TUNGSTEN AND COBALT IN WORKPLACE ATMOSPHERES (ICP
ANALYSIS)
Method Number: |
ID-213 |
|
Matrix: |
Air |
|
OSHA Permissible Exposure Limits |
|
* Final Rule Limits |
|
Insoluble
Tungsten (Insol. W): |
5. mg/m3 Time Weighted
Average (TWA) 10. mg/m3
Short-Term Exposure Limit (STEL) |
Soluble Tungsten
(Sol. W): |
1. mg/m3 TWA 3.
mg/m3 STEL |
Cobalt (Co): |
0.05 mg/m3 TWA |
|
Transitional Limits |
|
Insol. and Sol.
W: |
Not Applicable (NA) |
Co: |
0.1 mg/m3 TWA |
|
Collection Device: |
An air sample is collected on a 37-mm
mixed-cellulose ester (MCE) filter using a calibrated sampling pump
and a 37-mm sampling cassette. Wipe and bulk samples
can also be taken. |
|
Recommended Sampling Rate TWA
and STEL: |
2. liters per minute (L/min) |
|
Recommended Air Volumes |
|
TWA (W and Co): |
480. L |
STEL (W): |
30. L |
|
Analytical Procedure: |
The MCE filter is subjected to a sequence of
digestion steps using ammonium hydroxide (aqueous ammonia), water,
hydrochloric acid, nitric acid, and phosphoric acid. The resulting
solution is analyzed by aspiration into the argon stream of an
inductively coupled plasma atomic emission spectrometer
(ICP-AES). |
|
Detection Limit (mg/m3) |
|
Qualitative: TWA
(W and Co) STEL
(W) Quantitative: TWA
(W and Co) STEL (W) |
W 0.0043 0.069
0.022 0.34 |
Co 0.0003
0.0017
|
| |
|
* Precision and Accuracy
(TWA): Validation Ranges
(mg/m3): CV
(pooled): Bias: Overall
Error (OE): |
Insol.
W 2.5 - 10.0 0.045 + 0.012 ±
10.2% |
Sol.
W 0.5 - 2.0 0.0097 + 0.073 ±
9.2% |
Co 0.025
- 0.10 0.027 + 0.046 ±
10.0% | |
|
Precision and Accuracy
(STEL): Validation Level
(mg/m3): CV: Bias: Overall
Error: |
Insol.
W 10.0 0.020 + 0.047 ±
8.7% |
Sol.
W 3.0 0.0070 + 0.062 ±
7.6% |
| |
|
Method Classification: |
Validated Method |
|
Chemist: |
Mike C. Rose |
|
Date: |
February, 1994 |
Branch of Inorganic Methods Development OSHA Salt Lake
Technical Center Salt Lake City, Utah
Commercial manufacturers and products mentioned in this method
are for descriptive use only and do not constitute endorsements
by
USDOL-OSHA.
Similar products from other sources can be substituted.
* Note: The U.S. Court of Appeals, Eleventh Circuit, recently
decided that the Final Rules of the Air Contaminants Standard (29 CFR
1910.1000) be vacated. The Federal Occupational Safety and Health
Administration (OSHA) enforcement would follow the "Transitional Limits"
of Tables Z-1-A, Z-2, and Z-3 in effect before
1989. At that time, cobalt (Co) was regulated at an OSHA Permissible
Exposure Limit (PEL) of 0.1 mg/m3. In the case
of tungsten (W), both soluble and insoluble W were regulated as nuisance
dusts at a PEL of 15 mg/m3 or potentially by
reference to Section 5(a)(1) of the Occupational Safety and Health Act of
1970 and subsequent NIOSH 1977 recommendations for an insoluble W
Recommended Exposure Limit (REL) of 5 mg/m3 and
a soluble W REL of 1 mg/m3 (5.1.). References to
PELs in this method refer to the Final Rule PELs. This method is
applicable for determining compliance to either Final Rule or Transitional
PELs.
1. Introduction
This method describes the sample collection and analysis of airborne
tungsten (W) and cobalt (Co). Most industrial exposures to W also involve
exposures to Co (5.1.). Tungsten carbide (WC or
W2C) materials used for abrasives and cutting
tools are frequently made containing Co metal which adheres to the surface
of the hard but brittle tungsten carbide grains and cements them together
into a tough composite. Other common components of cemented tungsten
carbide materials include intermetallic compounds of W and Co (e.g.,
Co6W6C) in addition to
other metals and their carbides. Because Co often occurs at significant
levels in these materials, it is potentially present in airborne dusts
along with W when tungsten carbide composites are used in the workplace.
The Co component of cemented tungsten carbides has often been attributed
as the etiological agent for certain occupational diseases observed in the
tungsten carbide industry (5.1., 5.2.). For these reasons, the sampling
and analysis of Co was included in this method. Air samples are taken in
the breathing zone of workplace personnel. Wipe and bulk samples can also
be taken. Analysis is performed for elemental W and Co by Inductively
Coupled Plasma Atomic Emission Spectrometry (ICP-AES).
During abrasive cutting or grinding, the matrix being cut or ground can
aerosolize in addition to the grinding agent itself. The method is also
suitable for screening other elements such as Al, As, Be, Ca, Cd, Cr, Cu,
Fe, Mg, Mn, Mo, Ni, Pb, Sb, Se, Si, Sn, Te, Ti, V, Zn, and Zr.
1.1. History
Prior to this method, several analytical approaches of W
particulates were used.
1.1.1. Atomic Absorption Spectroscopy (AAS)
Tungsten samples were digested by NIOSH Method P&CAM 271
(5.3.) and analyzed by AAS. The high AAS detection limit (DL) for W
limited its utility and the acid extraction of interferences such as
Fe and Co introduced sources of sample loss. The use of hydrofluoric
acid (HF) required the use of expensive Teflon labware and presented
a potential hazard to the analyst. Because of these limitations,
samples submitted for the analysis of W were often screened using
qualitative X-ray Fluorescence (XRF) to determine if
analysis was necessary or whether other analytes might be of greater
interest to the industrial hygienist.
1.1.2. Direct Current Plasma Atomic Emission Spectroscopy
(DCP-AES)
When DCP-AES became available, a modification of the
P&CAM 271 AAS procedure (5.3.) was tried by OSHA using two
different digestion techniques. Analysis was by DCP-AES
at a wavelength of 207.91 nm (5.4.). In this modified procedure,
mixtures containing WC, Co, and Fe powders were prepared using
microwave digestion that gave recoveries of 95% for W (ranging from
92% to 99%), 77% for Fe, and 84% for Co. Samples digested in a fume
hood in open Teflon beakers gave recoveries of 82% for W (ranging
from 68 to 100%), 89% for Fe, and 87% for Co. The Co recoveries were
disappointing, especially when considering its PEL (0.05
mg/m3) and toxicity. Cobalt is considered
more toxic than W (5.2.)
1.1.3. X-ray Fluorescence (XRF)
Although an initial recovery study looked promising, attempts to
validate an XRF technique using samples of various sieved
W-containing materials deposited on silver membranes
met with mixed success; thin-layer deposition
presentations occasionally tended to be too uneven for adequate
analytical precision to be realized. Because particle size can
affect XRF recoveries, materials used in the validation were sieved
to < 41 µm. Unfortunately, the necessity of sieving
tungsten carbide composites complicated the interpretation of the
results. Unlike stoichiometric compounds, the tungsten carbide
composites consist of mixtures of particles of differing
composition; therefore, it was uncertain whether the composition of
the sieved material was still comparable to the certified
composition (5.5.).
1.1.4. Inductively Coupled Plasma Atomic Emission Spectroscopy
The next methodology investigated was ICP-AES. An
in-house ICP instrument at the OSHA Salt Lake Technical
Center (SLTC) was configured with W and Co channels. The W and Co
channels offer excellent detection limits. Additional channels
included those necessary for potential interferences and channels to
screen for other elements which may also be present in the
workplace.
Attempts to validate an ICP-AES method for W and Co
using a potassium pyrosulfate fusion sample preparation were
unsuccessful; although dissolution of diverse W matrices appeared
complete, the salt content resulting from the fusion caused high
reflected power levels in the plasma chamber resulting in instrument
shutdown. While such solutions might be analyzed for Co by AAS, W
could not be effectively analyzed by AAS due to its high detection
limit.
In pursuing digestion techniques, a procedure for the
determination of Fe in bulk samples of tungsten carbide matrices was
found (5.6.) which offered the potential of digesting tungsten
carbide under relatively mild digestion conditions to produce a
final solution consisting of a 6% phosphoric acid
(H3PO4) matrix.
Additional reagents and steps were needed, and the digestion
procedure was adjusted until six diverse W matrices were completely
digested. A final matrix of 2%
H3PO4 in water
was selected as being adequate for samples.
1.2. Principle
Samples are collected in the workplace using classical monitoring
equipment and techniques. Sampling is further detailed in Section 2.
The wide variety of W matrices found in industry requires use of a
sequence of digestion steps in sample preparation. The chemical
principles involved in the digestion are detailed at the appropriate
steps in the procedure in Section 3.5.
Digested samples are analyzed by aspiration into an inductively
coupled argon plasma. This technique uses a high-power
radio frequency (rf) source to power a water-cooled coil
(inductor) around the end of the plasma torch. A
high-voltage source initiates the plasma by causing some
of the argon gas to ionize. The presence of ions renders it
conductive. The high electrical power induced in the conductive plasma
heats it to very high temperatures (5,000 to 8,000 K), and is capable
of producing atomic and ionic emission lines (spectra) of elements
aspirated into the argon stream.
The instrument currently in use at the OSHA-SLTC uses
a diffraction grating and photomultiplier tube (PMT) system to isolate
and quantitate the plasma-produced emissions. Light from
the atomic emissions passes into the optical system of the
simultaneous atomic emission spectrometer and is dispersed into lines
by a diffraction grating. This technique offers the opportunity to use
analytical lines from higher order diffraction spectra to improve
resolution and decrease the potential for interference. This
instrument is set up to use primarily second and third order
diffraction lines. Unwanted overlapping lines from other diffracted
orders are reduced using optical filters. The light for various lines
is detected simultaneously by an array of
especially-aligned PMTs. Each PMT represents an
analytical channel. Resolution is limited, so spectral overlap of
lines of other elements is potentially present in the sample matrix.
Two steps are taken to solve this problem:
(1) Analytical channels are selected which minimize
interferences so that normally only large amounts of interfering
elements must be present before significant interference is
observed.
(2) Additional analytical channels that monitor for the presence
of interfering elements are included in the analysis so that
interelement corrections (IECs) can be readily made.
See Table 16 in Section 4.3. for a listing of the channels,
wavelength, orders, and interferences used in this method.
1.3. Advantages and Disadvantages
1.3.1. This method has adequate sensitivity, accuracy, and
precision for determining compliance with the OSHA
Short-Term Exposure Limits (STEL) of W and the OSHA
Time Weighted Average Permissible Exposure Limits (TWA-PELs) of W
and Co.
1.3.2. Digestion conditions are surprisingly mild and the final
solution matrix is dilute 2%
H3PO4 reducing
potential matrix effects.
1.3.3. The use of HF is avoided. This acid was commonly used in
the past for W matrix digestions and required
Teflon® ware. Therefore, this method
should be compatible with common silica glass ICP torch and
nebulizer assemblies, and should be significantly safer for the
analyst.
1.3.4. This method screens for many other elements that could be
present in the workplace.
1.3.5. Pending further validation, Fe and Ni may also be analyzed
by this method. This digestion matrix should be suitable for the
analysis of these elements by either ICP-AES or AAS.
1.3.6. A disadvantage is the need to separately determine both
soluble and insoluble W when both are known to be present in the
workplace. See Table 7 in Section 1.6. for solubility designations
for various common materials used in industrial applications.
1.3.7. Another disadvantage is the long digestion time required.
1.4. Method Performance
An Applied Research Lab (ARL) Model 3560 Atomic Emission
Spectrometer with accompanying software (ARL, Sunland, CA) was used to
determine method performance. Uncorrected results from the computer of
the ARL Model 3560 were captured via a VAX 850 and networked to a
personal computer (PC). Custom in-house software was used
on the PC to perform iterative interelement corrections and produce
reports. The additional software performed several functions including
the following:
a) Speed the workup of data
b) Provide the ability to assess different interelement
correction algorithms without need to reanalyze the samples
c) Avoid a "bug" found in the manufacturer's software (5.7.)
d) Present the results in a form amenable to reporting samples
A synopsis of method performance is presented below. Further
information can be found in Section 4.
1.4.1. This method was validated for spiked MCE filter samples
corresponding to air concentrations of 0.025 to 1.41
mg/m3 Co and 0.50 to 10.3
mg/m3 W. See Sections 4.5. to 4.8. for
details.
1.4.2. The qualitative detection limits (2 × SD) are shown in
Table 1:
Table 1
Co |
W |
Assumptions |
0.0032 µg/mL |
0.041 µg/mL |
|
0.16 µg |
2.1 µg |
50 mL soln. vol. |
0.0003
mg/m3 |
0.0043
mg/m3 |
480 L air vol. |
0.0053
mg/m3 |
0.069
mg/m3 |
30 L air
vol. |
See Table 16 in Section 4.3. for qualitative detection limits of
the screening elements.
1.4.3. The quantitative detection limits (10 × SD) are shown in
Table 2 below:
Table 2
Co |
W |
Assumptions |
0.016 µg/mL |
0.21 µg/mL |
|
0.80 µg |
10.3 µg |
50 mL soln. vol. |
0.0017
mg/m3 |
0.022
mg/m3 |
480 L air vol. |
0.027
mg/m3 |
0.34
mg/m3 |
30 L air
vol. |
1.4.4. The sensitivity of the analytical method [(KPulse units
per 1 µg/mL), KPulse is an integration unit used by a
specific manufacturer of ICP instruments] was determined for W and
Co while using the instrumental parameters listed in Table 10
(Section 4.) and Table 16 (Section 4.3.). The calibration is
applicable over the analytical range shown. Above the analytical
range, detector saturation can occur. The sensitivity is the slope
of the calibration curve using the calibration range listed in Table
3:
Table 3
Element |
Sensitivity |
Calibration Range |
Analytical Range |
Co |
3.8 |
0 to 10 µg/mL |
0 to 200 µg/mL |
W |
1.0 |
0 to 100 µg/mL |
0 to 1,000
µg/mL |
1.4.5. The total pooled coefficients of variation
(CVT), biases, and OE for sample masses
expected at about 0.5, 1, and 2 times the OSHA TWA PEL (Sections
4.5.-4.7.) are listed in Table 4 below:
Table 4
TWA |
Element |
CVT |
Bias |
OE (%) |
Co |
0.027 |
+ 0.046 |
± 10.0 |
W (Insol.) |
0.045 |
+ 0.012 |
± 10.2 |
W (Sol.) |
0.0097 |
+ 0.073 |
± 9.2 |
The OE (5.8.) was calculated using the equation:
OEi =
±(|biasi| +
2CVi) × 100%
(95%
confidence level)
Where i is the respective sample pool being examined. Air
volumes of 480 L/min were used in calculating masses.
The results shown in Table 5 below were obtained using amounts
approximating the Final Rule STEL concentrations (Section
4.5.-4.6.):
Table 5
STEL |
Element |
CVT |
Bias |
OE (%) |
W (Insol.) |
0.020 |
+ 0.047 |
± 8.7 |
W (Sol.) |
0.0070 |
+ 0.062 |
± 7.6 |
An air volume of 30 L was used to calculate masses found near
STEL determinations.
When sampling cemented tungsten carbides, Co is potentially
present in amounts that greatly exceed the Co PEL when W is near its
PEL. For this reason, additional Co analyses were performed in the
evaluation of this method to ensure that quantitation of Co can be
performed at the higher levels. The total pooled coefficients of
variation (CVT) and biases for samples
analyzed for Co from 3.2 to 28 times the OSHA Final Rule PEL
(Section 4.8.) are listed in Table 6:
Table 6
High Concentration
Co |
Element |
CVT |
Bias |
OE (%) |
Co |
0.018 |
+ 0.028 |
± 6.5 |
Bias and overall error (OE) values were calculated from those
found analytically versus theoretical (known or certified) values.
1.4.6. Sample stability: No stability problems are
expected in the case of matrices involving insoluble forms of W and
Co. Some very soluble compounds of W, particularly sodium tungstate,
may absorb enough water from the air to dissolve and penetrate into
the backup pad. A procedure to stabilize such samples is presented
in Section 2.2.6.
1.5. Interferences
1.5.1. There are no known chemical interferences.
1.5.2. Spectral interferences depend on the selection of
wavelengths of the element channels (listed in Table 16 in Section
4.3.). In the case of this analytical method, W and Co spectrally
interfered slightly with one another when using the wavelengths
recommended by the instrument manufacturer. In addition, Ti [present
in some tungsten carbide matrices as titanium carbides (5.1.)]
interfered slightly with both W and Co. Additional elements of
interest to industrial hygienists which might be present in tungsten
carbide grinding operations (and elements interfering with the
analysis of W and Co) were included for use in performing
interelement corrections and screening. Also see Section 4.3. for
more details regarding spectral interferences.
1.5.3. This method includes the analysis of Co, W, and the 22
screening elements listed in Section 1. The industrial hygienist
should note the identity of additional elements suspected to be
present in the workplace atmosphere so the analyst can evaluate the
potential for possible interference.
1.6. Sources of Potential Exposure to W (CAS
7440-33-7) and W Materials Containing Co (CAS
7440-48-4) (5.1., 5.2., and 5.9.) are listed in Table 7:
Table 7
Material |
Application |
Tungsten (Insol. W) |
Welding electrodes, alloys, high speed
tool steels, filaments in incandescent lamps, heating elements,
rocket nozzles, and solar energy devices |
Tungsten carbide (Insol. W) |
Abrasive, cemented carbide tools,
dies, and wear-resistant parts |
Tungsten disulfide (Insol. W) |
Solid lubricant |
Tungsten carbonyl (Insol. W) |
Deposition of tungsten
coatings |
Tungsten chlorides and
fluorides (includes both Insol. W and Sol. W) |
Tungstic acid and tungsten oxide
(Insol. W) |
Textiles, ceramics, and plastics |
Sodium tungstate (Sol. W) |
Biological assays, fire proofing and
waterproofing fabrics |
Cobalt |
Component in cemented tungsten carbide
composites and alloys |
1.7. Physical and Chemical Properties of W and Co (5.2.):
Table 8 displays properties of the elemental forms of W and Co.
Compounds and alloys containing these elements are also regulated
under the OSHA PELs.
Table 8
Property |
W |
Co |
Atomic weight |
183.85 |
58.9332 |
Specific gravity |
19.3 |
8.92 |
Melting point (°C) |
3410 |
1493 |
Boiling point (°C) |
5927 |
3200 |
Vapor pressure (torr) |
non-volatile at room
temp. |
Aqueous solubility |
insoluble |
1.8. Toxicology (5.2., 5.10.)
Note: Information listed within this section is a synopsis of
current knowledge of the physiological effects of W and Co and is not
intended to be used as a basis for OSHA policy.
Tungsten
1.8.1. When ingested or given intravenously to guinea pigs, the
chief effects of W (insoluble) and
Na2WO4.2H2O
(soluble) are anorexia, colic, weight loss, incoordination of
movement, trembling, and dyspnea (5.2).
1.8.2. The toxic effects of W are greater for the more soluble
forms. In terms of W content, the compounds
Na2WO4,
WO, and
(NH4)6W7O24
were shown to have decreasing toxicity to rats in the order listed.
Insoluble WC was found biologically inert, although the metal W
produced toxic effects. Dust chamber exposures of animals to W,
WO2, and WC produced only minor changes
(5.2.). The major toxic component in many tungsten carbide cutting
tools is often the Co that provides the tough matrix in which the
relatively brittle WC is embedded (5.1.).
1.8.3. Reports of occupational effects from exposures to soluble
W compounds are not found in the literature (5.1., 5.2.).
Cobalt
1.8.4. Toxic response to Co often involves hypersensitivity.
Pulmonary involvement in humans has been observed to lead to chronic
interstitial pneumonitis. Effects on the lungs are often reversible.
However, acute exposures to levels below 1 to 2
mg/m3 have been fatal. Allergic dermatitis
has also been observed in humans (5.2.).
1.8.5. In the mineral assay and hard metal industries, human
exposures to Co have been associated with
non-inflammatory cardiomyopathy leading to heart
failure. Elevated levels of Co were present in the heart tissue
(5.11.).
1.8.6. Cardiomyopathy and elevated Co levels in human heart
tissue has also been found in beer-associated exposures
to Co (from Co salts added as foam enhancers by some breweries
during the 1960s). Of those who developed cardiomyopathy, 43% died.
Exposures ranged from 1.0 - 1.5 µg/mL Co in beer that was
ingested at a rate of 8 - 30 pints / day for a period of years (0.04
- 0.14 mg per kilogram body weight per day). The protein deficient
diets of beer drinkers may have contributed. Cardiac effects have
not been seen in people treated for anemia with 1 mg Co per kilogram
body weight (5.10.).
1.8.7. In guinea pigs, obliterative bronchiolitis adenomatosis is
observed when high doses of cobalt are injected intratracheally.
Animal studies appear to confirm hypersensitivity but do not
necessarily model human lung responses (5.2.).
2. Sampling
2.1. Equipment
2.1.1. Mixed-cellulose ester (MCE) filters
(0.8-µm pore size), cellulose backup pads, and cassettes,
37-mm diameter, part no. MAWP 037 A0 (Millipore Corp.,
Bedford, MA) are used. Cassettes, MCE filters, and backup pads of
25-mm diameter can also be used.
Note: Polyvinyl chloride (PVC) filters are applicable for
analysis if only soluble forms of W or Co are suspected. See Table 7
in Section 1.6. for solubility designations of materials in various
common industrial applications. See reference 5.1. for any materials
not listed in Section 1.6. Insoluble forms of W or Co collected on
PVC cannot be analyzed by this method, but may be analyzed
gravimetrically in the field and may also be submitted for
"Qualitative XRF" analysis (OSHA Method no. ID-204) for
elements present (atomic numbers 13-92 which include W
and Co). Always submit blank samples for any analysis.
2.1.2. Gel bands (Omega Specialty Instrument Co., Chelmsford, MA)
for sealing cassettes.
2.1.3. Calibrated personal sampling pumps capable of sampling
within "5% of the recommended flow rate of 2 L/min.
2.1.4. Various lengths of polyvinyl chloride tubing to connect
sampling cassettes to pumps.
2.1.5. A stopwatch and bubble tube or meter to calibrate pumps.
2.1.6. Smeartabs (part no. 225-24, SKC Inc., Eighty
Four, PA) for wipe sampling.
2.1.7. Scintillation vials, 20 mL, (part no. 74515 or 58515,
Kimble, Div. of Owens-Illinois Inc., Toledo, OH) with
polypropylene or Teflon liners for bulk and wipe sampling. See
Section 2.3. for details.
2.2. Sampling Procedure - Air Samples (5.1.)
Collect samples on 0.8-µm pore size, 37-mm
diameter MCE filters. Conduct the particulate sampling with the filter
cassette located within the breathing zone of the employee.
Note: Some welding operations involve W. If a welding operation is
being sampled for W, conduct the welding fume sampling with the MCE
filter cassette located inside the welding helmet. If the
free-space inside the hood precludes the use of
37-mm diameter cassettes and MCE filters,
25-mm sampling assemblies with MCE filters can be used.
2.2.1. Place a MCE filter and a cellulose backup pad in each
two- or three-piece cassette. Seal each
cassette with a gel band.
2.2.2. Prepare at least one blank for each batch of ten samples.
Use the same lot of MCE filters for blank and collected samples.
Handle the blank filters in the same manner as the samples except
that no air is drawn through blanks.
2.2.3. Calibrate each personal sampling pump with a prepared
cassette in-line to approximately 2 L/min flow rate.
2.2.4. Attach prepared cassettes to calibrated sampling pumps
(the backup pad should face the pump) and place in appropriate
positions on the employee or in the workplace area. For STEL
samples, use a flow rate of 2 L/min and a minimum sampling time of
15 min. For TWA determinations, take two consecutive
480-L samples at a flow rate of 2 L/min for 240 min
each.
Note: If other regulated soluble compounds (e.g.,
Cr2+, Cr3+,
soluble salts of Al, Fe, Mo, Ni, ZnCl2,
etc.) are suspected to be present in the sampled air, take separate
samples. Request analysis for the specific regulated compound(s).
These samples are analyzed using OSHA Method No. ID-121
and not by this method.
2.2.5. If the filter becomes overloaded while sampling, another
filter cassette should be prepared. Consecutive samples using
shorter sampling periods should be taken if overloading occurs.
After sampling, place plastic end caps tightly on both ends of the
cassette.
2.2.6. Special instructions for soluble W samples: Prudent
precautions should be used to stabilize certain air samples after
collection. In humid air, some soluble W compounds are deliquescent,
i.e., they can absorb water vapor from humid air and form a
solution. Such solution formation on a filter could spread to the
backup pad (which is not analyzed) resulting in low recoveries. For
soluble W samples:
1) Open the cassette. 2) Carefully lift out the MCE filter
with forceps. Do not disturb any of the particulate on the filter.
3) Remove and discard the backup pad. 4) Replace the MCE
filter in its original orientation.
Apply OSHA 21 seals in such a way as to secure the end caps.
Record the sampling conditions. Treat the blank(s) in the same way.
For air samples on MCE filters that are suspected of containing only
soluble W, request "Soluble W." Otherwise, request "Soluble and
Insoluble W" analysis on the OSHA 91A. Ship air and blank samples to
the lab. Cobalt results will be reported in addition to the
requested W analysis.
2.2.7. Instructions for air samples not containing soluble
W: Apply OSHA 21 seals in such a way as to secure the end caps.
Record the sampling conditions. Specify "W+Co" analysis and ship air
and blank samples to the laboratory. When "W+Co" is requested,
soluble W will not be reported.
2.2.8. When other compounds are known or suspected to be present
in the air, such information should be transmitted with the sample.
This is particularly important for elements not included among the
22 screened elements listed in Section 1. Other more exotic elements
may produce spectral interferences in the analysis of W and Co. The
industrial hygienist should note the identity of additional elements
suspected to be present in the workplace atmosphere so that the
analyst can evaluate possible interferences.
2.3. Bulk Samples and Wipe Samples
Very finely divided bulk material can be analyzed by this method.
Settled dust samples are preferred. The laboratory may not be able to
grind coarse samples of many of the hard industrial W matrices to a
fine enough particle size for digestion by this method. As an
alternative for bulk analysis, OSHA Method ID-204 may be
used to assess the surface composition of bulk samples. If there are
any questions about bulk sampling for W or Co, call the OSHA SLTC.
Wipe samples taken on smear tabs may be submitted for analysis.
Do not collect wipe samples using PVC filters, large Whatman
filters, or backup pads.
2.3.1. Bulks: Place bulk samples in 20-mL
scintillation vials. Fill 20-mL scintillation vials at
least half full. Large pieces that do not fit inside
20-mL scintillation vials may be shipped in larger
containers.
2.3.2. Wipes: Prepare at least one blank for every ten
wipe samples.
1) Wear clean, impervious, disposable gloves when taking each
wipe sample. 2) Moisten the wipe filters with deionized water
prior to use. 3) If possible, wipe a surface area covering 100
cm2. 4) Fold the wipe sample with the exposed side in. 5)
Transfer the wipe sample into a 20-mL scintillation
vial, seal with a cap having an inert plastic liner, and wrap with
vinyl or electrical tape. Securely wrap an OSHA-21
seal length-wise (top to bottom) around the vial.
2.3.3. Complete the OSHA Form 91A. See Table 7 in Section 1.6.
for solubility designations for materials in various common
industrial applications. Specify "W+Co", "Soluble W", "Soluble and
Insoluble W" and/or "Qualitative XRF" analyses as appropriate and
ship the bulk and wipe samples to the laboratory separately from air
samples.
3. Analysis
3.1. Safety Precautions
3.1.1. Refer to appropriate ICP-AES instrument
manuals and the standard operating procedures (SOP) for proper
instrument operation (5.12., 5.13.).
3.1.2. Observe laboratory safety regulations and practices.
3.1.3. The reagents NH3(aq), HCl,
HNO3, and
H3PO4 are
corrosive. Use appropriate personal protective equipment such as
safety glasses, goggles, splash shield, gloves, and lab coat when
handling corrosive chemicals. Prepare solutions in an exhaust hood.
3.2. Equipment
3.2.1. Inductively coupled plasma atomic emission spectrometer
(ARL Model 3560 or equivalent configured with channels Co, W, and
major interferant elements).
3.2.2. Automatic sampler, if desired.
3.2.3. Laboratory computer system for data reduction, if desired.
3.2.4. Filtering apparatus for extractions: vacuum system or
aspirator, assorted hoses, trap, suction flask, filtering stand,
filtering funnel, clamp, wash bottle, and container to receive
filtrate.
3.2.5. Miscellaneous volumetric ware: Pipettes; micropipettes;
50-1,000 µL pipette tips* (such as part no.
RC-200, Rainin Instrument Co., Inc., Woburn, MA);
volumetric flasks; Erlenmeyer flasks; graduated cylinders; and
beakers.
* Note: Concentrated
H3PO4 is syrupy
and wets glass. To deliver aliquots of
H3PO4
quantitatively and reproducibly, use a plastic pipette tip that is
not appreciably wetted by
H3PO4.
3.2.6. Parafilm®.
3.2.7. Forceps.
3.2.8. Plastic wrap or aluminum foil.
3.2.9. Thermostated hot plate(s) 70 to 100 °C and 180 to 200 °C.
3.3. Reagents
All chemicals should be at least reagent grade. Refer to Material
Safety Data Sheets (MSDSs) for safe and proper handling of reagents.
3.3.1. Reagents:
Deionized water (DI H2O). Ammonium
hydroxide (NH3(aq)), concentrated (28.6%
as NH3). Hydrochloric acid (HCl),
concentrated (36.5 - 38.0%). Nitric acid
(HNO3), concentrated
(70.9%). Phosphoric acid
(H3PO4),
concentrated (85.5%).
3.3.2. ICP stock standards:
The stock standards used in the validation are listed in Table 11
in Section 4. All calibration standards are prepared from 1,000
µg/mL ICP stock standards (except for the 10,000 µg/mL
Fe standard). Atomic spectral (ICP) standards may be used. Standards
normally classified as "atomic absorption" may not be appropriate
due to the presence of contaminants. Stock standards prepared in
water, dilute nitric acid, and dilute hydrochloric acid matrices are
suitable provided that they do not cause precipitation when mixed
together to prepare calibration standards.
3.4. Preparation of Standards
Table 9 below summarizes the preparation of standards. All the
calibration standards and the reagent blank use a final 2%
H3PO4 matrix
except for the 1,000 µg/mL solution of Ti. The 1,000
µg/mL Ti stock standard is used neat (without any added
H3PO4) because a
high concentration of Ti in 2%
H3PO4 tends to
produce a precipitate after several hours. Prepare the calibration
standards (except Ti) in 100-mL volumetric flasks
containing 2 mL
H3PO4 and about 25
mL DI H2O.
The ICP stock solutions listed in Table 11 in Section 4. were
compatible when mixed together at the dilutions used in preparing the
standards listed below. If substitutions for these stock solutions are
made, different combinations for standards may be necessary in order
to avoid possible precipitation reactions, (e.g., do not mix together
ICP stock solutions containing
SO42- and
Pb2+). After adding the aliquots of ICP
stock standards, dilute to 100 mL with DI
H2O. Mix thoroughly.
Immediately transfer the standard which contains the silicon
standard (labelled "4" in Table 9 below) to a plastic bottle. The
other standards may be stored in glass volumetric flasks.
Table 9
Label |
ICP Stock Standards |
mL |
Final Concentration
(µg/mL) |
Acid |
none |
NA |
0 (reagent blank) |
1 |
Be, Cd, Co*, Cu, Pb, Zn |
0.200 |
2.00 each |
2 |
Cr, Mo, Ni, Sb, V, Fe |
0.200 |
2.00 each and 20.00 for Fe |
3 |
Ca, Mg, Mn, Al, As, Se, Te |
0.200 |
2.00 each |
4 |
Sn, Si, Zr |
1.00 |
10.00 each |
W |
W |
10.00 |
100.0 |
Co |
Co* |
1.000 |
10.00 |
Ti |
Ti |
NA |
1,000. |
* |
A three point calibration
curve is used for Co; two point calibration curves are used for
the other elements. |
3.5. Sample Preparation
A flow chart is shown in Figure 1 to assist the analyst in
preparing samples. Observe laboratory safety rules. Use personal
protective equipment necessary to conduct safe acid digestions and a
laboratory exhaust hood appropriate for acid use when preparing
samples and glassware.
Figure 1 Overview of Tungsten Air Sample
Preparation.
3.5.1. Glassware cleaning.
1) Clean 125-mL Phillips beakers to be used for
sample digestions by refluxing with 1:1
HNO3 in an exhaust hood. Rinse well with
DI H2O.
2) Rinse the 50- and 100-mL
volumetric flasks with 10% HNO3. Rinse
well with DI H2O.
3.5.2. Soluble W air sample preparation.
Note: It is often possible to determine by the nature of the
operation that only soluble W or only insoluble W may be present.
(See Table 7 in Section 1.6. for solubility designations of various
common tungsten materials in industrial applications.) In such cases
where only one solubility designation applies, the separate
extraction and analysis of soluble W is not necessary. For example,
grinding operations using cemented tungsten carbides normally need
not be analyzed for soluble W.
1) Carefully remove the filter from the polystyrene cassette
with forceps and place the sample filter, dust side up, on the
moistened platform of a filtering apparatus. Secure the filtering
funnel, and add 3-mL DI
H2O.
Note: If the sample filter is torn, place the filter on the
filtering stand on top of a laboratory 0.8-µm pore size MCE
filter. (If insoluble W is requested, combine filters. For
corresponding blank corrections, combine a field and laboratory
blank.)
Allow to stand for 3 minutes. (See Section 3.5.3. step 2
regarding treatment of contaminated backup pads.)
2) Apply vacuum while supplying several mL of DI
H2O from a wash bottle to quantitatively
transfer the solution to a receiving container.
3) Quantitatively transfer the solution from the receiving
container to a clean 50-mL volumetric containing 1 mL
H3PO4 in about
10 mL DI H2O. Dilute to volume with DI
H2O and mix well.
4) If insoluble W is also requested, carefully place each
extracted filter in a clean labelled 125-mL Phillips
beaker so that it rests on the bottom of the beaker.
Note: Care should be exercised to avoid getting the filter or
filter fragments up on the glass beaker where digestion will be
inefficient.
5) For the soluble portion, proceed to Section 3.6. below.
6) For the insoluble portion, proceed to Section 3.5.3., steps
2 to 11.
3.5.3. Insoluble W air sample preparation.
Complete sample digestion of certain insoluble W matrices
requires reaction times for steps 3 and 8 listed below that exceed
the analyst's normal 8-h workday. For this reason,
overnight reaction times are indicated for these two steps. If less
time is given for digestion, low recoveries are observed. To avoid
unexpected complications, it is important to observe both the
sequence of operations and the hot-plate temperature
ranges given in the digestion procedure. For example, excessive or
prolonged heating (i.e., far beyond the endpoint indicators
described in the associated note) in the final digestion may allow
the H3PO4 to
dissolve some of the silica in the Pyrex Phillips beaker and
complicate the analysis by producing a phosphosilicic acid gel. No
significant concentrations of Si were found when this method was
validated using the procedure stated below. In the unlikely event
that a gel forms, transfer the suspension to transparent plastic
ware and try clearing the gel with the minimum
drop-wise addition of HF.
Note: Use extreme caution in handling concentrated HF.
Hydrofluoric acid is both corrosive and toxic. It is particularly
hazardous because it rapidly penetrates deep below intact skin and
causes extensive tissue damage. Check first aid procedures before
using. Cleanup spills promptly with appropriate spill cleanup
materials.
Observe laboratory safety rules. Use personal protective
equipment and an appropriate laboratory exhaust hood when preparing
samples.
1) Carefully remove the filter from the polystyrene cassette
with forceps.
2) Transfer each sample filter into a clean labelled
125-mL Phillips beaker so that it rests on the bottom
of the beaker. If dust collected during sampling no longer adheres
to the filter and is found on the inside walls of the cassette,
carefully rinse out the insides of the cassette into the
appropriate beaker with a few mL of DI
H2O.
Note: Care should be exercised to avoid getting the filter or
filter fragments up on the glass beaker where digestion will be
inefficient.
Contaminated backup pads may be analyzed in the manner of MCE
filters with the additional treatment noted in step 10 below. If
analysis of contaminated backup pads is performed, also analyze
the backup pad of the blank so that blank corrections may be
performed.
3) In an exhaust hood, add 1 mL of concentrated
NH3(aq) to each Phillips beaker and
quickly seal the tops of the Phillips beakers with Parafilm
followed by plastic wrap or aluminum foil. Allow to stand
overnight.
Note: The concentrated NH3(aq)
weakens MCE filters and dissolves any
WO3 to form soluble tungstate. The
WO3 matrix is not soluble in the acid
digestion steps used in this method. Some Co in samples containing
cemented tungsten carbides may also dissolve as a result of air
oxidation and complexation of Co by NH3.
The minimum practical volume of concentrated
NH3(aq) to use in this step is 1 mL.
Concentrated NH3(aq) tends to evaporate
readily; therefore, to avoid loss of NH3
and speed the process, the plastic wrap and Parafilm should be cut
to size before the aqueous NH3(aq) is
added. More NH3(aq) (2 mL total) could
be used if desired because the excess is removed in step 4. Be
consistent in the volume of NH3(aq) used
for samples and blanks.
The H3PO4
used in Step 6 dissolves any remaining insoluble
tungsten-containing dust; it complexes both the
ferric iron and the tungstates that form during digestion.
4) After standing overnight, the sealed beakers should still
contain liquid. If no liquid is present, the previous step must be
repeated (including allowing it to stand overnight) because the
ammonia may have evaporated before the reaction was complete.
Note: If Co is present in sufficient amounts, a purple or brown
solution may result.
Uncover the beakers and quickly evaporate the remaining liquid
in an exhaust hood on a hot-plate at 70 to 100 °C.
Remove from the hot plate and cool to room temperature.
5) Add 1 mL DI H2O or distilled
H2O (to dissolve soluble tungstates and
prevent dehydration to insoluble tungstic acid when the acids are
added). Swirl to dissolve soluble components; some Co complexes
present may decompose to a brownish mass during the gentle heating
in the previous step but will quickly dissolve in subsequent steps
when the acids are added.
6) Add 1 mL of
H3PO4. Swirl
to mix well.
7) Add 1 mL of HCl. Swirl. A blue
[CoCl4]2-
complex may form.
8) Add 1 mL of HNO3. Swirl. Seal the
beakers with Parafilm (or cover with watch glasses) and allow to
stand overnight.
Note: The low viscosity of this mixed acid matrix solution is
important in dissolving the residual Co out of the tungsten
carbide composites. Diffusion is limited in these matrices; the
formation of both slightly soluble tungstates and gas bubbles
formed during the digestion may clog the passages between the
carbide grains. It takes time for the
H3PO4 to
diffuse between the grains and complex any tungstic acid that
forms. Gaseous oxidants present in the solution are less soluble
at elevated temperatures. This digestion is hindered if high
temperatures are applied too soon; therefore, the samples are
allowed to sit overnight for the reaction to take place.
9) Add 5 mL of HNO3 to the samples
and digest at 180 to 200 °C until the volume is reduced to
approximately 1 mL (mostly
H3PO4) and
digestion is complete. Cool to room temperature.
Note: To determine completion of digestion ( 1 mL left), the
1 mL tends to draw away from the center because Phillips beakers
have slightly raised centers. Completion is also evidenced by the
following events: The reddish-brown fumes above the
solution dissipate, and the golden solution turns colorless. The
solution becomes viscous and ceases to bubble. If Co is present in
a large amount, the solution may be slightly bluish when hot and
very pale pink when cool.
10) To digest any backup pads that may be contaminated with
sample, add 2 mL DI H2O and swirl to mix
well. Add 5 mL HNO3 and repeat step 9
once more.
11) Point the mouth of the flask away from yourself. Slowly and
cautiously add DI H2O to the
H3PO4 solution
and transfer quantitatively to 50-mL volumetric
flasks using DI H2O rinses. Dilute to
volume with DI H2O and mix well.
3.5.4. Bulk sample preparation:
Prepare bulk samples by weighing 3 to 5 mg of each dry bulk
sample on individual tared 0.8-µm pore size,
37-mm diameter, MCE filters that have reached constant
weight. Record the weights. Transfer the filter containing the
sample to a 125-mL Phillips beaker and prepare
according to either Section 3.5.2. or 3.5.3. above. Also prepare a
blank from the same lot of filters.
3.5.5. Wipe sample preparation:
Transfer each Smeartab wipe sample to individual
125-mL Phillips beakers and prepare according to either
Section 3.5.2. or 3.5.3. above. Also prepare a blank from the same
lot of Smeartabs.
3.6. Standard and Sample Analysis
Note: If particulate remains after digestion, the sample(s) should
be filtered using a disposable-syringe-type
Teflon® filter(s) prior to analysis to
prevent plugging the nebulizer.
1) Analyze the standards and reagent blank by
ICP-AES. Calibrate the instrument and analyze samples
according to the appropriate ICP instrument SOP (5.13.). Instrument
parameters used during this validation are listed in Table 10 in
Section 4.
2) If necessary, analyze standard and reagent blanks to determine
the detection limits using the appropriate manufacturer's software.
ICP-AES detection limits normally do not vary
significantly unless the instrument is altered through maintenance
or abuse.
3) Analyze the samples, calibration standards, and blanks by
ICP-AES.
4) Follow the SOP for further instructions regarding analysis
(5.13.).
5) Some samples require reanalysis. These include the following:
a) Dilute and reanalyze any samples which have calculated
concentrations above the analytical range. This especially
includes samples which saturate (overdrive) any channels. Evidence
for overdriven channels may vary, and is described further in the
ICP SOP for the particular instrument (5.13.). Pipet an
appropriate aliquot of the high concentration sample into a clean
volumetric flask and dilute to volume with 2%
H3PO4.
b) Reanalyze any samples and corresponding blanks if results
indicate a high degree of imprecision . This problem is evidenced
by a high relative standard deviation (CV > 10%) of the three
exposure measurements/sample.
3.7. Calculations
3.7.1. Perform interelement corrections. According to the
equation below, a final concentration for each analyte is
calculated:
Where: |
|
Cc |
= |
corrected concentration of an element |
Cunc |
= |
uncorrected concentration of the element |
Ci |
= |
concentration from each contributing
interference for the element |
Ki |
= |
respective Interelement Correction (IEC) factor
for the interference |
Application of this equation is an iterative process described in
greater detail in Section 4.3. These corrections are calculated
either "real-time" using software provided by the instrument
manufacturer or post-analysis using custom software.
3.7.2. Obtain hard copies of the data and results from a printer.
3.7.3. The amount of analyte in each sample or blank is
calculated as:
µg A |
= |
(µg/mL A) × (mL S) × (DF) |
|
Where: |
|
µg A |
= |
Total µg of analyte in the sample or
blank |
µg/mL A |
= |
Measured concentration of analyte in solution
(derived from calibration curve) |
mL S |
= |
Total volume of the solution analyzed |
DF |
= |
Amount of dilution applied to an aliquot of the
original solution (ratio of final volume divided by the
aliquot volume) |
3.7.4. The blank value, if any, is subtracted from each sample:
µgc A |
= |
(µg A) -
(µgb A) |
|
Where: |
|
µgc A |
= |
µg of analyte, blank corrected |
µg A |
= |
µg of analyte from equation in Section
3.7.3. above |
µgb A |
= |
µg of analyte in
blank |
Wipe samples are reported in terms of µg of analyte.
3.7.5. For air samples, calculate mg/m3
exposures for all screened elements, Co, soluble W, and insoluble W:
mg/m3 A = |
(µgc A) × (GF)
(L) |
Where: |
|
GF |
= |
Gravimetric factor |
L |
= |
Air volume in liters |
For those elements having a PEL listed as an oxide, the
gravimetric factors are calculated from formula weights (FW):
GF |
= |
1.2447 for
FWZnO/FWZn |
GF |
= |
1.4298 for FWFe2O3
/2FWFe |
GF |
= |
1.7852 for
FWV2O5/2FWV |
3.7.6. Compare the calculated mg/m3
exposures to the various PELs or OSHA action limits. Evaluate the
reliability of the results for all detected elements whose air
concentration may exceed a PEL or an action limit. With the custom
software available at the OSHA SLTC, the analytical results reported
for each sample includes a listing of the total interelement
correction and the corrected concentration for each element detected
in the sample.
3.7.7. Convert bulk sample analytes to % composition using:
% A = |
(µgc A) × (100%)
(mg B) × (1,000 µg/mg) |
Where: |
|
mg B |
= |
milligrams of bulk
sample |
3.8. Reporting Results
3.8.1. When evidence of migration of the sample to the backup
pad is observed on any air sample, advise the industrial hygienist
that part of the sample was collected on the backup pad. When this
occurs, report the sum of exposure results for the air filter sample
and the corresponding backup pad.
3.8.2. Report exposures in terms of
mg/m3 as appropriate for Co, soluble W,
and insoluble W.
3.8.3. Discuss high exposures of screened elements with your
supervisor and the industrial hygienist. Resubmit air samples and
blanks for additional confirmation and quantitation of any screened
elements near or over the PEL, or discuss resampling with the
industrial hygienist. Include copies of sampling information and
results when resubmitting samples.
4. Backup Data
This analytical method has been validated for the spiked MCE filters
specified in Section 2.1.1. The method validation was performed
according to guidelines (5.14., 5.15.) with the exception of the
sampling validation. The analytical method validation was conducted at
the respective OSHA Final Rule TWA-PELs and STELs for W and
Co. Several reasons precluded aerosol sampling validation.
a) Aerosol generation of cemented tungsten carbides is very
difficult because of the high density of this matrix. The great
hardness of this material precludes grinding to a smaller particle
size range.
b) The generation of cemented tungsten carbide aerosols of known
composition is not feasible because cemented tungsten carbides are
non-stoichiometric.
c) The non-stoichiometric nature of cemented tungsten
carbides makes the generation of uniform dispersions of aerosol of
known composition infeasible.
d) Equipment to generate and measure aerosols was not available at
the time of this validation.
Instrument parameters used during validation are listed in Table 10
below. (Additional information regarding this instrument may be found in
reference 5.7.):
Table 10
The Specifications for the
ARL 3560 Simultaneous ICP-AES |
RF GENERATOR: |
Snout Gas Flow (L/min) |
1.5 |
Generator (Model No. Henry) |
2.5 kW * |
Nebulizer Uptake Rate (mL/min) |
2.3 |
Incident Power (W) |
1,175 |
Nebulizer type |
Meinhard-Type A |
Reflected RF Power (W) |
0-20 |
Peristaltic Pump Used? (yes/no) |
no |
P.A. Plate Supply Voltage (V) |
4,100 |
Pre-Integration Flow Time (s) |
30 |
P.A. Plate Current (mA) |
700 |
Integration Time (s) |
5 |
P.A. Grid Current (mA) |
125 |
Number of Exposures |
† |
P.A. Filament Voltage (V) |
7.5 |
SPECTROMETER: |
P.A. Tune (Setting) |
190 |
Instrument Model |
3560 |
P.A. Load (Setting) |
050 |
Grating (lines/mm) |
1,080 |
(where P.A. = Power Amplifier) |
|
Primary Slit Size (µm) |
20 |
EXCITATION: |
Profiling Element |
Mn |
Plasma Observation Height (mm) |
15 |
Profile Point (peak dial div.) |
496 ‡ |
Coolant Gas Flow (L/min) |
12 |
Vacuum (µmHg) |
20 |
Coolant Gas Pressure (psi) |
25.5 |
Path Length (m) |
1 |
Plasma Gas Flow (L/min) |
0.8 |
COMPUTER
CONFIGURATION: |
Plasma Gas Pressure (psi) |
21.5 |
Disk Drive Capacity |
30 Megabyte fixed 1.2 Megabyte 5.25 in.
floppy |
Carrier Gas Flow (L/min) |
1 |
Computer System |
Micro 11/53 DEC Computer |
Carrier Gas Pressure (psi) |
30.5 |
Software (operating system) |
DPS/TSX+ |
* Indicates peak rf power capability in kW. † Dependent on
software program used; 10 for DL determinations, and 3 for calibrations
and analyses. ‡ From spectral scanning device (SAMI). This value can
change over time.
The instrument calibration standards were those listed in Table 11
below:
Table 11
ICP Stock Standards* |
Matrix |
ICP Stock Standards* |
Matrix |
Al |
0.3 M HCl |
Ni |
0.3 M
HNO3 |
As |
0.3 M HCl |
Pb |
0.3 M
HNO3 |
Be |
0.3 M HCl |
Sb |
H2O |
Ca |
0.3 M
HNO3 |
Se |
0.3 M
HNO3 |
Cd |
0.3 M
HNO3 |
Si |
H2O |
Co |
0.3 M
HNO3 |
Sn |
2.5 M HCl |
Cr |
H2O |
Te |
10% HCl |
Cu |
0.3 M
HNO3 |
Ti |
H2O |
Fe |
5% HNO3 |
V |
0.3 M
HNO3 |
Mg |
0.3 M
HNO3 |
W |
H2O |
Mn |
0.3 M
HNO3 |
Zn |
0.3 M
HNO3 |
Mo |
H2O |
Zr |
2%
HNO3 |
* All calibration standards are prepared from 1,000
µg/mL ICP stock standards (except for the 10,000
µg/mL Fe standard). |
The method validation consisted of the following experimental
protocol:
1. Preliminary assessments of reference materials.
2. Preliminary determination of W and Co analytical recoveries of
diverse tungsten-containing matrices. Preparation and
analysis of two spiked samples each of six different
solid-phase tungsten-containing matrices representative
of industrial exposures individually spiked on MCE filters. These
studies included two different cemented tungsten carbide matrices
containing Co.
3. Determination of interelement corrections.
4. Determination of the qualitative and quantitative detection
limits and background equivalent concentrations.
5. Determination of analytical method precision and accuracy for
insoluble W. Preparation and analysis of spiked MCE filter samples of
three samples each of the two most difficult insoluble matrices (94%
WC/6% Co and 88% W2C/12% Co) at W exposures
at 0.5 ×, 1 ×, and 2 × TWA PEL. Preparation and analysis of six spiked
MCE filter samples of pure W metal which represent concentrations of
insoluble W near the STEL.
6. Determination of analytical method precision and accuracy for
soluble W. Preparation and analysis of spiked MCE filter samples of
six samples each of soluble W
(Na2WO4.2H2O)
representing W exposures at 0.5 ×, 1 ×, and 2 × TWA PEL. Preparation
and analysis of six spiked MCE filter samples which represent
concentrations of soluble W near the STEL.
7. Determination of analytical method precision and accuracy for
Co. Preparation and analysis of solution-spiked MCE
filter samples of six samples each of Co at 0.5 ×, 1 ×, and 2 × TWA
PEL. These were also spiked with W to give a 100 µg/mL W
solution matrix.
8. Determination of analytical method precision and accuracy for Co
at higher levels. Analysis of solution-spiked MCE filter
samples of three samples each of Co at 3.2 ×, 6.5 ×, 7.4 ×, 12.3 ×,
14.9 ×, and 28.0 × TWA PEL.
9. Assessment of typical levels of blank contamination, analyte
carry-over, and noise. Preparation and analysis of four
blank MCE filter samples for W and Co.
10. Summary.
All analytical results were calculated from
concentration-response curves, corrected for reagent blank
and interelement interferences as identified in Section 4.3., and
statistically examined. In addition, the analytical recoveries were
tested for possible outliers using the Treatment of Outliers test
(5.8.). For samples tested at different levels, the appropriateness of
pooling was tested by checking the homogeneity of variance using the
Bartlett's test (5.17.). Statistical evaluation was conducted according
to Inorganic Methods Evaluation Protocol (5.14.). To reduce cumulative
round-off errors during calculations in this method,
additional non-significant digits were preserved in the
intermediate results shown in the backup data in Section 4. Results
shown in the main body of the method are shown rounded off to the
correct number of significant figures.
4.1. Preliminary Assessment of Reference Materials:
In order to develop a method for measuring atmospheric exposures in
the workplace, unsieved reference materials of six diverse common
industrial tungsten-containing matrices were selected as
models
(Na2WO4.2H2O,
WC, 94% WC/6% Co, 88% W2C/12% Co,
WO3, and W).
Procedure
The cemented tungsten carbide matrices were
manufacturer-labelled with nominal descriptions "88%
W2C/12% Co" and "94% WC/6% Co" indicating
only the approximate elemental composition of these matrices.
Certificates of Analysis provided accurate analytical results for
certain major and trace elements. For the purpose of this validation,
these matrices were further characterized by qualitative powder
X-ray diffraction analysis to identify the crystalline
phases of tungsten carbide actually present. Appropriate
stoichiometric factors could then be calculated based on the
information presented on the respective Certificates of Analysis.
Particle sizing for each of the six materials was also assessed
based on product information, measurement, and visual inspection.
Results
X-ray Diffraction Analyses
The powder diffraction patterns corresponding to WC and Co were
identified by X-ray diffraction analysis of the 94% WC/6%
Co matrix. X-ray diffraction analysis results for this
matrix were not reported on the Certificate of Analysis provided by
the manufacturer.
The powder diffraction patterns corresponding to WC, Co, and
Co6W6C were
identified in the 88% W2C/12% Co matrix. The
compound with the formula W2C (suggested by
the label 88% W2C/12% Co) was not identified
by X-ray diffraction analysis. The Certificate of
Analysis for this matrix indicated that X-ray diffraction
analysis had identified the major phase as WC, with traces of
Co6W6C and
Co3W3C. The
compound W2C was not listed on the
Certificate of Analysis.
The use of these materials in the validation offers an excellent
opportunity to determine recoveries for chemically bound and
interstitial elemental Co.
Particle Size Assessments
The
Na2WO4.2H2O
was particle-sized using a 41 µm sieve and a
metric ruler. It was found to consist primarily of granular
crystalline material with particles in the size range of 41 µm
to 1 mm. Particle sizing on the other materials was not performed
because the manufacturer product descriptions of the particle size
ranges were available. These descriptions are included in the Tables
12 to 15 in Section 4.2.
4.2. Preliminary Determination of W and Co Analytical Recoveries of
Diverse Tungsten-Containing Matrices:
This study was performed to test the matrix independence of the
digestion procedure of this analytical method. Six diverse
tungsten-containing matrices described in Section 4.1.
were digested and analyzed. To ensure that the digestion procedure was
compatible with the MCE filter used for sampling and would also have
adequate capacity, several of these samples were spiked with amounts
of reference material approaching 10 mg.
Because this method is based on the analysis of elemental
composition, gravimetric conversion factors must be used to convert
from the mass of a matrix spike to the mass of an analyte element. In
order to accurately evaluate recoveries, the manufacturer's reagent
label information and Certificates of Analysis were used to calculate
the gravimetric factors listed in Tables 12 to 15 below. In addition
to chemical formula information, these data included percent purity
and, in some cases, elemental composition. For example,
Na2WO4.2H2O
was listed as 100.4% pure based on tungsten content (presumably due to
water of hydration lost during manufacture). In the case of the
tungsten carbide matrices, the respective gravimetric factors implied
by the nominal compositions 88% W2C/12% Co
and 94% WC/6% Co were in good agreement with the gravimetric factors
calculated from the certification provided by the manufacturer.
Procedure
Matrix spikes were prepared from reagent and certified sources
independent of calibration standards. Two samples of each of the six
dry matrices were prepared. Known amounts of each dry sample matrix
were added to individual tared 0.8-µm pore size,
37-mm diameter, MCE filters that had reached constant
weight in a controlled low humidity lab environment (35% RH). All 12
samples were digested and analyzed according to this method using
50-mL solution volumes and the same ICP-AES
instrument described at the beginning of Section 4. Because of their
importance to the method, the tungsten carbide/Co samples were
re-analyzed. Interelement interferences were noted and
were corrected for the affect of Co on W and for W on Co. The details
for determining these corrections are further discussed in Section
4.3. Interelement corrections were not performed on screened elements
for this preliminary test because they were present in these materials
at trace levels.
The amounts of W and Co in these analyte spikes were calculated
from assay information provided by the chemical manufacturer. All W
and Co results were calculated from
concentration-response curves, corrected for reagent
blank, corrected for interelement interferences between W and Co, and
statistically examined.
Results
Results for the preliminary analyses for spiked W and Co samples
are shown in Table 12 (repeat analyses of tungsten carbide/Co samples
are tabulated separately):
Table 12
Tungsten Matrix Digestion
Evaluation - Tungsten Results |
Matrix in 2%
H3PO4 |
Amount (mg) |
% W in Matrix |
µg/mL Theory (T) |
µg/mL Found (F) |
W Recovery F/T |
W, powder |
2.446 |
100.00 |
48.92 |
49.72 |
1.0164 |
W, powder |
4.769 |
100.00 |
95.38 |
96.30 |
1.0097 |
WO3,
powder |
3.068 |
79.10 |
48.54 |
48.61 |
1.0014 |
WO3,
powder |
6.121 |
79.10 |
96.83 |
95.00 |
0.9811 |
WC, < 1 µm |
3.496 |
93.40 |
65.31 |
65.83 |
1.0080 |
WC, < 1 µm |
3.637 |
93.40 |
67.94 |
69.37 |
1.0211 |
94% WC, < 45
µm |
4.201 |
87.87 |
73.83 |
71.67 |
0.9708 |
94% WC, < 45
µm |
9.922 |
87.87 |
174.37 |
167.70 |
0.9618 |
88%
W2C, 5 to 45 µm |
5.742 |
84.08 |
96.56 |
97.37 |
1.0084 |
88%
W2C, 5 to 45 µm |
8.551 |
84.08 |
143.79 |
145.49 |
1.0118 |
Na2WO4.2H2O |
2.401 |
55.96 |
26.87 |
27.07 |
1.0074 |
Na2WO4.2H2O |
7.638 |
55.96 |
85.48 |
87.88 |
1.0280 |
Na2WO4.2H2O |
9.684 |
55.96 |
108.38 |
106.50 |
0.9826 |
Average |
1.0007 |
(SD =
0.0202) CV |
±0.0202 |
The tungsten carbide spiked samples were reanalyzed for W
confirming the recoveries obtained above. These results are shown in
Table 13:
Table 13
Tungsten Matrix Digestion
Evaluation - Tungsten Results (Repetitions) |
Matrix in 2%
H3PO4 |
Amount (mg) |
% W in Matrix |
µg/mL Theory (T) |
µg/mL Found (F) |
W Recovery F/T |
94% WC (repeat) |
4.201 |
87.87 |
73.83 |
72.26 |
0.9788 |
94% WC (repeat) |
9.922 |
87.87 |
174.37 |
166.61 |
0.9555 |
88% W2C
(repeat) |
5.742 |
84.08 |
96.56 |
96.71 |
1.0016 |
88% W2C
(repeat) |
8.551 |
84.08 |
143.79 |
145.05 |
1.0087 |
Average |
0.9862 |
(SD =
0.0241) CV |
±0.0244 |
The same tungsten carbide spiked samples were also analyzed on the
ICP-AES instrument giving the Co recoveries shown in
Table 14:
Table 14
Tungsten Matrix Digestion
Evaluation - Cobalt Results |
Matrix in 2%
H3PO4 |
Amount (mg) |
% Co in Matrix |
µg/mL Theory (T) |
µg/mL Found (F) |
Co Recovery F/T |
94% WC, < 45 µm |
4.201 |
5.45 |
4.58 |
4.93 |
1.0766 |
94% WC, < 45 µm |
9.922 |
5.45 |
10.82 |
11.56 |
1.0689 |
88% W2C, 5
to 45 µm |
5.742 |
11.70 |
13.44 |
13.65 |
1.0159 |
88% W2C, 5
to 45 µm |
8.551 |
11.70 |
20.01 |
20.44 |
1.0215 |
Average |
1.0457 |
(SD =
0.0314) CV |
±0.0301 |
These samples were reanalyzed for Co confirming the recoveries
shown above. The results of the Co reanalyses are listed in the Table
15:
Table 15
Tungsten Matrix Digestion
Evaluation - Cobalt Results (Repetitions) |
Matrix in 2%
H3PO4 |
Amount (mg) |
% Co in Matrix |
µg/mL Theory (T) |
µg/mL Found (F) |
Co Recovery F/T |
94% WC (repeat) |
4.201 |
5.45 |
4.58 |
4.96 |
1.0832 |
94% WC (repeat) |
9.922 |
5.45 |
10.82 |
11.53 |
1.0661 |
88% W2C
(repeat) |
5.742 |
11.70 |
13.44 |
13.53 |
1.0070 |
88% W2C
(repeat) |
8.551 |
11.70 |
20.01 |
20.44 |
1.0215 |
Average |
1.0445 |
(SD =
0.0361) CV |
±0.0345 |
4.3. Determination of Interelement Corrections (IECs):
Spectral interferences in ICP-AES generally occur when
weak lines of one or more interfering elements overlap the analytical
line of another element. The configuration of 25 channels included in
this method (2 Fe channels and 23 other channels) provides the
opportunity to correct for up to 575 potential interferences (2 × 23 +
23 × 23 = 575). Significant interferences normally need to be
corrected for by using interelement correction factors. The equation
used to calculate IEC factors is:
Ki =
Bi /
Ci
Where: |
|
Ki |
= |
IEC factor for a particular pairing of
affected element with affecting element |
Bi |
= |
apparent concentration determined on the
channel of the affected element produced by aspiration of
the single-element standard of the interference
element |
Ci |
= |
concentration of the interference
element |
These IEC factors are typically small. A typical factor of 2 ×
10-4 means that a 1,000 µg/mL
standard of the interferant produces a signal on the channel
comparable to 0.2 µg/mL of the element represented by the
channel. The determination of interelement correction factors requires
the presence of large amounts of the interfering elements; therefore,
the stock solutions used in the interference study have concentrations
that far exceed levels expected in actual samples. The solutions used
were of high purity and expected to give reliable indications of
interferences. However, interferences on the Si channel may not be as
reliable due to contamination from Si in the glassware used in
manufacture and preparation.
Spectral corrections are minimal for most screened elements but are
included to avoid misleading indications of overexposures to screened
elements. Interferences from matrix background and contamination are
minimized by the experimental design. During calibration, a reagent
blank is measured on all channels. The intercept of the resulting
calibration curve therefore corrects for the reagent matrix background
and for any contamination from the aqueous ammonia and reagent acids.
When the MCE filter or Smeartab wipe sample medium contributes
background, the blank sample media results can also be subtracted.
When the sample dust matrix contributes to the background, only
interferences from those elements measured on other channels can be
corrected in this method. Spectral interferences on an affected line
are corrected based on measurements at the analytical lines of known
interferant elements. Most elements potentially present in W matrices
are included in the list.
Background corrections were not applied in this method other than
for intercept and blanks. Other approaches to background corrections
are often provided by instrument manufacturers. Background
interference corrections are sometimes performed by background
modelling. The simplest model is to assume that the background level
measured at a preselected wavelength near the affected line represents
the background level "under" a peak signal. Such an approach considers
the elevated background without regard to cause and requires a
familiarity with the sample matrix for all samples to be analyzed. A
wide variety of sample matrices may be problematic using this
approach, unless each sample is scanned to determine that the
background model is appropriate. It is an appropriate strategy if the
background is known to be level across the ICP scan, and the
selected background wavelength does not overlap spectral interference
peaks of any other elements (not just screening elements). Such
assumptions regarding background corrections may be inappropriate for
matrices submitted for routine analyses by this method; therefore,
that approach was not used.
A few interferences having an elevated background component were
noted (such as the interference of W on the Co channel). In this
method, both background and spectral interferences were treated as
linear functions of the interfering element and are corrected as
spectral interferences.
Once determined, the IEC factors are used in the iterative
interelement correction equation given in Section 3.7.1. When the
channels for interference elements do not have interferences from
other elements; the measured concentrations of the interferant
elements are accurate, and the corrections are first order and can be
made in one step. Second and higher order interferences are common in
ICP; interferences themselves often have interferences, and the
initial estimates of interference element concentrations may be
progressively corrected in an iterative process. In such a process,
the concentrations of interference elements having interferences are
progressively corrected so that the Cc of
the interferant element resulting from the previous step substitutes
for Ci in making corrections in the
subsequent step. (Cunc in the equation does
not change.) Estimates of concentrations of both analytes and
interferences are improved at each step and the process is continued
until there are no significant differences in concentrations of the
elements calculated at each step or until the differences are less
than the corresponding detection limits. The iterative process
converges to an internally consistent set of results.
Procedure
For W, Co, and all screening elements, one or more standard
solutions each containing 1,000 µg/mL of a single interferant
element (10,000 µg/mL in the case of Fe) were analyzed, and
apparent concentrations were measured on the other element channels
for indications of interference. These stock solutions are listed in
Table 11 in Section 4.
The analysis of high concentration standards can result in
carry-over (See Section 4.9. for an evaluation of
carry-over). To reduce the possibility of misidentifying
minor carry-over between measurements as interference,
the solutions of the elements were analyzed twice: first in (near)
alphabetical order and then in reverse order. The lower of these two
measurements was used to calculate the interference correction
factors.
Archived scans for interferences on W and Co were visually examined
to assess the types of interferences affecting the analysis of these
elements and to help assess the appropriateness of using a linear
model to correct for these interferences.
Results
The IEC factors for each interference noted were calculated as
described in the introductory remarks above. These IEC factors are
instrument dependent; for brevity, only a relative ranking of
interferences is shown in Table 16 below along with pertinent channel
parameters, qualitative DL, and background equivalent concentration
(BEC) data (DL and BEC results are further discussed in Section 4.4.).
Interferences shown in bolded type have correction factors
exceeding 10-4 expressed as µg/mL
channel per µg/mL interferant. The range of interest for
interference correction factors extends below
10-4 for certain elements; these additional
detected interferences are indicated by element symbols in small type.
Interferences are included without distinction to background or
spectral type and are listed by element symbol in order of decreasing
magnitude. Only significant interferences are shown and are corrected
for. The cutoff criterion for determining significant interelement
correction factors was whether an element, at a concentration of 100
µg/mL, would produce an apparent concentration on another
element's channel exceeding 5% of the Final Rule Limit for that
element (50-mL solution volume and 480-L air volume). The
conservative criterion used to determine significance functions well
with this configuration of channels selected for use in sampling W and
Co in workplace atmospheres. The 100 µg/mL criterion is applied
to individual interfering elements, whereas 100 µg/mL
for a 50-mL solution volume corresponds to 5 mg material
on the MCE filter which represents a high estimate of the total amount
of all interfering and non-interfering elements
likely to result from air sampling. The 5% criterion represents a
small fraction of the PEL. In this method, the largest IEC factor used
was 0.09255 [(µg/mL Fe)/(µg/mL Co)] for the interference
of Co on the less sensitive Fe channel. The smallest
non-zero IEC factor used in this method was 0.00001
[(µg/mL Be)/(µg/mL Ti)]. The W contained in the matrices
used in the validation produces an Sb interference
(Ki = 0.01431) which illustrates the
importance in performing interelement corrections; as a
worst-case example, a 100 µg/mL W solution
produces a signal on the Sb screening channel indicating 1.4
µg/mL Sb (corresponding to approximately 0.25 × OSHA PEL for
Sb).
Although an interference was noted for Ti on both W and Co
determinations, in the absence of additional complexing agents (e.g.,
oxalate ion) any digested titanium would be converted to insoluble
titanium phosphate. For this reason, the interference from titanium
may be minimal. (Spot tests of various titanium compounds using the
digestion procedure indicate that the anatase and rutile forms of TiO2
are only slightly soluble, while titanium oxalate is completely
soluble.) The instrument in use at the OSHA SLTC was not equipped with
a Ti channel, but Ti could be monitored adequately on the Os channel
(225.585 nm) where Ti strongly interferes. Titanium has a peak
centered at 225.580 nm. Osmium is a very rare element that is not
expected to be present in workplace atmospheres sampled for W and Co.
In addition to those listed, other potential interferences were
identified in this study, but they were judged insignificant for the
expressed purposes of this method.
Table 16
ICP-AES Instrumental
Parameters and Observed Interferences |
Channel |
Wavelength (nm) |
Order * |
DL, µg/mL (2 × SD) |
BEC, µg/mL |
Interferences |
Al |
308.215 |
2 |
0.11050 |
6.32 |
Be |
As |
189.042 |
3 |
0.03150 |
1.30 |
Si Mo Cr Al Be W Fe1 |
Be |
313.042 |
2 |
0.00055 |
0.0345 |
V Zr Ti |
Ca |
393.366 |
2 |
0.00040 |
0.0891 |
|
Cd |
226.502 |
3 |
0.00425 |
0.232 |
Fe1 Ni Mo As Ti Co Al W |
Co |
228.616 |
2 |
0.00320 (ID-213) 0.00850
(ID-125G) |
0.569 |
Ti W Cr Ni Cd Mg Fe1 Mo Zr
Be |
Cr |
267.720 |
3 |
0.01000 |
0.766 |
|
Cu |
324.754 |
2 |
0.01000 |
0.665 |
Mo Cd |
Fe1 |
259.940 |
3 |
0.00600 |
0.431 |
|
Fe2 |
271.440 |
3 |
0.16150 |
9.34 |
Co |
Mg |
279.080 |
2 |
0.10250 |
5.86 |
|
Mn |
257.610 |
3 |
0.00195 |
0.0680 |
|
Mo |
202.030 |
3 |
0.01200 |
0.495 |
|
Ni |
231.604 |
3 |
0.01600 |
0.649 |
|
Pb |
220.353 |
3 |
0.06500 |
2.44 |
Al Ni W Co Zr Se Mn Cu Cd Zn
Fe1 |
Sb |
217.581 |
3 |
0.06250 |
2.46 |
W Be Zn |
Se |
196.090 |
3 |
0.05150 |
2.52 |
W Al Mn Sb |
Si |
288.158 |
2 |
0.06550 |
3.52 |
Sb Al W |
Sn |
189.980 |
3 |
0.01800 |
0.877 |
|
Te |
214.275 |
2 |
0.06450 |
3.79 |
V Cd Mo Ti Mg W Zn Al Fe1 Mn
Cu |
Ti † |
(225.585) |
3 |
(0.2 estimated) |
7.39 |
Fe1 Ni W |
V |
310.230 |
2 |
0.03000 |
1.21 |
Be Zr W |
W |
239.709 |
3 |
0.041 (ID-213) 0.04600
(ID-125G) |
2.56 |
Co Mo Fe1 Be Zr Ni Ti |
Zn |
213.856 |
3 |
0.00400 |
0.181 |
|
Zr |
343.823 |
1 |
0.01800 |
1.95 |
* See Section 1.2. for a
discussion of the use of higher-order diffraction
lines. |
† With the exception of
the Fe channels (Fe1 and Fe2), the symbols used in the Channel
and Interferences columns above are interchangeable with the
elements having the same symbol. The Fe1 channel was used to
monitor the Fe concentration in the samples, because it
generally provides more accurate corrections in the
concentration range expected in air samples. The Fe2 channel is
an alternate, less sensitive, channel that may be used at high
Fe concentrations to assist in determining appropriate dilutions
when the Fe1 channel is saturated. The third order osmium (Os)
channel at 225.585 nm was the channel used to monitor the Ti
concentration (see discussion in Section 1.5.2.) Titanium was
evaluated as a possible interference because it is common in
industrial operations where W and Co are found; Osmium was not
evaluated as an interference because it is an exceedingly rare
element. |
4.4. Determination of Qualitative and Quantitative Detection Limits
and Background Equivalent Concentrations (5.7.):
A modification of the International Union of Pure and Applied
Chemistry (IUPAC) qualitative detection limit equation (5.16.) was
used for this study. According to the IUPAC convention, 3 × SD defines
the qualitative detection limit; however, ICP-AES
manufacturers (including ARL) generally use 2 × SD in their software.
Procedure:
The instrument was calibrated for W, Co, and the screening
elements. BECs were calculated from the calibration information. The
detection limits for W and Co were measured during the validation.
Detection limit estimates for screened elements were obtained from
reference 5.7. Reagent blanks and calibration standards of 100
µg/mL W and 10 µg/mL Co were aspirated to obtain the
slope of the calibration curve. The reagent blank was measured ten
times at the W and Co wavelengths to determine the SDs of the
measurements. The ARL software (5.13.) was used to perform these
calculations. Each DL was calculated as follows:
m = (Istd -
Irblank) / Cstd
Qualitative DL = 2 × SDrblank / m
Quantitative DL = 10 × SDrblank / m
Where: |
|
SDrblank |
= |
standard deviation of the reagent blank
signal |
Istd |
= |
signal intensity of the aspirated calibration
standard |
Irblank |
= |
signal intensity of the aspirated reagent
blank |
Cstd |
= |
concentration of the aspirated standard |
m |
= |
slope of the calibration
curve |
Note: Some ICP-AES users and manufacturers prefer to
use BECs to estimate DLs. The BEC is defined as the concentration of
an analyte that is equal to the net intensity of the background signal
for that analyte:
BEC = Irblank / m
According to this latter approach, the DL is estimated to be a
small fraction, e.g., 2%, of the BEC.
Results:
The results for qualitative and quantitative detection limits are
shown in terms of solution and exposure concentrations in Tables 1 (in
Sections 1.4.2.) and 2 (in Section 1.4.3.). Qualitative DLs as well as
BECs for Co, W, and all the screening elements are shown in Table 16
(Section 4.3.) in terms of solution concentration. The DLs averaged 2%
of the corresponding BECs with a range of 0.5% to 3%.
4.5. Determination of Analytical Method Precision and Accuracy for
Insoluble W:
Tungsten carbide composites with Co binder were the most
representative of previous analyses requested at the OSHA Salt Lake
Technical Center and they also presented the most severe test of this
analytical method. Tungsten carbide and many other tungsten compounds
are very dense. The pure tungsten carbides WC and
W2C have densities of 15.63 and 17.15
respectively (5.9.). At 19.3 g/mL, the density of W is too high to
obtain stable aerosols above 5 µm diameter (5.2.). The
commercially available reference materials for these composites exceed
5 µm. For this validation the materials were not sieved because
they may not be sufficiently homogeneous for sieved material to match
the certified composition. In addition, larger particles are generally
less easily digested because they offer a small
surface-to-mass ratio. The use of the unsieved matrix
therefore, posed a more accurate and more difficult test for recovery
determinations than could be obtained using either aerosols or sieved
powders of these materials.
Procedure
Three spiked samples of each of the two most difficult insoluble
matrices, 94% WC/6% Co and 88% W2C/12% Co,
(certified materials described in Section 4.1.) were prepared by
spiking tared 0.8-µm pore size, 37-mm diameter MCE
filters that had reached constant weight in a controlled low humidity
lab environment (33 - 36% RH). These were prepared at three levels
representing W exposures at 0.5 ×, 1 ×, and 2 × TWA PEL for W.
Six spiked MCE filter samples representing concentrations of
insoluble W near the STEL were made from pure W in order to avoid
problems associated with small sample weight and the inhomogeneity in
the tungsten carbide matrices. All the reagents and certified
materials used for spiking were independent of the stock standards
used to prepare calibration standards.
Results
The theoretical amounts of W in all insoluble analyte spikes were
calculated from assay information provided by the chemical
manufacturer. The analytical method precision and accuracy for
insoluble W are shown in Tables 17 and 18.
Table 17
Analysis - Insoluble W TWA
Determination Tungsten Carbide Matrices * (2 L/min, 4 h,
480 L, 50-mL solution volume) |
OSHA-PEL Level |
N † |
Mean F/T |
Std Dev |
CV |
OE (%) |
0.5 × |
6 |
1.0190 |
0.0160 |
0.0157 |
± 5.04 |
1 × |
6 |
1.0248 |
0.0139 |
0.0135 |
± 5.18 |
2 × |
6 |
0.9935 |
0.0744 |
0.0748 |
± 15.61 ‡ |
Summary |
18 |
1.0124 |
NA |
0.0448 |
± 10.2 |
F/T =
Found/Theoretical
OE = Overall error (± %)
* Composition of certified materials: 87.87 for W in 94%
WC 84.08 for W in 88% W2C
† At each exposure level, three 94% WC and three 88%
W2C spiked samples were analyzed.
‡ One of the 94% WC samples at 2 × PEL gave a recovery of 0.854 and
was not deleted in the statistical analysis. Reanalysis of this sample
also gave a low recovery (F/T = 0.841). This was not an outlier by the
Grubb's test (99% confidence level). Nothing unusual was observed in
the analysis of the Co content of this sample. Variation was probably
due to the inhomogeneity of the WC/Co composite.
Table 18
Analysis - Insoluble W
STEL Determination Pure Tungsten Matrix * (2 L/min, 15
min, 30 L, 50-mL solution volume) |
OSHA-STEL Level |
N |
Mean FT |
Std Dev |
CV |
OE (%) |
1 × |
6 |
1.0467 |
0.0212 |
0.0202 |
± 8.71 |
F/T = Found/Theoretical
OE
= Overall error (± %) * The composition indicated on
reagent certification is 100.00% W.
4.6. Determination of Analytical Method Precision and Accuracy for
Soluble W:
Cobalt is not commonly present when sampling for soluble W, so it
was not included in this experiment.
Procedure
Analyte spikes of W were prepared from reagents independent of
calibration standards. Spikes of soluble W were made by delivery of
solutions of known concentration using pipettes and micropipettes.
The 0.8-µm pore size, 37-mm diameter, MCE
filters were placed in Phillips beakers and spiked at levels
representing 0.5 ×, 1 ×, and 2 × TWA PEL using respectively 240, 480,
and 960 µL aliquots of a 1,000 µg/mL W solution (178.72
mg of 100.4%
Na2WO4.2H2O
diluted to 100 mL with DI H2O). The MCE
filters spiked at the STEL used 900 µL aliquots of a 100
µg/mL W solution prepared by a 1:10 dilution of the 1,000
µg/mL W solution. The gravimetric factor of 0.5596 was
calculated from (100.4%/100%) × (183.85 g W)/(329.86 g
Na2WO4.2H2O)
based on reagent label analysis and formula weights. The samples were
air dried prior to digestion. The samples were then diluted to 50 mL,
and analyzed according to the method.
Results
The analytical method precision and accuracy data for soluble W are
shown in Tables 19 and 20:
Table 19
Analysis - Soluble W TWA
Determination Sodium Tungstate Dihydrate Matrix (2 L/min,
4 h, 480 L, 50-mL solution volume) |
OSHA-PEL Level |
N |
Mean F/T |
Std Dev |
CV |
OE (%) |
0.5 × |
6 |
1.0892 |
0.0100 |
0.0092 |
± 10.76 |
1 × |
6 |
1.0652 |
0.0087 |
0.0082 |
± 8.16 |
2 × |
6 |
1.0640 |
0.0121 |
0.0114 |
± 8.68 |
Summary |
18 |
1.0728 |
NA |
0.0097 |
± 9.22 |
F/T = Found/Theoretical
OE
= Overall error (± %)
Table 20
Analysis - Soluble W STEL
Determination Sodium Tungstate Dihydrate Matrix (2 L/min,
15 min, 30 L, 50-mL solution volume) |
OSHA-STEL Level |
N |
Mean F/T |
Std Dev |
CV |
OE (%) |
1 × |
6 |
1.0615 |
0.0074 |
0.0070 |
± 7.55 |
F/T = Found/Theoretical
OE
= Overall error (± %)
4.7. Determination of Analytical Method Precision and Accuracy for
Co:
Procedure
The 0.8-µm pore size, 37-mm diameter, MCE
filters for this experiment were placed in Phillips beakers and spiked
at levels representing Co exposures at 0.5 ×, 1 ×, and 2 × TWA PEL
using 120-, 240-, and 480-µL aliquots
of a 100 µg/mL Co solution respectively. To match expected
matrix conditions where spectral interferences from W may be
significant, these samples were also spiked with 0.5-mL aliquots of
10,000 µg/mL W (1.79 g of 100.4%
Na2WO4.2H2O
diluted to 100 mL with DI H2O). The W
reagent and calibration standards were independent sources of W. The
spiked filters in the Phillips beakers were then air dried prior to
digestion. The samples were digested, diluted to 50 mL, and analyzed
according to the method.
Note: The 100 µg/mL Co spiking solution was a 1:10 dilution
made from a different bottle of atomic spectral standard than was used
to make the two Co calibration standards. Because both bottles had the
same lot number, this experiment did not provide a completely
independent measure of Co recovery but did provide an adequate measure
of precision. When new ICP calibration standards are prepared at SLTC,
a quality assurance check is made by analyzing and comparing the old
and new calibration standards. This Co ICP stock standard has checked
out fine on this quality assurance test and on subsequent quality
control tests using independent Co sources. A fully independent test
of Co recovery is provided by the additional Co recovery experiments
performed in Section 4.8.
After preparation, the resulting final solution matrix was 100
µg/mL W and would represent exposures to soluble W at 10 × TWA
PEL or exposures to insoluble W at 2 × TWA PEL.
Results
Analytical method data and precision and accuracy for Co are shown
below in Table 21. The analytical method data, precision, and accuracy
for 100 µg/mL W matrix is provided below Table 21:
Table 21
Analysis - Co TWA
Determination (In 100 µg/mL W Solution Matrix) * (2
L/min, 480 L, 50-mL solution volume) |
OSHA-PEL Level |
N |
Mean F/T |
Std Dev |
CV |
OE (%) |
0.5 × |
6 |
1.0627 |
0.0484 |
0.0456 |
± 15.39 † |
1 × |
6 |
1.0440 |
0.0014 |
0.0014 |
± 4.68 |
2 × |
6 |
1.0328 |
0.0063 |
0.0061 |
± 4.50 |
Summary |
18 |
1.0465 |
NA |
0.0266 |
± 9.97 |
F/T = Found/Theoretical
OE
= Overall error (± %)
* |
Analysis for W (100 µg/mL)
matrix: |
|
Bias |
= |
+ 0.0411 |
|
CV (Pooled) |
= |
0.0079 |
|
Overall Error (Total) |
= |
± 5.69% |
† One of the samples at 0.5 × PEL was overlooked on the day all
the other samples were analyzed; it was analyzed two days later at
the first opportunity. The result was not an outlier according to
Grubb's test (99% confidence level).
4.8. Determination of Analytical Method Precision and Accuracy for
Co at Higher Levels:
The spiked MCE filter samples of the cemented tungsten carbide
matrices prepared in Section 4.5. also contained substantial amounts
of Co (3.2 ×, 6.5 ×, 7.4 ×, 12.3 ×, 14.9 ×, and 28.0 × TWA PEL). These
samples were also analyzed for Co. This supplementary experiment
serves two purposes:
1) To provide an independent test of Co recovery not provided in
Section 4.7.
2) To extend the validation to include levels of Co potentially
found in sampling for W in cemented tungsten carbide operations.
Procedure
As described in Section 4.5., analyte spikes were prepared from
certified sources independent of calibration standards. The amounts of
W and Co in all insoluble analyte spikes were calculated from assay
information provided by the chemical manufacturer. The weights of the
spikes of the cemented tungsten carbide matrices were determined by
weighing on tared 0.8-µm pore size, 37-mm
diameter, MCE filters that had reached constant weight in a controlled
low humidity lab environment (33 - 36% RH). The spikes filters were
transferred to Phillips beakers, digested, diluted to 50 mL, and
analyzed according to the method.
Results
This experiment gave an average Co recovery of 1.028 which compares
well to the average Co recovery of 1.046 obtained in Section 4.7. The
Co results are listed in Table 22:
Table 22
Analysis - Co TWA
Determinations Above Usual Validation Range (In Insoluble
Tungsten Carbide Samples at 0.5, 1, and 2 × W PEL)* (2 L/min,
4 h, 480 L, 50-mL solution volume) |
OSHA-PEL Level † |
Matrix |
Mean F/T |
Std Dev |
CV |
OE (%) |
3.2 × |
Co in 94% WC |
1.0653 |
0.0190 |
0.0179 |
± 10.11 |
6.5 × |
Co in 94% WC |
1.0663 |
0.0083 |
0.0078 |
± 8.19 |
7.4 × |
Co in 88%
W2C |
1.0067 |
0.0099 |
0.0098 |
± 2.63 |
12.3 × |
Co in 94% WC |
1.0330 |
0.0380 |
0.0368 |
± 10.66 |
14.9 × |
Co in 88%
W2C |
0.9997 |
0.0101 |
0.0101 |
± 2.05 |
28.0 × |
Co in 88%
W2C |
1.0000 |
0.0056 |
0.0056 |
± 1.12 |
Summary |
18 samples (Pooled) |
1.0285 |
NA |
0.0181 (df=12) |
± 6.47 |
18 samples (Combined) |
1.0285 |
NA |
0.0325 (df=17) |
± 9.35 |
F/T = Found/Theoretical Overall
error (± %) df = degrees of freedom
* Composition of the certified materials: 5.45% Co in 94% WC,
11.70% Co in 88% W2C † N = 3 samples for
each OSHA-PEL Level tested.
4.9. Assessment of Typical Levels of Blank Contamination, Analyte
Carry-over, and Noise:
Typical blank contamination, carry-over, and noise
between samples was surveyed by analyzing blanks interspersed with
samples.
Procedure
Four blanks were prepared from the same lot of 0.8-µm pore
size, 37-mm diameter, MCE filters used to prepare the
spiked samples. These were subjected to the same digestion steps as
the spiked samples and were analyzed along with the spiked samples
described in Sections 4.5. to 4.8. All four blanks were used to
evaluate contaminant levels in the sampling medium. Some of these were
also analyzed after high samples and standards in order to evaluate
carry-over. Carry-over can result from physical
carry-over of solution between analyses. Additionally,
carry-over of electronic origin can result after PMTs or
other sensitive electronic components are stressed by strong signals;
it may take additional time for certain channels to settle down to the
same noise level measured during calibration, and the resulting
excursions in electronic noise can be incorrectly interpreted as
signals.
All results were calculated from
concentration-response curves, corrected for reagent
blank, and corrected for the interelement interferences identified in
Section 4.3.
Results
Table 23 shows the blank contamination levels (µg/mL)
observed:
Table 23
Element |
N, µg/mL* |
Element |
N, µg/mL |
Element |
N, µg/mL |
Element |
N, µg/mL |
Al |
2, 0.12 |
Cr |
1, 0.020 |
Ni |
ND |
Te |
1, 0.080 |
As |
ND |
Cu |
ND |
Pb |
ND |
Ti |
ND |
Be |
ND |
Fe |
ND |
Sb |
ND |
V |
ND |
Ca |
4, 0.018 |
Mg |
2, 0.11 |
Se |
2, 0.064 |
W |
2, 0.070 |
Cd |
ND |
Mn |
ND |
Si |
4, 3.3 |
Zn |
ND |
Co |
4, 0.015 |
Mo |
ND
| Sn |
1, 0.020 |
Zr |
2,
0.020 |
* The levels of contamination are reported in the following format:
number detected, mean of detected. ND indicates all four blanks were
non-detected. Blank analyses above excluded replicate
checks used to test for carry-over.
Carry-over and noise were observed during the validation. The worst
instances are listed in Table 24:
Table 24
Preceding Sample Element(s) |
Apparent Carry-Over of
Element |
2 µg/mL Cr,Mo,Ni,Sb,V 20
µg/mL Fe * |
0.075 µg/mL As |
" |
0.020 µg/mL Co |
" |
0.13 µg/mL Mg |
" |
0.079 µg/mL Sb |
" |
0.21 µg/mL Se |
" |
0.080 µg/mL Sn |
" |
0.12 µg/mL Te |
99.26 µg/mL W, 56
µg/mL Co |
0.18 µg/mL
W |
* Carry-over, noise, or interferences on As, Co, Se,
Sn, Te, and W were also observed when this standard was analyzed. This
standard was preceded by six 2 × PEL Co samples (0.96 µg/mL Co
containing 100 µg/mL W). Undercorrected As, Mg, Se, Sn, and Te
interferences were also noted during the analyses of several of these
2 × PEL Co samples.
The long delay in re-achieving normal background noise
levels indicates that the type of carry-over noise
observed in these experiments probably arises from electronic origin.
The potential for background contamination and carry-over
demonstrates the need for blank analyses. This will correct for
sampling media contamination and provide a quality assurance check for
possible carry-over effects, faltering electronic
circuits, or noisy channels. Carry-over can be diminished
by lengthening the amount of time between analysis of each sample or
standard. When using the parameters described in this method, the time
between samples was approximately 85 seconds to change sample and 30
seconds pre-integration time to allow sample entry into
the plasma.
4.10. Summary:
The validation results indicate the method meets both the NIOSH and
OSHA criteria for accuracy and precision (5.5., 5.17.). The results
are within an acceptable range (OE < ± 25%). Detection limits are
adequate when samples are taken for TWA determinations for 480 min at
2 L/min, or for 15-min STEL determinations. The method is
adequate for monitoring TWA, STEL, and indoor air types of exposures.
5. References
5.1. National Institute for Occupational Safety and Health:
Criteria for a recommended standard .... occupational exposure to
Tungsten and Cemented Tungsten Carbide (DHEW/NIOSH Pub. No.
77-127). Cincinnati, OH: National Institute for
Occupational Safety and Health, 1977. pp. 4, 28, 48,
94-96, 102, 130, 146-149, 166-170.
5.2. American Conference of Governmental Industrial Hygienists
Inc.: Documentation of the Threshold Limit Values and
Biological Exposure Indices; 5th ed. Cincinnati, OH: ACGIH, 1986.
pp. 144, 614.
5.3. National Institute for Occupational Safety and Health:
NIOSH Manual of Analytical Methods by D. Taylor (DHEW/NIOSH
Pub. No. 78-175). Cincinnati, OH: National Institute for
Occupational Safety and Health, 1978. pp. 271-1 -
271-8.
5.4. Occupational Safety and Health Administration Analytical
Laboratory: Insoluble Tungsten and Compounds (WC). Special
Project by P. Giles (In house report). Salt Lake City, UT: USDOL/OSHA
Salt Lake Technical Center, November, 1989.
5.5. Occupational Safety and Health Administration Analytical
Laboratory: Insoluble and Soluble Tungsten in Workplace
Atmospheres by XRF by M. C. Rose (In house report). Salt Lake
City, UT: USDOL/OSHA Salt Lake Technical Center, June, 1988 (revised
February, 1993).
5.6. Perkin-Elmer: Analysis of Tungsten Carbide,
ID-3, In Analytical Methods for Atomic Absorption
Spectrophotometry, Norwalk, CT: Perkin-Elmer, March
1971.
5.7. Occupational Safety and Health Administration Analytical
Laboratory: ICP Backup Data Report (ARL 3560) for Metal and
Metalloid Particulates in Workplace Atmospheres (ICP Analysis),
(USDOL/OSHA Method No. ID-125G). Salt Lake City, UT:
USDOL/OSHA Salt Lake Technical Center, February, 1992 (revised
December, 1992).
5.8. Mandel, J.: Accuracy and Precision, Evaluation and
Interpretation of Analytical Results, The Treatment of Outliers. In
Treatise On Analytical Chemistry, 2nd ed., Vol.1, edited by I.
M. Kolthoff and P. J. Elving. New York: John Wiley and Sons, 1978. pp.
282-285.
5.9. Weast, R.C., ed.: Handbook of Chemistry and
Physics. 60th ed. Boca Raton, FL: Chemical Rubber Company Press,
1979. pp B138 - B139.
5.10. Public Health Service: Toxicological Profile for
Cobalt; (U.S., Department of Health & Human Services, Public
Health Service, Agency for Toxic Substances and Disease Registry
publication ATSDR/TP-91/10), 1992.
5.11. Jarvis, J.Q., E. Hammond, R. Meier, and C. Robinson:
Cobalt Cardiomyopathy. Journal of Occupational Medicine.
34:620-626 (1992).
5.12. Occupational Safety and Health Administration Salt Lake
Technical Center: ARL ICP Analysis Reference Guide - OSHA LAB
(Comprehensive Version) by J. Septon. Salt Lake City, UT:
USDOL/OSHA Salt Lake Technical Center, November, 1991 (unpublished).
5.13. Occupational Safety and Health Administration Salt Lake
Technical Center: ARL ICP Tungsten + Cobalt Standard Operating
Procedure by M.C. Rose. Salt Lake City, UT: USDOL/OSHA Salt Lake
Technical Center (in progress), 1993 (unpublished).
5.14. National Institute for Occupational Safety and Health:
Documentation of the NIOSH Validation Tests by D. Taylor, R.
Kupel, and J. Bryant (DHEW/NIOSH Pub. No. 77-185).
Cincinnati, OH: National Institute for Occupational Safety and Health,
1977. pp. 1-12.
5.15. Occupational Safety and Health Administration Analytical
Laboratory: Precision and Accuracy Data Protocol for Laboratory
Validations. In OSHA Analytical Methods Manual 1st ed. (Pub. No.
ISBN: 0-936712-66-X). Cincinnati, OH: American
Conference of Governmental Industrial Hygienists, 1985.
5.16. Long, G.L. and J.D. Winefordner: Limit of Detection -
A Closer Look at the IUPAC Definition. Anal. Chem. 55:712A-724A
(1983).
5.17. Occupational Safety and Health Administration Analytical
Laboratory: OSHA Analytical Methods Manual. Vol III
(USDOL/OSHA-SLTC Method Evaluation Guidelines), (Pub. No. ISBN:
0-936712-66-X). Cincinnati, OH: American Conference of
Governmental Industrial Hygienists, 1985.
|