METHYLENE CHLORIDE
Method no.: |
80 |
|
Matrix: |
Air |
|
Target concentrations: |
500 ppm (1736 mg/m3) OSHA PEL 10
ppm (34.7 mg/m3) |
|
Procedure: |
Samples are collected by drawing air through glass sampling
tubes containing Carbosieve S-III (carbon based
molecular sieve) adsorbent. Samples are desorbed with a mixture of
99:1 (v/v) carbon disulfide
(CS2)/dimethyl-formamide (DMF) in the
presence of anhydrous sodium sulfate and are analyzed by GC using a
flame ionization detector. |
|
Recommended air volume and sampling rate: |
3 L at 0.05 L/min |
|
Reliable quantitation limit: |
0.201 ppm (0.697 mg/m3) |
|
Standard errors of estimate at the target concentration:
(Section
4.7.) |
5.8% (10 ppm) 6.5% (500 ppm) |
|
Status of method: |
Evaluated method. This method has been subjected to the
established evaluation procedures of the Organic Methods Evaluation
Branch. |
|
Date: February 1990 |
Chemist: Donald
Burright |
Organic Methods Evaluation Branch OSHA Analytical
Laboratory Salt Lake City, Utah
1. General Discussion
1.1. Background
1.1.1. History
This procedure allows the use of a single standard size (6-mm
o.d. × 4-mm i.d. × 7 cm) adsorbent tube for the
collection of methylene chloride. Because methylene chloride readily
migrates on charcoal, earlier methods required the use of two
standard size charcoal tubes connected in series, which could be
isolated from each other after sampling, (Ref.
5.1.) or the use of a custom designed jumbo size adsorbent tube
containing an inconveniently large amount of charcoal (Ref.
5.2.). These two procedures overcome the effects of migration by
either isolating the backup section from the sampling section so
that it would not be contaminated by the migrating methylene
chloride or by using such a large quantity of charcoal that
migration is slowed to an acceptable rate. The benefit of a single
standard size adsorbent tube is made possible by the use of an
adsorbent material, Carbosieve S-III, which
demonstrates negligible methylene chloride migration. Carbosieve
S-III, available through Supelco, Inc., is a spherical
carbon molecular sieve and has a surface area of about 800
m2/g, about the same as charcoal.
This evaluation was performed at two target concentrations
because the current OSHA PEL for methylene chloride is 500 ppm
(TC-500) and OSHA is now in the process of 6(b)
rule-making to reduce the occupational exposure to methylene
chloride. For the purpose of this evaluation, 10 ppm was selected as
the lower target concentration (TC-10). A goal of this
evaluation was to develop a sampling procedure that will accommodate
both the current PEL and a new lower limit if one is promulgated.
The current ACGIH TLV is 50 ppm and the current NIOSH REL states
that exposure to methylene chloride be controlled to the lowest
feasible level. (Refs. 5.3.-5.5.)
1.1.2. Toxic effects (This section is for information only and
should not be taken as the basis of OSHA policy.)
Methylene chloride can affect the body if it is inhaled,
swallowed or comes in contact with the eyes or skin. Methylene
chloride is an anesthetic and inhaling the vapor may cause mental
confusion, light-headedness, nausea, vomiting, and
headache. Continued exposure may cause increased
light-headedness, staggering, unconsciousness, and
death. High vapor concentrations may cause irritation of the eyes
and respiratory tract. Exposure to the liquid may cause irritation
to the skin and eyes. Extended exposure to the liquid by the skin
may cause skin burns. (Ref.
5.6.)
Carbon monoxide is a metabolite of methylene chloride in animals
and humans. NIOSH has recommended that if the carbon monoxide
concentration is above 9 ppm, either the concentration of methylene
chloride or carbon monoxide should be reduced. (Ref.
5.5.)
On the basis of carcinogenic and tumorigenic responses in rats
and mice, NIOSH recommends that methylene chloride be regarded as a
potential occupational carcinogen. (Ref.
5.3.)
1.1.3. Workplace exposure
Methylene chloride is used mainly as a solvent in paint removers,
aerosol mixtures, chemical reaction processes and as a blowing agent
for flexible foams. Photography, metal-work,
refrigeration, chemical processing, pharmaceutical, flexible foam,
and food industries use methylene chloride. (Ref.
5.8.) Methylene chloride has been used to extract
heat-sensitive naturally occurring substances, such as
cocoa, edible fats, spices, and beer hops. It has also been used in
the decaffeination of coffee. (Ref.
5.7.)
Approximately 230 million kilograms of methylene chloride were
produced in the United States in 1988. (Ref.
5.9.) In 1986, NIOSH estimated that 1 million workers were
potentially exposed to methylene chloride. (Ref.
5.5.)
1.1.4. Physical properties and other descriptive information (Ref.
5.10., unless otherwise stated)
CAS no.: |
75-09-2 |
molecular weight: |
84.93 |
molecular formula: |
CH2Cl2 |
melting point: |
-97°C |
boiling point: |
39.8°C |
vapor pressure: |
53.3 kPa (400 mmHg) at 24.1°C |
vapor density: |
2.93 (air=1) |
specific gravity: |
1.3266 (water=1) |
explosive limit: |
12% (lower) 19% (upper) (Ref.
5.11.) |
self-ignition temperature: |
640°C (Ref.
5.11.) |
odor: |
penetrating ether-like |
odor threshold: |
205-307 ppm |
appearance: |
colorless liquid |
solubility: |
1-2 g/100 mL of water; soluble in most common organic
liquids |
synonyms: |
dichloromethane; Aerothene MM; methylene dichloride;
Narkotil; R30; Solaesthin; Solmethine; DCM; methane
dichloride; methylene bichloride; NCI-C50102; UN
1593 (Ref.
5.11.) |
The analyte air concentrations throughout this method are based on
the recommended sampling and analytical parameters. Air concentrations
listed in ppm are referenced to 25°C and 101.3 kPa (760 mmHg).
1.2. Limit defining parameters
1.2.1. Detection limit of the analytical procedure
The detection limit of the analytical procedure is 0.0746 ng per
injection (1.0-µL injection with a 28:1 split). This is
the amount of analyte that will give a peak whose height is
approximately 5 times the height of the baseline noise. (Section
4.1.)
1.2.2. Detection limit of the overall procedure
The detection limit of the overall procedure is 2.09 µg per
sample (0.201 ppm or 0.697 mg/m3). This is
the amount of analyte spiked on the sampling device that allows
recovery of an amount of analyte equivalent to the detection limit
of the analytical procedure. (Section
4.2.)
1.2.3. Reliable quantitation limit
The reliable quantitation limit is 2.09 µg per sample (0.201 ppm
or 0.697 mg/m3). This is the smallest
amount of analyte spiked on the sampling device that can be
quantitated within the requirements of a recovery of at least 75%
and a precision (±1.96 SD) of ±25% or better. (Section
4.3.)
The reliable quantitation limit and detection limits reported in
the method are based upon optimization of the instrument for the
smallest possible amount of analyte. When the target concentration
of analyte is exceptionally higher than these limits, they may not
be attainable at the routine operating parameters.
1.2.4. Instrument response to the analyte
The instrument response over the concentration range of 0.5 to 2
times the TC-10 and TC-500 target
concentrations is linear. (Section
4.4.)
1.2.5. Recovery
The recovery of methylene chloride from samples used in 17-day
storage tests remained above 99.6% and 97.3% at the
TC-10 and TC-500 levels, respectively,
when the samples were stored at about 22°C. The recovery of analyte
from the collection medium during storage must be 75% or greater.
(Section
4.5., regression lines of Figures 4.5.1.1.
and 4.5.2.1.)
1.2.6. Precision (analytical procedure only)
The pooled coefficients of variation obtained from replicate
determinations of analytical standards at 0.5, 1 and 2 times the
target concentrations are 0.0330 and 0.0267 at the
TC-10 and TC-500 levels respectively. (Section
4.6.)
1.2.7. Precision (overall procedure)
The precisions at the 95% confidence level for the 17-day ambient
temperature storage tests are ±11.3% and ±12.7% at the
TC-10 and TC-500 levels respectively. (Section
4.7.) These each include an additional ±5% for pump error. The
overall procedure must provide results at the target concentration
that are ±25% or better at the 95% confidence level.
1.2.8. Reproducibility
Six samples (TC-10) collected from a controlled test atmosphere
and a draft copy of this procedure were given to a chemist
unassociated with this evaluation. The samples were analyzed after
11 days of refrigerated storage. No individual sample result
deviated from its theoretical value by more than the precision
reported in Section
1.2.7. (Section
4.8.)
1.3. Advantages
Because methylene chloride does not readily migrate on Carbosieve
S-III, a single standard size Carbosieve S-III sampling
tube may be used. This is not permissible with activated charcoal due
to the migration of methylene chloride from the front section to the
back section.
1.4. Disadvantages
1.4.1. The fine mesh size of Carbosieve S-III (60/80) results in
a greater pressure drop across the sample tube than occurs with the
conventional coconut shell charcoal sampling tube. This results in
the need for the 0.05 L/min sampling rate.
1.4.2. The recommended sample size is 3 L as opposed to the 10 L
sample size of previous methods.
2. Sampling Procedure
2.1. Apparatus
2.1.1. Samples are collected using a personal sampling pump
calibrated to within ±5% of the recommended flow rate with the
sampling device attached.
2.1.2. Samples are collected with 4-mm i.d. × 6-mm o.d. × 70 mm
glass sampling tubes packed with two sections of 60/80 mesh
Carbosieve S-III. The front section contains 130 mg and
the back section contains 65 mg of adsorbent. The sections are held
in place with glass wool plugs. For this evaluation, Supelco's
ORBO-91 Carbosieve S-III tubes (catalog
no. 2-0360) were used.
2.2. Reagents
No sampling reagents are required.
2.3. Technique
2.3.1. Immediately before sampling, break off the ends of the
Carbosieve S-III sampler. All samplers should be from
the same lot.
2.3.2. Attach the sampler to the sampling pump with plastic
tubing such that the large front section of the sampler is exposed
directly to the atmosphere. Do not place any tubing in front of the
sampler.
2.3.3. Attach the sampler vertically in the worker's breathing
zone in such a manner that it does not impede work performance or
safety.
2.3.4. After sampling for the appropriate time, remove the
sampler and seal the tube with plastic end caps. Wrap each sample
end-to-end with a Form OSHA-21 seal.
2.3.5. Submit at least one blank sampler with each set of
samples. Handle the blank sampler in the same manner as the other
samples except draw no air through it.
2.3.6. Ship any bulk samples in a container separate from the air
samples.
2.4. Sampler capacity
The sampling capacity of the front section of a Carbosieve S-III
sampling tube was determined by sampling a test atmosphere containing
20 ppm (69.5 mg/m3, 80% relative humidity)
methylene chloride at ambient temperature. The sampling rate was
0.0503 L/min. The 5% breakthrough air volume was 7.6 L. (Section
4.9.)
The 5% breakthrough air volume was 3.34 L when sampling a test
atmosphere containing 1018 ppm (3535 mg/m3,
83% relative humidity) of methylene chloride at 0.053 L/min. (Section
4.9.)
2.5. Desorption efficiency
2.5.1. The average desorption efficiencies for methylene
chloride from Carbosieve S-III adsorbent were
quantitative over the range of 0.5 to 2 times the TC-10
and TC-500 target concentrations. (Section
4.10.)
2.5.2. Desorbed samples remain relativity stable for at least 24
h. (Section
4.10.)
2.6. Recommended air volume and sampling rate
2.6.1. For time-weighted average samples collect 3 L at 0.05
L/min.
2.6.2. For ceiling or peak samples (Table
Z-2 of Title 29 CFR Part 1910.1000) collect the sample for at
least 5 min at 0.05 L/min.
2.6.3. When ceiling or peak samples are required, the reliable
quantitation limit becomes larger. For example, the reliable
quantitation limit is 2.41 ppm (8.36
mg/m3) for methylene chloride when 0.25 L
is collected.
2.7. Interferences (sampling)
2.7.1. It is not known if any compounds will severely interfere
with the collection of methylene chloride on Carbosieve
S-III. In general, the presence of other contaminant
vapors in the air will reduce the capacity of Carbosieve
S-III to collect methylene chloride.
2.7.2. Suspected interferences should be reported to the
laboratory with submitted samples.
2.8. Safety precautions (sampling)
2.8.1. The sampling equipment should be attached to the worker
in such a manner that it will not interfere with work performance or
safety.
2.8.2. All safety practices that apply to the work area being
sampled should be followed.
3. Analytical Procedure
3.1. Apparatus
3.1.1. A GC equipped with a flame ionization detector (FID). A
Hewlett-Packard 5890 Gas Chromatograph equipped with a
7673A Autosampler, split/splitless injection port and an FID was
used in this evaluation.
3.1.2. A GC column capable of separating methylene chloride from
the desorbing solvent, the internal standard and any potential
interferences. A 30-m × 0.32-mm i.d.
SUPELCOWAX 10 (0.25 µm film thickness) capillary column (Supelco
Inc.) was used in this evaluation.
3.1.3. An electronic integrator or some other suitable means of
measuring detector response. A Hewlett-Packard 5895A GC
ChemStation was used in this evaluation.
3.1.4. Two-milliliter vials with PTFE-lined caps.
3.1.5. A dispenser capable of delivering 1.0 mL of desorbing
solution is used to prepare standards and samples. If a dispenser is
not available, a 1.0-mL volumetric pipet may be used.
3.2. Reagents
3.2.1. Methylene chloride. Reagent grade or better should be
used. The methylene chloride (b&j brand HIGH PURITY SOLVENT)
used in this evaluation was purchased from American Burdick &
Jackson (Muskegon, MI).
3.2.2. Carbon disulfide, CS2. Reagent
grade or better CS2 should be used. The
CS2 (REAGENT ACS) was purchased from
Fisher Scientific (Fair Lawn, NJ). In this evaluation, benzenefree
CS2 was used. The
CS2 had been passed through Molecular
Sieve 13X (45/60 mesh) to remove the benzene contamination. Fifty
grams of molecular sieve should remove the benzene from 1 L of
carbon disulfide.
3.2.3. Dimethylformamide, DMF. Reagent grade or better should be
used. The DMF (b&j brand HIGH PURITY SOLVENT) used in this
evaluation was purchased from American Burdick & Jackson
(Muskegon, MI).
3.2.4. Sodium sulfate, anhydrous. Sodium sulfate is used as a
drying agent. The sodium sulfate (AR grade) used in this evaluation
was purchased from Mallinckrodt (Paris, KY).
3.2.5. Desorbing solution. This consists of a solution of 99:1
(v/v) benzene-free CS2/DMF.
An internal standard such as ethyl benzene can be used.
3.2.6. Ethyl benzene. This was used as the internal standard in
the desorbing solution. The solution is prepared by adding 250 µL of
ethyl benzene to 1 L of desorbing solution. The ethyl benzene
(reagent grade) used in this evaluation was purchased from Eastman
Kodak (Rochester, NY).
3.3. Standard preparation
3.3.1. Prepare concentrated stock standards by diluting the
methylene chloride with DMF. Prepare working analytical standards by
injecting microliter amounts of concentrated stock standards into
2-mL vials containing 1 mL of desorbing solution
delivered from the same dispenser used to desorb samples. For
example, to prepare a TC-10 standard, inject 10 µL of a
stock solution containing 10.42 mg/mL methylene chloride in DMF into
1 mL of desorbing solution.
3.3.2. Prepare at least three standards at different
concentrations to generate a calibration curve. Ensure that the
amount of methylene chloride found in the samples is within the
range of the standards. Prepare additional standards if necessary.
3.4. Sample preparation
3.4.1. Remove the plastic caps from the sample tube and
carefully transfer each section of the adsorbent to separate vials.
Discard the glass tube and glass wool plugs.
3.4.2. Add approximately 150 mg of anhydrous sodium sulfate to
each sample.
3.4.3. Add 1.0 mL of desorbing solution to each vial and
immediately seal the vials with PTFE-lined caps.
3.4.4. Shake the vials vigorously by hand several times during
the next 30 min.
3.5. Analysis
3.5.1. Analytical conditions
GC conditions
|
|
temperatures: |
40°C (column) 200°C (injector) 220°C
(detector) |
temp program: |
hold initial temp 1.0 min, increase temp at 5°C/min to
65°C, then increase temp at 25°C/min to 190°C. |
column gas flow: |
1.20 mL/min (hydrogen) |
septum purge: |
1.5 mL/min (hydrogen) |
injection size: |
1.0 µL (with an appropriate split) |
column: |
30 m × 0.32-mm i.d. capillary SUPELCOWAX 10 (0.25 µm
film thickness) column |
retention times: |
4.3 min (methylene chloride) 7.2 min (ethyl
benzene) |
|
FID conditions
|
hydrogen flow: |
34 mL/min |
air flow: |
450 mL/min |
nitrogen makeup flow: |
33 mL/min |
chromatogram: |
Figure
3.5.1. |
3.5.2. Measure detector response using a suitable method such as
electronic integration.
3.5.3. An internal standard (ISTD) calibration method is used. A
calibration curve can be constructed by plotting micrograms of
methylene chloride per sample versus ISTD-corrected
response of standard injections. Bracket the samples with freshly
prepared analytical standards over a range of concentrations.
3.6. Interferences (analytical)
3.6.1. Any compound that responds on an FID and has a similar
retention time as the analyte or internal standard is a potential
interference. Generally, chromatographic conditions can be altered
to separate an interference from the analyte.
3.6.2. Retention time on a single column is not considered proof
of chemical identity. Analysis by an alternate GC column or
confirmation by mass spectrometry are additional means of
identification.
3.7. Calculations
The analyte concentration for samples is obtained from the
appropriate calibration curve in terms of micrograms per sample,
uncorrected for desorption efficiency. The air concentration is
calculated using the following formulae. The back (65-mg)
section is analyzed to determine if there was any breakthrough from
the front (130-mg) section during sampling. If any
analyte is found on the back section, it is added to the amount on the
front section. This total amount is then corrected by subtracting the
total amount (if any) found on the blank. If a significant amount of
analyte is found on the back section (e.g., greater than 25% of the
amount found on the front section), this should be reported with
sample results.
mg/m3 = |
(micrograms of analyte per sample)
(liters of air sampled)(desorption efficiency)
|
where desorption efficiency = 1.05 for TC-10 or 0.989
for TC-500
ppm = |
(mg/m3)(24.46)
(molecular weight of analyte) |
where 24.46 = molar volume (liters) at 101.3 pKa (760
mmHg) and 25°C molecular
weight = 84.93
3.8. Safety precautions (analytical)
3.8.1. Restrict the use of all chemicals to a fume hood if
possible.
3.8.2. Avoid skin contact and inhalation of all chemicals.
3.8.3. Wear safety glasses and a lab coat at all times while in
the laboratory areas.
4. Backup Data
4.1. Detection limit of the analytical procedure
The detection limit of the analytical procedure is 0.0746 ng per
injection, based on a 1.0-µL injection (with a 28:1
split) of a 2.09 µg/mL standard. This amount produced a methylene
chloride peak whose height is about 5 times the height of the baseline
noise in the chromatogram. A chromatogram of the detection limit of
the analytical procedure is shown in Figure
4.1.
4.2. Detection limit of the overall procedure
The detection limit of the overall procedure is 2.09 µg per sample
(0.201 ppm or 0.697 mg/m3). The injection
size listed in the analytical procedure (1.0 µL) was used in the
determination of the detection limit of the overall procedure. Eight
vials containing 130 mg of Carbosieve S-III resin were
spiked with 10 µL of a solution containing 0.209 µg/µL. The samples
were desorbed about 24 h after being spiked.
Table 4.2. Detection Limit of the Overall Procedure
for Methylene Chloride
|
sample no. |
µg spiked |
µg recovered |
|
1 2 3 4 5 6 7 8 |
2.09 2.09 2.09 2.09 2.09 2.09 2.09 2.09 |
2.40 2.28 2.29 2.31 2.33 2.31 2.25 2.13 |
|
4.3. Reliable quantitation limit data
The reliable quantitation limit is 2.09 µg per sample (0.201 ppm or
0.697 mg/m3). The injection size listed in
the analytical procedure (1.0 µL) was used in the determination of the
reliable quantitation limit. Eight vials containing 130 mg of
Carbosieve S-III resin were liquid-spiked
with 10 µL of a solution containing 0.209 µg/µL or 2.09 µg of
methylene chloride. Because the recovery of methylene chloride from
the spiked samples was greater than 75% and had a precision of ±25 or
better, the detection limit of the overall procedure and reliable
quantitation limit are the same.
Table 4.3. Reliable Quantitation Limit (Based on
samples and data of Table 4.2.)
|
percent recovered |
statistics |
|
114.8 109.1 109.6 110.5 111.5 110.5 107.7 101.9 |
= SD
= Precision = = |
109.5 3.7 (1.96)(±3.7) ±7.3 |
|
4.4. Instrument response to methylene chloride
The instrument response to methylene chloride over the range of 0.5
to 2 times the TC-10 and TC-500 target
concentrations is linear with a slope of 5132 and 1474 respectively
(in ISTD-corrected area counts per microgram per
milliliter). The precision of the response to methylene chloride was
determined by multiple injections of methylene chloride standards. The
data below is presented graphically in Figure 4.4.1.
and 4.4.2.
Table 4.4.1. Instrument Response to Methylene
Chloride Injection Split = 8:1
|
× TC-10 µg/mL ppm/sample |
0.5× 52.1 5.00 |
1× 104.2 10.00 |
2× 208.4 20.01 |
|
area counts |
261 278 277 276 276 266 264 272 |
550 520 581 532 542 539 535 530 |
1073 1105 1055 1063 1033 1030 1159 1074 |
|
|
271.3 |
541.1 |
1074.0 |
|
Table 4.4.2. Instrument Response to Methylene
Chloride Injection Split = 28:1
|
× TC-10 mg/mL ppm/sample |
0.5× 2.65 254.5 |
1× 5.30 509 |
2× 10.6 1018 |
|
area counts |
3935 4005 3977 3944 4191 4188 4116 4070 |
8105 7750 7685 7801 8191 8151 8119 8190 |
15786 15637 15354 15846 15392 16088 15395 16638 |
|
|
4053.3 |
7999.0 |
15767.0 |
|
4.5. Storage data
4.5.1. Storage test at 10 ppm
Thirty-six storage samples were collected by sampling a
dynamically generated atmosphere containing 57.4
mg/m3 or 16.5 ppm of methylene chloride
and 80% relative humidity for 30 min at 0.05 L/min.
One-half of the tubes was stored in a freezer
(-20°C) and the other half was stored in a closed
drawer at ambient temperature (about 22°C). At 3-4 day
intervals, three samples were selected from each of the two storage
sets and analyzed. The results are listed below and shown
graphically in Figures 4.5.1.1.
and 4.5.1.2.
Table 4.5.1. Storage Test at TC-10
|
storage time (days) |
% recovery (ambient) |
|
% recovery (refrigerated) |
|
0 4
7 11 14 17 |
101.1 99.4 98.5 101.2 106.3 104.9 109.4 |
101.1 100.1 98.4 99.6 103.9 96.9 99.1 |
98.9 100.0 99.5 99.5 101.2 101.2 98.7 |
|
101.1 99.4 100.8 100.6 98.7 102.3 94.9 |
101.1 100.1 103.3 94.8 108.6 100.4 95.1 |
98.9 100.0 95.9 99.6 102.4 97.7 104.9 |
|
4.5.2. Storage test at 500 ppm
Thirty-six storage samples were collected by sampling a
dynamically generated atmosphere containing 3516
mg/m3 or 1013 ppm of methylene chloride
and 86% relative humidity for 30 min at 0.05 L/min.
One-half of the tubes was stored in a freezer
(-20°C) and the other half was stored in a closed
drawer at ambient temperature (about 22°C). At 3-5 day
intervals, three samples were selected from each of the two storage
sets and analyzed. The results are listed below and shown
graphically in Figures 4.5.2.1.
and 4.5.2.2.
Table 4.5.2. Storage Test at TC-500
|
storage time (days) |
% recovery (ambient) |
|
% recovery (refrigerated) |
|
0 3
6 11 14 17 |
99.8 96.8 94.7 93.6 93.4 102.9 100.3 |
98.3 97.9 103.6 96.3 92.3 94.0 102.7 |
104.4 102.9 90.4 97.4 94.7 96.5 98.9 |
|
99.8 96.8 95.7 99.6 89.3 87.5 94.3 |
98.3 97.9 96.0 92.2 97.2 95.5 92.9 |
104.4 102.9 97.9 89.4 94.1 87.8 100.3 |
|
4.6. Precision (analytical method)
The precision of the analytical procedure is defined as the pooled
coefficient of variation determined from replicate injections of
methylene chloride standards at 0.5, 1 and 2 times the
TC-10 and TC-500 target concentrations.
Based on the data of Tables 4.4.1.
and 4.4.2.,
the coefficients of variation (CV) for the three levels and the pooled
coefficient of variation () were calculated and are listed below.
Table 4.6.1. Precision of the Analytical Method
(Based on the Data of Table
4.4.1.)
|
× TC-10 µg/mL ppm/sample |
0.5× 52.1 5.00 |
1× 104.2 10.00 |
2× 208.4 20.01 |
|
SD1 CV = 0.0330 |
6.64 0.02451 |
18.38 0.03397 |
41.89 0.03901 |
1 standard deviation is in area
counts |
Table 4.6.2. Precision of the Analytical Method
(Based on the Data of Table
4.4.2.)
|
× TC-10 mg/mL ppm/sample |
0.5× 2.65 254.5 |
1× 5.30 509 |
2× 10.6 1018 |
|
SD1 CV = 0.0267 |
103.8 0.02561 |
214.4 0.02681 |
436.7 0.02770 |
1 standard deviation is in area
counts |
4.7. Precision (overall procedure)
The precision of the overall procedure is determined from the
storage data. The determination of the standard error of estimate
(SEE) for a regression line plotted through the graphed storage data
allows the inclusion of storage time as one of the factors affecting
overall precision. The SEE is similar to the standard deviation except
it is a measure of dispersion of data about a regression line instead
of about a mean. It is determined with the following equation:
where |
n = k
= k = |
total no. of data points 2 for linear
regression 3 for quadratic regression |
|
Yobs
= |
observed % recovery at a given time |
|
Yest
= |
estimated % recovery from the regression line at the same
given time |
An additional 5% for pump error is added to the SEE by the addition
of variances. The precision at the 95% confidence level is obtained by
multiplying the SEE (with pump error included) by 1.96 (the
z-statistic from the standard normal distribution at the
95% confidence level). The 95% confidence intervals are drawn about
their respective regression lines in the storage graphs as shown in
Figure 4.5.1.1. The data for Figures 4.5.1.1.
and 4.5.2.1.
was used to determine the SEE of ±5.75% and ±6.48% at the
TC-10 and TC-500 levels respectively.
4.8. Reproducibility data
Six samples, collected from a dynamically generated atmosphere
containing methylene chloride, were given to a chemist unassociated
with this study. The samples were generated by drawing the
20-ppm test atmosphere through sampling tubes for
30-45 min at approximately 0.050 L/min. The samples were
analyzed after being stored for 11 days at 5°C. No sample result had a
deviation greater than the precision of the overall procedure, which
is ±11.3%.
Table 4.8. Reproducibility Data at TC-10
|
µg spiked |
µg recovered |
% recovered |
% deviation |
|
147.7 149.8 102.6 100.2 172.7 172.0 |
145.6 148.8 94.4
97.1 162.3 154.2 |
98.6 99.3 92.0 96.9 94.0 89.6 |
-1.4 -0.7 -8.0 -3.1 -6.0 -10.4 |
|
4.9. Sampler capacity
4.9.1. Breakthrough for TC-10
Sampler capacity was determined by sampling from a dynamically
generated atmosphere of 20 ppm (69.5
mg/m3) methylene chloride with a
Carbosieve S-III sampling tube that contained only the
front section. The tube was followed by a whole Carbosieve
S-III sampling tube. The backup tube was periodically
changed over a 4 h time. The relative humidity of the test
atmosphere was 80%. The sampling rate was 0.0503 L/min. The air
volumes listed below are the midpoints of each sampling interval.
The data is graphically shown in Figure
4.9.1.
Table 4.9.1. Breakthrough on the Carbosieve S-III
Tube
|
air vol (L) |
amt found (µg) |
downstream mg/m3 |
breakthrough (%) |
|
0.75 2.26 3.77 4.91 5.66 6.41 7.17 7.92 8.68 9.43 10.2
10.9 11.7 |
2.56 0 0 0 0 2.39 0 2.50 4.32 6.99 8.49 12.6
14.5 |
1.696 0 0 0 0 3.162 0 3.308 5.720 9.244 11.26
16.61 19.25 |
2.44 0 0 0 0 4.55 0 4.76 8.23 13.3
16.2 23.9
27.7 |
|
4.9.2. Breakthrough for TC-500
Sampler capacity was determined by sampling from a dynamically
generated atmosphere of 1018 ppm (3535
mg/m3, 83% relative humidity) methylene
chloride with a Carbosieve S-III sampling tube that
contained only the 130-mg front section. The sampling
rate was 0.053 L/min. The air passing through the sampler was
monitored with a GC equipped with a gas sampling valve. An injection
was made every three minutes. After sampling for 63 min or 3.34 L,
the concentration of methylene chloride in the air downstream of the
sampler became greater than 5% of the upstream concentration. The
data was collected (but not retained) by a computer that produced Figure
4.9.2.
4.10. Desorption efficiency and stability of
desorbed samples
4.10.1. Desorption efficiency
The desorption efficiency (DE) of methylene chloride was
determined by liquid-spiking 130-mg
portions of Carbosieve S-III adsorbent with methylene
chloride at 0.5 to 2 times the TC-10 and
TC-500 target concentrations. These samples were stored
overnight and then desorbed with desorbing solution and analyzed.
The average desorption efficiency over the studied range was 105.4%
and 98.9% at the TC-10 and TC-500 levels
respectively.
Table 4.10.1.1. Desorption Efficiency of Methylene
Chloride
|
× TC-10 µg/sample |
0.5× 52.1 |
1× 104.2 |
2× 208.4 |
|
DE, % |
101.9 106.7 108.6 100.5
97.1 104.6 105.9 108.7 |
103.0 112.9 103.2 108.9 106.2
99.7 101.0 108.0 |
108.4 105.4 109.6 106.5 105.0 106.2 107.7 102.9 |
|
|
104.3 |
105.4 |
106.4 |
|
Table 4.10.1.2. Desorption Efficiency of Methylene
Chloride
|
× TC-10 mg/sample |
0.5× 2.65 |
1× 5.31 |
2× 10.6 |
|
DE, % |
101.2 98.9 100.9 100.6
99.3 96.8 102.9 95.6 |
103.2 101.5
98.4 99.2 98.8 94.9 98.2 104.6
|
99.6 96.7 95.4 93.6 99.7 97.7 99.1 95.5 |
|
|
99.5 |
99.9 |
97.2 |
|
4.10.2. Stability of desorbed samples
The stability of desorbed samples was investigated by reanalyzing
the TC-10 samples (10 ppm) 24 h after initial analysis.
The original analysis was performed and the vials were recapped
after injection. The samples were reanalyzed with fresh standards.
The average recovery, compared to the average recovery of the
original analysis, was 100.4 or a +6.1% change.
Table 4.10.2.1. Stability of Desorbed TC-10
Samples
|
initial recovery (percent) |
recovery after 24 h (percent) |
percent change |
|
93.9 105.1 94.7 86.8 91.4 93.7 |
102.0 101.1 97.6 102.6
98.4 100.7 |
+8.1 -4.0 +2.9 +15.8 +7.0 +7.0 |
|
The stability of desorbed samples was investigated by reanalyzing
the TC-500 samples (500 ppm) 24 h after initial
analysis. The original analysis was performed and the vials were
recapped after injection. The samples were reanalyzed with fresh
standards. The average recovery, compared to the average recovery of
the original analysis, was 101.2% or a +2.7% change.
Table 4.10.2.2. Stability of Desorbed TC-500
Samples
|
initial recovery (percent) |
recovery after 24 h (percent) |
percent change |
|
95.8 100.8 97.9 99.0 98.5 98.9 |
103.6 100.3 101.5 102.4
94.8 104.4 |
+7.8 -0.5 +3.6 +3.4 -3.7 +5.5 |
|
Figure 3.5.1. Chromatogram of methylene chloride at the
TC-10 target concentration.
Figure 4.1. Chromatogram of methylene chloride at the
detection limit, 0.0746 ng per injection, injection split =
28:1.
Figure 4.4.1. Instrument response curve for methylene
chloride at 10 ppm, slope = 322 area counts per micrograms per
milliliter, injection split =
8:1.
Figure 4.4.2. Instrument response curve for methylene
chloride at 500 ppm, slopes = 1474 area counts per micrograms per
milliliter, injection split =
28:1.
Figure 4.5.1.1. Ambient storage test for methylene
chloride at 10 ppm.
Figure 4.5.1.2. Refrigerated storage test for methylene
chloride at 10 ppm.
Figure 4.5.2.1. Ambient storage test for methylene
chloride at 500 ppm.
Figure 4.5.2.2. Refrigerated storage test for methylene
chloride at 500 ppm.
Figure 4.9.1. Determination of the 5% breakthrough air
volume at 10 ppm.
Figure 4.9.2. Determination of the 5% breakthrough air
volume at 500 ppm.
5. References
5.1. "NIOSH Manual of Analytical Methods", 3rd
ed.; U.S. Department of Health and Human Services, Center for Disease
Control, NIOSH; Cincinnati, OH, 1984, Method 1005, DHHS (NIOSH) Publ.
No. 84-100.
5.2. Cummins, K.J. "OSHA Method No. 59; Methylene
Chloride", OSHA Analytical Laboratory, unpublished, Salt Lake City, UT
84165, April, 1986.
5.3. "Air Contaminants - Permissible Exposure
Limits", Code of Federal Regulations, Title 29; 1910.1000, U.S.
Department of Labor, OSHA; Washington, D.C., 1989, DOL (OSHA) Publ.
No. OSHA 3112.
5.4. "American Conference of Government
Industrial Hygienists: Threshold Limit Values and Biological Indices
for 1988-1989", p 26, Cincinnati, OH (1988).
5.5. "Current Intelligent Bulletin 46, Methylene
Chloride"; April 18, 1986, U.S Department of Health and Human
Services, Public Health Service, Center for Disease Control, NIOSH.
5.6. "Occupational Health Guidelines for
Methylene Chloride", U.S Department of Health and Human Services,
Center for Disease Control, NIOSH, September 1978.
5.7. International Agency for Research on Cancer,
"IARC Monographs on the Evaluation of the Carcinogenic Risk of
Chemicals to Humans: Some Halogenated Hydrocarbons and Pesticide
Exposures", IARC, Lyon, Switzerland, 1986, Vol. 41, pp.
43-85.
5.8. Anthony, T. in "Kirk-Othmer Encyclopedia of
Chemical Technology"; 3rd ed.; Grayson, M., Ed.; John Wiley &
Sons, New York, 1983, Vol. 5, pp. 686-693.
5.9. Chemical and Engineering News, Vol.
67, No. 25, Jun 19, 1989, p. 41.
5.10. ChemInfo Database on CCINFO CD-ROM disc
89-2, Canadian Centre for Occupational Health and Safety, Hamilton,
Ontario.
5.11. CAMEO Database, National Oceanic and
Atmospheric Administration Hazardous Materials Response Branch,
Seattle, WA.
|