2,4-DINITROTOLUENE (DNT) 2,4,6-TRINITROTOLUENE (TNT)
Method no.: |
44 |
|
Matrix: |
Air |
|
Target concentration: |
DNT or TNT 1.5 mg/m3 (OSHA
PEL) (skin notations apply) |
|
Procedure: |
Samples are collected by drawing known volumes of air
through laboratory modified commercial Tenax-GC resin
tubes. The modification consists of the placement of an
8-mm glass-fiber filter disc inside the
tube, ahead of the first resin bed. The samples are desorbed with
acetone and analyzed by gas chromatography using a Thermal Energy
Analyzer (TEA) equipped with an Explosives Analysis Package
(EAP). |
|
Recommended air volume and sampling rate: |
60 L at 1 L/min |
|
|
|
Reliable quantitation limit,
µg/m3: |
|
|
Standard error of estimate at the target
concentration, %: (Section 4.7.) |
|
|
Special requirements: |
The air sampling pump must be certified by NIOSH or
MSHA as intrinsically safe for use in coal mines. |
|
Status of method: |
A sampling and analytical method that has been
subjected to the established evaluation procedures of the Organic
Methods Evaluation Branch. |
|
Date: October 1983 |
Chemist: Warren
Hendricks |
Organic Methods Evaluation Branch OSHA Analytical
Laboratory Salt Lake City, Utah
1. General Discussion
1.1. Background
1.1.1. History
The fully validated NIOSH air sampling procedure for DNT
recommends the use of a 37-mm diameter mixed cellulose ester filter
connected in series with a midget bubbler containing ethylene glycol
(Ref. 5.1.).
NIOSH has evaluated a collection procedure for TNT which resulted
in a failure report. The failure report cited inadequate collection
of TNT vapors. The test method utilized filter collection because
initial data indicated that TNT would exist primarily as
particulate. However, it was determined that generated test
atmospheres contained a considerable vapor component which was not
retained by the filter. The failure report also indicated poor
storage stability for both generated and spiked samples.
Volatilization and chemical decomposition were given as possible
reasons for the low recoveries following storage. The failure report
concluded that a particulate/vapor sampling train should definitely
be used to collect TNT (Ref. 5.2.).
This work was undertaken because no adequate TNT sampling method
was available and also because the DNT sampling method employs a
bubbler which is inconvenient for field use. In addition, a common
sampling procedure for DNT and TNT is appropriate because the
analytes may be present together.
This method recommends the use of a commercial, large size,
two-section Tenax-GC sampling tube which has been
modified by the addition of an 8-mm glass fiber filter disc for the
collection of DNT and TNT. The filter is placed inside the tube
ahead of the first resin bed and is used to collect aerosols which
may otherwise penetrate the sorbent. The 100-mg
Tenax-GC adsorbent bed, located behind the filter,
serves to collect vapors and also any analyte which may volatilize
from the filter. The 50-mg Tenax-GC resin bed is used
as a backup section.
Tenax-GC resin was selected for evaluation as a
collection medium for DNT and TNT vapors because of published
recommendations (Ref. 5.3.) and also because initial laboratory
tests indicated that the material would prove to be adequate.
The air sampling device was evaluated by conducting experiments
using a TSI Model 3050 Bergland-Liu Vibrating Orifice Monodisperse
Aerosol Generator and a TSI Model 3076 Constant Output Atomizer
sub-micrometer aerosol generator. A TSI Model 3200 Particle Mass
Monitor was used to detect the presence of an aerosol in the test
atmospheres.
Glass fiber filters, midget bubblers containing toluene or
acetone, Tenax-GC resin tubes, and the recommended
filter disc/Tenax-GC sampling device were evaluated as
sampling media for DNT/TNT aerosol test atmospheres. Glass fiber
filters proved ineffective because DNT was not well retained. Midget
bubblers containing either toluene or acetone gave low results due
to the breakthrough of both analytes. Sampling tubes containing
Tenax-GC resin alone were not effective because
submicrometer aerosols of both analytes penetrated the resin beds.
Only the recommended sampling device provided consistent results
without breakthrough of the analytes onto a backup section or device
(Section 4.5.).
Several very adequate analytical techniques are available for DNT
and TNT. These techniques include high performance liquid
chromatography (HPLC) with ultraviolet detection (Ref. 5.1.), gas
chromatography (GC) with electron capture detection (Ref. 5.3.), GC
with flame ionization detection (Ref. 5.4.), and GC using a
specialized chemiluminescent (TEA/EAP) detector (Ref. 5.5.). A
GC/(TEA/ EAP) analytical procedure was selected because the TEA/EAP
has been shown to have a sensitive and selective response to the
analytes. The GC separation method was necessary because HPLC
solvents are not compatible with the TEA/EAP when it is operated in
the high temperature nitro mode.
1.1.2. Toxic effects (This section is for information only and
should not be taken as the basis of OSHA policy).
DNT Inhalation and skin absorption are both significant
means of occupational exposure to DNT. Intense headaches are
frequently the first reported symptom of overexposure to DNT.
Additional complaints may include fatigue, nausea, vomiting, chest
pain, and weight loss. These symptoms are caused by anoxia (loss of
oxygen-carrying capacity of the blood) due to the
formation of methemoglobin. Jaundice and anemia have been reported
as a result of chronic exposure to DNT (Ref. 5.6.).
Technical grade 2,4-DNT, and all six individual isomers of DNT
were reported to be mutagenic in the Ames Salmonella/microsome test
(Ref. 5.7.). The Salmonella/microsome mutagenicity test was
developed for use as a screening method to identify potential
carcinogens (Ref. 5.8.).
Practical-grade 2,4-DNT (purity greater than 95%) was
administered to rats and mice in a bioassay to test its possible
carcinogenicity. The compound was administered in the food to male
and female animals of each species for 78 weeks. The results of the
study show that dietary administration of 2,4-DNT resulted in
fibroma of the skin and subcutaneous tissue of male rats and
fibroadenoma of the mammary gland for female rats. No evidence was
observed for the carcinogenicity of the agent in mice of either sex
(Ref. 5.9.).
TNT Occupational exposure to TNT has been reported to
occur by inhalation, ingestion, and skin absorption. Symptoms of
overexposure to TNT include liver damage, cyanosis, sneezing, cough,
sore throat, peripheral neuritis, muscular pain, kidney damage,
cataracts, sensitization dermatitis, leukocytosis (large increase in
the number of white cells in the blood) or leukopenia (abnormally
low number of white cells in the blood), and aplastic anemia (Ref.
5.6.).
Toxic effects have been observed in humans at TNT levels well
below the current OSHA PEL of 1.5 mg/m3.
The effects included upper respiratory and gastrointestinal
complaints, anemia, liver function abnormalities, and possibly
aplastic anemia. A standard of 0.5 mg/m3
(eight hour time-weighted exposure) was suggested for protection
against adverse health effects due to TNT exposure (Ref. 5.10.).
TNT was reported to be mutagenic in the Ames Salmonella/microsome
test (Ref. 5.11.).
A literature search resulted in no evidence for the
carcinogenicity of TNT. Additional carcinogenicity testing of TNT is
recommended because the agent is a bacterial mutagen and exposure
has been shown to result in aplastic anemia. Aplastic anemia is a
condition characterized by defective functioning of the
blood-forming organs. Other chemicals which cause aplastic anemia
have been identified as carcinogens (Ref. 5.10.).
1.1.3. Potential workplace exposure
DNT In 1975, 308 million pounds of 2,4-DNT and 273 million
pounds of a mixture of 2,4- and 2,6-DNT were produced. The chemical
is used by the dye manufacturing and munitions industries. It is
also used as a chemical intermediate to produce toluene diisocyanate
which is used to make polyurethane foam (Ref. 5.12.).
TNT The production of TNT was estimated to be 48 million
pounds in 1976. TNT is used as a military explosive (Ref. 5.12.). It
is also used as an intermediate in dyestuffs and in photographic
chemicals (Ref. 5.13.).
1.1.4. Physical properties (Ref. 5.13. and 5.14.)
|
|
|
DNT |
TNT |
|
|
CAS no.: |
121-14-2 |
118-96-7 |
molecular weight: |
182.14 |
227.13 |
physical appearance: |
yellow solid |
pale yellow solid |
UV l maximum, nm: |
252 |
225 |
melting point (°C): |
71 |
82 |
boiling point (°C): |
300 (sl. dec.) |
240 (explodes) |
density (g/mL) (at 71°C): |
1.3208 |
1.654 |
|
solubility |
|
water: |
insoluble |
insoluble |
alcohol: |
soluble |
slightly soluble |
ether: |
soluble |
soluble |
acetone: |
very soluble |
soluble |
benzene: |
soluble |
soluble |
|
|
structures: Figure 1.1.4.
synonyms: (Ref. 5.15.) DNT
2,4-dinitrotoluene; 2,4-dinitrotoluol;
2,4-DNT; 1-methyl-2,4-dinitrobenzene.
TNT 2-methyl-1,3,5-trinitrobenzene;
entsufon; tolite; trinitrotoluene; s-trinitrotoluene;
2,4,6-trinitrotoluene; triton;
2,4,6-trinitrotoluol.
1.2. Limit Defining Parameters (The analyte air concentrations
listed throughout this method are based on an air volume of 60 L and a
desorption volume of 3.0 mL.)
1.2.1. Detection limits of the analytical procedure
The detection limits of the analytical procedures are 0.36 ng for
DNT and 0.37 ng for TNT per injection. These are the amounts of
analytes which gave peaks whose heights were about 5 times the
height of the baseline noise. (Section 4.1.)
1.2.2. Detection limits of the overall procedure
The detection limits of the overall procedure for DNT and TNT are
1.21 µg (20 µg/m3) and 1.23 µg (21
µg/m3) per sample respectively. These are
the amounts of analytes spiked on the sampling device which allow
recoveries approximately equivalent to the detection limits of the
analytical procedure. (Section 4.2.)
1.2.3. Reliable quantitation limits
The reliable quantitation limits for DNT and TNT are 1.21 µg (20
µg/m3) and 1.23 µg (21
µg/m3) per sample respectively. These are
the smallest amounts of the analytes which could be quantitated
within the requirements of a recovery of at least 75% and a
precision (1.96 SD) of ±25% or better. (Section 4.3.)
The reliable quantitation limit and detection limits reported in
the method are based upon optimization of the instrument for the
smallest possible amount of analyte. When the target concentration of
an analyte is exceptionally higher than these limits, they may not be
attainable at the routine operating parameters.
1.2.4. Sensitivity
The sensitivities of the analytical procedure over concentration
ranges representing 0.5 to 2 times the OSHA PEL, based on the
recommended air volume, are 13441 area units per µg/mL for DNT and
13199 area units per µg/mL for TNT. These were determined by the
slope of the calibration curves. (Section 4.4.) The sensitivity may
vary with the particular instrument used in the analysis.
1.2.5. Recovery
The recoveries of DNT and TNT from samples used in the 19-day
ambient temperature test are 95.0% and 93.7%, respectively, relative
to control samples. These were recoveries at day 19, determined from
the linear regression line of the storage data. (Section 4.7.) The
recovery of the analyte from the collection device following storage
must be at least 75%.
1.2.6. Precision (analytical procedure only)
The pooled coefficient of variation obtained from replicate
determinations of analytical standards at 0.5, 1, and 2 times the
target concentration are 0.021 for DNT and 0.015 for TNT. (Section
4.4.)
1.2.7. Precision (overall procedure)
The precisions at the 95% confidence level for the 19-day ambient
temperature storage test are ±15.6% for DNT and ±16.1% for TNT.
(Section 4.7.) These values each include an additional ±5% for
sampling error. The overall procedure must provide results at the
target concentration that are ±25% or better at the 95% confidence
level.
1.2.8. Reproducibility
Six spiked samples and a draft copy of this procedure were given
to a chemist unassociated with this evaluation. The samples were
analyzed after 6 days of storage at ambient temperature. The average
recoveries (corrected for desorption efficiencies) were 99.2% for
DNT and 98.0% for TNT. The standard deviations were 4.9% for DNT and
9.3% for TNT. (Section 4.9.)
1.3. Advantages
1.3.1. The sampling and analytical procedures are precise,
reliable, and convenient.
1.3.2. Air samples are stable even when stored at ambient
temperature for 19 days.
1.4. Disadvantages
1.4.1. This method has not been field tested.
1.4.2. The sampling device is not commercially available.
1.4.3. The TEA/EAP detector is expensive.
2. Sampling Procedure
2.1. Apparatus
2.1.1. Samples are collected by use of a personal sampling pump
that can be calibrated to within ±5% of the recommended flow rate
with the sampling device in line. The sampling pump must be
certified by NIOSH or MSHA as intrinsically safe for use in coal
mines.
2.1.2. Samples are collected on laboratory modified, commercial,
Tenax-GC resin sampling tubes. SKC, Inc.
Tenax-GC resin tubes (catalog no. 226-35-03) were used
to prepare the sampling device used in this evaluation. The SKC tube
has two sections of 35/60 mesh resin separated by a glass wool plug.
The front (sampling) section contains 100 mg of resin and the back
section 50 mg. The sections are held in place by glass wool plugs in
an 8-mm o.d. × 100-mm glass tube.
The laboratory modification of the sampling tubes is performed as
follows: remove the flame sealed tip of the glass sampling tube
nearest the 100-mg section of resin. Leave about 2.5 cm of glass
tubing in front of the 100-mg resin bed. Remove the steel lockspring
wire. Prepare Teflon-support rings by cutting Teflon tubing of 6-mm
o.d. and 4-mm i.d. into 0.5-cm lengths. Cut each Teflon ring along
its 0.5-cm length to permit its easy insertion into the sampling
tube. Place a Teflon-support ring on top of the exposed glass wool
plug of the sampling tube. Be careful not to compress the glass
wool. Severe compression of the glass wool will cause high back
pressures when sampling. Prepare 8-mm glass fiber
filter discs by using a number 4 cork borer to cut the discs from
Gelman Type A glass fiber filters. Place an 8-mm filter disc inside
the sampling tube by tamping the oversize filter on top of the
Teflon-support ring with a glass rod or similar object. Place
another Teflon-support ring on top of the filter so that the filter
disc is sandwiched between the two support rings. Fire polish the
cut end of the glass sampling tube, for safety. Cap the modified
device with one of the sealing caps that are included with the SKC
Tenax-GC resin tubes.
2.2. Reagents
None required
2.3. Technique
2.3.1. Break open the closed end of the laboratory modified
Tenax-GC resin sampling tube. Remove and save the
sealing cap on the front of the device. Connect the device to a
NIOSH or MSHA certified sampling pump with flexible tubing. Position
the tube so that sampled air first passes through the filter disc
and then into the larger resin bed. Sampled air should not pass
through any hose or tubing before entering the sampling tube.
2.3.2. Place the sampling tube vertically in the employee's
breathing zone.
2.3.3. After sampling, seal the tube immediately with plastic
caps and wrap it lengthwise with OSHA Form 21.
2.3.4. Submit at least one blank for each sample set. The blank
should be handled in the same manner as the samples, except that no
air is drawn through it.
2.3.5. List any potential interferences on the sample data sheet.
2.3.6. Ship bulk material samples in a separate container to
prevent contamination of the air samples. Shipping restrictions may
apply to DNT and TNT bulk samples.
2.4. Breakthrough
Several studies were performed to investigate breakthrough and the
collection efficiency of the air sampling device. No breakthrough from
the 100-mg to the 50-mg resin bed was observed when the recommended
air sampler was used. (Section 4.5.)
2.5. Desorption efficiency
The average desorption efficiencies for DNT and TNT from samples
spiked at 0.5, 1, and 2 times the OSHA PEL are 97.4% and 95.8%
respectively. (Section 4.6.)
2.6. Recommended air volume and sampling rate
2.6.1. The recommended air volume is 60 L. The recommended air
volume was not selected because of breakthrough problems but because
the filter disc was found to be somewhat susceptible to plugging. It
was observed that the filter could partially plug when DNT and TNT
concentrations were significantly higher than the PEL. When,
however, the levels were near the PEL, filter plugging was not
significant, even when the test atmosphere was sampled for extended
periods. The 60 L recommended air volume should provide an adequate
safety margin to prevent filter plugging. (Section 4.5.)
2.6.2. The recommended air sampling rate is 1 L/min.
2.7. Interferences (sampling)
2.7.1. There are no known interferences to the sampling method.
2.7.2. Suspected interferences should be reported to the
laboratory on the sampling data sheets.
2.8. Safety precautions (sampling)
2.8.1. The air sampling pump must be certified by NIOSH or MSHA
as intrinsically safe for use in coal mines.
2.8.2. Exercise due caution when breaking open the sampling
tubes. Take measures to prevent cuts from the sharp ends of the
broken glass tubes.
2.8.3. Attach the sampling equipment to the worker in such a
manner that it will not interfere with work performance or safety.
2.8.4. Follow all safety practices that apply to the work area
being sampled.
3. Analytical Procedure
3.1. Apparatus
3.1.1. A GC apparatus equipped with a TEA/EAP detector. For this
evaluation, a Hewlett-Packard Model 5840A gas chromatograph was used
in series with a Thermo Electron Corporation Model 502 A TEA/EAP.
Injections were made using a Hewlett-Packard Model
7671-A automatic sampler. The TEA/EAP cold trap was replaced with a
CTR gas stream filter which was purchased from Thermo Electron
Corporation.
3.1.2. A GC column capable of resolving the analytes from each
other and potential interferences. A 3-ft × 0.2-mm i.d. glass GC
column containing 3% OV 225 on 100/120 mesh Chromosorb W AW was used
in this evaluation.
3.1.3. Vials, 4-mL with Teflon-lined caps. Waters WISP vials were
used in this evaluation.
3.1.4. Volumetric flasks, pipets and syringes for preparing
standards, making dilutions and making injections.
3.2. Reagents
3.2.1. HPLC grade acetone.
3.2.2. GC grade helium, oxygen, and air.
3.2.3. DNT and TNT of known purity.
3.3. Standard preparation
3.3.1. Prepare stock standards by diluting known amounts of DNT
and TNT with acetone.
3.3.2. Prepare an intermediate standard mixture using known
volumes of the stock standards and diluting the mixture with
acetone. The intermediate standard should contain 1.5 mg/mL of each
analyte.
3.3.3. Prepare fresh working range standards daily by diluting
the intermediate standard mixture with acetone. A standard
representing the OSHA PEL was obtained by diluting the intermediate
standard mixture 1 to 50 with acetone.
3.3.4. Prepare standards at concentrations other than the OSHA
PEL in order to generate calibration curves.
3.3.5. Store the standards in a freezer using well-sealed dark
containers.
3.4. Sample preparation
3.4.1. Transfer both Teflon-support rings, the glass-fiber
filter disc, the front glass wool plug, and the front
Tenax-GC resin section to a 4-mL vial. Place the center
glass wool plug and the Tenax-GC backup section in a
separate vial. Discard the end glass wool plug.
3.4.2. Add 3.0 mL of acetone to each vial.
3.4.3. Seal the vials with Teflon-lined caps and allow them to
desorb for 1 h. Shake the vials by hand with moderate force several
times during the desorption time.
3.4.4. Wash the inside of the glass sampling tube into a separate
vial with three 1-mL volumes of acetone.
3.5. Analysis
3.5.1. GC conditions
injection temperature: |
175°C |
column temperature: |
temperature programmed from 150 to 210°C at
6°C/min |
helium flow rate: |
30 mL/min |
injection volume: |
0.9 µL |
GC column: |
3% OV 225 on 100/120 mesh Chromosorb W
AW |
3.5.2. TEA/EAP conditions
GC pyrolyzer temperature: |
800°C |
GC interface temperature: |
225°C |
oxygen flow rate: |
5 mL/min |
3.5.3. Chromatogram: Section 4.8.
3.5.4. Detector response is measured with an electronic
integrator.
3.5.5. Use an external standard method to prepare the calibration
curve with at least three standard solutions of different
concentrations. Prepare the calibration curve daily. Program the
integrator to report results in µg/mL.
3.5.6. Bracket sample concentrations with standards.
3.6. Interferences (analytical)
3.6.1. Any compound with the same general retention time as DNT
or TNT and which also gives a detector response is a potential
interference. Possible interferences should be reported to the
laboratory with submitted samples by the industrial hygienist.
3.6.2. GC parameters (temperature, column, etc.) may be changed
to possibly circumvent interferences.
3.6.3. A useful means of structural designation is GC/MS. It is
recommended this procedure be used to confirm samples whenever
possible.
3.7. Calculations
3.7.1. Results are obtained by use of a calibration curve. The
detector response, for each standard, is plotted against its
concentration in µg/mL and the best straight line through the data
points is determined by linear regression.
3.7.2. The concentration, in µg/mL, for a particular sample is
determined by comparing its detector response to the calibration
curve. If any DNT and/or TNT is found on the backup section or in
the tubing wash, it is added to the amount found on the sampling
section. This total amount is then blank corrected.
3.7.3. The DNT and/or TNT air concentration can be expressed
using the following equation:
mg/m3 DNT or TNT =
(A)(B)/(C)(D)
where |
A = µg/mL from Section 3.7.2. |
|
B = desorption volume |
|
C = liters of air sampled |
|
D = desorption efficiency (decimal
form) |
3.8. Safety precautions (analytical)
3.8.1. Avoid skin contact and inhalation of all chemicals used.
3.8.2. Restrict the use of all chemicals to a fume hood whenever
possible.
3.8.3. Check that the TEA exhaust is connected to a fume hood.
3.8.4. Wear safety glasses and a lab coat in all laboratory
areas.
4. Backup Data
4.1. Detection limits of the analytical procedure
The detection limits of the analytical procedure are 0.36 ng for
DNT and 0.37 ng for TNT per injection. These amounts produce peaks
whose heights were about 5 times the height of the baseline noise
(Figure 4.1.).
4.2. Detection limits of the overall procedure
The detection limits of the overall procedure are 1.21 µg (20
µg/m3) for DNT and 1.23 µg (21
µg/m3) for TNT per sample. These are the
amounts of analytes spiked on the sampling device which allow
recoveries approximately equivalent to the detection limits of the
analytical procedure (Table 4.3.1.). The 0.9-µL injection size
recommended in the analytical procedure was used in the determination
of the detection limits of the overall procedure.
4.3. Reliable quantitation limits
The reliable quantitation limits were determined by liquid spiking
six air samplers with 1.21 µg of DNT and 1.23 µg of TNT. These samples
were desorbed with 3.0 mL of acetone for 1 h. The 0.9-µL injection
volume recommended in the analytical procedure was used to determine
the reliable quantitation limits. The results of the analysis of the
spiked samples are presented in Table 4.3.1.
Table 4.3.1. Reliable Quantitation Limit Data
|
|
DNT |
TNT |
DNT |
TNT |
sample no. |
mass spiked, µg |
% recovery |
|
1 2 3 4 5 6
SD 1.96 ×
SD |
1.21 1.21 1.21 1.21 1.21 1.21 |
1.23 1.23 1.23 1.23 1.23 1.23 |
85.2 95.4 92.0 88.6 105.6 92.7
93.2 7.01 13.7 |
83.2 99.8 101.0 96.2 111.7 89.1
96.8 9.93 19.5 |
|
Since the recoveries were near 100% and the precisions were better
than ±25%, the detection limits of the overall procedure and the
reliable quantitation limits were the same.
4.4. Sensitivity and precision (analytical method only)
The data in Tables 4.4.1. and 4.4.2. were obtained from multiple
injections of analytical standards. The data are also presented
graphically in Figures 4.4.1. and 4.4.2. The sensitivity for DNT was
13441 area counts per µg/mL and that for TNT was 13199 area counts per
µg/mL.
Table 4.4.1. DNT Sensitivity and Precision Data
|
× target conc. µg/mL |
0.5× 15.2 |
1× 30.4 |
2× 60.7 |
|
area counts
SD CV
= 0.021 |
180900 181600 188400 179300 180500 180200
181816.7 3313.9 0.0182 |
375000 372500 383900 374900 396200
380500.0 9793.6 0.0257 |
797200 807000 782800 796400 769600 801200
792366.7 13723.2 0.0173 |
|
Table 4.4.2. TNT Sensitivity and Precision Data
|
× target conc. µg/mL |
0.5× 15.4 |
1× 30.8 |
2× 61.6 |
|
area counts
SD CV
= 0.015 |
176300 173600 169900 173100 171200 170900
172500.0 2324.7 0.0135 |
376400 370200 380800 369600 379500
375300.0 5186.5 0.0138 |
794000 779000 784200 798000 758000 780000
782200.0 14096.8 0.0180 |
|
4.5. Breakthrough
Breakthrough and collection efficiency studies were conducted using
micrometer and sub-micrometer aerosols containing DNT and
TNT. The micrometer aerosols were generated with a TSI Model 3050
Bergland-Liu Vibrating Orifice Monodisperse Aerosol Generator. The
submicrometer test atmospheres were generated with a TSI Model 3075
Constant Output Atomizer, used in the nonrecirculating mode. Both
generation devices were equipped with TSI electrostatic charge
neutralizers. The output of each generator was monitored with a TSI
Model 3200 Particle Mass Monitor. The aerosol was sampled by means of
ports connected to a sampling chamber.
Micrometer Aerosol Data
Test atmospheres were generated by pumping an acetone solution
containing 0.46 mg/mL DNT and 0.46 mg/mL TNT into the vibrating
orifice aerosol generator at 0.15 mL/min. The frequency of the 20 µm
orifice was set at 36 KHz. The monodisperse particle diameters were
calculated to be 3.2 µm.
Two runs were performed using both modified and unmodified
Tenax-GC tubes as sampling devices. The aerosols were
sampled for 1 h at 1 L/min. The results of the sampling device
comparison test are presented in Table 4.5.1. Each data point
represents the average of at least two separate air samples taken
using identical sampling devices.
Table 4.5.1. Sampling Device Comparison: Micrometer
Aerosol
|
device |
DNT,
µg |
TNT,
µg |
type |
filter |
GW |
A |
B |
filter |
GW |
A |
B |
|
run
1 filter/Tenax Tenax
run
2 filter/Tenax Tenax |
3.4 ---
15.1 --- |
0.4 2.7
I/A 9.2 |
18.8 23.8
22.4 28.0 |
ND ND
ND ND |
15.1 ---
26.6 --- |
1.1 16.9
I/A 29.4 |
2.6 4.2
5.3 4.8 |
ND ND
ND ND |
|
filter = filter disc in the
recommended air sampling device; GW = front glass wool
plug; A = Tenax-GC tube "A" section; B = Tenax-GC tube "B"
section; I/A = included A; ND = none
detected. |
The data in Table 4.5.1. show that both sampling devices provided
similar results when the test atmosphere contained 3 µm particles. The
front glass wool plug of the unmodified device acted as a partially
effective filter for both analytes. The common practice of discarding
front glass wool plugs from adsorbent tubes is not supported by these
data.
Submicrometer Aerosol Data
Test atmospheres were generated by pumping an acetone solution
containing DNT and TNT into the atomizer assembly at 0.7 mL/min.
Aerosols generated in this manner are polydisperse and are estimated
to have mean particle diameters of 0.02 to 0.3 µm.
Several runs were performed which demonstrated the effectiveness of
the recommended glass fiber filter/Tenax-GC resin tube
and, conversely, the inadequacy of Tenax-GC tubes without
filters. Samples were also taken with midget bubblers containing
either acetone or toluene. Most sampling was performed at 1 L/min for
60 min.
The data in Table 4.5.2. are the results obtained when a
submicrometer aerosol was sampled with two Tenax-GC tubes
connected in series, the recommended filter/Tenax-GC
tube, and two midget bubblers, containing toluene, connected in
series. Each data point is the average of two separate samples taken
using identical sampling devices.
Table 4.5.2. Sampling Device Comparison With Submicrometer
Aerosol
|
device |
device component |
DNT, µg |
TNT, µg |
|
toluene bubblers (two in series)
tubes
w/o filters (two in
series)
tube with
filter
|
bubbler 1 bubbler 2
tube 1 glass
wool Tenax A Tenax B
tube 2 glass wool Tenax
A Tenax B
filter Tenax A Tenax B |
84.8 22.6
6.0 73.0 6.3
ND 6.0 3.2
15.4 86.0 ND |
65.8 36.6
34.6 22.8 22.7
7.0 13.3 8.7
124.2 12.2 ND |
|
The data in Table 4.5.2. show that toluene bubblers and unmodified
Tenax-GC resin tubes are inadequate to sample atmospheres
containing submicrometer DNT and TNT aerosols. When compared to the
results obtained by use of the tube containing a filter: Toluene
bubbler #1 results were 16% low for DNT and 52% low for TNT. Tube 1
(w/o filter) results were 16% low for DNT and 41% low for TNT. The
bubbler pair results were 6% high for DNT and 25% low for TNT. The
tube pair (w/o filters) results were 7% low for DNT and 20% low for
TNT.
A study was performed to determine if midget bubblers containing
acetone would be more effective than toluene bubblers. The sampling
time for this run was 30 min. The bubbler results are the average of
two separate air samples.
Table 4.5.3. Sampling Device Comparison With Submicrometer
Aerosol
|
device |
device components |
DNT, µg |
TNT, µg |
|
acetone bubblers (two in
series)
tube with filter |
bubbler 1 bubbler 2
filter Tenax
A Tenax B |
27.6 4.4
21.2 28.1 ND |
20.8 5.8
69.6 1.5 ND |
|
The acetone bubbler results, presented in Table 4.5.3. show this
device to be especially ineffective when used to sample the generated
test atmosphere. The acetone bubbler pair results were 35% low for DNT
and 62% low for TNT when compared to results obtained with a tube
containing a filter.
The data presented in Table 4.5.4. are the results obtained when
the sub-micrometer aerosol was sampled with the recommended device and
also with three Tenax-GC tubes connected in series. The
third tube of this sampling train contained a filter disc. The glass
wool plugs were analyzed together with the appropriate Tenax section.
Each data point is the average of two separate samples taken using
identical sampling devices.
Table 4.5.4. Sampling Device Comparison With Sub-micrometer
Aerosols
|
device |
device components |
DNT, µg |
TNT, µg |
|
three tubes in series, with the third
tube containing a
filter
tube with
filter |
tube 1 Tenax A Tenax B
tube
2 Tenax A Tenax B
tube 3 filter Tenax
A Tenax B
filter Tenax A Tenax B |
233.1 48.7
37.7 5.1
0.9 23.1 ND
216.8 111.8 ND |
201.2 96.3
41.8 11.1
42.9 5.6 ND
396.6 5.2 ND |
|
The data in Table 4.5.4. show that the breakthrough from tube 1 to
the remainder of the multiple tube device was 19% for DNT and 25% for
TNT. These data also show that the results from two unmodified tubes
were 7% low for DNT and 12% low for TNT. These data agree with the
data presented in Table 4.5.2. and show conclusively that a filter
disc is required to sample sub-micrometer aerosols containing DNT and
TNT.
It was observed that the filter disc was somewhat susceptible to
plugging when the generated test atmosphere contained DNT and TNT at
concentrations significantly higher than the OSHA PEL. However, filter
plugging was not significant when the levels were near the PEL. The
data in Table 4.5.6. were taken from five separate studies.
Table 4.5.6. Filter Disc Plugging vs. Concentrations of
DNT and TNT in a Combined Atmosphere
|
DNT conc., mg/m3 |
TNT conc., mg/m3 |
sampling time,
min |
air flow rate
before sampling, L/min |
air flow rate
after sampling, L/min |
|
6.0 2.0 1.4 1.2 1.1 |
7.1 4.0 2.0 1.2 1.2 |
70 103 58 254 257 |
0.97 0.99 0.96 0.97 0.97 |
0.64 0.88 0.97 0.94 0.90 |
|
Filter disc plugging is likely caused by TNT because DNT is easily
air-stripped from the filter. When the recommended sampling device was
preceded by an unmodified Tenax-GC resin tube, the filter
disc did not plug. Used in the recommended configuration, the filter
removed about 95% of the incoming TNT. An unmodified
Tenax-GC tube removed about 75% of the incoming TNT.
Therefore, the filter, when preceded by a sampling tube, was not
challenged with the full amount of TNT and it did not plug.
It was decided not to evaluate filter discs preceded by
Tenax-GC tubes as sampling media for DNT and TNT for the
following reasons: The use of a device composed of a
Tenax-GC resin tube followed by a filter disc requires
multiple tubes because the analytes can be air-stripped
from the filter. Filter plugging was not significant at levels near
the OSHA PEL when the filter was in front of the tube.
The high affinity of Tenax-GC resin for DNT and TNT
was demonstrated by a retention efficiency experiment. The filter disc
of a sampling device was liquid spiked with 360 µg of DNT and 340 µg
of TNT. The device was then connected to a humid air generator and 335
L of air at 77% relative humidity and 24°C were drawn through the
spiked air sampler. At the end of the test the device was analyzed and
less than 0.1% of the DNT/TNT spiked on the filter disc was found on
the Tenax-GC B section of resin.
4.6. Desorption efficiency
The following data are the results of the analysis of modified
Tenax-GC tubes spiked with DNT and TNT at 0.5, 1, and 2
times the OSHA PEL. The analytes were liquid spiked on the filter, the
tubes were sealed and stored in a freezer to be analyzed the following
day.
Table 4.6.1. Desorption Efficiency From Sampler When the
Filter Was Spiked
|
× target conc. |
0.5× |
1× |
2× |
analyte µg/sample |
DNT 45.5 |
TNT 46.2 |
DNT 91.0 |
TNT 92.4 |
DNT 182 |
TNT 185 |
|
desorption efficiency, %
|
96.8 93.4 96.9 93.0 99.3 94.9
95.7 |
96.2 94.5 95.1 94.6 95.0 95.0
95.1 |
97.4 95.5 102 96.8 98.9 99.7
98.4 |
91.7 95.6 98.9 94.2 102 97.7
96.7 |
98.4 93.2 102 96.3 95.1 103
98.0 |
98.2 93.0 98.0 96.4 92.4 96.4
95.7 |
|
The average desorption
efficiency for DNT was 97.4% and that for TNT was
95.8%. |
To determine if the desorption efficiencies were different for DNT
and TNT spiked directly on Tenax-GC resin, six tubes were
liquid spiked at 2 times the OSHA PEL. The tubes were sealed and
stored overnight in a freezer.
Table 4.6.2. Desorption Efficiency From Sampler When the
Sorbent Bed Was Spiked
|
analyte × target conc. µg/sample |
DNT 2× 182 |
TNT 2× 185 |
|
desorption efficiency,
%
|
94.4 93.9 91.0 98.6 99.7 95.5
95.5 |
97.3 93.6 94.0 100.8 101.3 98.9
97.6 |
|
The difference between the means of the desorption efficiencies
obtained by spiking different components of the sampling device at 2×
the OSHA PEL was tested using a two-tailed Student t distribution. The
computations showed that there was no statistical difference between
the desorption efficiencies of the two media at the 0.05 level of
significance. Therefore, the average desorption efficiencies reported
following Table 4.6.1. (97.4% for DNT and 95.8% for TNT) are those
which should be used for this method.
4.7. Storage data
The data in Tables 4.7.1. and 4.7.2. represent the effects of
storage at ambient (21 to 26°C) and reduced (-20°C) temperature on
samples taken from submicrometer aerosol test atmospheres. The results
are not corrected for desorption efficiency. The data are presented
graphically in Figures 4.7.1. - 4.7.4.
Because some variability in the air concentrations of DNT and TNT
occurred during the generation process, the recoveries in Table 4.7.1.
are reported relative to control samples. Four sets of six samples
were collected for each temperature studied and then two samples from
each set were selected as controls to be analyzed immediately. The
remaining four samples from each set were put into the storage sample
pool and then, when analyzed, corrected by the appropriate control
samples. For the ambient temperature study, the average control sample
was 1.60 mg/m3 for DNT and 2.02
mg/m3 for TNT. The average control sample,
for the reduced temperature study, was 1.17
mg/m3 for DNT and 1.30
mg/m3 for TNT.
Table 4.7.1. Ambient Temperature Storage Tests
|
storage time |
DNT |
|
TNT |
(days) |
(% recovery) |
|
0 4 7 11 14 19 |
100 102 98.9 99.3 104 98.7 |
100 79.2 91.9 95.0 86.0 98.4 |
100 94.8 94.8 96.9 96.0 89.1 |
|
100 101 99.7 97.1 97.9 105 |
100 78.8 92.7 94.7 84.7 92.1 |
100 98.1 93.7 94.6 87.4 95.0 |
|
Table 4.7.2. Reduced Temperature Storage Tests
|
storage time |
DNT |
|
TNT |
(days) |
(% recovery) |
|
0 3 7 10 15 17 |
100 104 90.3 91.5 90.9 86.3 |
100 104 103 98.2 91.5 97.8 |
100 104 97.6 94.1 92.4 97.5 |
|
100 107 98.3 96.8 92.3 93.0 |
100 110 105 102 94.8 97.3 |
100 108 103 106 94.6 98.3 |
|
4.8. Chromatogram
Figure 4.8.1. is a chromatogram obtained by the injection of a
standard mixture containing the analytes. The GC column was 3 ft ×
0.2-mm i.d., constructed of glass and packed with 3% OV 225 on 100/120
mesh Chromosorb W AW. The injector temperature was maintained at 175°C
and the column was temperature programmed from 150 to 210°C at
6°C/min. A TEA Model 502 A (EAP) detector was used in the nitro mode.
The TEA/GC pyrolyzer was set at 800°C and the GC interface temperature
was 225°C.
4.9. Reproducibility study
Six liquid spiked air samplers and a draft copy of this evaluation
were given to a chemist unassociated with this work. The samples were
analyzed after 6 days storage at ambient temperature. The results are
corrected for desorption efficiency.
Table 4.9. Reproducibility Study
|
amount spiked, µg |
DNT 91.0 |
TNT 92.4 |
|
% recovery
SD |
101. 107. 99.0 99.2 92.0 96.7
99.2 4.9 |
93.8 111. 97.9 102. 83.0 100.
98.0 9.3 |
|
Figure 1.1.4. Molecular structures of the
analytes.
Figure 4.1. The detection limits of the analytical
procedure.
Figure 4.4.1. Calibration curve for
DNT.
Figure 4.4.2. Calibration curve for
TNT.
Figure 4.7.1. Ambient temperature storage test for
DNT.
Figure 4.7.2. Ambient temperature storage test for
TNT.
Figure 4.7.3. Refrigerated storage test for
DNT.
Figure 4.7.4. Refrigerated temperature storage test for
TNT.
Figure 4.8.1. GC/(TEA/EAP) chromatogram of DNT and TNT.
5. References
5.1. "NIOSH Manual of Analytical Methods", 2nd ed.; Department of
Health, Education and Welfare, National Institute for Occupational
Safety and Health; Cincinnati, OH 1977; Vol. 4, Method No. S215; DHEW
(NIOSH) Publ. (US), NO. 78-175.
5.2. "FAILURE REPORT NO. S226"; National Institute for Occupational
Safety and Health; Cincinnati, OH March 17, 1978.
5.3. Bishop, R.E.; Ayers, T.A.; Rinehart, D.S. American Ind.
Hygiene Asso. J. (1981), 42, 586.
5.4. Jurinski, N.B.; Podolak, G.E.; Hess, T.L. American Ind.
Hygiene Asso. J. (1975), 497.
5.5. Lafleur, A.L.; Mills, K.M. Anal. Chem. (1981), 53,
1202.
5.6. Proctor, N.H.; Hughes, J.P. "Chemical Hazards of the
Workplace"; J.B. Lippincott Company: Philadelphia, 1978.
5.7. Couch, D.P.; Allen, P.F.; Abernethy, D.J. Mutation
Research (1981), 90, 373.
5.8. McCann, J.; Choi, E.; Yamasaki, E.; Ames, B.N. Proc. Nat.
Acad. Sci. USA (1975), 72, 5235.
5.9. "Bioassay of 2,4-Dinitrotoluene for Possible Carcinogenicity"
U.S. Department of Health, Education and Welfare, Public Health
Service, National Institutes of Health, 1978, DHEW Publ. No. (NIH)
78-1360, NTIS Publ. No. PB-280990.
5.10. Hathaway, J.A. J. of Occupational Medicine (1977), 19,
341.
5.11. Spanggard, R.J.; Mortelmans, K.E.; Griffin, A.F.; Semmon,
V.F. Environmental Mutagenesis (1982), 4, 163.
5.12. "Informational Profiles on Potential Occupational Hazards"
U.S. Department of Commerce, NTIS, Springfield, VA., PB 276-678.
5.13. "The Condensed Chemical Dictionary" Eighth Edition, Van
Nostrand Reinhold Company: New York, 1971.
5.14. Weast, R.C., Ed., "CRC Handbook of Chemistry and Physics",
60th ed.; CRC Press: Boca Raton, FL, 1979.
5.15. "Registry of Toxic Effects of Chemical Substances, 1980 ed."
U.S. Department of Health and Human Services, Public Health Service,
Centers for Disease Control, National Institute for Occupational
Safety and Health, DHHS (NIOSH), Publ. No. 81-116.
|