2-NITROPROPANE
Method no.: |
15 |
|
Matrix: |
Air |
|
Target concentration: |
25 ppm (90 mg/m3) (OSHA
PEL) |
|
Procedure: |
Collection on treated Chromosorb 106 tubes,
desorption with ethyl acetate, analysis by GC using a flame
ionization detector. |
|
Detection limit based on recommended air
volume: (for analytical procedure only) |
0.27 ppm (0.98
mg/m3) |
|
Recommended air volume and sampling rate: |
2 L at 0.2 L/min |
|
Coefficient of variation: |
0.023 (for analytical procedure only over the range
of 0.54 ppm to 54 ppm based on the recommended air volume) |
|
Special requirements: |
Chromosorb 106 tubes are pretreated by placing them
in a GC oven overnight at 120°C with a carrier gas flowing through
them. |
|
Status of method: |
Evaluated method. This method has been subjected to
the established evaluation procedures of the Organic Methods
Evaluation Branch. |
|
Date: January 1980 |
Chemist: Duane Lee |
Organic Methods Evaluation Branch OSHA Analytical
Laboratory Salt Lake City, Utah
1. General Discussion
1.1. Background
1.1.1. History
A number of early analytical procedures for nitro paraffins are
reported in the literature (Ref. 5.1.). The only procedure found
which is directed towards industrial hygiene application is a
colorimetric procedure in which the samples are collected in
concentrated sulfuric acid (Ref. 5.2.). The
2-nitropropane is decomposed to nitrous acid which is
combined with resorcinol to form a product which is suitable for
quantitative measurement. Typically, this method suffers specificity
problems with secondary nitro compounds, as well as certain other
nitro compounds, which give positive responses.
The most common procedure for determining air concentrations of
solvent vapors is collection of the vapors with charcoal adsorbent
tubes and analysis by GC after desorption with carbon disulfide.
When this procedure is used to determine 2-nitropropane
there is a partial conversion of the analyte to acetone on the
charcoal surface. This can be confirmed by the analysis of charcoal
tubes spiked with 2-nitropropane. NIOSH produced a
sampling and analytical method which calls for collection of air
samples with Chromosorb 106 adsorbent tubes and analysis by GC after
desorption with ethyl acetate (Ref. 5.3.). This procedure was
selected for evaluation using OSHA evaluation procedures.
1.1.2. Toxic effects (This section is for information only and
should not be taken as the basis of OSHA policy.)
Exposure to concentrations of 2-nitropropane on the
order of 20 to 45 ppm are reported to cause nausea, vomiting,
diarrhea, loss of appetite, and severe headaches in humans. (Ref.
5.4.) Concentrations of 165 to 445 ppm, cause nausea, dizziness,
headaches, and diarrhea in humans (Ref. 5.5.). Cats, which died
within 17 days of exposure to 328 ppm 2-nitropropane,
sustained severe liver damage (Ref. 5.6.). Liver damage attributed
to high concentrations of 2-nitropropane in humans has
been reported (Ref. 5.7.). Liver cancer was found in rats exposed to
207 ppm 2-nitropropane for 6 months (Ref. 5.8.). These
data raise the question of carcinogenic effects in humans. While
additional studies are in progress, NIOSH believes it would be
prudent to handle 2-nitropropane as if it were a human
carcinogen (Ref. 5.8.).
1.1.3. Work population
An estimated 100,000 workers are potentially exposed to
2-nitropropane (Ref. 5.8.).
1.1.4. Use and operations where exposure occur
The following information is taken directly from Reference 5.8.
Solvent systems containing 2-nitropropane are
used in coatings (e.g., vinyl, epoxy, nitrocellulose, and
chlorinated rubber), printing inks, and adhesives. Occupational
exposure to these products may occur in various industries
including industrial construction and maintenance, printing
(rotogravure and flexographic inks), highway maintenance (traffic
markings), shipbuilding and maintenance (marine coatings),
furniture, food packaging, and plastic products.
1.1.5. Physical properties (Refs. 5.9. and 5.10.)
molecular weight: |
89.09 |
density at 25°C: |
0.98290 g/mL |
melting point: |
-91.32°C |
boiling point: |
120°C |
vapor pressure at 20°C: |
12.9 mm Hg |
vapor density (air = 1): |
3.06 |
flash point: |
103°F |
refractive index: |
1.39439 |
critical temperature: |
344°C |
ignition temperature: |
802°F |
lower flammability limit (% by vol. in
air): |
2.6 |
solubility at 25°C,(% by wt compound in
water): |
1.7 |
wt water in compound: |
0.5 |
molecular structure: |
|
other names: |
dimethylnitromethane; isonitropropane; 2-NP;
nitroisopropane |
1.2. Detection limit, precision, sensitivity and working range
1.2.1. The detection limit for the analytical procedure is 2 ng
per injection with a coefficient of variation of 0.055 at this
level. (Section 4.1.) The detection limit was determined using
1.0-µL injections.
1.2.2. The pooled coefficient of variation of the analytical
procedure over the range of 3.95 to 395 µg per sample is 0.023.
(Section 4.2.) This represents an air concentration range of 0.54 to
54 ppm based on the recommended sampling and analytical procedures.
1.2.3. The sensitivity of the analytical procedure over a
concentration range representing 0.02 to 2 times the target
concentration based on the recommended air volume is 916.6 area
units per µg/mL. The sensitivity is determined by the slope of the
calibration curve (Section 4.2.) The sensitivity will vary somewhat
with the particular instrumentation used in the analysis.
1.2.4. The lower limit of the estimated working range, assuming
adequate desorption efficiency, is 0.27 ppm. The upper limit of the
working range is dependent on the capacity of the collection medium.
1.3. Accuracy
1.3.1. The overall procedure must provide results that are
within 25% or better at the 95% confidence interval.
1.3.2. The recovery of analyte from the collection medium after
storage must be 75% or greater.
1.3.3. The overall procedure has met the above validation
criteria. (Section 4.3.)
1.4. Advantages
1.4.1. The sampling procedure is convenient.
1.4.2. The analytical procedure is quick, sensitive, and
reproducible.
1.5. Disadvantages
If other compounds are present, the GC run time must be lengthened
so the late eluters will not interfere with the next sample.
2. Sampling Procedure
2.1. Apparatus
2.1.1. An approved and calibrated personal sampling pump whose
flow can be determined within ±5% at the recommended flow.
2.1.2. Chromosorb 106 tubes: Glass tube with both ends heat
sealed, 7 cm × 6-mm o.d. × 4-mm i.d., containing 100-mg front and
50-mg backup sections of Chromosorb 106 (60/80 mesh). SKC tubes or
equivalent.
2.1.3. The Chromosorb 106 tubes are conditioned prior to use by
heating overnight in a GC oven at 120°C with a carrier gas of helium
or nitrogen flowing through the tubes. Conditioning removes residual
components from the Chromosorb 106 which may interfere with the
analysis.
2.2. Reagents
None required.
2.3. Sampling technique
2.3.1. Immediately before sampling, remove the caps of the
Chromosorb 106 tube. All tubes must be from the same lot.
2.3.2. Connect the Chromosorb 106 tube to the sampling pump with
flexible tubing. The short section of the Chromosorb 106 tube is
used as a backup and should be positioned nearer the sampling pump.
2.3.3. The tube should be placed in a vertical position during
sampling to minimize channeling.
2.3.4. Air being sampled should not pass through any hose or
tubing before entering the Chromosorb 106 tube.
2.3.5. Seal the Chromosorb 106 tube with plastic caps immediately
after sampling. Also, seal each sample with OSHA sealing tape
lengthwise.
2.3.6. With each batch of samples, submit at least one blank tube
from the same lot used for samples. This tube should be subjected to
exactly the same handling as the samples (open, seal, transport)
except that no air is drawn through it.
2.3.7. Transport the samples (and corresponding paperwork) to the
lab for analysis.
2.3.8. If bulk samples are submitted for analysis, they should be
transported in glass containers with Teflon-lined caps. These
samples must not be put in the same container used for the
Chromosorb 106 tubes.
2.4. Breakthrough
Breakthrough tests were performed with the primary section of
Chromosorb 106 tubes using a controlled test atmosphere containing 52
ppm 2-nitropropane with an average relative humidity of
82% at 21°C. The 5% breakthrough volume was 5 L. Breakthrough was
determined by monitoring the downstream effluent.
2.5. Desorption efficiency
The desorption efficiency from liquid injections on Chromosorb 106
tubes averaged 99.3% for 99 to 395 µg per tube, which is 13.6 to 54
ppm for a 2-L air volume. (Section 4.4.)
2.6. Recommended air volume and sampling rate
2.6.1. The recommended air volume is 2 L.
2.6.2. The recommended sampling rate is 0.2 L/min.
2.6.3. If a longer sampling time is required, a sampling rate of
0.05 L/min to 0.1 L/min can be used.
2.7. Interferences (sampling)
2.7.1. At the present time, it is unknown if any compound would
severely interfere with the collection of
2-nitropropane on Chromosorb 106. In general, the
presence of other solvents will decrease the breakthrough air volume
for a particular solvent.
2.7.2. Any compound which is suspected of interfering in the
collection or analysis should be listed on the sampling data sheet.
2.8. Safety precautions
2.8.1. The broken ends of the tubes should be protected to avoid
injury to the person being sampled.
2.8.2. When working in environments containing flammable vapors,
do not provide any spark source from equipment used or pumps.
2.8.3. Observe all safety practices for working in hazardous
areas.
3. Analytical Procedure
3.1. Apparatus
3.1.1. A GC equipped with a flame ionization detector.
3.1.2. A number of GC columns are available and adequate. The
column used for this study was a 1/8-in. × 10-ft, stainless steel,
7% Tetracyanoethylated Pentaerythritol on 100/120 mesh Chromosorb
P-AW.
3.1.3. An electronic integrator or other suitable method of
measuring peak areas.
3.1.4. Two-milliliter vials with Teflon-lined caps.
3.1.5. Microliter syringes, 10-µL for preparing standards, 1-µL
for sample injections.
3.1.6. Pipets for diluting standards. A 1.0-mL pipet for
dispensing solvent for desorption, or a 1.0-mL repipet dispenser.
3.1.7. Volumetric flasks, convenient sizes for preparing
standards.
3.2. Reagents
3.2.1. Ethyl acetate, chromatographic grade.
3.2.2. 2-nitropropane, reagent grade.
3.2.3. Purified GC grade helium, hydrogen, and air.
3.3. Standard preparation
3.3.1. Standards are prepared by diluting pure
2-nitropropane with ethyl acetate.
3.3.2. Five microliters of 2-nitropropane per 25 mL
of ethyl acetate is equivalent to 27.1 ppm for a 2-L air sample
desorbed with 1.0 mL of ethyl acetate.
3.4. Sample preparation
3.4.1. The front and back sections of each sample are
transferred to separate 2-mL vials.
3.4.2. Each section is desorbed with 1.0 mL of ethyl acetate.
3.4.3. The vials are sealed immediately and allowed to desorb for
30 min with intermittent shaking.
3.5. Analysis
3.5.1. GC conditions:
helium (carrier gas) flow rate: |
25 mL/min |
detector (flame ionization) |
|
hydrogen flow: |
30 mL/min |
air flow: |
240 mL/min |
injector temperature: |
150°C |
detector temperature: |
250°C |
column temperature: |
100°C |
injection size: |
1.0 µL |
3.5.2. Chromatogram
A typical chromatogram of 2-nitropropane, using the
recommended GC conditions, is shown in Figure 3.5.2. The
concentration was 197.5 µg/mL (27.1 ppm) and the injection size was
1 µL. The larger peak is ethyl acetate and the smaller peak (7.55
min) is 2-nitropropane.
3.5.3. Peak areas are measured by an electronic integrator or
other suitable means.
3.5.4. An external standard procedure is used. The integrator is
calibrated to report results in ppm for a 2-L air sample after
correction for desorption efficiency.
3.6. Interferences (analytical)
3.6.1. Any compound having the same general retention time as
2-nitropropane is an interference. Possible
interferences are listed on the sample data sheets. GC parameters
should be chosen so these interferences will pose no problems.
3.6.2. GC parameters may be changed to circumvent most
interferences.
3.6.3. Retention time on a single column is not considered proof
of chemical identity. Samples should be confirmed by GC/MS or other
suitable means.
3.7. Calculations
Usually the integrator is programmed to report for a 2-L air
sample. The following calculation is used:
ppm = A/(B/2) |
where |
A = ppm on report B = air volume
(L) |
3.8. Safety precautions
3.8.1. All work using solvents (preparation of standards,
desorption of samples, etc.) should be done in a hood.
3.8.2. Avoid any skin contact with all of the solvents.
3.8.3. Safety glasses should be worn throughout the procedure.
4. Backup Data
4.1. Detection limit
A small amount of analyte (1.98 ng/injection) which produced a well
defined peak was designated as the analytical detection limit. This
was determined with an analytical standard which contained 0.002 µL of
2-nitropropane per milliliter of ethyl acetate (1.98
µg/mL).
Reproducibility of the peak produced by 1.98 ng was good. Eight
injections resulted in a coefficient of variation of 5.5%.
A sample collected from 2 L of air which contained 1.98 ng/µL after
desorption with 1 mL of ethyl acetate would represent an air
concentration of 0.27 ppm.
4.2. Precision data
Multiple injections were made of the standards that were prepared
over a range of 0.02 to 2.2 times the target concentration. A standard
deviation and a coefficient of variation was determined at each
concentration. The pooled coefficient of variation was determined for
the range.
Table 4.2. Analytical Precision
|
× target conc. µg/sample |
0.02× 3.95 |
0.11× 19.57 |
0.54× 98.76 |
1.1× 197.52 |
2.2× 395.04 |
|
area counts
SD CV
= 0.023 |
3785 3620 3814 3672
3723 92 0.0247 |
19266 19397 19095 19126
19221 139 0.0072 |
93540 93869 96997 91388
93948 2311 0.0246 |
183170 185086 182714 186600
184393 1795 0.0097 |
372497 372586 360046 345140
362567 13026 0.0359 |
|
4.3. Storage data
Samples were collected on Chromosorb 106 from a generated
atmosphere containing 25.3 ppm 2-nitropropane with an
average relative humidity of about 81% at 22°C. A storage study was
then conducted in which the collected samples were divided into two
groups; one stored at ambient temperature and the other under
refrigeration. Every few days, three samples from each group were
analyzed. The results are shown in Table 4.3. and in Figures 4.3.1.
and 4.3.2.
Table 4.3. Storage Tests
|
storage time |
% recovery |
(days) |
(-1°C to 6°C) |
|
(21.3°C to 23°C) |
|
1 3 6 10 12 14 |
89.6 85.8 96.2 112.4 99.7 101.4 |
90.4 lost 95.0 112.9 100.0 94.7 |
94.0 87.2 96.8 112.3 104.9 100.6 |
|
88.9 86.9 84.4 103.4 90.0 90.6 |
88.5 91.9 91.8 92.4 92.6 77.2 |
90.0 96.5 89.4 89.7 97.8 78.8 |
|
4.4. Desorption efficiency
Liquid injections were made on the front section of conditioned
Chromosorb 106 tubes at approximately 0.5, 1, and 2 times the target
concentration. These were refrigerated overnight, desorbed and
analyzed the following day. The overall desorption for the
concentration studied is 99.3%.
Table 4.4. Desorption Efficiency
|
× target conc. µg/sample |
0.5× 98.76 |
1× 197.5 |
2× 395.0 |
|
desorption efficiency, %
|
111.7 104.9 92.8 103.7 106.6
102.7 |
98.4 100.4 99.2 92.1 95.1
97.7 |
97.0 97.2 95.7 95.9 99.0
97.5 |
|
Figure 3.5.2. Chromatogram of a standard of
2-nitropropane.
Figure 4.2. Calibration curve of instrument response to
2-nitropropane.
Figure 4.3.1. Ambient storage test of
2-nitropropane.
Figure 4.3.2. Refrigerated storage test of
2-nitropropane.
5. References
5.1. Standen, A. (Ed.), "Kirk-Othmer Encyclopedia of Chemical
Technology". 2nd ed., New York: Interscience Publishers, Vol. 13, pp
875-876, 1964.
5.2. Jones, L. R., American Industrial Hygiene Association
Journal, 24, 11-16, 1963.
5.3. Taylor, D. G. (Manual Coordinator) "NIOSH Manual of Analytical
Methods". 2nd Edition, U.S. DHEW/PHS/CDC/NIOSH, Vol. 4, Method No.
P&CAM 272, 1978.
5.4. Skinner, J. B., Ind. Med., 16, 441, 1974.
5.5. "Documentation of Threshold Limit Values", American Conference
of Governmental Industrial Hygienists, 3rd Ed., p. 188-189, 1971.
5.6. Freon and Dutra Arch. Ind. Hyg. and Occ. Med., 5, 52,
1952.
5.7. Gaultier, M., et.al., Arch. I. Mal. Prof., 25, 425,
1964.
5.8. U.S. DHEW/PHS/CDC/NIOSH "Current Intelligence Bulletin 17"
DHEW (NIOSH), April 25, 1977.
5.9. Patty, F. A. (Ed.) "Industrial Hygiene and Toxicology", 2nd
ed., New York: John Wiley and Sons, Inc., p. 2073, 1963.
5.10. Standen, A. (Ed.), "Kirk-Othmer Encyclopedia of Chemical
Technology". 2nd ed., New York: Interscience Publishers, Vol. 13, p.
866, 1964.
|