ENFLURANE HALOTHANE ISOFLURANE
Method number: |
103 |
|
Matrix: |
Air |
|
Target concentration, TC: |
|
|
enflurane |
halothane |
isoflurane |
TC |
low |
high |
low |
high |
low |
high |
|
ppm |
1 |
75 |
1 |
50 |
1 |
75 |
mg/m3 |
7.5 |
566 |
8 |
403 |
7.5 |
566 |
| |
|
OSHA PEL: ACGIH TLV: |
None 75 ppm (566 mg/m3)
for enflurane 50 ppm (403 mg/m3) for
halothane |
|
Procedure: |
Samples are collected by drawing a known volume of
air through standard size (6-mm o.d., 150/75 mg) Anasorb CMS or
(6-mm o.d., 140/70 mg) Anasorb 747 tubes. Samples are desorbed with
CS2 and analyzed by GC using
a flame-ionization detector (FID). |
|
Air volume and sampling rate: |
12 L at 0.05 L/min |
|
Reliable quantitation limit: |
|
|
enflurane |
halothane |
isoflurane |
Anasorb |
CMS |
747 |
CMS |
747 |
CMS |
747 |
|
ppb |
25.0 |
40.7 |
24.4 |
21.3 |
23.0 |
23.5 |
µg/m3 |
189 |
307 |
197 |
172 |
174 |
177 |
| |
|
Standard error of estimate at the target
concentration: |
|
|
enflurane |
halothane |
isoflurane |
Anasorb |
CMS |
747 |
CMS |
747 |
CMS |
747 |
|
low TC |
0.072 |
0.058 |
0.076 |
0.054 |
0.078 |
0.059 |
high TC |
0.083 |
0.077 |
0.070 |
0.058 |
0.085 |
0.061 |
| |
|
| |
Special requirements: |
Samples collected on Anasorb CMS for halothane should
be stored at reduced temperature following receipt at the laboratory
until analysis. |
|
Status of method: |
Evaluated method. This method has been subjected to
the established evaluation procedures of the Organic Methods
Evaluation Branch. |
|
| |
Date: May 1994 |
Chemist: Donald Burright |
Organic Methods Evaluation Branch OSHA Salt Lake Technical
Center Salt Lake City, UT 84165-0200
1. General Discussion
1.1 Background
1.1.1 History
The objective of this method is to eliminate the need to use two
adsorbent tubes connected in series as specified for enflurane and
halothane in OSHA Method 29 (Ref. 5.1), and to expand the
methodology to include the newer anesthetic gases, isoflurane and
desflurane. (Desflurane will appear as a separate method because it
requires different analytical conditions.) Enflurane, halothane and
isoflurane were each evaluated at two target concentrations because
NIOSH recommended exposure limits (Refs. 5.2 and 5.3) are
considerably lower than the current ACGIH TLVs (Ref. 5.4). For this
reason, the method was evaluated at a lower target concentration of
1 ppm for all three analyses. Currently there are no OSHA PELs for
these substances. Preliminary studies were performed with the
following adsorbents: Anasorb CMS, Anasorb 747, Carbosieve S-III and
activated coconut charcoal. Anasorb CMS and Anasorb 747 were both
good candidates for an improved sampler as neither adsorbent
required two tube in series. Evaluation tests were begun with both
adsorbents in the anticipation that one would dearly surpass the
other in performance. Since this did not occur, both were evaluated
and are presented as sampling options.
ACGIH has recommended a TLV-TWA of 75 ppm for
enflurane and 50 ppm for halothane (Ref. 5.4). The TLV for enflurane
is based on the assumption enflurane is a safer anesthetic gas than
halothane. The TLV for halothane is based on a comparison of
toxicity and TLVs of trichloroethylene and chloroform. (Ref. 5.4)
The ACGIH recommendations are the basis for setting the higher
target concentrations of enflurane and halothane for the evaluation
of this method. A higher target concentration of 75 ppm was set for
isoflurane because it is a geometric isomer of enflurane. NIOSH has
recommended that exposure to these halogenated anesthetic gases
should be controlled with a 60-min ceiling value of 2
ppm (Ref. 5.2). The anesthetic gases are usually administered in
conjunction with nitrous oxide.
1.1.2 Toxic effects (This section is for information only and
should not be taken as the basis of OSHA policy.)
Current scientific evidence obtained from human and animal
studies suggest that chronic exposure to anesthetic gases increase
the risk of both spontaneous abortion and congenital abnormalities
in offspring among female workers and wives of male workers. Risks
of hepatic and renal diseases are also increased among exposed
workers. (Ref 5.2) IARC states there is inadequate evidence for the
carcinogenicity of enflurane, halothane and isoflurane in both
animals and humans (Ref. 5.5).
Enflurane and isoflurane have similar health effects for acute
exposure. An exposure may cause irritation and redness in eyes,
dryness and irritation of skin, and irritation of the mouth and
throat. If inhaled, headaches, dizziness, drowsiness,
unconsciousness, and death can occur. (Refs. 5.6 and 5.7)
Acute exposures of halothane can cause severe irritation to the
eyes, irritation of the skin, reduction of the blood pressure,
dizziness, drowsiness, and unconsciousness. Chronic exposures can
possibly cause cancer. (Ref. 5.8)
1.1.3 Workplace exposure
Enflurane, halothane and isoflurane are the most commonly used
organic anesthetic gases. Occupational exposure may occur whenever
anesthetics are used in operating rooms, dental offices and
veterinary hospitals. The number of people potentially exposed was
estimated to be 215,000 in 1977 (Ref. 5.2). This number is probably
much higher today if the increase in the health care industry since
1977 is considered.
1.1.4 Physical properties and other descriptive information
(Refs. 5.6 - 5.9)
|
enflurane |
halothane |
isoflurane |
|
CAS number: |
13838-16-9 |
151-67-7 |
26675-46-7 |
molecular weight: |
184.49 |
197.39 |
184.49 |
boiling point, °: |
56.5 |
50.2 |
48.5 |
color: |
colorless |
colorless |
colorless |
specific gravity: |
1.52 |
1.872 |
1.50 |
molecular formula: |
C3H2OClF5 |
C2HBrClF3 |
C3H2OClF5 |
vapor pressure, kPa (mmHg): |
25.1(188.6) @22°C |
32.4(243.3) @20°C |
34.9(261.8) @22°C |
odor: |
odorless |
|
mild ethereal |
flash point, °: |
>200 |
none |
>200 |
solubility: |
miscible with organic solvents |
miscible with pet ether and other fat
solvents |
miscible with organic liquids |
synonyms: |
Ethrane;
2-chloro- 1,1,2-trifluoroethyl difluoromethyl ether;
methyl flurether; Efrane; Alyrane |
Fluothrane; 2-bromo-2-chloro- 1,1,1-trifluoro- ethane; Rhodialothan |
Forane;
1-chloro- 2,2,2-trigluoroethyl difluoromethyl ether;
Aerrane; Forene |
structural formulas: |
|
The analyte air concentrations throughout this method are based on the
recommended sampling and analytical parameters. Air concentrations listed
in ppm are referenced to 25° and 101.3 kPa (760 mmHg).
1.2 Limit defining parameters
1.2.1 Detection limit of the analytical procedure
The detection limits of the analytical procedure are 92.8, 87.3
and 44.7 pg for enflurane, halothane and isoflurane respectively.
These are the amounts of each analyte that will give a response that
is significantly different from the background response of a reagent
blank. (Sections 4.1 and 4.2)
1.2.2 Detection limit of the overall procedure
The detection limits of the overall procedure (mass per sample)
are listed below. These are the amounts of each analyte spiked on
the sampler that will give a response that is signfticantly
different from the background response of a sampler blank. (Sections
4.1 and 4.3)
Table 1.2.2 |
Detection Limits of the Overall
Procedure |
|
adsorbent enflurane halothane
isoflurane |
|
Anasorb CMS |
0.679 µg |
0.709 µg |
0.625 µg |
|
7.50 ppb |
7.32 ppb |
6.91 ppb |
|
56.6
µg/m3 |
59.1
µg/m3 |
52.1
µg/m3 |
Anasorb 747 |
1.105 µg |
0.620 µg |
0.639 µg |
|
12.2 ppb |
6.40 ppb |
7.06 ppb |
|
92.1
µg/m3 |
51.7
µg/m3 |
53.3
µg/m3 |
|
1.2.3 Reliable quantitation limit
The reliable quantitation limits (mass per sample) are listed
below. These are the amounts of analyses spiked on a sampler that
will give a signal that is considered the lower limit for precise
quantitative measurements. (Section 4.4)
Table 1.2.3 |
Reliable Quantitation Limits |
|
adsorbent |
enflurane |
halothane |
isonurane |
|
Anasorb CMS |
2.26 µg |
2.36 µg |
2.08 µg |
|
25.0 ppb |
24.4 ppb |
23.0 ppb |
|
188 µg/m3 |
197 µg/m3 |
173
µg/m3 |
Anasorb 747 |
3.68 µg |
2.07 µg |
2.13 µg |
|
40.7 ppb |
21.4 ppb |
23.5 ppb |
|
307 µg/m3 |
172 µg/m3 |
178
µg/m3 |
|
1.2.4 Precision (analytical Procedure)
The precisions of the analytical procedure are measured as the
pooled relative standard deviation over a concentration range
equivalent to the range of 0.5 to 2 times the target concentration.
(Section 4.5)
Table 1.2.4 |
Precisions of the Analytical
Procedure, % |
|
target concn |
enflurane |
halothane |
isoflurane |
1 ppm |
2.77 |
1.39 |
3.50 |
50 ppm |
|
1.37 |
|
75 ppm |
2.27 |
|
2.83 |
|
1.2.5 Precision (overall procedure)
The precisions of the overall procedure at the 95% confidence
level for the ambient temperature 15-18 day storage
tests (at the target concentration) are listed below. This includes
an additional 5% for sampling error. (Section 4.6)
Table 1.2.5.1 |
Precision of the Overall Procedure
on Anasorb CMS, % |
|
target concn |
enflurane |
halothane |
isoflurane |
|
1 ppm |
14.1 |
14.9‡ |
15.3 |
50 ppm |
|
13.7 |
|
75 ppm |
16.3 |
|
16.6 |
|
‡ - refrigerated storage test at
4° |
Table 1.2.5.2 |
Precision of the Overall Procedure
on Anasorb 747, % |
|
target concn |
enflurane |
halothane |
isoflurane |
|
1 ppm |
11.3 |
10.6 |
11.5 |
50 ppm |
|
11.4 |
|
75 ppm |
15.1 |
|
12.0 |
|
1.2.6 Recovery
The recoveries of enflurane, halothane and isoflurane from
samples used in the 15-18 day storage tests remained
above the values listed below when the samples were stored at 22°.
(Section 4.7)
Table 1.2.6.1 |
Recovery from Anasorb CMS, % |
|
target concn |
enflurane |
halothane |
isoflurane |
|
1 ppm |
94.6 |
97.1‡ |
93.7 |
50 ppm |
|
94.5 |
|
75 ppm |
96.2 |
| 97.2 |
|
‡ - refrigerated storage test at
4° |
Table 1.2.6.2 |
Recovery from Anasorb 747, % |
|
target concn |
enflurane |
halothane |
isoflurane |
|
1ppm |
99.6 |
99.8 |
98.7 |
50 ppm |
|
99.3 |
|
75 ppm |
98.0 |
|
97.9 |
|
1.2.7 Reproducibility
Forty-eight samples collected from controlled test atmospheres,
along with a draft copy of this procedure, were submitted for
analysis by one of the OSHA Salt Lake Technical Center's service
branch laboratories. The samples were analyzed after
17-23 days of storage at 4°. No indiividual sample
result deviated from its theoretical value by more than the
precision reported in Section 1.2.5. (Section 4.8)
2. Sampling Procedure
2.1 Apparatus
2.1.1
Samples are collected using a personal sampling pump calibrated,
with the sampling device attached, within ±5% at the recommended
flow rate.
2.1.2 Samples are collected with 7-cm ×
4-mm i.d. × 6-mm o.d. glass sampling tubes
packed with two sections of (150/75 mg) Anasorb CMS or (140/70 mg)
Anasorb 747. The sections are held in place with a glass wool plug
and two urethane foam plugs. For this evaluation, commercially
prepared sampling tubes were purchased from SKC, Inc. (catalog nos.
226-121 and 226-81).
2.2 Reagents
None required.
2.3 Technique
2.3.1 Only properly trained personnel can sample in an operating
room or dental office, this is necessary to be in compliance with
OSHA's Exposure Control Plan for bloodborne pathogens. (Ref. 5.10)
2.3.2 Immediately before sampling, break off the ends of the
sampling tube. All tubes should be from the same lot.
2.3.3 Attach the sampling tube to the sampling pump with
flexible, non-crimping tubing. It is desirable to
utilize sampling tube holders which have a protective cover to
shield the employee from the sharp, jagged end of the sampling tube.
Position the tube so that the sampled air first passes through the
larger section.
2.3.4 Air being sampled should not pass through any hose or
tubing before entering the sampling tube.
2.3.5 To avoid channeling, attach the sampler vertically with the
larger section pointing downward, in the worker's breathing zone.
Position the sampler so it does not impede work performance or
safety.
2.3.6 After sampling for the appropriate time, immediately remove
the sampling tube and seal it with plastic end caps.
2.3.7 In order to prevent occupational exposure to SLTC
personnel, sampling tubes that may become contaminated with blood or
other potentially infectious materials are to be examined prior to
shipping and decontaminated (e.g., wiped off with bleach or other
disinfectant) as necessary. Contaminated items are not to be placed
or stored in areas where food is kept, and decontamination should be
accomplished as soon as possible following the inspection where
contamination occurred. Decontamination is not to take place in any
area where food or drink is consumed. (Ref. 5.10)
2.3.8 Wrap each sample end-to-end with a Form
OSHA-21 seal.
2.3.9 Submit at least one blank sample with each set of samples.
Handle the blank sampling tube
in the same manner as the other samples, except draw no air
through it.
2.3.10 Record sample air volumes (in liters) for each sample,
along with any potential interferences.
2.3.11 Ship any bulk sample(s) in a container separate from the
air samples.
2.3.12 Submit the samples to the laboratory for analysis as soon
as possible after sampling. If delay is unavoidable, store the
samples at reduced temperature.
2.4 Sampler capacity
Sampler capacity is determined by measuring how much air can be
sampled before the analyte breaks through the sampler, i.e., the
sampler capacity is exceeded. Breakthrough is considered to
occur when the effluent from the sampler contains a concentration
of analyte that is 5% of the upstream concentration (5% breakthrough).
Testing for breakthrough was performed by using an FID to monitor the
effluent from sampling tubes containing only either the
150-mg section of Anasorb CMS or 140-mg
section of Anasorb 747. Dynamically generated test atmospheres, which
were about two times the higher target concentration of each analyte,
were used for the capacity tests. The samples were collected at 0.05
L/min and the relative humidity was about 80% at 25°. The 5%
breakthrough air volumes were calculated from the data of duplicate
determinations and are listed below. (Section 4.9)
Table 2.4 |
Sampler Capacity |
|
analyte |
atmospheric |
Anasorb |
Anasorb |
|
concentration |
CMS |
747 |
|
enflurane |
165 ppm |
28.8 L |
14.2 L |
|
(1247
mg/m3) |
halothane |
93 ppm |
15.8 L |
19.9 L |
|
(753 mg/m3) |
isoflurane |
165 ppm |
24.0 L |
17.5 L |
|
(1246 mg/m3) |
|
2.5 Desorption efficiency
2.5.1 The average desorption efficiencies for the analyses from
the sampling media over the range of 0.5 to 2.0 times the target
concentrations (TC) are listed below. (Section 4.10)
Table 2.5.1 |
Desorption Efficiencies, % |
|
analyte |
Anasorb CMS |
Anasorb 747 |
|
low TC |
high TC |
low TC |
high TC |
|
enflurane |
100.3 |
99.8 |
103.7 |
100.5 |
halothane |
99.7 |
99.5 |
99.6 |
99.3 |
isoflurane |
99.4 |
99.2 |
100.8 |
100.2 |
|
2.5.2 The desorption efficiencies at 0.05, 0.1 and 0.2 times the
target concentrations (TC) were found to be very high and are listed
below. (Section 4.10)
Table 2.5.2.1 |
Desorption Efficiencies at 0.05 to
0.2 times Low TC, % |
|
analyte |
Anasorb CMS |
Anasorb 747 |
|
0.05×Tc |
0.1×TC |
0.2×TC |
0.05×TC |
0.1×TC |
0.2×TC |
|
enflurane |
100.2 |
100.4 |
99.5 |
101.3 |
99.0 |
99.0 |
halothane |
99.6 |
100.5 |
99.8 |
84.3 |
92.6 |
94.7 |
isoflurane |
99.3 |
98.4 |
99.9 |
96.8 |
100.0 |
101.1 |
|
Table 2.5.2.2 |
Desorption Efficiencies at 0.05 to
0.2 times High TC, % |
|
analyte |
Anasorb CMS |
Anasorb 747 |
|
0.05×TC |
0.1×TC |
0.2×TC |
0.05×TC |
0.1×TC |
0.2×TC |
|
enflurane |
100.1 |
99.8 |
99.8 |
100.2 |
100.0 |
100.3 |
halothane |
99.3 |
98.9 |
98.5 |
99.6 |
98.6 |
100.4 |
isoflurane |
100.0 |
99.3 |
99.2 |
98.0 |
97.2 |
99.0 |
|
2.5.3 Desorbed samples remain stable for at least 22.5 h.
2.6 Recommended air volume and sampling rate
2.6.1 For long-term samples, collect 12 L at 0.05
L/min.
2.6.2 For short-term samples, collect 0.75 L at 0.05
L/min.
2.6.3 When short-term samples are collected, the air
concentration equivalent to the reliable quantitation limit becomes
larger.
Table 2.6.3 |
Reliable Quantitabon Limits at 0.75
L |
|
adsorbent |
enflurane |
halothane |
isoflurane |
|
Anasorb CMS |
2.26 µg |
2.36µg |
2.08 µg |
|
399 ppb |
390 ppb |
368 ppb |
|
3013
µg/m3 |
3147
µg/m3 |
2773
µg/m3 |
Anasorb 747 |
3.68 µg |
2.07 µg |
2.13 µg |
|
651 ppb |
342 ppb |
377 ppb |
|
4707
µg/m3 |
2760
µg/m3 |
2840
µg/m3 |
|
2.7 Interferences (sampling)
2.7.1 It is not known if any compounds will severely interfere
with the collection of enflurane, halothane and isoflurane on
Anasorb CMS or Anasorb 747. In general, the presence of other
contaminant vapors in the air will reduce the capacity of Anasorb
CMS or Anasorb 747 to collect the three analyses.
2.7.2 Nitrous oxide was tested as an interferant to the
collection of halothane and it does not interfere. (Section 4.12)
2.7.3 Suspected interferences should be reported to the
laboratory with submitted samples.
2.8 Safety precautions (sampling)
2.8.1 The sampling equipment should be attached to the worker in
such a manner that it will not interfere with work performance or
safety.
2.8.2 All safety practices that apply to the work area being
sampled should be followed.
2.8.3 Protective eyewear should be worn when breaking the ends of
the glass sampling tubes.
3. Analybcal Procedure
3.1 Apparatus
3.1.1 Gas chromatograph equipped with an FID. For this
evaluation, a Hewlett-Packard 5890A Gas Chromatograph
equipped with a 7673A Automatic Sampler was used. A Forma Scientific
Model 2006 refrigerated circulator was used to cool the sample tray
of the HP 7673A to 10° to minimize evaporation.
3.1.2 A GC column capable of separating the analyte of interest
from the desorption solvent, internal standard and any
interferences. A 60-m × 0.32-mm i.d. fused silica
Stabilwax-D8419 column with a 1-µm df (Restek
Corp., Bellefonte, PA) was used in the evaluation.
3.1.3 An electronic integrator or some other suitable means of
measuring peak areas. A Waters 860 Networking Computer System was
used in this evaluation.
3.1.4 Two-milliliter vials with
polytetrafluoroethylene-lined caps.
3.1.5 A dispenser capable of delivering 1.0 mL of desorbing
solvent to prepare standards and samples. If a dispenser is not
available, a 1.0-mL volumetric pipes may be used.
3.2 Reagents
Enflurane, USP. The enflurane used in this evaluation was
manufactured by Anequest (Madison, WI), and purchased from a local
hospital.
3.2.2 Halothane, reagent grade or better. The halothane used in
this evaluation was purchased from Aldrich Chemical (Milwaukee, WI).
3.2.3 Isoflurane, USP. The isoflurane used in this evaluation was
manufactured by Anequest (Madison, WI), and purchased from a local
hospital.
3.2.4 Carbon disulfide
(CS2), reagent grade or
better. The CS2 used in this
evaluation was purchased from JT Baker Chemical (Phillipsburg, NJ).
3.2.5 A suitable internal standard, reagent grade. The
n-decane used in this evaluation was purchased from ICN
Pharmaceuticals, Inc. (Plainview, NY).
3.2.6 Desorption solvent. The desorption solvent contains 500
µL of n-decane per 1 L of
CS2.
3.2.7 GC grade nitrogen, air, and hydrogen.
3.2.8 Toluene, chromatographic grade or better. The toluene used
in this evaluation was Optima Grade and was purchased from Fisher
Scientific (Fair Lawn, NJ).
3.3 Standard preparation
3.3.1 Prepare concentrated stock standard of enflurane,
halothane and isoflurane in toluene. Prepare working analytical
standards by injecting microliter amounts of concentrated stock
standards into 2-mL vials containing 1.0 mL of
desorption solvent delivered from the same dispenser used to desorb
samples. For example, to prepare a target level standard of
isoflurane, inject 10 µL of a stock solution containing 672
mg/mL of isoflurane in toluene into 1 mL of desorption solvent.
3.3.2 Bracket sample concentrations with working standard
concentrations. If samples fall outside the concentration range of
prepared standards, prepare and analyze additional standards or
dilute the sample.
3.4 Sample preparation
3.4.1 Remove the plastic end caps from the sample tube and
carefully transfer each section of the adsorbent to separate
2-mL vials. Discard the glass tube, urethane foam plugs
and glass wool plug.
3.4.2 Add 1.0 mL of desorption solvent to each vial using the
same dispenser as used for preparation of standards.
3.4.3 Immediately seal the vials with
polytetrafluoroethylene-lined caps.
3.4.4 Shake the vials vigorously several times during the next 30
min.
3.5 Analysis
3.5.1 Analytical conditions
GC conditions |
zone temperatures: |
60° (column) 250° (injector) 300°
(detector) |
run time: |
15 min |
column gas flow: |
1.2 mL/min (hydrogen) |
septum purge: |
1.5 mL/min (hydrogen) |
injector size: |
1.0 µL (11.3:1 split) |
column: |
60-m × 0.32-mm i.d. capillary
Stabilwax-DB (1.0-µm df) |
retention times: |
5.50 min (isoflurane) 5.97 min
(halothane) 6.51 min (enflurane) 8.34 min
(n-decane) |
FID conditions |
hydrogen flow: |
34 mL/min |
air flow: |
450 mL/min |
makeup flow: |
33 mL/min (nitrogen) |
Figure 3.5.1.1. Chromatogram obtained at the high TC with the
recommended condibons. Peak identification: (1) carbon disutfide,
(2) isoflurane, (3) halothane, (4) enflurane, (5) benzene -
contaminant in CS2, (6)
n-decane, (7) toluene - from spiking solution.
Figure 3.5.1.2. Chromatogram obtained at the low TC with the
recommended conditions. Peak identification: (1) carbon disulfide,
(2) isoflurane, (3) halothane, (4) enflurane, (5) benzene -
contaminant in CS2, (6)
n-decane, (7) toluene - from spiking solution.
3.5.2 An internal standard (ISTD) calibration method is used. A
calibration curve can be constructed by plotting micrograms of
analyte per sample versus ISTD-corrected response of
standard injections. Bracket the samples with freshly prepared
analytical standards over a range of concentrations.
Figure 3.5.2.1. Calibrabon curve of enflurane at low TC made from
data of Table 4.5.1.
Figure 3.5.2.2. Calibrabon curve of enflurane at high TC made
from data of Table 4.5.2.
Figure 3.5.2.3. Calibrabon curve of halothane at low TC made from
data of Table 4.5.3.
Figure 3.5.2.4. Calibration curve of halothane at high TC made
from data of Table 4.5.4.
Figure 3.5.2.5. Calibration curve of isoflurane at low TC made
from data of Table 4.5.5.
Figure 3.5.2.6. Calibration curve of isoflurane at high TC made
from data of Table 4.5.6.
3.6 Interferences (analytical)
3.6.1 Any compound that produces an FID response and has a
similar retention time as the analyses or internal standard is a
potential interference. If any potential interferences were
reported, they should be considered before the samples are desorbed.
3.6.2 Generally, chromatographic conditions can be altered to
separate an interference from the analyte.
3.6.3 When necessary, the identity or purity of an analyte peak
may be confirmed with additional analytical data. (Section 4.11)
3.7 Calculations
The amount of analyte per sampler is obtained from the appropriate
calibration curve in terms of micrograms per sample, uncorrected for
desorption efficiency. The back (70-75 mg) section is analyzed
primarily to determine if there was any breakthrough from the front
(140-150 mg) section during sampling. If a significant amount of
analyte is found on the back section (e.g., greater than 25% of the
amount found on the front section), this fact should be reported with
the sample results. If any analyte is found on the back section, it is
added to the amount on the front section. This amount is then
corrected by subtracting the total amount (if any) found on the blank.
The air concentration is calculated using the following formulae.
mg/m3 = |
micrograms of analyte per sample
liters of air sampled × desorption efficiency
|
ppm = |
24.46 × mg/m3
molecular weight of analyte
|
where |
24.46 is the molar volume at 25° and 101.3 kPa (760
mmHg) |
|
184.49 = molecular weight of enflurane and isoflurane |
|
197.39 = molecular weight of halothane |
3.8 Safety precautions (analytical)
3.8.1 Adhere to the rules set down in your Chemical Hygiene
Plan.
3.8.2 Avoid skin contact and inhalation of all chemicals.
3.8.3 Wear safety glasses, gloves and a lab coat at all ffmes
while in the laboratory areas.
4. Backup Data
4.1 Determination of detection limits
Detection limits, in general, are defined as the amount (or
concentration) of analyte that gives a response
(YDL) that is significantly
different [three standard deviations
(SDBR)] from the background
response (YBR).
YDL - YBR
= 3(SDBR)
The measurement of YBR and
SDBR in chromatographic methods
is typically inconvenient and difficult because
YBR is usually extremely low.
Estimates of these parameters can be made with data obtained from the
analysis of a series of analytical standards or samples whose
responses are in the vicinity of the background response. The
regression curve obtained for a plot of instrument response versus
concentration of analyte will usually be linear. Assuming
SDBR and the precision of the
data about curve are similar, the standard error of estimate (SEE) for
the regression curve can be substituted for
SDBR in the above equation. The
following calculations derive a formula for DL:
Yobs
= |
observed response |
Yest
= |
estimated response from regression curve |
n = |
total number of data points |
k = |
2 for linear regression curve |
At point YDL on the regression curve
YDL - A(DL) +
YBR |
A = analytical sensitivity
(slope) |
therefore
Substituting 3(SEE) + YBR for
YDL gives
4.2 Detection limit of the analytical procedure (DW)
The DW is measured as the mass of analyte actually introduced into
the chromatographic columns. Ten analytical standards were prepared in
equal descending increments with the highest standard containing
10.02, 10.68 and 9.89 µg/mL of enflurane, halothane and
isoflurane respecffvely. This is the concentration that would produce
a peak approximately 10 times the baseline noise of a reagent blank
near the elution time of the analyte. These standards, and the reagent
blank, were analyzed with the recommended analytical parameters
(1-µL injection with a 11.3:1 split), and the data obtained
were used to determine the required parameters (A and SEE) for the
calculation of the DLAP.
Table 4.2.1 |
DLAP Data for Enflurane |
A = 3.81 SEE = 117.9 |
|
concentration |
mass on column |
area counts |
(µg/mL) |
(pg) |
(µV-s) |
|
0 |
0 |
0 |
0.956 |
84.4 |
455 |
1.90 |
168 |
840 |
2.84 |
251 |
1002 |
3.77 |
333 |
1145 |
5.60 |
494 |
2090 |
6.50 |
574 |
2105 |
7.39 |
653 |
2610 |
8.28 |
731 |
2869 |
9.15 |
808 |
3041 |
10.02 |
884 |
3519 |
|
Figure 4.2.1 Plot of the data from Table 4.2.1 to determine the
DLAP of enflurane, DLAP = 92.8 pg.
Table 4.2.2 |
DLAP Data for Halothane |
A = 4.08 SEE = 118.7 |
|
concentration |
mass on column |
area counts |
(µg/mL) |
(pg) |
(µV-s) |
|
0 |
0 |
0 |
1.02 |
90 |
413 |
2.03 |
179 |
916 |
3.03 |
267 |
1219 |
5.97 |
527 |
2366 |
6.93 |
612 |
2501 |
7.88 |
696 |
2907 |
8.82 |
779 |
3219 |
9.76 |
861 |
3416 |
10.68 |
943 |
4035 |
|
Figure 4.2.2. Plot of the data from Table 4.2.2 to determine the DW
of halothane, DLAP = 87.3 pg.
Table 4.2.3 |
DLAP Data for Isoflurane |
A = 2.34 SEE = 34.90 |
|
concentration |
mass on column |
area counts |
(µg/mL) |
(pg) |
(µV-s) |
|
0 |
0 |
0 |
0.944 |
83 |
288 |
1.88 |
166 |
421 |
2.80 |
247 |
627 |
3.72 |
328 |
794 |
5.53 |
488 |
1230 |
6.41 |
566 |
1307 |
7.30 |
644 |
1537 |
8.17 |
721 |
1711 |
9.03 |
797 |
1921 |
9.89 |
873 |
2090 |
|
Figure 4.2.3 Plot of the data from Table 4.2.3 to determine the
DLAP of isoflurane, DLAP = 44.7 pg.
4.3 Detection limit of the overall procedure (DLOP)
The DLOP is measured as mass per sample and expressed as equivalent
air concentration, based on the recommended sampling parameters. Ten
samplers were spiked with equal descending increments of analyte, such
that the highest sampler loading was 9.15, 9.76 and 9.03
µg/sample of enflurane, halothane and isoflurane respectively.
This is the amount, when spiked on a sampler, that would produce a
peak approximately 10 times the baseline noise for a sample blank.
These spiked samplers, plus a sample blank, were analyzed with the
recommended analytical parameters, and the data obtained used to
calculate the required parameters (A and SEE) for the calculation of
the DLOP.
Table 4.3.1 |
DLOP Data for Enflurane |
|
mass per |
area counts on |
area counts on |
sample |
Anasorb CMS |
Anasorb 747 |
(µg) |
(µV-s) |
(µV-s) |
|
0.956 |
441 |
292 |
1.90 |
651 |
735 |
2.84 |
1042 |
1109 |
3.77 |
1423 |
1449 |
4.69 |
1618 |
1690 |
5.60 |
1858 |
1786 |
6.50 |
2084 |
2430 |
7.39 |
2378 |
2393 |
8.28 |
2635 |
2749 |
9.15 |
3035 |
3238 |
|
Figure 4.3.1.1. Plot of the data to determine the DLOP of enflurane
on Anasorb CMS, (SEE = 71.55).
Figure 4.3.1.2. Plot of the data to determine the DLOP of enflurane
on Anasorb 747, (SEE = 124.5).
Table 4.3.2 |
DLOP Data for Halothane |
|
mass per |
area counts on |
area counts on |
sample |
Anasorb CMS |
Anasorb 747 |
(µg) |
(µV-s) |
(µV-s) |
|
0 |
0 |
0 |
1.02 |
512 |
382 |
2.03 |
761 |
812 |
3.03 |
1224 |
1263 |
4.02 |
1554 |
1670 |
5.00 |
1870 |
1999 |
5.97 |
2192 |
2481 |
6.93 |
2526 |
2795 |
7.88 |
2721 |
3059 |
8.82 |
3288 |
3319 |
9.76 |
3677 |
3761 |
|
Figure 4.3.2.1. Plot of the data to determine the DLOP of halothane
on Anasorb CMS, (SEE = 85.02).
Figure 4.3.2.2 Plot of the data to determine the DLOP of halothane
on Anasorb 747, (SEE = 79.37).
Table 4.3.3 |
DLOP Data for Isodurane |
|
mass per |
area counts on |
area counts on |
sample |
Anasorb CMS |
Anasorb 747 |
(µg) |
(µV-s) |
(µV-s) |
|
0 |
0 |
0 |
0.944 |
274 |
245 |
1.88 |
444 |
442 |
2.80 |
654 |
630 |
3.72 |
845 |
813 |
4.63 |
963 |
1041 |
5.53 |
1195 |
1088 |
6.41 |
1349 |
1326 |
7.30 |
1564 |
1488 |
8.17 |
1622 |
1615 |
9.03 |
1877 |
1825 |
|
Figure 4.3.3.1. Plot of the data to determine the DLOP of
isoflurane on Anasorb CMS, (SEE = 41.65).
Figure 4.3.3.2. Plot of the data to determine the DLOP of
isoflurane on Anasorb 747, (SEE = 41.51).
4.4 Reliable quantitation limit (RQL)
The RQL is considered the lower limit for precise quantitative
measurements. It is determined from the regression line parameters
obtained for the calculations of the DLOP (Section 4.3), providing at
least 75% of the analyte is recovered. The RQL is defined as the
amount of analyte that gives a response
(YRQL) such that
YRQL - YBR
= 10(SDBR)
therefore
Figure 4.4.1. Chromatogram of the RQL for all three analyses on
Anasorb CMS.
Figure 4.4.2. Chromatogram of the RQL for halothane and isoflurane
on Anasorb 747.
Figure 4.4.3. Chromatogram of the RQL for enflurane on Anasorb 747.
Table 4.4 |
Reliable Quantitation Limits |
|
adsorbent |
enflurane |
halothane |
isoflurane |
|
Anasorb CMS |
2.26 µg |
2.36 µg |
9 2.08 µg |
|
225 ppb |
219 ppb |
207 ppb |
|
1695 µg/m3 |
1770 µg/m3 |
1560µg/m3 |
|
91.7% |
95.9% |
104.1% |
Anasorb 747 |
3.68 µg |
2.07 µg |
2.13 µg |
|
366 ppb |
192 ppb |
212 ppb |
|
2760 µg/m3 |
1553 µg/m3 |
1598 µg/m3 |
|
109.2% |
102.9% |
103.6% |
|
The RQL for each analyte was calculated and listed above along with
the recovery of the analyte peak near the RQL.
4.5 Precision (analytical method)
The precision of the analytical procedure is measured as the pooled
relative standard deviation
(RSDP). Relative standard
deviations are determined from six replicate injections of analyte
standards at 0.5, 0.75, 1, 1.5 and 2 times the target concentration.
After assuring that the RSDs satisfy the Cochran test for homogeneity
at the 95% confidence level,
RSDP was calculated.
Table 4.5.1 |
Instrument Response to Enflurane at
Low TC |
|
× target concn |
0.5× |
0.75× |
1× |
1.5× |
2× |
(µg/mL) |
45.6 |
68.4 |
91.2 |
136.8 |
182.4 |
|
area counts |
5802 |
8044 |
11502 |
15685 |
21003 |
(µV-s) |
5762 |
7971 |
11748 |
15514 |
20685 |
|
5592 |
8178 |
11513 |
15995 |
19527 |
|
5604 |
8024 |
10887 |
16911 |
20585 |
|
5898 |
8173 |
11441 |
15579 |
19114 |
|
5902 |
8146 |
11336 |
15502 |
19871 |
|
5760 |
8089 |
11405 |
15864 |
20131 |
SD |
136.8 |
87.6 |
287.4 |
544.1 |
740.3 |
RSD (%) |
2.37 |
1.08 |
2.52 |
3.42 |
3.67 |
|
Table 4.5.2 |
Instrument Response to Enflurane at
High TC |
|
× target concn |
0.5× |
0.75× |
1× |
1.5× |
2× |
(µg/mL) |
3405 |
5108 |
6810 |
10215 |
13620 |
|
area counts |
313768 |
442725 |
588885 |
909618 |
1266303 |
(µV-s) |
295385 |
449118 |
614061 |
884040 |
1231585 |
|
306119 |
444388 |
600184 |
888369 |
1239847 |
|
302927 |
452003 |
605174 |
873998 |
1195472 |
|
302101 |
463954 |
583097 |
872158 |
1232590 |
|
315239 |
450679 |
623083 |
934700 |
1220325 |
|
305923 |
450478 |
602414 |
893814 |
1231020 |
SD |
7523.3 |
7523.5 |
15044.1 |
24117.5 |
23253.5 |
RSD (%) |
2.45 |
1.67 |
2.48 |
2.69 |
1.88 |
|
Table 4.5.3 |
Instrument Response to Halothane at
Low TC |
|
× target concn |
0.5× |
0.75× |
1× |
1.5× |
2× |
(µg/mL) |
48.62 |
72.93 |
97.24 |
145.9 |
194.5 |
|
area counts |
4652 |
6892 |
9563 |
14067 |
19459 |
(µV-s) |
4572 |
6936 |
9438 |
14324 |
18888 |
|
4781 |
6802 |
9773 |
14181 |
18732 |
|
4625 |
6931 |
9501 |
14180 |
19216 |
|
4630 |
6802 |
9694 |
14534 |
18964 |
|
4609 |
6960 |
9404 |
14271 |
19499 |
|
4645 |
6887 |
9562 |
14260 |
19126 |
SD |
71.8 |
69.5 |
145.6 |
160.7 |
315.0 |
RSD (%) |
1.54 |
1.00 |
1.52 |
1.12 |
1.64 |
|
Table 4.5.4 |
Instrument Response to Halothane at
High TC |
|
× target concn |
0.5× |
0.75× |
1× |
1.5× |
2× |
(µg/mL) |
2431 |
3646 |
4862 |
7293 |
9724 |
|
area counts |
214963 |
31920 |
427008 |
631533 |
876733 |
(µV-s) |
211987 |
326882 |
426550 |
625176 |
856581 |
|
218236 |
322866 |
430225 |
620733 |
870877 |
|
215433 |
325338 |
429946 |
632545 |
849740 |
|
214585 |
329467 |
436427 |
647793 |
860103 |
|
207597 |
319337 |
433215 |
623791 |
847012 |
|
213800 |
323849 |
430562 |
630262 |
860174 |
SD |
3635.3 |
4144.4 |
3759.2 |
9723.2 |
11694.2 |
RSD (%) |
1.70 |
1.27 |
0.87 |
1.54 |
1.35 |
|
Table 4.5.5 |
Instrument Response to Isoflurane at
Low TC |
|
× target concn |
0.5× |
0.75× |
1× |
1.5× |
2× |
(µg/mL) |
45.0 |
67.5 |
90.0 |
135.0 |
180.0 |
|
area counts |
3164 |
4584 |
6793 |
9544 |
14068 |
(µV-s) |
3083 |
4690 |
6928 |
9750 |
13672 |
|
3239 |
4700 |
6791 |
10564 |
13408 |
|
2983 |
4510 |
6209 |
10076 |
12670 |
|
3120 |
4546 |
6809 |
9707 |
13257 |
|
3147 |
4757 |
6621 |
9820 |
12437 |
|
3123 |
4631 |
6692 |
9910 |
13252 |
SD |
85.9 |
98.2 |
256.0 |
364.5 |
611.5 |
RSD (%) |
2.75 |
2.11 |
3.82 |
3.67 |
4.61 |
|
Table 4.5.6 |
Instrument Response to Isoflurane at
High TC |
|
× target concn |
0.5× |
0.75× |
1× |
1.5× |
2× |
(µg/mL) |
3360 |
5040 |
6720 |
10080 |
13440 |
|
area counts |
192010 |
285648 |
377707 |
560192 |
831803 |
(µV-s) |
198666 |
302263 |
389365 |
536323 |
808032 |
|
196337 |
286004 |
390013 |
576197 |
832183 |
|
197101 |
283263 |
392882 |
581411 |
807540 |
|
184540 |
294184 |
416447 |
572522 |
811381 |
|
202374 |
287224 |
410104 |
564809 |
803188 |
|
95171 |
289764 |
396086 |
565242 |
815688 |
SD |
6199.9 |
7149.2 |
14430.4 |
16102.5 |
12896.4 |
RSD (%) |
3.17 |
2.46 |
3.64 |
2.84 |
1.58 |
|
The Cochran test for homogeneity:
The critical value of the g-statistic, at the 95% confidence
level, for five variances, each associated with six observations is
0.5065. Because the g-statistic does not exceed this value, the
RSDs can be considered equal and they can be pooled
(RSDP) to give an estimated RSD for the
concentration range studied.
Table 4.5.7 |
Cochran Test Results and Pooled
Relative Standard Deviations |
|
|
enflurane |
halothane |
isoflurane |
|
TC |
low |
high |
low |
high |
low |
high |
g |
0.3515 |
0.2808 |
0.2785 |
0.3058 |
0.3465 |
0.3317 |
RSDP % |
2.77 |
2.27 |
1.39 |
1.37 |
3.50 |
2.83 |
|
4.6 Precision (overall procedure)
The precision of the overall procedure is determined from the
storage data in Section 4.7. The determination of the standard error
of estmate (SEER) for a
regression line ptotted through the graphed storage data allows the
inclusion of storage time as one of the factors affecting overall
precision. The SEER is similar
to the standard deviation, except it is a measure of the dispersion of
data about a regression line instead of about a mean. It is determined
with the following equation:
Yobs |
= |
observed % recovery at a given time |
Yest |
= |
estimated % recovery from the regression line at
the same given time |
n |
= |
total number of data points |
k |
= |
2 for linear regression |
k |
= |
3 for quadratic regression |
An additional 5% for pump error (SP) is added to the
SEER by the addition of
variances to obtain the total standard error of the estimate.
The precision at the 95% confidence level is obtained by
multiplying the standard error of estimate (with pump error included)
by 1.96 (the z-statistic from the standard normal
distribution at the 95% confidence level). The 95% confidence
intervals are drawn about their respective regression lines in the
storage graphs, as shown in figures 4.7.1.1.1 through 4.7.2.6.2. The
precisions of the overall procedure and the assodated figures are
listed below.
Table 4.6.1 |
Precision of the Overall Procedure on
Anasorb CMS, % |
|
target concn |
enflurane |
halothane |
isoflurane |
|
1 ppm |
14.1 |
14.9 |
15.3 |
|
Fig. 4.7.2.1.1 |
Fig. 4.7.2.3.2 |
Fig. 4.7.2.5.1 |
50 ppm |
|
13.7 |
|
|
|
Fig 4.7.1.3.1 |
|
75 ppm |
16.3 |
|
16.6 |
|
Fig. 4.7.1.1.1 |
|
Fig. 4.7.1.5.1 |
|
Table 4.6.2 |
Precision of the Overall Procedure on
Anasorb 747, % |
|
target concn |
enflurane |
halothane |
isoflurane |
|
1 ppm |
11.3 |
10.6 |
11.5 |
|
Fig. 4.7.2.2.1 |
Fig. 4.7.2.4.1 |
Fig. 4.7.2.6.1 |
50 ppm |
|
11.4 |
|
|
|
Fig. 4.7.1.4.1 |
|
75 ppm |
15.1 |
|
12.0 |
|
Fig. 4.7.1.2.1 |
|
Fig. 4.7.1.6.1 |
|
4.7 Storage test
4.7.1 Analyte storage at high target concentration
4.7.1.1 Storage samples were generated by sampling from a
controlled test atmosphere containing 2120
mg/m3 of enflurane, about 3.7
times the 75-ppm target concentration. Anasorb CMS
tubes were used to sample for 60 min at 0.05 L/min, the relative
humidity was about 80% at 22°. Thirty-six storage
samples were prepared. Six samples were analyzed immediately after
generation, fifteen tubes were stored at reduced temperature (4°)
and the other fifteen were stored in a closed drawer at ambient
temperature (about 22°). At 2-4 day intervals, three
samples were selected from each of the two sets and analyzed.
Table 4.7.1.1 |
Storage Test for Enflurane on
Anasorb CMS |
|
time |
ambient storage |
refrigerated storage |
(days) |
recovery (%) |
recovery (%) |
|
0 |
90.0 |
101.5 |
98.0 |
90.0 |
101.5 |
98.0 |
|
101.9 |
103.4 |
95.3 |
101.9 |
103.5 |
95.3 |
3 |
101.0 |
101.6 |
94.6 |
90.9 |
88.3 |
95.6 |
6 |
102.1 |
105.1 |
107.3 |
111.1 |
103.6 |
98.7 |
9 |
101.8 |
104.1 |
95.4 |
90.7 |
97.0 |
100.5 |
13 |
97.8 |
88.4 |
91.9 |
98.3 |
112.7 |
103.3 |
15 |
91.1 |
102.7 |
90.7 |
92.0 |
100.2 |
102.3 |
|
Figure 4.7.1.1.1. Ambient storage test for enflurane on Anasorb
CMS.
Figure 4.7.1.1.2. Refrigerated storage test for enflurane on
Anasorb CMS.
4.7.1.2 Storage samples were generated by sampling from a
controlled test atmosphere containing 2118
mg/m3 of enflurane, about 3.7 times the
75-ppm target concentration. Anasorb 747 tubes were
used to sample for 60 min at 0.05 L/min, the relative humidity was
about 80% at 22°. Thirty-six storage samples were
prepared. Six samples were analyzed immediately after generation,
fifteen tubes were stored at reduced temperature (4°) and the
other fifteen were stored in a closed drawer at ambient
temperature (about 22°). At 2-4 day intervals, three
samples were selected from each of the two sets and analyzed.
Table 4.7.1.2 |
Storage Test for Enflurane on
Anasorb 747 |
|
time |
ambient storage |
refrigerated storage |
(days) |
recovery (%) |
recovery (%) |
|
0 |
96.3 |
97.4 |
106.9 |
96.3 |
97.4 |
106.9 |
|
102.1 |
101.7 |
102.5 |
102.1 |
101.7 |
102.5 |
2 |
92.4 |
84.7 |
100.8 |
96.9 |
101.4 |
98.2 |
6 |
92.0 |
104.1 |
97.6 |
95.1 |
96.9 |
99.6 |
9 |
102.2 |
96.7 |
106.3 |
93.1 |
104.2 |
99.0 |
12 |
97.2 |
112.2 |
105.2 |
102.1 |
103.0 |
105.6 |
16 |
102.0 |
102.6 |
105.7 |
99.1 |
112.2 |
106.9 |
|
Figure 4.7.1.2.1 Ambient storage test for enflurane on Anasorb
747.
Figure 4.7.1.2.2. Refrigerated storage test for enflurane on
Anasorb 747.
4.7.1.3 Storage samples were generated by sampling from a
controlled test atmosphere containing 2146
mg/m3 of halothane, about 5.3 times the
50-ppm target concentration. Anasorb CMS tubes were
used to sample for 60 min at 0.05 L/min, the relative humidity was
about 80% at 22°. Thirty-six storage samples were
prepared. Six samples were analyzed immediately after generation,
fifteen tubes were stared at reduced temperature (4°) and the
other fifteen were stored in a closed drawer at ambient
temperature (about 22°). At 24 day intervals, three samples were
selected from each of the two sets and analyzed.
Table 4.7.1.3 |
Storage Test for Halothane on
Anasorb CMS |
|
time |
ambient storage |
refrigerated storage |
(days) |
recovery (%) |
recovery (%) |
|
0 |
103.3 |
104.3 |
90.8 |
103.3 |
104.3 |
90.8 |
|
98.3 |
95.5 |
98.7 |
98.3 |
95.5 |
98.7 |
4 |
101.5 |
95.4 |
96.9 |
106.6 |
91.4 |
99.6 |
6 |
103.2 |
88.7 |
96.0 |
99.1 |
95.8 |
98.8 |
8 |
103.5 |
101.1 |
97.7 |
108.3 |
92.9 |
100.4 |
12 |
106.0 |
91.0 |
94.5 |
108.7 |
102.0 |
102.3 |
15 |
96.06 |
91.1 |
93.6 |
103.6 |
90.6 |
101.1 |
|
Figure 4.7.1.3.1. Ambient storage test for halothane on Anasorb
CMS.
Figure 4.7.1.3.2. Refrigerated storage test for halothane on
Anasorb CMS
4.7.1.4 Storage samples were generated by sampling from a
controlled test atmosphere containing 2305
mg/m3 of halothane, about 5.7 times the
50-ppm target concentration. Anasorb 747 tubes were
used to sample for 60 min at 0.05 L/min, the relative humidity was
about 80% at 22°. Thirty-six storage samples were
prepared. Six samples were analyzed immediately after generation,
fifteen tubes were stored at reduced temperature (4°) and the
other fifteen were stored in a closed drawer at ambient
temperature (about 22°). At 3-4 day intervals, three
samples were selected from each of the two sets and analyzed.
Table 4.7.1.4 |
Storage Test for Halothane on
Anasorb 747 |
|
time |
ambient storage |
refrigerated storage |
(days) |
recovery (%) |
recovery(%) |
|
0 |
99.1 |
100.0 |
103.9 |
99.1 |
100.0 |
103.9 |
|
100.5 |
99.7 |
96.9 |
100.5 |
99.7 |
96.9 |
3 |
99.4 |
99.2 |
96.2 |
98.6 |
104.2 |
98.9 |
6 |
98.8 |
104.2 |
98.8 |
101.5 |
100.1 |
97.3 |
10 |
99.1 |
103.5 |
94.9 |
100.9 |
105.0 |
101.3 |
13 |
95.8 |
104.5 |
98.7 |
101.3 |
95.7 |
95.5 |
17 |
103.8 |
104.8 |
99.6 |
102.7 |
105.9 |
99.2 |
|
Figure 4.7.1.4.1. Ambient storage test for halothane on Anasorb
747.
Figure 4.7.1.4.2. Refrigerated storage test for halothane on
Anasorb 747.
4.7.1.5 Storage samples were generated by sampling from a
controlled test atmosphere containing 3050
mg/m3 of isoflurane, about 5.4 times the
75-ppm target concentration. Anasorb CMS tubes were
used to sample for 60 min at 0.05 L/min the relative humidity was
about 80% at 22°. Thirty-six storage samples were
prepared. Six samples were analyzed immediately after generation,
fifteen tubes were stored at reduced temperature (4°) and the
other fifteen were stored in a closed drawer at ambient
temperature (about 22°). At 2-4 day intervals, three
samples were selected from each of the two sets and analyzed.
Table 4.7.1.5 |
Storage Test for Isoflurane on
Anasorb CMS |
|
time |
ambient storage |
refrigerated storage |
(days) |
recovery (%) |
recovery (%) |
|
0 |
99.7 |
103.6 |
93.9 |
99.7 |
103.6 |
93.6 |
|
99.0 |
85.4 |
97.8 |
99.0 |
85.4 |
97.8 |
4 |
111.9 |
104.8 |
97.5 |
106.7 |
91.7 |
92.1 |
7 |
99.2 |
94.2 |
90.2 |
108.8 |
103.0 |
100.6 |
11 |
113.1 |
102.9 |
102.3 |
106.5 |
99.1 |
91.4 |
13 |
103.9 |
96.4 |
92.0 |
108.0 |
105.3 |
103.6 |
15 |
108.0 |
105.1 |
106.1 |
105.0 |
100.3 |
96.8 |
|
Figure 4.7.1.5.1. Ambient storage test for isoflurane on
Anasorb CMS.
Figure 4.7.1.5.2. Refrigerated storage test for isoflurane on
Anasorb CMS.
4.7.1.6 Storage samples were generated by sampling from a
controlled test atmosphere containing 2992
mg/m3 of isoflurane, about 5.3 times the
75-ppm target concentration. Anasorb 747 tubes were
used to sample for 60 min at 0.05 L/min, the relative humidity was
about 80% at 22°. Thirty-six storage samples were
prepared. Six samples were analyzed immediately after generation,
fifteen tubes were stored at reduced temperature (4°) and the
other fifteen were stored in a dosed drawer at ambient temperature
(about 22°). At 3-4 day intervals, three samples were
selected from each of the two sets and analyzed.
Table 4.7.1.6 |
Storage Test for Isoflurane on
Anasorb 747 |
|
time |
ambient storage |
refrigerated storage |
(days) |
recovery (%) |
recovery (%) |
|
0 |
92.3 |
99.3 |
99.1 |
92.3 |
99.3 |
99.1 |
|
99.6 |
100.6 |
101.5 |
99.6 |
100.6 |
101.5 |
4 |
99.6 |
97.4 |
102.6 |
97.0 |
92.1 |
98.2 |
7 |
92.5 |
94.9 |
101.0 |
103.2 |
102.8 |
103.5 |
10 |
100.5 |
103.2 |
102.5 |
98.0 |
97.0 |
100.7 |
14 |
103.1 |
100.6 |
108.3 |
106.8 |
102.4 |
105.9 |
18 |
107.7 |
100.1 |
105.1 |
103.1 |
100.5 |
104.1 |
|
Figure 4.7.1.6.1. Ambient storage test for isoflurane on
Anasorb 747.
Figure 4.7.1.6.2. Refrigerated storage test for isoflurane on
Anasorb 747.
4.7.2 Analyte storage at low target concentration
4.7.2.1 Storage samples were generated by sampling from a
controlled test atmosphere containing 58.2
mg/m3 of enflurane, about 7.7 times the
1-ppm target concentration. Anasorb CMS tubes were
used to sample for 30 min at 0.05 L/min, the relative humidity was
about 80% at 22°. Thirty-six storage samples were
prepared. Six samples were analyzed immediately after generation,
fifteen tubes were stored at reduced temperature (4°) and the
other fifteen were stored in a closed drawer at ambient
temperature (about 22°). At 2-4 day intervals, three
samples were selected from each of the two sets and analyzed.
Table 4.7.2.1 |
|
|
time |
ambient storage |
refrigerated storage |
(days) |
recovery (%) |
recovery (%) |
|
0 |
82.6 |
88.7 |
94.1 |
82.6 |
88.7 |
94.1 |
|
98.8 |
98.3 |
102.9 |
98.8 |
98.3 |
102.9 |
4 |
90.1 |
92.9 |
97.0 |
90.4 |
101.0 |
101.7 |
7 |
91.6 |
99.9 |
102.2 |
96.6 |
103.9 |
99.0 |
11 |
99.1 |
96.6 |
100.9 |
90.0 |
100.8 |
96.9 |
13 |
88.3 |
95.9 |
97.3 |
104.9 |
97.7 |
98.1 |
15 |
96.3 |
92.1 |
95.6 |
86.6 |
93.0 |
102.0 |
|
Figure 4.7.2.1.1. Ambient storage test for enflurane on Anasorb
CMS.
Figure 4.7.2.1.2. Refrigerated storage test for enflurane on
Anasorb CMS.
4.7.2.2 Storage samples were generated by sampling from a
controlled test atmosphere containing 59.4
mg/m3 of enflurane, about 7.9 times the
1-ppm target concentration. Anasorb 747 tubes were
used to sample for 30 min at 0.05 L/min, the relative humidity was
about 80% at 22°. Thirty-six storage samples were
prepared. Six samples were analyzed immediately after generation,
fifteen tubes were stored at reduced temperature (4°) and the
other fifteen were stored in a closed drawer at ambient
temperature (about 22°). At 2-6 day intervals, three
samples were selected from each of the two sets and analyzed.
Table 4.7.2.2 |
Storage Test for Enflurane on
Anasorb 747 |
|
time |
ambient storage |
refrigerated storage |
(days) |
recovery (%) |
recovery (%) |
|
0 |
101.8 |
93.4 |
96.5 |
101.8 |
93.4 |
96.5 |
|
100.0 |
103.4 |
98.3 |
100.0 |
103.4 |
98.3 |
3 |
102.3 |
98.9 |
97.3 |
99.0 |
96.3 |
101.1 |
6 |
101.7 |
104.0 |
102.9 |
98.8 |
101.8 |
100.3 |
12 |
98.3 |
100.2 |
96.4 |
98.0 |
98.2 |
101.3 |
14 |
101.1 |
101.3 |
102.7 |
102.2 |
104.2 |
96.7 |
18 |
98.6 |
99.0 |
95.8 |
106.9 |
100.7 |
103.0 |
|
Figure
4.7.2.2.1. Ambient storage test for enflurane on Anasorb 747.
Figure
4.7.2.2.2. Refrigerated storage test for enflurane on Anasorb 747.
4.7.2.3 Storage samples were generated by sampling from a
controlled test atmosphere containing 70.1
mg/m3 of halothane, about 8.7 times the
1-ppm target concentration. Anasorb CMS tubes were used to sample
for 30 min at 0.05 L/min, the relative humidity was about 80% at
22°. Thirty-six storage samples were prepared. Six
samples were analyzed immediately after generation, fifteen tubes
were stored at reduced temperature (4°) and the other fifteen were
stored in a closed drawer at ambient temperature (about 22°). At
2-5 day intervals, three samples were selected from
each of the two sets and analyzed.
Table 4.7.2.3 |
Storage Test for Halothane on
Anasorb CMS |
|
time |
ambient storage |
refrigerated storage |
(days) |
recovery (%) |
recovery (%) |
|
0 |
91.7 |
93.8 |
98.7 |
91.7 |
93.8 |
98.7 |
|
101.1 |
103.6 |
106.9 |
101.1 |
103.6 |
106.9 |
3 |
80.6 |
88.4 |
95.5 |
89.3 |
95.4 |
105.5 |
8 |
78.8 |
85.6 |
90.1 |
92.6 |
99.3 |
100.5 |
12 |
79.2 |
85.3 |
92.4 |
86.5 |
107.4 |
100.0 |
14 |
77.2 |
86.9 |
98.0 |
97.0 |
98.6 |
101.2 |
16 |
73.4 |
80.0 |
92.0 |
92.5 |
94.4 |
100.2 |
|
Figure
4.7.2.3.1. Ambient storage test for halothane on Anasorb CMS.
Figure
4.7.2.3.2. Refrigerated storage test for halothane on Anasorb CMS.
4.7.2.4 Storage samples were generated by sampling from a
controlled test atmosphere containing 71.2
mg/m3 of halothane, about 8.8 times the
1-ppm target concentration. Anasorb 747 tubes were
used to sample for 30 min at 0.05 L/min, the relative humidity was
about 80% at 22°. Thirty-six storage samples were
prepared. Six samples were analyzed immediately after generation,
fifteen tubes were stored at reduced temperature (4°) and the
other fifteen were stored in a closed drawer at ambient
temperature (about 22°). At 2-4 day intervals, three
samples were selected from each of the two sets and analyzed.
Table 4.7.2.4 |
Storage Test for Halothane on
Anasorb 747 |
|
time |
ambient storage |
refrigerated storage |
(days) |
recovery (%) |
recovery (%) |
|
0 |
97.8 |
104.5 |
97.5 |
97.8 |
104.5 |
97.5 |
|
99.4 |
101.0 |
99.8 |
99.4 |
101.0 |
99.8 |
3 |
99.2 |
102.0 |
98.0 |
97.3 |
99.0 |
103.4 |
6 |
- |
100.2 |
100.8 |
98.4 |
105.3 |
97.1 |
9 |
104.5 |
103.7 |
101.0 |
98.9 |
111.2 |
102.6 |
13 |
101.0 |
102.7 |
100.6 |
106.0 |
102.3 |
99.7 |
15 |
105.7 |
103.0 |
101.4 |
113.3 |
101.0 |
101.1 |
|
Figure
4.7.2.4.1. Ambient storage test for halothane on Anasorb 747.
Figure
4.7.2.4.2. Refrigerated storage test for halothane on Anasorb 747.
4.7.2.5 Storage samples were generated by sampling from a
controlled test atmosphere containing 50.4
mg/m3 of isoflurane, about 6.7 times the
1-ppm target concentration. Anasorb CMS tubes were
used to sample for 30 min at 0.05 L/min, the relative humidity was
about 80% at 22°. Thirty-six storage samples were
prepared. Six samples were analyzed immediately after generation,
fifteen tubes were stored at reduced temperature (4°) and the
other fifteen were stored in a closed drawer at ambient
temperature (about 22°). At 24 day intervals, three samples were
selected from each of the two sets and analyzed.
Table 4.7.2.5 |
Storage Test for Isoflurane on
Anasorb CMS |
|
time
| ambient storage
| refrigerated storage |
(days)
| recovery (%)
| recovery (%) |
|
0
| 79.1
| 88.1
| 91.8
| 79.1
| 88.1
| 91.8 |
| 98.9
| 98.0
| 103.2
| 98.9
| 98.0
| 103.2 |
4
| 97.8
| 96.5
| 101.7
| 90.7
| 101.8
| 105.7 |
7
| 91.5
| 97.4
| 102.2
| 94.5
| 103.7
| 100.8 |
11
| 97.6
| 94.0
| 98.7
| 89.3
| 102.3
| 100.0 |
13
| 85.4
| 95.1
| 98.0
| 105.3
| 98.9
| 96.1 |
15
| 92.1
| 87.4
| 92.7
| 85.0
| 89.4
| 99.4 |
|
Figure
4.7.2.5.1. Ambient storage test for isonurane on Anasorb CMS.
Figure
4.7.2.5.2. Refrigerated storage test for isoflurane on Anasorb
CMS.
4.7.2.6 Storage samples were generated by sampling from a
controlled test atmosphere containing 52.7
mg/m3 of isoflurane, about 7 times the
1-ppm target concentration. Anasorb 747 tubes were
used to sample for 30 min at 0.05 L/min, the relative humidity was
about 80% at 22°. Thirty-six storage samples were
prepared. Six samples were analyzed immediately after generation,
fifteen tubes were stored at reduced temperature (4°) and the
other fifteen were stored in a closed drawer at ambient
temperature (about 22°). At 2-6 day intervals, three
samples were selected from each of the two sets and analyzed.
Table 4.7.2.6 |
Storage Test for Isoflurane on
Anasorb 747 |
|
time |
ambient storage |
refrigerated storage |
(days) |
recovery (%) |
recovery (%) |
|
0 |
102.3 |
90.5 |
96.7 |
102.3 |
90.5 |
96.7 |
|
100.0 |
104.2 |
96.8 |
100.0 |
104.2 |
96.8 |
3 |
101.8 |
98.0 |
98.4 |
100.7 |
97.4 |
102.0 |
6 |
102.5 |
102.6 |
101.2 |
98.4 |
102.3 |
100.6 |
12 |
97.7 |
98.5 |
96.7 |
99.9 |
100.5 |
102.3 |
14 |
99.5 |
99.5 |
100.9 |
102.8 |
101.6 |
96.0 |
16 |
98.4 |
96.9 |
97.0 |
107.6 |
101.8 |
103.2 |
|
Figure
4.7.2.6.1. Ambient storage test for isoflurane on Anasorb 747.
Figure
4.7.2.6.2. Refrigerated storage test for isoflurane on Anasorb
747.
4.8 Reproducibility
4.8.1 Analyte reproducibility at high target concentration
4.8.1.1 Six samples for each adsorbent were prepared by
collecting them from a 75-ppm controlled test atmosphere
containing enflurane and isoflurane for 4 h at 0.05 L/min. The
samples were submitted to an OSHA Salt Lake Technical Center
service branch. The samples were analyzed after being stored for
21 days at 4°. Sample results were corrected for desorption
efficiency. No sample result for enflurane or isoflurane had a
deviation greater than the precision of the overall procedure
determined in Section 4.6.
Table 4.8.1.1.1 |
Reproducibility Data for
Enflurane |
|
|
Anasorb CMS |
Anasorb 747 |
sample |
expected |
reported |
recovery |
deviation |
reported |
recovery |
deviation |
|
(mg/m3) |
(mg/m3) |
(%) |
(%) |
(mg/m3) |
(%) |
(%) |
|
1 |
570 |
546.3 |
95.9 |
-4.1 |
517.6 |
90.8 |
-9.2 |
2 |
570 |
517.3 |
90.9 |
-9.1 |
533.1 |
93.5 |
-6.5 |
3 |
570 |
544.1 |
95.5 |
-4.5 |
520.3 |
91.3 |
-8.7 |
4 |
570 |
556.9 |
97.7 |
-2.3 |
532.1 |
93.4 |
-6.6 |
5 |
570 |
518.7 |
91.0 |
-9.0 |
499.6 |
87.7 |
-12.3 |
6 |
570 |
536.9 |
94.2 |
-5.8 |
524.0 |
91.9 |
-8.1 |
|
Table 4.8.1.1.2 |
Reproducibility Data for
Isoflurane |
|
|
Anasorb CMS |
Anasorb 747 |
sample |
expected |
reported |
recovery |
deviabon |
reported |
recovery |
deviation |
|
(mg/m3) |
(mg/m3) |
(%) |
(%) |
(mg/m3) |
(%) |
(%) |
|
1 |
562.5 |
596.6 |
106.1 |
+6.1 |
562.3 |
100.0 |
0 |
2 |
562.5 |
550.2 |
97.8 |
-2.2 |
576.6 |
102.5 |
+2.5 |
3 |
562.5 |
591.1 |
105.1 |
+5.1 |
564.0 |
100.3 |
+0.3 |
4 |
562.5 |
609.0 |
108.3 |
+8.3 |
578.1 |
102.8 |
+2.8 |
5 |
562.5 |
564.8 |
100.3 |
+0.3 |
540.3 |
96.1 |
-3.9 |
6 |
562.5 |
587.4 |
104.4 |
+4.4 |
569.2 |
101.2 |
1.2 |
|
4.8.1.2 Six samples for each adsorbent were prepared by
collecting them from a 50-ppm controlled test atmosphere
containing halothane for 4 h at 0.05 L/min. The samples were
submilted to an OSHA Salt Lake Technical Center service branch.
The samples were analyzed after being stored for 17 days at 4°.
Sample results were corrected for desorption efficiency. No sample
result for halothane had a deviation greater than the precision of
the overall procedure determined in Section 4.6.
Table 4.8.1.2 |
Reproducibility Data for
Halothane |
|
|
Anasorb CMS |
Anasorb 747 |
sample |
expected |
reported |
recovery |
deviation |
reported |
recovery |
deviation |
|
(mg/m3) |
(mg/m3) |
(%) |
(%) |
(mg/m3) |
(%) |
(%) |
|
1 |
408 |
411.0 |
100.7 |
+0.7 |
406.2 |
99.6 |
-0.4 |
2 |
408 |
406.6 |
99.7 |
-0.3 |
411.1 |
100.8 |
+0.8 |
3 |
408 |
402.8 |
98.7 |
-1.3 |
411.5 |
100.8 |
+0.8 |
4 |
408 |
409.8 |
100.4 |
+0.4 |
410.6 |
100.6 |
+0.6 |
5 |
408 |
399.7 |
98.0 |
-2.0 |
401.7 |
98.5 |
-1.5 |
6 |
408 |
406.4 |
99.6 |
-0.4 |
408.1 |
100.0 |
0 |
|
4.8.2 Analyte reproducibility at low target concentration
4.8.2.1 Six samples for each adsorbent were prepared by
collecting them from a 1-ppm controlled test atmosphere containing
enflurane and isoflurane for 4 h at 0.05 L/min. The samples were
submitted to an OSHA Salt Lake Technical Center service branch.
The samples were analyzed after being stored for 22 days at 4°.
Sample results were corrected for desorption efficiency. No sample
result for enflurane and isoflurane had a deviation greater than
the precision of the overall procedure determined in Section 4.6.
Table 4.8.2.1.1 |
Reproducibility Data for
Enflurane |
|
|
Anasorb CMS |
Anasorb 747 |
sample |
expected |
reported |
recovery |
deviation |
reported |
recovery |
deviation |
|
(mg/m3) |
(mg/m3) |
(%) |
(%) |
(mg/m3) |
(%) |
(%) |
|
1 |
7.62 |
8.02 |
105.2 |
+5.2 |
7.81 |
102.5 |
+2.5 |
2 |
7.62 |
7.58 |
99.5 |
-0.5 |
7.82 |
102.6 |
+2.6 |
3 |
7.62 |
7.62 |
100.0 |
0 |
7.82 |
102.6 |
+2.6 |
4 |
7.62 |
6.55 |
86.0 |
-14.0 |
6.95 |
91.2 |
-8.8 |
5 |
7.62 |
7.58 |
99.5 |
-0.5 |
7.73 |
101.4 |
+1.4 |
6 |
7.62 |
7.65 |
100.4 |
+0.4 |
8.48 |
111.3 |
+11.3 |
|
Table 4.8.2.3 |
Reproducibility Data for
Isoflurane |
|
|
Anasorb CMS |
Anasorb 747 |
sample |
expected |
reported |
recovery |
deviation |
reported |
recovery |
deviation |
|
(mg/m3) |
(mg/m3) |
(%) |
(%) |
(mg/m3) |
(%) |
(%) |
|
1 |
7.52 |
7.74 |
102.9 |
+2.9 |
7.91 |
105.2 |
+5.2 |
2 |
7.52 |
7.47 |
99.3 |
-0.7 |
8.22 |
109.3 |
+9.3 |
3 |
7.52 |
7.65 |
101.7 |
+1.7 |
8.32 |
110.6 |
+10.6 |
4 |
7.52 |
6.58 |
87.5 |
-12.5 |
7.09 |
94.3 |
-5.7 |
5 |
7.52 |
7.15 |
95.1 |
-4.9 |
8.22 |
109.3 |
+9.3 |
6 |
7.52 |
7.59 |
100.9 |
+0.9 |
7.70 |
102.4 |
+2.4 |
|
4.8.2.2 Six samples for each adsorbent were prepared by
collecting them from a 1-ppm controlled test atmosphere containing
halothane for 4 h at 0.05 L/min. The samples were submitted to an
OSHA Salt Lake Technical Center service branch. The samples were
analyzed after being stored for 23 days at 4°. Sample results were
corrected for desorption efficiency. No sample result for
halothane had a deviation greater than the precision of the
overall procedure determined in Section 4.6.
Table 4.8.2.2 |
Reproducibility Data for
Halothane |
|
|
Anasorb CMS |
Anasorb 747 |
sample |
expected |
reported |
recovery |
deviation |
reported |
recovery |
deviation |
|
(mg/m3) |
(mg/m3) |
(%) |
(%) |
(mg/m3) |
(%) |
(%) |
|
1 |
8.40 |
7.38 |
37.9 |
-12.1 |
8.79 |
104.6 |
+4.6 |
2 |
8.40 |
7.80 |
92.9 |
-7.1 |
8.53 |
101.5 |
+1.5 |
3 |
8.40 |
8.32 |
99.0 |
-1.0 |
9.03 |
107.5 |
+7.5 |
4 |
8.40 |
7.93 |
94.4 |
-5.6 |
8.81 |
104.9 |
+4.9 |
5 |
8.40 |
8.20 |
97.6 |
-2.4 |
8.70 |
103.6 |
+3.6 |
6 |
8.40 |
8.67 |
103.2 |
+3.2 |
8.93 |
106.3 |
+6.3 |
|
4.9 Sampler capacity
4.9.1 Anasorb CMS
4.9.1.1 The sampling capacity of the front section of an
Anasorb CMS sampling tube was tested by sampling from a
dynamically generated test atmosphere of enflurane (1247
mg/m3 or 165 ppm). The samples were
collected at 0.05 L/min and the relative humidity was about 80% at
22°. A GC with a gas sampling valve was placed
in-line behind the 150-mg front test
section. The valve was rotated to measure the amount of enflurane
passing through the sampler at the time of rotation. The 5%
breakthrough air volume was determined to be 28.8 L.
Table 4.9.1.1 |
Capacity of Enflurane on Anasorb CMS |
|
first test |
second test |
air volume |
breakthrough |
air volume |
breakthrough |
(L) |
(%) |
(L) |
(%) |
|
17.54 |
0 |
16.39 |
0 |
18.24 |
0.79 |
17.11 |
0.27 |
19.62 |
1.57 |
20.76 |
0.69 |
21.93 |
2.07 |
20.97 |
1.04 |
23.18 |
2.50 |
22.18 |
1.34 |
24.19 |
2.87 |
22.90 |
1.57 |
25.06 |
3.23 |
24.20 |
1.98 |
26.12 |
3.93 |
25.55 |
2.38 |
26.99 |
4.41 |
26.46 |
2.83 |
28.48 |
5.03 |
27.57 |
3.33 |
29.23 |
5.59 |
27.72 |
3.81 |
29.99 |
6.25 |
28.44 |
4.24 |
|
Figure 4.9.1.1. Five percent breakthrough air volume for
enflurane on Anasorb CMS.
4.9.1.2 The sampling capacity of the front section of an
Anasorb CMS sampling tube was tested by sampling from a
dynamically generated test atmosphere of halothane (753
mg/m3 or 93.3 ppm). The samples were
collected at 0.05 L/min and the relative humidity was about 80% at
22°. A GC with a gas sampling valve was placed
in-line behind the 150-mg front test
section. The valve was rotated to measure the amount of halothane
passing through the sampler at the time of rotation. The 5%
breakthrough air volume was determined to be 15.8 L.
Table 4.9.1.2 |
Capacity of Halothane on Anasorb CMS |
|
first test |
second test |
air volume |
breakthrough |
air volume |
breakthrough |
(L) |
(%) |
(L) |
(%) |
|
5.88 |
0 |
6.06 |
0 |
7.10 |
0 |
7.26 |
0 |
8.37 |
0 |
8.52 |
0 |
12.42 |
3.11 |
12.53 |
3.08 |
12.93 |
3.28 |
13.03 |
3.21 |
13.44 |
3.71 |
13.53 |
3.65 |
13.94 |
4.11 |
14.03 |
4.00 |
14.55 |
4.53 |
14.63 |
4.29 |
15.06 |
4.64 |
15.13 |
4.47 |
15.56 |
4.97 |
15.63 |
4.72 |
16.07 |
5.26 |
16.13 |
4.96 |
16.58 |
5.74 |
16.63 |
5.19 |
17.09 |
5.83 |
17.73 |
5.58 |
17.59 |
6.24 |
17.64 |
5.87 |
18.10 |
6.32 |
|
Figure 4.9.1.2. Five percent breakthrough air volume for
haothane on Anasorb CMS.
4.9.1.3 The sampling capacity of the front section of an
Anasorb CMS sampling tube was tested by sampling from a
dynamically generated test atmosphere of isoflurane (1246
mg/m3 or 165 ppm). The samples were
collected at 0.05 L/min and the relative humidity was about 80% at
22°. A GC with a gas sampling valve was placed
in-line behind the 150-mg front test
sector. The valve was rotated to measure the amount of isoflurane
passing through the sampler at the time of rotation. The 5%
breakthrough air volume was determined to be 24.0 L.
Table 4.9.1.3 |
Capacity of Isoflurane on Anasorb CMS |
|
first test |
second test |
air volume |
breakthrough |
air volume |
breakthrough |
(L) |
(%) |
(L) |
(%) |
|
9.10 |
0 |
10.10 |
0 |
13.48 |
0.42 |
14.11 |
0.96 |
14.51 |
0.73 |
15.12 |
1.42 |
15.50 |
1.03 |
16.13 |
1.76 |
16.48 |
1.39 |
17.14 |
1.96 |
17.76 |
1.75 |
18.50 |
2.42 |
19.19 |
2.36 |
20.06 |
3.15 |
20.41 |
2.86 |
21.17 |
3.74 |
21.50 |
3.24 |
22.27 |
4.24 |
22.68 |
3.67 |
23.49 |
5.21 |
23.86 |
4.32 |
24.70 |
6.16 |
24.85 |
4.65 |
25.70 |
6.89 |
|
Figure 4.9.1.3. Five percent breakthrough air volume for
isoflurane on Anasorb CMS.
4.9.2 Anasorb 747
4.9.2.1 The sampling capacity of the front section of an
Anasorb 747 sampling tube was tested by sampling from a
dynamically generated test atmosphere of enflurane (1247
mg/m3 or 165 ppm). The samples were
collected at 0.05 L/min and the relative humidity was about 80% at
22°. A GC with a gas sampling valve was placed
in-line behind the 140-mg front test
section. The valve was rotated to measure the amount of enflurane
passing through the sampler at the time of rotation. The 5%
breakthrough air volume was determined to be 14.2 L.
Table 4.9.2.1 |
Capacity of Enflurane on Anasorb 747 |
|
first test |
second test |
air volume |
breakthrough |
air volume |
breakthrough |
(L) |
(%) |
(L) |
(%) |
|
9.85 |
0 |
10.10 |
0 |
13.13 |
1.40 |
13.43 |
2.22 |
13.69 |
3.12 |
13.94 |
3.98 |
14.24 |
5.61 |
14.49 |
6.29 |
14.75 |
7.77 |
15.00 |
8.12 |
15.25 |
9.13 |
15.50 |
9.95 |
|
Figure 4.9.2.1. Five percent breakthrough air volume for
enflurane on Anasorb 747.
4.9.2.2 The sampling capacity of the front section of an
Anasorb 747 sampling tube was tested by sampling from a
dynamically generated test atmosphere of halothane (753
mg/m3 or 93.3 ppm). The samples were
collected at 0.05 L/min and the relative humidity was about 80% at
22°. A GC with a gas sampling valve was placed
in-line behind the 140-mg front test
section. The valve was rotated to measure the amount of halothane
passing through the sampler at the time of rotation. The 5%
breakthrough air volume was determined to be 19.9 L.
Table 4.9.2.2 |
Capacity of Halothane on Anasorb 747 |
|
first test |
second test |
air volume |
breakthrough |
air volume |
breakthrough |
(L) |
(%) |
(L) |
(%) |
|
16.07 |
0 |
16.13 |
0 |
17.34 |
0 |
17.38 |
0 |
18.86 |
0.52 |
18.89 |
0.45 |
19.87 |
2.56 |
19.94 |
2.37 |
20.18 |
5.06 |
20.19 |
4.86 |
20.43 |
8.67 |
20.44 |
8.70 |
20.69 |
12.58 |
20.69 |
12.70 |
|
Figure 4.9.2.2. Five percent breakthrough air volume for
halothane on Anasorb 747.
4.9.2.3 The sampling capacity of the front section of an
Anasorb 747 sampling tube was tested by sampling from a
dynamically generated test atmosphere of isoflurane (1246
mg/m3 or 165 ppm). The samples were
collected at 0.05 L/min and the relative humidity was about 80% at
22°. A GC with a gas sampling valve was placed
in-line behind the 140-mg front test
section. The valve was rotated to measure the amount of isoflurane
passing through the sampler at the time of rotation. The 5%
breakthrough air volume was determined to be 17.5 L.
Table 4.9.2.3 |
Capacity of Isoflurane on Anasorb 747 |
|
first test |
second test |
air volume |
breakthrough |
air volume |
breakthrough |
(L) |
(%) |
(L) |
(%) |
|
11.21 |
0 |
10.66 |
0 |
12.20 |
0 |
11.62 |
0 |
13.20 |
0.18 |
12.68 |
0.12 |
13.94 |
0.35 |
13.64 |
0.61 |
14.44 |
0.59 |
14.39 |
1.12 |
14.94 |
0.93 |
14.90 |
1.59 |
15.44 |
1.36 |
15.40 |
2.13 |
15.94 |
1.94 |
15.91 |
2.86 |
16.43 |
2.50 |
16.41 |
3.72 |
16.93 |
3.29 |
16.92 |
4.62 |
17.43 |
4.07 |
17.42 |
5.56 |
17.93 |
4.85 |
17.93 |
6.69 |
18.43 |
5.73 |
18.43 |
7.69 |
|
|
18.94 |
8.76 |
|
Figure 4.9.2.3 Five percent breakthrough air volume for
isoflurane on Anasorb 747.
4.10 Desorption efficiency and stability of desorbed samples
4.10.1 Anasorb CMS at high target concentration (TC)
4.10.1.1 Enflurane
The desorption efficiencies (DE) of enflurane were determined
by liquid-spiking150 mg portions of Anasorb CMS with
amounts equivalent to 0.05 to 2 times the 75-ppm
target concentration. These samples were stored overnight at
ambient temperature and then desorbed and analyzed. The average
desorption emciency over the working range of 0.5 to 2 times the
target concentration is 99.8%.
Table 4.10.1.1.1 |
Desorbtion Efficiency of
Enflurane from Anasorb CMS at High TC |
|
× target concn |
0.05× |
0.1× |
0.2× |
0.5× |
1.0× |
2.0× |
(µg/sample) |
340.5 |
681 |
1362 |
3403 |
6810 |
13620 |
|
DE (%) |
101.4 |
98.5 |
100.2 |
100.2 |
99.8 |
99.3 |
|
98.7 |
98.3 |
97.2 |
99.6 |
98.2 |
98.9 |
|
100.5 |
100.3 |
100.1 |
99.9 |
100.6 |
98.9 |
|
98.9 |
100.7 |
101.6 |
101.2 |
100.7 |
100.1 |
|
100.8 |
100.0 |
99.6 |
100.3 |
99.7 |
97.8 |
|
100.1 |
100.7 |
100.2 |
101.0 |
101.0 |
99.2 |
|
100.1 |
99.8 |
99.8 |
100.4 |
100.0 |
99.0 |
|
The stability of desorbed samples was investigated by
reanalyzing the target concentration samples 22.5 h after intial
analysis. After the original analysis was performed, three vials
were recapped with new septa while the remaining three retained
their punctured septa. The samples vials were stored in the
refrigerated sampling tray for the GC injector. The samples were
reanalyzed with fresh standards. The average percent change was
-2.0% for samples that were resealed with new septa, and -4.2% for
those that retained their punctured septa.
Table 4.10.1.1.2 |
Stability of Desorbed Samples for
Enflurane from Anasorb CMS |
|
punctured septa
replaced |
punctured septa
retained |
initial |
DE after |
|
initial |
DE after |
|
DE |
one day |
difference |
DE |
one day |
difference |
(%) |
(%) |
|
(%) |
(%) |
|
|
99.8 |
97.3 |
-2.5 |
100.7 |
98.7 |
-2.0 |
98.2 |
96.7 |
-1.5 |
99.7 |
95.4 |
-4.3 |
100.6 |
98.6 |
-2.0 |
101.0 |
94.8 |
-6.2 |
|
(averages) |
|
|
(averages) |
|
99.5 |
97.5 |
-2.0 |
100.5 |
96.3 |
-4.2 |
|
4.10.1.2. Halothane
The desorption efficiencies (DE) of halothane were determined
by liquid-spiking 150-mg portions of Anasorb CMS with
amounts equivalent to 0.05 to 2 times the 50-ppm
target concentration. These samples were stored overnight at
ambient temperature and then desorbed and analyzed. The average
desorption efficiency over the working range of 0.5 to 2 times the
target concentration is 99.5%.
Table 4.10.1.2.1 |
Desorption Efficiency of
Halothane from Anasorb CMS at High TC |
|
×target concn |
0.05× |
0.1× |
0.2× |
0.5× |
1.0× |
2.0× |
(µg/sample) |
243.1 |
486.2 |
972.4 |
2431 |
4862 |
9724 |
|
DE (%) |
99.4 |
98.6 |
98.9 |
99.1 |
99.8 |
99.6 |
|
99.6 |
98.5 |
96.5 |
98.8 |
98.0 |
99.9 |
|
97.4 |
99.1 |
98.5 |
101.8 |
100.0 |
99.4 |
|
101.0 |
99.3 |
99.5 |
99.6 |
99.7 |
100.1 |
|
99.0 |
98.9 |
98.3 |
99.5 |
99.2 |
98.6 |
|
99.2 |
99.0 |
99.1 |
99.8 |
99.9 |
98.5 |
|
99.3 |
98.9 |
98.5 |
99.8 |
99.4 |
99.4 |
|
The stability of desorbed samples was investigated by
reanalyzing the target concentration samples 22.5 h after initial
analysis. After the original analysis was performed, three vials
were recapped with new septa while the remaining three retained
their punctured septa. The samples vials were stored in the
refrigerated sampling tray for the GC injector. The samples were
reanalyzed with fresh standards. The average percent change was
-1.5% for samples that were resealed with new septa, and -3.4% for
those that retained their punctured septa.
Table 4.10.1.2.2 |
Stability of Desorbed Samples for
Halothane from Anasorb CMS |
|
punctured septa
replaced |
punctured septa
retained |
initial |
DE after |
|
initial |
DE after |
|
DE |
one day |
difference |
DE |
one day |
difference |
(%) |
(%) |
|
(%) |
(%) |
|
|
99.8 |
97.4 |
-2.4 |
99.7 |
97.7 |
-2.0 |
98.0 |
97.1 |
-0.9 |
99.2 |
96.1 |
-3.1 |
100.0 |
98.8 |
-1.2 |
99.9 |
94.8 |
-5.1 |
|
(averages) |
|
|
(averages) |
|
99.3 |
97.8 |
-1.5 |
99.6 |
96.2 |
-3.4 |
|
4.10.1.3 Isoflurane
The desorption efficiencies (DE) of isoflurane were determined
by liquid-spiking 150-mg portions of Anasorb CMS with
amounts equivalent to 0.05 to 2 times the 75-ppm
target concentration. These samples were stored overnight at
ambient temperature and then desorbed and analyzed. The average
desorption efficiency over the working range of 0.5 to 2 times the
target concentration is 99.2%.
Table 4.10.1.3.1 |
Desorption Efficiency of
Isoflurane from Anasorb CMS at High TC |
|
× target concn |
0.05× |
0.1× |
0.2× |
0.5× |
1.0× |
2.0× |
(µg/sample) |
336 |
672 |
1344 |
3360 |
6720 |
13440 |
|
DE (%) |
100.2 |
97.3 |
99.6 |
99.8 |
99.1 |
97.6 |
|
99.2 |
99.3 |
99.4 |
99.3 |
97.6 |
97.3 |
|
98.7 |
99.9 |
100.9 |
99.4 |
99.7 |
97.8 |
|
101.3 |
100.2 |
99.3 |
100.7 |
99.7 |
98.5 |
|
100.9 |
99.1 |
99.5 |
100.2 |
103.3 |
96.8 |
|
99.9 |
99.9 |
96.5 |
100.8 |
99.5 |
98.1 |
|
100.0 |
99.3 |
99.2 |
100.0 |
99.8 |
97.7 |
|
The stability of desorbed samples was investigated by
reanalyzing the target concentration samples 22.5 h after initial
analysis. After the original analysis was performed, three vials
were recapped with new septa while the remaining three retained
their punctured septa. The samples vials were stored in the
refrigerated sampling tray for the GC injector. The samples were
reanalyzed with fresh standards. The average percent change was
-2.3% for samples that were resealed with new septa, and -4.1% for
those that retained their punctured septa.
Table 4.10.1.3.2 Stability of
Desorbed Samples for Isoflurane from Anasorb CMS |
|
|
punctured septa
replaced |
punctured septa
retained |
initial |
DE after |
|
initial |
DE after |
|
DE |
one day |
difference |
DE |
one day |
difference |
(%) |
(%) |
|
(%) |
(%) |
|
|
99.1 |
96.2 |
-2.9 |
99.7 |
98.0 |
-1.7 |
97.6 |
95.5 |
-2.1 |
103.3 |
98.1 |
-5.2 |
99.7 |
97.8 |
-1.9 |
99.5 |
93.9 |
-5.4 |
|
(averages) |
|
|
(averages) |
|
98.8 |
96.5 |
-2.3 |
100.8 |
96.7 |
-4.1 |
|
4.10.2 Anasorb 747 at high target concentration (TC)
4.1 0.2.1 Enflurane
The desorption efficiencies (DE) of.enflurane were determined
by liquid-spiking 140-mg portions of Anasorb 747 with
amounts equivalent to 0.05 to 2 times the75-ppm
target cocentration. These samples were stored overnight at
ambient temperature and then desorbed and analyzed. The average
desorption efficiency over the working range of 0.5 to 2 times the
target concentration is 100.5%.
Table 4.10.2.1.1 |
Desorption Effidency of Enflurane
from Anasorb 747 at High TC |
|
× target concn |
0.05× |
0.1× |
0.2× |
0.5× |
1.0× |
2.0× |
(µg/sample) |
340.5 |
681 |
1362 |
3405 |
6810 |
13620 |
|
DE (%) |
100.9 |
100.6 |
101.4 |
98.9 |
103.9 |
99.5 |
|
99.1 |
99.1 |
98.0 |
100.0 |
100.9 |
100.3 |
|
101.1 |
99.5 |
98.4 |
98.9 |
105.3 |
102.6 |
|
101.7 |
99.6 |
102.5 |
98.3 |
99.5 |
101.0 |
|
99.9 |
101.8 |
102.0 |
100.6 |
100.2 |
102.3 |
|
98.6 |
99.6 |
99.7 |
97.6 |
100.5 |
97.5 |
|
100.2 |
100.0 |
100.3 |
99.1 |
101.8 |
100.5 |
|
The stability of desorbed samples was investigated by
reanalyzing the target concentration samples 22.5 h after initial
analysis. After the original analysis was perfommed, three vials
were recapped with new septa while the remaining three retained
their punctured septa. The samples vials were stored in the
refrigerated sampling tray for the GC injector. The samples were
reanalyzed with fresh standards. The average percent change was
-0.5% for samples that were resealed with new septa, and -3.7% for
those that retained their punctured septa.
Table 4.10.2.1.2 |
Stability of Desorbed Samples for
Enflurane from Anasorb 747 |
|
punctured septa
replaced |
punctured septa
retained |
initial |
DE after |
|
initial |
DE after |
|
DE |
one day |
difference |
DE |
one day |
difference |
(%) |
(%) |
|
(%) |
(%) |
|
|
99.5 |
99.4 |
-0.1 |
105.3 |
99.3 |
-6.0 |
100.2 |
98.9 |
-1.3 |
103.9 |
99.8 |
-4.1 |
100.5 |
100.3 |
-0.2 |
100.9 |
99.8 |
-1.1 |
|
(averages) |
|
|
(averages) |
|
100.1 |
99.5 |
-0.5 |
103.4 |
99.6 |
-3.7 |
|
4.10.2.2 Halothane
The desorption efficiencies (DE) of halothane were determined
by liquid-spiking 140-mg portions of Anasorb 747 with
amounts equivalent to 0.05 to 2 times the 5-ppm
target concentration. These samples were stored overnight at
ambient temperature and then desorbed and analyzed. The average
desorption efficiency over the working range of 0.5 to 2 times the
target concentration is 99.3%.
Table 4.10.2.2.1 |
Desorption Efficiency of
Halothane from Anasorb 747 at High TC |
|
× target concn |
0.05× |
0.1× |
0.2× |
0.5× |
1.0× |
2.0× |
(µg/sample) |
243.5 |
486.2 |
972.4 |
2431 |
4862 |
9724 |
|
DE (%) |
98.9 |
98.9 |
100.0 |
99.9 |
99.3 |
99.1 |
|
98.7 |
98.2 |
98.1 |
-100.1 |
100.6 |
98.4 |
|
100.4 |
98.2 |
101.5 |
100.2 |
100.1 |
100.7 |
|
100.6 |
98.3 |
100.4 |
100.5 |
96.5 |
99.6 |
|
100.4 |
98.1 |
101.3 |
99.3 |
96.6 |
101.1 |
|
98.7 |
99.9 |
100.8 |
100.7 |
97.0 |
97.1 |
|
99.6 |
98.6 |
100.4 |
100.1 |
98.4 |
99.3 |
|
The stability of desorbed samples was investigated by
reanalyzing the target concentration samples 22.5 h after infflal
analysis. After the original analysis was performed, three vials
were recapped with new septa while the remaining three retained
their punctured septa. The samples vials were stored in the
refrigerated sampling tray for the GC injector. The samples were
reanalyzed with fresh standards. The average percent change was
+1.5% for samples that were resealed with new septa, and -1.6% for
those that retained their punctured septa.
Table 4.10.2.2.2 |
Stability of Desorbed Samples for
Halothane from Anasorb 747 |
|
punctured septa
replaced |
punctured septa
retained |
initial |
DE after |
|
initial |
DE after |
|
DE |
one day |
difference |
DE |
one day |
difference |
(%) |
(%) |
|
(%) |
(%) |
|
|
96.5 |
96.6 |
+0.1 |
99.3 |
98.0 |
-1.36 |
96.6 |
98.6 |
+2.0 |
100.6 |
99.0 |
-1.6 |
97.0 |
99.3 |
+2.3 |
100.1 |
98.3 |
-1.8 |
|
(averages) |
|
|
(averages) |
|
96.7 |
98.2 |
+1.5 |
100.0 |
98.4 |
-1.6 |
|
4.10.2.3 Isoflurane
The desorption effciencies (DE) of isoflurane were determined
by liquid-spiking 140-mg portions of Anasorb 747 with
amounts equivalent to 0.05 to 2 times the 75-ppm
target concentration. These samples were stored overnight at
ambient temperature and then desorbed and analyzed. The average
desorption efficiency over the working range of 0.5 to 2 times the
target concentration is 100.2%.
Table 4.10.2.3.1 |
Desorption Efficiency of
Isoflurane from Anasorb 747 at High TC |
|
× target concn |
0.05× |
0.1× |
0.2× |
0.5× |
1.0× |
2.0× |
(µg/sample) |
336 |
672 |
1344 |
3360 |
6720 |
13440 |
|
DE (%) |
97.9 |
97.7 |
99.3 |
100.3 |
102.8 |
98.3 |
|
98.5 |
96.6 |
96.4 |
101.5 |
104.0 |
99.5 |
|
99.3 |
97.1 |
100.2 |
98.2 |
99.6 |
101.7 |
|
98.9 |
96.7 |
99.9 |
100.4 |
100.1 |
100.2 |
|
96.9 |
98.6 |
101.3 |
100.3 |
99.5 |
101.3 |
|
96.7 |
96.6 |
97.1 |
99.6 |
99.1 |
96.5 |
|
98.0 |
97.2 |
99.0 |
100.1 |
100.9 |
99.6 |
|
The stability of desorbed samples was investigated by
reanalyzing the target concentration samples 22.5 h after initial
analysis. After the original analysis was performed, three vials
were recapped with new septa while the remaining three retained
their punctured septa. The samples vials were stored in the
refrigerated sampling tray for the GC injector. The samples were
reanalyzed with fresh standards. The average percent change was
-3.4% for samples that were resealed with new septa, and -4.8% for
those that retained their punctured septa.
Table 4.10.2.3.2 |
Stability of Desorbed Samples for
Isoflurane from Anasorb 747 |
|
punctured septa
replaced |
punctured septa
retained |
initial |
DE after |
|
initial |
DE after |
|
DE |
one day |
difference |
DE |
one day |
difference |
(%) |
(%) |
|
(%) |
(%) |
|
|
100.1 |
94.0 |
-6.1 |
102.8 |
97.2 |
-5.6 |
99.5 |
96.4 |
-3.1 |
104.0 |
97.2 |
-6.8 |
99.1 |
98.1 |
-1.0 |
99.6 |
97.6 |
-2.0 |
|
(averages) |
|
|
(averages) |
|
99.6 |
96.2 |
-3.4 |
102.1 |
97.3 |
-4.8 |
|
4.10.3 Anasorb CMS at low target concentration (TC)
4.10.3.1 Enflurane
The desorption efficiencies (DE) of enflurane were determined
by liquid-spiking 150 mg portions of Anasorb CMS with
amounts equivalent to 0.05 to 2 times the 1-ppm
target concentration. These samples were stored overnight at
ambient temperature and then desorbed and analyzed. The average
desorption efficiency over the working range of 0.5 to 2 times the
target concentration is 100.3%.
Table 4.10.3.1.1 |
Desorption Efficiency of
Enflurane from Anasorb CMS at Low TC |
|
× target concn |
0.05× |
0.1× |
0.2× |
0.5× |
1.0× |
2.0× |
(µg/sample) |
4.56 |
9.12 |
18.24 |
45.6 |
91.2 |
182.4 |
|
DE (%) |
99.4 |
100.2 |
99.5 |
101.0 |
98.7 |
99.1 |
|
99.2 |
98.7 |
99.3 |
102.2 |
98.7 |
100.2 |
|
101.4 |
101.2 |
97.7 |
101.1 |
97.7 |
101.8 |
|
101.5 |
100.0 |
99.0 |
101.3 |
98.7 |
100.4 |
|
99.4 |
101.6 |
100.0 |
99.4 |
99.4 |
102.8 |
|
99.4 |
100.5 |
101.7 |
100.9 |
100.3 |
101.8 |
|
100.2 |
100.4 |
99.5 |
101.0 |
98.9 |
101.0 |
|
The stability of desorbed samples was investigated by
reanalyzing the target concentration samples 22.5 h after initial
analysis. After the original analysis was perfommed, three vials
were recapped with new septa while the remaining three retained
their punctured septa. The samples vials were stored in the
refrigerated sampling tray for the GC injector. The samples were
reanalyzed with fresh standards. The average percent change was
+3.3% for samples that were resealed with new septa, and +0.9% for
those that retained their punctured septa.
Table 4.10.3.1.2 |
Stability of Desorbed Samples for
Enflurane from Anasorb CMS |
|
punctured septa
replaced |
punctured septa
retained |
initial |
DE after |
|
initial |
DE after |
|
DE |
one day |
difference |
DE |
one day |
difference |
(%) |
(%) |
|
(%) |
(%) |
|
|
98.7 |
101.4 |
+2.7 |
98.7 |
98.7 |
0 |
98.7 |
101.9 |
+3.2 |
99.4 |
99.8 |
+0.4 |
97.7 |
101.7 |
+4.0 |
100.3 |
102.8 |
+2.5 |
|
(averages) |
|
|
(averages) |
|
98.4 |
101.7 |
+3.3 |
99.5 |
100.4 |
+0.9 |
|
4.10.3.2 Halothane
The desorption efficiencies (DE) of enflurane were determined
by liquid-spiking 150-mg portions of Anasorb CMS with
amounts equivalent to 0.05 to 2 times the 1-ppm
target concentration. These samples were stored overnight at
ambient temperature and the desorbed and analyzed. The average
desorption efficiency over the working range of 0.5 to 2 times the
target concentration is 99.7%.
Table 4.10.3.2.1 |
Desorpbon Efficiency of Halothane
from Anasorb CMS at Low TC |
|
× target concn |
0.05× |
0.1× |
0.2× |
0.5× |
1.0× |
2.0× |
(µg/sample) |
4.862 |
9.724 |
19.45 |
48.62 |
97.24 |
194.5 |
|
DE (%) |
101.1 |
100.3 |
100.8 |
99.1 |
98.7 |
98.2 |
|
99.7 |
103.2 |
99.1 |
98.3 |
99.5 |
98.7 |
|
98.8 |
99.5 |
100.8 |
100.3 |
99.1 |
100.1 |
|
99.7 |
100.0 |
98.5 |
100.0 |
100.2 |
100.5 |
|
99.5 |
100.5 |
99.4 |
98.9 |
102.4 |
100.1 |
|
98.9 |
99.7 |
100.2 |
97.5 |
101.1 |
101.7 |
|
99.6 |
100.5 |
99.8 |
99.0 |
100.2 |
99.9 |
|
The stability of desorbed samples was investigated by
reanalyzing the target concentration samples 2.5 h after initial
analysis. After the original analysis was performed, three vials
were recapped with new septa while the remaining three retained
their punctured septa. The samples vials were stored in the
refrigerated sampling tray for the GC injector. The samples were
eanalyzed with fresh standards. The average percent change was
+1.1% for samples that were resealed with new septa, and -1.6% for
those that retained their punctured septa.
Table 4.10.3.2.2 |
Stability of Desorbed Samples for
Halothane from Anasorb CMS |
|
punctured septa
replaced |
punctured septa
retained |
initial |
DE after |
|
initial |
DE after |
|
DE |
one day |
difference |
DE |
one day |
difference |
(%) |
(%) |
|
(%) |
(%) |
|
|
98.7 |
100.9 |
+2.2 |
100.2 |
99.3 |
-0.9 |
99.5 |
99.5 |
0 |
102.4 |
99.2 |
-3.2 |
99.1 |
100.1 |
+1.0 |
101.1 |
100.4 |
-0.7 |
|
(averages) |
|
|
(averages) |
|
99.1 |
100.2 |
+1.1 |
101.2 |
99.6 |
-1.6 |
|
4.10.3.3 Isoflurane
The desorption efficiencies (DE) of isoflurane were determined
by liquid-spiking 150-mg portions of Anasorb CMS with
amounts equivalent to 0.05 to 2 times the 1-ppm
target concentration. These samples were stored overnight at
ambient temperature and then desorbed and analyzed. The average
desorption efficiency over the working range of 0.5 to 2 times the
target concentration is 99.4%.
Table 4.10.3.3.1 |
Desorption Efficiency of
Enflurane from Anasorb CMS at Low TC |
|
× target concn |
0.05× |
0.1× |
0.2× |
0.5× |
1.0× |
2.0× |
(µg/sample) |
4.50 |
9.00 |
18.0 |
45.0 |
90.0 |
183.0 |
|
DE% |
98.6 |
99.3 |
100.2 |
100.3 |
96.9 |
98.6 |
|
99.3 |
98.1 |
98.1 |
98.9 |
97.6 |
100.2 |
|
99.1 |
98.5 |
101.1 |
101.3 |
97.1 |
101.5 |
|
99.0 |
98.3 |
100.1 |
99.6 |
97.9 |
98.7 |
|
100.1 |
97.2 |
98.5 |
100.2 |
98.5 |
101.6 |
|
99.5 |
99.0 |
101.4 |
100.6 |
98.4 |
100.4 |
|
99.3 |
98.4 |
99.9 |
100.2 |
97.7 |
100.2 |
|
The stability of desorbed samples was investigated by
reanalyzing the target concentration samples 22.5 h after initial
analysis. After the original analysis was perfomned, three vials
were recapped with new septa while the remaining three retained
their punctured septa. The samples vials were stored in the
refrigerated sampling tray for the GC injector. The samples were
reanalyzed with fresh standards. The average percent change was
+3.6% for samples that were resealed with new septa, and +1.9% for
those that retained their punctured septa.
Table 4.10.3.3.2 |
Stability of Desorbed Samples for
Enflurane from Anasorb CMS |
|
punctured septa
replaced |
punctured septa
retained |
initial |
DE after |
|
initial |
DE after |
|
DE |
one day |
difference |
DE |
one day |
difference |
(%) |
(%) |
|
(%) |
(%) |
|
|
96.9 |
101.6 |
+4.7 |
97.9 |
97.1 |
-0.8 |
97.6 |
99.8 |
+2.2 |
98.5 |
101.1 |
+2.6 |
97.1 |
101.1 |
+4.0 |
98.4 |
102.5 |
+4.1 |
(averages) |
(averages) |
97.2 |
100.8 |
+3.8 |
98.3 |
100.2 |
+1.9 |
|
4.10.4 Anasorb 747 at low target concentration (TC)
4.10.4.1 Enflurane
The desorption efficiencies (DE) of enflurane were determined
by liquid-spiking 140-mg portions of Anasorb 747 with
amounts equivalent to 0.05 to 2 times the 1-ppm
target concentration. These samples were stored overnight at
ambient temperature and then desorbed and analyzed. The average
desorption efficiency over the working range of 0.5 to 2 times the
target concentration is 103.7%.
Table 4.10.4.1.1 |
Desorption Efficiency of
Enflurane from Anasorb 747 at Low TC |
|
× target concn |
0.05× |
0.1× |
0.2× |
0.5× |
1.0× |
2.0× |
(µg/sample) |
4.56 |
9.12 |
18.24 |
45.6 |
91.2 |
182.4 |
|
DE (%) |
102.0 |
100.5 |
99.1 |
103.2 |
102.3 |
101.9 |
|
102.2 |
100.2 |
98.4 |
102.3 |
104.2 |
105.3 |
|
96.9 |
97.9 |
99.6 |
102.2 |
101.6 |
103.7 |
|
102.4 |
98.2 |
98.2 |
100.9 |
104.6 |
103.8 |
|
101.2 |
97.7 |
99.7 |
107.8 |
106.3 |
103.6 |
|
102.9 |
99.4 |
99.0 |
107.4 |
100.6 |
104.6 |
|
101.3 |
99.0 |
99.0 |
104.0 |
103.3 |
103.8 |
|
The stability of desorbed samples was investigated by
reanalyzing the target concentration samples 22.5 h after initial
analysis. After the original analysis was performed, three vials
were recapped with new septa while the remaining three retained
their punctured septa. The samples vials were stored in the
refrigerated sampling tray for the GC injector. The samples were
reanalyzed with fresh standards. The average percent change was
-2.1% for samples that were resealed with new septa, and -2.9% for
those that retained their punctured septa.
Table 4.10.4.1.2 |
Stability of Desorbed Samples for
Enflurane from Anasorb 747 |
|
punctured septa
replaced |
punctured septa
retained |
initial |
DE after |
|
initial |
DE after |
|
DE |
one day |
difference |
DE |
one day |
difference |
(%) |
(%) |
|
(%) |
(%) |
|
|
102.3 |
98.4 |
-3.9 |
104.6 |
99.4 |
-5.2 |
104.2 |
100.9 |
-3.3 |
106.3 |
101.9 |
-4.4 |
101.6 |
102.6 |
+1.0 |
100.6 |
101.4 |
+0.8 |
|
(averages) |
|
|
(averages) |
|
102.7 |
100.6 |
-2.1 |
103.8 |
100.9 |
-2.9 |
|
4.10.4.2 Halothane
The desorption efficiencies (DE) of halothane were determined
by liquid-spiking 140-mg portions of Anasorb 747 with
amounts equivalent to 0.05 to 2 times the1-ppm target
concentration. These samples were stored overnight at ambient
temperature and then desorbed and analyzed. The average desorption
efficiency over the working range of 0.5 to 2 times the target
concentration is 99.6%.
Table 4.10.4.2.1 |
Desorption Efficiency of
Halothane from Anasorb 747 at Low TC |
|
× target concn |
0.05× |
0.1× |
0.2× |
0.5× |
1.0× |
2.0× |
(µg/sample) |
4.862 |
9.724 |
19.45 |
48.62 |
97.24 |
194.5 |
|
DE(%) |
85.2 |
95.7 |
94.8 |
98.0 |
98.3 |
101.0 |
|
84.5 |
93.3 |
97.2 |
98.3 |
100.2 |
100.9 |
|
85.2 |
92.7 |
94.3 |
99.7 |
97.9 |
99.0 |
|
83.1 |
91.7 |
94.2 |
99.8 |
100.9 |
98.9 |
|
84.5 |
90.7 |
93.7 |
99.7 |
101.6 |
99.1 |
|
83.5 |
91.4 |
94.2 |
98.9 |
100.7 |
100.5 |
|
84.3 |
92.6 |
94.7 |
99.1 |
99.9 |
99.9 |
|
The stability of desorbed samples was investigated by
reanalyzing the target concentration samples 22.5 h after initial
analysis. After the original analysis was performed, three vials
were recapped with new septa while the remaining three retained
their punctured septa. The samples vials were stored in the
refrigerated sampling tray for the GC injector. The samples were
reanalyzed with fresh standards. The average percent change was
±0% for samples that were resealed with new septa, and -3.0% for
those that retained their punctured septa.
Table 4.10.4.2.2 |
Stability of Desorbed Samplies
for Halothane from Anasorb 747 |
|
punctured septa
replaced |
punctured septa
retained |
initial |
DE after |
|
initial |
DE after |
|
DE |
one day |
difference |
DE |
one day |
difference |
(%) |
(%) |
|
(%) |
(%) |
|
|
98.3 |
98.4 |
+0.1 |
100.9 |
97.6 |
-3.3 |
100.2 |
98.1 |
-2.1 |
101.6 |
97.8 |
-3.8 |
97.9 |
99.9 |
+2.0 |
100.7 |
98.9 |
-1.8 |
|
(averages) |
|
|
(averages) |
|
98.8 |
98.8 |
±0 |
101.1 |
98.1 |
-3.0 |
|
4.10.4.3 Isoflurane
The desorption efficiencies (DE) of isoflurane were determined
by liquid-spiking 140-mg portions of Anasorb 747 with
amounts equivalent to 0.05 to 2 times the 1-ppm
target concentration. These samples were stored overnight at
ambient temperature and then desorbed and analyzed. The average
desorption efficiency over the working range of 0.5 to 2 times the
target concentration is 100.8%.
Table 4.10.4.3.1 |
Desorption Efficiency of
Isoflurane from Anasorb 747 at Low TC |
|
× target concn |
0.05× |
0.1× |
0.2× |
0.5× |
1.0× |
2.0× |
(µg/sample) |
4.50 |
9.00 |
18.0 |
45.0 |
90.0 |
180.0 |
|
DE (%) |
97.1 |
100.0 |
98.1 |
103.2 |
99.3 |
99.4 |
|
92.3 |
99.6 |
102.2 |
97.6 |
99.9 |
102.8 |
|
96.4 |
100.2 |
101.1 |
99.4 |
99.2 |
100.8 |
|
97.8 |
100.0 |
102.5 |
98.5 |
101.5 |
101.7 |
|
99.4 |
100.0 |
101.0 |
104.0 |
102.9 |
100.8 |
|
97.9 |
100.1 |
101.4 |
103.8 |
97.5 |
101.5 |
|
96.8 |
100.0 |
101.1 |
101.1 |
100.1 |
101.2 |
|
The stability of desorbed samples was investigated by
reanalyzing the target concentration samples 22.5 h after initial
analysis. After the original analysis was performed, three vials
were recapped with new septa while the remaining three retained
their punctured septa. The samples vials were stored in the
refrigerated sampling tray for the GC injector. The samples were
reanalyzed with fresh standards. The average percent change was
-1.3% for samples that were resealed with new septa, and -4.4% for
those that retained their punctured septa.
Table 4.10.4.3.2 |
Stability of Desorbed Samples for
Isoflurane from Anasorb 747 |
|
punctured septa
replaced |
punctured septa
retained |
initial |
DE after |
|
initial |
DE after |
|
DE |
one day |
difference |
DE |
one day |
difference |
(%) |
(%) |
|
(%) |
(%) |
|
|
99.3 |
97.4 |
-1.9 |
101.5 |
96.8 |
-4.7 |
99.9 |
97.0 |
-2.9 |
102.9 |
95.0 |
-7.9 |
99.2 |
100.2 |
+1.0 |
97.5 |
96.8 |
-0.7 |
|
(averages) |
|
|
(averages) |
|
99.5 |
98.2 |
-1.3 |
100.6 |
96.2 |
-4.4 |
|
4.11 Qualitative analysis
The anesthetic gases can be easily separated and identified by
GCA\MS. Mass spectra for enflurane, halothane and isoflurane, which
were separated using conditions similar to the information given in
Section 3.5, were obtained from a Perkin-Elmer Ion Trap
Detector interfaced to a Hewlett-Packard Series II GC.
Figure 4.11.1. Mass spectrum of enflurane.
Figure 4.11.2. Mass spectrum of halothane.
Figure 4.11.3. Mass spectrum of isoflurane.
4.12 Nitrous oxide interference
A test was developed to study the ability of nitrous oxide to
interfere with the collection of anesthetic gases on the recommended
sampling tubes. A 100-L gas-sampling bag was filled with
dry air and 1.50 mL of water was added to raise the humidity to 80% at
22°. Halothane was selected as a typical anesthetic gas and
20µL (37.4 mg) was added to the bag. Nitrous oxide (430.5 mg)
was also added. This produced an atmosphere containing 46.4 ppm of
halothane and 2390 ppm of nitrous oxide. A second bag was prepared to
dupiclate the first one except no nitrous oxide was added. Air samples
were drawn at 0.05 L/min for 4 h from both bags using Anasorb CMS and
Anasorb 747. No halothane was detected on any of the
back-up sections. The results show that nitrous oxide
does not substantially interfere with the collection of halothane from
an atmosphere containing both gases.
Table 4.12 |
Parts-per-million recovered from
Gas-Sampling Bags |
|
Anasorb |
halothane |
halothane with |
|
|
nitrous oxide |
|
CMS |
44.9 |
40.3 |
747 |
45.6 |
45.7 |
|
5. References
5.1 OSHA Analytical Methods Manual, 2nd ed., U.S. Department
of Labor, Occupational Safety and Health Administration; Salt Lake
Technical Center; Salt Lake City, UT 1993; "Method 29 - Enflurane and
Halothane" (1981); American Conference of Governmental Industrial
Hygienists (ACGIH); Cincinnati, OH, Publ. No. 4542.
5.2 NIOSH Criteria for a Recommended Standard: Occupational
Exposure to Waste Anesthetic Gases and Vapors, U.S. Department of
Health and Human Services, Public Health Service, Center for Disease
Control, National Institute for Health for Occupational Safety and
Health, Cincinnati, OH, 1977, DHHS (NIOSH) Publ.
77-140.
5.3 NIOSH Recommendations for Occupational Safety and Health:
Compendium of Policy Documents and Statements, U.S. Department of
Health and Human Services, Public Health Service, Center for Disease
Control, National Institute for Health for Occupational Safety and
Health, Cincinnati, OH, 1992, DHHS (NIOSH) Publ.
92-100.
5.4 Documentation of the Threshold Limit Values and Biological
Exposure Indices, 5th ed., American Conference of Governmental
Industrial Hygienists (ACGIH); Cincinnati, OH, 1986.
5.5 MRC Monograph on the Evaluation of Carcinogenic Risks to
Humans: Overall Evaluation of Carcinogenicity: An Update of IARC
Monographs Volumes 1 to 42, International Agency for Research on
Cancer (IARC), Lyon, France, 1987, Supplement 7, pp.
93-95.
5.6 Material Safety Data Sheet: Ethrane, Anaquest, Liberty Corner,
NJ, March 1992.
5.7 Material Safety Data Sheet: Forane, Anaquest, Liberty Corner,
NJ, March 1992.
5.8 Material Safety Data Sheet
2-Bromo-2-chloro-1,1,1-trifluoroethane, Aldrich Chemical
Co., Milwaukee, WI, May 1992.
5.9 Merck Index, Budavari, S. Ed., 11th ed., Merck &
Co., Rahway, NJ, 1989.
5.10 OSHA Instruction CPL 2-2.60, Exposure Control
Plan for Federal OSHA Personnel with Occupational Exposure to
Bloodbome Pathogens, March 7,1994; Occupational Safety and Heath
Administration, U.S. Department of Labor, Washington, D.C.
|