|
Method no.: |
PV2122 |
|
|
Control no.: |
T-PV2122-01-0304-M |
|
|
Target concentration: |
3 ppm (9.2 mg/m³) |
|
|
Procedure: |
Samples are collected by drawing
a known volume of air through glass sampling tubes containing
XAD-2 resin coated with 10% (w/w) 1-naphthylisothiocyanate
(NITC). Samples are extracted with 2 mL N,N-dimethylformamide (DMF) and
analyzed by LC using a UV detector. |
|
|
Recommended air volume and
sampling rate: |
200 min at 0.1 L/min (20
L) |
|
|
Reliable quantitation limit: |
35 ppb |
|
|
Status of method: |
Partially evaluated method. This
method has been subjected to established evaluation procedures
of the Methods Development Team and is presented for
information and trial use. |
|
|
Date: |
April 2003 |
|
|
Chemist: |
Mary E. Eide |
|
|
Methods Development Team Industrial Hygiene
Chemistry Division OSHA Salt Lake Technical
Center Sandy UT 84070-6407
|
1.
General Discussion
1.1 Background
1.1.1 History
Air
samples were received at SLTC requesting analysis for
1-amino-2-propanol (APOL) collected on tubes containing
XAD-2 resin coated with 10% 1-naphthylisothiocyanate
(NITC). The toxicology of APOL is similar to ethanolamine,
therefore 3 ppm was used as the target concentration in
this study. Ethanolamine, OSHA Method PV2111, was
collected on the same tubes, so the analytical parameters
were used as a starting point1.
The extraction and retention studies were performed using
the Bakerbond CN LC column with a mobile phase of 90:10
isooctane:isopropanol. The quantitation was performed at
280 nm because the N,N-dimethylformamide (DMF) peak did
not respond much at this wavelength and therefore did not
interfere with the integration of the APOL peak. This
column became irreparably clogged and could not be
replaced soon, therefore a Restek Pinnacle TO-11 LC column
and a mobile phase of 55:45:0.2
acetonitrile:water:phosphoric acid was used for the
remaining tests. The peak shape on the TO-11 column was
sharper, giving greater sensitivity. The samples were
extracted with 2 mL DMF, and had good extraction
efficiencies averaging 99.6%. The retention efficiency
study showed no APOL on the back up section of the spiked
tube or back up tube, for tubes spiked with 308 µg, that
had 20 L humid air drawn through them. The storage study
showed little loss for samples stored for up to 14 days
under both refrigerated and ambient
conditions.
1.1.2 Toxic effects (This section is
for information only and should not be taken as the basis
of OSHA policy.)2,3
APOL
is a moderate skin irritant, severe eye irritant, and
moderate mucous membrane irritant. It is moderately toxic
by ingestion and mucous membrane absorption. It is a
poison by intraperitoneal route. It causes a delayed
pulmonary edema, 4-6 hours after exposure, resulting in
chemical pneumonia.
1.1.3 Workplace exposure4,5
APOL
is used as an emulsifying agent, in dry-cleaning soaps,
soluble textile oils, wax removers, metal cutting oils,
cosmetics, emulsion paints, plasticizers, and
insecticides. National Occupational Exposure Survey
(1981-1983) found about 132,873 American workers
potentially exposed to APOL.
1.1.4 Physical
properties and other descriptive information6,7
CAS number: |
78-96-6 |
IMIS8: |
A606 |
molecular weight: |
75.13 |
vapor density: |
2.6 |
melting point: |
1.4°C |
boiling point: |
159.9°C |
appearance: |
clear liquid |
vapor pressure: |
0.0013 kPa @20°C |
odor: |
mild ammoniacal |
flash point: |
77°C (165 °F)(cc) |
autoignition |
|
density: |
0.9619 |
temperature: |
374°C (705°F) |
molecular formula: |
C3H9NO |
solubility: |
water, alcohol, acetone |
|
|
synonyms: |
1-aminopropan-2-ol;
2-hydroxypropylamine;
isopropanolamine; threamine |
structural formula: |
|
|
|
This method was evaluated according to the
OSHA SLTC “Evaluation Guidelines for Air Sampling Methods
Utilizing Chromatographic Analysis”9.
The Guidelines define analytical parameters, specify required
laboratory tests, statistical calculations and acceptance
criteria. The analyte air concentrations throughout this
method are based on the recommended sampling and analytical
parameters.
1.2 Detection limit of the
overall procedure (DLOP) and reliable quantitation limit
(RQL)
The DLOP is measured as mass per sample and
expressed as equivalent air concentrations, based on the
recommended sampling parameters. Ten samplers were spiked
with equal descending increments of analyte, such that the
highest sampler loading was 9.2 µg of APOL. This is the
amount spiked on a sampler that would produce a peak at
least 10 times the response for a sample blank. These spiked
samplers were analyzed with the recommended analytical
parameters, and the data obtained used to calculate the
required parameters (standard error of estimate (SEE) and
slope) for the calculation of the DLOP. The slope was
1.67×104 and the SEE was 3572. The RQL is
considered the lower limit for precise quantitative
measurements. It is determined from the regression line
parameters obtained for the calculation of the DLOP,
providing 75% to 125% of the analyte is recovered. The DLOP
and RQL were 0.64 µg and 2.14 µg, respectively. The recovery
at the RQL was 99.3%.
Table 1.2 Detection Limit of the
Overall Procedure for APOL
|
mass
per sample (µg) |
area
counts (µV-s) |
|
0.00 0.92 1.84 2.76 3.68 4.60 5.52 6.44 7.36 8.28 9.2 |
4170 20235 36189 48602 65229 80445 95767 119654 132564 139170 154875 |
| |
Figure 1.2.1 Plot of data to determine
the DLOP/RQL for APOL at 280 nm using a TO-11 column.
(y = 1.67×104x + 4666; SEE = 3572) |
Figure 1.2.2 Chromatogram of the APOL
peak in a standard near the RQL at 254 nm using a
TO-11 column. Key: (1) DMF; (2) APOL |
Figure 1.2.3. Chromatogram of the APOL
peak in a standard near the RQL at 280 nm using a
TO-11 column. Key: (1) DMF; (2) APOL; (3) & (4)
interferences from NITC. | Below are the chromatograms of the RQL
level. 2. Sampling
Procedure
All safety practices that apply to the work
area being sampled should be followed. The sampling equipment
should be attached to the worker in such a manner that it will
not interfere with work performance or safety.
2.1 Apparatus
2.1.1 Samples are
collected using a personal sampling pump calibrated, with
the sampling device attached, to within ±5% of the
recommended flow rate.
2.1.2 Samples are collected
with 7-cm × 4-mm i.d. × 7-mm o.d. glass sampling tubes
packed with two sections (80/40 mg) of XAD-2 resin coated
with 10% by weight 1-naphthylisothiocyanate. The sections
are held in place and separated with a glass wool plugs.
For this evaluation, commercially prepared sampling tubes
were purchased from SKC, Inc. (catalog no.
226-30-18). 2.2
Reagents
None required.
2.3 Technique
2.3.1 Immediately before
sampling, break off the ends of the flame-sealed tube to
provide an opening approximately half the internal
diameter of the tube. Wear eye protection when breaking
ends. Use tube holders to minimize the hazard of broken
glass. All tubes should be from the same lot.
2.3.2
The smaller section of the adsorbent tube is used as a
back-up and is positioned nearest the sampling pump.
Attach the tube holder to the sampling pump so that the
adsorbent tube is in an approximately vertical position
with the inlet facing down during sampling. Position the
sampling pump, tube holder and tubing so they do not
impede work performance or safety.
2.3.3 Draw the
air to be sampled directly into the inlet of the tube
holder. The air being sampled is not to be passed through
any hose or tubing before entering the sampling
tube.
2.3.4 After sampling for the appropriate
time, remove the adsorbent tube and seal it with plastic
end caps. Seal each sample end-to-end with an OSHA-21 form
as soon as possible.
2.3.5 Submit at least one
blank sample with each set of samples. Handle the blank
sample in the same manner as the other samples except draw
no air through it.
2.3.6 Record sample air volumes
(liters), sampling time (minutes) and sampling rate
(L/min) for each sample, along with any potential
interferences on the OSHA-91A form.
2.3.7 Submit
the samples to the laboratory for analysis as soon as
possible after sampling. If delay is unavoidable, store
the samples at refrigerator temperature. Ship any bulk
samples separate from the air samples.
2.4 Extraction efficiency
The extraction efficiency was determined by spiking
NITC-coated XAD-2 tubes with APOL at 0.1 to 2 times the
target concentration. These samples were stored overnight at
ambient temperature and then extracted for 30 minutes with
shaking, and analyzed. The mean extraction efficiency over
the studied range was 99.6%. The wet extraction efficiency
was determined at 1 times the target concentration by
spiking the analyte onto NITC-coated XAD-2 tubes which had
20-L humid air (absolute humidity of 15.9 mg/L of water,
about 80% relative humidity at 22.2°C) drawn through them
immediately before spiking. The mean recovery for the wet
samples was 99.1%.
Table
2.4 Extraction Efficiency (%) of APOL
|
level |
sample
number |
|
× target concn |
µg per sample |
1 |
2 |
3 |
4 |
5 |
6 |
mean |
|
0.1 0.25 0.5 1.0 1.5 2.0
1.0
(wet) |
18.4 46 92 184 276 368
184 |
100.6 98.9 97.4 100.8 99.8 100.8
100.0 |
98.6 99.9 99.2 99.9 98.9 99.3
98.9 |
98.5 100.5 100.4 99.5 99.3 99.9
99.2 |
99.9 98.7 99.1 100.3 100.4 100.2
98.5 |
99.0 99.1 98.3 99.3 99.9 98.7
98.7 |
100.1 100.2 100.2 99.7 98.9 100.5
99.0 |
99.5 99.6 99.1 99.9 99.5 99.9
99.1 |
| 2.5
Retention efficiency
Six NITC-coated XAD-2 tubes were
spiked with 368 µg (6.0 ppm) of APOL and allowed to
equilibrate for 4 h. Each spiked tube was placed in series
with a second NITC-coated XAD-2 tube. Each sampling train
had 20-L humid air (absolute humidity of 15.9 mg/L of water,
about 80% relative humidity at 22.2°C) pulled through them
at 0.1 L/min. The samples were extracted and analyzed. The
mean recovery was 99.4%. There was no analyte found on the
backup section of any of the tubes or on the second, backup
tube.
Table
2.5 Retention Efficiency (%) of APOL
|
|
sample
number |
|
section |
1 |
2 |
3 |
4 |
5 |
6 |
mean |
|
front of
spiked tube rear of spiked tube front of series
tube back of series tube
total |
98.9 0.0 0.0 0.0
98.9 |
99.5 0.0 0.0 0.0
99.5 |
100.1 0.0 0.0 0.0
100.1 |
98.8 0.0 0.0 0.0
98.8 |
99.9 0.0 0.0 0.0
99.9 |
99.3 0.0 0.0 0.0
99.3 |
99.4 0.0 0.0 0.0
99.4 |
| 2.6 Sample
storage
Fifteen NITC-coated XAD-2 tubes were each
spiked with 184 µg (3.0 ppm) of APOL. They were allowed to
equilibrate for 4 h, then 20 L of air, with an absolute
humidity of 15.7 milligrams of water per liter of air (about
80% relative humidity at 23°C), was drawn through them.
Three samples were analyzed immediately, and the rest were
sealed. Six were stored at room temperature (23 °C), while
the other six were stored at refrigerated temperature (4°C).
Three samples stored at room temperature and three samples
stored at refrigerated temperature were analyzed after 7
days and the remaining three after 14 days. The amounts
recovered, indicate good storage stability for the time
period studied.
Table
2.6 Storage Test APOL
|
time (days) |
ambient
storage recovery (%) |
|
refrigerated
storage recovery (%) |
|
0 7 14 |
100.1 99.8 99.1 |
99.5 99.0 98.9 |
98.5 98.6 99.7 |
|
100.1 98.9 99.7 |
99.5 99.8 99.0 |
98.5 99.1 98.3 |
| 2.7
Recommended air volume and sampling rate
Based on the
data collected in this evaluation, 20-L air samples should
be collected at a sampling rate of 0.1 L/min for 200
minutes.
2.8 Interferences (sampling)
2.8.1 There are no known compounds which will
severely interfere with the collection of APOL. Other
primary and secondary amines will collect on this media,
and form derivatives with the NITC, affecting the ability
of the tube to collect APOL, so sampling time should be
adjusted if high concentrations of amines are
expected.
2.8.2 Suspected interferences should be
reported to the laboratory with submitted samples.
3. Analytical
Procedure
Adhere to the rules set down in your Chemical
Hygiene Plan. Avoid skin contact and inhalation of all
chemicals and review all appropriate MSDSs.
3.1 Apparatus
3.1.1 A liquid
chromatograph equipped with a UV detector. For this
evaluation, a Waters 600 controller and pump were used,
with a Waters 2487 dual wavelength absorbance detector,
and a Waters 717 plus autosampler was used in this
evaluation.
3.1.2 An LC column capable of
separating APOL from the desorption solvent, interferences
in the NITC, and any potential interferences. A 4.6 × 250
cm column packed with 5µ Bakerbond cyanopropyl (JT Baker,
Phillipsburg, NJ), and a 4.6 × 250 cm column packed with
5µ Pinnacle TO-11 (Restek, Bellefonte, PA) were used in
the evaluation.
3.1.3 An electronic integrator or
some other suitable means of measuring peak areas. A
Waters Millennium32 Data System was used in this
evaluation.
3.1.4 Glass vials with
poly(tetrafluoroethylene)-lined caps. For this evaluation
4-mL vials were used.
3.1.5 A dispenser capable of
delivering 2.0 mL of desorbing solvent to prepare
standards and samples. If a dispenser is not available, a
2.0-mL volumetric pipet may be used.
3.1.7
Volumetric flasks - 10-mL and other convenient sizes for
preparing standards.
3.2 Reagents
3.2.1
1-Amino-2-propanol. Aldrich lot 03417PU 93% (7%
2-amino-1-propanol) was used in this
evaluation.
3.2.2 N,N-Dimethyl formamide, reagent
grade. Fisher 99.5%+ (lot 933764) was used for this
evaluation.
3.2.3 1-Naphthylisothiocyanate, reagent
grade. Aldrich 95%+ (lot 09925MY) was used in this
evaluation.
3.2.4 Isopropyl alcohol, HPLC grade.
Fisher 99.9% (lot 022995) was used in this
evaluation.
3.2.5 Isooctane, HPLC grade. Fisher
99.0%+ (lot 025050) was used in this
evaluation.
3.2.6 Acetonitrile, HPLC grade. Fisher
99.9%+ (lot 023721) was used in this
evaluation.
3.2.7 Deionized water (DI water). A
Barnstead NANOpure Diamond water deionizer was used in
this evaluation.
3.2.8 Phosphoric acid, Baker
Analyzed Reagent grade. Baker 85.9% (lot D25821) was used
in this evaluation.
3.2.9 Mobile phase for the
normal phase analysis using the Bakerbond CN column was
95:5 isooctane:isopropyl alcohol.
3.2.10 Mobile
phase for the reverse phase analysis using the Restek
Pinnacle TO-11 column was 50:50:0.2
acetonitrile:water:phosphoric acid.
3.3 Standard
preparation
3.3.1 Prepare two stock
standards. A stock standard of a concentration of 2 mg/mL
may be prepared by weighing out about 50 mg of NITC in a
10-mL flask, then weigh out 20 mg APOL placing the drops
on top of the NITC in the flask, then weigh out about 50
mg more NITC on top of the APOL. Allow the amine to react
with the NITC for 10 minutes (if other aliphatic amines
are being prepared at the same time it is necessary to
allow them to react 1 hour, other alcohol amines take 10
minutes). Partially fill the volumetric flask with DMF and
allow to sit at least 10 minutes to dissolve the
derivative, swirl to dissolve, and fill to the mark with
DMF. Do not place the flask in a sonic bath to try to get
the derivative to go into solution, as this will destroy
the derivative. There must always be an excess of the NITC
for the derivative to be completely formed. There is one
amine group which will react with the NITC, so this mole
ratio must be used in calculating the amount of NITC to be
added. For example, the amount of NITC needed for the
above stock standard would be calculated:
20 mg APOL × (NITC
MW=185.25/APOL MW=75.13) = 49.3 mg NITC
In the above stock
standard preparation a total of 100 mg NITC was weighed
out so that an excess of NITC was present.
3.3.2
Diluted standards are prepared with a solution of 1 mg/mL
NITC in DMF, so that an excess of NITC is always present.
Bracket sample concentrations with working standard
concentrations. If sample concentrations are higher than
the concentration range of prepared standards, either
analyze higher standards, or dilute the sample. The higher
standards should be at least as high in concentration as
the highest sample. Diluted samples should be prepared
with a solution of 1 mg/mL NITC in the DMF. The range of
standards used in this study was from 0.5 to 368
µg/mL. 3.4 Sample
preparation
3.4.1 Remove the plastic
end caps from the sample tubes and carefully transfer the
adsorbent sections to separate 4-mL vials. Discard the
glass tube, urethane foam plug and glass wool
plug.
3.4.2 Add 2.0 mL of DMF to each vial using
the same dispenser as used for preparation of
standards.
3.4.3 Immediately seal the vials with
poly(tetrafluoroethylene)-lined caps.
3.4.4 Shake
the vials on a shaker for 30 minutes.
3.5 Analysis
3.5.1 Liquid chromatograph
conditions.
LC conditions
normal phase
|
column: |
Bakerbond cyanopropyl
(CN) 5-µ column 4.6 × 250 cm |
injection size: |
10 µL |
mobile phase: |
2 mL/min 95:5 isooctane:
isopropyl alcohol |
detector: |
UV at 254 and 280
nm |
run time: |
10 min |
retention
times: |
1.98 min NITC; 4.63 min
DMF (at 280 nm this peak is minimal); 6.09 min
APOL |
Figure 3.5.1.1 A chromatogram of 184
µg/mL APOL in DMF with NITC at 254 nm using a
Bakerbond CN column. Key: (1) NITC; (2) DMF; (3)
APOL. |
Figure 3.5.1.2 A chromatogram of 184
µg/mL APOL in DMF with NITC at 280 nm using a
bakerbond CN column. Key: (1) NITC; (2)
APOL. |
LC conditions
reverse phase
|
column: |
Restek Pinnacle TO-11
5-µ column, 4.6 × 250 cm |
injection
size: |
10 µL |
mobile
phase: |
1.5 mL/min 50:50:0.2
acetonitrile:water:phosphoric acid |
detector: |
UV at 254 and 280
nm |
run time: |
14 min |
retention
times: |
1.92 min DMF (at 280 nm
this peak is minimal); 2.49 min APOL; 30.4 min
NITC |
Figure 3.5.1.3 A chromatogram of 184
µg/mL APOL in DMF with NITC at 254 nm using a TO-11
column. Key: (1) APOL; (2) interference from NITC;
(3) NITC; (4) NTIC. |
Figure 3.5.1.4 A chromatogram of 184
µg/mL APOL in DMF with NITC at 280 nm using a TO-11
column. Key: (1) APOL; (2) interference from NITC;
(3) NITC. | 3.5.2 Peak areas are measured by an
integrator or other suitable means.
3.5.3 An
external standard (ESTD) calibration method is used. A
calibration curve can be constructed by response of
standard injections versus micrograms of analyte per
sample. Bracket the samples with freshly prepared
analytical standards over a range of
concentrations.
Figure 3.5.3. Calibration curve of APOL
at 280 nm on TO-11 column. (Y =
3.31×104x -
2.98×104). | 3.6 Interferences (analytical)
3.6.1 Any compound that
produces a LC response and has a similar retention time as
the analyte is a potential interference. If any potential
interferences were reported, they should be considered
before samples are extracted. Generally, chromatographic
conditions can be altered to separate an interference from
the analyte.
3.6.2 When necessary, the identity or
purity of an analyte peak may be confirmed by a photodiode
array scan of the peak, by wavelength ratioing, or by
LC-mass spec. 3.7
Calculations
The amount of analyte per sampler is
obtained from the appropriate calibration curve in terms of
micrograms per sample, uncorrected for extraction
efficiency. This total amount is then corrected by
subtracting the total amount (if any) found on the blank.
The air concentration is calculated using the following
formulas.
CM = |
M |
where: |
CM is
concentration by weight (mg/m³) M is micrograms per
sample V is liters of air sampled EE
is extraction efficiency, in decimal form |
|
VEE |
CV = |
VMCM |
where: |
CV is
concentration by volume (ppm) VM is
molar volume at 25°C and 1 atm =
24.46 CM is concentration by
weight Mr is molecular weight =
75.13 |
|
VMr | 4. Recommendations for further study
Collection, reproducibility, and other detection limit
studies need to be performed to make this a validated
method.
1. OSHA Sampling and Analytical Methods.
http://www.osha-slc.gov/
(accessed 3/25/2003).
2. Lewis, R., Sax’s Dangerous
Properties of Industrial Materials, Van Nostrand Reinhold: New
York, 2002, p 212.
3. International Chemical Safety
Card. http://www.cdc.gov/niosh (accessed 3/25/2003).
4. Lewis, R., Ed, Hawley’s Condensed Chemical
Dictionary, John Wiley & Sons: New York, 2001, p
628.
5. National Occupational Exposure Survey
(1981-1983). http://www.cdc.gov/noes1 (accessed
3/25/2003).
6. Lewis, R., Sax’s Dangerous Properties of
Industrial Materials, Van Nostrand Reinhold: New York, 2002, p
212.
7. Lewis, R., Ed, Hawley’s Condensed Chemical
Dictionary, John Wiley & Sons: New York, 2001, p
628.
8. OSHA Chemical Sampling Information http://www.osha-slc.gov/
(accessed 3/25/2003).
9. Burright, D.; Chan, Y.; Eide,
M.; Elskamp, C.; Hendricks, W.; Rose, M. C. Evaluation
Guidelines for Air Sampling Methods Utilizing
Chromatographic Analysis; OSHA Salt Lake Technical Center,
U.S. Department of Labor: Salt Lake City, UT, 1999.
|
|
| |