ETHYLENE DICHLORIDE
Method no.: |
03 |
|
Matrix: |
Air |
|
OSHA PEL: |
50 ppm |
|
Target concentration: |
5 ppm |
|
Procedure: |
Collection on charcoal adsorbent, desorption with
o-xylene, analysis by gas chromatography using electron
capture detector. |
|
Detection limit based on recommended air
volume: |
0.05 ppm (for analytical procedure only) |
|
Recommended air volume and sampling rate: |
10 L at 0.2 L/min |
|
Standard error of estimate: (Figure 4.3.1.) |
6.7% |
|
Status of method: |
Evaluated method. This method has been subjected to
the established evaluation procedures of the Organic Methods
Evaluation Branch. |
|
Date: April 1979 |
Chemist: Duane E. Lee |
Organic Methods Evaluation Branch OSHA Analytical
Laboratory Salt Lake City, Utah
1. General Discussion
1.1. Background
1.1.1. History
In 1946, the American Conference of Governmental Industrial
Hygienists adopted a Maximum Allowable Concentration (MAC) of 100
ppm for ethylene dichloride. In 1947, they changed it to 75 ppm; and
in 1948, they changed the nomenclature to Threshold Limit Value
(TLV) for allowable concentrations. The TLV remained at 75 ppm until
1952 when it was changed to 100 ppm. Subsequently the TLV was
changed to 50 ppm in 1962. (Ref. 5.1.)
OSHA has adopted a 50 ppm TWA and a ceiling concentration of 100
ppm with a maximum peak of 200 ppm for 5 min in any 3-h period.
(Ref. 5.2.) In April 1978, NIOSH issued a "Current Intelligence
Bulletin" recommending an exposure limit of 5 ppm based on adverse
effects on the nervous system and being a potential carcinogen.
(Ref. 5.3.) For this reason, the analytical and sampling procedures
must be evaluated to assure that an adequate method exists for lower
levels. In 1977 and 1978, the OSHA Analytical Laboratory analyzed
approximately 50 samples each year for ethylene dichloride.
1.1.2. Toxic effects (This section is for information only and
should not be used as basis for OSHA policy.)
Animal studies have evaluated the toxic effects of ethylene
dichloride. The adverse effects of acute exposure include extreme
lowering of blood pressure, cardiac impairment, pulmonary edema,
fatty degeneration of the liver and kidney, and degeneration of the
adrenal cortex. (Ref. 5.3.) Studies also indicate it causes cancer
in rats and mice by producing tumors such as adenomas,
adenocarcinomas, and fibromas. (Ref. 5.4.)
The effects of ethylene dichloride on humans are similar for
ingestion, inhalation, and skin absorption. Acute exposures result
in nausea, vomiting, dizziness, internal bleeding, bluish-purple
discoloration of the mucous membranes and skin, rapid but weak
pulse, and unconsciousness. (Ref. 5.3.) There is no definitive
evidence that ethylene dichloride causes cancer in humans. (Ref.
5.4.)
1.1.3. Worker exposure
NIOSH estimates that as many as 2 million workers may be exposed
to ethylene dichloride. Of these workers, an estimated 34,000 have
been exposed to ethylene dichloride for 4 h or more per day. (Ref.
5.3.)
1.1.4. Use and operations where exposure occurs
Ethylene dichloride is the earliest known chlorinated
hydrocarbon, having been produced in 1795. (Ref. 5.5.) It is one of
the highest volume chemicals used in the United States with
approximately 10 billion pounds produced annually. Most of the
ethylene dichloride produced is used as an intermediate in the
production of vinyl chloride, but it is also used in the production
of other chlorinated hydrocarbons. Besides production uses, it is
used as a solvent and is used in leaded fuels because it is a lead
scavenger. (Ref. 5.3.)
Industries where ethylene dichloride may be used include: medical
and health services, automotive dealers and service stations,
machinery, printing and publishing, eating and drinking places,
chemical and allied products, and miscellaneous business services.
(Ref. 5.4.)
1.1.5. Physical properties (Ref. 5.5. and 5.6.)
molecular weight: |
98.97 |
boiling point: |
83.5°C |
melting point: |
-35.3°C |
specific gravity: |
1.2529 at 20/4°C |
self-ignition temp: |
449°C |
flash point: |
18.3°C open cup 13.0°C closed cup |
explosive limits |
|
% by volume: |
6.2 - 16.9 |
color: |
colorless liquid |
odor: |
resembles chloroform |
molecular formula: |
CH2ClCH2Cl |
|
other names: |
ethylene chloride; brocide; EDC;
1,2-dichloroethane; dutch liquid; s-dichloroethane;
1,2-bichloroethane |
1.2. Detection limit, precision, sensitivity, and working range
1.2.1. The detection limit for the analytical procedure is 2 ng
with a coefficient of variation equal to 0.05 at this level.
(Section 4.1.) The detection limit was determined using 1-µL
injections.
1.2.2. The pooled coefficient of variation of the analytical
procedure over the range of 100 to 402 µg per sample is 0.017.
(Section 4.2.) This represents an air concentration of 2.5 to 10 ppm
based on the recommended sampling and analytical procedures.
1.2.3. The sensitivity of the analytical procedure over a
concentration range representing 0.5 to 2 times the target
concentration based on the recommended air volume is 7.69946 ×
107 area units per mg/mL. The sensitivity
is determined by the slope of the calibration curve. (Section 4.2.)
The sensitivity will vary somewhat with the particular
instrumentation used in the analysis.
1.2.4. The lower limit of the estimated working range, assuming
adequate desorption efficiency, is 0.05 ppm. The upper limit of the
working range is dependent on the capacity of the collection medium.
1.3. Accuracy
1.3.1. The overall procedure must provide results that are
within 25% or better at the 95% confidence interval.
1.3.2. The recovery of analyte from the collection medium after
storage must be 75% or greater.
1.3.3. The overall procedure has met the above validation
criteria for a 15-day period. (Section 4.3.)
1.4. Advantages
1.4.1. The sampling procedure is convenient.
1.4.2. The analytical procedure is quick, sensitive, and
reproducible.
1.4.3. Reanalysis of samples is possible.
1.4.4. Automation of the procedure is possible.
1.4.5. The electron capture detector is sensitive to ethylene
chloride and not very sensitive to hydrocarbons, therefore making
the method somewhat selective.
1.5. Disadvantages
1.5.1. It may be difficult to analyze for hydrocarbons because
of the solvent used and the electron capture detector.
1.5.2. The analysis time is lengthened because a temperature
program is required to get the o-xylene off the GC column.
2. Sampling Procedure
2.1. Apparatus
2.1.1. An approved and calibrated personal sampling pump whose
flow can be determined within ±5% at the recommended flow.
2.1.2. Charcoal tubes: Glass tube, with both ends heat sealed,
7.0 cm × 6-mm o.d. × 4-mm i.d., containing 100-mg front
and 50-mg backup sections of 20/40 mesh charcoal SKC tubes or
equivalent.
2.2. Reagents
None required.
2.3. Sampling technique
2.3.1. Immediately before sampling, break the ends of the
charcoal tube. All tubes must be from the same lot.
2.3.2. Connect the charcoal tube to the sampling pump with
flexible tubing. The short section of the charcoal tube is used as a
backup and should be positioned nearer the sampling pump.
2.3.3. The tube should be placed in a vertical position during
sampling to minimize channeling.
2.3.4. Air being sampled should not pass through any hose or
tubing before entering the charcoal tube.
2.3.5. Seal the charcoal tube with plastic caps immediately after
sampling. Also, seal each sample lengthwise with an OSHA sealing
tape.
2.3.6. With each batch of samples, submit at least one blank tube
from the same lot used for samples. This tube should be subjected to
exactly the same handling as the samples (break, seal, transport)
except that no air is drawn through it.
2.3.7. Transport the samples (and corresponding paperwork) to the
lab for analysis.
2.3.8. If bulk samples are submitted for analysis, they should be
transported in glass containers with Teflon-lined caps. These
samples must not be put in the same container used for the charcoal
tubes.
2.4. Breakthrough
Three separate samples were drawn at 0.195 L/min from an atmosphere
containing 10 ppm ethylene dichloride and varying relative humidity.
The breakthrough (5% breakthrough) volume was: 15.8 L for 75% relative
humidity, 14.3 L for 82% relative humidity, and 11.9 L for 85%
relative humidity.
2.5. Desorption efficiency
2.5.1. The desorption efficiency, from liquid injections on
charcoal tubes SKC Lot 107, averaged 87.9% for 0.1 to 0.4 mg per
tube, which is 2.5 to 10 ppm for a 10-L air volume. (Section 4.4.)
2.5.2. The desorption efficiency may vary from laboratory to
laboratory and for each lot of charcoal.
2.6. Recommended air volume and sampling rate
2.6.1. The recommended air volume is 10 L.
2.6.2. The recommended sampling rate is 0.2 L/min.
2.7. Interferences
2.7.1. At the present time, it is unknown if any compound would
severely interfere with the collection of EDC on charcoal. In
general, the presence of other solvents will decrease the
breakthrough volume for a particular solvent.
2.7.2. Any compound which is suspected of interfering in the
collection or analysis should be listed on the sampling data sheet.
2.8. Safety precautions
2.8.1. Safety glasses should be worn when breaking the ends of
the tubes.
2.8.2. The broken ends of the tubes should be protected to avoid
injury to the person being sampled.
2.8.3. When working in environments containing flammable vapors,
do not provide any spark source from equipment used or pumps.
2.8.4. Observe all safety practices for working in hazardous
areas.
3. Analytical Procedure
3.1. Apparatus
3.1.1. A gas chromatograph equipped with an electron capture
detector.
3.1.2. GC column capable of separating 1,1-dichloroethane,
1,2-dichloroethane, and 1,1,2-trichloroethane. The column used for
this study was a 10-f t × 1/8-in. stainless steel 0.1% SP-1000
coated on 80/100 mesh Carbopack C.
3.1.3. An electronic integrator or other suitable method of
measuring detector response.
3.1.4. Two-milliliter vials with Teflon-lined caps.
3.1.5. Microliter syringes, 10-µL for preparing standards, 1-µL
for sample injections.
3.1.6. Pipets for diluting standards. A 1-mL pipet for dispensing
solvent for desorption, or a 1-mL dispenser pipet.
3.1.7. Volumetric flasks, for preparing standards.
3.2. Reagents
3.2.1. o-Xylene, chromatographic grade.
3.2.2. Ethylene dichloride, reagent grade.
3.2.3. Argon/methane, 95/5; purified GC grade carrier gas.
3.3. Standard preparation
3.3.1. Standards are prepared by diluting pure ethylene
dichloride with o-xylene.
3.3.2. Eight microliters of ethylene dichloride per 50 mL of
o-xylene equals 4.95 ppm for a 10-L air sample desorbed with 1 mL of
o-xylene.
3.4. Sample preparation
3.4.1. The front and back sections of each sample are
transferred to separate 2-mL vials.
3.4.2. Each section is desorbed with 1.0 mL of o-xylene.
3.4.3. The vials are sealed immediately and allowed to desorb for
30 min with intermittent shaking.
3.5. Analysis
3.5.1. GC conditions
argon/methane flow rate: |
25 mL/min |
injector temperature: |
200°C |
detector temperature: |
275°C |
initial column temperature: |
80°C for 15 min |
temperature program rate: |
10°C/min |
final column temperature: |
210°C for 5 min |
EC detector: |
Ni-63 |
3.5.2. Chromatogram (Figure 4.5.)
3.5.3. Peak areas are measured by an electronic integrator or
other suitable means.
3.5.4. Because an external standard method is used, duplicate
injections of all standards and samples are made.
3.6. Interferences
3.6.1. Any substance having the same general retention time as
ethylene dichloride is an interference.
3.6.2. GC parameters may be changed to circumvent most
interferences.
3.6.3. Retention time on a single column is not considered proof
of chemical identity. Samples should be confirmed by GC/MS or other
suitable means.
3.7. Calculations
Usually the integrator is programmed to report results in ppm
(corrected for desorption efficiency) for a 10-L air volume. The
following calculation is used:
ppm = A/(0.1)(B) |
where |
A |
= |
ppm on report |
|
B |
= |
air volume (L) |
3.8. Safety precautions
3.8.1. All work using solvents should be done in a hood.
3.8.2. Avoid any skin contact with all of the solvents.
3.8.3. Safety glasses should be worn throughout the procedure.
4. Backup Data
4.1. Detection limit
Standard used was 0.0016 µL ethylene dichloride per milliliter of
o-xylene (2.0 µg/mL).
Table 4.1. Detection Limit Data
|
injection |
area |
injection |
area |
|
1 |
200600 |
7 |
203900 |
2 |
189400 |
8 |
200300 |
3 |
191300 |
9 |
193900 |
4 |
198000 |
10 |
194600 |
5 |
189800 |
11 |
200600 |
6 |
194800 |
12 |
230200 |
|
= 198950 |
SD
=
10889.236 CV
= 0.0547 |
|
Detection limit is (2.0 ng/µL)(1 µL) = 2.0 ng per injection, or
0.05 ppm for a 10-L air volume desorbed with 1.0 mL of o-xylene.
4.2. Instrument response to the analyte and analytical precision
Table 4.2. Precision of the Analytical Procedure
|
× target conc. |
0.5× |
1× |
2× |
µg/mL |
100.6 |
201.1 |
402.2 |
|
area |
9932000 |
17330000 |
33150000 |
counts |
10000000 |
16840000 |
33130000 |
|
9468000 |
17100000 |
32440000 |
|
9450000 |
17090000 |
32690000 |
|
9760000 |
17470000 |
33250000 |
|
9668000 |
17590000 |
32690000 |
|
|
9713000 |
17236667 |
32891667 |
SD |
229677.16 |
277680.87 |
327805.84 |
CV |
0.0236 |
0.0161 |
0.00997 |
|
= 0.017 |
|
|
The response data from above are used to determine the calibration
curve in Figure 4.2.
4.3. Storage tests
Thirty-six samples were collected on charcoal tubes (SKC Lot 107)
from a dynamic test atmosphere containing 5 ppm ethylene dichloride
and 80% relative humidity. The recommended sampling conditions were
used. Six samples were analyzed immediately. The remainder were
divided into two equal groups, one stored at room temperature (20 to
23°C) and the other under refrigeration (-5 to 0°C). Three samples
from each group were analyzed at 2 to 4 day intervals over 15 days.
The results shown below are also shown in Figures 4.3.1. and 4.3.2.
Table 4.3. Storage Tests
|
storage time |
% recovery |
|
% recovery |
|
(refrigerated) |
|
(ambient) |
|
0 |
88.7 |
86.9 |
88.6 |
|
85.1 |
88.5 |
85.5 |
3 |
83.8 |
84.6 |
83.6 |
|
82.6 |
80.0 |
84.0 |
7 |
99.4 |
87.9 |
95.2 |
|
95.8 |
95.8 |
93.3 |
9 |
86.5 |
88.0 |
89.1 |
|
84.9 |
88.0 |
87.9 |
12 |
91.8 |
89.6 |
90.8 |
|
86.7 |
82.9 |
86.5 |
15 |
89.4 |
92.7 |
95.1 |
|
92.2 |
93.2 |
89.3 |
|
4.4. Desorption efficiency
Table 4.4. Desorption Efficiency of Ethylene Dichloride
from Charcoal with o-Xylene
|
× target conc. |
0.5× |
1× |
2× |
µg/sample |
100.6 |
201.1 |
402.2 |
|
desorption |
88.9 |
87.2 |
90.7 |
efficiency, |
87.4 |
84.5 |
91.7 |
% |
77.1 |
87.4 |
90.5 |
|
87.6 |
87.8 |
89.4 |
|
87.7 |
87.6 |
91.2 |
|
86.8 |
88.2 |
90.5 |
|
|
85.9 |
87.1 |
90.7 |
|
= 87.9 |
|
|
4.5. Chromatogram
A typical chromatogram of ethylene dichloride under the GC
conditions used in the analysis is shown in Figure 4.5.
Figure 4.2. Calibration curve of instrument response to
EDC.
Figure 4.3.1. Ambient storage test of
EDC.
Figure 4.3.2. Refrigerated storage test of
EDC.
Figure 4.5. Chromatogram of a standard of EDC.
5. References
5.1. "Criteria for a Recommended standard. . .Occupational
Exposure to Ethylene Dichloride" NIOSH April, 1976.
5.2. General Industrial Safety and Health standards, OSHA 2206 (29
CFR 1910), Revised January, 1976.
5.3. "Current intelligence Bulletin 25" NIOSH, April 19,1978.
5.4. "Current Intelligence Bulletin 27" NIOSH, August 21, 1978.
5.5. Encyclopedia of Chemical Technology, Second Edition, Vol. 5,
1967.
5.6. Industrial Hygiene and Toxicology, 2nd Revised Edition, Vol.
11, Frank A. Patty, editor, page 1280.
|