OSHA banner
Technical Links > Sampling & Analytical Methods > Index
  DOL logo

TETRACHLOROETHYLENE
TRICHLOROETHYLENE


Method number: 1001
 
Matrix: Air
TWA
Ceiling (5 min)
Peak
Target concentration:
  Tetrachloroethylene
  Trichloroethylene
100 ppm (678 mg/m3)
100 ppm (537 mg/m3)
200 ppm (1356 mg/m3)
200 ppm (1074 mg/m3)
300 ppm (2034 mg/m3)
300 ppm (1612 mg/m3)
OSHA PEL:
  Tetrachloroethylene
  Trichloroethylene
100 ppm (678 mg/m3)
100 ppm (537 mg/m3)
200 ppm (1356 mg/m3)
200 ppm (1074 mg/m3)
300 ppm (2034 mg/m3)
300 ppm (1612 mg/m3)
ACGIH TLV:
  Tetrachloroethylene
  Trichloroethylene
25 ppm (169 mg/m3)
50 ppm (269 mg/m3)
100 ppm (678 mg/m3)(STEL)
100 ppm (537 mg/m3)(STEL)
none
none
 
Procedure: Adsorbent tube samples are collected by drawing workplace air through coconut shell charcoal tubes using personal sampling pumps. Diffusive samples are collected by exposing SKC 575-002 Passive Samplers to workplace air. Samples are extracted with carbon disulfide and the extracts are analyzed by gas chromatography using flame ionization detection.
 
Recommended
sampling parameters:
Charcoal tubes

    Sampling rate:
    Sampling time:
TWA Samples

50 mL/min
< 240 min
Ceiling Samples

50 mL/min
> 5 min
Peak Samples

50 mL/min
> 1 min
SKC 575-002 Samplers

    Exposure time:
 
 
< 240 min
 
 
> 5 min
 
 
> 5 min
 
Charcoal Tubes
SKC 575-002 Samplers
Reliable quantitation limit: (240-min samples)
  Tetrachloroethylene
  Trichloroethylene
14 ppb (98 µg/m3)
13 ppb (68 µg/m3)
104 ppb (705 µg/m3)
61 ppb (325 µg/m3)


Standard error
of estimate at:
Tetrachloroethylene
Trichloroethylene

 
 
  TWA: (240 min)
  Ceiling: (5 min)
  Peak: (1 min)
Char. Tubes

5.1%
5.0%
5.2%
SKC 575-002

 9.0%*
9.6%*
----
Char. Tubes

5.1%
5.0%
5.1%
SKC 575-002

 9.1%*
9.1%*
----
 
*For samples where sampling site atmospheric pressure and temperature are known. When either or both of these values are unknown, see Table 4.4.2 for applicable standard errors of estimate.
 
Status of method: Evaluated method. This method has been subjected to the established evaluation procedures of the OSHA SLTC Methods Development Team.
 
May 1999 Carl J. Elskamp



Methods Development Team
Industrial Hygiene Chemistry Division
OSHA Salt Lake Technical Center
Salt Lake City UT 84115-1802




1. General Discussion

1.1 Background

    1.1.1 History

    The determinations of tetrachloroethylene and trichloroethylene in workplace air have consistently been among the top analyses for solvent vapors done at the OSHA Salt Lake Technical Center (SLTC). Until this present evaluation was completed, OSHA did not have fully validated methodologies to determine workplace exposures to vapors of either of these solvents. This method is an extension of the evaluation work recently completed for toluene where diffusive as well as active samplers were investigated.(1) As with toluene, exposure limits to both of these solvents are covered in Table Z-2, thus tests were conducted at TWA, ceiling and peak target concentrations. After the toluene method was completed, by consensus it was decided that future evaluation work by the OSHA SLTC Methods Development Team for solvent vapors would first address charcoal tubes for active sampling as well as SKC 575-002 Samplers (which contain Anasorb 747) for diffusive sampling. It was also decided that carbon disulfide (CS2) would be the first choice for the extraction solvent for each of these samplers. Using these two samplers and CS2 as the extraction solvent, sampling and analytical methods for both tetrachloroethylene and trichloroethylene were fully validated at TWA, ceiling and peak target concentrations.

    1.1.2 Toxic effects (This section is for information only and should not be taken as the basis of OSHA policy.)

    Overexposure to tetrachloroethylene through inhalation can affect the central nervous system and liver. Prolonged exposure to 200 ppm causes dizziness, headache, confusion, nausea, and eye and mucous membrane irritation. These effects are intensified and include incoordination and drunkenness when concentrations exceed 600 ppm. When concentrations exceed 1000 ppm, anesthetic and respiratory depression effects can cause unconsciousness and death. One single, brief exposure to greater than 6000 ppm can be immediately dangerous to life. Prolonged exposures to greater than 200 ppm have reportedly caused reversible changes to the liver. Alcohol consumption before or after exposures may increase adverse effects. Repeated exposure to liquid tetrachloroethylene may defat the skin, causing dermatitis. It can cause significant discomfort if splashed in the eyes and can cause transient, reversible corneal damage.(2)

    Exposure to high concentrations of trichloroethylene can cause headaches, vertigo, tremors, nausea with vomiting, fatigue, intoxication, unconsciousness, and even death. It is acutely toxic, primarily because of its anesthetic effect on the central nervous system. Exposure by inhalation is followed by rapid absorption into the bloodstream. Concentrations of 3000 ppm can cause unconsciousness in less than 10 minutes and anesthetic effects can occur after 20-minute exposure to 400 ppm. Skin contact may cause dermatitis.(3)

    1.1.3 Workplace exposure

    About 50% of the demand for tetrachloroethylene is used in the dry cleaning industry where around 80% of all dry cleaners use it as their primary cleaning agent. It is also used as a feedstock in the production of fluorocarbons, which accounts for 30% of the current demand. About 12% of the demand is used as a degreaser for metals and the remaining 8% is utilized in miscellaneous applications such as transformer insulating fluid, chemical maskant formulations, and as a process solvent for desulfurizing coal.(4)

    Around 85% of the trichloroethylene produced in the United States is used in degreasing of fabricated metal parts with the remaining 15% divided equally between exports and use in miscellaneous applications. Some of these applications include use as a component in adhesive and paint-stripping formulations, a low temperature heat-transfer medium, a nonflammable solvent carrier in industrial paint systems, and a solvent base for metal phosphatizing systems. It is also used in the textile industry as a carrier solvent for spotting fluids and as a solvent in waterless preparation dyeing and finishing operations. Because of the environmental regulations that have been placed on the use of trichloroethylene, for many of these applications it is gradually being replaced with alternative solvents such as 1,1,1-trichloroethane.(5)

    1.1.4 Physical properties

    Tetrachloroethylene:(6)

    CAS number: 127-18-4
    IMIS number: 2020
    molecular weight: 165.83
    boiling point: 121.2°C
    melting point: -22.7°C
    appearance: colorless liquid
    density: 1.623
    molecular formula: C2Cl4
    vapor pressure: 5.47 kPa at 40°C
    flash point: none
    odor: etheral(7)
    explosive limits: not applicable
    solubility: 15 mg is soluble in 100 g of water at 25°C; miscible with chlorinated solvents and most other common solvents.
    synonyms: Perchloroethylene; ethylene tetrachloride; tetrachloroethylene; Nema; Tetracap; Tetropil; Perclene; Ankilostin; Didakene.(8)
    structural formula: Cl2C=CCl2


    Trichloroethylene:(9)
     
    CAS number: 79-01-6
    IMIS number: 2490
    molecular weight: 131.39
    boiling point: 87.3°C
    melting point: -86.5°C
    appearance: colorless liquid
    density: 1.464
    molecular formula: C2HCl3
    vapor pressure: 9.3 kPa at 25°C
    flash point: none
    odor: sweet smelling, chloroform-like
    explosive limits: 8% to saturation at 25°C
    solubility: 107 mg is soluble in 100 g of water at 20°C; miscible with many organic liquids.
    synonyms: Trichloroethene; ethinyl trichloride; Tri-Clene; Trilene; Trichloren; Algylen; Trimar; Triline; Trethylene; Westrosol; Chlorylen; Gemalgene; Germalgene.(10)
    structural formula: Cl2C=CHCl



This method was evaluated according to the OSHA SLTC "Evaluation Guidelines for Air Sampling Methods Utilizing Chromatographic Analyses" (11) in which the terms used and evaluation procedures followed for this method are fully discussed and explained. The analyte air concentrations throughout this method are based on the recommended sampling and analytical parameters. TWA target concentration samples are based on 240 minutes and ceiling samples on 5 minutes of charcoal tube sampling and diffusive sampler exposure. Peak samples are based on 1 minute of sampling with charcoal tubes. Air concentrations listed in ppb and ppm are referenced to 25°C and 101.3 kPa (760 mmHg).


1.2 Limit defining parameters

    1.2.1 Detection limit of the analytical procedure

    The detection limits of the analytical procedure are 16.2 pg of tetrachloroethylene and 11.2 pg of trichloroethylene. These are the amounts of analytes that will give detector responses that are significantly different from the responses of a reagent blank. (Section 4.1)

    1.2.2 Detection limit of the overall procedure

    The detection limits of the overall procedure are 0.35 µg of tetrachloroethylene per sample (4.3 ppb or 29 µg/m3) and 0.24 µg of trichloroethylene per sample (3.7 ppb or 20 µg/m3) for charcoal tubes and 0.66 µg of tetrachloroethylene per sample (31 ppb or 211 g/m3) and 0.33 µg of trichloroethylene per sample (18 ppb or 97 µg/m3) for SKC 575-002 Samplers. These are the amounts of analytes spiked on the respective samplers that will give detector responses that are significantly different from the responses of respective sampler blanks. (Section 4.2)

    1.2.3 Reliable quantitation limit

    The reliable quantitation limits are 1.17 µg of tetrachloroethylene per sample (14 ppb or 98 µg/m3) and 0.81 µg of trichloroethylene per sample (13 ppb or 68 µg/m3) for charcoal tubes and 2.21 µg of tetrachloroethylene per sample (104 ppb or 705 µg/m3) and 1.11 µg of trichloroethylene per sample (61 ppb or 325 µg/m3) for SKC 575-002 Samplers. These are the amounts of analytes spiked on the respective samplers that will give detector responses that are considered the lower limits for precise quantitative measurements. (Section 4.2)

    1.2.4 Instrument calibration

    The coefficient of determination (r2) is 0.99996 for tetrachloroethylene and 0.99993 for trichloroethylene over the ranges of 21.1 to 16,230 µg/mL and 19.0 to 12,874 µg/mL respectively. The coefficient of non-determination (k2) is 4×10-5 for tetrachloroethylene and 7×10-5 for trichloroethylene over the same ranges. These ranges correspond to 0.5 times the ceiling target concentrations for SKC 575-002 Samplers to 2 times the TWA target concentrations for charcoal tubes. (Section 4.3)

    1.2.5 Precision (overall procedure)

    Charcoal tubes

    The precisions of the overall procedure at the 95% confidence level from the ambient temperature storage tests for TWA, ceiling and peak samples for charcoal tubes are given in Table 1.2.5.1. The TWA samples are 240-min samples taken from 100-ppm atmospheres, the ceiling samples are 5-min samples taken from 200-ppm atmospheres and the peak samples are 1-min samples taken from 300-ppm atmospheres. (Section 4.4)

    Table 1.2.5.1
    Precision of the Overall Procedure at the 95% Confidence Interval for Charcoal Tubes
    analyte TWA samples ceiling samples peak samples

    tetrachloroethylene
    trichloroethylene
    ±9.9%
    ±9.9%
    ±9.9%
    ±9.9%
    ±10.2%
    ±10.0%

    SKC 575-002 Samplers

    The precisions of the overall procedure at the 95% confidence level from the ambient temperature storage tests for TWA and ceiling samples for SKC 575-002 Samplers are given in Table 1.2.5.2. The TWA samples are 240-min samples taken from 100-ppm atmospheres and the ceiling samples are 5-min samples taken from 200-ppm atmospheres. There are different values given for each analyte at each of the levels, depending on whether both, either, or neither temperature or atmospheric pressure are known at the sampling site. If the temperature is not known, it is assumed to be 22.2 ± 15°C (72 ± 27°F) and a variability of ±7.7% is included. If the atmospheric pressure is not known, it is estimated from the sampling site elevation and a variability ±3% is included. (Section 4.4)

    Table 1.2.5.2
    Precision of the Overall Procedure at the 95% Confidence Interval for SKC 575-002 Samplers When Sampling Site Temperature (T) or Atmospheric Pressure (P) are Known or Unknown
    analyte TWA samples ceiling samples

    tetrachloroethylene
      both T&P known
      only T known
      only P known
      neither T nor P known

    trichloroethylene
      both T&P known
      only T known
      only P known
      neither T nor P known
     
    ±17.7%
    ±18.6%
    ±23.3%
    ±23.9%
     
     
    ±17.9%
    ±18.8%
    ±23.3%
    ±24.1%
     
    ±18.9%
    ±19.8%
    ±24.1%
    ±24.9%
     
     
    ±17.9%
    ±18.9%
    ±23.5%
    ±24.1%

    1.2.6 Recovery

    The recovery of tetrachloroethylene from TWA samples in 17-day storage tests remained above 99.5% and 98.6% for charcoal tubes and SKC 575-002 Samplers respectively when the samples were stored at ambient temperatures. The recovery of trichloroethylene from TWA samples in 17-day storage tests remained above 98.5% and 96.8% for charcoal tubes and SKC 575-002 Samplers respectively when the samples were stored at ambient temperatures. The recovery of tetrachloroethylene from ceiling samples in 17-day storage tests remained above 100% and 97.3% for charcoal tubes and SKC 575-002 Samplers respectively when the samples were stored at ambient temperatures. The recovery of trichloroethylene from ceiling samples in 17-day storage tests remained above 98.8% and 97.4% for charcoal tubes and SKC 575-002 Samplers respectively when the samples were stored at ambient temperatures. The recoveries of tetrachloroethylene and trichloroethylene from charcoal tube peak samples in 17-day storage tests remained above 97.3% and 96.5% respectively when the samples were stored at ambient temperatures. (Section 4.5)

    1.2.7 Reproducibility

    Six samples for both samplers at the TWA and ceiling target concentrations and six charcoal tube samples at the peak target concentrations that were collected from controlled test atmospheres, along with a draft copy of this procedure, were submitted to an SLTC service branch for analysis. All samples were analyzed 22 days after generation and had been stored at room temperature. No individual sample result deviated from its theoretical value by more than the precisions reported in Section 1.2.5. (Section 4.6)

2. Sampling Procedure

All safety practices that apply to the work area being sampled should be followed. The sampling equipment should be attached to the worker so that it will not interfere with work performance or safety.

2.1 Apparatus

    2.1.1 Adsorbent tubes

    Samples are collected with 7-cm × 4-mm i.d. × 6-mm o.d. glass sampling tubes packed with two sections of coconut shell charcoal. The tubes contain 100 mg of adsorbent in the front section and 50 mg in the back section. The adsorbent sections are held in place with glass wool plugs and are separated by urethane foam plugs. The ends of the glass sampling tubes are heat sealed. SKC Lot 2000 (SKC Inc., Fullerton, CA, Cat. No. 226-01) charcoal tubes were used for this evaluation.

    Samples are collected using personal sampling pumps that have been calibrated, with sampling devices attached, to within ±5% at the recommended flow rate of 50 mL/min. The charcoal tubes are contained in commercially available tube holders (SKC Cat. No. 222-3-1) which are connected to the pumps with flexible, non-crimpable tubing.

    2.1.2 Diffusive samplers

    Samples are collected with SKC 575-002 Passive Samplers (SKC, Inc., Fullerton, CA). The samplers contain 500 mg of Anasorb 747 adsorbent. Lot 263 samplers were used in this evaluation.

    A thermometer and barometer to determine the sampling site conditions.

2.2 Reagents

None required

2.3 Technique

    2.3.1 Charcoal tubes

    Immediately before sampling, break off the ends of the charcoal tube. All tubes should be from the same lot.

    Connect the sampling tube to the sampling pump with flexible, non-crimpable tubing. It is desirable to utilize a sampling tube holder that shields the employee from the sharp, jagged end of the sampling tube. Position the tube so that sampled air first passes through the larger adsorbent section.

    Air being sampled should not pass through any hose or tubing before entering the sampling tube.

    To avoid channeling, place the sampling tube vertically in the employee's breathing zone. Position the sampler so it does not impede work performance or safety.

    After sampling for the appropriate time, immediately remove the sampling tube and seal it with plastic caps. Wrap each sample lengthwise with a Form OSHA-21 seal.

    Submit at least one blank sampling tube with each sample set. Blanks should be handled in the same manner as samples, except no air is drawn through them.

    Record air volume (in liters), sampling time (minutes) and sampling rate (mL/min) for each sample on Form OSHA-91A.

    Also list any compounds that could be considered potential interferences, especially solvents, that are being used in the sampling area.

    Ship any bulk sample(s) in a container separate from the air samples.

    2.3.2 SKC 575-002 Samplers (In general, follow the manufacturer's instructions supplied with the samplers.)

    Remove the sampler, which is enclosed in an airtight clear bag, from the container. Keep the O-ring, press-on cover, cover retainer, port plugs and Teflon tube for later use.

    Remove the sampler from the clear bag when ready to begin sampling. CAUTION - The monitor immediately begins to sample when it is removed from this bag.

    Record the start time on the sampler label or on a Form OSHA-91A.

    Attach the sampler to the worker near his/her breathing zone with the perforations in the sampler facing out. Assure that the area directly in front of the sampler is unobstructed throughout the sampling period.

    At the end of the sampling period, immediately detach the sampler from the worker and with the O-ring in place, attach the cover onto the sampler securing it with the cover retainer. Visually inspect the O-ring to be sure it is forming a proper seal around the entire circumference of the sampler. Record the stop time on sampler label or on a Form OSHA-91A.

    Prepare a blank by removing an unused sampler from its clear bag and immediately attaching a cover, with the O-ring in place, onto it. Secure the cover with a retainer.

    Seal each sampler with an OSHA-21 form.

    Verify that the sampling times are properly recorded on the Form OSHA-91A for each sample. Also, identify blank samples on this form.

    Record the room temperature and atmospheric pressure (station pressure) of the sampling site on the Form OSHA-91A.

    List any compounds that could be considered potential interferences, especially solvents, that are being used in the sampling area.

    Submit the samplers to the laboratory for analysis as soon as possible. Include all port plugs and Teflon tubes which will be used in the laboratory analyses.

    Ship any bulk sample(s) in a container separate from the air samples.

2.4 Sampler capacity (charcoal tubes) and sampler rate/capacity (SKC 575-002 Samplers)

    2.4.1 Charcoal tubes

    The sampling capacity of the front section of charcoal sampling tubes was tested by sampling from a dynamically generated test atmosphere of tetrachloroethylene at 192.7 ppm (1306 mg/m3) and from another atmosphere containing 192.0 ppm (1031 mg/m3) of trichloroethylene. The samples were collected at nominal flow rates of 50 mL/min and the mean relative humidity of the tetrachloroethylene atmosphere was 65% at 26.1°C and the trichloroethylene atmosphere was 64.5% RH at 25.8°C. The breakthrough from the front sections of charcoal tubes did not exceed 0.1% when sampling from the tetrachloroethylene atmosphere for 9.5 hours (approximately equal to28.5 liters), indicating the capacity of charcoal tubes is about 37 mg of tetrachloroethylene. The mean 5% breakthrough volume was determined to be approximately 23.3 liters (approximately equal to7.8 hours) when sampling from the trichloroethylene atmosphere, indicating a capacity of about 24 mg of trichloroethylene before 5% breakthrough occurred. (Section 4.7.1)

    2.4.2 SKC 575-002 Samplers

    The sampling rate and capacity of SKC 575-002 Samplers were determined by taking samples from a dynamically generated test atmosphere of tetrachloroethylene (nominal concentration of 200 ppm or 1356 mg/m3) and trichloroethylene (nominal concentration of 200 ppm or 1074 mg/m3) for increasing time intervals. A sampling rate of 13.06 mL/min (at 760 mmHg, 25°C) and capacity of greater than 12 mg per sample for tetrachloroethylene and a sampling rate of 14.24 mL/min (at 760 mmHg, 25°C) and capacity of greater than 10 mg per sample for trichloroethylene were obtained from this test. (Section 4.7.2)

2.5 Extraction efficiency

It is the responsibility of each analytical laboratory to determine the extraction efficiency for each analyte because the internal standard, adsorbent material, reagents and laboratory techniques may influence the results and be different than the those listed in this evaluation.

    2.5.1 Charcoal tubes (Section 4.8.1)

    The mean respective extraction efficiencies for tetrachloroethylene and trichloroethylene from SKC Lot 2000 charcoal tubes over the range of the RQL to 2 times the target concentrations are 98.8% and 99.3% for TWA levels, 98.3% and 98.8% for ceiling levels and 98.5% and 99.0% for peak levels. The mean extraction efficiencies for the RQL and all of the eighteen other concentrations are 98.5% for tetrachloroethylene and 99.1% for trichloroethylene.

    The mean respective extraction efficiencies for tetrachloroethylene and trichloroethylene from SKC Lot 2000 charcoal tubes that had humid air drawn through them before they were spiked at the target concentrations are 98.9% and 99.5% for TWA levels, 98.6% and 98.9% for ceiling levels and 98.6% and 98.8% for peak levels.

    Extracted samples remain stable for at least 24 h.

    2.5.2 SKC 575-002 Samplers (Section 4.8.2)

    The mean respective extraction efficiencies for tetrachloroethylene and trichloroethylene from SKC 575-002 Samplers over the ranges of the RQL to 2 times the target concentrations are 96.2% and 97.7% for TWA levels and 94.9% and 97.3% for ceiling levels. The mean extraction efficiencies for the RQL and all of the twelve other concentrations are 95.4% for tetrachloroethylene and 97.5% for trichloroethylene.

    The mean respective extraction efficiencies for tetrachloroethylene and trichloroethylene from SKC 575-002 Samplers that had been exposed to humid air before they were spiked at the target concentrations are 93.2% and 95.3% for TWA levels and 94.0% and 96.0% for ceiling levels.

    Extracted samples remain stable for at least 24 h.

2.6 Recommended air volume and sampling rate for charcoal tube samples and recommended exposure time for SKC 575-002 Samplers

    2.6.1 For TWA (long-term) samples, sample up to 12 L of air at 50 mL/min (up to 240 min) when using charcoal tubes. When using SKC 575-002 Samplers, sample for as long as 240 minutes.

    2.6.2 For ceiling samples, sample greater than 0.25 L of air at 50 mL/min (greater than 5 min) when using charcoal tubes. When using SKC 575-002 Samplers, sample for greater than 5 minutes.

    2.6.3 For peak samples, sample at least 0.05 L of air at 50 mL/min (at least 1 min) when using charcoal tubes. When using SKC 575-002 Samplers, sample for 5 minutes or longer.

    2.6.4 When short-term samples are collected, the air concentrations equivalent to the reliable quantitation limits become larger. For example, the respective reliable quantitation limits for tetrachloroethylene and trichloroethylene samples collected on charcoal tubes become 0.71 ppm (4.8 mg/m3) and 0.60 ppm (3.2 mg/m3) for 5-min samples and 3.6 ppm (24 mg/m3) and 3.0 ppm (16 mg/m3) for 1-min samples.

2.7 Interferences (sampling)

    2.7.1 The presence of other contaminants in the sampled air can potentially reduce the capacity of the samplers to collect the analyte of interest. Also, the sampling rates of the SKC 575-002 Samplers could possibly be altered. Interference studies were performed by sampling for 240 minutes from a test atmosphere (70% RH, 24°C, 659 mmHg) containing 100 ppm of tetrachloroethylene with 83 ppm of methylene chloride, 70 ppm of isopropyl alcohol, 71 ppm of 2-butanone (MEK), 26 ppm of butyl acetate, 17 ppm of dioxane and 16 ppm of amyl acetate. Another test was done by sampling for 240 minutes from a test atmosphere (59% RH, 27°C, 659 mmHg) containing 100 ppm of trichloroethylene plus the aforementioned solvent vapors. The presence of these compounds, which may represent typical substances that may be collected with tetrachloroethylene or trichloroethylene, did not have a significant effect on sample results using either of the samplers. (Section 4.9.1)

    2.7.2 A reverse diffusion study for the SKC 575-002 Samplers and a stripping study for the charcoal tubes was performed by sampling an atmosphere (65% RH, 26C, 651 mmHg) containing 193 ppm of tetrachloroethylene and 194 ppm trichloroethylene for 60 minutes with six of each samplers. Three samplers from each set were additionally subjected to 180 minutes of the same atmosphere without the analytes present to determine if any of the collected solvents diffused off of the SKC 575-002 Samplers and also whether it was stripped off of the charcoal tubes. Upon analysis of the samples, the mean recovery of the removed samplers versus the mean recovery of the samplers that were additionally exposed to the atmosphere without analyte was greater than 90% for both samplers, indicating that reverse diffusion and stripping are not significant. (Section 4.9.2)

    2.7.3 Sampling from relatively dry atmospheres was investigated by sampling from an atmosphere (3% RH, 26°C, 650 mmHg) containing 198 ppm of tetrachloroethylene and 199 ppm of trichloroethylene for 240 minutes with both samplers. Sampling from dry atmospheres did not have a significant effect on the results for either of the analytes using either of the samplers. (Section 4.9.3)

    2.7.4 Sampling from atmospheres containing low concentrations of analytes was investigated by sampling from an atmosphere (67% RH, 26°C, 649 mmHg) containing 9.0 ppm of tetrachloroethylene and 10.7 ppm of trichloroethylene for 240 minutes with both samplers. Sampling these low concentrations did not have a significant effect on results using either of the samplers for either of the analytes. (Section 4.9.4)

    2.7.5 Suspected interferences should be reported to the laboratory with submitted samples.

3. Analytical Procedure

Adhere to the rules set down in your Chemical Hygiene Plan, which is mandated by the OSHA laboratory standard. Avoid skin contact and inhalation of all chemicals.

3.1 Apparatus

    3.1.1 A GC equipped with a flame ionization detector. For this evaluation, a Hewlett-Packard 6890 Series Gas Chromatograph equipped with an Automatic Liquid Sampler was used.

    3.1.2 A GC column capable of separating analyte from the extraction solvent, internal standard and any interferences. A 60-m × 0.32-mm i.d. fused silica Stabilwax capillary column (Cat. No. 10657) with a 1.0-µm df from Restek Corporation (Bellefonte, PA) was used in this evaluation.

    3.1.3 An electronic integrator or some other suitable means of measuring peak areas. A Waters Millennium 2020 Networking Computer System was used in this evaluation.

    3.1.4 Two-milliliter vials with Teflon-lined caps.

    3.1.5 A dispenser capable of delivering 1.0 mL of extraction solvent to prepare standards and samples. If a dispenser is not available, a 1.0-mL volumetric pipet may be used.

    3.1.6 A sampler rack (SKC Cat. No. 226-04-5) and a specialized shaker (SKC Cat. No. 226D-03-1) to facilitate the extraction of SKC 575-002 Samplers.

3.2 Reagents

    3.2.1 Tetrachloroethylene (CAS 127-18-4), reagent grade. Aldrich A.C.S. reagent grade (Cat. No. 44,378-6) Lot TR06115EN, was used in this evaluation.

    3.2.2 Trichloroethylene (CAS 79-01-6), reagent grade. Aldrich A.C.S. reagent grade (Cat. No. 25,140-2) Lot DR0306BQ, was used in this evaluation.

    3.2.3 Carbon disulfide (CS2) [CAS 75-15-0], chromatographic grade. Aldrich 99.9+% low benzene grade (Cat. No. 34,227-0) Lot 1054JQ, was used in this evaluation.

    3.2.4 A suitable internal standard, reagent grade. Supelco Neat EPA Standard, Lot LA59304, ethylbenzene (CAS 100-41-4) was used in this evaluation.

    3.2.5 The extraction solvent consists of CS2 containing 1.0 milliliter of internal standard per liter of solution (1µL/mL).

    3.2.6 GC grade nitrogen, air and hydrogen.

3.3 Standard preparation

    3.3.1 Prepare standards by injecting microliter amounts of analyte into vials containing 1.0 mL (for charcoal tubes) or 2.0 mL (for SKC 575-002 Samplers) of extraction solvent delivered from the same dispenser used to extract samples. For example, inject 5.00 µL of tetrachloroethylene into a vial containing 1.0 mL of extraction solvent. Assuming the density of tetrachloroethylene is 1.623 g/mL, (which is dependent on the temperature of the tetrachloroethylene) this standard contains 8115 µg of tetrachloroethylene per milliliter.

    3.3.2 Bracket sample concentrations with standard concentrations. If upon analysis, sample concentrations fall outside the range of prepared standards, prepare and analyze additional standards to ascertain the linearity of instrument response or dilute high samples with extraction solvent and reanalyze the diluted samples.

3.4 Sample preparation

    3.4.1 Charcoal tube samples

    Transfer each section of adsorbent from the sampling tubes to separate labeled vials. Discard the glass tubes, urethane foam plugs and glass wool plugs.

    Add 1.0 mL of extraction solvent to each vial using the same dispenser as used for preparation of standards.

    Immediately cap the vials.

    Allow the adsorbent sections to extract for 30 minutes. Periodically apply gentle agitation to the vials during the extraction period.

    3.4.2 SKC 575-002 Samplers (In general, follow the manufacturer's instructions supplied with the samplers.)

    Cut off the ends of the two protruding tubes of each sampler with a razor blade or sharp knife.

    Slowly add 1.0 mL of extraction solvent through one of the protruding tubes (ports). After about 30 seconds, slowly add another 1.0 mL of extraction solvent and then immediately insert plugs into the ports.

    Mount the samplers in the sampler rack (SKC Cat. No. 226-04-5) of a specialized shaker (SKC Cat. No. 226D-03-1) and shake the samplers for 1 hour.

    Transfer each extracted sample by removing the plugs from the sampler ports, firmly inserting the tapered end of a supplied Teflon tube into the outer port and carefully pouring the solution through the Teflon tube into a labeled autosampler vial.

3.5 Analysis

    3.5.1 GC conditions

    zone temperatures: column-
    injector-
    detector-
    100°C (isothermal)
    200°C
    250°C
    gas flows: hydrogen (carrier)-
    nitrogen (makeup)-
    hydrogen (flame)-
    air-
    2.5 mL/min (89 kPa head pressure)
    50 mL/min
    38 mL/min
    450 mL/min
    injection volume: 1.0 µL (with a 15:1 split)
    signal range: 0
    column: 60-m × 0.32-mm i.d. fused silica, Stabilwax 1.0-µm df
    retention times: trichloroethylene-
    tetrachloroethylene-
    ethylbenzene-
    (CS2- 2.9 min)
    4.7 min
    5.3 min
    7.6 min (internal standard)
Figure 3.5.1.1.
Figure 3.5.1.1. Chromatogram of a standard near the TWA target concentrations for charcoal tubes.
Key: (1) CS2, (2) trichloroethylene, (3) tetrachloroethylene, (4) ethylbenzene.
  Figure 3.5.1.2.
Figure 3.5.1.2. Chromatogram of a standard near the TWA target concentrations for SKC 575-002 Samplers. Key: (1) CS2, (2) trichloroethylene, (3) tetrachloroethylene, (4) ethylbenzene.

 
 
Figure 3.5.1.3.
Figure 3.5.1.3. Chromatogram of a standard near the ceiling target concentrations for charcoal tubes. Key: (1) CS2, (2) trichloroethylene, (3) tetrachloroethylene, (4) ethylbenzene.
Figure 3.5.1.4.
Figure 3.5.1.4. Chromatogram of a standard near the ceiling target concentrations for SKC 575-002 Samplers. Key: (1) CS2, (2) trichloroethylene, (3) tetrachloroethylene, (4) ethylbenzene.
Figure 3.5.1.5.

Figure 3.5.1.5. Chromatogram of a standard near the peak target concentrations charcoal tubes. Key: (1) CS2, (2) trichloroethylene, (3) tetrachloroethylene, (4) ethylbenzene.

    3.5.2 Peak areas are measured by an integrator or other suitable means.

    3.5.3 An internal standard (ISTD) calibration method is used. A calibration curve is prepared by analyzing standards and plotting micrograms of analyte per milliliter versus ISTD-corrected area counts of the analyte peaks. Sample concentrations must be bracketed by standards.

Figure 3.5.3.1.
Figure 3.5.3.1. Calibration curve for tetrachloroethylene constructed from the data in Table 4.3.1. The equation of the line is Y = 216.8X + 10648.
  Figure 3.5.3.2.
Figure 3.5.3.2. Calibration curve for trichloroethylene constructed from the data in Table 4.3.2. The equation of the line is Y = 224.8X - 827.

3.6 Interferences (analytical)

    3.6.1 Any compound that produces a response on a flame ionization detector and has the same general retention time of the analyte of interest or the internal standard is a potential interference. Possible interferences should be reported to the laboratory with submitted samples by the industrial hygienist. These interferences should be considered before samples are extracted.

    3.6.2 GC parameters (i.e. column and column temperature) may be changed to possibly circumvent interferences.

    3.6.3 The extraction efficiencies from "wet" samplers for both analytes at the various target concentrations were investigated. The extraction efficiencies were comparable to those obtained from samplers that had not been exposed to humid air. (Section 2.5)

    3.6.4 When necessary, the identity or purity of an analyte peak may be confirmed with additional analytical data, such as mass spectrometry. (Section 4.10)

3.7 Calculations

    3.7.1 Charcoal tube samples

    The analyte concentration for samples is obtained from the appropriate calibration curve in terms of micrograms per milliliter, uncorrected for extraction efficiency. The air concentration is calculated using the following formulae. The back (50-mg) section is analyzed primarily to determine if there was any breakthrough from the front section during sampling. If a significant amount of analyte is found on the back section (e.g., greater than 25% of the amount found on the front section), this fact should be reported with sample results. If any analyte is found on the back section, it is added to the amount found on the front section. This total amount is then corrected by subtracting the total amount (if any) found on the blank.

    mg/m3 = (µg of analyte per mL)(1.0 mL) / ((L of air sampled)(extraction efficiency))

    where   L of air sampled is [(sampling time, min)(sampling rate, mL/min)] / 1000
    extraction efficiency is expressed in decimal fraction

    ppm  =  (mg/m3)(24.46) / (molecular weight of analyte)

    where 24.46 is the molar volume at 25°C and 101.3 kPa (760 mmHg)
    MW of tetrachloroethylene is 165.83
    MW of trichloroethylene is 131.39

    3.7.2 SKC 575-002 Samplers

    The analyte concentration for samples is obtained from the appropriate calibration curve in terms of micrograms per milliliter, uncorrected for extraction efficiency. The air concentration is calculated using the following formulae.

    mg/m3  =  (µg of analyte per mL)(2 mL) / ((L of air sampled)(extraction efficiency))

    where L of air sampled is [(sampling time, min)(sampling rate, mL/min)]/1000
    extraction efficiency is expressed in decimal fraction
    Sampling rate for tetrachloroethylene is (13.06)(760/P)(T/298.2)1.5
    Sampling rate for trichloroethylene is (14.24)(760/P)(T/298.2)1.5
    P is the sampling site atmospheric pressure (mmHg)
    T is the sampling site temperature (K)
    If the sampling site temperature is not provided, assume that it is 22.2°C (295.4 K, 72°F). If the sampling site atmospheric pressure is not given, calculate an approximate value (P) based on the sampling site elevation (E) from the following equation.
    P  =  (3.887 × 10-7)(E2) - (2.7467 × 10-2)(E) + 760

    ppm  =  (mg/m3)(24.46) / (molecular weight of analyte)

    where 24.46 is the molar volume at 25°C and 101.3 kPa (760 mmHg)
    MW of tetrachloroethylene is 165.83
    MW of trichloroethylene is 131.39

4. Backup Data

This method was evaluated according to the OSHA SLTC "Evaluation Guidelines for Air Sampling Methods Utilizing Chromatographic Analyses"(12) in which the terms used and evaluation procedures followed for this method are fully discussed and explained.

4.1 Detection limit of the analytical procedure (DLAP)

The DLAP is measured as the mass of analyte introduced onto the chromatographic column. Ten analytical standards were prepared in equal increments with the highest standard containing 2.43 µg of tetrachloroethylene and 1.90 µg of trichloroethylene per mL. Injection of the highest standards produced peaks approximately 10 times the baseline noise of a reagent blank when a 1-µL injection with a 1:15 split was made onto the GC column. Standards, plus a reagent blank, were analyzed and the data obtained were used to determine the required parameters (A and SEE) for the calculation of the DLAP for each analyte. SEE values of 18.3 and 15.3 and A values of 3.392 and 4.089 were obtained for tetrachloroethylene and trichloroethylene respectively. The DLAP was calculated to be 16.2 pg for tetrachloroethylene and 11.2 pg for trichloroethylene.

Table 4.1.1
DLAP for Tetrachloroethylene

concentration
(µg/mL)
mass on column
(pg)
peak area
(µV • s)

0.000
0.243
0.487
0.730
0.974
1.22
1.46
1.70
1.95
2.19
2.43
0.0
16.2
32.5
48.7
64.9
81.3
97.3
113
130
146
162
0
55.9
123
178
198
276
328
364
486
571
614

  Figure 4.1.1

Figure 4.1.1 Plot of data from Table 4.1.1 to determine the DLAP. The equation of the line is Y = 3.392X + 2.79.




Table 4.1.2
DLAP for Trichloroethylene

concentration
(µg/mL)
mass on column
(pg)
peak area
(µV • s)

0.000
0.190
0.380
0.571
0.761
0.951
1.14
1.33
1.52
1.71
1.90
0.0
12.7
25.3
38.1
50.7
63.4
76.0
88.7
101
114
127
0
56.9
94.5
133
170
248
320
374
409
448
512

  Figure 4.1.2

Figure 4.1.2 Plot of data from Table 4.1.2 to determine the DLAP. The equation of the line is Y = 4.089X - 7.68.


4.2 Detection limit of the overall procedure (DLOP) and reliable quantitation limit (RQL)

The DLOP is measured as mass per sample and expressed as equivalent air concentrations, based on the recommended sampling parameters. Ten samplers of each type were spiked with equal descending increments of tetrachloroethylene and trichloroethylene such that the highest sampler loadings were 2.43 µg/sample and 1.90 µg/sample for the charcoal tubes and 4.87 µg/sample and 3.80 µg/sample for SKC 575-002 Samplers of tetrachloroethylene and trichloroethylene respectively. (The SKC 575-002 Samplers were spiked with twice the amounts of analytes compared to charcoal tubes because they are extracted with 2.0 mL of solvent versus 1.0 mL for the charcoal tubes.) These are the amounts, when spiked on the samplers, that would produce peaks approximately 10 times the baseline noise for sample blanks. These spiked samplers, plus blanks, were analyzed with the recommended analytical parameters, and the data obtained used to calculate the required parameters (A and SEE) for the calculation of the DLOPs. Respective A and SEE values of 238.1 and 27.9 for tetrachloroethylene and 264.8 and 21.5 for trichloroethylene were obtained from charcoal tube samples to give DLOPs of 0.35 µg of tetrachloroethylene per sample (4.3 ppb or 29µg/m3) and 0.24 µg of trichloroethylene sample (3.7 ppb or 20 g/m3). Respective A and SEE values of 107.2 and 23.7 for tetrachloroethylene and 118.1 and 13.1 for trichloroethylene were obtained from SKC 575-002 Samplers to give DLOPs of 0.66 µg of tetrachloroethylene per sample (31 ppb or 211 µg/m3) and 0.33 µg of trichloroethylene sample (18 ppb or 97 µg/m3).

Table 4.2.1
Detection Limit of the Overall Procedure for Tetrachloroethylene from Charcoal Tubes

mass (µg) per
sample
peak area
(µV • s)

0.0
0.243
0.487
0.730
0.974
1.22
1.46
1.70
1.95
2.19
2.43
0
26.0
109
161
221
317
319
384
402
499
611

  Figure 4.2.1

Figure 4.2.1 Plot of data from Table 4.2.1 to determine the DLAP. The equation of the line is Y = 3.392X + 2.79.




Table 4.2.2
Detection Limit of the Overall Procedure for Trichloroethylene from Charcoal Tubes

mass (µg) per
sample
peak area
(µV • s)

0.0
0.190
0.380
0.571
0.761
0.951
1.14
1.33
1.52
1.71
1.90
0
44.3
82.3
188
203
230
257
349
411
452
508

  Figure 4.2.2

Figure 4.2.2 Plot of data from Table 4.2.2 to determine the DLAP. The equation of the line is Y = 4.089X - 7.68.




Table 4.2.3
Detection Limit of the Overall Procedure for Tetrachloroethylene from SKC 575-002 Samplers

mass (µg) per
sample
peak area
(µV • s)

0.0
0.487
0.974
1.46
1.95
2.43
2.92
3.41
3.90
4.38
4.87
0
65.5
124
175
207
239
310
374
386
462
574

  Figure 4.2.3

Figure 4.2.3. Plot of data from Table 4.2.3 to determine the DLOP/RQL for tetrachloroethylene from SKC 575-002 Samplers. The equation of the line is Y = 107.2X + 4.1.




Table 4.2.4
Detection Limit of the Overall Procedure for Trichloroethylene from SKC 575-002 Samplers

mass (µg) per
sample
peak area
(µV • s)

0.0
0.380
0.761
1.14
1.52
1.90
2.28
2.66
3.04
3.42
3.80
0
44.9
124
134
194
243
264
329
359
425
450

  Figure 4.2.4

Figure 4.2.4. Plot of data from Table 4.2.4 to determine the DLOP/RQL for trichloroethylene from SKC 575-002 Samplers. The equation of the line is Y = 118.1X + 9.0.


The RQL is considered the lower limit for precise quantitative measurements. It is determined from the regression line data obtained for the calculation of the DLOP. The RQLs were calculated to be 1.17 µg per sample (14 ppb or 98 µg/m3) and 0.81 µg per sample (13 ppb or 68 µg/m3) respectively for tetrachloroethylene and trichloroethylene from charcoal tubes and 2.21 µg per sample (104 ppb or 705 µg/m3) and 1.11 µg per sample (61 ppb or 325 µg/m3) respectively for tetrachloroethylene and trichloroethylene from SKC 575-002 Samplers. The recoveries at these levels are 99.5%, 98.8%, 97.2% and 97.6% respectively.

Figure 4.2.5
Figure 4.2.5. Chromatogram near the RQLs.
Key: (1) CS2, (2) trichloroethylene,
(3) tetrachloroethylene, (4) ethylbenzene.

4.3 Instrument calibration

The instrument was calibrated using standards over concentration ranges equivalent to 0.5 times the ceiling target concentrations for SKC 575-002 Samplers to 2 times the TWA target concentrations for charcoal tube samples. Calibration curves were constructed as shown in Section 3.5.3 from six replicate injections of five standards equally spaced over these ranges. The coefficient of determination (r2) is 0.99996 for tetrachloroethylene and 0.99993 for trichloroethylene. The coefficient of non-determination (k2) is 4×10-5 for tetrachloroethylene and 7×10-5 for trichloroethylene.

Table 4.3.1
Tetrachloroethylene Standards used for Instrument Calibration

standard concn
(µg/mL)
  response in peak areas (µV • s) × 10-3

21.1
4058
8115
12172
16230
  4.6945
904.56
1775.2
2649.1
3529.2
4.6979
903.81
1770.0
2645.5
3531.8
4.7172
904.63
1771.2
2644.1
3524.8
4.6936
903.90
1772.3
2650.8
3525.0
4.6995
902.18
1769.5
2643.3
3527.8
4.6970
904.97
1772.2
2642.9
3521.7




Table 4.3.2
Trichloroethylene Standards used for Instrument Calibration

standard concn
(µg/mL)
  response in peak areas (µV • s) × 10-3

19.0
3219
6437
9656
12874
  4.3280
715.66
1461.8
2177.6
2891.2
4.3264
714.21
1451.5
2175.1
2898.6
4.3216
711.91
1455.5
2174.7
2887.3
4.2951
711.87
1457.2
2184.8
2880.9
4.2926
709.61
1456.7
2170.6
2879.5
4.2970
713.48
1455.2
2169.3
2878.0

4.4 Precision (overall procedure)

The precision of the overall procedure is determined from the storage data in Section 4.5. The SEEs and precisions of the overall procedure are given in the following tables for charcoal tubes and SKC 575-002 Samplers. All values are based on the ambient storage tests in Section 4.5.

Table 4.4.1
SEEs and Precisions of the Overall Procedure at the 95% Confidence Interval for Charcoal Tube Samples

TWA samples
ceiling samples
peak samples
analyte SEE precision SEE precision SEE precision

tetrachloroethylene
trichloroethylene
5.06%
5.06%
±9.9%
±9.9%
5.05%
5.04%
±9.9%
±9.9%
5.21%
5.09%
±10.2%
±10.0%




Table 4.4.2
SEEs and Precisions of the Overall Procedure at the 95% Confidence Interval for SKC 575-002 Samplers when Sampling Site Temperature (T) or Atmospheric Pressure (P) are Known or Unknown

TWA samples
ceiling samples
analyte SEE precision SEE precision

tetrachloroethylene
  both T&P known
  only T known
  only P known
  neither T nor P known

trichloroethylene
  both T&P known
  only T known
  only P known
  neither T nor P known
 
9.03%
9.51%
11.9%
12.2%


9.13%
9.61%
11.9%
12.3%
 
±17.7%
±18.6%
±23.3%
±23.9%


±17.9%
±18.8%
±23.3%
±24.1%
 
9.63%
10.1%
12.3%
12.7%


9.14%
9.62%
12.0%
12.3%
 
±18.9%
±19.8%
±24.1%
±24.9%


±17.9%
±18.9%
±23.5%
±24.1%

4.5 Storage tests

Charcoal tube storage samples were prepared by sampling at 50 mL/min from controlled test atmospheres that were at approximately 80% RH and 22.2°C and at an atmospheric pressure of 655 mmHg. Storage samples for SKC 575-002 Samplers were prepared by exposing them to the same atmospheres as were generated for the charcoal tube samples. The flow of the atmospheres through the exposure chamber provided for face velocities of approximately 0.4 m/s on the SKC 575-002 Samplers. TWA samples were prepared by sampling from 100-ppm atmospheres for 240 minutes, ceiling samples from 200-ppm atmospheres for 5 minutes and peak samples from 300-ppm atmospheres for 1 minute. Six samples for each set were analyzed immediately after generation, fifteen were stored in a refrigerator at 0°C and fifteen were stored in a closed drawer at ambient temperatures of 20-25°C. At approximately three-day intervals, three samples were selected from each of the storage sets and analyzed.

Table 4.5.1
TWA Storage Tests for Tetrachloroethylene (Charcoal Tubes)

time
(days)
refrigerated storage
recovery (%)
  ambient storage
recovery (%)

0
0
6
10
13
15
17
98.7
99.6
101.8
99.4
99.3
99.6
99.0
98.2
99.9
99.4
99.8
100.5
99.6
100.7
100.5
99.6
100.4
100.9
100.2
99.3
99.8
  100.6
102.8
101.7
99.4
99.3
99.6
100.0
102.0
100.4
98.7
99.3
100.0
99.0
99.3
99.9
101.3
100.0
100.8
99.8
99.3
98.7





Figure 4.5.1.1

Figure 4.5.1.1. Tetrachloroethylene refrigerated storage test on charcoal tubes, 240-minute samples at 100 ppm.
Figure 4.5.1.2

Figure 4.5.1.2. Tetrachloroethylene ambient storage test on charcoal tubes, 240-minute samples at 100 ppm.




Table 4.5.2
TWA Storage Tests for Trichloroethylene (Charcoal Tubes)

time
(days)
refrigerated storage
recovery (%)
  ambient storage
recovery (%)

0
0
6
10
13
15
17
99.7
99.6
101.6
98.8
98.8
99.2
98.4
98.6
100.2
98.9
99.4
99.8
99.3
100.4
100.7
100.2
100.0
100.4
99.8
98.9
99.3
  99.7
99.6
100.9
98.8
98.2
99.0
99.2
98.6
100.2
97.8
98.6
99.2
98.5
98.4
100.7
100.2
99.3
100.5
99.0
98.7
98.0





Figure 4.5.2.1

Figure 4.5.2.1. Trichloroethylene refrigerated storage test on charcoal tubes, 240-minute samples at 100 ppm.
Figure 4.5.2.2

Figure 4.5.2.2. Trichloroethylene ambient storage test on charcoal tubes, 240-minute samples at 100 ppm.




Table 4.5.3
Ceiling Storage Tests for Tetrachloroethylene (Charcoal Tubes)

time
(days)
refrigerated storage
recovery (%)
  ambient storage
recovery (%)

0
0
6
10
13
15
17
101.8
100.9
101.8
100.9
101.1
102.4
99.8
100.0
101.1
100.7
102.3
99.4
100.5
100.1
101.2
99.7
95.1
99.3
99.1
102.2
103.1
  101.8
100.9
101.0
99.8
100.6
101.0
100.9
100.0
101.1
100.1
100.1
99.5
99.4
100.6
101.2
99.7
101.7
100.2
100.0
98.8
99.9





Figure 4.5.3.1

Figure 4.5.3.1. Tetrachloroethylene refrigerated storage test on charcoal tubes, 5-minute samples at 200 ppm.
Figure 4.5.3.2

Figure 4.5.3.2. Tetrachloroethylene ambient storage test on charcoal tubes, 5-minute samples at 200 ppm.




Table 4.5.4
Ceiling Storage Tests for Trichloroethylene (Charcoal Tubes)

time
(days)
refrigerated storage
recovery (%)
  ambient storage
recovery (%)

0
0
6
10
13
15
17
99.3
100.0
100.3
99.9
99.4
101.2
98.5
99.8
99.7
99.4
101.2
98.2
100.5
99.0
100.0
98.5
99.4
97.9
97.4
100.8
101.6
  99.3
100.0
99.6
98.4
99.5
99.8
99.4
99.8
99.7
99.2
98.9
98.3
98.8
99.0
100.0
98.5
100.4
100.0
98.4
98.4
98.3





Figure 4.5.4.1

Figure 4.5.4.1. Trichloroethylene refrigerated storage test on charcoal tubes, 5-minute samples at 200 ppm.
Figure 4.5.4.2

Figure 4.5.4.2. Trichloroethylene ambient storage test on charcoal tubes, 5-minute samples at 200 ppm.




Table 4.5.5
Peak Storage Tests for Tetrachloroethylene
(Charcoal Tubes)

time
(days)
refrigerated storage
recovery (%)
  ambient storage
recovery (%)

0
0
6
10
13
15
17
99.6
96.5
97.6
95.9
95.3
100.1
98.5
98.6
96.9
96.0
97.0
97.8
99.2
97.5
96.9
97.1
96.8
97.7
96.3
99.1
95.7
  99.6
96.5
97.4
97.2
95.2
99.1
97.5
98.6
96.9
96.4
97.6
95.4
100.0
98.7
96.9
97.1
96.7
97.3
94.3
96.2
97.4





Figure 4.5.5.1

Figure 4.5.5.1. Tetrachloroethylene refrigerated storage test on charcoal tubes, 1-minute samples at 300 ppm.
Figure 4.5.5.2

Figure 4.5.5.2. Tetrachloroethylene ambient storage test on charcoal tubes, 1-minute samples at 300 ppm.




Table 4.5.6
Peak Storage Tests for Trichloroethylene
(Charcoal Tubes)

time
(days)
refrigerated storage
recovery (%)
  ambient storage
recovery (%)

0
0
6
10
13
15
17
97.5
94.7
98.0
95.7
96.8
98.3
96.6
97.5
95.2
96.8
97.4
99.0
96.9
95.5
96.6
95.5
98.9
98.1
96.2
97.1
96.2
  97.5
94.7
97.6
97.7
96.3
97.2
96.6
97.5
95.2
96.3
98.0
96.9
97.5
96.1
96.6
95.5
97.2
97.4
95.9
97.3
95.5





Figure 4.5.6.1

Figure 4.5.6.1. Trichloroethylene refrigerated storage test on charcoal tubes, 1-minute samples at 300 ppm.
Figure 4.5.6.2

Figure 4.5.6.2. Trichloroethylene ambient storage test on charcoal tubes, 1-minute samples at 300 ppm.




Table 4.5.7
TWA Storage Tests for Tetrachloroethylene
(SKC 575-002 Samplers)

time
(days)
refrigerated storage
recovery (%)
  ambient storage
recovery (%)

0
0
6
10
13
15
17
102.3
99.1
96.6
102.2
99.3
99.3
98.2
100.7
99.4
100.2
96.1
98.6
95.2
101.5
100.3
96.6
97.4
95.3
98.4
98.6
102.1
  102.3
99.1
99.9
94.6
99.6
100.2
97.4
100.7
99.4
99.8
96.9
98.1
103.6
99.2
100.3
96.6
94.1
97.0
100.3
96.2
100.0





Figure 4.5.7.1

Figure 4.5.7.1. Tetrachloroethylene refrigerated storage test SKC 575-002, 240-minute samples at 100 ppm.
Figure 4.5.7.2

Figure 4.5.7.2. Tetrachloroethylene ambient storage test on SKC 575-002, 240-minute samples at 100 ppm.




Table 4.5.8
TWA Storage Tests for Trichloroethylene
(SKC 575-002 Samplers)

time
(days)
refrigerated storage
recovery (%)
  ambient storage
recovery (%)

0
0
6
10
13
15
17
103.0
99.1
94.7
100.7
97.4
97.7
97.0
100.2
98.8
98.6
94.2
96.9
94.0
100.1
99.2
95.1
95.0
93.3
96.4
97.2
100.6
  103.0
99.1
97.4
92.9
98.2
98.6
96.8
100.2
98.8
97.0
95.2
96.4
102.2
97.9
99.2
95.1
92.0
95.2
98.0
93.8
98.3





Figure 4.5.8.1

Figure 4.5.8.1. Trichloroethylene refrigerated storage test SKC 575-002, 240-minute samples at 100 ppm.
Figure 4.5.8.2

Figure 4.5.8.2. Trichloroethylene ambient storage test on SKC 575-002, 240-minute samples at 100 ppm.




Table 4.5.9
Ceiling Storage Tests for Tetrachloroethylene
(SKC 575-002 Samplers)

time
(days)
refrigerated storage
recovery (%)
  ambient storage
recovery (%)

0
0
6
10
13
15
17
102.1
100.2
97.5
94.1
93.4
98.2
105.0
114.1
113.9
99.1
95.3
95.2
99.4
103.8
113.0
103.5
105.2
100.9
105.7
105.8
102.8
  102.1
100.2
98.0
98.8
97.5
99.4
99.1
114.1
113.9
100.4
104.0
96.6
100.8
97.1
113.0
103.5
108.2
100.2
98.7
98.3
99.4





Figure 4.5.9.1

Figure 4.5.9.1. Tetrachloroethylene refrigerated storage test SKC 575-002 Samplers, 5-minute at 200 ppm.
Figure 4.5.9.2

Figure 4.5.9.2. Tetrachloroethylene ambient storage test on SKC 575-002 Samplers, 5-minute at 200 ppm.




Table 4.5.10
Ceiling Storage Tests for Trichloroethylene
(SKC 575-002 Samplers)

time
(days)
refrigerated storage
recovery (%)
  ambient storage
recovery (%)

0
0
6
10
13
15
17
102.5
102.0
99.9
94.5
97.5
99.6
106.8
106.1
105.6
102.1
98.0
100.1
101.4
104.4
102.0
97.6
96.3
99.1
104.1
105.1
105.4
  102.5
102.0
100.8
95.3
98.2
99.2
98.9
106.1
105.6
103.7
101.0
96.7
100.4
99.4
102.0
97.6
96.5
94.7
94.8
97.8
99.3





Figure 4.5.10.1

Figure 4.5.10.1. Trichloroethylene refrigerated storage test SKC 575-002 Samplers, 5-minute at 200 ppm.
Figure 4.5.10.2

Figure 4.5.10.2. Trichloroethylene ambient storage test on SKC 575-002 Samplers, 5-minute at 200 ppm.


4.6 Reproducibility

Six samples for both samplers at the TWA and ceiling target concentrations and six charcoal tube samples at the peak target concentrations that were collected from controlled test atmospheres, along with a draft copy of this procedure, were submitted to an SLTC service branch for analysis. All of samples were stored at room temperature and analyzed 22 days after generation. No sample result had a deviation greater than the precisions of the overall procedure determined in Section 4.4.

Table 4.6.1
Reproducibility Data for Tetrachloroethylene
Charcoal Tube Samples at the TWA Level

sample
no.
mg/m3
reported
mg/m3
expected
percent
found
deviation
(%)

1
2
3
4
5
6
652.7
647.6
649.2
647.2
634.6
649.4
668.5
668.5
668.5
668.5
668.5
668.5
97.6
96.9
97.1
96.8
94.9
97.1
-2.4
-3.1
-2.9
-3.2
-4.1
-2.9

Table 4.6.2
Reproducibility Data for Tetrachloroethylene
Charcoal Tube Samples at the Ceiling Level

sample
no.
mg/m3
reported
mg/m3
expected
percent
found
deviation
(%)

1
2
3
4
5
6
1365
1314
1324
1326
1323
1313
1337
1337
1337
1337
1337
1337
102.1
98.3
99.0
99.2
99.0
98.2
+2.1
-1.7
-1.0
-0.8
-1.0
-1.8

 


Table 4.6.3
Reproducibility Data for Tetrachloroethylene
Charcoal Tube Samples at the Peak Level

sample
no.
mg/m3
reported
mg/m3
expected
percent
found
deviation
(%)

1
2
3
4
5
6
1961
1970
1957
1969
1974
1976
2005
2005
2005
2005
2005
2005
97.8
98.3
97.6
98.2
98.5
98.6
-2.2
-1.7
-2.4
-1.8
-1.5
-1.4

Table 4.6.4
Reproducibility Data for Trichloroethylene
Charcoal Tube Samples at the TWA Level

sample
no.
mg/m3
reported
mg/m3
expected
percent
found
deviation
(%)

1
2
3
4
5
6
459.1
457.4
457.4
455.4
445.5
456.5
479.2
479.2
479.2
479.2
479.2
479.2
95.8
95.5
95.5
95.0
93.0
95.3
-4.2
-4.5
-4.5
-5.0
-7.0
-4.7

 


Table 4.6.5
Reproducibility Data for Trichloroethylene
Charcoal Tube Samples at the Ceiling Level

sample
no.
mg/m3
reported
mg/m3
expected
percent
found
deviation
(%)

1
2
3
4
5
6
1001
954.2
965.2
964.2
958.4
951.2
990.0
990.0
990.0
990.0
990.0
990.0
101.1
96.4
97.5
97.4
96.8
96.1
+1.1
-3.6
+-2.5
-2.6
-3.2
-3.9

Table 4.6.6
Reproducibility Data for Trichloroethylene
Charcoal Tube Samples at the Peak Level

sample
no.
mg/m3
reported
mg/m3
expected
percent
found
deviation
(%)

1
2
3
4
5
6
1385
1380
1388
1397
1404
1404
1463
1463
1463
1463
1463
1463
94.7
94.3
94.9
95.5
96.0
96.0
-5.3
-5.7
-5.1
-4.5
-4.0
-4.0

 


Table 4.6.7
Reproducibility Data for Tetrachloroethylene
SKC 575-002 Samples at the TWA Level

sample
no.
mg/m3
reported
mg/m3
expected
percent
found
deviation
(%)

1
2
3
4
5
6
663.2
638.4
646.4
644.4
635.1
649.8
668.5
668.5
668.5
668.5
668.5
668.5
99.2
95.5
96.7
96.4
95.0
97.2
-0.8
-4.5
-3.3
-3.6
-5.0
-2.8

Table 4.6.8
Reproducibility Data for Tetrachloroethylene
SKC 575-002 Samples at the Ceiling Level

sample
no.
mg/m3
reported
mg/m3
expected
percent
found
deviation
(%)

1
2
3
4
5
6
1415
1326
1382
1499
1320
1360
1337
1337
1337
1337
1337
1337
105.8
99.2
103.4
112.1
98.7
101.7
+5.8
-0.8
+3.4
+12.1
-1.3
+1.7

 


Table 4.6.9
Reproducibility Data for Trichloroethylene
SKC 575-002 Samples at the TWA Level

sample
no.
mg/m3
reported
mg/m3
expected
percent
found
deviation
(%)

1
2
3
4
5
6
471.5
450.4
455.7
456.2
446.6
458.1
479.2
479.2
479.2
479.2
479.2
479.2
98.4
94.0
95.1
95.2
93.2
95.6
-2.6
-6.0
-4.9
-4.8
-6.8
-4.4

Table 4.6.10
Reproducibility Data for Trichloroethylene
SKC 575-002 Samples at the Ceiling Level

sample
no.
mg/m3
reported
mg/m3
expected
percent
found
deviation
(%)

1
2
3
4
5
6
1027
969.2
1006
1062
979.1
982.1
990.0
990.0
990.0
990.0
990.0
990.0
103.7
97.9
101.6
107.3
98.9
99.2
+3.7
-2.1
+1.6
+7.3
-1.1
-0.8



4.7 Sampler capacity and sampling rate

    4.7.1 Charcoal tubes

    The sampling capacity of the front section of charcoal sampling tubes was tested by sampling from a dynamically generated test atmosphere of tetrachloroethylene at 192.7 ppm (1306 mg/m3) and from another atmosphere containing 192.0 ppm (1031 mg/m3) of trichloroethylene). The samples were collected at a nominal flow rate of 50 mL/min and the mean relative humidity of the tetrachloroethylene atmosphere was 65% at 26.1°C and the trichloroethylene atmosphere was 64.5% RH at 25.8°C. Complete charcoal tubes were placed in-line behind the front test sections and changed at measured intervals to check for breakthrough. No significant breakthrough (<0.1%) occurred when sampling from the tetrachloroethylene atmosphere for 9.5 hours (approximately equal to28.5 liters) which is equivalent to about 37 mg of tetrachloroethylene. The mean 5% breakthrough volume from the trichloroethylene atmosphere was determined to be about 23.3 liters (approximately equal to7.8 hours), which is equivalent to approximately 24 mg.

Table 4.7.1
Breakthrough of Trichloroethylene from
Charcoal Sampling Tubes when Sampling
from a 1031 mg/m3 Atmosphere

sampler
tube
no.
air
vol
(L)
sampling
time
(min)
downstream
concn
(mg/m3)
break-
through
(%)

1






2






3
6.1
13.7
16.7
19.7
22.4
24.3

5.8
13.1
16.0
18.9
21.4
23.2

6.3
14.2
17.4
20.5
23.3
25.3
0-240
240-300
300-360
360-420
420-465
465-495

0-240
240-300
300-360
360-420
420-465
465-495

0-240
240-300
300-360
360-420
420-465
465-495
0
0
0.06
5.19
44.6
119.7

0
0
0
0.56
5.86
21.4

0
0.02
0.30
8.03
50.3
119.6
0
0
0.01
0.50
4.33
11.6

0
0
0
0.05
0.57
2.08

0
0.002
0.03
0.78
4.88
11.6

Figure 4.7.1

    Figure 4.7.1. Determination of the 5%
    breakthrough volume for trichloroethylene
    collected with charcoal tubes.


    4.7.2 SKC 575-002 Samplers

    The sampling rates and capacities were determined by taking samples from a dynamically generated test atmosphere of tetrachloroethylene (nominal concentration of 200 ppm or 1356 mg/m3) and trichloroethylene (nominal concentration of 200 ppm or 1074 mg/m3) for increasing time intervals. The atmospheres were at approximately 70% relative humidity, 25°C and 660 mmHg. The flow of the atmospheres through the exposure chamber provided for face velocities of approximately 0.4 m/s on the SKC 575-002 Samplers. Three samples were taken for each sampling interval. Sampler capacity is exceeded when the sampling rate decreases rapidly. Because this did not occur for the time period tested, the capacity is determined to be greater than 12 mg for tetrachloroethylene and greater than 10 mg for trichloroethylene. Sampling rates standardized to 760 mmHg and 25°C of 13.06 mL/min for tetrachloroethylene and 14.24 mL/min for trichloroethylene were determined from the mean sample rates from samples collected from 5.0 to 600 minutes.


Table 4.7.2.1
Sampling Rate and Capacity
for Tetrachloroethylene using SKC 575-002 Samplers

sampling time
(min)
sampling rate
(mL/min)
RSD
(%)

5
10
30
60
120
180
240
360
480
600
12.61
12.99
12.63
12.73
13.13
12.77
13.55
13.23
13.43
13.50
1.29
1.01
1.02
1.71
3.04
3.06
2.53
1.38
1.16
0.94

Figure 4.7.2.1

Figure 4.7.2.1 Determination of sampling rate and capacity for tetrachloroethylene using SKC 575-002 Samplers.
 



Table 4.7.2.2
Sampling Rate and Capacity
for Trichloroethylene using SKC 575-002 Samplers

sampling time
(min)
sampling rate
(mL/min)
RSD
(%)

5
10
30
60
120
180
240
360
480
600
14.18
14.10
13.99
13.87
14.10
13.73
14.75
14.41
14.60
14.65
3.97
1.11
4.20
1.48
2.71
3.15
2.66
1.60
0.54
0.81

Figure 4.7.2.2

Figure 4.7.2.2 Determination of sampling rate and capacity for trichloroethylene using SKC 575-002 Samplers.


4.8 Extraction efficiency

    4.8.1 Charcoal tubes

    The extraction efficiencies of tetrachloroethylene and trichloroethylene were determined by liquid-spiking 100-mg portions of charcoal with the analytes at the RQL and 0.05 to 2 times the target concentrations for each of three levels. These samples were stored overnight at ambient temperature and then extracted and analyzed. The mean extraction efficiency over the ranges of the RQL to 2 times the TWA, ceiling and peak target concentration levels are 98.8%, 98.3% and 98.5% respectively for tetrachloroethylene and 99.3%, 98.8% and 99.0% for trichloroethylene. The mean extraction efficiencies for the RQL and all of the eighteen other concentrations are 98.5% for tetrachloroethylene and 99.1% for trichloroethylene.

    Extraction efficiencies from wet charcoal were also determined by drawing the recommended volume (12, 0.25 and 0.05 liters for TWA, ceiling and peak levels respectively) of humid air at approximately 80% RH, 22.2°C before the charcoal was spiked with the analytes. The mean respective extraction efficiencies for tetrachloroethylene and trichloroethylene are 98.9% and 99.5% for TWA levels, 98.6% and 98.9% for ceiling levels and 98.6% and 98.8% for peak levels.


Table 4.8.1.1
Extraction Efficiency of Tetrachloroethylene from Charcoal Tubes (TWA Levels)

level
sample number
× target
concn
µg per
sample
1 2 3 4 5 6 mean

(approximately equal toRQL)
0.05
0.1
0.2
0.5
1.0
2.0

1.0 (wet)
1.22
405.8
811.5
1623
4058
8115
16230

8115
97.5
97.6
98.4
98.5
98.5
98.8
99.4

98.6
100.1
97.9
99.0
98.5
98.5
98.4
99.4

98.8
97.8
97.8
98.1
98.8
98.6
98.9
99.4

98.8
100.0
98.0
98.6
98.9
98.5
98.7
99.7

98.8
101.5
97.7
98.3
98.6
98.6
98.8
99.3

98.8
96.8
98.2
98.5
98.7
98.5
98.7
99.5

99.1
99.5
97.9
98.5
98.7
98.5
98.7
99.4

98.9




Table 4.8.1.2
Extraction Efficiency of Tetrachloroethylene from Charcoal Tubes (Ceiling Levels)

level
sample number
× target
concn
µg per
sample
1 2 3 4 5 6 mean

(approximately equal toRQL)
0.05
0.1
0.2
0.5
1.0
2.0

1.0 (wet)
1.22
16.88
33.76
67.52
168.8
337.6
675.2
337.6
97.5
97.7
98.9
97.6
97.9
98.2
98.8

98.7
100.1
98.0
97.7
98.5
96.7
98.6
98.9

98.4
97.8
97.9
99.0
98.2
97.9
98.5
98.6

98.9
100.0
98.7
98.6
97.9
96.7
98.7
99.3

98.7
101.5
96.1
99.0
94.6
97.1
98.3
99.1

98.3
96.8
95.6
98.8
98.4
96.4
98.5
98.8

98.5
99.5
97.3
98.7
98.2
97.1
98.5
98.9

98.6




Table 4.8.1.3
Extraction Efficiency of Tetrachloroethylene from Charcoal Tubes (Peak Levels)

level
sample number
× target
concn
µg per
sample
1 2 3 4 5 6 mean

(approximately equal toRQL)
0.05
0.1
0.2
0.5
1.0
2.0

1.0 (wet)
1.22
5.03
10.06
20.13
50.30
100.6
201.3

100.6
97.5
98.1
99.4
97.2
97.6
99.0
98.5

98.8
100.1
101.8
96.3
97.5
98.0
98.9
99.0

98.8
101.3
98.0
95.8
99.3
97.4
98.8
99.2

98.4
100.0
97.3
96.0
98.1
98.9
98.7
99.1

98.4
101.5
99.6
97.6
97.9
98.3
96.6
99.2

98.6
96.8
100.2
98.3
98.8
98.2
98.4
98.6

98.9
99.5
99.2
97.2
98.1
98.4
98.4
98.9

98.6




Table 4.8.1.4
Extraction Efficiency of Trichloroethylene from Charcoal Tubes (TWA Levels)

level
sample number
× target
concn
µg per
sample
1 2 3 4 5 6 mean

(approximately equal toRQL)
0.05
0.1
0.2
0.5
1.0
2.0

1.0 (wet)
0.96
321.9
643.7
1287
3219
6437
12870

6437
94.9
98.6
100.0
99.4
98.2
98.6
100.0

99.2
97.5
98.9
100.3
99.2
98.9
99.4
100.0

99.4
101.5
98.8
99.2
99.6
98.8
99.5
100.2

99.7
94.6
98.8
99.8
100.2
98.3
99.3
100.0

99.8
102.4
98.7
99.8
99.7
99.4
99.2
99.8

99.4
101.6
99.2
99.5
99.8
98.2
99.1
99.9

99.6
98.8
98.8
99.8
99.6
98.6
99.2
100.0

99.5




Table 4.8.1.5
Extraction Efficiency of Trichloroethylene from Charcoal Tubes (Ceiling Levels)

level
sample number
× target
concn
µg per
sample
1 2 3 4 5 6 mean

(approximately equal toRQL)
0.05
0.1
0.2
0.5
1.0
2.0

1.0 (wet)
0.96
13.39
26.78
53.56
133.9
267.8
535.6
267.8
94.9
96.7
98.7
98.2
98.7
99.2
99.4

99.4
97.5
98.3
99.5
99.7
98.0
99.3
99.6

98.8
101.5
98.2
99.8
99.5
99.2
99.4
99.4

98.9
94.6
98.6
98.4
98.3
97.6
99.4
99.9

99.0
102.4
99.5
98.3
98.9
97.9
99.1
99.9

98.6
98.6
96.7
98.7
98.7
97.2
99.7
99.4

98.9
98.8
98.0
98.9
98.9
98.1
99.4
99.6

98.9




Table 4.8.1.6
Extraction Efficiency of Trichloroethylene from Charcoal Tubes (Peak Levels)

level
sample number
× target
concn
µg per
sample
1 2 3 4 5 6 mean

(approximately equal toRQL)
0.05
0.1
0.2
0.5
1.0
2.0

1.0 (wet)
0.96
3.99
7.98
15.96
39.91
79.82
159.6

79.82
94.9
103.2
98.2
100.0
99.2
100.3
99.3

99.8
97.5
100.3
97.5
99.6
98.3
99.7
99.8

98.8
101.5
93.7
100.8
100.0
98.9
99.6
99.9

98.9
94.6
98.3
100.7
98.4
98.0
99.2
99.7

98.6
102.4
101.1
98.8
98.6
97.4
97.7
100.2

98.4
98.6
96.4
99.1
98.1
98.5
99.4
99.2

98.6
98.8
98.8
99.2
99.1
98.4
99.3
99.7

98.8




    Stability of extracted charcoal tubes samples

    The stability of extracted samples was investigated by reanalyzing the target concentration samples 24 h after the initial analyses. After the original analyses for each target concentration level were performed, three vials from each set were recapped with new septa, while the remaining three retained their punctured septa. The samples were reanalyzed with fresh standards. The mean percent change was 0.0%, -0.5% and -0.2% for TWA, ceiling and peak tetrachloroethylene samples respectively that were resealed with new septa and -0.7%, -1.6% and -1.4% for those that retained their punctured septa. The mean percent change was +0.7%, -0.5% and -0.5% for TWA, ceiling and peak trichloroethylene samples respectively that were resealed with new septa and -3.6%, -5.6% and -5.4% for those that retained their punctured septa.

    Table 4.8.1.7
    Stability of Extracted Tetrachloroethylene Samples (TWA)

    punctured septa replaced
      punctured septa retained
    initial
    DE
    (%)
    DE after
    one day
    (%)
    difference initial
    DE
    (%)
    DE after
    one day
    (%)
    difference

    98.8
    98.4
    98.9
     
    98.7
    98.6
    98.8
    98.8
    (means)
    98.7
    -0.2
    +0.4
    -0.1
     
    0.0
    98.7
    98.8
    98.7
     
    98.7
    97.7
    98.1
    98.2
    (means)
    98.0
    -1.0
    -0.7
    -0.5
     
    -0.7




    Table 4.8.1.8
    Stability of Extracted Tetrachloroethylene Samples (Ceiling)

    punctured septa replaced
      punctured septa retained
    initial
    DE
    (%)
    DE after
    one day
    (%)
    difference initial
    DE
    (%)
    DE after
    one day
    (%)
    difference

    98.2
    98.6
    98.5
     
    98.4
    97.9
    98.0
    98.0
    (means)
    98.0
    -0.3
    -0.6
    -0.5
     
    -0.5
    98.7
    98.3
    98.5
     
    98.5
    96.8
    97.0
    96.8
    (means)
    96.9
    -1.9
    -1.3
    -1.7
     
    -1.6




    Table 4.8.1.9
    Stability of Extracted Tetrachloroethylene Samples (Peak)

    punctured septa replaced
      punctured septa retained
    initial
    DE
    (%)
    DE after
    one day
    (%)
    difference initial
    DE
    (%)
    DE after
    one day
    (%)
    difference

    99.0
    98.9
    98.8
     
    98.9
    98.8
    98.8
    98.4
    (means)
    98.7
    -0.2
    -0.1
    -0.4
     
    -0.2
    98.7
    96.6
    98.4
     
    97.9
    97.1
    95.0
    97.3
    (means)
    96.5
    -1.6
    -1.6
    -1.1
     
    -1.4




    Table 4.8.1.10
    Stability of Extracted Trichloroethylene Samples (TWA)

    punctured septa replaced
      punctured septa retained
    initial
    DE
    (%)
    DE after
    one day
    (%)
    difference initial
    DE
    (%)
    DE after
    one day
    (%)
    difference

    98.6
    99.4
    99.5
     
    99.2
    98.8
    100.6
    100.2
    (means)
    99.9
    +0.2
    +1.2
    +0.7
     
    +0.7
    99.3
    99.2
    99.1
     
    99.2
    94.8
    95.6
    96.4
    (means)
    95.6
    -4.5
    -3.6
    -2.7
     
    -3.6




    Table 4.8.1.11
    Stability of Extracted Trichloroethylene Samples (Ceiling)

    punctured septa replaced
      punctured septa retained
    initial
    DE
    (%)
    DE after
    one day
    (%)
    difference initial
    DE
    (%)
    DE after
    one day
    (%)
    difference

    99.2
    99.3
    99.4
     
    99.3
    98.7
    98.9
    98.8
    (means)
    98.8
    -0.5
    -0.4
    -0.6
     
    -0.5
    99.4
    99.1
    99.7
     
    99.4
    93.6
    94.5
    93.4
    (means)
    93.8
    -5.8
    -4.6
    -6.3
     
    -5.6




    Table 4.8.1.12
    Stability of Extracted Trichloroethylene Samples (Peak)

    punctured septa replaced
      punctured septa retained
    initial
    DE
    (%)
    DE after
    one day
    (%)
    difference   initial
    DE
    (%)
    DE after
    one day
    (%)
    difference

    100.3
    99.7
    99.6
     
    99.9
    99.4
    99.4
    99.4
    (means)
    99.4
    -0.9
    -0.3
    -0.2
     
    -0.5
      99.2
    97.7
    99.4
     
    98.8
    94.4
    91.8
    93.9
    (means)
    93.4
    -4.8
    -5.9
    -5.5
     
    -5.4




    4.8.2 SKC 575-002 Samplers

    The extraction efficiencies of tetrachloroethylene and trichloroethylene were determined by liquid-spiking 500-mg portions of Anasorb 747 with the analytes at 0.05 to 2 times the target concentrations. These samples were stored overnight at ambient temperature and then extracted and analyzed. The mean extraction efficiency over the ranges of the RQL to 2 times the TWA and ceiling levels are 96.2% and 94.9% respectively for tetrachloroethylene and 97.7% and 97.3% for trichloroethylene. The mean extraction efficiencies for the RQL and all of the twelve other concentrations are 95.4% for tetrachloroethylene and 97.5% for trichloroethylene.

    Extraction efficiencies from wet Anasorb 747 were also determined by exposing the samplers for the recommended times (240 and 5 minutes for TWA and ceiling levels respectively) to humid air at approximately 80% RH, 22.2°C before the samplers were spiked with the analytes. The mean respective extraction efficiencies for tetrachloroethylene and trichloroethylene are 93.2% and 95.3% for TWA levels and 94.0% and 96.0% for ceiling levels.

    Table 4.8.2.1
    Extraction Efficiency of Tetrachloroethylene from SKC 575-002 Samplers (TWA Levels)

    level
    sample number
     
    × target
    concn
    µg per
    sample
    1 2 3 4 5 6 mean

    (approximately equal toRQL)
    0.05
    0.1
    0.2
    0.5
    1.0
    2.0

    1.0 (wet)
    2.43
    107.1
    214.2
    428.5
    1071
    2142
    4285

    2142
    96.0
    96.5
    96.1
    95.3
    96.9
    95.8
    96.7

    93.3
    96.6
    95.9
    95.6
    95.2
    96.8
    96.4
    96.2

    92.9
    97.5
    95.7
    95.1
    94.8
    96.7
    96.0
    96.6

    93.1
    101.4
    95.6
    95.4
    94.6
    96.1
    96.4
    96.4

    92.9
    97.7
    96.4
    95.8
    95.0
    96.5
    96.5
    96.6

    93.1
    94.1
    96.0
    95.6
    95.0
    96.7
    96.0
    96.7

    93.8
    97.2
    96.0
    95.6
    95.0
    96.6
    96.2
    96.5

    93.2




    Table 4.8.2.2
    Extraction Efficiency of Tetrachloroethylene from SKC 575-002 Samplers (Ceiling Levels)

    level
    sample number
     
    × target
    concn
    µg per
    sample
    1 2 3 4 5 6 mean

    (approximately equal toRQL)
    0.05
    0.1
    0.2
    0.5
    1.0
    2.0

    1.0 (wet)
    2.43
    4.38
    8.76
    17.5
    43.8
    87.6
    175.2

    87.6
    96.0
    94.5
    96.4
    96.3
    90.0
    94.7
    92.5

    94.3
    96.6
    95.6
    93.3
    95.3
    91.1
    94.6
    91.5

    94.4
    97.5
    98.3
    94.0
    98.0
    91.5
    95.2
    90.6

    93.9
    101.4
    95.0
    98.2
    96.4
    93.2
    94.9
    91.0

    94.2
    97.7
    97.1
    92.8
    98.0
    91.8
    93.8
    92.6

    93.5
    94.1
    98.0
    96.6
    97.1
    93.3
    95.1
    92.4

    93.6
    97.2
    96.4
    95.2
    96.8
    91.8
    94.7
    91.8

    94.0




    Table 4.8.2.3
    Extraction Efficiency of Trichloroethylene from SKC 575-002 Samplers (TWA Levels)

    level
    sample number
     
    × target
    concn
    µg per
    sample
    1 2 3 4 5 6 mean

    (approximately equal toRQL)
    0.05
    0.1
    0.2
    0.5
    1.0
    2.0

    1.0 (wet)
    1.23
    96.56
    193.1
    386.2
    965.6
    1931
    3862

    1931
    101.2
    98.1
    97.9
    97.6
    98.7
    97.6
    98.0
    95.4
    94.1
    97.7
    97.4
    97.4
    98.1
    98.2
    97.0

    94.7
    98.6
    97.7
    96.7
    97.3
    98.2
    97.4
    98.4

    95.2
    99.2
    97.1
    97.1
    96.8
    97.6
    98.1
    97.9

    95.1
    94.5
    98.1
    97.7
    97.3
    98.0
    98.1
    98.2

    95.4
    98.0
    98.0
    97.8
    97.2
    98.3
    97.4
    98.5

    95.8
    97.6
    97.9
    97.4
    97.3
    98.2
    97.8
    98.0

    95.3




    Table 4.8.2.4
    Extraction Efficiency of Trichloroethylene from SKC 575-002 Samplers (Ceiling Levels)

    level
    sample number
     
    × target
    concn
    µg per
    sample
    1 2 3 4 5 6 mean

    (approximately equal toRQL)
    0.05
    0.1
    0.2
    0.5
    1.0
    2.0

    1.0 (wet)
    1.23
    3.95
    7.90
    15.80
    39.50
    79.00
    158.0

    79.00
    101.2
    95.8
    99.6
    96.7
    93.5
    97.1
    94.8

    96.1
    94.1
    102.7
    99.8
    97.1
    92.8
    96.4
    94.3

    97.4
    98.6
    99.0
    97.1
    96.7
    95.5
    96.4
    93.5

    97.7
    99.2
    102.9
    102.4
    97.8
    96.7
    96.3
    94.0

    96.0
    94.5
    102.1
    99.6
    98.5
    94.9
    95.8
    95.0

    95.7
    98.0
    98.9
    101.0
    98.9
    96.4
    96.8
    94.7

    95.2
    97.6
    100.2
    99.9
    97.6
    95.0
    96.5
    94.4

    96.0




    Stability of extracted SKC 575-002 Samplers

    The stability of extracted samples was investigated by reanalyzing the target concentration samples 24 h after the initial analyses. After the original analyses for each target concentration level were performed, three vials from each set were recapped with new septa, while the remaining three retained their punctured septa. The samples were reanalyzed with fresh standards. The mean percent change was -0.4% and +1.0% for TWA and ceiling tetrachloroethylene samples respectively that were resealed with new septa and -1.0% and +0.3% for those that retained their punctured septa. The mean percent change was -0.4% and +0.9% for TWA and ceiling trichloroethylene samples respectively that were resealed with new septa and 4.0% and 1.6% for those that retained their punctured septa.

    Table 4.8.2.5
    Stability of Extracted Tetrachloroethylene Samples (TWA)

    punctured septa replaced
    punctured septa retained
    initial
    EE
    (%)
    EE after
    one day
    (%)
    difference initial
    EE
    (%)
    EE after
    one day
    (%)
    difference

    95.8
    96.4
    96.0
     
    96.1
    95.1
    96.2
    95.8
    (means)
    95.7
    -0.7
    -0.2
    -0.2
     
    -0.4
    96.4
    96.5
    96.0
     
    96.3
    95.0
    95.5
    95.4
    (means)
    95.3
    -1.4
    -1.0
    -0.6
     
    -1.0




    Table 4.8.2.6
    Stability of Extracted Tetrachloroethylene Samples (Ceiling)

    punctured septa replaced
    punctured septa retained
    initial
    EE
    (%)
    EE after
    one day
    (%)
    difference initial
    EE
    (%)
    EE after
    one day
    (%)
    difference

    94.7
    94.6
    95.2
     
    94.8
    95.6
    95.8
    96.0
    (means)
    95.8
    +0.9
    +1.2
    +0.8
     
    +1.0
    94.9
    93.8
    95.1
     
    94.6
    95.1
    95.0
    94.7
    (means)
    94.9
    +0.2
    +1.2
    -0.4
     
    +0.3




    Table 4.8.2.7
    Stability of Extracted Trichloroethylene Samples (TWA)

    punctured septa replaced
    punctured septa retained
    initial
    EE
    (%)
    EE after
    one day
    (%)
    difference initial
    EE
    (%)
    EE after
    one day
    (%)
    difference

    97.6
    98.2
    97.4
     
    97.7
    97.0
    97.9
    97.1
    (means)
    97.3
    -0.6
    -0.3
    -0.3
     
    -0.4
    98.1
    98.1
    97.4
     
    97.9
    92.2
    94.9
    94.7
    (means)
    93.9
    -5.9
    -3.2
    -2.7
     
    -4.0




    Table 4.8.2.8
    Stability of Extracted Trichloroethylene Samples (Ceiling)

    punctured septa replaced
    punctured septa retained
    initial
    EE
    (%)
    EE after
    one day
    (%)
    difference initial
    EE
    (%)
    EE after
    one day
    (%)
    difference

    97.1
    96.4
    96.4
     
    96.6
    97.4
    97.4
    97.7
    (means)
    97.5
    +0.3
    +1.0
    +1.3
     
    +0.9
    96.3
    95.8
    96.8
     
    96.3
    94.5
    94.8
    94.7
    (means)
    94.7
    -1.8
    -1.0
    -2.1
     
    -1.6




4.9 Interferences (sampling)

    4.9.1 Interference studies were performed by sampling for 240 minutes from a test atmosphere (70% RH, 24°C, 659 mmHg) containing 100 ppm of tetrachloroethylene with 83 ppm of methylene chloride, 70 ppm of isopropyl alcohol, 71 ppm of 2-butanone (MEK), 26 ppm of butyl acetate, 17 ppm of dioxane and 16 ppm of amyl acetate. Another test was done by sampling for 240 minutes from a test atmosphere (63% RH, 27°C, 659 mmHg) containing 100 ppm of trichloroethylene plus the aforementioned solvent vapors. The presence of these compounds, which may represent typical substances that may be collected with tetrachloroethylene or trichloroethylene did not have a significant effect on sample results using either of the samplers.

    Table 4.9.1
    Recovery (%) from the Atmosphere Described in 4.9.1 for each Analyte

      tetrachloroethylene trichloroethylene

    sample
    no.
    charcoal
    tubes
    SKC 575-002
    Samplers
    charcoal
    tubes
    SKC 575-002
    Samplers

    1
    2
    3
     
    mean
    101.6
    100.0
    102.0
     
    101.2
    104.9
    108.4
    105.4
     
    106.2
    96.8
    95.8
    97.7
     
    96.8
    98.4
    100.0
    100.6
     
    99.7



    4.9.2 Reverse diffusion studies for the SKC 575-002 Samplers and stripping studies for the charcoal tubes was performed by sampling a 193-ppm atmosphere of tetrachloroethylene and a 194-ppm atmosphere of trichloroethylene (65% RH, 26°C, 651 mmHg) for 60 minutes with six of each samplers. Three samplers from each set were additionally subjected to 180 minutes of the same atmosphere without the analytes present to determine if any of the collected solvents diffused off of the SKC 575-002 Samplers and also whether it was stripped off of the charcoal tubes. Upon analysis of the samples, the mean recovery of the removed samplers versus the mean recovery of the samplers that were additionally exposed to the atmosphere without analyte was greater than 90% for both samplers, indicating that reverse diffusion and stripping is insignificant. The first three samples in Table 4.9.2 for each sampler were used to sample the atmosphere for 60 minutes, while the last three were additionally exposed to a blank atmosphere for another 180 minutes.

    Table 4.9.2
    Recovery (%) for the Circumstances Described in 4.9.2 for Each Analyte
    (Samples 1-3 used to sample atmosphere for 60 minutes,
    samples 4-6 for an additional 180 minutes of blank atmosphere)

      tetrachloroethylene trichloroethylene

    sample
    no.
    charcoal
    tubes
    SKC 575-002
    Samplers
    charcoal
    tubes
    SKC 575-002
    Samplers

    1
    2
    3
    mean (1-3)
     
    4
    5
    6
    mean (4-6)
     
    mean (4-6) ÷ mean (1-3)
    100.2
    99.0
    100.6
    99.9
     
    101.7
    101.8
    101.7
    101.7
     
    1.02
    104.6
    100.1
    103.2
    102.6
     
    97.3
    106.8
    102.7
    102.3
     
    1.00
    100.7
    99.1
    101.1
    100.3
     
    101.4
    101.8
    101.4
    101.5
     
    1.01
    104.8
    100.2
    102.3
    102.4
     
    96.7
    106.3
    102.6
    101.9
     
    1.00



    4.9.3 The effects from sampling from relatively dry atmospheres were investigated by sampling from an atmosphere (3% RH, 26°C, 650 mmHg) containing 198 ppm of tetrachloroethylene and 199 ppm of trichloroethylene for 240 minutes with both samplers. The mean results for the both analytes collected on both samplers was greater than 90% of the theoretical amounts.

    Table 4.9.3
    Recovery (%) for 240-minute Samples from a
    Dry Atmosphere for Each Analyte

      tetrachloroethylene trichloroethylene

    sample
    no.
    charcoal
    tubes
    SKC 575-002
    Samplers
    charcoal
    tubes
    SKC 575-002
    Samplers

    1
    2
    3
     
    mean
    100.0
    99.3
    100.0
     
    99.8
    100.4
    106.8
    102.4
     
    103.2
    98.4
    98.1
    97.9
     
    98.1
    98.5
    105.3
    101.0
     
    101.6



    4.9.4 The effects from sampling from atmospheres containing low concentrations of analytes were investigated by sampling from an atmosphere (67% RH, 26°C, 649 mmHg) containing 9.0 ppm of tetrachloroethylene and 10.7 ppm of trichloroethylene for 240 minutes with both samplers. The mean results for both analytes collected on both samplers was greater than 90% of the theoretical amounts.

    Table 4.9.4
    Recovery (%) for 240-minute Samples from a
    Atmosphere Containing 9.0 ppm of Tetrachloroethylene
    and 10.7 ppm of Trichloroethylene

      tetrachloroethylene trichloroethylene

    sample
    no.
    charcoal
    tubes
    SKC 575-002
    Samplers
    charcoal
    tubes
    SKC 575-002
    Samplers

    1
    2
    3
     
    mean
    100.2
    100.2
    100.8
     
    100.4
    101.7
    101.5
    100.2
     
    101.1
    98.5
    99.3
    99.5
     
    99.1
    100.2
    100.6
    98.2
     
    99.7



4.10 Qualitative analysis

Tetrachloroethylene and trichloroethylene can easily be identified by GC/mass spectrometry. The following figures represent typical mass spectra of tetrachloroethylene and trichloroethylene.

Figure 4.10.1
Figure 4.10.1 representative mass spectrum of tetrachloroethylene.
Figure 4.10.2
Figure 4.10.1 representative mass spectrum of trihloroethylene.





1. Elskamp, C.J. "OSHA Method No. 111, Toluene"; OSHA Salt Lake Technical Center, Salt Lake City, UT. Unpublished work, 1998.

2. Kirk-Othmer Encyclopedia of Chemical Technology, 4th ed.; Kroschwitz, J., Ed.; John Wiley & Sons: New York, 1993, Vol. 6; pp 50-59.

3. Kirk-Othmer Encyclopedia of Chemical Technology, pp 40-49.

4. Kirk-Othmer Encyclopedia of Chemical Technology, pp 50-59.

5. Kirk-Othmer Encyclopedia of Chemical Technology, pp 40-49.

6. Kirk-Othmer Encyclopedia of Chemical Technology, pp 50-59.

7. The Merck Index, 12th ed.; Budavari, S., Ed.; Merck & Co., Whitehouse Station, NJ, 1996; 9332.

8. The Merck Index, 12th ed.; Budavari, S., Ed.; Merck & Co., Whitehouse Station, NJ, 1996; 9332.

9. Kirk-Othmer Encyclopedia of Chemical Technology, pp 40-49.

10. The Merck Index, 12th ed.; Budavari, S., Ed.; Merck & Co., Whitehouse Station, NJ, 1996; 9769.

11. "Evaluation Guidelines for Air Sampling Methods Utilizing Chromatographic Analyses", OSHA Salt Lake Technical Center, Salt Lake City, UT. Unpublished work, 1999.

12. "Evaluation Guidelines for Air Sampling Methods Utilizing Chromatographic Analyses", OSHA Salt Lake Technical Center, Salt Lake City, UT. Unpublished work, 1999.



 

US Department of LaborContact InformationDisclaimer
OSHA Home PageOSHA IndexOSHA Search