DIPROPYLENE GLYCOL METHYL ETHER
1. General Discussion
Dipropylene glycol methyl ether (DPGME) is one of the most commonly used propylene glycol ethers in industry and is discussed in a recently published NEG/NIOSH document. (Ref. 5.1) DPGME is a collective term describing a mixture of structural isomers. In the past, OSHA has determined airborne concentrations based on a method validated by NIOSH (Ref. 5.2). The method specifies collection of the vapors on activated charcoal, desorption of the charcoal with carbon disulfide, and analysis by GC using flame ionization detection. An examination of the Backup Data Report for the NIOSH method (Ref. 5.3) revealed that the desorption efficiency was not constant, the desorption efficiency of the individual isomers of DPGME was not investigated, and the desorption efficiency from wet charcoal was not addressed. The reported desorption efficiency ranged from 60.4% at 2.954 mg to 89.1% at 12.01 mg of DPGME. In cases where the desorption efficiency is not constant, calculations to determine analyte concentrations are complicated through the use of a desorption efficiency curve. Also, a desorption efficiency less than 75% does not meet one of the evaluation requirements used by the Organic Methods Evaluation Branch of the OSHA Salt Lake Technical Center (SLTC). For analytes such as DPGME, which are comprised as mixtures of related compounds, quantitation is accomplished by summing the peak areas of each component and treating the summed areas as one analyte. This is an accepted and convenient practice when using a flame ionization detector because the responses for all of the isomers of DPGME are identical. But if the desorption efficiencies are not the same for each isomer, they must be quantitated separately with individual desorption efficiency corrections, and then the resulting amounts are summed to determine the total amount of DPGME. This procedure is necessary for any method using charcoal collection and carbon disulfide desorption because the relative proportion of isomers in DPGME can vary by lot and manufacturer. Because charcoal will always collect some water from sampled air,
the desorption of DPGME from wet charcoal is an important
consideration as evidenced by evaluations done at SLTC for other
chemically similar analytes. (Refs. The present evaluation was accomplished using a desorption
solvent consisting of 95/5 (v/v) methylene chloride/methanol, which
is used for other chemically similar compounds evaluated at SLTC.
(Refs. The use of 99/1 (v/v) carbon
1.1.2 Toxic effects (This section is for information only and should not be taken as the basis of OSHA policy.) In the review presented in the previously mentioned NEG/NIOSH document, it was concluded that DPGME seems to lack reproductive toxicity, unlike some other chemically similar compounds. (Ref. 5.1) At very high air concentrations, DPGME causes narcosis in
animals. It is expected that severe exposure would produce similar
effects in humans, but high concentrations are disagreeable and not
tolerated. Also, concentrations over 200 ppm (40% saturated
atmosphere) are difficult to attain, which suggests these high
concentrations would not likely be found in workplace air. DPGME at
300 ppm caused eye and nasal irritation to humans. There was no
evidence of skin irritation from prolonged or repeated contact with
the pure liquid. High vapor concentrations or direct contact of the
eyes with the liquid causes transient irritation. (Ref.
5.7) The OSHA 1.1.3 Workplace exposure DPGME is used as a solvent for paints, lacquers, resins, dyes,
oil/greases, cleaners and cellulose and as a
1.1.4 Physical properties (Ref.
5.1 unless otherwise noted)
The analyte air concentrations throughout this method are based on the recommended sampling and analytical parameters. Air concentrations listed in ppm and ppb are referenced to 25°C and 101.3 kPa (760 mmHg).
The detection limit of the analytical procedure is 0.13 ng. This is the amount of analyte that will give a response that is significantly different from the background response of a reagent blank. (Sections 4.1 and 4.2) 1.2.2 Detection limit of the overall procedure The detection limit of the overall procedure is 1.5 µg per sample (25 ppb or 150 µg/m3). This is the amount of analyte spiked on the sampler that will give a response that is significantly different from the background response of a sampler blank. (Sections 4.1 and 4.3) 1.2.3 Reliable quantitation limit The reliable quantitation limit is 5.1 µg per sample (84 ppb or 510 µg/m3). This is the amount of analyte spiked on a sampler that will give a signal that is considered the lower limit for precise quantitative measurements. (Section 4.4) 1.2.4 Precision (analytical procedure) The precision of the analytical procedure, measured as the pooled relative standard deviation over a concentration range equivalent to 0.5 to 2 times the target concentration, is 0.14%. (Section 4.5) 1.2.5 Precision (overall procedure) The precision of the overall procedure at the 95% confidence
level for the ambient temperature 1.2.6 Recovery The recovery of DPGME from samples used in a 15-day storage test remained above 99% when the samples were stored at ambient temperatures. (Section 4.7) 1.2.7 Reproducibility Six samples collected from controlled test atmospheres, with a draft copy of this procedure, were submitted to an SLTC service branch for analysis. The samples were analyzed after 27 days of storage. No individual sample result deviated from its theoretical value by more than the precision reported in Section 1.2.5. (Section 4.8) 2. Sampling Procedure
2.1.2 Samples are collected with 7-cm × 4-mm i.d. × 6-mm o.d.
glass sampling tubes packed with two sections of coconut shell
charcoal. The front section contains 100 mg and the back section
contains 50 mg of charcoal. The sections are held in place with
glass wool plugs and are separated by a urethane foam plug. For this
evaluation, commercially prepared sampling tubes were purchased from
SKC, Inc. (Fullerton, CA, Catalog No. 2.2 Reagents None required 2.3 Technique
2.3.2 Connect the sampling tube to the sampling pump with
flexible, 2.3.3 Air being sampled should not pass through any hose or tubing before entering the sampling tube. 2.3.4 To avoid channeling, place the sampling tube vertically in the employee's breathing zone. Position the sampler so it does not impede work performance or safety. 2.3.5 After sampling for the appropriate time, immediately remove
the sampling tube and seal it with plastic caps. Wrap each sample
lengthwise with a Form 2.3.6 Submit at least one blank sampling tube with each sample set. Blanks should be handled in the same manner as samples, except no air is drawn through them. 2.3.7 Record sample volumes (in liters of air) for each sample. 2.3.8 List any compounds that could be considered potential interferences, especially solvents, that are being used in the sampling area. 2.3.9 Ship any bulk sample(s) in a container separate from the air samples. 2.4 Sampler capacity Sampler capacity is determined by measuring how much air can be
sampled before breakthrough of analyte through the sampler occurs,
i.e., the sampler capacity is exceeded. Breakthrough is considered to
occur when the effluent from the sampler contains a concentration of
analyte that is 5% of the upstream concentration (5% breakthrough).
Testing for breakthrough was performed by using a total hydrocarbon
analyzer to monitor the effluent from sampling tubes containing only
the 100-mg section of charcoal while sampling at 0.1 L/min from an
atmosphere containing 202 ppm of DPGME. The atmosphere was at
approximately 80% relative humidity and 2.5 Desorption efficiency
2.5.2 The desorption efficiency at 0.05, 0.1, and 0.2 times the target concentration was found to be 97.0%, 98.0%, and 98.2% respectively. (Section 4.10.1) 2.5.3 Desorbed samples remain stable for at least 24 h. (Section 4.10.2) 2.6 Recommended air volume and sampling rate
2.6.2 For short-term samples collect 1.5 L at 0.1 L/min
2.6.3 When short-term samples are collected, the air concentration equivalent to the reliable quantitation limit becomes larger. For example, the reliable quantitation limit is 560 ppb (3390 µg/m3) when 1.5 L is collected. 2.7 Interferences (sampling)
2.7.2 Suspected interferences should be reported to the laboratory with submitted samples. 2.8 Safety precautions (sampling)
2.8.2 Wear eye protection when breaking the ends of the charcoal tubes. 2.8.3 Follow all safety procedures that apply to the work area being sampled. 3. Analytical Procedure
3.1.2 A GC column capable of separating the analyte of interest
from the desorption solvent, internal standard and any
interferences. A 3.1.3 An electronic integrator or some other suitable means of measuring peak areas. A Waters 860 Networking Computer System was used in this evaluation. 3.1.4 Two-milliliter vials with Teflon®-lined caps. 3.1.5 A dispenser capable of delivering 1.0 mL of desorption
solvent to prepare standards and samples. If a dispenser is not
available, a 3.2 Reagents
3.2.2 Methylene chloride, chromatographic grade. Burdick and Jackson Lot BB551 was used in this evaluation. 3.2.3 Methanol, chromatographic grade. Fisher Lot 913607 was used in this evaluation. 3.2.4 A suitable internal standard, reagent grade. Aldrich
Chemical Lot 11329LW 3.2.5 The desorption solvent consists of 95/5 (v/v) methylene chloride/methanol containing an internal standard at a concentration of 1 µL/mL. 3.2.6 GC grade nitrogen, air, and hydrogen. 3.3 Standard preparation
3.3.2 Bracket sample concentrations with working standard concentrations. If samples fall outside of the concentration range of prepared standards, prepare and analyze additional standards to ascertain the linearity of response. 3.4 Sample preparation
3.4.2 Add 1.0 mL of desorption solvent to each vial using the same dispenser as used for preparation of standards. 3.4.3 Immediately cap the vials and shake them several times over the next 15 min. 3.5 Analysis 3.5.1 GC conditions
3.5.2 Peak areas are measured by an integrator or other suitable means. The areas of the isomers of DPGME are summed together and treated as one analyte. 3.5.3 An internal standard (ISTD) calibration method is used. A
calibration curve is prepared by plotting micrograms of DPGME per
sample versus
Figure 3.5.3. Calibration curve constructed from the data in Table 4.5. The equation of the line is 3.6 Interferences (analytical)
3.6.2 GC parameters (i.e. column and column temperature) may be changed to possibly circumvent interferences. 3.6.3 When necessary, the identity or purity of an analyte peak may be confirmed with additional analytical data. (Section 4.11) 3.7 Calculations The DPGME concentration for samples is obtained from the
appropriate calibration curve in terms of micrograms of analyte per
sample, uncorrected for desorption efficiency. The air concentration
is calculated using the following formulae. The back
mg/m3 = (micrograms of DPGME per sample)/((liters of air sampled)(desorption efficiency)) where the desorption efficiency = 0.994 ppm = (mg/m3)(24.46)/(molecular weight of analyte) = (mg/m3)(0.1650) where 24.46 is the molar volume at 25°C and 101.3 kPa (760
mmHg) 3.8 Safety precautions (analytical)
3.8.2 Avoid skin contact and inhalation of all chemicals. 3.8.3 Wear safety glasses and a lab coat at all times while in the lab area. 4. Backup Data
Detection limits (DL), in general, are defined as the amount (or concentration) of analyte that gives a response (YDL) that is significantly different (three standard deviations (SDBR)) from the background response (YBR). YDL - YBR = 3(SDBR) The direct measurement of YBR and SDBR in chromatographic methods is typically inconvenient and difficult because YBR is usually extremely low. Estimates of these parameters can be made with data obtained from the analysis of a series of analytical standards or samples whose responses are in the vicinity of the background response. The regression curve obtained for a plot of instrument response versus concentration of analyte will usually be linear. Assuming SDBR and the precision of data about the curve are similar, the standard error of estimate (SEE) for the regression curve can be substituted for SDBR in the above equation. The following calculations derive a formula for DL:
At point YDL on the regression
curve
therefore
Substituting 3(SEE) + YBR for YDL gives
4.2 Detection limit of the analytical procedure (DLAP) The DLAP is measured as the mass of analyte actually introduced
into the chromatographic column. Ten analytical standards were
prepared in equal descending increments with the highest standard
containing 12.0 µg/mL. This is the concentration that would
produce a peak approximately 10 times the baseline noise of a reagent
blank. These standards, plus a solvent blank, were analyzed with the
recommended analytical parameters
4.3 Detection limit of the overall procedure (DLOP) The DLOP is measured as mass per sample and expressed as equivalent air concentrations, based on the recommended sampling parameters. Ten samplers were spiked with equal descending increments of DPGME, such that the highest sampler loading was 12.0 µg/sample. This is the amount, when spiked on a sampler, that would produce a peak approximately 10 times the baseline noise for a sample blank. These spiked samplers, plus a sample blank, were analyzed with the recommended analytical parameters, and the data obtained used to calculate the required parameters (A and SEE) for the calculation of the DLOP. Values of 203.9 and 104.3 were obtained for A and SEE respectively. The DLOP was calculated to be 1.5 µg/sample (25 ppb, 150 µg/m3).
4.4 Reliable quantitation limit (RQL) The RQL is considered the lower limit for precise quantitative measurements. It is determined from the regression line data obtained for the calculation of the DLOP (Section 4.3). The RQL is defined as the amount of analyte that gives a response (YRQL) such that YRQL - YBR = 10(SDBR) therefore
The RQL was calculated to be 5.1 µg/sample (84 ppb, 510 µg/m3). Recovery at this concentration is 92.1%.
4.5 Precision (analytical method) The precision of the analytical procedure is defined as the pooled relative standard deviation (RSDP). Relative standard deviations were determined from six replicate injections of DPGME standards at 0.5, 0.75, 1, 1.5, and 2 times the target concentration. After assuring that the RSDs satisfy the Cochran test for homogeneity at the 95% confidence level, the RSDP was calculated to be 0.14%.
The Cochran test for homogeneity: = 0.4433 Because the g statistic does not exceed the critical value of 0.5065, the RSDs can be considered equal and they can be pooled (RSDP) to give an estimated RSD for the concentration range studied. = 0.14% 4.6 Precision (overall procedure) The precision of the overall procedure is determined from the storage data in Section 4.7. The determination of the standard error of estimate (SEER) for a regression line plotted through the graphed storage data allows the inclusion of storage time as one of the factors affecting overall precision. The SEER is similar to the standard deviation, except it is a measure of dispersion of data about a regression line instead of about a mean. It is determined with the following equation:
An additional 5% for pump error (SP) is added to the SEER by the addition of variances to obtain the total standard error of estimate. The precision at the 95% confidence level is obtained by
multiplying the standard error of estimate (with pump error included)
by 1.96 (the 4.7 Storage test Storage samples were generated by sampling from test atmospheres
containing DPGME at the target concentration. Six samples were
analyzed immediately after generation, fifteen were stored in a
refrigerator at 0°C, and fifteen were stored in a closed drawer at
ambient temperatures of
4.8 Reproducibility Six samples were prepared by collecting them from a controlled test
atmosphere similar to that which was used in the collection of the
storage samples. The samples were submitted to an SLTC service branch
for analysis. Samples
4.9 Sampler capacity Sampler capacity was determined by using a total hydrocarbon
analyzer to monitor the effluent from sampling tubes containing only
the
4.10 Desorption efficiency and stability of desorbed samples
The desorption efficiencies (DE) for DPGME were determined by
liquid-spiking the
4.10.2 Stability of desorbed samples The stability of desorbed samples was investigated by reanalyzing
the target concentration samples 24 h after initial analysis. After
the original analysis was performed three vials were recapped with
new septa while the remaining three retained their punctured septa.
The samples were reanalyzed with fresh standards. The average
percent change was +0.3% for samples that were resealed with new
septa, and +0.2% for those that retained their punctured
septa.
4.11 Qualitative analysis The isomers of DPGME can easily be separated and identified by
GC/MS. Mass spectra for six of the isomers, which were separated using
similar conditions given in Section 3.5, were obtained from a
&nsbsp;
5. References
5.2. "NIOSH Manual of Analytical Methods",
2nd ed. Vol. 2; U.S. Department of Health and Human Services, Public
Health Service, Centers for Disease Control, National Institute for
Occupational Safety and Health; Cincinnati, OH, 1977, Method S69, DHEW
(NIOSH) Publication No. 5.3. "Documentation of the NIOSH Validation
Tests"; U.S. Department of Health and Human Services, Public
Health Service, Centers for Disease Control, National Institute for
Occupational Safety and Health, Division of Physical Sciences and
Engineering; Cincinnati, OH, 1977, Backup Data Report No. S69, DHEW
(NIOSH) Publication No. 5.4. "OSHA Analytical Methods Manual" U.S.
Department of Labor, Occupational Safety and Health Administration;
OSHA Salt Lake Technical Center: Salt Lake City, UT, 1990; Method 79;
American Conference of Governmental Industrial Hygienists (ACGIH):
Cincinnati, OH, ISBN: 5.5. Elskamp, C.J. "OSHA Method No. 99;
5.6. Elskamp, C.J. "OSHA Method No. 83;
2-Butoxyethanol and 5.7. Hathaway, G.J., et al., Ed., "Proctor and Hughes' Chemical Hazards of the Workplace", 3rd ed., Van Nostrand Reinhold, New York, 1991. 5.8. Department of Labor Fed. Regist. 1993, 58 (No. 124, Wed. June 30), 35344. 5.9. "American Conference of Governmental Industrial Hygienists: Documentation of the Threshold Limit Values"; 5th ed., p. 221, Cincinnati, OH (1986). 5.10. Cheminfo Database on CCINFO
5.11 "Hawley's Condensed Chemical Dictionary"; 12th ed.; Van Nostrand Reinhold Company, New York, NY, 1993.
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||