1. Introduction
This method describes the collection and
analysis of airborne cyanide and hydrogen cyanide gas.
1.1
History
Several methods for CN¯ determination have been developed
which are based on colorimetric reactions.
1.2
Uses
HCN gas is used for exterminating rodents and insects in
ships and for killing insects on trees, etc. NaCN is used for extracting
Au and Ag from ores, electroplating baths, fumigating citrus
and other fruit trees, ships, railway cars, warehouses, making
other cyanides and case hardening steel. (8.4)
1.3
Physical Properties (8.4)
For HCN gas: Colorless gas or
liquid; characteristic odor; very weakly acidic (does not redden
litmus): burns in air with a blue flame: intensely poisonous even when
mixed with air.
Physical Constants for
HCN:
Specific Gravity: 0.941 Melting Point:
-13.4°C Boiling Point: 25.6°C Molecular Weight:
27.03 For NaCN as a representative CN salt, physical
constants are:
Melting Point: 563°C Boiling Point:
1496°C Molecular Weight: 49.02 2. Working Range and Detection Limit
2.1 The working range is 0.25 ppm to 25 ppm
cyanide.
2.2 The detection limit is about 0.25 ppm cyanide for
linear response. 3. Precision and Accuracy
3.1 The following precision and accuracy data are based on
57 quality control samples prepared by the QC department of the OSHA
Analytical Laboratory and analyzed there. They were mixed cellulose
ester filters spiked with a CN¯ salt solution in the range of 200-400 µg
CN¯. The calculated (CVT) = 0.11 with a standard deviation =
0.10. The mean recovery 0.997. 4. Interferences
4.1 Ions that form insoluble silver salts such as Cl¯, I¯,
Br¯ (8.1) (i.e., I- conc. greater than 1/10 the CN -
conc.).
4.2 Many transition and heavy metals that form very
stable cyanide complexes such as Fe+2 , Fe+3 ,
Co+3 , Ni+2 , Cu+, Ag+,
Cd+, Hg+2 etc. The extent of complexation is
dependent upon the concentration of metal ions present, cyanide
concentration and the solution pH (8.1).
4.3 Sulfide ion.
Presence of sulfide ion in the sample solution will poison the
electrode. (8.3) 5. Advantages and Disadvantages
5.1 The procedure is quick and simple and has a large
concentration range.
5.2 High concentration samples can
quickly damage the electrode.
5.3 Impinger solutions can easily
be lost during collection.
5.4 Particulate CN¯ collected on
a filter converts to HCN upon exposure to moisture in the air.
Studies show approximately 16% conversion. (8.1)
5.5 Strong
reducing solutions in the atmosphere or sulfide contaminants cause
damage to the electrode. (8.1) 6. Sampling Procedure
Summary
6.1 The sample is drawn through a cassette containing a
37-mm mixed cellulose ester membrane filter ahead of a midget impinger
containing 10 mL of 0.1 N NaOH. A flow rate of 1.0 L/min and a minimum
sampling time of 60 minutes are recommended. the filter is used to
collect particulate cyanide while the impinger collects HCN
gas.
6.2 The filter cassettes are plugged. Within 1 hour of
sampling, transfer the filter to a screw-cap vial. The vial is sealed
with OSHA tape, and sent to the laboratory for
analysis.
6.3 The impinger solutions are transferred to
leak-proof vials with Teflon-lined caps. Vinyl or waterproof tape is
used around the caps to prevent leakage during shipment. The tape is
wrapped around the cap in the direction the cap is turned. The vials are
also sealed with an official OSHA seal and sent to the laboratory for
analysis.
6.4 Analysis should be completed as soon as possible to
avoid losses due to decomposition or volatilization of HCN.
7. Analytical Procedure
7.1 Apparatus
7.1.1 Cyanide Ion Specific Electrode - Orion Model 94-06
or equivalent.
7.1.2 Orion single-junction reference electrode
Model 90-01-00 or equivalent.
7.1.3 Magnetic stirrer and Teflon
stirring bars or battery operated stirring rod.
7.1.4 Millivolt
meter (with expanded scale)-Orion Ionalyzer 940 or
equivalent.
7.1.5 Calibrated personal sampling pump.
7.1.6 Midget impinger and filter cassette assembly.
7.1.7 Associated laboratory glassware including volumetric
flasks, plastic beakers, pipettes, etc. 7.2 Reagents
7.2.1 0.1 N NaOH solution - 4.0 g NaOH is added to
approximately 600 mL deionized water and the solution is diluted to 1
liter. The solution is stored in a polyethylene bottle.
7.2.2
1000 ppm CN¯ standard solution - 1.88 g NaCN reagent grade or better
is added to approximately 600 mL 0.1 N NaOH. The solution is diluted
to 1 liter with 0.1 N NaOH and stored in a polyethylene bottle. The
solution is stable for at least 6 months.
7.2.3 Lead acetate
paper.
7.2.4 Cadmium carbonate. ACS reagent grade or better.
7.3 Safety Precautions
7.3.1 Cyanide compounds are powerful poisons which prevent
the utilization of oxygen by the body tissues. Cyanides which contact
the skin or eyes should be washed off immediately since they can be
absorbed through the skin.
7.3.2 Disposal of samples should be
accomplished by pouring down a sink with copious amounts of tap water.
In no case should cyanide standards come in contact with acids since
this will produce deadly hydrogen cyanide gas. 7.4.
Standard Preparation
7.4.1. All previously cleaned glassware is rinsed with 0.1
N NaOH and deionized water prior to use.
7.4.2. Standard
solutions are prepared by appropriate dilution of stock 1000 ppm CN¯
solution. All dilutions are made using 0.1 N NaOH. 100 ppm, 25 ppm, 10
ppm, 1 ppm, and 0.1 ppm solutions are made in the following
manner:
Standard Solution
used |
Dilution |
Final Conc. |
|
|
|
1000 ppm CN |
10 mL:100 mL |
100 ppm CN¯ |
100 ppm |
25 mL:100 mL |
25 ppm |
100 ppm |
10 mL:100 mL |
10 ppm |
10 ppm |
10 mL:100 mL |
1 ppm |
1 ppm |
10 mL:100 mL |
0.1
ppm | 7.5 Sample
Preparation
7.5.1 The volume of each impinger sample solution is
measured and recorded. If an analysis other than hydrogen cyanide
is to be done, then an aliquot of the sample is measured out for
cyanide analysis into a 25 mL volumetric flask and diluted to volume
with 0.1 N NaOH. The entire sample may be used as the aliquot if no
other analyses are required.
7.5.2 Filter samples are
transferred to a 50 mL plastic beaker and desorbed with 25 mL 0.1 N
NaOH for at least 30 minutes with occasional agitation. If an analysis
other than cyanide is to be done, then an aliquot of the sample is
measured out for cyanide analysis into a 25 mL volumetric flask and
diluted to volume with 0.1 N NaOH. The entire sample may be used as
the aliquot if no other analyses are required.
7.5.3 A bulk
sample is prepared by weighing out 1 g of the sample and dissolving it
in enough 0.1 N NaOH to give a total volume of 25 mL.
7.5.4
Sulfide ion contamination of the sample solution is determined by
touching a drop of the sample solution to a piece of lead acetate
paper. Discoloration of the test paper indicates sulfide ion
contamination. If no sulfide is indicated, proceed to 7.6. (8.2)
7.5.5 Sulfide ion contamination can be removed from the sample
solutions by addition of a small amount (spatula tip) of cadmium
carbonate to the sample. The solution pH must be in the range 11-13.
After each addition of cadmium carbonate, the sample is swirled and
rechecked for sulfide ion presence prior to addition of more cadmium
carbonate. (Note: Avoid a large excess of cadmium carbonate and long
contact times.)
7.5.6 When a drop of liquid no longer discolors
the strip of lead acetate paper, solids are removed by
filtering the sample solution through a plug of glass wool contained
in a small glass funnel. The funnel is not rinsed since the sample has
already been diluted to volume. 7.6 Analysis
7.6.1 The cap is removed from the CN¯ ion specific
electrode. (Note: The electrode should be stored dry with a cap over
the end.) Polish the end of the electrode with some deionized water
and a polishing strip (94-82-01) for about 30 seconds. This will
remove build-up and improve electrode response. Do this
before and after electrode use.
7.6.2 The single-junction
reference electrode is filled with saturated KC1 filling solution
(90-00-01). (Note: The reference electrode should be stored empty and
dry as response time will be quicker.) It should be stored dry
only after it has been thoroughly cleaned and rinsed.
7.6.3
The potentiometer, Orion 940 Ionalyzer or Altex SelectIon 5000, is
used following manufacturer's instructions. Plug the electrodes into
the appropriate jacks on the back of the instrument.
7.6.4
Samples and standards are placed in 50 mL polyethylene beakers, Each
is analyzed in the relative mV or concentration mode. (Note: See
instrument instructions for use of either mode. See table 1 for the
instructions for the Orion 940.) A battery operated stirring rod is
inserted into the solution before each analysis. The reproducibility
of sample concentration is dependent on constant electrode temperature
and constant stirring rate.
7.6.5 The electrodes are placed
into the solution at about the same depth each time, making sure the
stirring rod does not contact the electrodes.
7.6.6 Electrodes
should be removed from solutions reading higher than the 25 ppm CN¯
standard and an aliquot should be taken and diluted to 25 mL with 0.1
N NaOH for analysis. Failure to remove the electrode immediately will
cause damage to the electrode@ and shorten its life. If the electrode
is exposed to a sample of high concentration it should be quickly
removed from the solution, polished with a polishing strip and rinsed
well. Check its response in a standard solution to be sure it is
giving a reproducible reading before proceeding.
7.6.7
Electrode voltage readings are allowed to stabilize for each sample.
They stabilize faster in higher concentration solutions than in lower
ones. The stable reading is recorded. Between samples, the electrodes
are rinsed with deionized water and carefully blotted dry prior to
introduction into the next one to avoid contamination. If the next
sample is not immediately ready for determination, the electrodes
should be placed in 0.1 N NaOH solution to reduce electrode
stabilization time.
7.6.8 Typical voltage readings for 1, 10,
and 25 ppm CN standard solutions are -115 mV, -175 mV, and -200 mV ±20
mV. Standards should be read at the beginning of the sample run and
also at the end. Selected standards should be read between sample
determinations to bracket sample values obtained.
7.6.9 The 1
ppm CN¯ standard mV reading should be logged into the QC worksheet
near the bench along with the I.S.E. and reference electrode ID
numbers. The analyst initials and date should be included.
7.7 Calculations
7.7.1 Since the instrument was calibrated in the
concentration mode the readings are simply plotted versus
concentration of standards in ppm using the colorimetric or AA least
squares plotting programs. Both of these curves give the sample ppm of
CN.
7.7.2 The equation for calculating the air concentration
of CN¯ in mg/m³ is:
mg
|
= |
(sample ppm × dilution
factor - blank ppm) × 25 ml
|
m³ |
Air Volume
(L) | 7.7.3 The
equation for calculating the air concentration of HCN gas in ppm is:
(result from equation in 7.7.3) × (1.038) × (0.906)
where the gravimetric factor = 1.038 for HCN from CN and the
ppm in air conversion factor = 0.906 for HCN.
This may be
entered in the Auto AA program by combining the factors and entering
the result as a "gravimetric factor" of 0.9404. [0-9404 =
(1-038)(0-906)] 8.
References
8.1 Instrument Manual Cyanide Ion Activity Electrode Model
94-06, Orion Research Inc.
8.2 Draft Report Cyanide (as CN)
Method Number S-250, validation date Jan. 30, 1976.
8.3
Cyanide in Air, P&CAM 116 NIOSM Manual of Analytical Methods, issued
Sept. 7, 1972, revised Dec. 1, 1973.
8.4 Merck Index, Tenth Edition 1983 page 696.
Analysis using the Orion 940 follows a menu array. Answer the
questions of the menu appropriately by using the "Yes", "No" and the
numeric keys as follows:
TABLE 1
OPERATOR
MENU? |
Answer "YES" |
CHANGE ELECTRODE
ID? |
Answer "YES" |
ELECTRODE 1 =
NH3 IS THIS CORRECT? |
Answer "NO" to change the
ID. Continue to answer "NO" until CN¯ is displayed as a choice. Use
the numeric keys to select CN¯ . The display will change
to... |
ELECTRODE 1 =
CN- IS THIS CORRECT? |
Answer "YES". |
SET ABSOLUTE OR
RELATIVE MILLIVOLTS? |
Answer "NO". |
CHANGE THE pH OR
ISE LIMITS? |
Answer "NO". |
SET
TIMER? |
Answer "NO". |
CHANGE PRINT
INTERVAL? |
Answer "NO". |
SET
TEMPERATURE? |
Answer "NO". |
CHANGE THE TIME
AND DATE? |
Answer "NO". |
ENTER STANDBY
MODE? |
Enter "SPEED 0" to
continue. |
CALIBRATE CN-
? |
Answer "YES". |
CALIBRATE BY
DIRECT MEASUREMENT? |
Answer "YES". |
ENTER
NUMBER OF STANDARDS (1-5)
|
Enter 2 to standardize on
the 1 and 10 ppm standards. |
DO BLANK
CORRECTION? |
Answer "NO". |
1-CN-
ELECTRODE IN STANDARD 1? |
Place the electrodes into
the 1 ppm standard and answer "YES". Wait for the reading
to stabilize. |
STD 1 =
1.16 CAL AS 1.00? |
Use the numeric keys to
change to "CAL AS 1.0011 if necessary then answer
"YES". |
1-CN-
ELECTRODE IN STANDARD 2? |
Place the the electrodes
into the 10 ppm standard and answer "YES". Wait for the reading to
stabilize. |
STD 2 =
10.00 CAL AS 10.00? |
Correct if necessary and
answer "YES". |
SLOPE = -59.2
mV/DEC YES TO CONTINUE |
Record the slope value in
your log book and press "YES". |
MEASURE
1-CN-? 10:04
11-21-88 |
Place electrodes in the
first standard and answer "YES". |
1-CN- =
1.00 10:05 READY 23.1°C |
Record this concentration
value for the standard. Rinse the electrodes and blot dry. Place the
electrodes in the next sample or standard. Record the values shown
(when the READY message is given) as the concentration of
the appropriate sample. Enter "NO" to exit the measurement
mode. |
|