DIETHYLNITROSAMINE
Method no.: |
13 |
|
Matrix: |
Air |
|
Target concentration: |
6 µg/m3 (1.4 ppb) |
|
Procedure: |
Collection on two Florisil adsorbent tubes in series,
each of which have been pretreated with 11 mg of DL-a-tocopherol, desorption with a solution
composed of equal volumes methylene chloride and methyl alcohol,
analysis by GC with chemiluminescence detection. |
|
Detection limit based on recommended air
volume: |
0.4 µg/m3 (96 ppt) |
|
Recommended air volume and sampling rate: |
25 L at 0.2 L/min |
|
Standard error of estimate at the target
concentration: (Section 4.5.) |
7.4% |
|
Special requirements: |
The air sampler must be kept from light during and
after sampling. |
|
Status of method: |
A sampling and analytical method which has been
subjected to the established evaluation procedures of the Organic
Methods Evaluation Branch. |
|
Date: August 1979 |
Chemist: Warren
Hendricks |
Organic Methods Evaluation Branch OSHA Analytical
Laboratory Salt Lake City, Utah
1. General Discussion
1.1. Background
1.1.1. History
Sample Collection: Diethylnitrosamine (DENA) vapors have
been collected using the same methods used for dimethylnitrosamine.
These methods include cryogenic techniques that involve the use of
successive cold traps, ambient temperature KOH bubblers and Tenax GC
cartridges (Ref. 5.1.).
Analytical: Analytical procedures for DENA are similar to
those used for dimethylnitrosamine. The separation is usually by gas
chromatography and detection by one or more of the following
techniques: mass spectrometry, Coulson Electrolytic Conductivity
Detector, Hall Electrolytic Conductivity Detector, nitrogen
selective alkali flame-ionization detector, and Thermal Energy
Analysis (Ref. 5.1.).
For several reasons, the existing sampling procedures are not
adequate for use by OSHA personnel. Some of these reasons are - poor
collection efficiency, low extraction efficiency and artifactual
formation of DENA on the sampling device. Therefore, the primary
emphasis of this work has been to develop new air sampling
techniques. The analytical method utilizes separation by gas
chromatography and detection with the Thermal Energy Analyzer (TEA).
The TEA detector was selected because it is both sensitive and
selective for nitrosamines.
1.1.2. Toxic effects (This section is for information only and
should not be taken as the basis of OSHA policy.)
Acute: The LD50 for DENA, administered by
intraperitoneal injection to the rat, is 216 mg/kg. The lowest
published toxic dose is 100 mg/kg also given by intraperitoneal
injection to rats (Ref. 5.2.). The acute toxic effects from exposure
to DENA are similar to those produced by dimethylnitrosamine with
serious destruction of liver tissue as the most important result
(Ref. 5.3.).
Chronic: There is no correlation between acute toxic
effects and carcinogenic potential for nitrosamines. This is
demonstrated by the fact that even though DENA has only about
one-sixth the acute toxicity of dimethylnitrosamine (Ref. 5.2.), if
administered continuously to rats, it is probably a more active
liver carcinogen (Ref. 5.4.).
DENA has been shown to be carcinogenic to the mouse, the rat, the
hamster, the guinea-pig, the rabbit, the dog, the pig,
the monkey, and to aquarium fish. The agent induces tumors primarily
in the nasal cavity, trachea, esophagus, and liver. It causes cancer
after different modes of exposure which include ingestion,
inhalation, and skin painting. It is carcinogenic in single doses
and following prenatal exposure. In lifetime feeding studies with
rats in which daily doses between 1 and 10 mg/kg body weight were
administered, tumor yields approaching 100% have been obtained (Ref.
5.5.).
A dose-response study has been conducted using rats. DENA was
administered in drinking water and the daily exposure was between
0.075 and 14.2 mg/kg body weight in 9 groups of animals. The total
dose, until death occurred, was between 64 and 965 mg/kg body
weight. The tumor induction time was between 68 and 840 days. All
daily doses higher than 0.15 mg/kg body weight gave a tumor
incidence of 100%. When a dose of 0.15 mg/kg body weight per day was
administered, a tumor yield of 90% was obtained. At 0.075 mg/kg body
weight per day, 20 rats survived for more than 600 days and 11 of
the 20 animals had tumors of the liver, esophagus, or the nasal
cavity. All 4 of the animals that lived longer than 940 days at this
dose level had tumors (Ref. 5.5.).
1.1.3. Worker exposure
The chemical reaction, in the condensed phase, between nitrous
acid and diethylamine or triethylamine to form DENA is well known
(Ref. 5.6.). Any tertiary amine that contains the diethyl moiety may
react with a nitrosating agent to form DENA. Recently, it has been
shown that both amines can react with oxides of nitrogen in the
vapor phase to give DENA as a reaction product (Ref. 5.7.). This
means that even though DENA is not used at a particular location, it
may be formed from its precursors and therefore be found in the
occupational environment.
Exposure to DENA can occur during operations in which
diethylamine or triethylamine is utilized or produced. The exposure
results because both amines are often contaminated with the
nitrosamine. Further, if the amine is used as a chemical
intermediate, DENA can possibly appear in the reaction product (Ref.
5.8.).
DENA is extensively used in cancer research facilities. Human
exposure occurs when unchanged DENA is excreted by the laboratory
animals (Ref. 5.9.).
DENA is not widely used by industry today. Uses or proposed uses
of DENA include: as a solvent in the fiber industry, as a softener
for copolymers, as an additive for lubricants, in condensers to
increase the dielectric constant and for the synthesis of
1,1-diethylhydrazine (Ref. 5.5.).
Non-occupational exposure to DENA can be a result of eating a
normal western diet. It has been estimated that the weekly intake of
preformed dialkyl and heterocyclic nitrosamines per person on a
normal English diet is approximately 1 and 3 µg/week respectively
(Ref. 5.10.). Another source of exposure to DENA is the endogenous
formation of the agent in the gastrointestinal tract. Diethylamine
has been shown to react with nitrite to form DENA in human gastric
juice (Ref. 5.5.). DENA has been reported to be a component of
tobacco and tobacco smoke (Ref. 5.11.).
1.1.4. Number of workers that face exposure - unknown
1.1.5. Physical properties (Ref. 5.5.)
CAS no.: |
55-18-5 |
synonyms: |
diethylamine, N-nitroso;
N,N-diethylnitrosamine; DEN; DENA; DANA |
structure: |
|
mol wt: |
102.1 |
physical appearance: |
a yellow, volatile liquid |
boiling point: |
177°C (760 mm Hg) 64-65°C (17 mm Hg) |
density: |
0.9422 (20/4°C) |
refractive index: |
1.4386 (20°C) |
absorp. spec.: |
max 230 nm; log e
3.86 |
(in water) |
max 332 nm; log e
1.98 |
solubility: |
about 10% in water; soluble in organic solvents
and in lipids |
1.2. Detection limit, precision, sensitivity and working range
1.2.1. The detection limit for the analytical procedure is 50 pg.
The coefficient of variation is 0.14 at this level (Section 4.1.).
The detection limit was determined using 5-µL
injections from standard solutions.
1.2.2. The pooled coefficient of variation for the analytical
procedure over the range of 80 to 300 ng per sample is 0.034
(Section 4.2.). This represents an air concentration range of from
3.2 to 12.0 µg/m3 based on the recommended sampling and
analytical procedures.
1.2.3. The sensitivity of the analytical procedure over a
concentration range of 80 to 300 ng (3.2 to 12.0 µg/m3
based upon the recommended air sampling volume of 25 L) is 61918
area units (HP-5840A) per µg/mL. The sensitivity is determined by
the slope of the calibration curve (Section 4.3.). The sensitivity
will vary somewhat with the particular instrumentation used in the
analysis.
1.2.4. The lower limit of the estimated working range, assuming
adequate desorption efficiency, is 0.4 µg/m3. The upper
limit of the working range is dependent on the capacity of the
treated Florisil tubes.
1.3. Accuracy
1.3.1. The overall procedure must provide results that are within
±25% of the true value or better at the 95% confidence interval.
1.3.2. The recovery of analyte from the collection medium during
storage must be 75% or greater.
1.3.3. The overall procedure has met the above validation
criteria (Section 4.5.).
1.4. Advantages
1.4.1. The sampling procedure is convenient.
1.4.2. The significance of artifactual formation of DENA upon the
sampling device has been eliminated through pretreatment of the air
sampler.
1.4.3. The analytical procedure is quick, sensitive, and
reproducible.
1.4.4. Reanalysis of the samples is possible.
1.4.5. The samples are stable, even when stored at room
temperature for 17 days.
1.4.6. Dimethylnitrosamine can be determined using the
recommended sampling and analytical techniques. The desorption
efficiency must be verified.
1.4.7. It may be possible to determine other nitrosamines
simultaneously.
1.4.8. The effects of potential interferences are reduced through
the use of a selective detector (the TEA) and can be further reduced
by proper selection of GC parameters.
1.5. Disadvantage
The relative humidity of the sampled air affects the ability of the
adsorbent to retain the analyte.
2. Sampling Procedure
2.1. Apparatus
2.1.1. An approved and calibrated personal sampling pump whose
flow can be determined to ±5% at the recommended flow rate.
2.1.2. Florisil adsorbent tubes: Glass tubes, 6-mm o.d., 4-mm
i.d., 7-cm length, containing 100-mg front and 50-mg rear (separated
by a 2-mm portion of urethane foam or silylated glass wool) sections
of 20/40 mesh Florisil. SKC, Inc. Catalog No. 226-39 or equivalent.
Each tube is pretreated with 11 mg DL-a-tocopherol. (Section 4.6.)
2.2. Reagents
None required
2.3 Sampling technique
2.3.1. The air sampler is composed of two treated Florisil tubes
in series. The tubes are easily connected with an end cap that has
been modified by cutting off the closed portion.
2.3.2. Connect the air sampler to the sampling pump with flexible
tubing. The 50-mg section of each tube should be positioned toward
the sampling pump. Cover each tube of the air sampler with masking
tape or other suitable material to prevent light from reaching the
adsorbent.
2.3.3. The air sampler should be placed in a vertical position
during sampling to minimize channeling.
2.3.4. Sampled air should not pass through any hose or tubing
before entering the sampling device.
2.3.5. Immediately after sampling, separate the air sampler into
its component tubes, identify each tube as front or backup and seal
each tube with plastic end caps. Also, wrap each samples end to end
with official OSHA seals.
2.3.6. With each batch of samples, submit at least one blank tube
from the same lot used for sampling. This tube should be subjected
to exactly the same handling as the samples (seal, transport) except
that no air is drawn through it.
2.3.7. Transport the samples (and corresponding paperwork) to the
lab for analysis.
2.3.8. If bulk samples are submitted for analysis, they should be
transported in glass vials with Teflon-lined caps. The samples must
be kept from light. Bulk samples must not be put in the same mailing
container used for the treated Florisil tubes.
2.4. Breakthrough
The relative humidity of the sampled air has a significant effect
on the ability of the air sampler to retain DENA. However, laboratory
studies indicate that 25 L of air containing the target concentration
of DENA, at 80% relative humidity and 22°C, can be sampled with no
loss of analyte. Breakthrough from the front to the rear tube, when
the air sampler was challenged with vapors containing 0.15 µg DENA at
these conditions, was about 30%.
No quantitative breakthrough information can be obtained from the
analysis of individual tubes from a particular air sampler, but, if
the amount found on the rear tube is 45% or greater of the total, then
it is possible that some of the analyte may have been lost.
2.5. Desorption efficiency
2.5.1. The average desorption from DL-a-tocopherol treated Florisil adsorbent
tubes spiked with 80, 150, and 300 ng DENA was 95.3% (Section 4.4.).
The spiked tubes represent an air concentration range of 3.0 to 12.0
µg/m3 based upon the recommended air volume.
2.5.2. The desorption efficiency for DENA may vary from one lab
to another, and from one lot of DL-a-tocopherol treated Florisil tubes to
another. Therefore, it is necessary for each laboratory to determine
the desorption efficiency for each lot of DL-a-tocopherol treated Florisil tubes.
2.6. Recommended air volume and sampling rate
2.6.1. The recommended sampling rate is 0.2 L/min.
2.6.2. The recommended air volume is 25 L.
2.7. Interferences (sampling)
2.7.1 Since it is possible that the precursors of DENA -
diethylamine, triethylamine, other tertiary amines with the diethyl
moiety, and various nitrosating agents (oxides of nitrogen,
nitrites, etc.), are present in the environment, it is conceivable
that DENA may be formed upon the sampling device and not be present
in the sampled air.
Laboratory studies indicate that it is possible to form DENA from
its precursors on untreated Florisil tubes. Further experiments show
that when the Florisil tube is treated with 11 mg of DL-a-tocopherol, the formation of DENA does
not take place.
2.7.2. At the present time, it is unknown if any compound would
severely interfere with the collection of DENA on treated Florisil
tubes. In general, the presence of other compounds will reduce the
breakthrough volume for a particular compound.
2.7.3. Any compound which is suspected of interfering with the
collection or analysis should be listed on the sampling data sheet.
2.7.4. Light will decompose DENA (Ref. 5.5.). The air sampler
must be protected from light during and after sampling.
2.8. Safety precautions (sampling)
2.8.1. Observe due care when working with the sharp ends of the
air sampler.
2.8.2. Attach the sampling equipment to the worker in such a
manner that it will not interfere with work performance.
2.8.3. Follow all safety practices that apply to the work area
being sampled.
3. Analytical Procedure
3.1. Apparatus
3.1.1. A gas chromatograph interfaced to a Thermal Energy
Analyzer.
3.1.2. A GC column capable of resolving DENA from the desorption
solvent and potential interferences. The column used in this work
was a 10-ft × 1/8-in. stainless steel column containing 10% Carbowax
20M with TPA on 80/100 mesh Chromosorb W AW.
3.1.3. An electronic integrator or other suitable method to
measure peak area.
3.1.4. An analytical balance capable of accurately weighing to 5
decimal places.
3.1.5. Vials. 2-mL vials with Teflon-lined caps.
3.1.6. Microliter syringes. 5-µL syringes for sample injections
and other convenient sizes.
3.1.7. Pipets. Convenient sizes for diluting standards. A 1-mL
dispenser pipet for desorbing solvent.
3.1.8. Volumetric flasks. Convenient sizes for diluting
standards.
3.1.9. Dewar flasks. Convenient sizes for liquid nitrogen.
3.2. Reagents
3.2.1. DENA, authentic primary standard, 98% minimum.
3.2.2. Methyl alcohol, chromatographic grade.
3.2.3. Methylene chloride, chromatographic grade.
3.2.4. Isopropyl alcohol, chromatographic grade.
3.2.5. Ethyl alcohol, U.S.P., 95%.
3.2.6. Gases, purified GC grade helium and medical grade oxygen.
3.2.7. Nitrogen, liquid.
3.3. Sample preparation
3.3.1. The status of the OSHA seal on each sample is noted and
recorded as intact, broken, or none.
3.3.2. The field and laboratory identification numbers on each
seal are checked against those on the sample identification sheets.
3.3.3. Avoid exposure of the samples to light. DENA will
photodecompose easily.
3.3.4. The front and rear tubes from each sampler are transferred
to separate 2-mL vials.
3.3.5. The contents of each vial are desorbed with 1.0 mL of
desorbing solution. The desorbing solution is composed of equal
parts by volume methyl alcohol and methylene chloride.
3.3.6. The vials are sealed immediately with Teflon-lined caps
and desorbed for 30 min with intermittent shaking.
3.4. Standard preparation
3.4.1. Stock standards are prepared by diluting a weighed amount
of DENA with isopropyl alcohol. The stock standard is diluted to the
working range with isopropyl alcohol (Section 3.8. Safety
Precautions).
3.4.2. A solution composed of 0.15 µg/mL DENA in isopropyl
alcohol is equivalent to an air concentration of 6.0
µg/m3 for a 25-L air sample desorbed with 1.0 mL of
desorbing solution. This amount is not corrected for the desorption
efficiency.
3.4.3. Standards are stored in dark bottles under refrigeration.
3.5. Analysis
3.5.1. GC conditions
helium (carrier gas) flow rate: |
30 mL/min |
injector temperature: |
200°C |
column temperature: |
160°C |
TEA transfer line temperature: |
205°C |
injection volume: |
5 µL |
elution time: |
3.5 min |
3.5.2. TEA conditions
oxygen pressure: |
9 PSI |
GC pyrolyzer furnace temperature: |
475°C |
chamber vacuum: |
1.8 mm Hg |
coarse zero: |
high |
calibrate: |
0.0 |
attenuator: |
4 |
cold trap temperature: |
-130°C (ethyl alcohol and liquid
nitrogen) |
Complete instructions for the TEA are found in its manual.
3.5.3. Chromatogram (Section 4.3.)
3.5.4. Peak areas are measured by an electronic integrator or
other suitable means.
3.5.5. An external standard procedure is used to prepare a
calibration curve from the analysis of at least three different
standard solutions. The calibration curve is prepared daily. The
integrator is calibrated to report results in micrograms per
milliliter after correction for desorption efficiency.
3.5.6. Bracket the samples with analytical standards.
3.6. Interferences
3.6.1. Any compound that has the same GC retention as DENA and
will elicit a response from the TEA detector is an interference.
Possible interferences are listed on the sample data sheets.
3.6.2. GC parameters may be changed to circumvent most
interferences.
3.6.3. Retention time on a single GC column is not proof of
chemical identity. Samples should be confirmed by GC/Mass
Spectrometry or other suitable means when required.
3.7. Calculations
3.7.1. The integrator value in micrograms per milliliter
(corrected for desorption efficiency) is used for reference only.
More reliable results are obtained by use of the calibration curve.
The peak area, for each standard, is compared to its concentration
in micrograms per milliliter (corrected for desorption efficiency)
and the equation for the best straight line through the data points
is determined by linear regression.
3.7.2. The concentration in micrograms per milliliter (corrected
for desorption efficiency) for a particular sample is determined by
comparing its area to the calibration curve.
3.7.3. Analytical results from the two tubes that compose an air
sampler are added together.
3.7.4. The air concentration for a sample result is calculated by
the following equation:
DENA, µg/m3 = (A)(B)(1000)/C
where |
A |
= |
µg/mL from 3.7.3. |
|
B |
= |
desorption volume |
|
C |
= |
air volume in
liters |
3.7.5. To convert micrograms per cubic meter to parts per billion
(ppb), the following relationship is used:
DENA, ppb = (µg/m3)(24.46)/102.1
where |
µg/m3 |
= |
result from 3.7.4. |
|
24.46 |
= |
molar volume at 25°C and 760 mm Hg |
|
102.1 |
= |
molecular weight of
DENA |
3.8. Safety precautions (analytical)
3.8.1. DENA is an extremely potent carcinogen and utmost care
must be exercised when working with this compound.
3.8.2. Avoid skin contact with liquid nitrogen and the solvents.
3.8.3. Confine the use of solvents to a fume hood.
3.8.4. Wear safety glasses in all laboratory areas.
3.8.5. Check to be sure that the TEA exhaust is connected to a
fume hood.
4. Backup Data Section
4.1. Detection limit
The following data were generated by replicate 5-µL injections of a
standard solution whose concentration was 0.01 µg/mL. The detection
limit was determined to be 50 pg which is equivalent to 0.4
µg/m3 based on the recommended air volume. Peak heights
were used because integrated data were unreliable at this level.
Table 4.1. Detection Limit Data
|
injection |
peak height, mm |
statistics |
|
1 2 3 4 5 6 |
87 82 67 62 88 80 |
= 77.67 SD = 10.74 CV
= 0.14
|
|
4.2. Precision
These data represent multiple injections of standard solutions. The
injection volume was 5 µL and the concentrations of the standards were
0.08 µg/mL, 0.15 µg/mL and 0.30 µg/mL. Peaks were integrated by a
Hewlett-Packard 5840A Gas Chromatograph.
Table 4.2. Analytical Precision
|
× target conc. pg |
0.5× 400 |
1.0× 750 |
2.0× 1500 |
|
SD CV
= 0.034 |
4635 4318 4184 4313 4281 4061
4298.67 191.65 0.0446 |
8714 8792 8954 8872 8982 9128
8907 147.22 0.0165 |
18090 18980 17220 17850 18250 17490
17980 619.61 0.0345 |
|
4.3. Sensitivity
A typical calibration curve is shown in Figure 4.3.1. The slope of
the curve indicates the sensitivity of the method.
A representative chromatogram is shown in Figure 4.3.2.
4.4. Desorption
Samples representing 3.2, 6.0 and 12.0 µg/m3 based on
25-L air volumes were prepared by injecting liquid standards on
treated Florisil tubes.
Table 4.4. Desorption Efficiency
|
µg/m3 |
3.2 |
6.0 |
12.0 |
|
desorption efficiency, %
|
95.0 82.5 87.5 112.5 95.0 95.0
94.6
|
100.7 90.7 89.3 92.7 92.0 94.0
93.2
|
87.0 91.3 105.7 99.7 102.0 103.0
98.1
|
overall average =
95.3% |
|
4.5. Storage
Samples were generated by the liquid injection of DENA on Polar
Partition tubes containing about 150 mg of adsorbent. The tubes were
allowed to equilibrate overnight and then were placed in front of the
treated Florisil tubes. Twenty-five liters of air, at 80% relative
humidity and 22°C, were drawn through the sampling train. The DENA on
the Polar Partition tube was desorbed by the humid air and was
deposited on the air sampler. Studies conducted at ambient relative
humidity and temperature indicated that desorption was essentially
complete after 1 L of air had passed through the Polar Partition tube
(Figure 4.5.1.). In these studies the spiked Polar Partition tube was
connected directly to the TEA.
The data in Table 4.5. represent the effects of storage at ambient
(21-26°C) and reduced (-20°C) temperatures on DENA
collected using treated Florisil tubes. The recoveries for the six
samples run on the designated days are not corrected for desorption
efficiency. A graphical representation of the data may be found in
Figures 4.5.2. and 4.5.3.
Table 4.5. Storage Tests
|
storage time |
% recovery |
(days) |
(refrigerated) |
|
(ambient) |
|
0 3 6 10 13 17 |
90.0 87.3 96.0 96.7 100.7 104.0 |
90.7 87.3 103.3 96.7 93.3 89.3 |
94.0 86.7 91.3 95.3 96.0 104.7 |
|
93.3 82.7 82.7 84.7 88.7 84.0 |
91.3 94.0 84.0 81.3 98.0 89.3 |
91.3 82.0 84.7 92.7 94.0 94.7 |
|
4.6. Preparation of sampling tubes
4.6.1. Reagents
Methylene chloride, chromatographic grade.
DL-a-tocopherol (Vitamin
E), 90% minimum - available from ICN Pharmaceutical Inc., K & K
Labs Division, Life Sciences Group (1979 Catalog No. 18787). Prepare
two solutions containing 50 and 100 mg/mL in methylene chloride.
Store under refrigeration in dark bottles.
4.6.2. Technique
Each Florisil tube is treated with 11 mg DL-a-tocopherol. This is accomplished by
first breaking open both ends of each tube and adding 200 µL of the
50 mg/mL DL-a-tocopherol
solution. Both sections of the tube should be uniformly wetted by
the solution. The methylene chloride is allowed to evaporate in a
fume hood overnight. The tube is further treated by placing 10 µL of
the 100-mg/mL DL-a-tocopherol
solution on the head of the front section of Florisil. The methylene
chloride is again evaporated and then both ends of the tube are
sealed with the plastic end caps supplied with the tubes. The tubes
should be stored in a freezer with suitable precautions taken to
prevent their contamination. The tubes are stable, when stored in a
freezer, for at least 2 months.
Figure 4.3.1. Calibration curve for diethylnitrosamine.
Figure 4.3.2. A typicdal chromatogram for
diethylnitrosamine.
Figure 4.5.1. The rate at which diethylnitrosamine is
desorbed from Polar Partition adsorbent by air at ambient temperature
and relative humidity.
Figure 4.5.2. Ambient temperature storage test for
diethylnitrosamine.
Figure 4.5.3. Refigerated temperature test for
diethylnitrosamine.
5. References
5.1. W. Hendricks, Dimethylnitrosamine (Method 06, Organic Methods
Evaluation Branch, OSHA Analytical Laboratory, Salt Lake City, Utah),
Unpublished.
5.2. Registry of Toxic Effects of Chemical Substances, 1976
Edition. (H.E. Christensen, E.J. Fairchild, Eds., B.S. Carroll, R.J.
Lewis, Project Coordinators) U.S. Department of Health, Education, and
Welfare, Public Health Service, Center for Disease Control, National
Institute for Occupational Safety and Health, U.S. Government Printing
Office, Washington, D.C. (1976).
5.3. P.N. Magee and J.M. Barnes, Carcinogenic Nitroso Compounds,
Advances in Cancer Research 10, 163-247 (1967).
5.4. D. Druckey, R. Preussmann, S. Ivankovic and D. Schmahl,
Organotrope Carcinogene Wirkany bei 65 Verschiedenen
N-Nitroso-ver-bindungen and BD-Ratten, Z. Krebsforsch, 69(103),
103-201 (1967).
5.5. IARC Monographs on the Evaluation of Carcinogenic Risk of
Chemicals to Man, Vol. 1, International Agency for Research on Cancer,
Lyon, France, World Health Organization, 107-124 (1971).
5.6. J. March, Advanced Organic Chemistry: Reactions, Mechanisms
and Structure, McGraw-Hill Book Company, New York, 486
(1968).
5.7. J.N. Pitts, P. Grosjean, K.V. Cauwenberghe, J.P. Schmid, and
D.R. Fitz, Photooxidation of Aliphatic Amines Under Simulated
Atmospheric Conditions: Formation of Nitrosamines, Nitramines, Amides
and Photochemical Oxidant, Environmental Science and
Technology, 12(8) 946-953 (1978).
5.8. B. Spiegelhander, G. Eisenbrand and R. Preussman,
Contamination of Amines With N-Nitrosoamines, Angew.
Chem. Int. Ed. Engl., 17(5) 367-368 (1978).
5.9. P. Issenberg and H. Sornson, A Monitoring Method for Volatile
Nitrosamine Levels in Laboratory Atmospheres, Environmental N-Nitroso
Compounds Analysis and Formation, E.A. Walker and P. Bogovski, Eds.,
Lyon, France, IARC Scientific Publication No. 14 (1976).
5.10. T. Kakizoe, T.T. Wang, VW.S. Eng., R. Farrer, P. Dion and R.
Bruce, Volatile N-Nitrosoamines in the Urine of Normal
Donors and of Bladder Cancer Patients, Cancer Research 39,
829-832 (1979).
5.11. K. D. Brunnemann, L. Ye and D. Hoffmann, Assessment of
Carcinogenic Volatile N-Nitrosamines in Tobacco and in
Mainstream and Sidestream Smoke from Cigarettes, Cancer
Research 37, 3218-3333 (1977).
|