N-NITROSOMORPHOLINE
Method no.: |
17 |
|
Matrix: |
Air |
|
Target concentration: |
1.2 ppb (5.5 µg/m3) |
|
Procedure: |
Samples are collected by drawing a known volume of
air through two sampling tubes connected in series. The first tube
contains Polar Partition and the second tube contains Florisil; both
adsorbents are coated with ascorbic acid. Samples are desorbed with
75/25 (v/v) methylene chloride/methanol. Analysis is by gas
chromatography using a chemiluminescence detector. |
|
Recommended air volume and sampling rate: |
40 L at 0.2 L/min |
|
Detection limit of the overall procedure: |
84 parts per trillion (ppt) (0.4
µg/m3) |
|
Reliable quantitation limit: |
116 ppt (0.6 µg/m3) |
|
Standard error of estimate at the target
concentration: (Section 4.6.) |
6.0% |
|
Special requirements: |
The treated adsorbent tubes must be protected from
light during and after sampling. The samples should be stored in a
freezer. |
|
Status of method: |
A sampling and analytical method which has been
subjected to the established evaluation procedures of the Organic
Methods Evaluation Branch. |
|
Date: January 1980 |
Chemist: Warren
Hendricks |
Organic Methods Evaluation Branch OSHA Analytical
Laboratory Salt Lake City, Utah
1. General Discussion
1.1. Background
1.1.1. History of analysis
Sample collection: N-Nitrosomorpholine (NMOR)
vapors have been collected using impingers containing 1 N KOH as the
trapping solution (Ref. 5.1.) and on Tenax GC cartridges (Ref.
5.2.).
Analytical: Analytical procedures for NMOR are similar to
those used for dimethylnitrosamine (DMN) and diethylnitrosamine
(DENA). The separation is usually by gas chromatography and
detection is by one or more of the following: mass spectrometry,
Coulson Electrolytic Conductivity Detector, Hall Electrolytic
Conductivity Detector, nitrogen selective alkali flame ionization
detector and Thermal Energy Analyzer (TEA). (Refs. 5.3. and 5.4.)
Liquid chromatographic procedures that utilize both ultraviolet and
TEA detection systems have been reported. (Ref. 5.1.)
1.1.2. Discussion concerning the sampling and analytical method
This method recommends the use of two treated adsorbent tubes in
series as an air sampling device for NMOR. The analytical
separations are performed by gas chromatography and detection is
accomplished using the TEA. The TEA detector was selected because it
is both sensitive and selective for nitrosamines.
An important requirement of any air sampling technique for
nitrosamines is that the artifactual formation of the analyte on the
sampling device from its precursors must be prevented. The sampler
recommended for DENA (Florisil coated with Vitamin E) was found to
be inadequate for the collection of NMOR because of artifact
problems. Artifact formation experiments showed that Vitamin E did
not prevent nitrosation from occurring in the presence of nitrogen
oxides and the NMOR precursor amine morpholine. Not only did
Florisil effectively collect (and thus concentrate) morpholine,
Florisil coated with Vitamin E was an even more effective collector
of morpholine than Florisil alone. Therefore, morpholine was
available to react with nitrogen oxides to form NMOR. This reaction
occurred before the nitrogen oxides could be removed by the Vitamin
E. These artifact formation experiments involved drawing air
containing nitrogen oxides through the sampler after it had received
a loading of morpholine.
Further experiments showed that Vitamin C coated Polar Partition
(OV-275 on Chromosorb A) when placed ahead of Vitamin C coated
Florisil prevented the artifactual formation of NMOR. When a coated
Polar Partition tube was placed in series ahead of a coated Florisil
tube, NMOR passed through the first tube and was collected on the
second tube. Morpholine was not effectively collected on the first
tube, and experiments showed that the artifactual formation of NMOR
on the first tube did not occur. Also, experiments with the two
tubes in series showed that even if morpholine did reach the second
tube, artifactual formation of NMOR did not occur there since the
nitrosating agent was removed by the first tube. Therefore, the
sampler which was evaluated for this method consisted of two
adsorbent tubes connected in series. The first contained Polar
Partition and the second contained Florisil. The contents of each
tube were coated with 10 mg of Vitamin C.
The recommended NMOR air sampler is probably not suitable to
sample DMN or DENA vapors because the device does not have the
quantity of Florisil adsorbent that is necessary to collect and
retain the more volatile nitrosamines. An air sampler with
sufficient capacity for all three analytes could be assembled,
however, it is unlikely that this device would be a universal air
sampler for all nitrosamines because each analyte has unique
properties that must be considered. It may be possible, when we have
evaluated several samplers and nitrosamines, to develop a more
general sampling device.
1.1.3. Toxic effects (This section is for information only and
should not be taken as the basis for OSHA policy)
Acute: The LD50 for NMOR,
administered by intraperitoneal injection to the rat, was 282 mg/kg.
(Ref. 5.6.) The LD50, following
administration of NMOR with an esophageal probe to rats, was 320
mg/kg. (Ref. 5.7.) The agent was much more toxic when administered
by intravenous injection - the LD50 was 98
mg/kg for rats (Ref. 5.7.). The most important result of acute
exposure to NMOR is extreme destruction of centrilobulor liver
tissue. (Ref. 5.8.) The acute toxic effects normally develop within
one or two days and a lethal dose usually kills the animal within a
week. (Ref. 5.8.)
Chronic: NMOR has been reported to be carcinogenic to the
European hamster (Ref. 5.9.), the mouse (Ref. 5.10.), the Syrian
golden hamster (Ref. 5.11.), aquarium fish (Ref. 5.12.), to the
guinea pig (Ref. 5.13.), and to the rat (Ref. 5.14.). The agent
induced tumors in the lungs, liver, nasal cavity, trachea, and upper
digestive tract following different modes of administration which
included subcutaneous injection, oral ingestion, inhalation, skin
painting, and intravenous injection. In all cases, animal survival
decreased with increasing NMOR dosage and tumor yields of up to 100%
were obtained.
NMOR was reported to be the most rapidly acting liver carcinogen
of the 65 different nitrosamines administered to rats by Druckey, et
al. When 16 rats were given 8 µg/kg (1/40 of the
LD50) NMOR daily in the drinking water, 14
animals died exclusively of liver cancer after an average tumor
induction period (t50) of 165 days. The
average total dose which was applied up to the appearance of cancers
in 50% of the animals (D50) was 1.3 g/kg.
In contrast to the oral administration, when 5 rats were given 10
mg/kg (1/10 of the LD50) once per week by
intravenous injection, 2 animals died of cancer of the nasal region
and 1 of liver cancer. The D50 was 0.5
g/kg and the t50 was 440 days. Perhaps the
reason for the apparent weaker carcinogenic effect following
intravenous administration was that much lower doses were given in
consideration of the higher toxicity (lower
LD50) of intravenous as compared to oral
administration. (Ref. 5.7.)
Data regarding the toxicity of NMOR to humans are unavailable,
however, the wide range of species that are subject to
carcinogenesis by the agent indicates that humans may also be
susceptible. While there is no direct evidence that exposure to
nitrosamines leads to cancer in humans, indirect evidence obtained
from studies that measured the relative metabolic rates of DMN by
rat and human liver slices, indicate that man is probably about as
sensitive to the carcinogenic action of DMN (and perhaps other
nitrosamines) as is the rat. (Ref. 5.15.)
1.1.4. Operations where exposure occurs
NMOR is extensively used in cancer research facilities. Human
exposure can occur when unchanged NMOR is excreted by the laboratory
animals.
Many amines have been shown to be contaminated with the
corresponding N-nitrosamine. (Ref. 5.16.) Exposure to
NMOR can occur in operations where morpholine (the parent amine for
NMOR) is used or produced. Further, if the amine is used as a
chemical intermediate, NMOR can appear in the reaction product.
It has been shown that secondary and tertiary amines can react
with oxides of nitrogen in the vapor phase to produce the
corresponding nitrosamines. (Ref. 5.17.) Studies conducted in our
laboratory confirm that morpholine is very easily nitrosated by
nitrogen oxides and that ambient winter Salt Lake City air is
sufficient to nitrosate the amine.
Potential exposure to NMOR exists whenever morpholine vapors are
present because of the ease with which the agent can be nitrosated.
Further, any tertiary amine which contains the morpholine moiety can
be nitrosated to give NMOR. Industrial uses for morpholine include:
as a rubber accelerator; a solvent; a reactant for organic
synthesis; an additive to boiler water; in waxes and polishes; an
optical brightener for detergents; a corrosion inhibitor and for
preservation of book paper. (Ref. 5.18.)
The fairly common practice of the addition of steam to ambient
winter air in order to increase the relative humidity of the air for
its subsequent use in heated areas is an example of potential
exposure to NMOR. Morpholine is a corrosion inhibitor that is often
added to boiler water and steam generated from such boiler water
contains morpholine which can be nitrosated by the nitrogen oxides
found in ambient air. (Ref. 5.19.)
NMOR has been reported to be present in the tire and rubber
industry. The agent was found as an air pollutant in a tire chemical
plant and an aircraft tire factory. NMOR, found in the chemical
plant, was present in waste water (0.003 µg/g), a dirt scraping from
a stairway (730 µg/g), a soil sample taken outside the plant (4.4
µg/g). Air samples taken in the chemical plant varied from 0.07
µg/m3 in the lunchroom to 5.1
µg/m3 near a diphenylnitrosamine reactor.
The agent was found to be present in every site sampled at the tire
factory. The lowest air concentration was 0.6
µg/m3 in the finishing and inspection area
and the highest, 27 µg/m3 adjacent to a
tire tread extruder. The source of the compound is unknown, however,
morpholine levels of 230 and 42 µg/m3 were
found at the tire factory and chemical plant, respectively. (Ref.
5.1.)
NMOR has been determined to be present in morpholine-nitrite
salts used as vapor-phase corrosion inhibitors (Ref. 5.20.) and in
certain hydraulic fluids. (Ref. 5.21.)
A non-occupational source of exposure to NMOR is the endogenous
formation of the agent in the gastrointestinal tract. Morpholine has
been shown to react with nitrite to form NMOR in the stomachs of
Syrian hamsters. (Ref. 5.22.) The amine has been reported to react
with human saliva to form NMOR. (Ref. 5.23.) Morpholine is used as a
food additive, it is permitted in food intended for human
consumption. (Ref. 5.24) Sources of dietary nitrite include the
reduction of nitrate by microorganisms and nitrite cured food
products. (Ref. 5.23.)
1.1.5. Number of workers that are exposed - unknown
1.1.6. Physical properties (Refs. 5.6. and 5.7.)
synonyms: |
Morpholine, N-nitroso; 4-nitrosomorpholine;
n-nitrosomorpholin (German) |
CAS no.: |
59-89-2 |
molecular formula: |
C4H8N2O2
(MOR-NO) |
molecular structure: |
Figure 1.1.6. |
molecular weight: |
116.14 |
boiling point: |
96°C (6 mm Hg) |
melting point: |
29°C |
absorp. spec.: |
|
(in water) |
|
solubility in water: |
unlimited |
distribution coefficient (concentration in
hexane to concentration in pH 7.4 phosphate buffer): |
0.039 |
1.2. Limit defining parameters
1.2.1. Detection limit of the analytical procedure
The detection limit of the analytical procedure is 80 pg (0.016
µg/mL × 5 µL injection volume) per injection. This is the amount of
analyte which will give a peak whose height is about five times the
amplitude of the baseline noise. (Section 4.1.)
1.2.2. Detection limit of the overall procedure
The detection limit of the overall procedure is 0.016 µg/ sample
(84 ppt or 0.4 µg/m3). This is the amount
of NMOR spiked on the treated adsorbent tubes which allows recovery
of an amount of the analyte equivalent to the detection limit of the
analytical procedure. (Section 4.2.)
1.2.3. Reliable quantitation limit
The reliable quantitation limit for the analytical method was
0.022 µg/sample (116 ppt or 0.6 µg/m3).
This is the smallest amount of analyte which can be quantified
within the requirements of 75% recovery and 95% confidence limits of
±25%. The reliable quantitation limit was determined by spiking the
sampling device with the analyte. The average recovery from treated
Florisil tubes was 96.8% and the 95% confidence interval limits were
±17.2%. The average recovery from treated Polar Partition tubes was
98.7% and the 95% confidence interval was ±18.5%. (Section 4.3.)
The reliable quantitation limit and detection limits reported in
the method are based upon optimization of the instrument for the
smallest possible amount of analyte. When the target concentration of
an analyte is exceptionally higher than these limits, they may not be
attainable at the routine operating parameters.
1.2.4. Sensitivity
The sensitivity of the analytical procedure over a concentration
range representing 0.5 to 2 times the target concentration based on
the recommended air volume is 44085 area units (HP-5840A) µg/mL. The
sensitivity is determined by the slope of the calibration curve.
(Figure 4.4.1.) The sensitivity will vary somewhat with the
particular instrument used in the analysis. A representative
chromatogram is shown in Figure 4.4.2.
1.2.5. Recovery
The average recovery over the range of 0.5 to 2 times the target
concentration was 96.3% for treated Florisil tubes and 96.7% for
treated Polar Partition tubes. (Section 4.5.) The recovery remained
above 81% during the 19-day ambient temperature storage period.
(Section 4.6.) The recovery of analyte from the collection medium
must be 75% or greater.
1.2.6. Precision of the analytical method
The pooled coefficient of variation obtained from replicate
determinations of analytical standards at 0.5, 1, and 2 times the
target concentration was 0.028. (Section 4.7.)
1.2.7. Precision of the overall procedure
The overall procedure must provide results at the target
concentration that are ±25% or better at the 95% confidence level.
The 95% confidence limits for the 19 day ambient temperature storage
test are ±12.6%. (Section 4.6.). This includes an additional ±5% for
sampling error.
1.2.8. Reproducibility
The analytical method was shown to be reproducible by the
analysis of liquid-spiked samples by a second chemist who was not
associated with this evaluation. The average recovery from treated
Polar Partition tubes spiked with 0.22 µg of NMOR was 97.9%. The
average recovery from treated Florisil tubes spiked with 0.11 µg of
NMOR was 100.3%. (Section 4.8.)
1.3. Advantages
1.3.1. The sampling procedure is convenient.
1.3.2. The significance of artifactual formation of NMOR upon the
sampling device has been eliminated through pretreatment of the air
sampler.
1.3.3. The analytical procedure is quick, sensitive, and
reproducible.
1.3.4. Reanalysis of the samples is possible.
1.3.5. It may be possible to determine other nitrosamines
simultaneously.
1.3.6. The effects of potential interferences are reduced through
the use of a selective detector (the TEA) and can be further reduced
by proper selection of GC parameters.
1.3.7. The air sampler is composed of commercially available
materials.
1.4. Disadvantages
1.4.1. At this time the sampling method has not been field
tested.
1.4.2. The cost of the TEA may be prohibitive to small
laboratories.
2. Sampling Procedure
2.1. Apparatus
2.1.1. An approved and calibrated personal sampling pump whose
flow can be determined to ±5% at the recommended flow rate.
2.1.2. Florisil and Polar Partition adsorbent tubes: Glass tubes,
6-mm o.d., 4-mm i.d., 7-cm length, containing 100-mg
front and 50-mg rear (separated by a 2-mm portion of urethane foam
or silylated glass wool) sections of 20/40 mesh Florisil and Polar
Partition adsorbent. SKC Inc. Catalog Nos. 226-39 and 226-26 or
equivalent. Each tube is pretreated with 10 mg
L-(+)-ascorbic acid. Instructions for the preparation
of the coated sampling tubes are given in Section 4.9.
2.2. Reagents
None required.
2.3. Sampling technique
2.3.1. The air sampler is composed of a treated Polar Partition
tube followed by a treated Florisil tube in series. The tubes are
easily connected with an end cap that has been modified by cutting
off the closed portion.
2.3.2. Connect the air sampler to the sampling pump with flexible
tubing. The 50-mg section of each tube should be positioned toward
the sampling pump. Cover each tube of the air sampler with masking
tape or other suitable material to prevent light from reaching the
adsorbent.
2.3.3. The air sampler should be placed in a vertical position
during sampling to minimize channeling.
2.3.4. Sampled air should not pass through any hose or tubing
before entering the sampling device.
2.3.5. Immediately after sampling, separate the air sampler into
its component tubes, identify each tube as front or backup and seal
each tube with plastic end caps. Also, wrap each sample end to end
with official OSHA seals.
2.3.6. With each batch of samples, submit at least one blank tube
of each adsorbent material from the same lot used for samples. These
tubes should be subjected to exactly the same handling as the
samples (seal, transport) except that no air is drawn through them.
2.3.7. Transport the samples (and corresponding paperwork) to the
lab for analysis.
2.3.8. If bulk samples are submitted for analysis, they should be
transported in glass containers with Teflon-lined caps. The samples
must be kept from light. These samples must not be put in the same
container used for the treated adsorbent tubes.
2.4. Sampler capacity
Breakthrough, from the Polar Partition tube to the front section of
the Florisil tube, was 75% after about 40 L of air containing target
level concentrations of NMOR at 80% relative humidity and 29°C were
sampled at 0.2 L/min.
Breakthrough from the front to the rear section of the Florisil
tube was 5% after about 85 L of air containing twice the target level
of NMOR at 80% relative humidity and 21°C were sampled at 0.2 L/min.
The breakthrough volume remained constant when the flow rate was
increased from 0.2 to 0.5 L /min. The pressure drop across the air
sampler was less than 1 in. of mercury at 0.5 L/min. If larger air
volumes or shorter sampling times are desired, the flow rate can be
increased to 0.5 L/min. The air volume should be no greater than 85 L.
The possibility of interferences will increase as the air volume
sampled is increased.
2.5. Desorption efficiency
2.5.1. The average desorption from L-(+)-ascorbic
acid treated Polar Partition tubes spiked with 112, 224, and 448 ng
of NMOR was 96.7% (Section 4.5.).
2.5.2. The average desorption from L-(+)-ascorbic
acid treated Florisil tubes spiked with 112, 224, 448 ng of NMOR was
96.3% (Section 4.5.).
2.5.3. The spiked tubes represent an air concentration range of
2.8 to 11.3 µg/m3 (0.6 to 2.4 ppb) based
on the recommended air volume.
2.5.4. The desorption efficiency for NMOR may vary somewhat from
one laboratory to another and also from one lot of
L-(+)-ascorbic acid treated Polar Partition and
Florisil tubes to another. Therefore, it is necessary for each
laboratory to determine the desorption efficiency for each lot of
treated adsorbent tubes.
2.6. Recommended air volume and sampling rate
2.6.1. The recommended air volume is 40 L.
2.6.2. The recommended sampling rate is 0.2 L/min.
2.7. Interferences
2.7.1. Since it is possible that the precursors of NMOR, which
are morpholine, and various nitrosating agents (oxides of nitrogen,
nitrites, etc.), are present in the environment, it is conceivable
that NMOR may be formed upon the sampling device and not be present
in the sampled air. Laboratory experiments indicate that it is
possible to form NMOR from its precursors on an untreated air
sampler. Further experiments show that when the tubes are treated
with 10 mg of L-(+)-ascorbic acid, the formation of
NMOR does not take place.
2.7.2. At the present time, it is unknown if any compound would
severely interfere with the collection or retention of NMOR on
treated Polar Partition and Florisil tubes. In general, the presence
of other compounds will reduce the breakthrough volume for a
particular compound.
2.7.3. Any compound which is suspected of interfering with the
collection or analysis should be listed on the sampling data sheet.
2.7.4. Light will decompose NMOR (Ref. 5.7.). The air sampler
must be protected from light during and after sampling.
2.8. Safety precautions
2.8.1. Observe due care when working with the sharp ends of the
air sampler.
2.8.2. Attach the sampling equipment to the worker in such a
manner that it will not interfere with work performance.
2.8.3. Follow all safety practices that apply to the work area
being sampled.
3. Analytical Procedure
3.1. Apparatus
3.1.1. A gas chromatograph interfaced to a Thermal Energy
Analyzer.
3.1.2. A GC column capable of resolving NMOR from the desorption
solvent and potential interferences. The column used in this work
was a 10-ft × 1/8-in. stainless steel column containing 10% Carbowax
20M with TPA on 80/100 mesh Chromosorb W AW.
3.1.3. An electronic integrator or other suitable method to
measure peak area.
3.1.4. An analytical balance capable of accurately weighing to 5
decimal places.
3.1.5. Vials, 2-mL vials with Teflon-lined caps.
3.1.6. Microliter syringes, 5-µL syringe for sample injections
and other convenient sizes.
3.1.7. Pipets, convenient sizes for diluting standards. A 1-mL
dispenser pipet for the desorbing solvent.
3.1.8. Volumetric flasks, convenient sizes for diluting
standards.
3.1.9. Dewar flasks, convenient sizes for liquid nitrogen.
3.1.10. Heating tape, high temperature, heavily insulated.
3.1.11. Variable voltage transformer.
3.2. Reagents
3.2.1. NMOR, authentic primary standard, 98% minimum.
3.2.2. Methyl alcohol, chromatographic grade.
3.2.3. Methylene chloride, chromatographic grade.
3.2.4. Isopropyl alcohol, chromatographic grade.
3.2.5. Ethyl alcohol, U.S.P., 95%.
3.2.6. Gases, purified GC grade helium and medical grade oxygen.
3.2.7. Nitrogen, liquid.
3.3. Sample preparation
3.3.1. The status of the OSHA seal on each sample is noted and
recorded as intact, broken, or none.
3.3.2. The field and laboratory identification numbers on each
sample are checked against those on the sample identification
sheets.
3.3.3. Suitable precautions must be undertaken to prevent the
exposure of the samples to light. NMOR will photodecompose easily.
3.3.4. The contents of the treated Polar Partition (front) tube
(solid sorbent, glass wool and urethane plugs) are added to a 2-mL
vial. The front glass wool plug and the 100-mg section
of the treated Florisil (rear) tube are added to a second 2-mL vial.
The remaining 50 mg of treated Florisil, the glass wool and urethane
plug are added to a third vial.
3.3.5. Analysis of the third vial will indicate if breakthrough
has occurred. If 20% or more of the total NMOR found on the air
sampler is present in the third vial, then breakthrough has occurred
and an appropriate comment is made on the analysis sheet.
3.3.6. Each vial is desorbed with 1.0 mL of desorbing solution.
The desorbing solution is composed of 75 parts methylene chloride
and 25 parts methyl alcohol by volume.
3.3.7. The vials are sealed immediately with Teflon-lined caps
and allowed to desorb for 30 min with intermittent shaking.
3.4. Standard preparation
3.4.1. Stock standards are prepared by diluting a weighed amount
of NMOR with isopropyl alcohol. The stock standard is diluted to the
working range with isopropyl alcohol (See Section 3.8. Safety
Precautions).
3.4.2. A solution composed of 0.22 µg/mL NMOR in isopropyl
alcohol equals an air concentration of 5.5
µg/m3 for a 40-L air sample desorbed with
1.0 mL of desorbing solution. This amount is not corrected for
desorption efficiency.
3.4.3. Standards are stored in dark bottles under refrigeration.
3.5. Analysis
3.5.1. GC conditions
helium (carrier gas) flow rate: |
30 mL/min |
injector temperature: |
205°C |
column temperature: |
205°C |
TEA transfer line temperature: |
210°C |
elution time: |
2.5 min |
3.5.2. TEA conditions
oxygen pressure: |
9 psi |
GC pyrolyzer furnace temperature: |
475°C |
coarse zero: |
high |
calibrate: |
0.0 |
attenuator: |
4 |
cold trap temperature: |
-130°C (ethyl alcohol and liquid
nitrogen) |
Complete instructions for the TEA are found in its manual.
3.5.3. The transfer line between the GC and TEA must be
maintained at about 210°C with a heating tape. The temperature is
controlled with a variable voltage transformer.
3.5.4. Chromatogram (Section 4.4.)
3.5.5. Peak areas are measured by an electronic integrator or
other suitable means.
3.5.6. An external standard procedure is used to prepare a
calibration curve from the analysis of at least three different
standard solutions. The calibration curve is prepared daily. The
integrator is calibrated to report results in micrograms per
milliliter.
3.5.7. Bracket the samples with analytical standards.
3.6. Interferences
3.6.1. The TEA is a highly selective detector for
N-nitrosamines but false positive results have been
reported. (Ref. 5.25.) The sample is separated into its components
by gas or liquid chromatography. The N-nitroso compound, in the
carrier gas of a GC or the mobile phase of an HPLC, enters a flash
catalytic heater or pyrolyzer where the bonds between the two
nitrogen atoms (N-NO) are ruptured forming nitrosyl radicals
(.NO). Organic compounds, solvent and
fragmentation products are collected in a cold trap.
The nitrosyl radicals in the carrier gas are then reacted with
ozone, under vacuum, to form electronically excited nitrogen dioxide
(NO2*). The excited nitrogen dioxide
quickly decays to its ground state emitting light at a
characteristic wavelength, which is measured by a photomultiplier
tube.
Any compound which will form the nitrosyl radical will elicit a
response from the TEA. In addition to N-nitroso compounds, the TEA
will respond to organic and inorganic nitrites, some nitrates,
certain C-nitroso compounds and certain C-nitro compounds.
Any compound that has the same GC column retention time as NMOR
and has a TEA response is an interference.
3.6.2. GC parameters may be changed to circumvent most
interferences.
3.6.3. Retention time on a single GC column is not proof of
chemical identity. Samples should be confirmed by GC/MS or other
suitable means when required.
3.7. Calculations
3.7.1. The integrator value in µg/mL is used for reference only.
More reliable results are obtained by use of the calibration curve.
The peak area, for each standard, is compared to its concentration
in µg/mL and the equation for the best straight line through the
data points is determined by linear regression.
3.7.2. The concentration in µg/mL for a sample is determined by
comparing the area of a particular sample to the calibration curve.
3.7.3. The result obtained from the analysis of each vial is
corrected by the appropriate desorption efficiency and then the
corrected results from the two tubes that compose a particular air
sample are added together.
3.7.4. The air concentration for a sample result is calculated by
the following equation:
NMOR, µg/m3 = (1000)(A)(B)/C
where |
A |
= |
µg/mL from Section 3.7.3. |
|
B |
= |
desorption volume |
|
C |
= |
air volume in
liters |
3.7.5. To convert the results from Section 3.7.4. in
µg/m3 to parts per billion the following
relationship is used:
NMOR, ppb = (µg/m3)(24.46)/116.1
where |
µg/m3 |
= |
result from Section 3.7.4. |
|
24.46 |
= |
molar volume at 25°C and 760 mm Hg |
|
116.1 |
= |
molecular weight of
NMOR |
3.8. Safety precautions
3.8.1. NMOR is an extremely potent animal carcinogen and utmost
care must be exercised when working with this compound.
3.8.2. Avoid skin contact with liquid nitrogen and the solvents.
3.8.3. Confine the use of solvents to a fume hood.
3.8.4. Wear safety glasses in all laboratory areas.
3.8.5. Be certain that the TEA exhaust is connected to a fume
hood.
4. Backup Data
4.1. Detection limit of the analytical procedure
Figures 4.1.1., 4.1.2., and 4.1.3. were used to determine the
detection limit of the analytical procedure.
4.2. Detection limit of the overall procedure
The data in Table 4.2. are graphically represented in Figures
4.2.1. and 4.2.2. in order to determine the detection limit of the
overall procedure.
Table 4.2 Detection Limit Data
|
µg NMOR spiked on treated adsorbent
tubes |
avg. µg NMOR recovered from treated
Florisil tubes |
avg. µg NMOR recovered from treated
Polar Partition tubes |
|
0.110 0.0450 0.0220 0.0160 |
0.104 0.0446 0.0213 0.0159 |
0.108 0.0437 0.0217 0.0170 |
|
4.3. Reliable quantitation limit
Treated adsorbent tubes were each spiked with 0.022 µg of NMOR and
the recoveries were as follows:
Table 4.3. Reliable Quantitation Limit Data
|
trial |
% recovery from treated
Florisil tubes |
% recovery from treated
Polar Partition tubes |
|
1 2 3 4 5
SD 1.96(SD) |
102.3 102.3 88.6 98.9 92.1
96.8 6.2 12.2 |
100.0 96.2 106.7 88.9 101.8
98.7 6.7 13.1 |
|
4.4. Sensitivity
A typical calibration curve is shown in Figure 4.4.1. The slope of
the curve indicates the sensitivity of the method.
A representative chromatogram is shown in Figure 4.4.2.
4.5. Desorption efficiency
The following data represent the recovery of NMOR spiked on treated
Polar Partition and on treated Florisil adsorption tubes. The
desorption study for the Florisil tubes was conducted with the
reference (B) portion removed.
Table 4.5. Desorption Efficiencies from Two
Adsorbents
|
|
Polar Partition |
|
Florisil |
× target conc. ng/sample |
0.5× 112 |
1× 224 |
2× 448 |
|
0.5× 112 |
1× 224 |
2× 448 |
|
desorption efficiency, %
|
96.6 98.3 96.5 101.0 99.0 99.2
98.4 |
91.8 100.0 93.9 98.0 98.0 95.2
96.2 |
99.1 93.3 100.0 88.7 96.4 96.2
95.6 |
|
91.2 100.4 91.7 93.9 93.7 98.5
94.9 |
97.2 100.5 95.5 99.9 94.5 99.2
97.8 |
95.6 101.1 98.8 87.7 93.5 99.7
96.1 |
|
|
Polar Partition = 96.7 |
|
Florisil = 96.3 |
|
4.6. Storage
Samples were generated by the liquid injection of NMOR on untreated
polar partition tubes containing only 50 mg of adsorbent. The tubes
were allowed to equilibrate overnight and then were placed in front of
the treated air sampler. Forty liters of air at 80% relative humidity
and 20°C were drawn through the sampling train. The NMOR on the Polar
Partition tube was desorbed by the humid air and was deposited on the
air sampler. Studies conducted at ambient relative humidity and
temperature indicated that desorption was essentially complete after
25 L of air had passed through the Polar Partition tube (Figure
4.6.1.). In these studies, the spiked polar partition tube was
connected directly to the TEA. The data in Table 4.6. represent the
effects of storage at ambient (21-26°C) and reduced (-20°C)
temperatures on NMOR collected using treated adsorption tubes. The
recoveries are not corrected for desorption efficiency. A graphical
representation of the data may be found in Figures 4.6.2. and 4.6.3.
Table 4.6 Storage Tests
|
storage time |
% recovery |
(days) |
(refrigerated) |
|
(ambient) |
|
0 3 6 7 10 14 19 20 |
97.3 92.5 97.7
100.5 98.0
94.1 |
98.0 97.5 97.3
99.8 97.7
104.5 |
101.4 95.0 98.6
98.9 100.0
100.4 |
|
97.3 101.4
92.0 85.4 90.7 80.2 |
91.8 93.6
92.3 93.0 81.8 82.5 |
99.1 93.6
90.0 89.5 83.4 81.4 |
|
4.7. Precision of the analytical method
These data represent multiple injections of standard solutions. The
injection volume was 5 µL and the concentrations of the standards were
0.11, 0.22, and 0.45 µg/mL. Peaks were integrated by a Hewlett-Packard
5840A Gas Chromatograph.
Table 4.7. Precision of the Analytical Method
|
× target conc. pg/injection |
0.5× 550 |
1× 1100 |
2× 2250 |
|
area
counts
SD CV
= 0.0282 |
5117 5017 5017 4842 5072 4918 4742 4788 5371
4987.11 193.26 0.0388 |
9900 10210 10170 10070 10070 9680 9582 9582 9494
9853.78 290.60 0.0295 |
19830 20530 20490 19490 19210 20280 19530 20008 20450
19979.78 491.46 0.0245 |
|
4.8. Reproducibility
Three treated Polar Partition and three treated Florisil tubes were
spiked (liquid injection) with 0.22 and 0.11 µg NMOR. These samples
were analyzed by a chemist unassociated with the development of this
procedure. The results are shown below.
Table 4.8. Reproducibility
|
µg/sample |
Polar Partition 0.22 |
Florisil 0.11 |
|
% NMOR recovered
|
96.4 105.5 91.8
97.9 |
95.0 112.5 93.3
100.3 |
|
4.9. Sampling tube preparation
4.9.1. Reagents
Methyl alcohol, chromatographic grade.
L-(+)-Ascorbic acid (Vitamin C), USP Grade, 98% min. Prepare a
solution containing 40 mg/mL in methyl alcohol. Store under
refrigeration in a dark bottle.
4.9.2. Technique
Each adsorbent tube is treated with 10 mg of
L-(+)-ascorbic acid. This is accomplished by first
breaking open both ends of each tube and adding 250 µL of the
L-(+)-ascorbic acid solution. Both sections of the tube
should be uniformly wet by the solution. The methyl alcohol is
allowed to evaporate and then both ends of the tubes are sealed with
the plastic end caps supplied with the tubes. The tubes should be
stored in a freezer with suitable precautions taken to prevent their
contamination. The tubes are stable, when stored in a freezer, for
at least a month.
Figure 1.1.6. The molecular structure of
N-Nitrosomorpholine.
Figure 4.1.1. Typical TEA detector
noise.
Figure 4.1.2. A chromatogram of the desorption
solvent.
Figure 4.1.3. Detection limit of the analytical procedure for
N-Nitrosomorpholine.
Figure 4.2.1. Detection limit of the overall procedure for
N-Nitrosomorpholine desorbed from treated Florisil tubes.
Figure 4.2.2. Detection limit of the overall procedure for
N-Nitrosomorpholine desorbed from treated Polar Partition
tubes.
Figure 4.4.1. N-Nitrosomorpholine calibration
curve.
Figure 4.4.2. A representative N-Nitrosomorpholine
chromatogram at the target concentration.
Figure 4.6.1. The rate at which N-Nitrosomorpholine is
desorbed from 50 mg of Polar Partition adsorbent at ambient relative
humidity and temperature.
Figure 4.6.2. Refrigerated temperature storage test for
N-Nitrosomorpholine.
Figure 4.6.3. Ambient temperature storage test for
N-Nitrosomorpholine.
5. References
5.1. J.M. Fajen, G.A. Carson, D.P. Roundbehler, T.Y. Fan, R. Vita,
V.E. Goff, M.H. Wolf, G.S. Edwards, D.H. Fine, V. Reinhold, and K.
Biemann, N-nitrosamines in the Rubber and Tire Industry,
SCIENCE, 205, 1262-1264 (1979).
5.2. J.M. Fajen, G.A. Carson, S. Fan, D.P. Roundbehler, J.
Morrison, I. Krull, G. Edwards, A. Lafleur, W. Herbst, V. Godd, R.
Vita, K. Mills, D.H. Fine and V. Reinhold, N-Nitroso Compounds as Air
Pollutants, presented to the Air Pollution Control Association, Annual
Meeting, Houston, June 26, 1978.
5.3. W. Hendricks, Dimethylnitrosamine (Method 06, Organic Methods
Evaluation Branch, OSHA Analytical Laboratory, Salt Lake City, Utah),
Unpublished, (1979).
5.4. W. Hendricks, Diethylnitrosamine, (Method 13, Organic Methods
Evaluation Branch, OSHA Analytical Laboratory, Salt Lake City, Utah),
Unpublished, (1979).
5.5. M.C. Archer, S.R. Tannenbaum, T.Y. Fan, and M. Weisman,
Reaction of Nitrite with Ascorbate and its Relations to Nitrosamine
Formation, J. Nat. Cancer Inst., 54(5), 1203-1205 (1975).
5.6. Registry of Toxic Effects of Chemical Substances, 1976
edition. (H.E. Christensen, E.J. Fairchild, Eds., B.S. Carroll, R.J.
Lewis, Project Coordinators) U.S. Department of Health, Education and
Welfare, Public Health Service, Center for Disease Control, National
Institute for Occupational Safety and Health, U.S. Institute for
Occupational Safety and Health, U.S. Government Printing Office,
Washington, D.C. (1976).
5.7. H. Druckey, R. Preussmann, S. Ivankovic and D. Schmahl,
Organotrope Carcinogene Wirkungen bei 65 Verschiedenen
N-Nitroso-verbindungen an Bd-Ratten, Z. Krebsforsch, 69(103),
103-201 (1967).
5.8. D.F. Heath and P.N. Magee, Toxic Properties of
Dialkylnitrosamines and Some Related Compounds, Brit. J. Industr.
Med., 19, 276-282 (1962).
5.9. U. Mohr, G. Reznikand, H. Reznik-Schuller, Carcinogenic
Effects of N-Nitrosomorpholine and N-Nitrosopiperidine on
European Hamser (Cricetus Cricetus), J. Natl. Cancer Inst.,
53(1), 231-235 (1974).
5.10. S.S. Mervish, A. Cardesa, L. Wallcave, and P. Shubik,
Induction of Mouse Lung Adenomas by Amines or Ureas Plus Nitrite and
by N-Nitroso Compounds: Effect of Gallic Acid, Thiocyanate and
Caffeine, J. Natl. Cancer Inst., 55(3), 633-636 (1975).
5.11. H. Haas, U. Mohr and F.W. Kruger, Comparative Studies With
Different Doses of N-Nitrosomorpholine,
N-Nitrosopiperidine, N-Nitrosomethylurea and Dimethylnitrosamine in
Syrian Golden Hamsters, J. Natl. Cancer Inst. 52(4), 1295-1300
(1973).
5.12. G.B. Pliss and V.V. Khudoley, Tumor Induction by Carcinogenic
Agents in Aquarium Fish, J. Natl. Cancer Institute, 55(1),
129-132 (1975).
5.13. F.J. Akin and A.E. Wasserman, Effect on Guinea Pigs of
Feeding N-Nitrosomorpholine and its Precursors in
Combination With Ascorbic Acid, Fd. Cosmet. Toxicol., 13,
239-242 (1975).
5.14. S.S. Mirvish, A.F. Pelfrene, H. Garcia and P. Shubik, Effect
of Sodium Ascorbate on Tumor Induction in Rats Treated With Morpholine
and Sodium Nitrite, and With Nitrosomorpholine, Cancer Letters,
2, 101-108 (1976).
5.15. P.N. Magee, Toxicity of Nitrosoamines: Their Possible Human
Health Hazards, Fd. Cosmet. Toxicol., Great Britain, 9, 207-218
(1971).
5.16. B. Spiegelhalder, G. Eisenbrand and R. Preussman,
Contamination of Amines With N-nitrosamines, Angew.
Chem. Int. Ed. Engl., 17(5) 367-368 (1978).
5.17. J.N. Pitts, P. Grosjean, K.V. Cauwenberghe, J.P. Schmid, and
D.R. Fitz, Photooxidation of Aliphatic Amines Under Simulated
Atmospheric Conditions: Formation of Nitrosamines, Nitramines, Amides
and Photochemical Oxidant, Environmental Science and
Technology, 12(8) 946-953 (1978).
5.18. G.G. Hawley (9th ed.), "The Condensed Chemical Dictionary",
Van Nostrand Reinhold Company: New York, 591 (1977).
5.19. Personal Communication between W. Hendricks and D.P.
Roundbehler (January, 1980).
5.20. M.C. Archer and J.S. Wishnok, Nitrosamine Formation in
Corrosion Inhibiting Compositions Containing Nitrite Salts of
Secondary Amines, J. Environ. Sci. Health All(10 & 11),
583-590 (1976).
5.21. J.J. Behen, Determination of N-Nitrosomorpholine
in Hydrolic Fluids: Analytical Methods and Results, Report prepared
for - The U.S. Environmental Protection Agency, Office of Toxic
Substances, Washington, D.C. 20460, by Union Carbide Corporation,
Chemicals and Plastics, Tarrytown Technical Center, Tarrytown, NY
(May, 1978).
5.22. N. Invi, Y. Nishi, M. Taketomi, and T. Yamada, A Short Term
Simple Method for Detection of N-Nitroso Compounds Produced From
Sodium Nitrite and Morpholine in Stomach, Biochemical and
Biophysical Research Communication 81(2) 310-31 (1978).
5.23. J.S. Wishnok and S.R. Tannenbaum, An Unknown Salivary
Morpholine Metabolite, Anal. Chem., 49(8) 715-178 (1977).
5.24. N.I. Sax, Dangerous Properties of Industrial Materials, Van
Nostrand Reinhold Company: New York, 946 (1975).
5.25. I.S. Krull, T.Y. Fan and D.H. Fine, Problem of Artifacts in
the Analysis of N-Nitroso Compounds, Anal. Chem., 50(6),
698-700 (1978).
|