DIISOCYANATES
1,6-HEXAMETHYLENE DIISOCYANATE (HDI) TOLUENE-2,6-DIISOCYANATE
(2,6-TDI) TOLUENE-2,4-DIISOCYANATE (2,4-TDI)
Method no.: |
42 |
|
Matrix: |
Air |
|
Procedure: |
Samples are collected by drawing a known volume of
air through glass fiber filters coated with 0.1 mg of
1-(2-pyridyl)piperazine (1-2PP) which are contained in open-face
cassettes. Samples are extracted with 90/10 (v/v)
acetonitrile/dimethyl sulfoxide (ACN/DMSO) and analyzed by high
performance liquid chromatography (HPLC) using an ultraviolet or
fluorescence detector. (The coated filters used in Method 47 for MDI
are also acceptable for this procedure. Those filters are coated
with 1 mg instead of 0.1 mg of 1-2PP.) |
|
Recommended air volume and sampling rate: |
15 L at 1 L/min |
|
Special requirements: |
It is recommended that coated glass fiber filters be
stored at reduced temperature until used for sampling. |
|
Status of method: |
Evaluated method. This method has been subjected to
the established evaluation procedures of the Organic Methods
Evaluation Branch. |
|
Date: |
February 1983 March 1989
(Revised) | |
Chemist: Donald
Burright |
Carcinogen and Pesticide Branch OSHA Analytical
Laboratory Salt Lake City, Utah
|
Analyte |
2,6-TDI |
HDI |
2,4-TDI |
|
Target concentration, µg/m3
(ppb): |
140(20) |
140(20) |
1401(20) |
|
Detection limit of the overall procedure,
µg/m3 (ppb): |
1.6(0.23) |
2.3(0.32) |
1.3(0.17) |
|
Reliable quantitation limit,
µg/m3 (ppb): |
2.3(0.32) |
2.9(0.43) |
2.5(0.36) |
|
Standard error of estimate at target
concentration, %: (Section 4.9.) |
7.63
|
7.79
|
6.89
|
|
1 OSHA PEL (Air
concentrations are based on 15-L air sample
volume.) |
1. General Discussion
1.1. Background
1.1.1. History
Some of the earliest procedures to determine atmospheric
diisocyanate concentrations were developed by Ranta and Marcali
(Ref. 5.1.). Both of these procedures are inconvenient because they
use a bubbler for sampling and their colorimetric analyses are
non-specific. A later sampling procedure uses
p-nitrobenzyl-N-n-propylamine (nitro reagent) in toluene bubblers
(Ref. 5.2.). While this method is specific for diisocyanates, it
still retains the use of the bubbler and nitro reagent which is
unstable when stored for long periods of time, even if it is kept at
reduced temperature. The past couple of years have seen several new
derivatizing reagents being used; they include
N-methyl-1-naphthalenemethylamine (Ref. 5.3.),
9-(n-methylaminomethyl)-anthracene (Ref. 5.4.) and
1-(2-pyridyl)piperazine (1-2PP) (Refs.
5.5.-5.7.). The collection procedure of these new studies all
involve the use of toluene bubblers. The purpose of this study was
to find a collection system that does not use a bubbler, yet retains
the sensitivity, precision and accuracy of the nitro reagent method.
1-2PP is a suitable derivatizing reagent, when coated on a glass
fiber filter, for several reasons:
1) The high boiling liquid is retained on a glass fiber filter
and stability is not a problem.
2) The rapid and exothermic reaction with both aromatic and
aliphatic diisocyanates results in derivatization on the filter
(Ref. 5.7.).
3) The derivatives have higher molar absorptivities in the UV
region than those formed with nitro reagent which allows the
extraction volume to be larger without loss of sensitivity (Ref.
5.5.).
This procedure compares favorably when tested side-by-side with
the nitro reagent method by Cummins (Ref. 5.10.) for
2,4-TDI. (Section 4.10.) Additional work is being done
to study 4,4'-methylenediphenylisocyanate and isophorone
diisocyanate using 1-2PP as the derivatizing reagent.
Additional work was performed on this procedure to reflect to
change in Title 29 CFR 1910.1000, Table Z-1-A in 1989. The Ceiling
PEL of 0.14 mg/m3 for 2,4-TDI
was replaced with an 8-h TWA PEL of 0.04
mg/m3. The sampling time can be increased
to 240 min at a sampling rate of 1 L/min. (Sections 4.6. and 4.12.)
1.1.2. Toxic effects (This section is for information only and
should not be taken as a basis for OSHA policy.)
Continued inhalation of diisocyanate vapors or mists can cause
nausea, headache, coughing, irritation of the nose and throat,
shortness of breath and chest discomfort. Massive exposure can cause
severe coughing spasms, bronchitis and chemical pneumonitis. Some
people can become sensitized to isocyanates and may suffer asthmatic
attacks and respiratory distress when subsequently exposed to very
low concentrations (Ref. 5.9.). Recent studies have produced
conflicting results about the mutagenicity of TDI (Refs. 5.1. and
5.9.). No data has been found to indicate that diisocyanates are
carcinogenic or teratogenic (Refs. 5.1. and 5.9.).
1.1.3. Operations where exposure may occur
The manufacture of polyurethane foams, coatings, and elastomers
potentially exposes a minimum of 100,000 workers to diisocyanates
(Ref. 5.2.). Diisocyanates can be found in paints, insulation,
adhesives, automobile bumpers, shoe soles, and hundreds of other
applications (Refs. 5.2. and 5.8.). Over 700 million pounds of
diisocyanates were produced in 1975 (Ref. 5.2.).
1.1.4. Physical properties
|
analyte |
2,6-TDI |
HDI |
2,4-TDI |
|
CAS no.: |
91-08-7 |
822-06-0 |
584-84-9 |
MW: |
174.16 |
168.20 |
174.16 |
bp, °C(mm Hg): |
96(1.5) |
213(760) |
251(760) |
mp, °C: |
8 |
-55 |
22 |
sp gr(75°C): |
NA1 |
1.05 |
1.22 |
vp, mm Hg: |
NA1 |
0.05 |
0.025 |
color: |
all colorless to pale yellow |
odor: |
all sharp pungent |
flash point(closed cup), °C: |
NA1 |
140 |
127 |
synonyms and structures: |
Figure 1.1.4. |
|
1 not
available |
1.2. Limit defining parameters (the analyte air concentrations
listed through this method are based on an air volume of 15 L and an
extraction volume of 2 mL.)
1.2.1. Detection limit of the analytical procedure
The detection limit of the analytical procedure is the mass of
analyte per injection which will result in a peak whose height is
about 5 times the amplitude of the baseline noise. (Section 4.1.)
Analytical Detection Limit
|
analyte |
2,6-TDI |
HDI |
2,4-TDI |
|
ng/injection |
0.18 |
0.18 |
0.18 |
|
1.2.2. Detection limit of the overall procedure
The detection limit of the overall procedure is the amount of
analyte spiked on the sampling device which allows recovery of an
amount of analyte equivalent to the detection limit of the
analytical procedure. (Section 4.2.)
Detection Limit of the Overall Procedure
|
analyte |
2,6-TDI |
HDI |
2,4-TDI |
|
ng/sample µg/m3 ppb |
24 1.6 0.23 |
33 2.3 0.32 |
19 1.3 0.17 |
|
1.2.3. Reliable quantitation limits
The reliable quantitation limit is the smallest amount of analyte
which can be quantitated within the requirements of at least 75%
recovery and a precision (1.96 SD) of ±25% or better. The reliable
quantitation limits are higher than the detection limits of the
overall procedure to satisfy the precision requirement. (Section
4.3.)
Reliable Quantitation Limits
|
analyte |
2,6-TDI |
HDI |
2,4-TDI |
|
ng/sample µg/m3 ppb |
34 2.3 0.32 |
44 2.9 0.43 |
39 2.5 0.36 |
|
The reliable quantitation limit and detection limits reported in
the method are based upon optimization of the instrument for the
smallest possible amount of analyte. When the target concentration of
an analyte is exceptionally higher than these limits, they may not be
attainable at the routine operating parameters.
1.2.4. Sensitivity
The sensitivity of the analytical procedure is determined by the
slope of the calibration curve over a concentration range 0.5 to 2
times the target concentration. The sensitivity will vary somewhat
with the particular instrument used in the analysis. (Section 4.5.)
Sensitivity of the Analytical Procedure
|
analyte |
2,6-TDI |
HDI |
2,4-TDI |
|
area units per µg/mL |
85600 |
84300 |
159000 |
|
1.2.5. Recovery
The recoveries of the analytes from samples used in the 18-day
storage tests remained above the values presented below. These
values are determined from the calculated regression lines of the
storage graphs. (Section 4.9.) The recovery of analyte from the
collection medium after storage must be 75% or greater.
Recovery, %
|
temp, °C |
2,6-TDI |
HDI |
2,4-TDI |
|
-25 22 |
86.3 86.4 |
81.1 83.0 |
81.3 80.3 |
|
1.2.6. Precision (analytical method)
The pooled coefficients of variation obtained from replicate
determinations of analytical standards at 0.5, 1 and 2 times the
target concentration are presented below. (Section 4.4.)
Pooled Coefficients of Variation
|
2,6-TDI |
HDI |
2,4-TDI |
|
0.009 |
0.013 |
0.009 |
|
1.2.7. Precision (overall procedure)
The overall procedure must provide results at the target
concentrations that are ±25% or better at the 95% confidence level.
The precisions at the 95% confidence level for the
18-day storage test are presented below. (Section 4.9.)
The reported values each include an additional ±5% for sampling
error.
Precision at the 95% Confidence Level, %
|
2,6-TDI |
HDI |
2,4-TDI |
|
14.9 |
15.2 |
13.5 |
|
1.2.8. Reproducibility
Five samples, prepared by vapor spiking, and a draft copy of this
procedure were given to a chemist unassociated with this evaluation.
The samples were analyzed after 6 days of storage at -25°C. The data
listed below are corrected for extraction efficiency (Section 4.8.).
Recovery %
|
|
2,6-TDI |
HDI |
2,4-TDI |
|
SD |
101.5 1.6 |
100.4 2.0 |
105.4 2.4 |
|
1.3. Advantages
1.3.1. The sampling and analytical procedures are specific and
sensitive for several diisocyanates employed in industry (Ref.
5.7.).
1.3.2. The collection system is less cumbersome than the use of a
bubbler.
1.3.3. 1-2PP is more stable and less expensive than
p-nitrobenzyl-N-n-propylamine (nitro reagent).
1.4. Disadvantages
The use of peak ratios to confirm low concentrations of
diisocyanates is impractical due to the small response at 313 nm.
2. Sampling Procedure
2.1. Apparatus
2.1.1. Samples are collected by use of a personal sampling pump
that can be calibrated to within ±5% at the recommended flow rate
with the sampling device in line.
2.1.2. A three-piece styrene cassette containing a glass fiber
filter coated with 0.1 mg of 1-2PP and a backup pad.
(Figure 2.1.2.)
2.1.3. Coated filters are prepared by applying 0.5 mL of a
solution of 0.2 mg/mL 1-2PP in methylene chloride to
each glass fiber filter. The wet filters are allowed to air dry
before placing them in a jar. Vacuum is applied to the jar to remove
residual methylene chloride. (The coated filters used in Method 47
for MDI are also acceptable for this procedure. These filters are
coated with 1 mg of 1-2PP and are prepared as above
except a 2.0 mg/mL solution of 1-2PP in methylene
chloride is used.)
2.1.4. Coated filters should be stored at reduced temperature as
a precaution.
2.2. Reagents
None required.
2.3. Sampling technique
2.3.1. Remove the inlet cover from the three-piece cassette.
Save the cover for installation after sampling.
2.3.2. Attach the cassette in the breathing zone of the employee
to be monitored.
2.3.3. The recommended flow rate is 1 L/min with a recommended
total air volume of 15 L. A longer 240-min sampling time is
permissible to comply with the 1989 change of the PEL.
2.3.4. After sampling for the appropriate time, remove the
sampling device and reinstall the small plug and inlet cover.
2.3.5. Wrap each sample end-to-end with an OSHA Form 21 seal.
2.3.6. With each set of samples, submit at least one blank
sample. The blank should be subjected to the same handling as the
samples except that no air is drawn through it.
2.3.7. Bulk samples submitted for analysis must be shipped in
sealed vials and in a separate container.
2.4. Retention efficiency
2.4.1. Experimental design
Due to present laboratory limitations, controlled test
atmospheres of diisocyanates cannot effectively be generated.
However, the following procedure using a vapor spiking technique was
used as an alternative to study analyte retention. This was done to
approximate the recommended open-face collection of diisocyanates.
A glass syringe barrel equipped with a Luer taper tip was
silanized and silanized glass wool was placed into the syringe. The
Luer tip was inserted into the inlet part of a cassette so that the
tip was flush with the inside surface of the cassette. The other end
of the syringe was attached to a sampling port. The outlet of the
cassette was attached to a vacuum pump. A critical orifice between
the cassette and the pump maintained a constant 1 L/min flow rate.
Dry air samples were prepared by attaching a dry air source to a
manifold inlet. Humid air samples were generated by passing air
through water in a controlled temperature water bath. The humidity
was monitored in the sampling manifold via a humidity probe. The
glass wool was spiked with diisocyanate in methylene chloride. The
desired quantity of air was then drawn through the glass wool, at a
flow rate of 1 L/min, and onto the coated filter, which was analyzed
to determine analyte loss.
2.4.2. Retention results
Humidity affects the ability of a glass fiber filter to retain
derivatized diisocyanates. When a sample ten times the target
concentration is vapor generated and 200 L of dry air (12% humidity)
is drawn through the filter, an average of 95.4% of the
diisocyanates is found on the coated filter. Only 1.2% is found on
the backup pad.
When higher relative humidity (R.H.) is added to the sampling
system, a different result is obtained. After samples were vapor
spiked at the target concentration using 20 L of dry air, several
known volumes of humid air (78% R.H.) pulled through them. The
samples showed increasing losses of diisocyanate derivative with
increasing volumes of humid air. (Section 4.6.)
2.5. Extraction efficiency
The average extraction efficiency for each of the analytes spiked
at the target concentration on a coated glass fiber filter is
presented below. (Section 4.7.)
Average Extraction Efficiencies, %
|
2,6-TDI |
HDI |
2,4-TDI |
|
91.2 |
93.3 |
90.8 |
|
2.6. Recommended air volume and sampling rate
2.6.1. The recommended air volume is 15 L for the OSHA Ceiling
PEL.
2.6.2. The recommended air sampling rate is 1 L/min.
2.6.3. To comply with the 1989 PEL changes, the air volume can be
increased to 240 L to sample for the OSHA TWA-PEL.
2.7. Interferences (sampling)
Any compound, that could be collected on the glass fiber filter
that could react with the 1-2PP or compete with it in the
reaction to derivatize the diisocyanate, should be considered as an
interference. Potential interferences include anhydrides, amines,
alcohols and carboxylic acids.
2.8. Safety precautions (sampling)
The sampling equipment should be attached to the worker in such a
manner that it will not interfere with work performance or safety.
3. Analytical Procedure
3.1. Apparatus
3.1.1. High performance liquid chromatograph equipped with UV
detector, manual or automatic sample injector, and chart recorder.
3.1.2. HPLC stainless steel column capable of separating
diisocyanate derivatives. The column employed in this study was a
25-cm × 4.6-mm i.d. Alltech C8 (10 µm)
stainless steel column.
3.1.3. An electronic integrator, or some other suitable method of
determining peak areas.
3.1.4. Vials, 4-mL with Teflon-lined caps.
3.1.5. Syringes, of convenient sizes for sample and standard
preparations and injections.
3.1.6. Volumetric pipettes and flasks for preparation of
standards.
3.1.7. Suitable glassware for preparation of diisocyanate urea
derivatives.
3.1.8. Micro-analytical balance used to weigh standard
preparations.
3.2. Reagents
3.2.1. Methylene chloride, hexane, acetonitrile, and dimethyl
sulfoxide, HPLC grade.
3.2.2. Water, HPLC grade. Our laboratory employs a commercially
available water filtration system for the preparation of HPLC grade
water.
3.2.3. 1-(2-Pyridyl)piperazine, Aldrich, Milwaukee, WI.
3.2.4. 2,6-TDI, Carbolabs, Inc., New Haven, CT.
3.2.5. HDI, Aldrich, Milwaukee, WI.
3.2.6. 2,4-TDI, Eastman Chemicals, Rochester, NY.
3.2.7. Ammonium acetate, HPLC grade.
3.2.8. Glacial acetic acid.
3.3. Standard preparation
3.3.1. A solution containing 3.5 g of 2,4-TDI in 25
mL of methylene chloride is slowly added to a stirred solution of
7.25 g of 1-2PP in 100 mL of methylene chloride. The
solution is then heated to 35°C for 10 min. Reduce the volume of
methylene chloride to about 10 mL with a stream of dry nitrogen. The
product is precipitated with hexane, (precipitation may start
without adding hexane), filtered, redissolved in a minimal volume of
methylene chloride and reprecipitated. The precipitate is filtered
and washed with hexane (approximate yield is 9 g of the derivative
after being dried by vacuum). This preparation is a modification of
the procedure reported by Goldberg et al (Ref. 5.7.). Derivatives of
the two other diisocyanates are prepared by a similar procedure.
3.3.2. Preparation of working range standards
A stock standard solution is prepared by dissolving the
diisocyanate derivatives into DMSO. To express the derivative as
free diisocyanate, the amount of 2,4-TDI and
2,6-TDI ureas weighed is multiplied by the conversion
factor 0.3479.
MW TDI
MW urea |
= |
174.16
500.61 |
=
0.3479 |
Similarly, the conversion factor for HDI urea is 0.3400.
MW HDI
MW urea |
= |
168.20
494.64 |
=
0.3400 |
All dilutions of the stock solutions are made with acetonitrile
to arrive at the working range.
3.4. Sample preparation
3.4.1. The styrene cassette is opened and the glass fiber filter
is placed into a 4-mL vial so that the filter is flat against the
inside surface of the vial, not folded or crumpled.
3.4.2. Two milliliters of the extracting solution, 90/10 (v/v)
ACN/DMSO, are added.
3.4.3. A cap equipped with a Teflon liner is installed.
3.4.4. The vial is shaken to remove large air bubbles from
between the filter and the glass. Let the vial set for 1 h.
3.5. Analysis
3.5.1. Reverse phase HPLC conditions
column: |
25-cm × 4.7-mm i.d. stainless steel column
packed with 10-µm Alltech C8 or
suitable equivalent. |
mobile phase: |
0.01 M ammonium acetate in 37.5/ 62.5 ACN/water
(v/v) adjusted to pH 6.2 with acetic acid |
flow rate: |
1 mL/min |
UV detector: |
254 and 313 nm |
fluorescence detector: |
240 nm excitation 370 nm emission |
injection size: |
10-25 µL |
chromatogram: |
Figure 3.5.1. |
3.5.2. An external standard procedure is used to prepare a
calibration curve using at least 2 stock solutions from which
dilutions are made. The calibration curve is prepared daily. The
samples are bracketed with analytical standards.
3.6. Interferences (analytical)
3.6.1. Any compound having the same retention time as the
analyte is a possible interference. Benzaldehyde is an interference
for 2,4-TDI urea using the aforementioned analytical
conditions but is not normally expected to be found. Generally,
chromatographic conditions can be altered to separate an
interference.
3.6.2. Retention time on a single column is not proof of chemical
identity. Analysis by an alternate column system, ratioing of
wavelength response, and mass spectrometry are additional means of
identity. (UV spectra for diisocyanate derivatives are shown in
Figures 4.11.1.-4.11.3.)
3.7. Calculations
The concentration in µg/mL of diisocyanate present in a sample is
determined from the area response of the analytes as measured by an
electronic integrator or peak heights. Comparison of sample response
with a least squares curve fit for standards allows the analyst to
determine the concentration of diisocyanate in µg/mL for the sample.
Since the sample volume is 2 mL, the results in
µg/m3 of air are expressed by the following
equation:
µg/m3 = (µg/mL)(2
mL)/(m3 of air sampled)(Extraction Eff.)
3.8. Safety precautions (analytical)
3.8.1. Avoid skin contact with all solvents.
3.8.2. Wear safety glasses at all times.
3.8.3. Avoid exposure to the diisocyanates standards.
4. Backup Data
4.1. Detection limit of the analytical procedure
The detection limit of the analytical procedure was 0.18
ng/injection for all three analytes. This amount produced a peak whose
height was about 5 times the height of the baseline noise. An
injection size of 10 µL was used in the determination of the detection
limits for the analytical procedure. (Figure 4.1.)
4.2. Detection limit of the overall procedure
4.2.1. The following data were obtained by vapor spiking
increasing amounts of the analytes onto sampling devices. An
injection size of 25 µL was used to determine the detection limits
of the overall procedure.
Table 4.2.1. Recoveries Near the Detection Limit
|
analyte |
2,6-TDI |
HDI |
2,4-TDI |
|
spiked |
found |
spiked |
found |
spiked |
found |
|
ng/sample
|
16.9 25.4 33.8 42.2 67.6 84.5 101.4 |
3.5 14.0 27.8 33.9 54.2 61.9 85.8 |
33.9 44.2 66.2 88.2 132.4 |
3.9 44.9 61.0 82.6 133.7 |
19.3 29.0 38.6 57.9 77.2 96.6 115.8 |
12.6 21.1 39.0 61.8 68.7 93.8 120.8 |
|
4.2.2. Graphical presentation of the above data are shown in
Figures 4.2.1.-4.2.3. The detection limits of the overall procedure
determined from the Figures were 24.4 ng/sample for
2,6-TDI, 33.3 ng/sample for HDI, and 19.2 ng/sample for
2,4-TDI.
4.3. Reliable quantitation limit
The following data were obtained by vapor spiking the analytes onto
sampling devices. An injection size of 25 µL was used to determine the
reliable quantitation limits.
Table 4.3. Extraction Efficiency at the Reliable
Quantitation Limit
|
analyte ng/sample |
2,6-TDI 33.8 |
HDI 44.2 |
2,4-TDI 38.6 |
|
%
recovery
SD 1.96
SD |
117.4 103.6 103.6 103.6 103.6 103.6 103.6 103.6
105.3 4.9 9.6 |
124.8 114.7 96.8 114.7 114.7 114.7 96.8 96.8
109.3 10.9 21.4 |
82.9 74.6 70.8 82.9 74.6 70.8 82.9 74.6
76.8 5.3 10.4 |
|
4.4. Sensitivity and precision (analytical method only)
The following data were obtained from multiple injections of
analytical standards.
Table 4.4.1. 0.5× Target Concentration
|
analyte µg/mL |
2,6-TDI 0.700 |
HDI 0.722 |
2,4-TDI 0.704 |
|
area counts
SD CV |
69054 69310 69380 68824 68117 67271 68701 68643 67196
68499.6 811 0.0118 |
70015 70643 70996 70340 68751 68445 69385 69036 68454
69562.8 967 0.0139 |
127935 127591 127408 125457 124953 124032 126054 125588 124185
125911.4 1454 0.0115 |
|
Table 4.4.2. 1× Target Concentration
|
analyte µg/mL |
2,6-TDI 1.400 |
HDI 1.443 |
2,4-TDI 1.407 |
|
area counts
SD CV |
127643 126872 126332 127445 126896 126037 127077 126384 127033
126857.7 526 0.0041 |
129539 130474 128313 128379 129521 128186 129882 125878 128370
128726.9 1346 0.0105 |
236004 235664 233651 234337 234274 231355 234258 229449 234524
233723.7 2076 0.0089 |
|
Table 4.4.3. 2× Target Concentration
|
analyte µg/mL |
2,6-TDI 2.800 |
HDI 2.886 |
2,4-TDI 2.814 |
|
area counts
SD CV |
249771 244922 248641 246677 246986 245615 252601 248169 248014
247932.9 2309 0.0093 |
252219 249296 259363 252678 252581 250940 247011 249906 251679
251741.4 3396 0.0135 |
459331 457553 458572 461448 461119 457897 463557 460536 459259
459919.1 1925 0.0042 |
|
Table 4.4.4. Pooled Coefficients of Variation
|
2,6-TDI |
HDI |
2,4-TDI |
|
0.0090 |
0.0127 |
0.0087 |
|
4.5. Sensitivity
The data in Tables 4.4.1.-4.4.3. are presented graphically in
Figures 4.4.1.-4.4.3.
4.6. Retention efficiency
4.6.1. Two retention studies were conducted, the first at 12%
relative humidity and the second at 78% relative humidity. The
samples were vapor spiked and removed from the sample generator
after a known volume of air had passed through the cassette.
Table 4.6.1.1. Percent Retention at 10×
Target Concentration with 200-L Air Volume (12% R.H.)
|
analyte µg/sample |
2,6-TDI 27.92 |
HDI 36.44 |
2,4-TDI 31.84 |
|
filter backup
filter backup |
96.9 1.0
95.6 0.9 |
97.2 2.0
95.6 1.8 |
94.4 0.8
92.9 0.6 |
|
Table 4.6.1.2. Percent Retention at 1× Target
Concentration (78% R.H.)
|
air volume, L |
2,6-TDI |
HDI |
2,4-TDI |
|
5.25 5.25 10.5 15.75 15.75 21.0 21.0 26.25 26.25 31.5 36.75 42.0 42.0 47.25 47.25 52.5 52.5 |
90.8 90.3 91.2 89.7 89.7 89.8 85.1 88.8 84.0 84.5 84.7 86.8 85.9 84.9 84.0 87.4 86.4 |
91.5 88.4 89.8 92.0 86.7 90.0 88.4 93.8 92.4 87.5 89.1 90.3 90.0 84.7 84.4 90.9 87.2 |
85.1 84.0 84.5 82.6 78.9 82.3 77.4 81.7 78.2 77.1 80.0 80.1 79.7 79.2 75.7 80.8 79.4 |
|
4.6.2. The following data are presented to show that the
diisocyanate derivatives, liquid spiked, are retained on the coated
glass fiber filter at the recommended air volume.
Table 4.6.2. Percent Retention at 1×
Target Concentration with 20-L Air Volume (80% R.H)
|
analyte µg/sample |
2,6-TDI 2.792 |
HDI 3.644 |
2,4-TDI 3.184 |
|
%
recovery
SD |
83.7 93.1 90.1 95.8 89.4 83.6 78.9 88.6
87.9 5.5 |
79.6 81.4 81.1 81.7 80.8 78.9 75.0 82.3
80.1 2.3 |
76.0 88.5 86.3 91.4 86.5 78.9 73.0 82.7
82.4 6.4 |
|
4.6.3. Ten liters of 80% R.H. air were drawn through a filter to
moisten it and then it was vapor spiked with 20 L of dry air to
observe the retention of the derivative on the wet filter.
Table 4.6.3. Recoveries From a Wet Filter
|
analyte µg/sample |
2,6-TDI 2.792 |
HDI 3.644 |
2,4-TDI 3.184 |
|
%
recovery
SD |
100.5 99.6 97.8 104.2 97.8
100.0 2.6 |
91.6 90.6 88.8 95.9 89.7
91.4 2.8 |
84.4 79.4 77.8 84.4 81.7
81.5 3.0 |
|
4.6.4. Retention efficiencies at the 1989 TWA-PEL
The following data are presented to show that the diisocyanate
derivatives, liquid spiked, are retained on the coated glass fiber
filter at the recommended air volume when sampling for the long
periods of time needed to determine the TWA exposure. No isocyanate
derivative was detected on any of the glass fiber filters placed
0.25 in. behind the coated filters.
Table 4.6.4 Percent Retention at 1× 1989 TWA PEL with
240-L Air Volume (71% R.H)
|
analyte µg/sample |
2,6-TDI 8.412 |
HDI 8.240 |
2,4-TDI 8.376 |
|
%
recovery
SD |
103.1 100.3 102.7 98.7 97.1 96.7
99.8 2.7 |
103.5 103.1 102.3 102.6 102.0 102.0
102.6 0.6 |
106.6 106.1 105.9 106.6 105.3 105.4
106.0 0.6 |
|
4.7. Extraction efficiency
The following data represent the analysis of coated glass fiber
filters vapor spiked with the analytes at 0.05 and 1 times the target
concentrations.
Table 4.7.1. Extraction Efficiency at 0.05× Target
Concentration
|
analyte µg/sample |
2,6-TDI 0.1396 |
HDI 0.1822 |
2,4-TDI 0.1592 |
|
%
recovery
|
86.0 92.8 80.2 84.2 69.3 89.4 91.7 95.1 77.4 91.7 103.2 94.6
88.0 |
93.9 90.0 91.7 92.2 91.3 104.9 96.1 91.7 85.6 96.6 107.6 99.6
95.1 |
98.6 102.1 98.5 100.9 100.1 111.3 96.1 95.6 87.7 101.6 108.2 100.0
100.1 |
|
Table 4.7.2. Extraction Efficiency at 1× Target
Concentration
|
analyte µg/sample |
2,6-TDI 2.792 |
HDI 3.644 |
2,4-TDI 3.184 |
|
%
recovery
|
92.0 95.6 92.6 92.4 91.8 93.7 88.3 89.6 90.2 90.8 87.7 89.9
91.2 |
92.2 98.9 94.1 92.9 92.9 94.9 94.5 92.8 94.3 91.5 88.6 92.3
93.3 |
93.0 98.1 92.9 94.4 92.0 93.9 85.8 85.5 88.6 90.5 87.5 87.6
90.8 |
|
4.8. Reproducibility data
Five samples were spiked with the three diisocyanates and had 20 L
of humid air drawn through the cassettes. The samples were analyzed by
a chemist unassociated with this evaluation after being stored for 6
days at -26°C. The results are corrected for extraction efficiencies.
Table 4.8. Reproducibility Results, % Recovery
|
analyte µg/sample |
2,6-TDI 2.792 |
HDI 3.644 |
2,4-TDI 3.184 |
|
SD |
102.5 98.8 102.7 102.5 101.2
101.5 1.6 |
101.3 97.0 102.0 101.3 100.6
100.4 2.0 |
106.2 103.4 108.6 106.2 102.6
105.4 2.4 |
|
4.9. Storage data
The data in Tables 4.9.2.-4.9.4. show the effects of storage at
ambient (22°C) and reduced (-20°C) temperatures on vapor
spiked cassettes, which were generated with 20 L of dry air followed
by 3 L of humid air to moisten the system. Except for day zero, three
samples for each of the two storage conditions were analyzed at
intervals over an 18-day period. The results are not corrected for
extraction efficiency. The data are also presented graphically in
Figures 4.9.1.-4.9.6.
Table 4.9.1. Amount Vapor Spiked, µg/Cassette
|
2,6-TDI |
HDI |
2,4-TDI |
|
2.792 |
3.644 |
3.184 |
|
Table 4.9.2 Storage Tests for 2,6-TDI
|
storage time |
% recovery |
(days) |
(refrigerated) |
|
(ambient) |
|
0 0 4 7 11 14 18 |
77.8 89.5 84.5 91.0 80.9 89.6 75.7 |
83.5 84.2 88.3 92.4 85.7 83.7 85.2 |
90.9 87.8 83.5 99.0 81.4 94.6 89.7 |
|
77.8 89.5 89.8 94.6 95.1 103.7 95.3 |
85.5 84.2 89.0 86.5 97.1 99.4 95.5 |
90.9 87.8 91.0 90.5 87.3 103.9 102.0 |
|
Table 4.9.3 Storage Tests for HDI
|
storage time |
% recovery |
(days) |
(refrigerated) |
|
(ambient) |
|
0 0 4 7 11 14 18 |
75.9 91.2 79.6 86.9 76.4 87.5 71.4 |
82.3 81.9 80.8 86.2 80.0 81.7 81.1 |
89.7 83.8 79.3 95.4 75.4 91.2 83.0 |
|
75.9 91.2 84.4 82.1 82.2 85.9 81.8 |
82.3 81.9 83.9 75.1 82.5 82.8 84.4 |
89.7 83.8 81.1 81.9 77.7 88.4 85.0 |
|
Table 4.9.4. Storage Tests for
2,4-TDI
|
storage time |
% recovery |
(days) |
(refrigerated) |
|
(ambient) |
|
0 0 4 7 11 14 18 |
74.6 87.7 80.5 83.3 75.1 83.9 73.8 |
78.6 81.3 87.8 84.0 80.9 78.4 82.6 |
84.7 82.9 79.5 89.3 76.4 88.4 86.2 |
|
74.6 87.7 83.6 80.4 81.4 84.2 79.8 |
78.6 81.3 82.7 72.9 79.8 78.4 82.0 |
84.7 82.9 81.7 78.6 72.1 82.3 82.9 |
|
4.10. Side-by-side sampling
A simple experiment was designed which allowed a bubbler containing
nitro reagent and a glass fiber filter coated with 1-2PP
to be simultaneously vapor spiked from the same 2,4-TDI
atmosphere. This was accomplished by leaching a known amount of
2,4-TDI off a glass wool plug contained in a glass tube
with dilution air which is then passed through a "Y" to each sampler.
The air flow was controlled by calibrated orifices of similar flow
rate down stream from the samplers.
Each sample was analyzed twice and its average was plotted in
Figure 4.10. The differences between the bubbler samples and the
filter samples appear to be random with no discernible bias between
them. The amount of scatter observed in both collection systems was
not expected and probably can be attributed to the experimental
design. The average line plotted in Figure 4.10. represents the
average of all the collected samples and the data is presented below.
Table 4.10. Analysis of Side-By-Side Samples,
µg/m3
|
spike |
average |
collection system |
average |
collection system |
|
1 2
3 4
5 6 7 8 9
10 11 12 13 14 15 16
17 18
19 20 |
192 197.5
164.5 172.5
208.5 231 230 222.5 233.5
226 221.5 226.5 212 212 223.5 225
202.5 219.5
174 331.5 |
F F
B B
F F F F F
F F F F F F F
B B
F F |
207 209.5
162.5 179
224.5 181 244.5 223 216
250.5 146.5 199.5 240.5 218.5 245 296.5
230 176.5
248 269 |
F F
B B
B B B B B
B B B B B B B
B B
F F |
|
F = Glass Fiber Filter with
1-2PP B = Toluene Bubbler with nitro
reagent |
4.11. UV Spectra
Figures 4.11.1.-4.11.3. are the UV spectra of the
1-2PP derivatives of the diisocyanates used in this
study. The three compounds are named below:
CAS no. |
name |
|
2,6-Bis(4-(2-pyridyl)-1-piperazinylcarbamyl)
toluene |
72375-27-0 |
1,6-Bis(4-(2-pyridyl)-1-piperazinylcarbamyl)
hexane |
72375-21-4 |
2,4-Bis(4-(2-pyridyl)-1-piperazinylcarbamyl)
toluene |
4.12. Capacity of an 1-mg coated glass fiber filter
A coated glass fiber filter was challenged with a 65/35 mixture of
2,4-TDI/2-6,TDI. The glass fiber filter
coated with 1 mg of 1-2PP was suspended on an adapter
ring of a standard 37-mm cassette. Another coated filter
was placed on a backup pad in the bottom of the cassette. Four more
adapter rings were placed in front of the suspended filter to allow
the incoming isocyanate to cover the entire filter face and not just
hit the center of the filter. The isocyanate mixture was liquid spiked
onto glass wool that had been placed inside a 13-mm stainless steel
filter holder. The metal filter holder was inserted into the Luer-Lok
fitting of the cassette top. The glass wool was then spiked with 8.52
µg of the TDI mixture. Air was pulled through the cassette and holder
at 1 L/min (72% relative humidity). After 15 min, the air flow was
stopped and the rear filter was changed and replaced with a new one.
The glass wool was spiked again with 8.52 µg of the TDI mixture. This
procedure was repeated until a total of 10 rear filters had been
removed and a total of 85.2 µg of isocyanate had been spiked onto the
glass wool.
When the rear filters were analyzed, none of the showed the
presence of any TDI. The front filter, which was not changed during
the sampling, had collected a total of 66.1 µg of TDI. The other 19.1
µg of TDI was probably lost on the sides of the adapter rings of the
cassette. The 66.1-µg collected represents 7.9 times the 1989 TWA-PEL
when sampling for 4 h.
A second cassette containing filters was also tested in the same
manner and again none of the rear filters contained any isocyanate. In
this test the front filter collected 59.4 µg of TDI or 7.1 times the
1989 TWA-PEL.
toluene-2,6-diisocyanate; 2,6-toluene diisocyanate;
2,6-diisocyanato-1-methylbenzene; isocyanic acid, 2-methyl-1,3-phenylene
ester; 2,6-TDI
1,6-hexamethylene diisocyanate; HDI
toluene-2,4-diisocyanate; 2,4-toluene diisoycanate;
2,4-diisocyanato-1-methylbenzene; isocyanic acid, 4-methyl-1,3-phenylene
ester; 2,4-TDI
Figure 1.1.4. Structures and synonyms of the
diisocyanates.
Figure 2.1.2. A drawing of a sample
cassette.
Figure 3.5.1. Chromatogram of standards of the three
diisocyanates.
Figure 4.1. Analytical detection limit for the
diisocyanates.
Figure 4.2.1. Detection limit of the overall procedure for
2,6-TDI.
Figure 4.2.2. Detection limit of the overall procedure for
HDI.
Figure 4.2.3. Detection limit of the overall procedure for
2,4-TDI.
Figure 4.4.1. Calibration curve for
2,6-TDI.
Figure 4.4.2. Calibration curve for
HDI.
Figure 4.4.3. Calibration curve for
2,4-TDI.
Figure 4.9.1. Ambient storage test for
2,6-TDI.
Figure 4.9.2. Ambient storage test for
HDI.
Figure 4.9.3. Ambient storage test for
2,4-TDI.
Figure 4.9.4. Refrigerated storage test for
2,6-TDI.
Figure 4.9.5. Refrigerated storage test for
HDI.
Figure 4.9.6. Refrigerated storage test for
2,4-TDI.
Figure 4.10. Side-by-side comparison of coated filters and
bubblers.
Figure 4.11.1. UV spectrum of 2,6-TDI derivative in
acetonitrile.
Figure 4.11.2. UV spectrum of HDI derivative in
acetonitrile.
Figure 4.11.3. UV spectrum of 2,4-TDI derivative in
acetonitrile.
5. References
5.1. "Criteria for a Recommended Standard...Occupational Exposure
to Toluene Diisocyanate", Department of Health, Education and Welfare,
National Institute for Occupational Safety and Health: Cincinnati, OH,
1973; HMS 73-11022.
5.2. "Criteria for a Recommended Standard...Occupational Exposure
to Diisocyanates"; Department of Health, Education and Welfare,
National Institute for Occupational Safety and Health: Cincinnati, OH,
1978; DHEW (NIOSH) Publ. (U.S.), No. 78-215.
5.3. Kormos, L.H.; Sandridge, R.L.; Keller, Anal. Chem.
1981, 53, 1125.
5.4. Sango, C.; Zimerson, E. J. Liq. Chromatogr. 1980, 3,
971.
5.5. Hardy, H.L.; Walker, R.F. Analyst 1979, 104,
890.
5.6. Ellwood, P.A., Hardy, H.L.; Walker, R.F. Analyst 1981,
106, 85.
5.7. Goldberg, P.A.; Walker, R.F.; Ellwood, P.A.; Hardy, H.L. J.
Chromatogr. 1981, 212, 93.
5.8. Hosein, H.R.; Farkes, S. Am. Ind. Hyg. Assoc. J. 1981,
42, 663.
5.9. Woolrich, P.F. Am. Ind. Hyg. Assoc. J. 1982, 43,
89.
5.10. Cummins, K., Diisocyanates, 2,4-TDI and MDI
(Method 18, Organic Methods Evaluation Branch, OSHA Analytical
Laboratory, Salt Lake City, Utah) unpublished (2-80).
|