ETHYLENE OXIDE
Method no.: |
50 |
|
Matrix: |
Air |
|
Target concentration: |
1 ppm (1.8 mg/m3) |
|
Procedure: |
Samples are collected by drawing a known volume of
air through hydrobromic acid-coated charcoal tubes to produce
2-bromoethanol. Following desorption with
dimethylformamide (DMF), an aliquot of the sample is derivatized to
its heptafluorobutyrylimidazole (HFBI), and analyzed by gas
chromatography using electron capture detection. |
|
Recommended air volume and sampling rate: |
24 L at 0.1 L/min |
|
Reliable quantitation limit: |
3.0 ppb (5.4 µg/m3) |
|
Standard error of estimate: (Figure 4.5.1.) |
6.3% |
|
Special requirements: |
Sampling tubes currently must be obtained from the
laboratory. Collected samples should be stored at reduced
temperature to minimize possible storage losses. (Section
4.5.) |
|
Status of method: |
Evaluated method. This method has been subjected to
the established evaluation procedures of the Organic Methods
Evaluation Branch. |
|
Date: January 1985 |
Chemist: Kevin J.
Cummins |
Organic Methods Evaluation Branch OSHA Analytical
Laboratory Salt Lake City, Utah
1. General Discussion
1.1. Background
1.1.1. History
A number of different methods are currently in use for monitoring
ethylene oxide (EtO) exposures in the workplace. These consist of
several continuous monitoring devices, a number of active sampling
devices, and at least two commercial passive monitoring devices
which are designed specifically for monitoring EtO. The continuous
monitoring devices, despite their ability to instantly measure EtO
exposures, in some cases lack sensitivity (IR monitor), and in other
cases are subject to interferences (photo-ionization).
Currently the most widely accepted active sampling device for
measuring EtO exposures uses a large charcoal tube to trap EtO
vapors (Qazi-Ketcham) (Ref. 5.1.). While this sampling method can
give reliable results with careful attention to details, the
inherent instability of EtO on charcoal can present serious
problems. In addition, the method is not as sensitive as may be
necessary for measuring sub-ppm exposure levels.
Acid bubblers have also been used to sample EtO in the work
environment. Analysis of the resulting ethylene glycol is performed
colorimetrically, or more recently, by gas chromatography (Ref.
5.2.). This method is inconvenient to use in the field, and may lack
adequate sensitivity to monitor at the new OSHA PEL of 1 ppm (8-h
TWA) and at the 0.5 ppm action level (8-h TWA).
Both 3M and DuPont currently manufacture passive monitoring
devices for sampling EtO. These sampling devices do not require a
sampling pump since diffusion principles determine the sampling
rate. Both of these devices require adequate air movement to ensure
that the air around the sampling device is not depleted of EtO. The
3M badge converts EtO to 2-bromoethanol in a manner
similar to the procedure described in this method with subsequent
analysis by gas chromatography. The DuPont monitor uses a small
pouch containing aqueous acid solution to trap EtO as ethylene
glycol and the analysis is by colorimetry.
The current OSHA method uses two standard size charcoal tubes in
series to trap EtO directly on the charcoal surface (Ref. 5.3.).
Because of the low capacity of charcoal for EtO, the recommended air
sample volume is limited to 1 L to avoid breakthrough. Samples are
recommended to be analyzed within 15 days of collection to minimize
sample loss upon storage. This evaluation was undertaken to improve
the sampling capability and to reduce possible storage losses of
EtO. The procedure described in this method uses a hydrobromic
acid-coated sampling tube to collect EtO as its
2-bromoethanol reaction product. The reaction is fast
and complete and produces a stable product which has a high affinity
for charcoal. This reaction scheme was first reported for the
determination of EtO residues in medical devices using a purge and
trap technique (Ref. 5.4.). The current OSHA method uses this same
reaction for the analysis of EtO air samples collected on
conventional charcoal tubes. (Ref. 5.3.) The direct conversion of
EtO to 2-bromoethanol on the acid-charcoal surface
offers a convenient means by which greater sample capacity and
stability are obtained over the conventional charcoal tube
collection method. The 5% breakthrough volume for sampling a 16-ppm
atmosphere of EtO at 0.15 L/min with the acid-coated tube is 39 L.
The 5% breakthrough volume for the current OSHA method, determined
by sampling a 2-ppm test atmosphere at 0.05 L/min, is only 2.6 L.
Sample stability with storage is also improved with the acid-coated
charcoal tube. No significant storage effects are observed for
samples collected in the 0.1- to 16-ppm range at high humidity and
stored at ambient temperature for a minimum of 2 weeks.
Efforts to analyze directly for the 2-bromoethanol
product collected on the acid-coated charcoal tube were unsuccessful
by gas chromatography with electron capture detection since the acid
matrix of the sample gave a non-reproducible detector response. This
problem was alleviated by first derivatizing an aliquot of the
sample with HFBI in isooctane to 2-bromoethyl heptafluorobutyrate.
Following hydrolysis of the excess reagent with water, the ester
product in the isooctane layer was analyzed by GC with electron
capture detection.
The method described in this procedure has been shown in
laboratory studies and in field study to provide a reliable,
convenient, and accurate means of measuring EtO exposures. This
method has been tested under a broad range of conditions in the
laboratory. Test atmospheres of 0.1, 0.5, 1.0, and 16 ppm at 70-80%
relative humidity and ambient temperature were sampled for 4 h at
0.1 L/min with no breakthrough. Average percent recoveries of 96.8,
94.6, 90.6, and 90.4 respectively were obtained upon same-day
analysis, and average percent recoveries of 102, 92.3, 84.6, and
85.6, respectively were obtained upon analysis for the same test
atmospheres after storage for a minimum of 2 weeks. (Table 4.9.)
The effects of storage on sample stability appear to be minor for
samples collected for 2 h at 0.1 L/min from a 2-ppm test atmosphere
at 80% R.H. and ambient temperature. (Figures 4.5.1. and 4.5.2.) The
average recovery of 90.2% was obtained upon storage of the samples
for 17 days at ambient temperature. (Section 4.5.)
The effects of low humidity on the sampling method were also
evaluated. Samples collected at 0.1 L/min from a 2-ppm test
atmosphere for 2 h at <5% R.H. and ambient temperature result in
high initial recoveries, but demonstrate a statistically valid
decrease in recovery with ambient storage. (Table 4.10.) Although
this effect is not understood, the recoveries still remain above 75%
after storage. This effect was not observed with high humidity as
described earlier.
The method can be used to accurately monitor short-term exposures
in the workplace. Fifteen-minute and 30-min air samples were
collected from a constant 5-ppm test atmosphere (80% R.H. and
ambient temperature) at 0.1 L/min. The results were reported in
Table 4.11. The high recoveries obtained for these samples indicate
that the sampling tube can be used to effectively monitor short-term
exposures.
The sampling tube can be used to accurately measure transient,
high exposures to EtO which frequently occur in hospital
sterilization facilities. Sample tubes which were spiked with 540 µg
of pure EtO gas, from a gas-tight syringe over a 30-s time period
(equivalent to a 6000 ppm exposure for 30 s) either before or after
24 L of EtO-free air at 80% R.H. and ambient temperature were drawn
through them at 0.1 L/min, resulted in average recoveries of 102%
and 105% respectively upon analysis the same day.
Field comparison samples were collected at a local hospital using
this method and the large charcoal tube method (Qazi-Ketcham). A
total of 15 pairs of samples were obtained from four separate
inspections by monitoring various areas of the hospital's
sterilization facility with side-by-side area samples. No
statistical difference in the two methods was observed over a range
of exposures of 0.3 to 7 ppm EtO measured at the site. (Section
4.8.)
Based on the results of these laboratory studies, and on the
excellent field comparison data, it is anticipated that this
sampling and analytical method will offer a reliable, accurate, and
convenient means of monitoring EtO exposures in the workplace.
1.1.2. Toxic effects. (This section is for information only and
should not be taken as the basis of OSHA policy.)
It has long been recognized that exposure to high levels of EtO
can cause a variety of toxic effects including respiratory tract,
eye, and skin irritations, nausea, vomiting, central nervous system
depression, and even death. Intermittent high exposures are also
reported to cause neurological effects such as sensory motor
neuropathies and seizures. (Ref. 5.5.)
Recently major concern has focused on the mutagenic and
carcinogenic properties exhibited by EtO. Quoting the June 22, 1984
publication of the new EtO standard from the Federal Register:
The evidence suggests that EtO may cause cancers of the blood
(leukemia) as well as other organs in humans. In addition EtO
exposure causes mutations, increases the rate of chromosomal
aberration and sister chromatid exchange, and causes other
undesirable changes in the DNA of mammalian cells... EtO exposure
has also been associated with an increased risk of spontaneous
abortion among pregnant women and is capable of causing other
adverse reproductive effects in both men and women.
These conclusions are supported by animal experiments involving
exposure to EtO by a variety of different routes including:
inhalation, sub-cutaneous injection, dermal exposure, and
intragastric administration. Additional data are cited in the
publication of the new standard to indicate that "virtually every
mutagenicity test system applied to EtO has shown the chemical to be
mutagenic". Several epidemiological studies are also cited in the
Federal Register which indicate that excess cancers may be occurring
in the workplace due to EtO exposure. Additional studies involving
groups of workers exposed to varying levels of EtO are cited which
indicate an EtO dose-dependent increase in sister chromatid exchange
rates and increased chromosomal breaks and aberrations. Based on
this increasing body of evidence, OSHA has acted to reduce the
current permissible exposure limit (PEL) from 50 ppm to 1 ppm for an
8-h time weighted average exposure. (Ref. 5.6.)
1.1.3. Potential workplace exposure
EtO is a major industrial chemical with production volume ranking
in the top 25 among all chemicals produced in the United States.
Approximately 6.7 billion lbs. of EtO were produced domestically by
the most recent estimate. (Ref. 5.6.) Over 99% of the total EtO
produced in the United States is used in the manufacture of other
products. Approximately 70% of the total is used to produce ethylene
glycol. EtO is also used to produce non-ionic surface-active agents
(which are used in household detergents), ethanolamines, glycol
ethers, di-, tri-, tetra-, polyethylene glycols, and crown ether
compounds. (Ref. 5.5)
Although less than 1% of the total EtO produced in the United
States is used as a sterilizing agent, this small sector represents
the greatest number of potential work exposures. It is estimated
that some 62,370 employees in 6,237 hospitals in the U.S. are
potentially exposed to EtO. Another 5,000 workers are estimated to
be exposed to EtO in its use as a sterilizing agent in the medical
products manufacturing industry. (Ref. 5.6.)
A small number of workers in other industries are also
potentially exposed to EtO with its use as a fumigant and a
sterilizing agent. Spice manufacturing, libraries, museums, dairy
packing, and fur treating are some of the industries and work
settings in which EtO exposure can occur. (Ref. 5.5.)
1.1.4. Physical properties (Ref. 5.7. unless otherwise noted)
CAS no.: |
75-21-8 |
molecular weight: |
44.05 |
boiling point: |
10.4°C at 760 mm Hg |
color: |
colorless gas |
density: |
0.8697 g/mL at 20°C |
molecular formula: |
C2H4O |
vapor pressure: |
1094 mm Hg at 20°C |
flash point (tag open cup): |
<-18°C |
odor: |
ether-like (Ref. 5.3.) |
explosive limits in air: |
upper, 100 % by volume lower, 3 % by
volume |
synonyms (Ref. 5.6.): |
dimethylene oxide;
1,2-epoxyethane; EtO; EO; oxirane;
oxacyclopropane; dihydrooxirene |
1.2 Limit Defining Parameters (The analyte air concentration listed
throughout this method are based on a 24-L air sample
unless otherwise noted. Amounts are expressed as the equivalent weight
of EtO, although the 2-bromoethyl heptafluorobutyrate ester is
analyzed.)
1.2.1. Detection limit of the analytical procedure
The detection limit of the analytical procedure is 0.29 pg per
injection. This is the amount of analyte which will give a
measurable response with the amounts of interferences present in a
standard. (Section 4.1.)
1.2.2. Detection limit of the overall procedure
The detection limit of the overall procedure is 0.14 µg per
sample (3.0 ppb or 5.4 µg/m3). This is the
amount of analyte spiked on the sampling device which allows
recovery approximately equivalent to the detection limits of the
analytical procedure. (Section 4.2.)
1.2.3. Reliable quantitation limit
The reliable quantitation limit is 0.14 µg per sample (3.0 ppb or
5.4 µg/m3). This is the amount of analyte
which can be quantitated within the requirements of a recovery of at
least 75% and a precision (±1.96 SD) of ±25% or better. (Section
4.2.)
The reliable quantitation limit and detection limits reported in
this method are based upon optimization of the instrument for the
smallest possible amount of analyte. When the target concentration of
an analyte is exceptionally higher than these limits, they may not be
attainable at the routine operating parameters.
1.2.4. Sensitivity
The sensitivity of the analytical procedure over the
concentration range representing 0.5 to 2 times the target
concentration based on a 24-L air sample is
approximately 10,600 area units per µg/sample. This is determined by
the slope of the calibration curve. (Section 4.4.) The sensitivity
will vary with the particular instrument used in the analysis.
1.2.5. Recovery
The recoveries of EtO from samples used in a 17-day storage test
when the samples were stored at ambient conditions in the dark was
90.2%. This is the percent recovery at 17 days determined from the
linear least squares line from the storage data. (Section 4.5.)
1.2.6. Precision (analytical method only)
The pooled coefficient of variation obtained from replicate
determinations of analytical standards at 0.5, 1, and 2 times the
target concentration is 0.028. (section 4.3.)
1.2.7. Precision (overall procedure)
The precision at the 95% confidence level for the 17-day storage
test is ±13%. (Figure 4.5.2.) This includes an additional ±5% for
sampling error. The overall procedure must provide results that are
±25% or better at the 95% confidence level.
1.2.8. Reproducibility
Six samples taken from a controlled test atmosphere and a draft
copy of this procedure were given to a chemist unassociated with
this evaluation. The samples were analyzed after 14 days of storage
at 5°C. The average recovery was 84.8% with a standard deviation of
±2.1%. (Section 4.6.)
1.3 Advantages
1.3.1. The acid-coated sampling tube is convenient to use and
requires no special shipping or storage requirements.
1.3.2. This sampling method allows for a much longer sampling
period than the current OSHA sampling method. Only one or two
acid-coated sample tubes are needed to monitor an 8-h exposure.
1.3.3. The analytical method is more sensitive than direct
analysis by flame ionization detection.
1.4. Disadvantages
1.4.1. At this time the sampling tubes are not commercially
available and must be obtained from the laboratory.
1.4.2. The analysis involves the formation of a derivative of
2-bromoethanol which is more time consuming than direct
analysis.
2. Sampling Procedure
2.1. Apparatus
2.1.1. A constant flow personal sampling pump is used which can
be calibrated to within ±5% of the recommended 0.1 L/min flow rate
while the sampling train is in line.
2.1.2. The sampling tube consists of a 6-mm o.d. × 4-mm i.d. ×
45-mm glass tube packed with two sections of 24% by weight
hydrobromic acid-coated charcoal. These tubes are made from used,
clean, sampling tubes, which have had one end of the tube removed.
The open end of the tube is fire polished prior to use. The front
and back sections contain 100 and 50 mg of the coated charcoal
respectively, and are separated and contained within the tube with
silanized glass wool plugs.
2.1.3. The coated charcoal is prepared by slowly adding a mixture
of 25 mL of hydrobromic acid (48% aqueous, Alfa Products, Thiokol,
Inc. Danvers, MA) and 125 mL of acetonitrile (Burdick and Jackson,
Inc., Muskegon, MI) to 75 grams of lot 208 petroleum base charcoal
(SKC Inc., Eighty-four, Pa) contained in a 500-mL round bottom
flask. After allowing the slurry to cool to room temperature, the
charcoal is dried by rotary evaporation using gentle heat, and kept
overnight under vacuum at ambient temperature. This coated charcoal
is stable for at least 4 months when stored in a tightly sealed
amber glass jar at room temperature.
2.2. Reagents
None required
2.3. Technique
2.3.1. Properly label the sampling tube before sampling.
2.3.2. Attach the sampling tube to the pump using a section of
flexible, plastic tubing such that the large, front section of the
sample tube is exposed directly to the atmosphere. Do not place any
tubing in front of the sampling tube. The sampling tube should be
attached in the worker's breathing zone vertically such that it does
not impede work performance.
2.3.3. After sampling for the appropriate time, remove the
sampling tube from the pump, replace the plastic caps, and seal the
tube with an official OSHA seal (Form 21).
2.3.4. Include at least one blank for each sampling set. The
blank should be handled in the same manner as the samples with the
exception that air is not drawn through it.
2.3.5. List any potential interferences on the sample data sheet.
2.4. Breakthrough
Breakthrough studies were performed by sampling a 16 ppm atmosphere
at 70% R.H. and ambient temperature at 0.15 L/min with a sampling tube
containing a 100-mg front section of acid-coated charcoal. A second
tube, containing a similar section of coated charcoal was attached
behind the front section to monitor breakthrough. The backup sections
were periodically changed and analyzed while the atmosphere was being
sampled until breakthrough was observed. The 5% breakthrough volume,
that is, the volume of air sampled that results in a concentration of
EtO downstream from the sampling tube that is 5% of the upstream
concentration, is approximately 39 L. (Figure 2.4.)
2.5. Desorption efficiency
The average percent recovery of EtO form the acid-coated charcoal
was determined both with pure EtO gas spikes and with liquid spikes of
an equivalent weight of 2-bromoethanol in acetonitrile.
The average percent recovery over a range equivalent to 0.5 to 2.0
times the 1-ppm PEL for a 24-L air sample was 93.3%
relative to controls for the gas-spiked samples, and 99.8% for the
2-bromoethanol spiked samples over the same equivalent
range. Based on the results for the liquid-spiked samples no
desorption efficiency correction factor was applied to any of the
results reported in this method. The minor discrepancy in recovery for
the gas-spiked samples is not understood. (Section 4.7.) It will be
necessary to check the desorption efficiency for each new lot of
acid-coated charcoal using liquid injections of a standard of
2-bromoethanol diluted with acetonitrile.
2.6. Recommended air volume and sampling rate
A 24-L air sample obtained by sampling for 4 h at 0.1
L/min is recommended for EtO. The sensitivity of the method will
permit much shorter sampling periods at this same sampling rate. The
method can reliably measure a 5-ppm test atmosphere with a 15-min
sampling period (Table 4.11.).
2.7. Interferences
There are no known interferences to the sampling procedure.
2.8. Safety precautions
2.8.1. Attach the sampling equipment to the worker in such a
manner that it will not interfere with work performance or safety.
2.8.2. Follow all safety practices that apply to the work area
being sampled.
3. Analytical Procedure
3.1. Apparatus
3.1.1. Gas chromatograph equipped with an electron capture
detector is needed for the analysis of the 2-bromoethyl
heptafluorobutyrate ester. Both a Hewlett-Packard model HP5730A
(Palo Alto, CA) gas chromatograph equipped with an autosampler, and
a Tracor model 222 gas chromatograph (Austin, TX) were used in this
study.
3.1.2. An electronic integrator or other suitable means of
measuring detector response is needed. A Hewlett-Packard 3357 data
system was used in this study.
3.1.3. Small screw-cap vials fitted with Teflon-coated septa are
needed for the preparation of samples and standards. Waters Inc.
WISP-type vials (Sun Brokers Inc., Wilmington, NC) were used in this
study.
3.1.4. Two repetitive, 1-mL solvent dispensers are used for
dispensing DMF and isooctane directly from the solvent bottle. L/I
Repipet dispensers (Lab Industries, Berkeley, CA) were used in this
study.
3.1.5. Precision 1-, 2-, and 10-µL syringes are needed for
preparation of standards, sample and standard transfers, and GC
analysis.
3.2. Reagents
3.2.1. Dimethylformamide, Burdick and Jackson (Muskegon, MI).
3.2.2. Isooctane, Fisher HPLC Grade (Fairlawn, NJ).
3.2.3. n-Heptafluorobutyrylimidazole (HFBI), Pierce Chemical Co.
(Rockford, IL).
3.2.4. High purity water, Milli-Q filtered water Millipore Inc.
(Bedford, MA).
3.2.5. 2-Bromoethanol, 98% pure, Eastman-Kodak (Rochester, NY).
3.2.6. Anhydrous magnesium sulfate, Baker reagent-grade
(Phillipsburg, NJ).
3.3. Standard preparation
A stock solution of 2-bromoethanol is prepared by
weighing 1 mL of the standard into a clean, dry 10-mL volumetric flask
and diluting to volume with DMF. Dilutions of this stock give standard
solutions from which working standards are prepared. Working standards
are prepared by injecting a series of µL volumes of the above
standards into 4-mL screw-capped vials fitted with septa and
containing 1 mL of DMF. A sample calculation with 98% pure
2-bromoethanol standard expressed as its equivalent
weight in EtO is shown below:
1.7308 grams
10 mL |
× 0.98 × |
44.05
124.97 |
= 59.79 mg/mL as
EtO |
(44.05 and 124.97 are the molecular weights of EtO and
2-bromoethanol respectively).
3/25 dilution = 7174.5 µg/mL as EtO 1/10 dilution = 717.45 µg/mL
as EtO
Injections of 2.5 and 10 µL of 717.45 µg/mL standard, and
injections of 2.5, 5.0, 10, and 15 µL of 7174.5 µg/mL standard into
separate vials containing 1 mL of DMF produce the following working
standards: 1.79, 7.10, 17.9, 35.7, 71.0, and 106 µg as EtO. These
standards, along with the desorbed samples are derivatized as
described in Section 3.5.
3.4. Sample preparation
The front acid-coated charcoal section with the glass wool plug,
and the back section with the remaining two glass wool plugs are
transferred to separate 4-mL screw-cap vials. One milliliter of DMF is
then added to each vial and the vials are capped and vigorously shaken
for 5-10 s to ensure adequate desorption. The vials are then allowed
to sit for a minimum of 5 min prior to derivatization.
3.5. Derivatization of samples and standards
Ten-microliter aliquots of each working standard and of each sample
solution in DMF are spiked into separate screw-capped vials containing
1 mL of isooctane and 20 µL of HFBI derivatizing agent. The
septum-capped vials are then briefly shaken to ensure mixing, and
allowed to sit at room temperature for a minimum of 5 min. One
milliliter of filtered water is then pipetted into each vial, and the
vials are capped and shaken vigorously for several seconds to ensure
complete hydrolysis of the excess derivatizing agent.
3.6. Analysis
3.6.1. Analysis of the heptafluorobutyrate ester of
2-bromoethanol in isooctane can be performed either
manually, by direct 2-µL injections of the isooctane layer onto the
GC, or by automated analysis by first transferring the isooctane
layer into autosampler vials containing approximately 50 mg of
anhydrous magnesium sulfate.
3.6.2. Gas chromatographic conditions are listed below:
Manual GC (Tracor 222) |
column: |
6-ft × 2-mm i.d., glass column packed with 10%
SP 1000 on 80/100 Supelcoport (Supelco, Inc., Bellefonte,
PA) |
carrier gas: |
argon/methane (95/5) |
flow rate: |
15 mL/min |
purge rate: |
25 mL/min |
inlet/oven/ |
|
detector temp: |
260/85/305 (°C) |
injection volume: |
2 µL |
retention time: |
6.0 min |
|
Automated GC (HP 5730A) |
column: |
10-ft × 1/8-in o.d. stainless steel column
packed with 10% SP 1000 on 80/100 Supelcoport |
carrier gas: |
argon/methane (95/5) |
flow rate: |
20 mL/min |
inlet/oven/ |
|
detector temp: |
200/100/300 (°C) |
injection volume: |
0.4 µL |
retention time: |
7.2. min |
3.6.3. Chromatograms of standard and of an actual field sample
are shown in figures 3.6.1. and 3.6.2.
3.6.4. Both the front and back sections of all samples are
analyzed to ensure that no sample breakthrough has occurred. In the
event that a number of samples exceed the range of the working
standards prepared for the analysis, it is advisable to prepare
additional standards in order to ensure that all sample responses
fall within the range of the standard curve. For an occasional high
sample result, dilution of the original sample with DMF and
rederivatization is an appropriate means of analyzing the sample.
Re-analysis of samples at a later date is possible if
the original sample in DMF is kept stored in a freezer. No
significant change in sample results has been observed for samples
stored in this manner and reanalyzed several months later.
3.7. Interferences
No significant interferences to this analysis have bee observed
during the course of this study. Methanol, ethanol, n-propanol,
2-chloroethanol, ethylene glycol, and n-butanol, all of which form
esters with HFBI, and are chromatographed under the existing
conditions, are not interferences. In the event that an interference
is observed, selection of alternative GC conditions will be necessary.
Confirmation of the derivative by GC/MS is a highly use full means of
compound identification.
3.8. Calculations
3.8.1. A calibration curve is prepared by plotting µg of EtO per
sample versus area response. A least squares fit of a parabolic
curve through zero was used to obtain the best fit of the data since
the ECD response was not entirely linear.
3.8.2. The amount of EtO found on both the front and back
sections of the sample tube are added together and the resulting air
concentration is reported in ppm (at 760 mm Hg, 25°C) using the
following formula:
ppm = |
µg EtO/sample
L of air sampled |
× |
24.46
44.05 |
where |
24.46 is the molar volume at 760 mm and
25°C. 44.05 is the molecular weight of
EtO. |
3.9. Safety precautions
3.9.1. Minimize exposure to all reagents and solvents by
performing all sample and standard preparations in a well-ventilated
hood.
3.9.2. Avoid skin contact with all solvents and reagents.
3.9.3. Wear safety glasses in the laboratory at all times.
4. Backup Data (All data reported in this section were determined
using the HP 5730A gas chromatograph equipped with an autosampler.)
4.1. Detection limit of the analytical procedure
The detection limit for the analytical procedure is 0.29 pg per
injection. This is based on a 0.4-µL injection of a 0.72
pg/µL working standard. This is the amount of analyte which will give
a measurable response with the amounts of interferences present in a
standard. (Figure 4.1.)
4.2. Reliable quatitation limit and detection limit of the overall
procedure
The reliable quantitation limit for this method is 0.14 µg per
sampler or 3.0 ppb based on a 24-L air sample. Six vials
containing 100 mg of coated charcoal were each spiked with 1 µL of
143.6 µg/mL of EtO (as 2-bromoethanol in acetonitrile)
and capped and stored at room temperature overnight. The following day
the samples were analyzed and the % recovery is reported in Table 4.2.
Table 4.2. Reliable Quantitation Limit
|
% recovery |
statistics |
|
94.8 98.6 98.6 104.0 104.0 100.0 |
SD 1.96 SD
|
= = =
|
100.0 3.7 7.25 |
|
4.3. Precision of the analytical method
The pooled coefficient of variation for EtO is 0.028 over a range
of 0.5 to 2.0 times the target concentration of 1 ppm. This value was
determined from six injections each of three standards which
correspond to 17.89, 35.69, and 71.03 µg or EtO per sample
respectively.
Table 4.3. Precision of the Analytical Method
|
× target conc. µg/sample |
0.5× 17.89 |
1× 35.69 |
2× 71.03 |
|
area counts
SD CV
= 0.028 |
171863 177031 176889 176924 163554 163743
171667 6516 0.0380 |
384974 377399 371456 369710 394991 393891
382070 10976 0.0287 |
734904 732039 730219 744751 748800 752568
740547 9393 0.0127 |
|
4.4. Sensitivity
The slope of the calibration curve over the range of 0.5 to 2.0
times the target concentration for the analysis represents the
sensitivity for the method. The ECD response is approximately linear
in this region and the slope of the line is approximately 10,600 area
counts per µg EtO per standard (Figure 4.4.).
4.5. Storage
Storage of sample sets over a 17-day period was performed at both
ambient and refrigerated temperatures. The samples were collected by
sampling a 2-ppm test atmosphere at 80% R.H. and ambient temperature
for 2 h at 0.1 L/min. This sample load is equivalent to a 1 ppm, 4-h
exposure. A total of 36 samples were generated in this study. Eighteen
of the samples were collected on one day, and the remaining 18 samples
were collected three days later.
All of the samples generated in this study, with the exception of
six samples generated on the second generation day which were analyzed
without storage, were randomly split into equal sized groups and
stored either at ambient conditions in the dark, or at 5°C in a
refrigerator. Twelve of the samples were analyzed on the second
generation day and these include three samples each from ambient and
refrigerated storage along with six of the samples generated that same
day. The remaining 24 samples were analyzed in groups of 12 at one
week intervals over the next two weeks. Each group of samples
consisted of three samples each from ambient and refrigerted storage
prepated on the two different generation days.
The results of this study are presented in Table 4.5. and in
Figures 4.5.1. and 4.5.2. A slight decrease in recovery is observed
upon storage, although recoveries remain above 90% through the 17 days
stored. It is recommended that samples be stored at reduced
temperature following sampling to minimize any possible losses.
Shipment of samples on dry ice, or other precautionary measures, is
not considered necessary for the samples.
Table 4.5. Storage Tests
|
storage time |
% recovery |
(days) |
(ambient) |
|
(refrigerated) |
|
0 0 3 7 10 14 17 |
92.8 87.3 96.9 89.5 86.5 90.7 90.2 |
89.2 86.9 89.1 97.0 87.5 95.1 82.9 |
96.6 ---- 93.9 91.1 92.0 92.9 92.9 |
|
92.8 87.3 91.3 99.6 94.5 90.7 93.2 |
89.2 86.9 82.6 97.1 98.7 95.1 99.7 |
96.6 ---- 90.0 99.5 96.2 92.9 98.1 |
|
4.6. Reproducibility
Six sample tubes were each spiked with a stock solution of
2-bromoethanol to give an equivalent weight of EtO of
37.7 µg per sample and the samples were then stored in a refrigerator
prior to analysis. The samples were analyzed by a chemist unassociated
with the evaluation. The results are given in Table 4.6.
Table 4.6. Reproducibility
|
% recovery |
statistics |
|
57.51 84.8 84.8 87.8 81.7 84.8 |
SD |
= =
|
84.8 2.1
|
|
1
not used in average |
4.7. Desorption efficiency
4.7.1. The percent recovery of 2-bromoethanol
spiked onto 100-mg sections of coated charcoal at levels
corresponding to 0.5 to 2 times the target concentration for EtO was
determined. A total of 18 samples in three groups of six were used
in this study. Each group of six samples was spiked with either a
2.5-, 5.0-, or a 10-µL injection of 7181 µg/mL EtO in acetonitrile
(2-bromoethanol standard in acetonitrile expressed as
weight equivalent of EtO). The samples were stored overnight and
analyzed the next day. The percent recoveries are reported in Table
4.7.1.
Table 4.7.1. Desorption Efficiency of 2-Bromoethanol
on Acid-Coated Charcoal1
|
Equivalent weight
of EtO spiked (µg) |
% Recovery ± 1
SD |
Relative SD |
|
18.0 35.9 71.8 |
101.3 ±6.2 100.6
±3.1 97.4 ±2.5 |
6.1 2.5 2.6 |
|
1 six samples per data
point |
4.7.2. The desorption efficiency of coated charcoal was also
determined by spiking three groups of six samples with either 12.5,
25.0, or 50 µL of pure EtO gas (Union Carbide, Corp., Linde
Division, N.Y., NY). The samples were spiked while laboratory air
was being drawn through the tubes at 0.1 L/min. Controls were
prepared by spiking identical volumes of EtO gas into sample vials
containing 1 mL of a 1.5% HBr solution in DMF. Under the prevailing
atmospheric conditions, these volumes of EtO corresponded to 19.44,
38.74, and 77.31 µg of EtO respectively. Theses amounts correspond
approximately to 0.5 to 2 times the PEL concentration for a
24-L air sample. The samples were capped and stored
along with the controls for analysis the next day. The average
percent recovery of samples and controls is reported in Table 4.7.2.
The overall average recovery relative to controls is 93.3%.
Table 4.7.2. Desorption Efficiency of EtO from
Gas-Spiked Charcoal Tubes
|
no. of samples |
amount
spiked (µg) |
% recovery ± 1
SD |
% recovery (rel. to
controls) |
|
5 3 (controls) 6 2
(controls) 6 3 (controls) |
19.44 " " 38.74 "
" 77.31 " " |
92.9 ±3.6 102.7
±4.4 92.4 ±1.1 99.9
---- 100.1 ±3.7 104.0 ±1.7 |
90.5 ---- 92.5 ---- 96.9 ---- |
|
4.8. Field comparison data
Side-by-side area samples were collected by Ed Zimowski of the OSHA
Health Response Team at a local hospital sterilization facility using
this method and the large charcoal sampling tube method
(Qazi-Kethcham). A total of 15 pairs of samples were obtained from
four separate inspections by monitoring different areas of the
hospital's sterilization facility. The sampling times varied from 4 to
7.5 h for each comparison sample. The sampling rate for both sampling
tubes was about 0.05 L/min. Two large charcoal tubes were used during
each sampling period and the time-weighted average for the two
compared with the acid-coated tube result. The large charcoal tubes
were kept on dry ice after sampling and stored in a freezer prior to
analysis at the laboratory. The acid-coated tubes were kept at ambient
conditions overnight and then stored in a freezer upon receipt at the
laboratory the next day. Analysis of the large charcoal sampling tubes
was performed by Carl Elskamp of the Methods Evaluation Group. All of
the samples were analyzed within three days of sample collection. The
results for each pair of samples are reported in Table 4.8. The
excellent correlation (R=0.994) with a slope of 0.996 (SD=0.030) and
an intercept of -0.063 (SD = 0.079) indicates no statistical
difference in the two methods with no bias over the 0.3 to 7 ppm range
measured at the site.
Table 4.8. Field Comparison Sampling (ppm EtO
determined)
|
sampling method |
ACT1 |
QKT2 |
ACT |
QKT |
|
ppm EtO |
0.330 0.370 0.774 0.892 0.982 1.00 1.17 1.29 |
0.278 0.359 0.731 0.938 1.06 1.04 1.09 1.10 |
1.40 2.11 2.26 2.67 2.71 4.56 6.87 |
1.42 1.88 2.35 2.44 2.43 4.04 7.15 |
|
1 acid-coated
tube 2
Quazi-Ketchem tube |
Table 4.9. Additional Storage
Data1
|
test
atmos. concentration (ppm) |
% recovery ± SD (0 days
stored) |
% recovery ± SD (after
storage2) |
test
of significance3
|
|
0.10
0.50
1.0
16.0 |
96.8 ±4.9
94.6
±2.0
90.6 ±1.4
90.4 ±5.4 (4 samples) |
102 ±6.4 (18 days) 92.3
±3.5 (21 days) 84.6 ±2.1 (28 days) 85.6 ±3.4 (14
days) (4 samples) |
not significant at 0.05
level not significant at 0.05 level significant at
0.01 level not significant at 0.05 level |
|
1
six samples per data point unless otherwise
indicated 2 days of storage in
parenthesis 3 two-tailed Student
t-test of means |
Table 4.10. Low Humidity Sampling (2-h sampling of 2 ppm
EtO at <5% R.H.)
|
% recovery ±1 SD (0 days
stored) |
% recovery ±1 SD (with
storage) |
test
of significance1
|
|
99.4 ±2.8 (5 samples)
98.9
±2.5 (4 samples) |
80.4
±3.12 (6 samples)
88.8
±4.733 (4 samples) |
significant at 0.01
level
significant at 0.05 level |
|
1
two-tailed Student t-test of
means 2 twenty-four days of ambient
storage 3 fourteen days of ambient
storage |
Table 4.11. Short-Term Sampling at 5 ppm
EtO1
|
sampling time (min)
|
% recovery ±1 SD (0 days
stored) |
% recovery ±1 SD (21 days
stored) |
test
of significance2
|
|
15
30
|
95.6 ±2.2
93.8 ±4.3
|
90.6 ±5.3
88.0 ±4.3
|
not significant at 0.05
level not significant at 0.05 level |
|
1
three samples per data point 2
two-tailed Student t-test of
means |
Figure 2.4. Breakthrough study for EtO.
Figure 3.6.1. Analytical standard for EtO (approximately 50 µg
EtO/sample).
Figure 3.6.2. Air sample from hospital field
study.
Figure 4.1. Detection limit for EtO.
Figure 4.4. Calibration curve for EtO.
Figure 4.5.1. Ambient storage for EtO.
Figure 4.5.2. Refrigerated storage for EtO.
5. References
5.1. Qazi, A.H.; Ketcham, N.H. Am. Ind. Hyg. Assoc. J.
1977, (38), 635-647.
5.2. Roman, S.J.; Renner, J.A. Am. Ind. Hyg. Assoc. J.,
1979, (40), 742-745.
5.3. Potter, Wayne "OSHA Method No. 30, Ethylene Oxide", August
1981, OSHA Analytical Laboratory, Salt Lake City, Utah 84115.
5.4. Mori, K.; Nishida, S.; Harda, H. Eisei Kagaku, 1980,
(26), 107-111.
5.5. "Current Intelligence Bulletin No. 25, Ethylene Oxide", May
22, 1981, U.S. Dept. of Health and Human Services, Public Health
Service, Center for Disease and Control, NIOSH.
5.6. "Occupational Exposure to Ethylene Oxide", Federal
Register, June 2, 1984, (49), 25734-809.
5.7. J.N. Cawse, in "Kirk-Othmer Encyclopedia of Chemical
Technology", Vol. 9, pp. 432-471, 3rd Edition, John Wiley
and Sons, N.Y., 1980.
|