |