January 15, 1982
NIOSH
EMERGENCY INFORMATION BULLETIN
ON THE USE OF SELF-CONTAINED BREATHING
APPARATUS IN LOW TEMPERATURES
Extreme caution should be
exercised by all persons using open circuit self-contained breathing apparatus (SCBA) in
hazardous environments during sub-freezing weather. SCBAs are widely used by fire fighters
combatting winter fires. All users who wear SCBAs in cold temperatures should take
particular note of the following important precautions:
- Moisture in the air cylinders
must be kept at an absolute minimum since small amounts of moisture in the air supply may
freeze and result in failure of the breathing apparatus.
- Always use a nosecup in the
SCBA facepiece when temperatures are below freezing. Failure to use a nosecup under such
circumstances can result in facepiece fogging and severely impaired vision. Chemical
anti-fog agents may not perform adequately in low temperatures.
- Carefully read the approval
label on the respirator to determine if it is necessary to install special accessories
prior to use of the SCBA in sub-freezing weather. Certain older U.S. Bureau of Mines
approved SCBAs require such low temperature accessories (SCBAs approved prior to March 25,
1972).
- When leaving an extremely hot
environment, such as a fire scene, and entering cold air (below or near freezing), always
place the SCBA facepiece in your turnout coat to keep it warm if it is to be quickly
reused. SCBAs when not being actively breathed can freeze-up very quickly.
- Use special care after washing
SCBA facepieces and breathing tubes to remove all moisture to prevent water drainage and
freeze-up of the regulator.
- SCBA alarms can fail in low
temperatures; therefore, visual checks of remaining service time should be made when SCBAs
are used in sub-freezing conditions.
- Be familiar with procedures on
how to cope with exhalation valves which can freeze open or closed in low temperatures.
(Contact the manufacturer or the State Fire Training Officer for specific instructions.)
- SCBAs are NIOSH laboratory
approved for use in temperatures down to -25° F. Therefore, if SCBAs are to be used in
temperatures below -25° F, extreme caution should be used.
- Also observe the following
general precautions:
a.
Use G-7.1, Type I, Grade D air or air of equivalent specification.
b. Follow all information
listed on the NIOSH/MSHA or BOM approval label for the specific SCBA in use.
c. Follow the manufacturer's
recommendations included in their instruction and maintenance manual accompanying the
SCBA.
d. Follow all applicable
Federal, State, and Local regulations concerned with the use of SCBAs.
e. Keep SCBAs in a warm
location between uses.
James A. Oppold, Ph.D., PE
Director
Division of Safety Research
November 15,
1982
RESPIRATOR
INFORMATION NOTICE
ON
MSA Powered Air Purifying Respirator
Mine Safety Appliance Company, Pittsburgh, PA
Model Numbers: 463354, 466607, 466608
Approval Number: TC-21C-186
On April 24, 1981, NIOSH
issued a Respirator Information Notice which described the results of a NIOSH study of the
MSA high efficiency powered air purifying respirator (PAPR) during use in a silica flour
mill. The observed workplace protection factors (defined as the ratio of the concentration
of contaminant outside the facepiece to the concentration of contaminant inside the
facepiece measured while the respirator is worn) were significantly below the anticipated
workplace protection factor of 1000. As a result, NIOSH stated that workers wearing the
MSA PAPR may not receive the protection they anticipated. NIOSH stated further that the
Institute had no evidence that the problem discovered in that study existed in other
industries or situations of use. NIOSH also stated that the Institute would conduct
further studies to evaluate the performance of the MSA PAPR against substances physically
and chemically different from silica flour to determine whether results with silica flour
were indicative of a problem associated with conditions of exposure or related to the
malfunction of equipment.
Staff of NIOSH subsequently
conducted a field evaluation of the half-mask MSA high efficiency PAPR at a primary lead
smelter. The challenge aerosols contained predominantly lead dust and or lead fume. From
this and other NIOSH studies, additional information has been developed and this Notice
supersedes the Notice of April 24, 1981.
This field evaluation of the
MSA PAPR produced the following preliminary results. The workplace protection factors
associated with the respirator was found to be approximately lognormally distributed. The
MSA PAPR produced a geometric mean workplace protection factor of 376 with a geometric
standard deviation of 2.64 against lead fume and lead dust. Approximately 95% of the
observed workplace protection factors for the MSA PAPR exceeded 77 while 84% of the
observed workplace protection factors were below 1000. During this study no wearer of the
MSA PAPR was exposed to concentrations of lead exceeding the permissible exposure limit
(PEL).
Subsequent to issuance of the
Respirator Information Notice of April 24, 1981, NIOSH and MSHA commenced proceedings to
withdraw the certification of the MSA PAPR. That action was predicated upon the
determination by NIOSH that the MSA PAPR, during use in a silica flour mill, apparently
did not provide the anticipated level of protection, i.e., a workplace protection factor
of 1000. That action was subsequently voluntarily dismissed by the agencies pending the
results of further studies. This study and additional studies of the PAPR class conducted
by NIOSH indicate that the previously anticipated protection factor of 1000 expected of
the entire class of PAPRs is inappropriately high. In view of this, the certification
withdrawal proceedings against the MSA PAPR, which were previously dismissed will not be
reinstituted. However, NIOSH recommends that users of PAPRs not rely upon them to
consistently provide a workplace protection factor of 1000.
The results of the additional
PAPR studies will be addressed in a subsequent Respirator Information Notice. For more
information on this subject, contact the Testing and Certification Branch, Division of
Safety Research, NIOSH, 944 Chestnut Ridge Road, Morgantown, West Virginia 26505, (304)
291-4331.
James A. Oppold,
Ph.D., PE, CSP
Director
Division of Safety Research
March 3, 1983
RESPIRATOR
INFORMATION NOTICE
ON
3M Powered Air
Purifying Respirator
3M, St. Paul, Minnesota
Model Number: W-344
Approval Number: TC-21C-246
Racal Powered Air Purifying
Respirator
Racal Airstream, Inc., Frederick, Maryland
Model Number: AH3
Approval Number: TC-21C-212
In a Respirator Information
Notice dated November 15, 1982, NIOSH recommended that powered air purifying respirators
(PAPRs) with high efficiency filters not be relied upon to consistently provide a
workplace protection factor of 1000. That recommendation was based upon the results of the
two studies of PAPRs with tight fitting facepieces described in that Notice as well as the
additional NIOSH study of helmeted PAPRs described in this Notice.
The NIOSH study of helmeted
PAPRs with high efficiency filters was conducted by NIOSH on the 3M W-344 PAPR and the
Racal AH3 PAPR at a secondary lead smelter. In this study the challenge aerosols contained
lead dust and/or lead fume.
This study produced the
following preliminary results. The workplace protection factors associated with both
respirator models were found to be approximately lognormally distributed. The results of
the t-tests indicate that there is no significant difference (P <.05) between the mean
workplace protection factors of the 3M and Racal PAPRs under the particular circumstances
of these studies. For both the 3M and Racal PAPRs, approximately 98% of the observed
workplace protection factors were below 1000. Approximately 95% of the observed workplace
protection factors for both the 3M and Racal PAPRs exceeded 33. The geometric mean
workplace protection factor for 3M and Racal PAPRs was 182 with a geometric standard
deviation of 3.2.
As stated in the November 15,
1982, Respirator Information Notice, the preliminary results of the NIOSH studies of the
MSA, 3M and Racal PAPRs indicate that the protection factor expected from this class of
respirators is inappropriately high.
For more information on this
subject, contact Glendel J. Provost, Division of Safety Research, NIOSH, 944 Chestnut
Ridge Road, Morgantown, West Virginia 26505. Commercial telephone number is (304) 291-4595
and the FTS number is 923-4595.
James A. Oppold, Ph.D., PE,
CSP
Director
Division of Safety Research
December 16,
1983
RESPIRATOR
USER'S NOTICE
Effects of
Chemicals on Rubber and Plastic Parts
of Self-contained Breathing Apparatus
The National Institute for
Occupational Safety and Health (NIOSH) has received several reports of damage to parts of
self-contained breathing apparatus that have apparently been exposed to concentrations of
chemicals. These exposures have occurred during emergency response activities after
accidental chemical vapor release and/or chemical discharge. The most recent report
concerned a leak of dimethyl amine in Benicia, California, on August 12 and 13, 1983.
Self-contained breathing apparatus and other equipment used during control of this leak
were reportedly rendered unserviceable after exposure.
In view of these reports, fire
fighting personnel who are engaged in emergency response activities should be equipped
with proper chemical protective clothing in addition to respiratory protection.
Information on the protective capabilities of such clothing should be obtained from the
clothing manufacturer.
NIOSH is conducting a study of
permeation of protective clothing materials by chemicals. Part of this study involves
preparation of a data base of information on that subject. As part of this data base,
NIOSH would appreciate receiving information on further cases of reported damage to
self-contained breathing apparatus by chemicals. Reports should be addressed to the
Testing and Certification Branch, Division of Safety Research, NIOSH, 944 Chestnut Ridge
Road, Morgantown, WV 26505-2888. Reports should include the name of the chemical, Chemical
Abstracts Service (CAS) Registry number, if known, identification and/or type of material
damaged, extent of damage, and either the approximate concentration of the chemical or
details of the exposure (e.g., exposure to liquid and/or vapor, temperature, wind
conditions, and degree of enclosure of exposure).
Thomas C. Purcell, Ph.D.
Acting Director,
Division of Safety Research
December 16,
1983
RESPIRATOR
USER'S NOTICE
Effects of
Heat and Flames on Rubber and Plastic
Parts of Self-contained Breathing Apparatus
The National Institute for
Occupational Safety and Health (NIOSH) has received several reports of damage to parts of
self-contained breathing apparatus that have apparently been exposed to excessive heat
and/or flames during fire fighting activities. A preliminary investigation of these
reports indicates that development of new turnout gear for fire fighters permits them to
enter and remain in higher temperatures and flame exposures. These higher temperatures and
flame exposures can apparently damage some presently-used rubber and plastic parts of
self-contained breathing apparatus.
NIOSH is proposing to include
requirements for high-temperature performance of self-contained breathing apparatus in
Title 30, Code of Federal Regulations, Part ll (30 CFR ll), the regulations governing
approval of respirators. NIOSH has been advised by self-contained breathing apparatus
manufacturers that they are developing new materials with greater resistance to heat and
flames. NIOSH recommends that fire fighters avoid overexposure of breathing apparatus
parts to high heat and/or flames, where possible.
NIOSH requests that fire
fighting personnel and others report further incidents of heat and flame damage of
self-contained breathing apparatus. Such reports should be sent to the Testing and
Certification Branch, Division of Safety Research, NIOSH, 944 Chestnut Ridge Road,
Morgantown, WV 26505-2888.
Thomas C. Purcell, Ph.D.
Acting Director,
Division of Safety Research
November 6,
1984
RESPIRATOR
USER'S NOTICE
USE OF
UNAPPROVED SUBASSEMBLIES
The National Institute for
Occupational Safety and Health (NIOSH) has received many questions and complaints in
regard to interchangeability of respirator subassemblies and unapproved modifications to
MSHA/NIOSH certified respirators. Further, some problems reported to NIOSH have, upon
investigation, been found to have been caused by user's modifying certified respirators
which have resulted in the modified respirator failing to perform as anticipated, thus
jeopardizing the respirator user.
MSHA/NIOSH respirator
certification regulations, Title 30 Code of Federal Regulations Part 11 (30 CFR 11), state
that approved respirators are ones that "are maintained in an approved condition and
are the same in all respects as those respirators for which a certificate has been
issued." [30 CFR 11, 11.2(b)] In addition, the regulations permit NIOSH/MSHA to only
approve complete respirator assemblies and prohibit the approval of respirator
subassemblies such as cylinders or air supply hoses. These requirements are intended to
insure that one manufacturer has overall control and responsibility for the integrity of
the approved respirator.
In some cases even minor
modifications to respirators may make significant changes in the performance of the
respirator. Manufacturers who modify certified respirators must test the modification to
determine if the respirator continues to meet the minimum requirements of 30 CFR 11, and
must submit the modifications to NIOSH. A user who modifies a certified respirator may not
be able to determine whether a change will decrease respiratory protection. Several cases
have been reported to NIOSH where unapproved modifications or use of an unapproved
subassembly have resulted in respirator failures. Therefore, users of NIOSH/MSHA approved
respirators are cautioned against interchanging subassemblies or making unapproved
modifications to their respiratory protective devices.
John B. Moran
Director
Division of Safety Research
June 28, 1985
RESPIRATOR
USERS NOTICE
Use and
Maintenance of Pressure-demand
Self-contained Breathing Apparatus
Since July 1, 1983, the
Occupational Safety and Health Administration (OSHA) Fire Brigade Standard, Title 29, Code
of Federal Regulations, Part 1910.156, has required that pressure-demand or other positive
pressure self-contained breathing apparatus be worn by fire brigade members performing
interior structural fire fighting. Although this standard is only applicable to all
industrial fire brigades and to municipal fire departments in states with state-OSHA
plans, other fire service organizations and industrial users of self-contained breathing
apparatus (SCBA) have also recognized the superior protective capabilities of
positive-pressure SCBA. As a result, there has been a steady change from demand to
pressure-demand SCBA in the United States.
To provide the increased
respiratory protection afforded by pressure-demand SCBA, it is generally necessary to
increase the static pressure within the facepiece. The complex mechanics necessary to
maintain this increased pressure and to control air flow when the facepiece is removed,
together with the wearer's physiological response to the pressure-demand system, have
presented problems to SCBA users.
Pressure demand SCBA requires
more careful maintenance and different training, than is required for demand SCBA.
Manufacturers have been providing maintenance and use instructions and training for
purchasers of pressure-demand SCBA. The National Institute for Occupational Safety and
Health (NIOSH) recommends that users of pressure-demand SCBA read those instructions,
follow them carefully in apparatus use and maintenance, and take advantage of the
manufacturer's training assistance. In addition to the manufacturers, training courses are
offered by Fire Service organizations and by private organizations.
In the area of pressure-demand
SCBA maintenance and repair, NIOSH strongly recommends that users have this service
performed by a manufacturer-trained representative. This service is required to assure
continued safe performance of pressure-demand SCBA.
Please advise NIOSH of any
problems encountered in maintenance and use of pressure-demand self-contained breathing
apparatus. Call the NIOSH Respirator Problem Coordinator, (304) 291-4595 (FTS 923-4595).
Use and Maintenance of
Pressure-Demand SCBA/Page 2
To assist you, NIOSH has
prepared the following list of manufacturer's and fire service organization personnel who
can provide further information on pressure-demand breathing apparatus training:
Clifton Precision
5100 State Road
Drexel Hill, PA 19026
Mr. Robert Gray (215) 622-1718
Globe Safety Equipment, Inc.
P.O. Box 7248
Dayton, OH 45407
Mr. Steven Bates (513) 224-7468
International Safety
Instruments, Inc.
P.O. Box 846
Lawrenceville, GA 30246
Mr. Donald Dawson (404) 962-2552
MSA
600 Penn Center Boulevard
Pittsburgh, PA 15235
Mr. Jay Mears (412) 273-5145
North Safety Equipment
2000 Plainfield Pike
Cranston, RI 02920
Mr. Richard T. Flynn (401) 943-4400
Rexnord
45 Great Valley Parkway
Malvern, PA 19355
Mr. Justin Mills (215) 647-7200 *
Scott Aviation
225 Erie Street
Lancaster, NY 14086
Mr. Dennis Browner (716) 683-5100
U.S.D.
3323 West Warner Avenue
Santa Ana, CA 92702
Mr. Brian Miller (714) 241-4601
National Draeger, Inc.
P.O. Box 120
Pittsburgh, PA 15230
Mr. Les Boord/Ms. Karen Cox/Mr. Richard Weaver (412) 787-8383
International Association of
Fire Chiefs
1329 18th Street, NW
Washington, DC 20036
Mr. Jan Thomas (202) 833-3420
International Association of
Fire Fighters
1750 New York Avenue, NW
Washington, DC 20006
Mr. Richard Duffy (202) 737-8484
International Society of Fire
Service Instructors
20 Main Street
Ashland, MA 01721
Mr. Ed McCormack (617) 881-5800
John B. Moran
Director
Division of Safety Research
* New contact for reporting
respirator problems (replaced Mr. John Moffa)
January 17,
1986
RESPIRATOR
USERS NOTICE
Inspection
of Certain Aluminum Cylinders for Breathing-gas Pressure
The light weight and high
charging pressure of aluminum cylinders have resulted in their widespread acceptance and
use with self-contained breathing apparatus (SCBA). The National Institute for
Occupational Safety and Health (NIOSH) estimates that more than half of the SCBA of 30-
and 60-minute duration in regular use today are equipped with aluminum cylinders.
Since first receiving reports
of defective fiber-glass wrapped aluminum cylinders in 1983, NIOSH has advised users of
potential hazards associated with use of certain fiber-glass wrapped aluminum cylinders.
At this time, NIOSH believes there is sufficient evidence to warrant issuance of this
NOTICE regarding inspection of fiber-glass wrapped aluminum cylinders.
The presently available
evidence indicates that fiber-glass wrapped aluminum cylinders manufactured under
Department of Transportation (DOT) exemptions DOT-E 7235 and DOT-E 8059 (including 2216
and 4500 psi) may, upon aging, develop neck cracks and may leak breathing gas during
storage and use. This may result in significant loss of breathing gas from an unattended
cylinder. If undetected, this loss of breathing gas could be dangerous to the user.
Based on this, NIOSH
recommends that where SCBA are equipped with fiber-glass wrapped aluminum cylinders,
inspection for cylinder pressure should be made at least weekly, for stored units. When
used on a daily basis, as in fire fighting, cylinder pressure should be checked daily and
immediately before use.
If a leak is suspected, the
cylinder and cylinder valve should be tested as prescribed in American National Standard,
Z88.5-1981, Practices for Respiratory Protection for the Fire Service, Section 6.2.4.2.
Leaks in cylinders should be
reported to the SCBA manufacturer who will, in turn report them to the cylinder
manufacturer. The numbers and charging pressures of leaking cylinders should also be
reported to DOT (Mr. Art Mallen, DOT Office of Hazardous Materials, 400 7th St. SW,
Washington, DC 20590) and to NIOSH (Mr. John Moran at the address shown at the top of this
letter).
Aluminum cylinders used with
SCBA, with exemption numbers other than DOT-E 7235 and DOT-E 8059 are not covered in this
notice. Self-contained self rescuers used in mines are also not included.
R E M I N D
E R
January 17,
1986
Manufacturers
of MSHA/NIOSH-approved SCBA
Incorporating DOT-E 7235 4500 Fiber-glass Wrapped Aluminum Cylinders
The following manufacturers
incorporate DOT-E 7235 4500 cylinders in their MSHA/NIOSH-approved SCBA:
- Bendix
- Clifton Precision
- Draeger
- Siebe Gorman
- Scott
- U.S.D. (SurvivAir)
DOT-E 7235 4500 cylinders must
be retrofitted by Luxfer (Telephone: 714-684-5110) with steel neck rings, to prevent
explosive rupture. DOT regulations prohibit charging of any DOT-E 7235 4500 cylinder that
has not been fitted with a steel neck ring. Any apparatus utilizing a DOT-E 7235 4500
cylinder without a neck ring, is considered unapproved by MSHA/NIOSH.
Change in
Address of Manufacturer's Contact
The following address change
has been reported to NIOSH for manufacturer's personnel who are responsible for handling
reports of problems with MSHA/NIOSH-approved respirators:
Clifton Precision:
New Address: 750 West Sproul
Road, Springfield, PA 19064-4084
Contact: Mr. Martin Ziegler
John B. Moran
Director, Division of Safety Research
|