m-, o- and p-PHENYLENEDIAMINE
Method no.: |
87 |
|
|
|
Matrix: |
Air |
|
Procedure: |
Samples are collected closed-face by
drawing known volumes of air through sampling devices consisting of
three-piece cassettes, each containing two sulfuric
acid-treated glass fiber filters separated by the ring
section. The sample filters are extracted with an aqueous EDTA
solution and the extracts are analyzed for the free amines by HPLC
using a UV detector. |
|
Recommended air volume and sampling rate: |
100 L at 1 L/min |
|
|
|
|
phenylenediamine
|
|
meta- |
ortho- |
para- |
|
|
Target concentration: |
0.10 mg/m3 |
0.10 mg/m3 |
0.10 mg/m3 |
|
Reliable quantitation limit: (based on a 100-L air
volume) |
0.56 µg/m3 |
2.1 µg/m3 |
0.44 µg/m3 |
|
Standard error of estimate at the target
concentration: (Section
4.7.) |
5.3% |
7.9% |
5.7% |
|
|
|
Status of method: |
Evaluated method. This method has been subjected to
the established evaluation procedures of the Organic Methods
Evaluation Branch. |
|
Date: February 1991 |
Chemist: Carl J.
Elskamp |
Organic Methods Evaluation Branch OSHA Analytical
Laboratory Salt Lake City, Utah
1. General Discussion
1.1. Background
1.1.1. History
Phenylenediamines are particularly difficult to determine in air
because they are very susceptible to oxidation reactions, which
create sample stability and analytical problems. There are a number
of methods in the literature which utilize various bubbler or
impinger solutions for collection of air samples. (Refs. 5.1.-5.5.)
Sampling in this manner is very inconvenient and can potentially be
unsafe because some of the collection solutions are toxic. There are
also special shipping regulations which complicate the transport of
samples to the laboratory for analysis. There is a published method
for collection of o-phenylenediamine vapors using glass
tubes packed with Tenax-GC. (Ref.
5.6.) Sample stability problems may exist with this method
because the sampler is wrapped with aluminum foil during sampling to
protect it from light. No storage tests were reported in the method.
Also, this methodology does not address collection of aerosols or
dusts, which may not be effectively collected with a solid sorbent
tube.
Air sampling and analytical procedures have previously been
evaluated by the OSHA Salt Lake Analytical Laboratory for a number
of other aromatic amines which utilize a sampling device containing
glass fiber filters coated with dilute sulfuric acid. (Refs.
5.7.-5.11.)
With this sampling device, the amines are converted to amine salts
on the filter. Not only does this provide for good collection
efficiencies, it also eliminates stability problems because the
salts are very stable compared to the free amines. With the
exception of diphenylamine and N-isopropylaniline,
which are analyzed by HPLC, the analysis scheme involves converting
the amine salts to the free amines using excess sodium hydroxide and
extracting the amines into toluene. The amines are then derivatized
with heptafluorobutyric acid anhydride and the derivatives are
analyzed by gas chromatography. This procedure is more sensitive
than direct HPLC analysis of the free amines and was evaluated for
the analysis of phenylenediamines because the target concentrations
are fairly low. However, stability problems arose, especially for
p-phenylenediamine, when the amine salts were converted
to the free amines with sodium hydroxide. The instability was most
likely due to oxidation of the free amines. It became apparent that
the free phenylenediamines are too unstable and the extracted
samples must be kept acidic until analyzed. The acidic extract is
analyzed directly by HPLC. The amine salts are converted to the free
amines upon injection by utilizing a mobile phase buffered to pH 7.
A phosphate buffer was chosen because it has a high buffer capacity
at this pH. Although all three isomers were sufficiently stable on
the acid-treated filters, the ortho isomer was somewhat
unstable in the acidic extract when deionized water was used to
extract the filters. The stability of extracted samples for this
isomer was greatly improved by using an aqueous EDTA solution
instead. The stability of the other two isomers appeared to be the
same for both water and EDTA solution extractions.
1.1.2. Toxic effects (This section is for information only and
should not be taken as the basis of OSHA policy.)
Exposure to phenylenediamines has been reported to affect the
kidney, liver, and blood. Inhalation causes respiratory problems and
asthma, but the most common toxic effect is dermatitis. (Ref.
5.12.) The current OSHA PEL and ACGIH TLV are 0.1
mg/m3 for p-phenylenediamine
with skin notations. ACGIH is now considering the same TLV for m-
and o-phenylenediamine and is also considering adding
o-phenylenediamine to its suspected human carcinogen
list. Currently there are no OSHA exposure limits for m- or
o-phenylenediamine.
1.1.3. Workplace exposure
The major uses for phenylenediamines are in the manufacture of
dyes. They are also used to dye hair and fur, as photographic
development agents, curing agents for epoxy resins, vulcanization
accelerators, and as components of gasoline antioxidants. (Ref.
5.13.)
1.1.4. Physical properties and other descriptive information (Ref.
5.14.)
|
|
|
|
|
|
|
phenylenediamine
|
|
meta- |
ortho- |
para- |
|
|
CAS no.: |
108-45-2 |
95-54-5 |
106-50-3 |
molecular weight: |
108.14 |
108.14 |
108.14 |
melting point: |
62-63°C |
103-104°C |
145-147°C |
boiling point: |
284-287°C |
256-258°C |
267°C |
description: |
white crystals turning red on exposure of air |
brownish-yellow crystals |
white to slightly red crystals; darkens on exposure to
air |
solubility: |
soluble in water, methanol, ethanol, chloroform, acetone,
DMF, MEK, dioxane |
slightly soluble in water; freely soluble in alcohol,
chloroform, ether |
soluble in 100 parts cold water; soluble in alcohol,
chloroform, ether |
structural formula: |
meta |
ortho |
para |
synonyms: (Ref.
5.13.)
m-phenylenediamine- 1,3-diaminobenzene;
1,3-phenylenediamine; 3-aminoaniline;
benzene, 1,3-diamino;
m-aminoaniline;
m-benzenediamine;
m-diaminobenzene;
m-fenylendiamin (Czech.);
meta-aminoaniline;
meta-benzenediamine;
meta-diaminobenzene; metaphenylenediamine;
phenylenediamine, meta, solid; CI 76025; CI Developer 11;
Developer 11; Developer C; Developer H; Developer M;
Direct Brown BR; Direct Brown GG
o-phenylenediamine- 1,2-benzenediamine;
1,2-diaminobenzene; 1,2-phenyl-enediamine;
2-aminoaniline;
o-benzenediamine;
o-diaminobenzene; orthamine; CI 76010; CI
Oxidation Base
p-phenylenediamine- 1,4-diaminobenzene;
1,4-phenylenediamine; 4-amino-aniline;
p-aminoaniline;
p-benzenediamine;
p-diaminobenzene; Para; paraphenylenediamine;
phenylenediamine, para, solid; Pelagol D; Pelagol DR;
Pelagol Grey D; Peltol D; PPD; Renal PF; Santoflex LC;
Tertral D; Ursol D; USAF EK-394; Z0BA Black
D; BASF Ursol D; Benzofur D; CI 76060; CI Developer 13; CI
Oxidation Base 10; Developer 13; Developer PF; Durafur
Black R; fenylenodwuamina (Polish); Fouramine D; Fourine
1; Fur Black 41866; Fur Black 41867; Fur Yellow; Furro D;
Futramine D; Nako D; Orsin; Oxidation Base 10
|
The analyte air concentrations throughout this method are based on
the recommended sampling and analytical parameters.
1.2. Limit defining parameters
1.2.1. Detection limit of the analytical procedure
The detection limits of the analytical procedure are 0.14, 0.53,
and 0.11 ng per injection for m-, o-, and
p-phenylenediamine respectively. These are the amounts
of each analyte that will produce peaks with heights that are
approximately five times the baseline noise. (Section
4.1.)
1.2.2. Detection limit of the overall procedure
The detection limits of the overall procedure are 56, 211, and 44
ng per sample for m-, o-,and
p-phenylenediamine respectively. These are the amounts
of each analyte spiked on sample filters that allow recovery of
analytes equivalent to the respective detection limits of the
analytical procedure. These detection limits correspond to air
concentrations of 0.56, 2.1, and 0.44
µg/m3 for m-, o-,
and p-phenylenediamine respectively. (Section
4.2.)
1.2.3. Reliable quantitation limit
The reliable quantitation limits are 56, 211, and 44 ng per
sample for m-, o-, and p-phenylenediamine respectively.
These are the smallest amounts of each analyte spiked on sample
filters that can be quantitated within the requirements of a
recovery of at least 75% and a precision (±1.96 SD) of ±25% or
better. These reliable quantitation limits correspond to air
concentrations of 0.56, 2.1, and 0.44 g/m3
for m-, o-, and
p-phenylenediamine respectively. (Section
4.3.)
The reliable quantitation limits and detection limits reported
in this method are based upon optimization of the instrument for the
smallest possible amount of analyte. When the target concentration
of an analyte is exceptionally higher than these limits, they may
not be attainable at the routine operating parameters.
1.2.4. Instrument response to the analyte
The instrument response over concentration ranges representing
0.5 to 2 times the target concentrations is linear for the three
analytes. (Section
4.4.)
1.2.5. Recovery
The recoveries of m-, o-, and p-phenylenediamine from samples
used in 15-day storage tests remained above 98%, 84%, and 98%
respectively. The sample filters were stored in cassettes in a
closed drawer at approximately 21°C. (Section
4.5.)
1.2.6. Precision (analytical method only)
The pooled coefficients of variation obtained from replicate
injections of analytical standards at 0.5, 1, and 2 times the target
concentrations are 0.0063, 0.0095, and 0.0090 for m-,
o-, and p-phenylenediamine
respectively. (Section
4.6.)
1.2.7. Precision (overall procedure)
The precisions at the 95% confidence level for the 15-day storage
tests are ±10.3, ±15.4 and ±11.1% for m-, o-,
and p-phenylenediamine respectively. These
include an additional ±5% for sampling error. The sample filters
were stored in cassettes in a closed drawer at approximately 21°C.
(Section
4.7.)
1.2.8. Reproducibility
Six samples, spiked by liquid injection, and a draft copy of this
procedure were given to a chemist unassociated with this evaluation.
The samples were analyzed after 48 days of storage at approximately
0°C. No individual sample result deviated from its theoretical value
by more than the corresponding precision of the overall procedure as
reported in Section
1.2.7. (Section
4.8.)
1.3. Advantages
1.3.1. The acid-treated filter provides a convenient method of
sampling for a number of aromatic amines.
1.3.2. The analysis is rapid, sensitive, and precise.
1.4. Disadvantages
None
2. Sampling Procedure
2.1. Apparatus
2.1.1. Samples are collected using a personal sampling pump that
can be calibrated within ±5% of the recommended flow rate with the
sampling device attached.
2.1.2. Samples are collected closed-face using a sampling device
consisting of two sulfuric-acid treated
37-mm Gelman type A/E glass fiber filters contained in
a three-piece polystyrene cassette. The filters are
prepared by soaking each filter with 0.5 mL of 0.26 N sulfuric acid.
(0.26 N sulfuric acid can be prepared by diluting 1.5 mL of 36 N
sulfuric acid to 200 mL with deionized water.) The filters are dried
in an oven at 100°C for 1 h and then assembled into
three-piece 37-mm cassettes without
support pads. The front filter is separated from the back filter by
the ring section. The cassettes are sealed with shrink bands and the
ends are plugged with plastic plugs. An unassembled sampling device
is shown in Figure
2.1.2.
2.2. Reagents
None required
2.3. Sampling technique
2.3.1. Remove the plastic plugs from the sampling device
immediately before sampling.
2.3.2. Attach the sampling device to the sampling pump with
flexible tubing and place the device in the employee's breathing
zone.
2.3.3. Seal the sampling device with the plastic plugs
immediately after sampling.
2.3.4. Seal and identify each sampling device with an OSHA Form
21.
2.3.5. Submit at least one blank sampling device with each sample
set. Handle the blanks in the same manner as the air samples, but
draw no air through them.
2.3.6. Record the volume of air sampled (in liters) for each
sample, along with any potential interferences.
2.4. Collection efficiency
A collection efficiency study was conducted by drawing humid air
through a sampling device that was attached to an impinger containing
milligram amounts of the pure amines. The inlet of the impinger was
attached to a humid air generator so air at approximately 80% relative
humidity could be drawn through it. The impinger was heated to
approximately 60°C with a heat tape. After sampling for 4 h at 1
L/min, the filters were analyzed. None of the amines were found on the
back filter and approximately 230, 670, and 130 µg of m-,
o-, and p-phenylenediamine
respectively were found on the front filter. This corresponds to air
concentrations of 0.96, 2.8, and 0.54 mg/m3
for m-, o-, and
p-phenylenediamine respectively. Although this test
demonstrates that the sampler has more than adequate capacity to
collect larger air volumes at concentrations much higher than the
target concentrations of 0.10 mg/m3, a
recommended air volume of 100 L was chosen to assure a sufficient
safety margin and to maintain consistency with previously evaluated
methods for aromatic amines.
2.5. Extraction efficiency
2.5.1. The average extraction efficiencies from six filters for
each amine spiked at the target concentration are 100.8%, 97.6%, and
101.0% for m-, o-, and
p-phenylenediamine respectively. (Section
4.9.)
2.5.2. The stability of extracted samples was verified by
reanalyzing the extraction efficiency samples 24 h later using fresh
standards. The average recoveries for the reanalyzed samples were
99.2%, 93.5%, and 98.4% for m-, o-, and
p-phenylenediamine respectively. (Section
4.9.)
2.6. Recommended air volume and sampling rate
2.6.1. The recommended air volume is 100 L.
2.6.2. The recommended sampling rate is 1 L/min.
2.6.3. When short-term samples are required, the reliable
quantitation limits will be larger. For example, the reliable
quantitation limit for p-phenylenediamine for a
15-L air sample would be 2.9
µg/m3.
2.7. Interferences (sampling)
2.7.1. Any compound in the sampled air that will react with the
sulfuric acid on the treated filters or with the collected analyte
is a potential sampling interference.
2.7.2. Suspected interferences should be reported to the
laboratory with submitted samples.
2.8. Safety precautions (sampling)
2.8.1. Attach the sampling equipment to the employees so that it
will not interfere with work performance or safety.
2.8.2. Follow all safety procedures that apply to the work area
being sampled.
3. Analytical Procedure
3.1. Apparatus
3.1.1. An HPLC system equipped with an ultraviolet detector. A
Hewlett-Packard 1050 Series HPLC consisting of a
pumping system, programmable variable wavelength detector and an
autosampler was used in this evaluation.
3.1.2. An HPLC column capable of separating the analyte from the
solvent and interferences. A Waters Radial-Pak
100-mm × 8-mm i.d. cartridge containing
Nova-Pak C18
(end-capped 5-µm spherical particles) was
used in conjunction with a Waters RCM-100 radial compression module.
3.1.3. An electronic integrator or some other suitable means of
measuring peak areas or heights. A Hewlett-Packard
18652A A/D converter interfaced to a Hewlett-Packard
3357 Lab Automation Data System was used in this evaluation.
3.1.4. Small resealable glass vials with Teflon-lined caps
capable of holding 4 mL. WISP-type autosampler vials
were used in this evaluation.
3.1.5. A dispenser capable of delivering 2.0 mL of the EDTA
extraction solution.
3.1.6. A pH meter with a combination electrode is used in the
preparation of the HPLC mobile phase.
3.2. Reagents
3.2.1. m-, o-, and
p-Phenylenediamine, reagent grade. The amines used in
this evaluation were purchased from Aldrich Chemical Company, Inc.
(Milwaukee, WI).
3.2.2. HPLC grade acetonitrile and water. The acetonitrile used
in this evaluation was "Optima" brand from Fisher Chemical (Fair
Lawn, NJ) and the water was from an in-house Millipore
Milli-Q water purification system.
3.2.3. Ethylenediaminetetraacetic acid (EDTA), reagent grade. A
0.1 g/L EDTA aqueous solution is used to extract the sample filters.
3.2.4. Phosphoric acid, 10 N sulfuric acid, and dibasic sodium
phosphate
(Na2HPO4),
reagent grade.
3.3. Standard preparation
3.3.1. Restrict the use of pure compounds and concentrated
standards to regulated areas. Prepare concentrated stock standards
by accurately weighing approximately 50 mg of each amine into a
25-mL volumetric flask. Initially dissolve the amines
with about 20 mL of 10 N sulfuric acid.
m-Phenylenediamine is readily soluble. Sonication can
be used to expedite dissolution of the o- and
p-phenylenediamine. After the amines are totally
dissolved, dilute to the mark with additional 10 N sulfuric acid and
thoroughly mix the solution. Stock standards are stable for at least
six months when stored in brown bottles in a refrigerator.
3.3.2. Prepare analytical standards by injecting microliter
amounts of stock standards into 4-mL vials containing
2.0 mL of the EDTA extraction solution using a 10-µl
syringe. Immediately rinse the syringe with water after the
standards are prepared. If 5.00 µL of a 2.00 µg/µL (50.00 mg of
amine to 25.00 mL with 10 N sulfuric acid) stock solution is
injected into 2.0 mL of the EDTA extraction solution, the analytical
standard would be equivalent to 10.0 µg of amine per sample, or 0.10
mg/m3 for a 100-L air sample.
3.3.3. Bracket sample concentrations with analytical standard
concentrations. If sample concentrations are higher than the upper
range of prepared standards, prepare higher standards to ascertain
detector response or dilute the extract of the samples using the
EDTA extraction solution.
3.4. Sample preparation
3.4.1. Transfer the sample filters to individual 4-mL vials.
3.4.2. Add 2.0 mL of the aqueous EDTA extraction solution to each
vial.
3.4.3. Recap and periodically invert the vials over a period of
10 min.
3.4.4. Analyze by making direct injections of the extracts.
3.5. Analysis
3.5.1. HPLC conditions and information
mobile phase: |
0.05 M sodium phosphate in 95/5, water/acetonitrile at
pH 7.0. Prepare by adding 7.1 g of dibasic sodium phosphate
per 1 L of the final total volume of mobile phase to the
water. After the sodium phosphate has dissolved (expedited
using sonication) adjust the pH of this aqueous solution to
7.0 with phosphoric acid. Add the acetonitrile to the
pH-adjusted aqueous solution and mix
thoroughly. |
flow rate: |
2 mL/min |
UV detector wavelength: |
240 nm |
injection volume: |
5 µL |
column: |
Waters Radial-Pak 100-mm × 8-mm i.d.
cartridge containing Nova Pak
C18 |
retention times: |
p-phenylenediamine, 2.4 min
m-phenylenediamine, 3.8 min o-phenylenediamine, 7.0
min |
chromatogram: |
Section
4.10. |
3.5.2. Measure peak areas or heights by use of an integrator or
by other suitable means.
3.5.3. Construct a calibration curve by plotting response (peak
areas or heights) of standard injections versus micrograms of
analyte per sample. Bracket sample concentrations with standards.
3.6. Interferences (analytical)
3.6.1. Any compound that elutes in the same general time as the
amine of interest is a potential interference. Suspected
interferences reported to the laboratory with submitted samples by
the industrial hygienist must be considered before samples are
extracted.
3.6.2. HPLC parameters may be changed to possibly circumvent
interferences.
3.6.3. Retention time on a single column is not considered proof
of chemical identity. Analyte identity should be confirmed by mass
spectrometry if possible.
3.7. Calculations
The analyte concentration for samples is obtained from the
calibration curve in micrograms of analyte per sample. If any analyte
is found on any back filter, that amount is added to the amount found
on the corresponding front filter. If any analyte is found on the
blank filters, the combined amount is subtracted from the combined
sample amounts. The air concentrations are calculated using the
following formula.
mg/m3 = |
(micrograms of analyte per sample)
(liters of air sampled)(extraction efficiency)
|
3.8. Safety precautions (analytical)
3.8.1. Restrict the use of pure compounds and concentrated
standards to regulated areas. Avoid skin contact and inhalation of
all chemicals.
3.8.2. Restrict the use of all chemicals to a fume hood if
possible.
3.8.3. Wear safety glasses and a lab coat at all times while in
the lab area.
4. Backup Data
4.1. Detection limit of the analytical procedure
The injection volume (5 µL) listed in Section
3.5.1. was used in the determination of the detection limits of
the analytical procedure. The detection limits of 0.14, 0.53, and 0.11
ng per injection were determined by analyzing dilute standards
equivalent to 56, 211, and 44 ng per sample for m-,
o-, and p-phenylenediamine
respectively. These amounts were judged to give peaks with heights
approximately five times the baseline noise. A chromatogram is shown
in Figure
4.1.
4.2. Detection limit of the overall procedure
The detection limits of the overall procedure were determined by
analyzing filters spiked with loadings equivalent to the detection
limits of the analytical procedure. Samples were prepared by injecting
56, 211, and 44 ng of m-, o-, and
p-phenylenediamine respectively onto
acid-treated filters. These amounts are equivalent to
0.56, 2.1, and 0.44 µg/m3 for m-,
o-, and p-phenylenediamine
respectively.
Table 4.2.1. Detection Limit of the Overall Procedure
for m-Phenylenediamine
|
sample no. |
ng spiked |
ng recovered |
|
1 2 3 4 5 6 |
56 56 56 56 56 56 |
56.6 56.3 63.0 53.3 47.5 53.6 |
|
Table 4.2.2. Detection Limit of the Overall Procedure
for o-Phenylenediamine
|
sample no. |
ng spiked |
ng recovered |
|
1 2 3 4 5 6 |
211 211 211 211 211 211 |
204 213 200 197 195 193 |
|
Table 4.2.3. Detection Limit of the Overall Procedure
p-Phenylenediamine
|
sample no. |
ng spiked |
ng recovered |
|
1 2 3 4 5 6 |
44 44 44 44 44 44 |
39.3 42.3 47.2 45.6 42.4 44.2 |
|
4.3. Reliable quantitation limit
The reliable quantitation limits were determined by analyzing
filters spiked with loadings equivalent to the detection limits of the
analytical procedure. Samples were prepared by injecting 56, 211, and
44 ng of m-, o-, and
p-phenylenediamine respectively onto
acid-treated filters. These amounts are equivalent to
0.56, 2.1, and 0.44 µg/m3 for m-,
o-, and p-phenylenediamine
respectively.
Table 4.3.1. Reliable Quantitation Limit for
m-Phenylenediamine (Based on samples and data of Table
4.2.1.)
|
sample no. |
percent recovered |
|
|
|
1 2 3 4 5 6 |
101.1 100.5 112.5
95.2 84.8 95.7 |
=
SD = Precision
= = |
98.3 9.1 (1.96)(±9.1) ±17.8 |
|
Table 4.3.2. Reliable Quantitation Limit for
o-Phenylenediamine (Based on samples and data of Table
4.2.2.)
|
sample no. |
percent recovered |
|
|
|
1 2 3 4 5 6 |
96.7 100.9
94.8 93.4 92.4 91.5 |
=
SD = Precision
= = |
95.0 3.4 (1.96)(±3.4) ±6.7 |
|
Table 4.3.3. Reliable Quantitation Limit for
p-Phenylenediamine (Based on samples and data of Table
4.2.3.)
|
sample no. |
percent recovered |
|
|
|
1 2 3 4 5 6 |
89.3 86.1 107.3 103.6
96.4 100.5 |
=
SD = Precision
= = |
98.9 6.3 (1.96)(±6.3) ±12.3 |
|
4.4. Instrument response to the analyte
The instrument response to the analytes over the range of 0.5 to 2
times the target concentrations was determined from multiple
injections of analytical standards. The response is linear for the
three analytes with slopes (in area counts per micrograms of analyte
per sample) of 2120, 1663, and 2805 for m-, o-,
and p-phenylenediamine respectively. The
instrument response is shown graphically in Figure
4.4.
Table 4.4.1. Instrument Response to
m-Phenylenediamine
|
× target
concn µg/sample mg/m3 |
0.5× 4.918 0.0492 |
1× 9.837 0.0984 |
2× 19.67 0.197 |
|
area counts |
10260 10096 10037 10143 10146 9992 |
20563 20615 20471 20608 20362 20615 |
41330 41218 41425 41385 41399 41581 |
|
|
10112 |
20539 |
41390 |
|
Table 4.4.2. Instrument Response to
o-Phenylenediamine
|
× target
concn µg/sample mg/m3 |
0.5× 5.203 0.0520 |
1× 10.41 0.104 |
2× 20.81 0.208 |
|
area counts |
7838 7741 7832 7678 7686 7634 |
16545 16644 16472 16438 16634 16121 |
33874 33549 33612 33810 33763 33637 |
|
|
7735 |
16476 |
33708 |
|
Table 4.4.3. Instrument Response to
p-Phenylenediamine
|
× target
concn µg/sample mg/m3 |
0.5× 4.879 0.0488 |
1× 9.757 0.0976 |
2× 19.51 0.195 |
|
area counts |
13088 13079 13182 12781 13212 12901 |
26947 26891 26703 26439 26538 26866 |
54204 53772 54331 53866 53945 54348 |
|
|
13040 |
26731 |
54078 |
|
4.5. Storage test
Storage samples were generated by spiking sulfuric acid-treated
glass fiber filters with amounts of analyte equal to the target
concentrations (9.837, 10.41, and 9.757 µg of m,
o-, and p-phenylenediamine
respectively). The filters were then assembled in
three-piece cassettes with back filters.
Thirty-six samples were prepared. One hundred liters of
air at approximately 80% relative humidity and 21°C were then drawn
through each sampling device. Six samples were analyzed immediately,
fifteen were stored in a refrigerator at 0°C, and fifteen were stored
in a closed drawer at approximately 21°C. Six samples, three from
refrigerated and three from ambient storage, were analyzed at
intervals over a period of fifteen days. The results are shown
graphically in Figures 4.5.1.1.,
4.5.1.2.,
4.5.2.1.,
4.5.2.2.,
4.5.3.1.,
and 4.5.3.2.
The recoveries of m-, o-, and
p-phenylenediamine from samples stored at ambient
temperature remained above 98%, 84%, and 98% respectively.
Table 4.5.1. Storage Test for
m-Phenylenediamine
|
days of |
% recovery |
storage |
refrigerated |
|
ambient |
|
0 0 3 6 9 12 15
|
100.5 95.5 97.7 99.2 98.0 98.5 93.4 |
98.8 96.6 97.1 97.4 98.7 100.1 96.7 |
98.7 98.0 98.8 98.4 99.4 99.8 95.3 |
|
100.5 95.5 96.7 100.2 99.9 100.9 99.8 |
98.8 96.6 103.4 98.6 99.9 100.1 98.9 |
98.7 98.0 97.2 99.9 100.1 100.3 100.5 |
|
Table 4.5.2. Storage Test for
o-Phenylenediamine
|
days of |
% recovery |
storage |
refrigerated |
|
ambient |
|
0 0 3 6 9 12 15
|
94.8 92.3 95.3 88.1 97.0 90.7 92.0 |
96.8 92.3 92.7 90.6 92.0 96.2 94.6 |
102.1 95.6 100.6 90.7 95.6 89.3 91.5 |
|
94.8 92.3 103.0 79.7 89.7 87.6 92.6 |
96.8 92.3 104.1 86.2 89.2 82.6 85.7 |
102.1 95.6 92.9 85.0 91.0 75.9 89.5 |
|
Table 4.5.3. Storage Test for
p-Phenylenediamine
|
days of |
% recovery |
storage |
refrigerated |
|
ambient |
|
0 0 3 6 9 12 15
|
99.3 95.4 95.5 97.1 98.0 95.6
94.5 |
101.1 96.1 92.2 96.4 97.0 100.5 98.0 |
100.4 100.5 94.4 97.9 98.9 100.0 94.2 |
|
99.3 95.4 93.1 91.8 97.3 100.1 98.6 |
101.1 96.1 99.5 98.8 97.6 101.3 98.7 |
100.4 100.5 101.2 97.7 95.9 98.3 101.0 |
|
4.6. Precision (analytical method only)
The precision of the analytical method for each analyte is the
pooled coefficient of variation determined from replicate injections
of standards. The precisions are based on the data from Section
4.4.
Table 4.6.1. Precision of the Analytical Method for
m-Phenylenediamine
|
× target
concn µg/sample mg/m3 |
0.5× 4.918 0.0492 |
1× 9.837 0.0984 |
2× 19.67 0.197 |
|
SD (area counts) CV = 0.0063 |
94.1 0.0093 |
102.8 0.0050 |
119.2 0.0029 |
|
Table 4.6.2. Precision of the Analytical Method for
o-Phenylenediamine
|
× target
concn µg/sample mg/m3 |
0.5× 5.203 0.0520 |
1× 10.41 0.104 |
2× 20.81 0.208 |
|
SD (area counts) CV = 0.0095 |
84.7 0.0110 |
192.6 0.0117 |
126.9 0.0038 |
|
Table 4.6.3. Precision of the Analytical Method for
p-Phenylenediamine
|
× target
concn µg/sample mg/m3 |
0.5× 4.879 0.0488 |
1× 9.757 0.0976 |
2× 19.51 0.195 |
|
SD (area counts) CV = 0.0090 |
167.3 0.0128 |
206.8 0.0077 |
248.6 0.0046 |
|
4.7. Precision (overall procedure)
The precision of the overall procedure is determined from the
storage data. The determination of the standard error of estimate
(SEE) for a regression line plotted through the graphed storage data
allows the inclusion of storage time as one of the factors affecting
overall precision. The SEE is similar to the standard deviation,
except it is a measure of dispersion of data about a regression line
instead of about a mean. It is deterinined with the following
equation:
where |
n = k
= k = |
total no. of data points 2 for linear
regression 3 for quadratic regression |
Yobs
= |
observed % recovery at a given time |
Yest
= |
estimated % recovery from the regression line at the same
given time |
An additional 5% for pump error is added to the SEE by the addition
of variances. The precision at the 95% confidence level is obtained by
multiplying the SEE (with pump error included) by 1.96 (the
z-statistic from the standard normal distribution at the
95% confidence level). The 95% confidence intervals are drawn about
their respective regression lines in the storage graphs. The standard
errors of estimate are 5.3%, 7.9%, and 5.7% and the precisions of the
overall procedure (95% confidence intervals) are ±10.3%, ±15.4%, and
±11.1% for m, o-, and p-phenylenediamine
respectively. These values were obtained from Figures 4.5.1.2.,
4.5.2.2.,
and 4.5.3.2.
for the ambient storage samples.
4.8. Reproducibility
Six samples were prepared by injecting microliter quantities of
standards onto acid-treated filters. The samples were
stored at approximately 0°C for 48 days. The samples were analyzed by
a chemist unassociated with this evaluation. No individual sample
result deviated from its theoretical value by more than the
corresponding precision of the overall procedure. The precisions of
the overall procedure are ±10.3%, ±15.4%, and ±11.1% for m-,
o-, and p-phenylenediamine
respectively.
Table 4.8.1. Reproducibility for
m-Phenylenediamine
|
sample no. |
µg found |
µg expected |
% found |
% deviation |
|
1 2 3 4 5 6 |
17.70 9.692 4.661 9.224 18.99
5.130 |
19.67 9.837 4.918 9.837 19.67
4.918 |
90.0 98.5 94.8 93.8 96.5 104.3 |
-10.0 -1.5 -5.2 -6.2 -3.5 +4.3 |
|
Table 4.8.2. Reproducibility for
o-Phenylenediamine
|
sample no. |
µg found |
µg expected |
% found |
% deviation |
|
1 2 3 4 5 6 |
18.34 9.768 4.605 9.502 19.93
4.703 |
20.81 10.41
5.203 10.41 20.81 5.203 |
88.1 93.8 88.5 91.3 95.8 90.4 |
-11.9 -6.2 -11.5 -8.7 -4.2 -9.6 |
|
Table 4.8.3. Reproducibility for
p-Phenylenediamine
|
sample no. |
µg found |
µg expected |
% found |
% deviation |
|
1 2 3 4 5 6 |
17.63 9.523 4.601 9.256 18.83
4.990 |
19.51 9.757 4.879 9.757 19.51
4.879 |
90.4 97.6 94.3 94.9 96.5 102.3 |
-9.6 -2.4 -5.7 -5.1 -3.5 +2.3 |
|
4.9. Extraction efficiency
Six sample filters for each amine were spiked with the target
concentration amounts by liquid injection (9.837, 10.41, and 9.757 µg
of m-, o-, and
p-phenylenediamine respectively). These samples were
analyzed to determine the extraction efficiencies. To determine the
stability of extracted samples, these same samples were allowed to
remain at room temperature for 24 h after extraction and were
reanalyzed using fresh standards.
Table 4.9.1. Extraction Efficiency for
m-Phenylenediamine
|
sample no. |
% extracted |
% extracted (reanalyzed after 24 h) |
|
1 2 3 4 5 6
|
99.6
99.2 100.3 101.9 100.7 103.2 100.8 |
98.1 99.8 100.1 98.4 98.0 100.9
99.2 |
|
Table 4.9.2. Extraction Efficiency for
o-Phenylenediamine
|
sample no. |
% extracted |
% extracted (reanalyzed after 24 h) |
|
1 2 3 4 5 6
|
96.9 98.0 95.9 96.0 96.4 102.5
97.6 |
88.8 95.5 93.7 92.7 94.1 96.2 93.5 |
|
Table 4.9.3. Extraction Efficiency for
p-Phenylenediamine
|
sample no. |
% extracted |
% extracted (reanalyzed after 24 h) |
|
1 2 3 4 5 6
|
99.9 100.2
99.3 101.7 101.6 103.2 101.0 |
97.0 98.3 99.4 97.7 98.7 99.6 98.4 |
|
4.10. Chromatogram
A chromatogram of an analytical standard is shown in Figure
4.10. The chromatogram is from a 5-µL injection of a
standard approximately equal to the target concentration for each
analyte (9.837, 10.41, and 9.757 µg of m-, o-,
and p-phenylenediamine per sample respectively)
for a 100-L sample.
Figure 2.1.2. Unassembled sampling
device for phenyldenediamines.
Figure 4.1. Detection limit chromatogram. Key: 1 =
p-phenylenediamine, 2 = m-phenylenediamine, 3
= o-phenylenediamine.
Figure 4.4. Instrument
response.
Figure 4.5.1.1.
Refrigerated m-phenylenediamine storage
samples.
Figure 4.5.1.2. Ambient
m-phenylenediamine storage samples.
Figure 4.5.2.1.
Refrigerated o-phenylenediamine storage
samples.
Figure 4.5.2.2. Ambient
o-phenylenediamine storage samples.
Figure 4.5.3.1.
Refrigerated p-phenylenediamine storage
samples.
Figure 4.5.3.2. Ambient
p-phenylenediamine storage samples.
Figure 4.10. Chromatogram
of a standard at the target concentrations. Key: 1 =
p-phenylenediamine, 2 = m-phenylenediamine, 3 =
o-phenylenediamine.
5. References
5.1. Lugg, G.A. Anal. Chem., 1963,
35, 899-904.
5.2. Levin, V.; Nippoldt, B.W.; Rebertus, R.L.
Anal. Chem., 1967, 39, 581-584.
5.3. Simonov, V.A.; Bartenev, V.D.; Mikhailova,
I.A. Hyg. and Sanit., 1972, 36,
237-239; "NIOSHTIC Database", on-line
database provided and updated by the National Institute of
Occupational Safety and Health (NIOSH).
5.4. Burg, W.R.; Winner, B.E.; Elia, V.J. Am.
Ind. Hyg. Assoc. J., 1980, 41, 557-562.
5.5. Dmitriev, M.T.; Semyanistye, V.D.;
Shelyapina, E.N.; Piskunov, B.G. Gig. Sanit., 1988,
11, 48-50; Chem. Abstr. 1989,
110, 28410x.
5.6. Elia, V.J.; Powers, K.M.; Burg, W.R. Am.
Ind. Hyg. Assoc. J., 1982, 43, 196-200.
5.7. "OSHA Analytical Methods Manual", Second
Edition, U.S. Department of Labor, Occupational Safety and Health
Administration; OSHA Analytical Laboratory: Salt Lake City, UT, 1990;
Method 57; American Conference of Governmental Industrial Hygienists
(ACGIH): Cincinnati, OH, Publication No. 4542.
5.8. ibid. Method 65.
5.9. ibid. Method 71.
5.10. ibid. Method 73.
5.11. ibid. Method 78.
5.12. Jasmin, G. Can. Biol., 1961,
20, 37.
5.13. "Hazardous Substances Database", on-line
database from U.S. Department of Health and Human Services, National
Library of Medicine, Bethesda, MD.
5.14. Badavari, S., Ed. "Merck Index", 11th ed.;
Merck and Co.: Rahway, NJ, 1989.
|