Method no.: |
07 |
|
Matrix: |
Air |
|
OSHA PELs: |
Section
4 |
|
Procedure: |
Collection on charcoal,
extraction with an organic solvent, and analysis by gas
chromatography with flame ionization detector. |
|
Recommended air volume
and sample rate: |
Section
4 |
|
Status of method: |
This method has been
used extensively in the OSHA Salt Lake Technical Center. With
slight modification, this method is a generalized version of
validated NIOSH methodology. |
|
|
Date: May 1979 |
By: Organic
Methods Evaluation Branch |
Last Update: May 2000 |
By: Methods Development
Team | |
Methods Development
Team Industrial Hygiene Chemistry Division OSHA Salt Lake
Technical Center Salt Lake City, UT 84115-1802
|
1. General Discussion
1.1 Background
Background information on the analytes may be obtained
from a number of sources such as NIOSH Criteria Documents,
chemical dictionaries and industrial hygiene manuals. Solvents are
used for degreasing, for dry cleaning, and in the manufacture of
many materials ranging from paints, varnishes, shellacs, and
lacquers to rubber and synthetic resins. When not being used as
solvents, they may function as fuels or act as chemical
intermediates with or without regard to their ability to put
materials into solution. Toxic effects of the analytes vary with
many acting as irritants or causing narcosis, and some having more
hazardous effects.
1.2 Statistical parameters
1.2.1 Each analyte included in this general
procedure has a validated NIOSH method, (Ref. 5.1)
and/or a validated OSHA method. One of the NIOSH validation
requirements is that the results obtained be within ±25% of the
true values at the 95% confidence level at the air concentration
equal to the OSHA standard. Although the OSHA evaluation
procedure differs from that of NIOSH, the same validation
requirements are used.
1.2.2 Refer to the validated
NIOSH methods, (Ref. 5.1)
for detailed information on individual analytes. 1.3
Advantages
1.3.1 The sampling device is small, portable, and
involves no liquids.
1.3.2 The analysis is by a quick
instrumental method.
1.3.3 Interferences can be
eliminated by altering chromatographic conditions in most cases.
1.3.4 The method allows simultaneous analysis of two or
more analytes. 1.4 Disadvantages
1.4.1 The air volume sampled is limited by the
capacity of the charcoal tubes. Exceeding the capacity of the
charcoal tube results in loss of sample. The adsorptive capacity
is decreased by high humidity.
1.4.2 The method is
limited by the reproducibility of the pressure drop across the
tubes. The pressure drop affects the flow rate causing the air
volume to be imprecise.
1.4.3 The analyst must work with
toxic solvents.
1.4.4 When many components are present,
elimination of interferences becomes difficult.
2. Sampling
Procedure
2.1 Apparatus
2.1.1 A calibrated personal sampling pump whose flow
can be determined within ±5% at the recommended flow rate with
the sampling device attached.
2.1.2 Charcoal tubes:
Glass tubes with both ends flame sealed, 7 cm long with a 6-mm
o.d. and 4-mm i.d., containing two sections of 20/40 mesh
activated charcoal separated by a 2-mm portion of urethane foam.
The activated charcoal is prepared from coconut shells and is
fired at 600°C prior to packing. The adsorbing section contains
100 mg of charcoal, the backup section 50 mg. A3-mm portion of
urethane foam is placed between the outlet end of the tube and
the backup section. A plug of silylated glass wool is placed in
front of the absorbing section. The pressure drop across the
tube must be less than 1 in. of mercury at a flow rate of 1
L/min.
2.1.3 Certain analytes require petroleum base
charcoal instead of coconut base charcoal. This requirement is
specified in Section
4. 2.2 Reagents
None required in sampling
procedure.
2.3 Technique
2.3.1 Immediately before sampling, break the ends of
the tube to provide an opening at least one-half
the internal diameter of the tube (2 mm).
2.3.2
The smaller section of charcoal is used as a backup and should
be positioned nearest the sampling pump.
2.3.3 The
charcoal tube should be placed vertically during sampling to
minimize channeling.
2.3.4 Air being sampled should not
be passed through any hose or tubing before entering the
charcoal tube.
2.3.5 Do not exceed the recommended air
volume.
2.3.6 The charcoal tubes should be capped with
the supplied plastic caps immediately after sampling. Under no
circumstances should rubber caps be used.
2.3.7 One tube
should be handled in the same manner as the sample tube (break,
seal and transport) except that no air is sampled through this
tube. This tube should be labeled as a blank.
2.3.8
Capped charcoal tubes should be wrapped end to end with official
OSHA seals. They should be packed tightly and padded before they
are shipped to minimize tube breakage during shipping.
2.3.9 For certain analytes where migration on the
charcoal is a significant problem, it may be requested that two
charcoal tubes be used in series in order that breakthrough may
be distinguished from migration. These tubes must be separated
and individually capped and sealed before shipping.
2.4 Breakthrough
Breakthrough data is
presented on each analyte in its respective validated NIOSH method
(Ref. 5.1).
2.5 Extraction efficiency
2.5.1 The back end of a charcoal tube is opened and
the backup portion of activated charcoal is removed, leaving the
front 100-mg portion of activated charcoal intact in the tube.
The activated charcoal must be of the same lot as that in the
tubes used to collect the samples. A known amount of analyte is
injected directly into the activated charcoal with a microliter
syringe and the tube is capped.
2.5.2 Six tubes at each
of three concentration levels (0.5, 1, and 2 times the standard)
are prepared by adding an amount of analyte equivalent to that
present in a recommended air sample at the selected level. The
tubes are allowed to stand at least overnight to assure complete
adsorption of the analyte onto the charcoal. These tubes are
referred to as the samples. A parallel blank tube should be
treated in the same manner except that no analyte is added to
it. The sample and blank tubes are extracted and analyzed in
exactly the same manner as the sampling tube described in Section
3.
2.5.3 The extraction efficiency (EE) equals the average weight in
milligrams recovered from the tube divided by the weight in
milligrams added to the tube, or
where: |
EE
is extraction efficiency |
|
MR
is mass recovered |
|
MS
is mass spiked | 2.5.4 If there is a significant change in
extraction efficiency over the range of loadings studied, a
plotted curve of EE versus
mass recovered must be used to correct for adsorption losses.
2.5.5 If there is no significant change in EE over the range studied,
reconfirmations need only be carried out at one loading in the
middle of the range. 2.6
Recommended air volume and sample rate
See Section
4. for recommended air volume and sampling rate.
2.7
Interferences
2.7.1 It is important to be aware of other
components in the atmosphere which may interfere with the
collection of the analyte.
2.7.2 High relative humidity
may significantly affect the collection of some analytes.
2.8 Safety precautions
2.8.1 Care must be taken when opening the sealed
ends of charcoal tubes to avoid cuts to the hands.
2.8.2
Safety glasses should be worn when opening the sealed ends of
charcoal tubes to avoid injury to the eyes from glass
splinters. 3. Analytical procedure
3.1 Apparatus
3.1.1 Gas chromatograph equipped with flame
ionization detector.
3.1.2 Columns. A variety of columns
are suitable. Two good selections are a 60-m × 0.32 mm
DB-1 capillary column with 1m df or a
60-m × 0.32 mm DB-Wax capillary column
with 1 µm df. Similar columns from other manufactures are
acceptable.
3.1.3 A suitable method of measuring peak
areas, such as an electronic integrator or data system.
3.1.4 Two-milliliter vials with either screw-on or
crimp-on caps which contain PTFE-lined septa.
3.1.5
Microliter syringes; one-microliter for GC injections and 10-µL
for standard preparation, or other suitable sizes.
3.1.6
Pipets for dispensing extracting solvent (ES). A Glenco 1-mL
reagent dispenser is adequate and convenient.
3.1.7
Volumetric Flasks. Five-milliliter and other convenient sizes.
3.1.8 Glass tubing cutter. 3.2 Reagents
3.2.1 Chromatographic quality extracting solvent
(ES). Although carbon disulfide is commonly used as the ES,
certain analytes can be more effectively extracted with the use
of alternate solvents or solvent solutions. These alternate ESs
are listed in Chemical
Sampling Information located at http://www.osha-slc.gov/ and
are normally used when the single analyte is requested or when
the requested analytes are known to be effectively extracted
with that ES. When analysis for a number of analytes requiring
different extracting solutions is requested, the preferred ES
will usually be carbon disulfide.
3.2.2 Analyte
standard, reagent grade.
3.2.3 Internal standard,
(optional) reagent grade. p-Cymene and
n-hexylbenzene are
suitable internal standards for many solvents.
3.2.4
Chromatographic quality helium, nitrogen, hydrogen, and air.
3.3 Standard preparation
3.3.1 Prepare analyte standard at a concentration of
1 µL of analyte per milliliter of ES by adding 5 µL of analyte
to a 5-mL volumetric flask partially filled with
ES. Fill the volumetric flask to the mark and invert 3 or 4
times for proper mixing. Other size volumetric flasks may also
be used to prepare the 1 µL/mL analyte standards. At least two
standards at 1 µL/mL are prepared. Standards must be used the
day they are prepared. In some cases, analyte standards in
concentrations other than 1 µL/mL may be more suitable,
especially with analytes that have extremely high or low OSHA
standards.
3.3.2 Injection of standards is accomplished
with a 1-µL or other suitable syringe. The syringe is rinsed
thoroughly in carbon disulfide between standards. Injector septa
should be checked for wear daily.
3.3.3 Injection sizes
other than 1-µL and injection by means of a gas chromatograph
autosampler are acceptable in most cases. 3.4
Preparation of samples
3.4.1 The status of the seals on each charcoal tube
is noted and recorded as intact, broken, or none.
3.4.2
The field identification number, the laboratory identification
number and signature of the industrial hygienist on each sample
seal are checked with those on the sample identification sheets.
3.4.3 The seal is removed and the charcoal tube is
opened with a glass tubing cutter at the end containing the
larger portion of charcoal. The front and back sections of
charcoal are transferred to separate 2-mL capped
vials. The glass wool plug and the small wad of urethane foam
separating the two sections of charcoal are discarded.
3.4.4 The charcoal lot number is noted in order that the
proper extraction efficiency is used in later calculations.
3.4.5 Gas chromatography parameters are set as
recommended in the instruments manual. Oven temperature and
column are varied until an optimum chromatogram is produced by
the analyte standard.
3.4.6 Once the internal standard
has been verified as not interfering with other peaks in the
chromatogram, the samples are extracted. One milliliter of ES is
dispensed into each sample vial. The vial is immediately sealed.
Each vial is swirled periodically to increase the rate of
extraction. Twenty to thirty minutes is typical for the
extraction process. 3.5 Analysis
3.5.1 The data processor can be calibrated to
provide results directly in units of mass. With a few of the
analytes an additional similar correction may be necessary due
to extraction efficiencies that change with concentration. The
linear nature of the flame ionization detector allows the use of
a point calibration, but the bracketing of samples with
analytical standards is a good practice. The calculation of the
equivalent air concentration for an analytical standard is
detailed in Section
3.7.1.
3.5.2 Sample injection is accomplished with a
1-µL or other suitable syringe. The syringe is rinsed thoroughly
in carbon disulfide between samples. Injector septa should be
checked for wear periodically. Injection by means of a gas
chromatograph autosampler is acceptable.
3.5.3 Bracket
the samples with analytical standards if detected concentrations
are above the PEL.
3.5.4 When the identity of a
suspected analyte peak is in question, it should be confirmed by
GC/MS, GC/IR, or by retention time on at least two GC columns
containing different packing material. The identity of the
analyte should be considered suspect when detected
concentrations are above the PEL. 3.6 Interferences
Interferences to the analytical method will in most cases
appear as poor resolution of the analyte peak from other
components. This may be overcome by prudent selection of a more
suitable chromatographic condition or column.
3.7
Calculations
3.7.1 An equivalent air
concentration for analytical standards is used to calibrate the
data processor such that analytical results are obtained
directly in mass, mg.
where: |
W |
is weight of analyte in µg |
VS |
is volume of analyte in µL |
d |
is density of analyte in
µg/µL |
where: |
CV |
is air concentration reported to
IH |
VM |
is molar volume at 25°C and 760 mmHg,
24.46 L/mol |
W |
is weight of analyte |
Mr |
is molecular weight of the analyte |
V |
is air volume sampled |
EE |
is extraction
efficiency | 3.7.2 The following example is the calculation for
toluene:
W = 867 µg = 1 µL × 867
µg/µL |
CV = |
23.73 ppm = |
(24.46 L/mol)(867 µg)
(92.15 g/mol)(10 L)(0.97) | The calculations should be considered an example
only, and various parameters confirmed before used in actual
analysis. 3.8 Safety
precautions
3.8.1 Care must be taken when opening charcoal tubes
to avoid cuts to the hands.
3.8.2 Safety glasses must be
worn throughout the analytical procedure.
3.8.3 Work
involving solvents open to the atmosphere must be performed in a
hood. 3.9 Reporting results
3.9.1 When results uncorrected for air volume are
greater than 10 ppm, three significant digits will be reported.
For results below 10 ppm, the chemist will use his judgment, but
in no cases report more than three significant digits.
3.9.2 The estimated detection limit based on the lowest
mass per sample injected as a standard.
3.9.3 All
concentration levels down to the detection limit are reported.
3.9.4 If the concentration of analyte found on the back
section of the charcoal tube is equal to or greater than 25% of
the concentration found on the front section, the charcoal tube
is considered to be saturated and reported as such on the
analyst worksheet.
3.9.5 The presence of significant
peaks caused by unrequested components in the sample is noted on
the analyst worksheet and they are identified and quantitated if
possible.
3.9.6 All data processor print-outs and chart
recorder chromatograms are filed in a central file according to
laboratory sample identification number.
3.9.7
Analytical data and results are checked by a fellow chemist
before the completed analyst worksheets are given to the team
leader. 4. Analytes
The following table contains those
analytes which can be analyzed by this procedure. Standard size
charcoal tubes containing coconut base charcoal are used unless
specified otherwise in the table. Listed PELs are 8-h
time weighted averages unless denoted as a ceiling concentration
with a "(C)", before the PEL value. Before taking samples, the OSHA
Chemical Sampling Information at http://www.osha-slc.gov/index.html
should be consulted for additional and more detailed
information.
Table
4. Recommended Sampling Parameters for Analytes Covered by
This Procedure. |
|
ANALYTE |
PEL (ppm) |
air
vol (L) |
max
rate (L/min) |
NIOSH
Method |
|
Allyl alcohol |
2 |
10 |
0.2 |
1402 |
Allyl chloride |
1 |
48 |
0.2 |
1000 |
n-Amyl acetate |
100 |
10 |
0.2 |
1450 |
sec-Amyl acetate |
125 |
10 |
0.2 |
1450 |
Benzyl chloride |
1 |
10 |
0.2 |
1003 |
Bromoform |
0.5 |
10 |
0.2 |
1003 |
Butyl acetate |
150 |
10 |
0.2 |
1450 |
sec-Butyl acetate |
200 |
10 |
0.2 |
1450 |
tert-Butyl acetate |
200 |
10 |
0.2 |
1450 |
Butyl alcohol |
100 |
10 |
0.2 |
1401 |
sec-Butyl alcohol |
150 |
10 |
0.2 |
1401 |
tert-Butyl alcohol |
100 |
10 |
0.2 |
1400 |
n-Butyl glycidyl ether (BGE) |
50 |
10 |
0.2 |
1616 |
p-tert-Butyltoluene |
10 |
24 |
0.2 |
1501 |
Camphor |
2 mg/m3 |
24 |
0.2 |
1301 |
Carbon tetrachloride |
10 |
15 |
0.2 |
1003 |
Chlorobenzene (monochlorobenzene) |
75 |
10 |
0.2 |
1003 |
Chlorobromomethane |
200 |
5 |
0.2 |
1003 |
Cumene |
50 |
10 |
0.2 |
1501 |
Cyclohexane |
300 |
5 |
0.2 |
1500 |
Cyclohexanol |
50 |
10 |
0.2 |
1402 |
Cyclohexene |
300 |
5 |
0.2 |
1500 |
Diacetone alcohol
(4-hydroxy-4-methyl-2-pentanone) |
50 |
10 |
0.2 |
1402 |
o-Dichlorobenzene |
(C)50 |
3 |
0.2 |
1003 |
p-Dichlorobenzene |
75 |
|
0.05 |
1003 |
1,1-Dichloroethane |
100 |
10 |
0.2 |
|
1,2-Dichloroethylene |
200 |
3 |
0.2 |
1003 |
Dichloroethyl ether |
(C)15 |
15 |
1.0 |
1004 |
1,1-Dichloro-1-nitroethane** |
(C)10 |
15 |
1.0 |
1601 |
Difluorodibromomethane(F-12-B2)* |
100 |
10 |
0.2 |
1012 |
Diisobutyl ketone |
50 |
10 |
0.2 |
1300 |
Dioxane (diethylene dioxide) |
100 |
10 |
0.2 |
1602 |
Epichlorohydrin |
5 |
20 |
0.2 |
1010 |
Ethyl acetate |
400 |
6 |
0.2 |
1457 |
Ethyl sec-amyl ketone
(5-methyl-3-heptanone) |
25 |
25 |
0.2 |
1301 |
Ethyl bromide |
200 |
4 |
0.2 |
1011 |
Ethyl butyl ketone (3-heptanone) |
50 |
25 |
0.2 |
1301 |
Ethylene chlorohydrin** |
5 |
35 |
0.2 |
2513 |
Ethyl ether |
400 |
3 |
0.2 |
1610 |
Ethyl formate |
100 |
10 |
0.2 |
1452 |
Glycidol (2,3-epoxy-1-propanol) |
50 |
50 |
1.0 |
1608 |
n-Heptane |
500 |
4 |
0.2 |
1500 |
Hexachloroethane |
1 |
10 |
0.2 |
1003 |
n-Hexane |
500 |
4 |
0.2 |
1500 |
2-Hexanone (MBK) |
100 |
10 |
0.2 |
1300 |
sec-Hexyl acetate |
50 |
10 |
0.2 |
1450 |
Isoamyl acetate |
100 |
10 |
0.2 |
1450 |
Isoamyl alcohol |
100 |
10 |
0.2 |
1402 |
Isobutyl acetate |
150 |
10 |
0.2 |
1450 |
Isobutyl alcohol |
100 |
10 |
0.2 |
1401 |
Isophorone** |
25 |
12 |
0.2 |
2508 |
Isopropyl acetate |
250 |
8 |
0.2 |
1454 |
Isopropyl ether |
500 |
3 |
0.05 |
1618 |
Isopropyl glycidyl ether |
50 |
10 |
0.2 |
1620 |
Mesityl oxide |
25 |
25 |
0.2 |
1301 |
Methyl acetate |
200 |
7 |
0.2 |
1458 |
Methylal (dimethoxymethane) |
1000 |
2 |
0.2 |
1611 |
Methyl-(n-amyl)ketone |
100 |
25 |
0.2 |
1301 |
Methylcyclohexane |
500 |
4 |
0.2 |
1500 |
Methyl isobutyl carbinol |
25 |
10 |
0.2 |
1402 |
a-Methyl styrene |
(C)100 |
3 |
0.2 |
1501 |
Octane |
500 |
4 |
0.1 |
1500 |
Pentane |
1000 |
2 |
0.05 |
1500 |
2-Pentanone |
200 |
10 |
0.05 |
1300 |
Phenyl glycidyl ether |
10 |
50 |
0.1 |
1619 |
n-Propyl acetate |
200 |
10 |
0.2 |
1450 |
Propyl alcohol |
200 |
10 |
0.2 |
1401 |
Propylene dichloride |
75 |
10 |
0.2 |
1013 |
n-Propyl nitrate** |
25 |
70 |
0.1 |
S227 |
1,1,1,2-Tetrachloro-2, 2-difluoroethane |
500 |
2 |
0.035 |
1016 |
1,1,2, 2-Tetrachloro-1, 2-difluoroethane |
500 |
2 |
0.035 |
1016 |
1,1,2,2-Tetrachloroethane** |
5 |
10 |
0.2 |
1019 |
Tetrahydrofuran |
200 |
5 |
0.2 |
1609 |
Tetramethyl succinonitrile |
0.5 |
48 |
0.2 |
S155 |
1,2,3-Trichloropropane |
50 |
10 |
0.2 |
1003 |
Vinyl toluene |
100 |
24 |
0.2 |
1501 |
*Use two charcoal tubes in series for
sampling. **Use petroleum base charcoal for
sampling. | 5.
References
5.1 "NIOSH Manual
of Analytical Methods", ed. 4 Vol. 1-3 National Institute of
Occupational Safety and Health, U.S. Government Printing Office,
Washington, D.C. (1998)
| |