DESFLURANE
Method number: |
106 |
|
Matrix: |
Air |
|
Target concentrations: |
1 ppm (6.9 mg/m3) and 75
ppm (515 mg/m3) |
OSHA PEL: |
None |
ACGIH TLV: |
None |
|
Procedure: |
Samples are collected by drawing a known volume of
air through standard size (6-mm o.d., 140/70) Anasorb 747 tubes.
Samples are desorbed with toluene and analyzed by GC using a flame
ionization detector (FID). |
|
Recommended air volume and sampling rate: |
3 L at 0.05 L/min |
|
Reliable quantitation limit: |
33.1 ppb (228
µg/m3) |
|
Standard error of estimate at the target
concentrations: |
5.3% at 1 ppm 5.6% at 75 ppm |
|
Status of method: |
Evaluated method. This method has been subjected to
the established evaluation procedures of the Organic Methods
Evaluation Branch. |
|
|
Date: April 1995 |
Chemist: Donald
Burright |
Organic Methods Evaluation Branch OSHA Salt Lake
Technical Center Salt Lake City, UT 84165-0200
1. General Discussion
1.1 Background
1.1.1 History
This method is an extension of the work that was done earlier to
produce OSHA Method 103. (Ref. 5.1) Desflurane was identified as a
new anesthetic gas undergoing testing for FDA approval during the
earlier work. FDA approved desflurane for use in the general
population in September 1992. Desflurane was not included in Method
103 because it required different analytical conditions. Initial
studies were conducted using both Anasorb CMS and Anasorb 747 as
adsorbents for collection of desflurane to parallel the method used
for isoflurane, halothane and enflurane. Anasorb 747 was selected as
the adsorbent for collection because a higher desorption efficiency
was obtained. The desorbing solvent was changed from carbon
disulfide to toluene. This reduced the loss of desflurane from the
sample which was caused by the heat generated when carbon disulfide
was added to the adsorbent. The evaluation was performed at two
target concentrations, 1 and 75 ppm, to parallel the earlier work.
There is no OSHA PEL or ACGIH TWA for desflurane, but NIOSH has a
recommended exposure limit (REL) for halogenated anesthetic gases of
2 ppm as a 60-min ceiling value (Refs. 5.2 and 5.3). Because OSHA
sometimes sets the TWA concentration at about one-half of the
ceiling value, the REL is the basis for the lower target
concentration. The higher target concentration was chosen because of
the close structural similarity between desflurane and enflurane,
which has an ACGIH TLV of 75 ppm.
1.1.2 Toxic effects (This section is for information only and
should not be taken as the basis of OSHA policy.)
Acute exposure to desflurane may cause irritation and redness to
the eyes, and dryness and irritation to the skin. Overexposure by
inhalation can lead to headaches, dizziness, drowsiness,
unconsciousness or death. Irritation of the mouth and throat can
occur with an acute exposure by inhalation. Acute exposure by
ingestion may lead to unconsciousness or death. (Ref. 5.4)
1.1.3 Workplace exposure
Desflurane is a new organic anesthetic gas and may be found in
operating rooms, teaching hospitals, dental offices, and veterinary
hospitals. The number of people potentially exposed is not known but
will rise as desflurane is accepted into general use.
1.1.4 Physical properties and other descriptive information (Ref.
5.4)
CAS number: molecular
weight: boiling point, °C: color: specific
gravity: molecular formula: vapor pressure, kPa
(mmHg): odor: flash point: synonyms: |
57041-67-5 168.04 22.8 clear 1.47 @
15°C C3H2OF6
89.2
(669.2) @ 20°C mild, ethereal >93°C (CC) Suprane™;
1,2,2,2-tetrafluoroethyl difluoromethyl ether;
I-653 |
structural formula: |
|
The analyte air concentrations throughout this method are based on the
recommended sampling and analytical parameters. Air concentrations listed
in ppm and ppb are referenced to 25°C and 101.3 kPa (760 mmHg).
1.2 Limit defining parameters
1.2.1 Detection limit of the analytical procedure
The detection limit of the analytical procedure is 13.6 pg. This
is the amount of analyte that will give a response that is
significantly different from the background response of a reagent
blank. (Sections 4.1 and 4.2)
1.2.2 Detection limit of the overall procedure
The detection limit of the overall procedure is 0.205 µg
per sample (9.93 ppb or 68.4
µg/m3). This is the amount of
analyte spiked on the sampler that will give a response that is
significantly different from the background response of a sampler
blank. (Sections 4.1 and 4.3)
1.2.3 Reliable quantitation limit
The reliable quantitation limit is 0.683 µg per sample
(33.1 ppb or 228 µg/m3). This is
the amount of analyte spiked on a sampler that will give a signal
that is considered the lower limit for precise quantitative
measurements. (Section 4.4)
1.2.4 Precision (analytical procedure)
The precisions of the analytical procedure, measured as the
pooled relative standard deviation over a concentration range
equivalent to the range of 0.5 to 2 times the target concentration
are 0.58% and 0.46% for the lower and higher target concentrations,
respectively. (Section 4.5)
1.2.5 Precision (overall procedure)
The precisions of the overall procedure at the 95% confidence
level for the ambient temperature 18- and 16-day storage tests (at
the target concentration) are 10.4% and 11.0% for the lower and
higher target concentrations, respectively. This includes an
additional 5% for sampling error. (Section 4.6)
1.2.6 Recovery
The recoveries of desflurane from samples used in the 18- and
16-day storage tests remained above 99.7% and 100.2% for the lower
and higher target concentrations, respectively when the samples were
stored at 22°C. (Section 4.7)
1.2.7 Reproducibility
Twelve samples spiked by liquid injection were submitted for
analysis by one of the OSHA Salt Lake Technical Center's service
branch laboratories. The samples were analyzed according to the
instructions in a draft copy of this procedure after 10 days of
storage at 4°C. No individual sample result deviated from its
theoretical value by more than the precision reported in Section
1.2.5. (Section 4.8)
2. Sampling Procedure
2.1 Apparatus
2.1.1 Samples are collected using a personal sampling pump
calibrated, with the sampling device attached, within ±5% at the
recommended flow rate.
2.1.2 Samples are collected with 7-cm × 4-mm i.d. × 6-mm o.d.
glass sampling tubes packed with two sections of Anasorb 747 (140/70
mg). The sections are held in place with a glass wool plug and two
urethane foam plugs. For this evaluation, commercially prepared
sampling tubes were purchased from SKC, Inc. (catalog no. 226-81A).
2.2 Reagents
None required.
2.3 Technique
2.3.1 Only properly trained personnel can sample in an operating
room or dental office, this is necessary to be in compliance with
OSHA's Exposure Control Plan for bloodbourne pathogens. (Ref. 5.5)
2.3.2 Break off the ends of the sampling tube immediately before
sampling. All tubes should be from the same lot.
2.3.3 Attach the sampling tube to the sampling pump with
flexible, non-crimpable tubing. It is desirable to utilize sampling
tube holders which have a protective cover to shield the employee
from the sharp, jagged end of the sampling tube. Position the tube
so that the sampled air first passes through the larger section.
2.3.4 Sampled air should not pass through any hose or tubing
before entering the sampling tube.
2.3.5 Attach the sampler vertically with the larger section
pointing downward in the worker's breathing zone to avoid
channeling. Position the sampler so it does not impede work
performance or safety.
2.3.6 Remove the sampling tube and seal it with plastic end caps
immediately after sampling for the appropriate time.
2.3.7 In order t prevent occupational exposure to SLTC personnel,
sampling tubes that may become contaminated with blood or other
potentially infectious materials are to be examined prior to
shipping and decontaminated (e.g., wiped off with bleach or other
disinfectant) as necessary. Contaminated items are not to placed or
stored in areas where food is kept, and decontamination should be
accomplished as soon as possible following the inspection where
contamination occurred. Decontamination is not to take place in any
area where food or drink is consumed. (Ref. 5.5)
2.3.8 Wrap each sample end-to-end with a Form OSHA-21 seal.
2.3.9 Submit at least one blank sample with each set of samples.
Handle the blank sampling tube in the same manner as the other
samples, except draw no air through it.
2.3.10 Record sample air volumes (in liters) for each sample,
along with any potential interferences.
2.3.11 Ship any bulk sample(s) in a container separate from the
air samples.
2.3.12 Submit the samples to the laboratory for analysis as soon
as possible after sampling. If delay is unavoidable, store the
samples at reduced temperature.
2.4 Sampler capacity
Sampler capacity is determined by measuring how much air can be
sampled before the analyte breaks through the sampler, i.e., the
sampler capacity is exceeded. Breakthrough is considered to occur when
the effluent from the sampler contains a concentration of analyte that
is 5% of the upstream concentration (5% breakthrough). Testing for
breakthrough was performed by using an FID to monitor the effluent
from sampling tubes containing only the 140-mg section of Anasorb 747.
Dynamically generated test atmospheres, which were about two times the
higher target concentration, were used for the capacity tests. The
samples were collected at 0.05 L/min and the relative humidity was
about 80% at 25°C. The 5% breakthrough air volume was calculated from
the data of duplicate determinations and is 3.83 L. (Section 4.9)
2.5 Desorption efficiency
2.5.1 The average desorption efficiencies for desflurane from
Anasorb 747 over the range of 0.5 to 2.0 times the target
concentrations were 101.1% and 102.9% for the lower and higher
target concentration respectively. (Section 4.10)
2.5.2 The desorption efficiencies at 0.05, 0.1 and 0.2 times the
target concentrations (TC) were found to be very high and are listed
below. (Section 4.10)
Table 2.5.2 Desorption
Efficiencies at 0.05 to 0.2 times TC, %
|
TC |
0.05×TC |
0.1×TC |
0.2×TC |
|
low high |
98.0 100.0 |
103.2 100.2 |
100.2 101.0 |
|
2.5.3 Desorbed samples remain stable for at least 20 and 33 h for
the lower and higher target concentration respectively.
2.6 Recommended air volume and sampling rate
2.6.1 For long-term samples, collect 3 L at 0.05 L/min.
2.6.2 For short-term samples, collect 0.75 L at 0.05 L/min.
2.6.3 When short-term samples are collected, the air
concentration equivalent to the reliable quantitation limit becomes
larger. For example, the reliable quantitation limit is 133 ppb (911
µg/m3) when 0.75 L is sampled.
2.7 Interferences (sampling)
2.7.1 There are no known compounds that will severely interfere
with the collection of desflurane on Anasorb 747. In general, the
presence of other contaminant vapors in the air will reduce the
capacity of Anasorb 747 to collect desflurane.
2.7.2 Suspected interferences should be reported to the
laboratory with submitted samples.
2.8 Safety precautions (sampling)
2.8.1 The sampling equipment should be attached to the worker in
such a manner that it will not interfere with work performance or
safety.
2.8.2 All safety practices that apply to the work area being
sampled should be followed.
2.8.3 Protective eyewear should be worn when breaking the ends of
the glass sampling tubes.
3. Analytical Procedure
3.1 Apparatus
3.1.1 Gas chromatograph equipped with an FID. For this
evaluation, a Hewlett-Packard 5890A Gas Chromatograph equipped with
a 7673A Automatic Sampler was used. A Forma Scientific Model 2006
refrigerated circulator was used to cool the sample tray of the HP
7673A to 10°C.
3.1.2 A GC column capable of separating the analyte of interest
from the desorption solvent, internal standard and any
interferences. A 60-m × 0.32-mm i.d. fused
silica SPB-1® column with a 4-µm df (Supelco, Inc,
Bellefonte, PA) was used in the evaluation.
3.1.3 An electronic integrator or some other suitable means of
measuring peak areas. A Waters 860 Networking Computer System was
used in this evaluation.
3.1.4 Two-milliliter vials with poly(tetrafluoroethylene)-lined
caps.
3.1.5 A dispenser capable of delivering 1.0 mL of desorbing
solvent to prepare standards and samples. If a dispenser is not
available, a 1.0-mL volumetric pipet may be used.
3.2 Reagents
3.2.1 Desflurane, USP. The desflurane used in this evaluation
was manufactured by Anaquest (Liberty Corner, NJ).
3.2.2 Toluene, chromatographic grade or better. The toluene used
in this evaluation was Optima Grade and was purchased from Fisher
Scientific (Fair Lawn, NJ).
3.2.3 Desorption solvent. The desorption solvent was toluene and
the benzene contaminant was used as the internal standard.
3.2.4 GC grade nitrogen, air, and hydrogen.
3.3 Standard preparation
3.3.1 Prepare two concentrated stock standards of desflurane in
toluene. Prepare working analytical standards by injecting
microliter amounts of concentrated stock standards into 2-mL vials
containing 1 mL of desorption solvent delivered from the same
dispenser used to desorb samples. For example, to prepare a target
level standard of desflurane, inject 20 µL of a stock
solution containing 77.37 mg/mL of desflurane in toluene into 1 mL
of desorption solvent.
3.3.2 Bracket sample concentrations with working standard
concentrations. If samples fall outside of the concentration range
of prepared standards, prepare and analyze additional standards to
ascertain the linearity of response.
3.4 Sample preparation
3.4.1 Remove the plastic end caps from the sample tube and
carefully transfer each section of the adsorbent to separate 2-mL
vials. Discard the glass tube, urethane foam plugs and glass wool
plug.
3.4.2 Add 1.0 mL of desorption solvent to each vial using the
same dispenser as used for preparation of standards.
3.4.3 Immediately seal the vials with
poly(tetrafluoroethylene)-lined caps.
3.4.4 Shake the vials vigorously several times during the next 30
min.
3.5 Analysis
3.5.1 Analytical conditions
GC conditions |
zone |
|
temperatures: |
60°C (column), hold 4 min, ramp at 5°C/min to
90°C, hold 0 min, ramp at 20°C/min to 150°C, hold 5
min 250°C (injector) 300°C (detector) |
run time: |
18 min |
column gas flow: |
2.7 mL/min (hydrogen) |
septum purge: |
1.9 mL/min (hydrogen) |
injection size: |
1.0 µL (15.5 : 1 split) |
column: |
60-m × 0.32-mm i.d. capillary SPB-1
(4.0-µm df) |
retention times: |
3.05 min (desflurane) 10.7 min
(benzene) |
FID conditions |
|
hydrogen flow: |
38 mL/min |
air flow: |
450 mL/min |
makeup flow: |
30 mL/min (nitrogen) |
Figure 3.5.1.1. Chromatogram obtained at the low TC
with the recommended conditions. Peak identification: (1) air
peak, (2) desflurane, (3) benzene, (4)
toluene.
Figure 3.5.1.2. Chromatogram obtained at the hight TC
with the recommended conditions. Peak identification: (1) air
peak, (2) desflurane, (3) benzene, (4)
toluene.
3.5.2 An internal standard (ISTD) calibration method is used. A
calibration curve can be constructed by plotting
ISTD-corrected response of standard injections versus
micrograms of analyte per sample. Bracket the samples with freshly
prepared analytical standards over a range of concentrations.
Figure 3.5.2.1. Calibration curve at the low TC made
from data of Table 4.5.1.
Figure 3.5.2.2. Calibration curve at the high TC made
from data of Table 4.5.2.
3.6 Interferences (analytical)
3.6.1 Any compound that produces an FID response and has a
similar retention time as the analyte or internal standard is a
potential interference. If any potential interferences were
reported, they should be considered before the samples are desorbed.
Generally, chromatographic conditions can be altered to separate an
interference from the analyte.
3.6.2 When necessary, the identity or purity of an analyte peak
may be confirmed with additional analytical data (Section 4.11).
3.7 Calculations
The amount of desflurane per sample is obtained from the
appropriate calibration curve in terms of micrograms per sample,
uncorrected for desorption efficiency. The back (70-mg) section is
analyzed primarily to determine if there was any breakthrough from the
front (140-mg) section during sampling. If a significant amount of
analyte is found on the back section (e.g., greater than 25% of the
amount found on the front section), this fact should be reported with
the sample results. If any analyte is found on the back section, it is
added to the amount on the front section. This amount is then
corrected by subtracting the total amount (if any) found on the blank.
The air concentration is calculated using the following formulae.
mg/m3 = |
micrograms of analyte per sample, blank
corrected
liters of air sampled × desorption
efficiency |
ppm = |
mg/m3 ×
24.46
molecular weight of analyte |
where 24.46 is the molar volume at 25°C and 101.3 kPa (760
mmHg) 168.04 = molecular weight
of desflurane
3.8 Safety precautions (analytical)
3.8.1 Adhere to the rules set down in your Chemical Hygiene
Plan.
3.8.2 Avoid skin contact and inhalation of all chemicals.
3.8.3 Wear safety glasses, gloves and a lab coat while in the
laboratory areas and working with chemicals.
4. Backup Data
4.1 Determination of detection limits
Detection limits, in general, are defined as the amount (or
concentration) of analyte that gives a response
(YDL) that is significantly different (three
standard deviations (SDBR)) from the
background response (YBR).
YDL - YBR
= 3(SDBR)
The measurement of YBR and
SDBR in chromatographic methods is typically
inconvenient and difficult because YBR is
usually extremely low. Estimates of these parameters can be made with
data obtained from the analysis of a series of analytical standards or
samples whose responses are in the vicinity of the background
response. The regression curve obtained for a plot of instrument
response versus concentration of analyte will usually be linear.
Assuming SDBR and the precision of the data
about curve are similar, the standard error of estimate (SEE) for the
regression curve can be substituted for SDBR
in the above equation. The following calculations derive a formula for
DL:
Yobs |
= observed response |
Yest |
= estimated response from regression curve |
n |
= total number of data points |
k |
= 2 for linear regression
curve |
At point YDL on the regression curve
YDL = A(DL) +
YBR
A =
analytical sensitivity (slope)
therefore
Substituting 3(SEE) + YBR for
YDLgives
4.2 Detection limit of the analytical procedure (DLAP)
The DLAP is measured as the mass of analyte actually introduced
into the chromatographic column. Ten analytical standards were
prepared in equal descending increments with the highest standard
containing 2.573 µg/mL of desflurane. This is the concentration
that would produce a peak approximately 10 times the background noise
of a reagent blank near the elution time of the analyte. These
standards, and the reagent blank, were analyzed with the recommended
analytical parameters (1-µL injection with a 15.5 : 1 split),
and the data obtained were used to determine the required parameters
(A and SEE) for the calculation of the DLAP. Values of 3.071 and
13.965 were obtained for A and SEE respectively. DLAP was calculated
to be 13.6 pg.
Table 4.2 Detection Limit of the
Analytical Procedure
|
concentration |
mass on column |
area counts |
(µg/mL) |
(pg) |
(µV-s) |
|
0 0.257 0.515 0.772 1.029 1.286 1.544 1.801 2.058 2.315 2.573 |
0 16.58 33.23 49.81 66.39 82.97 99.61 116.2 132.8 149.4 166.0 |
0 48.2 120.7 154.9 196.2 285.9 291.3 365.8 424.7 460.4 505.8 |
|
Figure 4.2. Plot of the data from Table 4.2 to determine
the DLAP of desflurane.
4.3 Detection limit of the overall procedure (DLOP)
The DLOP is measured as mass per sample and expressed as equivalent
air concentration, based on the recommended sampling parameters. Ten
samplers were spiked with equal descending increments of analyte, such
that the highest sampler loading was 2.573 µg/sample. This is
the amount spiked on a sampler that would produce a peak approximately
10 times the background response for a sample blank. These spiked
samplers, and a sample blank, were analyzed with the recommended
analytical parameters, and the data obtained used to calculate the
required parameters (A and SEE) for the calculation of the DLOP.
Values of 200.32 and 13.69 were obtained for A and SEE, respectively.
The DLOP was calculated to be 0.205 µg/sample (9.93 ppb or 68.4
µg/m3).
Table 4.3 Detection Limit of the
Overall procedure
|
mass per sample (µg) |
area counts (µV-s) |
|
0 0.257 0.515 0.772 1.029 1.286 1.544 1.801 2.058 2.315 2.573 |
0 61.8 113.7 144.1 192.2 265.5 305.5 342.7 396.8 485.3 523.0 |
|
Figure 4.3. Plot of the data to determine the
DLOP.
4.4 Reliable quantitation limit (RQL)
The RQL is considered the lower limit for precise quantitative
measurements. It is determined from the regression line parameters
obtained for the calculations of the DLOP (Section 4.3), providing at
least 75% of the analyte is recovered. The RQL is defined as the
amount of analyte that gives a response
(YRQL) such that
YRQL - YBR
= 10(SDBR)
therefore
Figure 4.4. Chromatogram of the RQL for desflurane on
Anasorb 747. Peak identification: (1) contaminant form Anasob 747,
(2) desflurane.
The RQL for desflurane was calculated to be 0.683 µg/sample
(33.1 ppb or 228
µg/m3). The recovery
at this concentration is 91.2%.
4.5 Precision (analytical method)
The precision of the analytical procedure is measured as the pooled
relative standard deviation (RSDP). Relative
standard deviations are determined from six replicate injections of
desflurane standards at 0.5, 0.75, 1, 1.5 and 2 times the target
concentrations. After assuring that the RSDs satisfy the Cochran test
for homogeneity at the 95% confidence level,
RSDP was calculated to be 0.58% and 0.46%
for the lower and higher target concentration, respectively.
Table 4.5.1 Instrument Response to
Desflurane at Low TC
|
× target concn (µg/mL) |
0.5× 10.29 |
0.75× 15.44 |
1× 20.58 |
1.5× 30.87 |
2× 41.16 |
|
area counts (mV-s) |
2.391 2.394 2.381 2.416 2.417 2.388 |
3.554 3.559 3.549 3.548 3.550 3.530 |
4.743 4.746 4.703 4.729 4.716 4.763 |
6.747 6.777 6.827 6.763 6.863 6.880 |
8.930 8.847 8.793 8.917 8.906 8.886 |
|
SD RSD (%) |
2.398 0.015 0.63 |
3.548 0.010 0.28 |
4.733 0.022 0.46 |
6.810 0.055 0.81 |
8.880 0.051 0.57 |
|
Table 4.5.2 Instrument Response to
Desflurane at High TC
|
× target concn (µg/mL) |
0.5× 774 |
0.75× 1161 |
1× 1547 |
1.5× 2321 |
2× 3095 |
|
area counts (mV-s) |
182.784 182.971 181.180 180.916 182.850 180.309 |
271.647 273.385 271.537 270.495 270.950 270.262 |
361.916 360.981 360.914 358.954 359.275 358.821 |
520.504 523.876 527.304 523.414 524.745 524.994 |
697.847 699.750 693.565 696.571 700.723 694.286 |
|
SD RSD (%) |
181.835 1.168 0.64 |
271.379 1.126 0.41 |
360.144 1.293 0.36 |
524.140 2.233 0.43 |
697.124 2.877 0.41 |
|
The Cochran test for homogeneity:
The critical value of the g-statistic, at the 95% confidence
level, for five variances, each associated with six observations is
0.5065. The g-statistics are 0.3915 and 0.3773 for the low and
high target concentrations respectively. Because the
g-statistic does not exceed this value, the RSDs can be
considered equal and they can be pooled
(RSDP) to give an estimated RSD for the
concentration range studied.
The (RSDP)s are 0.58% and 0.46% for the
low and high target concentration respectively.
4.6 Precision (overall procedure)
The precision of the overall procedure is determined from the
storage data in Section 4.7. The determination of the standard error
of estimate (SEER) for a regression line
plotted through the graphed storage data allows the inclusion of
storage time as one of the factors affecting overall precision. The
SEER is similar to the standard deviation,
except it is a measure of the dispersion of data about a regression
line instead of about a mean. It is determined with the following
equation:
Yobs |
= |
observed % recovery at a given time |
Yest |
= |
estimated % recovery from the regression line at
the same given time |
n |
= |
total number of data points |
k |
= |
2 for linear regression |
k |
= |
3 for quadratic regression |
An additional 5% for pump error (SP) is added to the
SEER by the addition of variances to obtain
the total standard error of the estimate.
The precision at the 95% confidence level is obtained by
multiplying the standard error of estimate (with pump error included)
by 1.96 (the z-statistic from the standard normal distribution
at the 95% confidence level). The 95% confidence intervals are drawn
about their respective regression lines in the storage graphs, as
shown in Figures 4.7.1.1 through 4.7.2.2. The precisions of the
overall procedure are 10.4% and 11.0% for the low and high target
concentration, respectively.
4.7 Storage test
4.7.1 Storage test at the low target concentration
Storage samples were generated by spiking Anasorb 747 tubes with
a toluene solution containing desflurane while pulling air through
the tubes at 0.05 L/min. The relative humidity was approximately 80%
at 22°C. Humid air was then pulled through the tubes for 30 min.
Thirty-six storage samples were prepared. Six samples were analyzed
immediately after generation, fifteen tubes were stored at reduced
temperature (4°C) and the other fifteen were stored in the dark at
ambient temperature (about 22°C). At 2-5 day intervals, three
samples were selected from each of the two sets and analyzed.
Table 4.7.1 Storage Test at the
Low TC
|
time (days) |
ambient storage recovery (%) |
refrigerated storage recovery
(%) |
|
0
5 8 11 13 18 |
90.2 103.9 97.0 98.3 95.8 101.9 96.7 |
102.0 102.3 101.9 102.6 102.1 98.5 99.0 |
101.2 102.1 99.6 104.2 102.5 100.2 100.8 |
90.2 103.9 101.9 105.3 100.1 103.0 102.3 |
102.0 102.3 102.3 106.0 101.9 102.1 100.2 |
101.2 102.1 101.8 107.0 102.5 103.2 100.6 |
|
Figure 4.7.1.1. Ambient storage test at 1
ppm.
Figure 4.7.1.2. Refrigerated storage test at 1
ppm.
4.7.2 Storage test at the high target concentration
Storage samples were generated by sampling from a controlled test
atmosphere containing 1151.2
mg/m3 of desflurane, about
2 times the target concentration. Anasorb 747 tubes were used to
sample for 30 min at 0.05 L/min. The relative humidity was
approximately 80% at 22°C. Thirty-six storage samples were prepared.
Six samples were analyzed immediately after generation, fifteen
tubes were stored at reduced temperature (4°C) and the other fifteen
were stored in the dark at ambient temperature (about 22°C). At 2-5
day intervals, three samples were selected from each of the two sets
and analyzed.
Table 4.7.2 Storage Test at the
High TC
|
time (days) |
ambient storage recovery (%) |
refrigerated storage recovery
(%) |
|
0
3 8 11 14 16 |
87.7 102.3 94.8 106.1 94.0 111.4 94.5 |
102.6 105.3 103.4 108.2 107.5 105.0 104.0 |
96.7 106.0 97.2 101.5 94.9 104.6 95.2 |
87.1 102.3 105.4 90.2 110.9 94.4 104.7 |
102.6 105.3 110.0 107.0 104.8 101.4 107.5 |
96.7 106.0 100.8 94.0 105.5 96.9 113.0 |
|
Figure 4.7.2.1. Ambient storage test at 75
ppm.
Figure 4.7.2.2. Refrigerated storage test at 75
ppm.
4.8 Reproducibility
4.8.1 Reproducibility at low target concentration
Six samples were prepared by injecting microliter quantities of a
toluene solution containing desflurane into Anasorb 747 tubes while
pulling air through the tubes at 0.05 L/min. The relative humidity
was approximately 80% at 22°C. Humid air was then pulled through the
tubes for 60 min. The samples were submitted to an OSHA Salt Lake
Technical Center service branch. The samples were analyzed after
being stored for 10 days at 4°C. Sample results were corrected for
desorption efficiency. No sample result for desflurane had a
deviation greater than the precision of the overall procedure
determined in Section 4.6, which is ±10.4%.
Table 4.8.1 Reproducibility Data
at Low TC
|
sample
|
expected (mg/m3) |
reported (mg/m3) |
recovery (%) |
deviation (%) |
|
1 2 3 4 5 6 |
6.18 6.18 6.18 6.18 6.18 6.18 |
5.77 5.94 5.87 5.87 6.00 6.03 |
93.4 96.1 95.0 95.0 97.1 97.6 |
-6.6 -3.9 -5.0 -5.0 -2.9 -2.4 |
|
4.8.2 Reproducibility at high target concentration
Six samples were prepared by injecting microliter quantities of a
toluene solution containing desflurane into Anasorb 747 tubes while
pulling air through the tubes at 0.05 L/min. The relative humidity
was approximately 80% at 22°C. Humid air was then pulled through the
tubes for 60 min. The samples were submitted to an OSHA Salt Lake
Technical Center service branch. The samples were analyzed after
being stored for 10 days at 4°C. Sample results were corrected for
desorption efficiency. No sample result for desflurane had a
deviation greater than the precision of the overall procedure
determined in Section 4.6, which is ±11.0%.
Table 4.8.2 Reproducibility Data
at High TC
|
sample
|
expected (mg/m3) |
reported (mg/m3) |
recovery (%) |
deviation (%) |
|
1 2 3 4 5 6 |
489.4 489.4 489.4 489.4 489.4 489.4 |
451.0 457.6 453.4 448.4 451.0 453.3 |
92.1 93.5 92.6 91.6 92.1 92.6 |
-7.9 -6.5 -7.4 -8.4 -7.9 -7.4 |
|
4.9 Sampler capacity
The sampling capacity of the front section of an Anasorb 747
sampling tube was tested by sampling from a dynamically generated test
atmosphere of desflurane (1030 mg/m3 or 150
ppm). The samples were collected at 0.05 L/min and the relative
humidity was approximately 80% at 22°C. A GC with a gas sampling valve
was placed in-line behind the 140-mg front test section. The valve was
rotated to measure the amount of desflurane passing through the
sampler at the time of rotation. The 5% breakthrough air volume was
determined to be 3.83 L.
Table 4.9 Capacity of Desflurane
on Anasorb 747
|
first test |
second test |
air volume (L) |
breakthrough (%) |
air volume (L) |
breakthrough (%) |
|
1.60 2.58 3.35 3.87 4.39 |
0 0 0 1.26 37.3 |
1.78 2.72 3.46 3.95 4.56 |
0 0 0 1.56 34.2 |
|
Figure 4.9. Five percent breakthrough air volume for
desflurane on Anasorb 747.
4.10 Desorption efficiency and stability of desorbed samples
4.10.1 Anasorb 747 at low target concentration (TC)
The desorption efficiencies (DE) of desflurane were determined by
liquid-spiking 140-mg portions of Anasorb 747 with amounts
equivalent to 0.05 to 2 times the target concentration. These
samples were stored overnight at ambient temperature and then
desorbed and analyzed. The average desorption efficiency over the
working range of 0.5 to 2 times the target concentration is 101.1%.
Table 4.10.1.1 Desorption
Efficiency of desflurane from Anasorb 747 at Low TC
|
× target concn (µg/sample) |
0.05× 1.029 |
0.1× 2.058 |
0.2× 4.116 |
0.5× 10.29 |
1.0× 20.58 |
2.0× 41.16 |
|
DE (%) |
96.6 96.2 98.2 100.1 95.9 101.0 |
100.7 101.9 104.3 103.3 104.8 104.0 |
100.9 100.0 99.1 100.4 100.7 99.9 |
101.4 102.0 101.4 100.7 99.8 100.7 |
102.3 103.2 102.3 101.1 102.8 103.4 |
99.8 100.8 99.0 99.5 99.2 99.8 |
|
98.0 |
103.2 |
100.2 |
101.0 |
102.5 |
99.7 |
|
The stability of desorbed samples was investigated by reanalyzing
the target concentration samples 20 h after initial analysis. After
the original analysis was performed, three vials were recapped with
new septa while the remaining three retained their punctured septa.
The samples were reanalyzed with fresh standards. The average
percent change was 2.4% for samples that were resealed with new
septa and 0.9% for those that retained their punctured septa.
Table 4.10.1.2 Stability of
Desorbed Samples from Anasorb 747
|
punctured septa replaced |
punctured septa
retained |
initial DE (%) |
DE after one day (%) |
difference |
initial DE (%) |
DE after one day (%) |
difference |
|
102.3 103.2 102.3
102.6 |
100.5 99.4 100.6 averages 100.2 |
-1.8 -3.8 -1.7
-2.4 |
101.1 102.8 103.4
102.4 |
101.7 101.3 101.6 averages 101.5 |
+0.6 -1.5 -1.8
-0.9 |
|
4.10.2 Anasorb 747 at high target concentration (TC)
The desorption efficiencies (DE) of desflurane were determined by
liquid-spiking 140-mg portions of Anasorb 747 with amounts
equivalent to 0.05 to 2 times the target concentration. These
samples were stored overnight at ambient temperature and then
desorbed and analyzed. The average desorption efficiency over the
working range of 0.5 to 2 times the target concentration is 102.9%.
Table 4.10.2.1 Desorption
Efficiency of Desflurane from Anasorb 747 at High TC
|
× target concn (µg/sample) |
0.05× 77.37 |
0.1× 154.7 |
0.2× 309.5 |
0.5× 773.7 |
1.0× 1547 |
2.0× 3095 |
|
DE (%) |
99.3 100.4 100.2 99.7 100.5 100.1 |
99.4 100.0 100.0 100.6 101.0 99.9 |
101.6 101.2 100.4 97.4 103.1 102.0 |
105.4 102.9 104.5 103.8 105.2 103.7 |
104.7 102.6 103.0 102.6 103.6 103.4 |
102.0 101.5 101.2 101.4 100.4 100.3 |
|
100.0 |
100.2 |
101.0 |
104.3 |
103.3 |
101.1 |
|
The stability of desorbed samples was investigated by reanalyzing
the target concentration samples 33 h after initial analysis. After
the original analysis was performed, three vials were recapped with
new septa while the remaining three retained their punctured septa.
The samples were reanalyzed with fresh standards. The average
percent change was 0% for samples that were resealed with new septa
and +0.6% for those that retained their punctured septa.
Table 4.10.2.2 Stability of
Desorbed Samples for Desflurane from Anasorb 747
|
punctured septa replaced |
punctured septa
retained |
initial DE (%) |
DE after one day (%) |
difference |
initial DE (%) |
DE after one day (%) |
difference |
|
104.7 102.6 103.0
103.4 |
103.5 103.4 103.2 (averages) 103.4 |
-1.2 +0.8 +0.2
0 |
102.6 103.6 103.4
103.2 |
104.3 104.4 102.6 (averages) 103.8 |
+1.7 +0.8 -0.8
+0.6 |
|
4.11 Qualitative analysis
The mass spectrum for desflurane was obtained from an HP5988A Mass
Spec interfaced to a Hewlett-Packard 5890 Series II GC.
Figure 4.11. Mass spectrum of
desflurane.
5. References
5.1 Burright, D.D. OSHA Method No. 103; Enflurane, Halothane
and Isoflurane, OSHA Salt Lake Technical Center, unpublished, Salt
Lake City, UT 84165, May 1994.
5.2 NIOSH Criteria for a Recommended Standard: Occupational
Exposure to Waste Anesthetic Gases and Vapors, U.S. Department of
Health and Human Services, Public Health Service, Center for Disease
Control, National Institute for Occupational Safety and Health,
Cincinnati, OH, 1977, DHHS (NIOSH) Publ. 77-140.
5.3 NIOSH Recommendations for Occupational Safety and Health:
Compendium of Policy Documents and Statements, U.S. Department of
Health and Human Services, Public Health Service, Center for Disease
Control, National Institute for Occupational Safety and Health,
Cincinnati, OH, 1992, DHHS (NIOSH) Publ. 92-100.
5.4 Material Safety Data Sheet: Suprane™, Anaquest, Liberty Corner,
NJ, March 1992.
5.5 OSHA Instruction CPL 2-2.60, Exposure Control Plan for
Federal OSHA Personnel with Occupational Exposure to Bloodbourne
Pathogens, March 7, 1994; Occupational Safety and Health
Administration, U.S. Department of Labor, Washington, D.C.
|