CARBON DIOXIDE BACKUP DATA REPORT (ID-172)
This report was revised June, 1990
Introduction
The evaluation of OSHA Method No. ID-172, Carbon Dioxide in Workplace
Atmospheres (9.1.), was conducted when the time weighted average (TWA)
Permissible Exposure Limit (PEL) for carbon dioxide (CO2) was
5000 ppm (1985). The PEL has been changed to 10,000 ppm TWA and a 30,000
ppm Short-Term Exposure Limit (STEL) has been added. Any mention of PEL in
this report, unless specified otherwise, is in reference to the
Transitional limit of 5,000 ppm. (Note: Some of the data in this report is
presented as %CO2. To convert % to ppm, ppm CO2 =
%CO2 x 10,000)
1. Experimental Protocol
1. The validation consists of the following experimental protocol:
2. Analysis of 3 sets of 6 spiked samples having concentration
ranges of approximately 0.5, 1, and 2 times the PEL.
3. Analysis of 3 sets of 6 dynamically generated samples having
concentration ranges of approximately 0.5, 1, and 2 times the PEL.
4. Determination of the qualitative and quantitative detection
limit for analysis of CO2 by gas chromatography.
5. Determination of any variation in results when sampling at high
and low humidity levels.
6. Comparison of other methods used for CO2 workplace
determinations with the gas chromatographic method.
7. Determination of the storage stability of CO2 samples
collected in gas sampling bags.
8. Assessment of the performance of the gas chromatographic method
and conclusions.
Data for certain experiments were statistically examined for outliers
and homogeneous variance. Possible outliers were determined using the
American Society for Testing and Materials (ASTM) test for outliers
(9.2.). Homogeneity of the coefficients of variation was determined
using the Bartlett's test (9.3.).
2. Analysis
Procedure: Three sets of spiked samples were prepared and
analyzed as follows:
2.1. Gas sampling bags were flushed with CO2-free
compressed air [ambient CO2 and other potential
contaminants were removed from the compressed air by using an air
scrubber/filtration system (Figure 1) consisting mainly of a Ascarite
II/Drierite bed]. A vacuum was then applied to completely evacuate the
bags.
2.2. A known amount of CO2-free air was metered into
each sampling bag. Compressed air flow rates were measured immediately
before and after each experiment using a soap bubble flowmeter. Air
flow was regulated by using a regulator-rotameter system as shown in
Figure 1. Blank samples of the compressed air were periodically
collected and analyzed along with the samples and standards.
2.3. A known amount of CO2 was metered into each
sampling bag containing diluent air. Carbon dioxide flow rates for the
spiked samples were determined immediately before and after each
experiment using a soap bubble flowmeter. A gas cylinder containing
1.93% CO2 in air (Air Products, Long Beach, CA, certified
analytical standard) was used for spiking the samples.
2.4. Reference and analytical standards were analyzed along with
the spiked samples. The reference standard was purchased commerically
(Scott Specialty Gases, Houston, TX, 0.9968% CO2 in
nitrogen, methane, carbon monoxide, oxygen, and hydrogen, certified
analytical standard) and analytical standards were generated by
dilution of Bone Dry grade CO2 (Union Carbide, 99.8% min.
purity). Carbon dioxide flow for analytical standards was regulated
using a regulator-mass flow controller system as shown in Figure 2.
2.5. Spiked samples, blanks, reference, and generated standards
were analyzed by gas chromatography (9.1.). Samples were analyzed
within 2 days of preparation. Analytical instrument parameters are
displayed in Appendix 1.
Results: Spiked sample results were calculated using a
linear regression concentration-response curve. Integrated peak areas
were used as signal measurement. Spiked sample recoveries are listed
in Table 1. All Analysis data passed both the outlier and Bartlett's
tests. The data (Table 1) indicates good precision and accuracy. The
coefficient of variation for analysis (CV1) was 0.034 and
the average analytical or spiked recovery was 95.2%.
3. Sampling and Analysis
Procedure: Three sets of generated samples were prepared and
analyzed by:
3.1. Gas sampling bags were flushed with CO2-free
compressed air. Ambient CO2 and other potential
contaminants were removed from the compressed air by using an air
filtration system similar to the one shown in Figure 1. A vacuum was
then applied to completely evacuate the bags.
3.2. A dynamic gas generation system was assembled as shown in
Figure 2. A humidity, temperature, and flow control system
(Miller-Nelson Research Model 301) was used to control and monitor air
flow. A dry test meter (Singer Co., Model # DTM 115) was to measure
air flow immediately before, during, and after the experiments. The
flow control system was calibrated in-house temperature, humidity, and
flow prior to use. Calibration of the dry test meter was done using a
spirometer as a primary standard.
3.3. Carbon dioxide (Bone Dry grade, 99.8% min. purity) gas was
introduced into the flow system via a mixing chamber as shown in
Figure 2. Carbon dioxide flow rates were taken immediately before and
after each experiment using a soap bubble flowmeter. Flow rates were
controlled using mass flow controllers (Tylan Model FC260 mass flow
controller).
3.4. Generated samples, blanks, reference, and spiked standards
were analyzed by gas chromatography. Reference standards were
determined by direct injection of the gas from the canister into the
gas sampling valve of the gas chromatograph. Analytical standards were
prepared in gas sampling bags and then injected. Samples were analyzed
within 2 days of preparation. Instrument parameters used during the
analysis are displayed in Appendix 1.
Results: Generated sample results were calculated using a
linear regression concentration response curve. Integrated peak areas
were used as signal measurement. Generated sample recoveries are listed
in Table 2. As shown in Table 2, the gas chromatographic determinations
of CO2 using gas sampling bags are within NIOSH accuracy and
precision guidelines (9.3.). The CVT was 0.026 and the
overall recovery was 99.5%. The Sampling and Analysis data shows
excellent precision and accuracy. All data passed both the outlier and
Bartlett's test.
4. Detection Limit
Procedure: The qualitative detection limit for the analysis of
CO2 by gas chromatography was calculated using the Rank Sum
Test (9.4.). The International Union of Pure and Applied Chemistry
(IUPAC) method for detection limit determinations was used to determine
the quantitative limit (9.5.). The procedure used for sample preparation
for determining the detection limit is shown below:
4.1. Same as Section 2.1.
4.2. Blank samples were generated using the flow, humidity, and
temperature control system mentioned in Section 3.2.
4.3. Low concentration CO2 standards were prepared by
mixing CO2 (1.93% in air), via a mixing chamber, with the
treated air. Concentrations of 205.1, 398.9, and 662 ppm were used as
standards.
4.4. Samples were then analyzed by gas chromatography. Analytical
conditions used are given in Appendix 1.
Results: Qualitative and quantitative detection limits are
listed in Tables 3 and 4, respectively. The qualitative limit is 200
ppm. The quantitative limit is 500 ppm. A 1-mL sampling loop was used
for all analyses. Lower detection limits for CO2 are
possible with larger sampling loops, but should not be necessary for
workplace determinations. This assumption is based on the fact that
ambient air will always contain a certain amount of CO2. In
well ventilated areas, the level of CO2 is normally in the
range of 300 to 700 ppm.
5. Humidity Study
Procedure: Samples were generated at high (80%), medium (50%),
and low (25%) relative humidities using the same equipment and
conditions described in Section 3. Samples were taken side-by-side with
detector tube samples (9.6.).
Results: Gas sampling bag results at 80 and 25% RH are
presented in Table 5. Table 2 contains the 50% RH test. Data from
sampling at different humidities displayed no apparent effect on
collection efficiency. As shown in Table 5, an analysis of variance (F
test) was performed on the data to determine any significant difference
among or within the different humidity groups. Variance at each
concentration level (0.5, 1, and 2 times the PEL) was compared across
the 3 humidity levels (25, 50, and 80% RH). The variance among and
within the different concentration groups gave acceptable calculated F
values with the exception of the test conducted at the PEL. Recovery at
each humidity level was also considered. As also shown in Table 5, no
evidence of any constant increase or decrease in average recovery is
apparent across the humidity levels. The large calculated F value at the
PEL was judged to be due to variation in sample generation and analysis
and not to a humidity effect.
6. Comparison Methods
6.1. Detector tubes (in-house study)
A side-by-side (in-house) determination of CO2 was
performed using different types of short-term CO2 detector
tubes and simultaneous gas sampling bag-gas chromatography analysis.
Detector tubes were chosen since they were listed as the OSHA sampling
method for CO2 (9.8.). As mentioned in the Introduction of
the CO2 method (9.1.), an alternative titration method
(9.7.) was considered unsuitable to use for comparison at the
generated concentration levels. Gas chromatographic and detector tube
samples were taken at different humidity levels. A synopsis of the
side-by-side testing is shown in Table 6. The overall recovery and CV
for the gas sampling method displays an improvement over the detector
tube technique. Further information regarding the short-term detector
tube evaluation can be found in reference 9.6.
6.2. "Numbering error. This section contains no data."
6.3. A preliminary evaluation of long-term detector tubes was also
performed (9.9.). The Draeger model no. 6728611 long-term detector
tube and the Mine Safety Appliance Vaporgard Dosimeter were examined.
Preliminary testing revealed the Draeger tube unsatisfactory for
CO2 compliance determinations. Only 10 MSA dosimeters were
tested and all were from the first lot of production. Dosimeter
results were satisfactory; however, further testing and assessment of
lot-to-lot variability are necessary.
6.4. Detector tubes (field study)
A side by side field sampling evaluation was also performed. The
sampling was done at a food freezing plant which used liquid
CO2 as the refrigerant. Gas bag samples and detector tubes
were taken in various areas around the plant. Detector tubes were
taken at random times and in close vicinity to the sampling bags. A
log normal distribution was applied to the data to determine TWAs (for
a further discussion of grab samples used to determine TWAs, see
references 9.10. and 9.11.). Detector tube readings were also taken
directly from the personal gas bag samples during the gas
chromatographic analysis. Results of the field testing are also listed
in Table 6.
6.5. Miran 1A
The Miran 1A infrared gas analyzer was also assessed for possible
use in CO2 determinations. As a direct reading instrument,
the Miran 1A appeared too sensitive to assess large CO2
levels sometimes found in industrial settings. An off-scale reading
was given when CO2 concentrations were above 5,000 ppm.
However, this response characteristic of the MIRAN appears to make it
useful for indoor air quality investigations because CO2
levels less than 5,000 ppm are normally used to determine ventilation
system performance. Carbon dioxide is also used in air quality
assessments as a tracer gas to monitor ventilation efficiency.
An attempt was also made to use the 5.4-L sampling cell of the gas
analyzer as a closed-system analyzer. Samples were collected in gas
sampling bags and aliquots were taken from the bags using gas-tight
syringes. These aliquots were then injected into the closed system. It
was necessary to take 50 to 100-mL aliquots to achieve an adequate
signal for CO2 measurement at the generated levels. The
aliquots were considered very large and made accurate and precise
analysis difficult.
7. Stability Test
Procedure: A long-term evaluation of sample media stability
was performed to determine any potential problems if delays in sample
analyses occur. Five layered, 5-L sampling bags containing generated
samples, field samples, and reference standards were used to assess
CO2 storage stability. Samples were analyzed at various
times, up to 50 days, after sample collection. A few samples were stored
in a refrigerator at 5 °C and analyzed periodically over 31 days.
Recovery data are listed in Table 7 and graphically represented in
Figure 3.
Results:
Note: Previously, different types of gas sampling bags were evaluated
for stability, structural integrity, and compactness. Tedlar sampling
bags can be used for standard dilution, provided the standards are
analyzed within a 24-h period. A significant loss of CO2 was
noted if Tedlar bag standards were analyzed during longer periods.
The storage ability of the five-layered, 5-L gas sampling bag for
CO2 is unacceptable if stored for a period longer than 14
days. Table 7 individually lists each sample/result. A summary of the
stability data per time period is listed below:
Total Samples |
Day |
Ave %
Recovery |
10 |
1-5 |
95.6 |
8 |
9-18 |
89.7 |
8 |
26-31 |
77.2 |
7 |
>31 |
68.5 |
Figure 3 graphically depicts the recovery dropping below 90% after 14
days of storage. Sample refrigeration appears to slightly retard
CO2 loss; however, gas bag samples should be sent to the
laboratory and analyzed as soon as possible.
8. Method Performance - Conclusions
The data generated during the validation of the method indicate an
acceptable alternative for sampling and analyzing CO2. The
gas chromatographic method offers an accurate and precise assessment of
CO2 exposures in the workplace. Although no samples were
taken at concentration levels at the 30,000 ppm STEL, the storage
stability data at about 20,000 ppm indicates the stability, precision,
and accuracy were similar to validation range (2,000 to 10,000 ppm)
results. Standards prepared at 30,000 ppm to determine the linear
working range displayed excellent linearity with lower concentration
standards used to construct the concentration-response curve.
This method should be capable of accurate and precise measurements to
determine compliance with the 10,000 ppm PEL and also the 30,000 ppm
STEL.
9. References
9.1. Occupational Safety and Health Administration Technical
Center: Carbon Dioxide in Workplace Atmospheres (OSHA-SLTC
Method No. ID-172). Salt Lake City, UT. Revised 1990.
9.2. Mandel, J.: Accuracy and Precision, Evaluation and
Interpretation of Analytical Results, The Treatment of Outliers. In
Treatise On Analytical Chemistry, 2nd ed., Vol. 1, edited by
I.M. Kolthoff and P.J. Elving. New York, NY: John Wiley and Sons,
1978. pp. 282-285.
9.3. National Institute for Occupational Safety and Health:
Documentation of the NIOSH Validation Tests by D. Taylor, R.
Kupel and J. Bryant (DHEW/NIOSH Pub. No. 77-185). Cincinnati, OH:
National Institute for Occupational Safety and Health, 1977. pp. 1-12.
9.4. National Bureau of Standards: Experimental
Statistics by M.G. Natrella (NBS Handbook 91). Washington, DC:
U.S. Department of Commerce, National Bureau of Standards, 1966.
Chapter 16, pp. 8-14.
9.5. Long, G.L. and J.D. Winefordner: Limit of Detection --
A Closer Look at the IUPAC Definition. Anal. Chem. 55:
712A-724A (1983).
9.6. Occupational Safety and Health Administration Analytical
Laboratory: Carbon Dioxide Detector Tubes (PE-2). Salt Lake
City, UT. 1987.
9.7. Norton, J. F., ed.: Standard Methods for the
Examination of Water and Sewage. 9th ed. New York, NY: American
Public Health Association, 1946. pp. 33-40.
9.8. U.S. Department of Labor-Occupational Safety and Health
Administration: Chemical Information File. Online Database
-- OSHA Information System. Washington, DC: Directorate of Technical
Support, U.S. Dept. of Labor, OSHA, 1985.
9.9. Occupational Safety and Health Administration Analytical
Laboratory: Carbon Dioxide Detector Tubes -- Long Term
(PE-3). Salt Lake City, UT. 1987.
9.10. National Institute for Occupational Safety and Health:
Occupational Exposure Sampling Strategy Manual by N.A. Leidel,
K.A. Busch and J.R. Lynch (DHEW/NIOSH Pub. No. 77-173). Washington,
DC: Government Printing Office, 1977.
9.11. Leichnitz, K.: Detector Tube Measuring Techniques.
Ecomed, Federal Republic of Germany, 1983. pp. 251-264.
Table 1
Analysis
|
Level |
%CO2 Taken |
%CO2 Found |
F/T |
n |
Mean |
Std Dev |
CV1 |
|
|
0.251 |
0.248 |
0.986 |
|
|
0.251 |
0.234 |
0.931 |
|
0.5 X PEL |
0.251 |
0.236 |
0.939 |
|
|
0.251 |
0.236 |
0.939 |
|
|
0.251 |
0.239 |
0.951 |
|
|
0.251 |
0.236 |
0.939 |
|
|
6 |
0.947 |
0.020 |
0.021 |
|
|
0.497 |
0.505 |
1.016 |
|
|
0.497 |
0.470 |
0.946 |
|
1 X PEL |
0.497 |
0.467 |
0.940 |
|
|
0.497 |
0.452 |
0.909 |
|
|
0.497 |
0.477 |
0.960 |
|
|
0.497 |
0.452 |
0.909 |
|
|
6 |
0.947 |
0.040 |
0.042 |
|
|
1.007 |
1.008 |
1.001 |
|
|
1.007 |
0.964 |
0.957 |
|
2 X PEL |
1.007 |
0.995 |
0.988 |
|
|
1.007 |
0.984 |
0.977 |
|
|
1.007 |
0.953 |
0.946 |
|
|
1.007 |
0.911 |
0.904 |
|
|
6 |
0.962 |
0.035 |
0.036 |
F/T = Found/Taken
CV1(Pooled) = 0.034
Average Recovery = 0.952
Table 2
Sampling and Analysis (50% RH, 25 °C)
|
Level |
%CO2 Taken |
%CO2 Found |
F/T |
Airvol |
n |
Mean |
Std Dev |
CV2 |
|
|
0.270 |
0.270 |
0.998 |
1.7 |
|
|
0.270 |
0.266 |
0.983 |
2.1 |
|
0.5 X PEL |
0.270 |
0.269 |
0.994 |
2.7 |
|
|
0.270 |
0.275 |
1.016 |
3.5 |
|
|
0.270 |
0.274 |
1.013 |
3.9 |
|
|
0.270 |
0.269 |
0.994 |
2.5 |
|
|
6 |
1.000 |
0.013 |
0.013 |
|
|
0.544 |
0.563 |
1.036 |
3.6 |
|
|
0.544 |
0.560 |
1.030 |
3.4 |
|
1 X PEL |
0.544 |
0.565 |
1.039 |
3.1 |
|
|
0.544 |
0.560 |
1.030 |
1.7 |
|
|
0.544 |
0.546 |
1.004 |
2.1 |
|
|
0.544 |
0.556 |
1.023 |
4.1 |
|
|
6 |
1.027 |
0.013 |
0.012 |
|
|
1.005 |
0.981 |
0.976 |
3.0 |
|
|
1.005 |
0.935 |
0.930 |
3.1 |
|
2 X PEL |
1.005 |
0.975 |
0.970 |
1.6 |
|
|
1.005 |
0.969 |
0.964 |
3.2 |
|
|
1.005 |
0.961 |
0.956 |
3.2 |
|
|
1.005 |
0.952 |
0.947 |
4.1 |
|
|
6 |
0.957 |
0.017 |
0.018 |
Airvol |
= |
Air volume taken (L) |
F/T |
= |
Found/Taken |
CV2(Pooled)/FONT> |
= |
0.014 |
CVT |
= |
0.026 |
Bias |
= |
-0.005 |
Overall Error |
= |
±5.7% |
Table 3
Determination of Qualitative Detection Limit
ppm |
|
Integrated Area |
|
|
|
BLANK |
|
0, 0, 0, 0, 0, 0, 0, 0, 130, 275 |
205.1 |
|
221, 242, 314, 324, 426, 504, 548, 563 |
398.9 |
|
967, 1041, 1075, 1249, 1267, 1268, 1334, 1421 |
662 |
|
2090, 2107, 2335, 2561 |
Rank Sum Data
a |
= |
0.01 (two-tailed test) |
n1 |
= |
8 (# of 205.1 ppm determinations) |
n2 |
= |
10 (# of blank determinations) |
n |
= |
n1 + n2 = 18 |
R |
= |
114 (sum of ranks for 205.1 ppm) |
Rn |
= |
n1(n+1) - R = 38 |
R(table) |
= |
47 |
Therefore, Rn is not equal to or greater than
R(table), and both sample populations are significantly
different.
Qualitative detection limit = 205.1 ppm
Table 4
Determination of Quantitative Detection Limit
IUPAC Method
Using the equation: |
Cld =
k(sd)/m |
Where:
Cld |
= |
the smallest detectable concentration an
analytical instrument can determine at a given confidence
level. |
|
k |
= |
3, thus giving 99.86% confidence that any detectable
signal will be greater than or equal to an average blank reading
plus three times the standard deviation (area reading >
Blave + 3sd). |
|
sd |
= |
standard deviation of blank readings. |
|
m |
= |
analytical sensitivity or slope as calculated by
linear regression. |
Minimum detectable signal:
|
Cld = 3(91.97/1.771) |
|
Cld = 156 ppm |
For k = 10 (Quantitative detection limit, 99.9% Confidence):
|
Cld = 519 ppm as a reliable
detectable signal |
Table 5
Humidity Tests
25% RH (25 °C)
|
Level |
%CO2 Taken |
%CO2 Found |
F/T |
n |
Mean |
Std Dev |
CV |
|
|
0.266 |
0.293 |
1.100 |
|
|
0.266 |
0.277 |
1.040 |
|
0.5 X PEL |
0.266 |
0.266 |
1.000 |
|
|
0.266 |
0.282 |
1.060 |
|
|
0.266 |
0.282 |
1.060 |
|
|
5 |
1.052 |
0.036 |
0.035 |
|
|
0.538 |
0.483 |
0.898 |
|
|
0.538 |
0.506 |
0.941 |
|
1 X PEL |
0.538 |
0.509 |
0.947 |
|
|
0.538 |
0.499 |
0.929 |
|
|
4 |
0.929 |
0.022 |
0.023 |
|
|
1.002 |
1.001 |
0.999 |
|
|
1.002 |
1.012 |
1.010 |
|
2 X PEL |
1.002 |
0.957 |
0.955 |
|
|
1.002 |
0.982 |
0.980 |
|
|
1.002 |
0.995 |
0.993 |
|
|
5 |
0.987 |
0.020 |
0.021 |
F/T = Found/Taken
CV(Pooled) = 0.027
Average Recovery = 0.989
Table 5 (Cont.)
Humidity Tests
80% RH (25 °C)
|
Level |
%CO2 Taken |
%CO2 Found |
F/T |
n |
Mean |
Std Dev |
CV |
|
|
0.268 |
0.300 |
1.120 |
|
|
0.268 |
0.289 |
1.080 |
|
0.5 X PEL |
0.268 |
0.281 |
1.050 |
|
|
0.268 |
0.276 |
1.030 |
|
|
0.268 |
0.270 |
1.010 |
|
|
5 |
1.058 |
0.043 |
0.041 |
|
|
0.530 |
0.493 |
0.930 |
|
|
0.530 |
0.484 |
0.912 |
|
1 X PEL |
0.530 |
0.498 |
0.939 |
|
|
0.530 |
0.483 |
0.910 |
|
|
0.530 |
0.490 |
0.923 |
|
|
5 |
0.923 |
0.012 |
0.013 |
|
|
1.002 |
0.952 |
0.950 |
|
|
1.002 |
0.908 |
0.906 |
|
2 X PEL |
1.002 |
0.970 |
0.968 |
|
|
1.002 |
0.985 |
0.983 |
|
|
1.002 |
0.954 |
0.952 |
|
|
5 |
0.952 |
0.030 |
0.030 |
F/T = Found/Taken |
|
CV(Pooled) = 0.030 |
Average Recovery = 0.978 |
|
|
F Test Results |
|
Recoveries % |
|
|
|
|
Level |
F(calc) |
F(0.99) |
df |
|
25% |
50% |
80% |
RH |
0.5 X PEL |
5.55 |
6.70 |
2,13 |
|
105.2 |
100.0 |
105.8 |
|
1 X PEL |
79.4* |
6.93 |
2,12 |
|
92.9 |
102.7 |
92.3 |
|
2 X PEL |
3.73 |
6.70 |
2,13 |
|
98.7 |
95.7 |
95.2 |
|
|
Average |
|
|
98.9 |
99.5 |
97.8 |
|
df = degrees of freedom
* Large F value appears to be due to variability in sample generation
and not to any humidity effect.
Table 6
Comparison of Detector Tube and Gas Chromatograph Analyses
In-house Samples -
Side-by-Side |
|
|
Detector Tube
Recoveries |
Gas Chromatograph |
Recoveries |
Tube Mfg. |
N |
Recovery% |
Pooled CV |
N |
Recovery |
Pooled CV |
|
MSA (50% RH) |
9 |
111.8 |
0.069 |
5 |
107.9 |
0.012 |
Kitagawa (50% RH) |
9 |
108.0 |
0.039 |
8 |
103.0 |
0.050 |
Gastec (25% RH) |
18 |
101.3 |
0.063 |
11 |
98.8 |
0.031 |
Gastec (50% RH) |
17 |
109.9 |
0.058 |
7 |
106.5 |
0.014 |
Gastec (80% RH) |
18 |
106.2 |
0.057 |
13 |
97.8 |
0.032 |
Draeger (50% RH) |
9 |
101.8 |
0.076 |
6 |
100.2 |
0.010 |
|
Totals |
80 |
106.5 |
0.039-0.076 |
50 |
102.3 |
0.01-0.050 |
|
Detector Tube - Gas Chromatography Statistical Summary
Detector Tube Pooled CV2 (all
tubes) |
= |
0.025-0.076 |
Ave. Recovery, Detector Tubes (all tubes) |
= |
85.9-111.8% |
Gas Chromatograph (CV2 Pooled) |
= |
0.014 |
Gas Chromatograph (CVT Pooled) |
= |
0.026 |
Average Recovery, GC |
= |
99.5% |
Average Recovery, GC (all samples) |
= |
95-103% |
Field Samples |
|
Sample
# |
Detector Tube
Recoveries |
GC Recoveries |
|
Type |
N |
AV R% |
LAV R% |
Bagtube |
ppm CO2 |
|
1A |
P |
7 |
76 |
73 |
104% |
11,000 |
1B |
P |
7 |
81 |
77 |
93% |
15,000 |
2A |
P |
6 |
83 |
80 |
86% |
8,000 |
2B |
P |
7 |
83 |
74 |
83% |
7,600 |
3A |
A |
5 |
84 |
83 |
-- |
8,900 |
3B |
A |
6 |
(11000) |
(9100) |
-- |
LIS* |
4A |
A |
3 |
97 |
96 |
-- |
6,800 |
4B |
A |
6 |
72 |
62 |
-- |
5,300 |
5A |
A |
4 |
92 |
89 |
-- |
7,200 |
5B |
A |
5 |
92 |
84 |
-- |
9,700 |
N |
= |
Number of tubes or samples taken |
*LIS |
= |
Lost in shipment. Tube results for this sample are,
listed in ppm. |
AV R% |
= |
Average recovery in %, normalized to GC
results. |
LAV R% |
= |
Log normal average results in %, also normalized
(9.10., 9.11.). |
Bagtube |
= |
Detector tube sample taken on gas sampling bag prior
to gas chromatographic analysis. Bagtube samples were only taken
from the personal samples. |
P |
= |
Personal sample |
A |
= |
Area sample |
Table 7
Stability Test (Per Sample)
|
ppm Taken |
ppm Found |
Recovery % |
Day |
|
5260 |
5150 |
97.9 |
5 |
" " |
4760 |
90.5 |
18 |
" " |
3980 |
75.7 |
29 |
" " |
3650 |
68.9 |
50+ |
5440 |
4670 |
85.9 |
11* |
" " |
4370 |
80.4 |
31+ |
" " |
4840 |
89.1 |
11* |
" " |
4720 |
86.9 |
31*+ |
" " |
="2">4990 |
91.8 |
11* |
" " |
5010 |
92.1 |
31*+ |
6770 |
6550 |
96.8 |
5 |
7210 |
7020 |
t face="Arial" size="2">97.4 |
5 |
7570 |
7140 |
94.3 |
5 |
" " |
6880 |
90.9 |
18 |
" " |
6060 |
80.1 |
29 |
" " |
5730 |
75.7 |
50+ |
7990 |
7170 |
89.7 |
5 |
" " |
6210 |
77.7 |
18 |
" " |
5010 |
62.7 |
29 |
" " |
4490 |
56.2 |
50+ |
8940 |
8280 |
92.6 |
5 |
" " |
7810 |
87.4 |
18 |
" " |
6380 |
71.4 |
29 |
" " |
5040 |
56.4 |
50+ |
11010 |
10420 |
94.6 |
5 |
" " |
10750 |
97.6 |
18 |
" " |
9550 |
86.7 |
29 |
" " |
8850 |
80.4 |
50+ |
14600 |
14240 |
97.5 |
5 |
" " |
13940 |
95.5 |
18 |
" " |
12900 |
88.3 |
29 |
" " |
12040 |
82.5 |
50+ |
19300 |
19280 |
99.9 |
2 |
" " |
18410 |
95.4 |
5 |
" " |
17930 |
92.9 |
9 |
" " |
16460 |
85.3 |
14 |
" " |
13990 |
72.5 |
26 |
" " |
11330 |
58.7 |
37 |
* Samples stored in refrigerator at 5 °C. Data for refrigerated samples
is not included in final calculations. All other samples were stored at 20
°C. + All samples were analyzed using a 1 mL gas sampling loop, with
the exception of the (+) day stability study. On that day, a 5 mL loop was
used.
Appendix 1
Analysis Parameters for CO2 Determinations
Gas chromatograph (Hewlett-Packard 5730a gas
chromatograph)
Detector |
Thermal conductivity |
Sensitivity |
5 |
Helium flow rate |
15 - 25 mL/min |
Column temperature |
ambient (20 to 25 °C) |
Detector temperature |
ambient (20 to 25 °C) |
Valve manifold temperature |
ambient (20 to 25 °C) |
Column |
Chromosorb 102 (6 ft X 1/4 in. stainless steel,
80/100 mesh) |
Gas sampling loop |
1 mL |
Integrator (Hewlett-Packard 3385a automation system)
Attenuation |
4 |
Run time |
3.5 min |
Peak time |
2.6 - 2.9 min |
External valve switch |
0.01 s (from start of integration to valve
opening) |
Auxiliary signal |
a |
Chart speed |
1 |
Zero |
10 |
Area reject |
0 |
Generation of Dilution Air (CO2-free)
Figure 1
Carbon Dioxide - Air Flow
Generation System
Figure 2
Storage Stability - Carbon
Dioxide in Gas Sampling Bags
Figure 3
|