ENDOSULFAN
Method number: |
PV2023 |
|
Matrix: |
Air |
|
Target Concentration: |
0.1 mg/m3 (TLV time weighted
average) |
|
Procedure: |
Samples are collected by drawing known volumes of air through
OSHA versatile sampler tubes (OVS-2) containing a glass fiber filter
and two sections of XAD-2 adsorbent. Samples are
desorbed with toluene and analyzed by gas chromatography (GC) using
an electron capture detector (ECD). |
|
Recommended air volume and sampling rate: |
60 L at 1.0 L/min |
|
Detection limit of the overall procedure (based on the
recommended air volume): |
3.4 µg/m3 |
|
Status of method: |
Stopgap method. This method has been only partially evaluated
and is presented for information and trial use. |
|
|
|
|
Date: April, 1988 (final) |
Chemist: Duane Lee |
Carcinogen And Pesticide Branch OSHA Analytical
Laboratory Salt Lake City, Utah
1. General Discussion
1.1. Background
1.1.1. History of procedure
The OSHA Analytical Laboratory received a set of samples
requesting the analysis of endosulfan from glass fiber filters.
Retention and storage studies on glass fiber filters yielded poor
recoveries of endosulfan. Therefore, this report describes the
preliminary validation of a sampling and analytical method using
OVS-2 tubes.
1.1.2. Toxic effects (This section is for information only and
should not be taken as the basis of OSHA policy).
Technical endosulfan consists of about four parts of a-cis isomer, and one part of
ß-trans isomer. The a isomer,
which is somewhat more insecticidal, is slowly converted to the more
stable ß form at high temperatures and both isomers are oxidized
slowly in air and biological systems and rapidly by peroxides or
permanganates to endosulfan sulfate (Ref. 5.2.). Endosulfan has an
acute LD50 to the rat of 30 mg/kg in
alcohol suspension, 70 mg/kg in aqueous suspension, and 110 mg/kg in
oil (Ref. 5.1.). For fish the LC50 value
is 0.001-0.003 ppm (Ref. 5.2.).
There are reports of workers becoming ill from inhalation of dust
from endosulfan. Symptoms of slight nausea, confusion, excitement,
flushing and dry mouth were experienced (Ref. 5.3.). Also,
endosulfan is a central nervous system stimulant for which no
specific antidote is available (Ref. 5.1.).
Accordingly, a TLV of 0.1 mg/m3, as a
time-weighted average, is recommended for endosulfan
(Ref. 5.3.).
1.1.3. Potential workplace exposure
No estimate of worker exposure to endosulfan could be found.
Endosulfan is a broad spectrum insecticide for control of vegetable,
fruit, field crop, and ornamental pests (Ref. 5.1.).
1.1.4. Physical properties (Ref. 5.1. and 5.3.)
Molecular weight: |
406.95 |
|
Molecular formula: |
C9H6CL6O3S |
|
CAS #: |
115-29-7 |
|
Specific gravity: |
1.735 at 20 °C |
|
Melting point: |
70-100 °C |
|
Solubility: |
Insoluble in water; soluble in xylene,
kerosene, chloroform, acetone, and alcohol; decomposes in the
presence of acids and alkalies to form sulfur dioxide. |
|
Chemical name: |
6,7,8,9,10,10-hexachloro- 1,5,5a,6,9,9a-hexahydro-6,9- methano-2,4,3-benzodioxathiepin 3-oxide |
|
Synonyms: |
Thiodan, Benzoepin, Cyclodan, Beosit, Endocel,
Chlorthiepin, Crisulfan, Endosan, Endosol,Hildan,
Insectophene, Malix, Thifor, Thimul, Thiofor, Thionex,
Tiovel |
|
Structure: |
|
|
Description: |
Technical endosulfan is a tan,
semi-waxy solid that is a mixture of two isomers
with an odor of hexachlorocyclopentadiene and may have a
slight sulfur dioxide odor. |
1.2. Limit defining parameters
The detection limit of the analytical procedure is 13.5 pg per
injection. This is the amount of analyte which will give a peak whose
height is approximately five times the baseline noise. (Figure 1)
2. Sampling procedure
2.1. Apparatus
2.1.1. Samples are collected by using a personal sampling pump
that can be calibrated to within ±5% of the recommended flow rate
with the sampling device in line.
2.1.2. Samples are collected with OVS-2 tubes, which
are specially made 13 mm O.D. glass tubes that are tapered to 6 mm
O.D., packed with 140-mg backup section and a 270-mg
sampling section of cleaned XAD-2 and a
13-mm glass fiber filter. The backup section is
retained by two foam plugs and the sampling section is between one
foam plug and the glass fiber filter. The glass fiber filter is held
next to the sampling section by a polytetrafluoroethylene (PTFE)
retainer.
2.2. Reagents
None
2.3. Sampling technique
2.3.1. Attach the small end of the sampling tube to the sampling
pump with flexible plastic tubing such that the large front section
of the sampling tube is exposed directly to the atmosphere.
2.3.2. Attach the sampler vertically in the employee's breathing
zone in such a manner that it does not impede work performance.
2.3.3. After sampling for the appropriate time, remove the
sampling device and seal the tube with plastic end caps.
2.3.4. Wrap each sample end-to-end with an OSHA seal
(Form 21).
2.3.5. Submit at least one blank for each set of samples. Handle
the blank in the same manner as the samples, except no air is drawn
through it.
2.3.6. Record the air volume (in liters of air) for each sample,
and list any possible interferences.
2.3.7. Submit bulk samples for analysis in a separate container.
2.4. Desorption efficiency
Six OVS-2 tubes were each liquid spiked with 15 µL of
a 422 µg/mL endosulfan standard. The tubes were stored in a drawer
overnight at room temperature. The next day the samples were desorbed
in 3 mL of toluene by rotating them for 60 min and then analyzed. The
results are listed in table 2.4.
Table 2.4. Extraction Efficiency
|
Amount |
Amount |
% |
Sample # |
Spiked, µg |
Found, µg |
Recovered |
|
Ex1 Ex2 Ex3 Ex4 Ex5 Ex6 |
6.33 6.33 6.33 6.33 6.33 6.33 |
5.40 5.19 5.47 5.16 5.30 5.36 |
85.3 82.0 86.4 81.5 83.7 84.7 |
|
|
Average =
83.9 |
2.5. Retention efficiency
Five OVS-2 tubes were liquid spiked with 15 µL of a
422 µg/mL standard and humid air (~80% relative
humidity) was drawn through each tube at 1 L/min for 60 minutes. The
tubes were stored in a drawer overnight at room temperature. The next
day the tubes were desorbed with 3 mL of toluene by rotating them for
60 min and then analyzed. The results are listed in table 2.5.
Table 2.5. Retention Efficiency
|
Amount |
Amount |
% |
Sample # |
Spiked, µg |
Found, µg |
Recovered |
|
R1 R2 R3 R4 R5 |
6.33 6.33 6.33 6.33 6.33 |
5.68 5.85 5.65 5.59 5.36 |
89.7 92.4 89.3 88.3 84.7 |
|
|
Average =
88.9 |
2.6. Sample storage
Twelve tubes were liquid spiked with 15 µL of a 422 µg/mL standard
and humid air (~80% relative humidity) was drawn
through each tube at 1 L/min for 60 minutes. Six of the samples were
stored at ambient temperature in a drawer, and six were stored in a
refrigerator. After four days of storage, three samples from each
group were desorbed with 3 mL of toluene by shaking for 60 min and
then analyzed. The remaining samples were desorbed and analyzed after
seven days of storage. The results are given in the tables below.
Table 2.6.1. Ambient Storage
|
Amount |
Amount |
% |
Sample # |
Spiked, µg |
Found, µg |
Recovered |
|
4 4 4 7 7 7 |
6.33 6.33 6.33 6.33 6.33 6.33 |
6.08 5.72 5.77 5.99 5.74 5.41 |
96.0 90.4 91.2 94.6 90.7 85.5 |
|
|
Average of 4 days =
92.5 |
|
Average of 7 days =
90.3 |
Table 2.6.2 Refrigerated Storage
|
Amount |
Amount |
% |
Sample # |
Spiked, µg |
Found, µg |
Recovered |
|
4 4 4 7 7 7 |
6.33 6.33 6.33 6.33 6.33 6.33 |
6.08 5.85 5.48 5.67 5.77 5.69 |
96.0 92.4 86.8 89.6 91.2 89.9 |
|
|
Average of 4 days =
91.7 |
|
Average of 7 days =
90.2 |
2.7. Recommended air volume and sampling rate
2.7.1. The recommended air volume is 60 L.
2.7.2. The recommended flow rate is 1.0 L/min.
2.8. Interferences
It is not known if any compounds will interfere with the collection
of endosulfan.
2.9. Safety precautions
2.9.1. Attach the sampling equipment in such a manner that it
will not interfere with work performance or employee safety.
2.9.2. Follow all safety practices that apply to the work area
being sampled.
3. Analytical procedure
3.1. Apparatus
3.1.1. A balance capable of weighing to the nearest tenth of a
milligram. A Mettler HL52 balance was used in this evaluation.
3.1.2. Mechanical rotator.
3.1.3. A gas chromatograph (GC) equipped with an electron capture
detector (ECD). A Hewlett Packard 5890 was used in this evaluation.
3.1.4. A GC column capable of separating endosulfan from any
interferences. A 10 m × .32 mm i.d. (1.0 µm film) DB-5
column was used in this evaluation.
3.1.5. An electronic integrator, or some other suitable method
for measuring detector response. The Hewlett-Packard
3357 Laboratory Data System was used in this evaluation.
3.1.6. Volumetric flasks and pipets.
3.1.7. Vials, 4-mL with PTFE-lined
septum.
3.1.8. Vials, 2-mL suitable for use on GC
autosamplers.
3.2. Reagents
3.2.1. Toluene high purity Burdick and Jackson.
3.2.2. Endosulfan EPA 3180 97.6% purity.
3.2.3. Hexachlorobenzene, reagent grade.
3.2.4. Desorbing solvent, hexachlorobenzene (2µg/mL) in toluene
3.3. Standard preparation
Prepare stock standards by weighing 10 to 14 mg of endosulfan,
placing in 25-mL volumetric flasks, and diluting to
volume with toluene. Make working range standards by pipet dilutions
of the stock standards with the desorbing solvent. Store stock and
dilute standards in a freezer.
3.4. Sample preparation
3.4.1. Transfer the glass fiber filter and large section of the
adsorbent of each sample to a 4-mL vial. Place the
separating foam plug and small section of adsorbent of each sample
in a separate 4-ml vial.
3.4.2. Pipet 3.0 mL of desorbing solvent into each vial and seal
with a Teflon-lined septum.
3.4.3. Rotate the vials for 60 minutes.
3.5. Analysis
3.5.1. Instrument conditions
Column: |
DB-5, 1.0 µm film, 10 m × 0.32 mm i.d. |
|
Injector temperature: |
235 °C |
|
Column temperature: |
170 °C |
|
Detector temperature: |
300 °C |
|
Gas flows: |
Column 8.6 mL/min hydrogen Make up 42 mL/min
nitrogen |
|
Injector volume: |
1.0 µL |
|
Split ratio: |
5:1 |
|
Retention time: |
10.3 min endosulfan I 15.6 min endosulfan
II |
3.5.2. Chromatogram (see Figure 2)
3.6. Interferences
3.6.1. Any collected compound having a similar retention time and
responds to an ECD is an interference.
3.6.2. GC conditions may be varied to circumvent an interference.
3.6.3. Retention time alone is not proof of chemical identity.
Analysis by an alternate GC column and confirmation by mass
spectrometry are additional means of identification.
3.7. Calculations
3.7.1. A calibration curve (figure 3) is constructed by plotting
detector response versus standard concentration. The detector
response is the value calculated from an internal standard method
that sums the areas of the endosulfan I and endosulfan II peaks.
3.7.2. The concentration of endosulfan in a sample is determined
from the calibration curve.
3.7.3. The air concentration is then determined by the following
formula.
mg/m3 = |
(µg/mL in sample) × (extraction
volume in mL)
(air volume in liters) × (desorption
efficiency) |
3.8. Safety precautions
3.8.1. Avoid skin contact and air exposure to endosulfan.
3.8.2. Avoid skin contact with all solvents.
3.8.3. Wear safety glasses at all times.
4. Recommendations for further study
The method should be fully validated.
Figure 1 Chromatogram at the Detection Limit
Figure 2 Chromatogram of Endosulfan
Figure 3 Calibration Curve
5. References
5.1. Farm Chemicals Handbook; Meister Publishing:
Willoughly, Ohio, 1986, p C97.
5.2. Kirk-Othmer Encyclopedia of Chemical Technology; John
Wiley & Sons: New York, 1981, Volume 13, PP 435-7.
5.3. Documentation of Threshold Limit Values and Biological
Exposure Indices; American Conference of Governmental Industrial
Hygienists Inc., Fifth Edition, 1986, p 230.
|